Positive Geometric Vascular Remodeling Is Seen After Catheter-Based Radiation Followed by Conventional Stent Implantation but Not After Radioactive Stent Implantation
Jump to

Abstract
Background—Recent reports demonstrate that intracoronary radiation affects not only neointimal formation but also vascular remodeling. Radioactive stents and catheter-based techniques deliver radiation in different ways, suggesting that different patterns of remodeling after each technique may be expected.
Methods and Results—We analyzed remodeling in 18 patients after conventional stent implantation, 16 patients after low-activity radioactive stent implantation, 16 patients after higher activity radioactive stent implantation, and, finally, 17 patients who underwent catheter-based radiation followed by conventional stent implantation. Intravascular ultrasound with 3D reconstruction was used after stent implantation and at the 6-month follow-up to assess remodeling within the stent margins and at its edges. Preprocedural characteristics were similar between groups. In-stent neointimal hyperplasia (NIH) was inhibited by high-activity radioactive stent implantation (NIH 9.0 mm3) and by catheter-based radiation followed by conventional stent implantation (NIH 6.9 mm3) compared with low-activity radioactive stent implantation (NIH 21.2 mm3) and conventional stent implantation (NIH 20.8 mm3) (P=0.008). No difference in plaque or total vessel volume was seen behind the stent in the conventional, low-activity, or high-activity stent implantation groups. However, significant increases in plaque behind the stent (15%) and in total vessel volume (8%) were seen in the group that underwent catheter-based radiation followed by conventional stent implantation. All 4 groups demonstrated significant late lumen loss at the stent edges; however, edge restenosis was seen only in the group subjected to high-activity stent implantation and appeared to be due to an increase in plaque and, to a lesser degree, to negative remodeling.
Conclusions—Distinct differences in the patterns of remodeling exist between conventional, radioactive, and catheter-based radiotherapy with stenting.
In our enthusiasm to control vessel recoil and remodeling after balloon angioplasty (BA), stent implantation has become increasingly popular. With conventional stenting, we have eliminated recoil and remodeling as components of the restenotic process. However, this has been at the cost of exacerbating neointimal proliferation secondary to chronic vessel wall irritation, leading to in-stent restenosis.1 2
Intracoronary radiation has been developed in an attempt to decrease restenosis after BA and stent implantation. Two parallel technologies, one using radioactive stents3 4 5 6 7 and the other using catheter-based radiation,8 9 10 have been the subject of both animal and human studies. Given the different dose rates and total doses delivered by each method, one may intuitively expect different patterns of remodeling subsequent to each approach.
Whereas the effect of catheter-based radiation after BA on vascular remodeling has been described,11 the response of the arterial wall to catheter-based radiation and subsequent stent implantation has not been described. Preliminary studies have reported the effect at the stent edge after radioactive stent implantation.4 However, these reports did not encompass the response behind the stent in the arterial wall.
The aim of the present study was to describe the response of the coronary artery to radiation and stenting by examining the stent and its edges after radioactive stent implantation and also after catheter-based radiation with stent implantation.
Methods
Patient Selection
We analyzed geometric vascular remodeling in 4 groups of patients: (1) LA group, those who had undergone implantation of 32P-emitting radioactive stents at activity levels of 0.75 to 1.5 μCi (Isostent Inc); (2) HA group, those who had undergone implantation of 32P radioactive stents at activity levels of 6.0 to 12 μCi (Isostent Inc); (3) CBS group, those who had undergone conventional stent implantation after suboptimal BA (clinically significant dissection or residual stenosis >30%) and catheter-based radiation; and (4) C group, those who had undergone conventional stent implantation after suboptimal BA .
Stents analyzed were from patients with single–native-vessel coronary artery disease, normal left ventricular function, and objective evidence of ischemia. All groups were matched for patient baseline characteristics, vessel size, lesion, and stent length. Stents placed in the ostial position or adjacent to major side branches, such that the stent edges were unable to be analyzed, were excluded from analysis. Only patients who had completed 6-month angiographic and intravascular ultrasound (IVUS) follow-up were included.
Implantation Technique
The same group of cardiologists, using a similar technique, implanted all stents. Predilation of the lesion was performed, followed by stent implantation with use of either a premounted stent or the Johnson & Johnson delivery system (Johnson & Johnson Interventional Systems Co). A balloon shorter than the stent was then selected, and high-pressure balloon inflation was performed within the stent to ensure good stent apposition. Intravascular ultrasound was used to ensure optimal stent deployment.
Medication
Patients received 250 mg aspirin and 10 000 IU heparin at the initiation of the procedure, and the activated clotting time was maintained at >300 seconds. All patients received aspirin (80 mg daily) indefinitely and ticlopidine (250 mg BID) for 2 weeks (C group) or clopidogrel (75 mg daily) for 12 weeks (LA, HA, and CBS groups) after stent implantation.
Radioactive Stents
The BX stent (Isostent Inc) was the only radioactive stent implanted in this trial. It was 15 mm in length and available in diameters of 3.0 and 3.5 mm. The BX stent was made radioactive by 32P. The initial activity of the stents was measured; thereafter, it was calculated at the date on which the activity had decreased to 0.75 to 1.5 μCi or 6 to 12 μCi, levels suitable for implantation.
Catheter-Based Radiation Delivery System
The Beta-Cath System (Novoste Corp) was used to deliver localized β-radiation (90Sr/90Y) to a depth of 2 mm from the center of the source at the site of coronary intervention. The device consisted of 3 components: (1) the transfer device that stored the radiation source train and allowed the positioning of these sources within the catheter; (2) the delivery catheter, which was a 5F multilumen over-the-wire noncentered catheter that used saline solution to send and return the radiation source train; and (3) the radiation source train, which consisted of a series of 12 independent cylindrical seeds that contained the radioisotope 90Sr sources and was bordered by 2 gold radiopaque markers separated by 30 mm. Other device and procedural details have been previously published by this group.11
Definitions
Stent Edges
Stent edges were defined as those volumes axially 5 mm proximal and distal to the final stent strut. An edge restenosis was defined as an angiographic restenosis >50% at 6-month follow-up located at either stent edge. An edge effect was defined as any stent-edge renarrowing.
Patients with balloon-injured edges that failed to receive radiation in the catheter-based radiation group were excluded. In other words, no stents implanted in areas of geographical miss were included in the present study.
IVUS Image Acquisition Analysis
After the final balloon inflation and administration of intracoronary nitrates, ECG-gated IVUS pullback was performed. This was repeated at the 6-month follow-up.
The segment subjected to 3D reconstruction was examined with a mechanical IVUS system (ClearView, CVIS) with a sheath-based IVUS catheter incorporating a 30-MHz single-element transducer rotating at 1800 rpm. The IVUS transducer was withdrawn through the stationary imaging sheath by an ECG-triggered pullback device with a stepping motor.12 IVUS images coinciding with the peak of the R wave, which eliminates the artifacts caused by the movement of the heart during the cardiac cycle, were acquired. After each image acquisition, the transducer was withdrawn 0.2 mm to acquire the next image coincident with the R wave. The ECG-gated image acquisition and digitization was performed by a workstation designed for the 3D reconstruction of echocardiographic images12 (EchoScan, Tomtec). A Microsoft Windows–based contour detection program, developed at the Thoraxcenter, Rotterdam, was used for the automated 3D analysis of up to 200 IVUS images.13 The feasibility, reproducibility, and interobserver and intraobserver variability of this system have been previously validated in clinical protocols.11
Quantitative IVUS Analysis
At the stent edges, the area encompassed by the lumen-intima and media-adventitia boundaries defined the luminal volume (LV) and the total vessel volume (TVV), respectively. The difference between LV and TVV defined the plaque volume. TVV, stent volume, neointimal hyperplasia (NIH), plaque behind the stent (TVV−stent volume), and LV were obtained within the axial boundaries of the stent.
The assessment of TVV in stented patients has previously been reported.14 Although in the previous report the delineation of TVV was not possible in some patients because of stent shadowing, in the present study the delineation of the TVV boundary was possible in all stented patients. When the TVV boundary was not visible in a single cross-sectional view, the computer extrapolated it from the contours of the previous and subsequent cross sections. In addition, the use of 3D reconstruction with multiple longitudinal views facilitates the visualization of vessel structures outside the stent.
Statistical Analysis
Quantitative data are presented as mean±SD. Volumetric data derived from the 3D reconstruction of the IVUS imaging were compared immediately after treatment and at follow-up by the 2-tailed paired Student t test. Comparison between groups was performed by 1-way ANOVA. A value of P<0.05 was considered statistically significant.
The Medical Ethical Committee of the University Hospital Rotterdam approved the study, and all patients provided written informed consent before the procedure.
Results
Baseline Characteristics
Eighteen patients were enrolled in the conventional group (C group), 16 patients were enrolled in both the 0.75- to 1.5-μCi and the 6.0- to 12-μCi radioactive stent groups (LA and HA groups, respectively), and 17 patients were enrolled in the group subjected to catheter-based radiation plus a stent (CBS group). In the C group, 10 ACS Multi-Link (Guidant Corp) and 8 NIR (new intravascular rigid-flex stent; Boston Scientific/Scimed) stents were implanted, and in the CBS group, 8 NIR and 9 ACS Multi-Link stents were implanted. Baseline characteristics are similar between all groups and are described in Table 1⇓. Lesion and procedural characteristics are described in Table 2⇓. No statistically significant differences were seen between groups in the parameters described in Table 2⇓. Comparisons of volumetric data measured at the stent edges and within the margins of the stent are presented in Tables 3⇓ and 4⇓.
Clinical Characteristics
Procedural Characteristics
Volumes for Edge Proximal and Distal to Stent (10-mm Length)
Volumes for Stent
In-Stent Inhibition of NIH
Intrastent NIH was decreased after high-activity radioactive stent implantation and catheter-based radiation followed by conventional stent implantation (P=0.008). Lower activity radioactive stents had an effect similar to that of conventional stent implantation (see Table 4⇑ and Figure 1⇓).
Remodeling within margins of stent. PBS indicates plaque behind the stent. *P<0.05 for values immediately after treatment vs follow-up; †P<0.05 for values between groups.
Behind Stent
The C, LA, and HA groups demonstrated an absence of remodeling behind the stent, with no significant changes in TVV or plaque volumes. This is in contrast to the CBS group, which demonstrated a significant increase in plaque (immediately after treatment versus follow-up, 15%; P=0.002) and an increase in TVV (after treatment versus follow-up, 8%; P=0.003). Intergroup comparison showed that this change was significant (Table 4⇑, P=0.01). Further comparisons of changes within and between groups are demonstrated in Figure 1⇑. No chronic recoil of the stent was seen in any group.
Stent Edge
No significant difference between groups was seen at baseline (after stent implantation). All groups demonstrated late lumen loss at the stent edges. At the stent edges, remodeling is similar in the C and LA groups. In these groups, there is evidence of a decrease in TVV, with little change in plaque as a cause of late lumen loss (Figure 2⇓). In the HA group, a target segment restenosis (angiographically >50%) was observed in 7 patients at the stent edges. This was more common at the proximal edge (in 6 of 7 patients). The major mechanism of such a restenosis appears to be due to an increase in plaque at the stent edge. In nonrestenotic patients, the edge effect appears to be due to a decrease in TVV and, to a lesser degree, an increase in plaque (Figure 3⇓).
Changes in volumes at stent edge. PV indicates plaque volume. *P<0.05 for values immediately after treatment vs follow-up; †P<0.05 for values between groups; and ‡P=NS for ΔLV (all groups) and ΔPV (C, LA, and HA groups).
High-activity (6.0- to 12.0-μCi) stents. Restenotic edges are compared with nonrestenotic edges. Δ indicates change in volume (immediately after treatment vs follow-up). *P<0.05 for values immediately after treatment vs follow-up. Note greater lumen loss seen in restenotic group. This loss was caused by increase in plaque (P<0.05) and a less profound decrease in TVV (P<0.05). ΔTVV is similar in both groups.
In the CBS group, the edge effect is largely due to an increase in plaque, with no negative remodeling seen (P=0.045 for plaque increase in CBS versus LA, HA, and C groups). No patient with edge restenosis after catheter-based radiation was seen in our series of patients.
Stent Activity and Dose Prescribed
Mean stent activity at implantation (LA group) was 1.1±0.3 μCi. Mean stent activity at implantation (HA group) was 8.6±1.6 μCi. For the CBS group, the mean dose prescribed was 16.7±2.0 Gy.
Discussion
The development of NIH within the stent witnessed at the 6-month follow-up is well appreciated15 ; however, the changes that occur at the stent edges or indeed behind the stent struts have not been the focus of attention until recently.4 The present study is the first to describe the difference in vascular remodeling seen after radioactive stent implantation and catheter-based radiation plus stenting with the use of modern conventional stents as a benchmark. The key findings are as follows: (1) The degree of inhibition of NIH was similar in the HA and CBS groups. (2) There was no significant remodeling behind the stent after conventional or radioactive stent implantation; however, the CBS group demonstrated an increase in plaque behind the stent and in TVV. (3) At the stent edge, 3 patterns of remodeling are seen at the 6-month follow-up: first, a shrinkage in TVV and LV was noted in the C and LA groups. These 2 subgroups were not associated with stent edge restenosis in this series. After high-activity stent implantation, a pattern similar to conventional and low-activity stent implantation is seen in those edges that remain nonrestenotic; however, in the restenotic edges, plaque increase is the major contributor to lumen loss. In the CBS group, a lumen loss similar to that found in the other groups is seen; however, this occurs secondary to a relatively greater increase in plaque, without loss in TVV.
Neointimal Hyperplasia
In the present study, neointimal formation was inhibited after higher dose radioactive stent implantation and after catheter-based radiation plus stenting. The present study is in contrasts to the recent study by Carter et al,16 who used 32P stents in the porcine model, but is in keeping with earlier studies of Hehrlein et al,6 who used the rabbit model, and recent reports by Albiero et al,4 who noted a dose-dependent inhibition of NIH.
Mechanism of Remodeling Behind the Stent
Catheter-Based Radiation
After conventional and radioactive stent implantation, little positive or negative remodeling is witnessed behind the stent. In stark contrast to this is the increase in plaque behind the stent and TVV seen after catheter-based radiation and stenting. Part of the key to understanding this process may be acquired from understanding the healing process after BA. Wilcox and colleagues17 18 describe the presence of early proliferation of myofibroblasts expressing contractile proteins in the adventitia surrounding the porcine coronary artery after BA. Tracing studies have indicated that the same cells migrate and form part of the neointima. Wilcox and colleagues hypothesize that the adventitial myofibroblasts constrict the artery at the angioplasty site in much the same way as myofibroblasts participate in scar retraction in dermal healing. The source of these myofibroblasts may be distant to the immediate site of injury, including pericardial, adipose, and intramyocardial layers.19
Radiation treatment of porcine coronary arteries after BA upregulates p21 synthesis in adventitial cells, especially myofibroblasts. Such induction is dose dependent and is sustained for at least 7 days after radiation. Additionally, radiation inhibits the expression of growth factors, reduces the proliferation of adventitial myofibroblasts, and decreases the production of α-actin by the adventitial myofibroblasts, preventing the formation of the myofibroblast scar around the angioplasty site and negative vascular remodeling.17 20 Data from Fareh and et al 21 suggest that inhibition of migration but not of cellular proliferation may occur at lower doses of radiation. Therefore, cells may remain in situ, unable to migrate but able to grow in the presence of a weakened external elastic membrane. After 1 week, the effect of the radiation diminishes, and cellular proliferation, possibly as a reaction to the presence of the stent, continues behind the stent in the context of positive vascular remodeling. In our cohort of patients, no cases of stent malapposition were seen at follow-up, although our group has described this as a risk of ongoing positive vascular remodeling.22 23 A further concept to be explored is that relating to the sharp drop-off in radiation seen with the β-radiation source, which may cause underdosing deep in the adventitia and geographical miss24 in a radial sense rather than the more commonly described longitudinal sense.
Radioactive Stent
The objective of using the radioactive stent is not to neutralize myofibroblasts in the adventitia; it is the prevention of the migration and invasion of myofibroblasts from the adventitia through the stent struts and into the lumen. As is seen in the HA group, this is accomplished by the continuous and low dose rate provided by the radioactive stent. Because of the range of the “radioactive fence” created, adventitial cells remain intact without upregulation of growth factors and inhibition of contractile proteins. Consequently, no remodeling is seen behind the radioactive stent at either activity level.
Edge Remodeling
Hoffmann et al15 have previously described negative remodeling at the stent edge after conventional stent implantation. In the present study, we have been able to precisely describe the decrease in TVV as the dominant contributor to nonrestenotic lumen loss at the stent edge. Recent reports on radioactive stents suggest that the edge effect and edge restenosis may be due to an increase in plaque at the edge and to a component of negative remodeling as one moves axially from the stent.4 The contributing factors to radioactive stent edge restenosis have been discussed in detail recently by Serruys and Kay.25
It may be argued that stent-edge restenosis was not seen in the CBS group because no individuals with geographical miss were evaluated. However, our objective in the present study was to analyze the vascular response to appropriately applied catheter-based radiation, which necessitates the exclusion of all those in whom injury was not covered by radiation. Recent reports have suggested that the combination of suboptimal low-dose radiation and injury may make individuals with geographical miss vulnerable to edge restenosis.26
Study Limitations
This was a retrospective nonrandomized study of individuals who had completed 6 months of follow-up and in whom IVUS examination was possible. Individuals who had a total occlusion or in whom the IVUS catheter could not be passed under acceptable clinical circumstances were not included.
No edge restenosis was seen in the CBS group, unlike the HA group; however, both the CBS and the HA groups reflected the larger parent populations from which they were selected in all other features.
The dosimetry (catheter-based) described in the present study relates to prescribed doses only and does not necessarily reflect the dose delivered 2 mm from the source in the adventitia. Description of dosimetry is beyond the scope of the present study; however, previous work by the authors (Sabaté et al27 ), who used a similar radiation source and study population, suggests that delivered dose, residual plaque burden, and tissue composition play a fundamental role on the volumetric outcome at 6 months of follow-up after catheter-based β-radiation therapy and BA.
Conclusions
Distinct differences in the patterns of remodeling exist between conventional, radioactive, and catheter-based radiotherapy with stenting. Users of radiation need to be alerted to edge restenosis seen after higher activity radioactive stent implantation and positive remodeling behind the stent seen after catheter-based radiation and stenting. Radiation, whether it be catheter or stent-based, has forced the interventional community to look closely not only at effective inhibition of intimal proliferation but also at the adverse response of the artery to the combination of injury and radiation.
Acknowledgments
Dr Kay was supported by the National Heart Foundation of New Zealand and the Ian Gordon Mackenzie Scholarship, University of Otago, Dunedin, New Zealand. The Wenckebach award was made to Dr Serruys by the Dutch Heart Foundation and is used for brachytherapy research in the catheterization laboratory. We would like to thank Dr Josiah N. Wilcox of Emory University, Atlanta, Ga, for his valuable comments and criticism.
- Received December 30, 1999.
- Revision received April 12, 2000.
- Accepted April 14, 2000.
- Copyright © 2000 by American Heart Association
References
- ↵
Farb A, Sangiorgi G, Carter AJ, et al. Pathology of acute and chronic coronary stenting in humans. Circulation. 1999;99:44–52.
- ↵
Murphy JG, Schwartz RS, Edwards WD, et al. Percutaneous polymeric stents in porcine coronary arteries: initial experience with polyethylene terephthalate stents. Circulation. 1992 86;1596–1604.
- ↵
Wardeh AJ, Kay IP, Sabaté M, et al. β-Particle-emitting radioactive stent implantation: a safety and feasibility study. Circulation. 1999;100:1684–1689.
- ↵
Albiero R, Adamian M, Corvaja N, et al. Radioactive 32P β-particle emitting BX stent implantation in patients with coronary artery disease: a serial IVUS analysis of the ‘candy wrapper’ pattern of restenosis. Eur Heart J. 1999;20:408. Abstract,
- ↵
Laird JR, Carter AJ, Kufs WM, et al. Inhibition of neointimal proliferation with low-dose irradiation from a β-particle–emitting stent. Circulation. 1996;93:529–536.
- ↵
Hehrlein C, Stintz M, Kinscherf R, et al. Pure β-particle–emitting stents inhibit neointima formation in rabbits. Circulation. 1996;93:641–645.
- ↵
Carter AJ, Laird JR, Bailey LR, et al. Effects of endovascular radiation from a β-particle–emitting stent in a porcine coronary restenosis model: a dose-response study. Circulation. 1996;94:2364–2368.
- ↵
King SB III, Williams DO, Chogule P, et al. Endovascular β-radiation to reduce restenosis after coronary balloon angioplasty: results of the Beta Energy Restenosis Trial (BERT). Circulation. 1998;97:2025–2030.
- ↵
Condado JA, Waksman R, Gurdiel O, et al. Long-term angiographic and clinical outcome after percutaneous transluminal coronary angioplasty and intracoronary radiation therapy in humans. Circulation. 1997;96:727–732.
- ↵
- ↵
Sabaté M, Serruys PW, van der Giessen WJ, et al. Geometric vascular remodeling after balloon angioplasty and β-radiation therapy: a three-dimensional intravascular ultrasound study. Circulation. 1999;100:1182–1188.
- ↵
- ↵
- ↵
Prati F, Di Mario C, Moussa I, et al. In-stent neointimal proliferation correlates with the amount of residual plaque burden outside the stent: an intravascular ultrasound study. Circulation. 1999;99:1011–1014.
- ↵
Hoffmann R, Mintz GS, Dussaillant GR, et al. Patterns and mechanism of in-stent restenosis: a serial intravascular ultrasound study. Circulation. 1996;94:1247–1254.
- ↵
Carter AJ, Scott D, Bailey L, et al. Dose-response effects in an atherosclerotic porcine coronary model. Circulation. 1999;100:1548–1554.
- ↵
Wilcox JN, Nakahara K. Perivascular proliferative responses after angioplasty leading to post angioplasty restenosis. In: Vascular Brachytherapy: New Perspectives. London, UK: Remedica; 1999.
- ↵
Scott NA, Ross CR, Dunn B, et al. Identification of a potential role for the adventitia in vascular lesion formation after balloon stretch injury of porcine coronary arteries. Circulation. 1996;93:2178–2187.
- ↵
Nakahara K, Okamoto E, Galis ZS, et al. Adventitial expression of MMP-1, MMP-2, TIMP-1 and TIMP-2 during vascular remodeling after angioplasty of porcine coronary arteries. Circulation. 1999;100(suppl I):I-700. Abstract.
- ↵
Wang H, Griendling KK, Scott NA, et al. Intravascular radiation inhibits cell proliferation and vascular remodeling after angioplasty by increasing the expression of P21 in adventitial myofibroblasts. Circulation. 1999;100(suppl I):I-700. Abstract.
- ↵
Fareh J, Martel R, Kermani P, et al. Cellular effects of β-particle delivery on vascular smooth muscle cells and endothelial cells: a dose-response study. Circulation. 1999;99:1477–1484.
- ↵
- ↵
- ↵
Paterson R. The Treatment of Malignant Disease by Radiotherapy. 2nd ed. London, UK: Edward Arnold Publishers; 1963.
- ↵
Serruys PW, Kay IP. I like the candy, I hate the wrapper: the 32P radioactive stent. Circulation. 2000;101:3–7.
- ↵
Sabaté M, Costa MA, Kozuma K, et al. Geographical miss: a cause of treatment failure in radio-oncology applied to intracoronary radiation therapy. Circulation.. 2000;101:2467–2471.
- ↵
Sabaté M, Marijnissen JPA, Carlier SG, et al. Residual plaque burden, delivered dose, and tissue composition predict 6-month outcome after balloon angioplasty and β - radiation therapy. Circulation. 2000;101:2472–2477.
This Issue
Jump to
Article Tools
- Positive Geometric Vascular Remodeling Is Seen After Catheter-Based Radiation Followed by Conventional Stent Implantation but Not After Radioactive Stent ImplantationI. Patrick Kay, Manel Sabaté, Marco A. Costa, Ken Kozuma, Mariano Albertal, Willem J. van der Giessen, Alexander J. Wardeh, Jurgen M. R. Ligthart, Veronique M. A. Coen, Peter C. Levendag and Patrick W. SerruysCirculation. 2000;102:1434-1439, originally published September 19, 2000https://doi.org/10.1161/01.CIR.102.12.1434
Citation Manager Formats
Share this Article
- Positive Geometric Vascular Remodeling Is Seen After Catheter-Based Radiation Followed by Conventional Stent Implantation but Not After Radioactive Stent ImplantationI. Patrick Kay, Manel Sabaté, Marco A. Costa, Ken Kozuma, Mariano Albertal, Willem J. van der Giessen, Alexander J. Wardeh, Jurgen M. R. Ligthart, Veronique M. A. Coen, Peter C. Levendag and Patrick W. SerruysCirculation. 2000;102:1434-1439, originally published September 19, 2000https://doi.org/10.1161/01.CIR.102.12.1434