Submitted on July 14, 2004
From the Division of Cardiology, Department of Medicine and Therapeutics, Prince of Wales Hospital, Chinese University of Hong Kong, Hong Kong (C.-M.Y., S.L.-K.); Medtronic, Inc, Minneapolis, Minn (L.W., J.C., R.W.S.); Grantham Hospital, Hong Kong (E.C.); and Division of Cardiology, Department of Medicine, Queen Mary Hospital, University of Hong Kong, Hong Kong (R.H.-W.C., M.-D.T., C.-P.L.). * To whom correspondence should be addressed. E-mail: cmyu{at}cuhk.edu.hk.
Background--Patients with heart failure are frequently hospitalized for fluid overload. A reliable method for chronic monitoring of fluid status is therefore desirable. We evaluated an implantable system capable of measuring intrathoracic impedance to identify potential fluid overload before heart failure hospitalization and to determine the correlation between intrathoracic impedance and standard measures of fluid status during hospitalization. Methods and Results--Thirty-three patients with NYHA class III and IV heart failure were implanted with a special pacemaker in the left pectoral region and a defibrillation lead in the right ventricle. Intrathoracic impedance was regularly measured and recorded between the lead and the pacemaker case. During hospitalizations, pulmonary capillary wedge pressure and fluid status were monitored. Ten patients were hospitalized for fluid overload 25 times over 20.7±8.4 months. Intrathoracic impedance decreased before each admission by an average of 12.3±5.3% (P<0.001) over an average of 18.3±10.1 days. Impedance reduction began 15.3±10.6 days (P<0.001) before the onset of worsening symptoms. There was an inverse correlation between intrathoracic impedance and pulmonary capillary wedge pressure (r=-0.61, P<0.001) and between intrathoracic impedance and net fluid loss (r=-0.70, P<0.001) during hospitalization. Automated detection of impedance decreases was 76.9% sensitive in detecting hospitalization for fluid overload, with 1.5 false-positive (threshold crossing without hospitalization) detections per patient-year of follow-up. Conclusions--Intrathoracic impedance is inversely correlated with pulmonary capillary wedge pressure and fluid balance and decreased before the onset of patient symptoms and before hospital admission for fluid overload. Regular monitoring of impedance may provide early warning of impending decompensation and diagnostic information for titration of medication.
Revised on April 22, 2005
Accepted on April 26, 2005
Intrathoracic Impedance Monitoring in Patients With Heart Failure. Correlation With Fluid Status and Feasibility of Early Warning Preceding Hospitalization
Cheuk-Man Yu MD, FRCP*,
This article has been cited by other articles:
![]() |
M. R. Zile, T. D. Bennett, M. St. John Sutton, Y. K. Cho, P. B. Adamson, M. F. Aaron, J. M. Aranda Jr, W. T. Abraham, F. W. Smart, L. W. Stevenson, et al. Transition From Chronic Compensated to Acute Decompensated Heart Failure: Pathophysiological Insights Obtained From Continuous Monitoring of Intracardiac Pressures Circulation, September 30, 2008; 118(14): 1433 - 1441. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. R. Teerlink Learning the points of COMPASS-HF: assessing implantable hemodynamic monitoring in heart failure patients. J. Am. Coll. Cardiol., March 18, 2008; 51(11): 1080 - 1082. [Full Text] [PDF] |
||||
![]() |
M. A. Silver New Approaches to Hemodynamic Measurement: Cool Devices But a Shaky Infrastructure J. Am. Coll. Cardiol., December 18, 2007; 50(25): 2383 - 2384. [Full Text] [PDF] |
||||
![]() |
H. E. Verdejo, P. F. Castro, R. Concepcion, M. A. Ferrada, M. A. Alfaro, M. E. Alcaino, C. C. Deck, and R. C. Bourge Comparison of a Radiofrequency-Based Wireless Pressure Sensor to Swan-Ganz Catheter and Echocardiography for Ambulatory Assessment of Pulmonary Artery Pressure in Heart Failure J. Am. Coll. Cardiol., December 18, 2007; 50(25): 2375 - 2382. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Ritzema, I. C. Melton, A. M. Richards, I. G. Crozier, C. Frampton, R. N. Doughty, J. Whiting, S. Kar, N. Eigler, H. Krum, et al. Direct Left Atrial Pressure Monitoring in Ambulatory Heart Failure Patients: Initial Experience With a New Permanent Implantable Device Circulation, December 18, 2007; 116(25): 2952 - 2959. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Frassi, L. Gargani, S. Gligorova, Q. Ciampi, G. Mottola, and E. Picano Clinical and echocardiographic determinants of ultrasound lung comets Eur J Echocardiogr, December 1, 2007; 8(6): 474 - 479. [Abstract] [Full Text] [PDF] |
||||
![]() |
C.-P. Lau, H.-F. Tse, A. J. Camm, and S. S. Barold Evolution of pacing for bradycardias: sensors Eur. Heart J. Suppl., December 1, 2007; 9(suppl_I): I11 - I22. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. E. Wolfel Can We Predict and Prevent the Onset of Acute Decompensated Heart Failure? Circulation, October 2, 2007; 116(14): 1526 - 1529. [Full Text] [PDF] |
||||
![]() |
Authors/Task Force Members, P. E. Vardas, A. Auricchio, J.-J. Blanc, J.-C. Daubert, H. Drexler, H. Ector, M. Gasparini, C. Linde, F. B. Morgado, et al. Guidelines for cardiac pacing and cardiac resynchronization therapy: The Task Force for Cardiac Pacing and Cardiac Resynchronization Therapy of the European Society of Cardiology. Developed in Collaboration with the European Heart Rhythm Association Europace, October 1, 2007; 9(10): 959 - 998. [Full Text] [PDF] |
||||
![]() |
Authors/Task Force Members, P. E. Vardas, A. Auricchio, J.-J. Blanc, J.-C. Daubert, H. Drexler, H. Ector, M. Gasparini, C. Linde, F. B. Morgado, et al. Guidelines for cardiac pacing and cardiac resynchronization therapy: The Task Force for Cardiac Pacing and Cardiac Resynchronization Therapy of the European Society of Cardiology. Developed in Collaboration with the European Heart Rhythm Association Eur. Heart J., September 2, 2007; 28(18): 2256 - 2295. [Full Text] [PDF] |
||||
![]() |
D. Vollmann, H. Nagele, P. Schauerte, U. Wiegand, C. Butter, G. Zanotto, A. Quesada, A. Guthmann, M. R.S. Hill, B. Lamp, et al. Clinical utility of intrathoracic impedance monitoring to alert patients with an implanted device of deteriorating chronic heart failure Eur. Heart J., August 1, 2007; 28(15): 1835 - 1840. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Page, S. Cazeau, P. Ritter, D. Galley, and C. Casset Physiological approach to monitor patients in congestive heart failure: application of a new implantable device-based system to monitor daily life activity and ventilation Europace, August 1, 2007; 9(8): 687 - 693. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. H. Schoenfeld Contemporary Pacemaker and Defibrillator Device Therapy: Challenges Confronting the General Cardiologist Circulation, February 6, 2007; 115(5): 638 - 653. [Full Text] [PDF] |
||||
![]() |
W.H. W. Tang and G. S. Francis The Year in Heart Failure J. Am. Coll. Cardiol., December 6, 2005; 46(11): 2125 - 2133. [Full Text] [PDF] |
||||
![]() |
Intrathoracic Impedance Monitoring in HF Patients Journal Watch Cardiology, September 9, 2005; 2005(909): 3 - 3. [Full Text] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2005 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |