(Circulation. 1997;96:3369-3377.)
© 1997 American Heart Association, Inc.
Articles |
Correspondence to P.W. Serruys, Department of Coronary Diagnostics and Interventions, Thoraxcenter, PO Box 1738, 3000 Dr Rotterdam, Netherlands.
Background The aim of this prospective, multicenter study was the identification of Doppler flow velocity measurements predictive of clinical outcome of patients undergoing single-vessel balloon angioplasty with no previous Q-wave myocardial infarction.
Methods and Results In 297 patients, a Doppler
guidewire was used to measure basal and maximal hyperemic flow
velocities proximal and distal to the stenosis before and after
angioplasty. In 225 patients with an angiographically successful
percutaneous transluminal coronary angioplasty
(PTCA), postprocedural distal coronary flow reserve (CFR) and
percent diameter stenosis (DS%) were correlated with symptoms
and/or ischemia at 1 and 6 months, with the need for target
lesion revascularization, and with angiographic
restenosis (defined as DS
50% at follow-up). Logistic
regression and receiver operator characteristic curve analyses
were applied to determine the prognostic cutoff value of CFR and DS
separately and in combination. Optimal cutoff criteria for predictors
of these clinical events were DS, 35%; CFR, 2.5. A distal CFR after
angioplasty >2.5 with a residual DS
35% identified lesions with a
low incidence of recurrence of symptoms at 1 month (10% versus
19%, P=.149) and at 6 months (23% versus 47%,
P=.005), a low need for reintervention (16% versus 34%,
P=.024), and a low restenosis rate (16% versus
41%, P=.002) compared with patients who did not meet these
criteria.
Conclusions Measurements of distal CFR after PTCA, in combination with DS%, have a predictive value, albeit modest for the short- and long-term outcomes after PTCA, and thus may be used to identify patients who will or will not benefit from additional therapy such as stent implantation.
Key Words: stenosis balloon prognosis angioplasty
This article has been cited by other articles:
![]() |
T. Kurita, H. Sakuma, K. Onishi, M. Ishida, K. Kitagawa, T. Yamanaka, T. Tanigawa, T. Kitamura, K. Takeda, and M. Ito Regional myocardial perfusion reserve determined using myocardial perfusion magnetic resonance imaging showed a direct correlation with coronary flow velocity reserve by Doppler flow wire Eur. Heart J., February 2, 2009; 30(4): 444 - 452. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Beleslin, M. Ostojic, A. Djordjevic-Dikic, V. Vukcevic, S. Stojkovic, M. Nedeljkovic, G. Stankovic, D. Orlic, N. Milic, J. Stepanovic, et al. The value of fractional and coronary flow reserve in predicting myocardial recovery in patients with previous myocardial infarction Eur. Heart J., November 1, 2008; 29(21): 2617 - 2624. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Kawamoto, H. Okura, Y. Koyama, I. Toda, H. Taguchi, K. Tamita, A. Yamamuro, Y. Yoshimura, Y. Neishi, E. Toyota, et al. The Relationship Between Coronary Plaque Characteristics and Small Embolic Particles During Coronary Stent Implantation J. Am. Coll. Cardiol., October 23, 2007; 50(17): 1635 - 1640. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Costa, S. Shoemaker, H. Futamatsu, C. Klassen, D. J. Angiolillo, M. Nguyen, A. Siuciak, P. Gilmore, M. M. Zenni, L. Guzman, et al. Quantitative Magnetic Resonance Perfusion Imaging Detects Anatomic and Physiologic Coronary Artery Disease as Measured by Coronary Angiography and Fractional Flow Reserve J. Am. Coll. Cardiol., August 7, 2007; 50(6): 514 - 522. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Kern, A. Lerman, J.-W. Bech, B. De Bruyne, E. Eeckhout, W. F. Fearon, S. T. Higano, M. J. Lim, M. Meuwissen, J. J. Piek, et al. Physiological Assessment of Coronary Artery Disease in the Cardiac Catheterization Laboratory: A Scientific Statement From the American Heart Association Committee on Diagnostic and Interventional Cardiac Catheterization, Council on Clinical Cardiology Circulation, September 19, 2006; 114(12): 1321 - 1341. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Fenchel, A. Franow, N. I. Stauder, U. Kramer, U. Helber, C. D. Claussen, and S. Miller Myocardial Perfusion after Angioplasty in Patients Suspected of Having Single-Vessel Coronary Artery Disease: Improvement Detected at Rest-Stress First-Pass Perfusion MR Imaging--Initial Experience Radiology, October 1, 2005; 237(1): 67 - 74. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Gibson, A. J. Kirtane, K. Boundy, H. Ly, D. Karmpaliotis, S. A. Murphy, R. P. Giugliano, C. P. Cannon, E. M. Antman, E. Braunwald, et al. Association of a negative residual stenosis following rescue/adjunctive percutaneous coronary intervention with impaired myocardial perfusion and adverse outcomes among ST-segment elevation myocardial infarction patients J. Am. Coll. Cardiol., February 1, 2005; 45(3): 357 - 362. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Voci, F. Pizzuto, and F. Romeo Coronary flow: a new asset for the echo lab? Eur. Heart J., November 1, 2004; 25(21): 1867 - 1879. [Full Text] [PDF] |
||||
![]() |
G S Werner, U Emig, P Bahrmann, M Ferrari, and H R Figulla Recovery of impaired microvascular function in collateral dependent myocardium after recanalisation of a chronic total coronary occlusion Heart, November 1, 2004; 90(11): 1303 - 1309. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. K. Chugh, J. Koppel, M. Scott, L. Shewchuk, D. Goodhart, R. Bonan, J.-C. Tardif, S. G. Worthley, C. DiMario, M. J. Curtis, et al. Coronary flow velocity reserve does not correlate with TIMI frame count in patients undergoing non-emergency percutaneous coronary intervention J. Am. Coll. Cardiol., August 18, 2004; 44(4): 778 - 782. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Lethen and H. Lambertz Doppler echocardiographic assessment of coronary artery disease: a challenge becomes reality Eur J Echocardiogr, January 1, 2004; 5(1): 5 - 7. [Full Text] [PDF] |
||||
![]() |
H. Lethen, H. P. Tries, S. Kersting, and H. Lambertz Validation of noninvasive assessment of coronary flow velocity reserve in the right coronary artery: A comparison of transthoracic echocardiographic results with intracoronary Doppler flow wire measurements Eur. Heart J., September 1, 2003; 24(17): 1567 - 1575. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. S. Werner, P. Bahrmann, O. Mutschke, U. Emig, S. Betge, M. Ferrari, and H. R. Figulla Determinants of target vessel failure in chronic total coronary occlusions after stent implantation: The influence of collateral function and coronary hemodynamics J. Am. Coll. Cardiol., July 16, 2003; 42(2): 219 - 225. [Abstract] [Full Text] [PDF] |
||||
![]() |
D V Cokkinos, A Manginas, and V Voudris Coronary flow: clinical considerations Heart, April 1, 2003; 89(4): 361 - 363. [Full Text] [PDF] |
||||
![]() |
E. Barbato, J. Marco, and W. Wijns Direct stenting Eur. Heart J., March 1, 2003; 24(5): 394 - 403. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Kern The value of systolic flow velocity acceleration after PCI Eur. Heart J., December 1, 2002; 23(23): 1801 - 1802. [Full Text] [PDF] |
||||
![]() |
M. Albertal, E. Regar, G. Van Langenhove, S.G. Carlier, J.J. Piek, B. de Bruyne, C. di Mario, D. Foley, K. Kozuma, M.A. Costa, et al. Value of coronary stenotic flow velocity acceleration in prediction of angiographic restenosis following balloon angioplasty Eur. Heart J., December 1, 2002; 23(23): 1849 - 1853. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Ruscazio, R. Montisci, P. Colonna, C. Caiati, L. Chen, G. Lai, M. Cadeddu, R. Pirisi, and S. Iliceto Detection of coronary restenosis aftercoronary angioplasty by contrast-enhanced transthoracic echocardiographic Doppler assessment of coronary flow velocity reserve J. Am. Coll. Cardiol., September 4, 2002; 40(5): 896 - 903. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Voskuil, R. A. M. van Liebergen, M. Albertal, E. Boersma, J. G. P. Tijssen, P. W. Serruys, J. J. Piek, and the DEBATE II Investigators Coronary hemodynamics of stent implantation after suboptimal and optimal balloon angioplasty J. Am. Coll. Cardiol., May 1, 2002; 39(9): 1513 - 1517. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. D. Miller Coronary flow studies for risk stratification in multivessel disease: A physiologic bridge too far? J. Am. Coll. Cardiol., March 6, 2002; 39(5): 859 - 863. [Full Text] [PDF] |
||||
![]() |
T. Nishida, C. Di Mario, M.J. Kern, T.J. Anderson, I. Moussa, R. Bonan, T. Muramatsu, A.C. Jain, J. Suarez de Lezo, S.Y. Cho, et al. Impact of final coronary flow velocity reserve on late outcome following stent implantation Eur. Heart J., February 2, 2002; 23(4): 331 - 340. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Albertal, E. Regar, G. Van Langenhove, S.G. Carlier, P. Serrano, E. Boersma, B. Bruyne, C. Di Mario, J. Piek, and P.W. Serruys Flow velocity and predictors of a suboptimal coronary flow velocity reserve after coronary balloon angioplasty Eur. Heart J., January 2, 2002; 23(2): 133 - 138. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. S. Werner, M. Ferrari, B. M. Richartz, O. Gastmann, and H. R. Figulla Microvascular Dysfunction in Chronic Total Coronary Occlusions Circulation, September 4, 2001; 104(10): 1129 - 1134. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Seabra-Gomes Is there a future for coronary physiological evaluation in clinical decision making? Eur. Heart J., September 2, 2001; 22(18): 1633 - 1635. [PDF] |
||||
![]() |
J.J Piek, E Boersma, M Voskuil, C di Mario, E Schroeder, C Vrints, P Probst, B de Bruyne, C Hanet, E Fleck, et al. The immediate and long-term effect of optimal balloon angioplasty on the absolute coronary blood flow velocity reserve. A subanalysis of the DEBATE study Eur. Heart J., September 2, 2001; 22(18): 1725 - 1732. [Abstract] [PDF] |
||||
![]() |
M Albertal, G Van Langenhove, E Regar, I P Kay, D Foley, G Sianos, K Kozuma, T Beijsterveldt, S G Carlier, J A Belardi, et al. Uncomplicated moderate coronary artery dissections after balloon angioplasty: good outcome without stenting Heart, August 1, 2001; 86(2): 193 - 198. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Pizzuto, P. Voci, E. Mariano, P. Emilio Puddu, G. Sardella, and A. Nigri Assessment of flow velocity reserve by transthoracic Doppler echocardiography and venous adenosine infusion before and after left anterior descending coronary artery stenting J. Am. Coll. Cardiol., July 1, 2001; 38(1): 155 - 162. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Kern and B. Meier Evaluation of the Culprit Plaque and the Physiological Significance of Coronary Atherosclerotic Narrowings Circulation, June 26, 2001; 103(25): 3142 - 3149. [Full Text] [PDF] |
||||
![]() |
S. C. Smith Jr, J. T. Dove, A. K. Jacobs, J. Ward Kennedy, D. Kereiakes, M. J. Kern, R. E. Kuntz, J. J. Popma, H. V. Schaff, D. O. Williams, et al. ACC/AHA guidelines for percutaneous coronary intervention (revision of the 1993 PTCA guidelines): A report of the American College of Cardiology/ American Heart Association Task Force on practice guidelines (Committee to revise the 1993 guidelines for percutaneous transluminal coronary angioplasty) endorsed by the Society for Cardiac Angiography and Interventions J. Am. Coll. Cardiol., June 15, 2001; 37(8): 2239 - 2239. [Full Text] [PDF] |
||||
![]() |
W. J. Cantor, A. S. Hellkamp, E. D. Peterson, J. P. Zidar, P. A. Cowper, M. H. Sketch Jr, J. E. Tcheng, R. M. Califf, and E. M. Ohman Achieving optimal results with standard balloon angioplasty: can baseline and angiographic variables predict stent-like outcomes? J. Am. Coll. Cardiol., June 1, 2001; 37(7): 1883 - 1890. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Herrmann, M. Haude, A. Lerman, R. Schulz, L. Volbracht, J. Ge, A. Schmermund, H. Wieneke, C. von Birgelen, H. Eggebrecht, et al. Abnormal Coronary Flow Velocity Reserve After Coronary Intervention Is Associated With Cardiac Marker Elevation Circulation, May 15, 2001; 103(19): 2339 - 2345. [Abstract] [Full Text] [PDF] |
||||
![]() |
C Le Feuvre, G Dambrin, G Helft, F Beygui, M Touam, J P Grünfeld, A Vacheron, and J P Metzger Clinical outcome following coronary angioplasty in dialysis patients: a case-control study in the era of coronary stenting Heart, May 1, 2001; 85(5): 556 - 560. [Abstract] [Full Text] |
||||
![]() |
M. Haude, D. Baumgart, E. Verna, J. J. Piek, C. Vrints, P. Probst, and R. Erbel Intracoronary Doppler- and Quantitative Coronary Angiography-Derived Predictors of Major Adverse Cardiac Events After Stent Implantation Circulation, March 6, 2001; 103(9): 1212 - 1217. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. R. Ward, P. S. Tsao, A. Agrotis, R. J. Dilley, G. L. Jennings, and A. Bobik Low Blood Flow After Angioplasty Augments Mechanisms of Restenosis : Inward Vessel Remodeling, Cell Migration, and Activity of Genes Regulating Migration Arterioscler Thromb Vasc Biol, February 1, 2001; 21(2): 208 - 213. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. V. Anderson and B. A. Carabello Provisional Versus Routine Stenting : Routine Stenting Is Here To Stay Circulation, December 12, 2000; 102(24): 2910 - 2914. [Full Text] [PDF] |
||||
![]() |
P. W. Serruys, B. de Bruyne, S. Carlier, J. E. Sousa, J. Piek, T. Muramatsu, C. Vrints, P. Probst, R. Seabra-Gomes, I. Simpson, et al. Randomized Comparison of Primary Stenting and Provisional Balloon Angioplasty Guided by Flow Velocity Measurement Circulation, December 12, 2000; 102(24): 2930 - 2937. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Di Mario, J. W. Moses, T. J. Anderson, R. Bonan, T. Muramatsu, A. C. Jain, J. Suarez de Lezo, S. Y. Cho, M. Kern, I. T. Meredith, et al. Randomized Comparison of Elective Stent Implantation and Coronary Balloon Angioplasty Guided by Online Quantitative Angiography and Intracoronary Doppler Circulation, December 12, 2000; 102(24): 2938 - 2944. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Verna, L. Ceriani, L. Giovanella, G. Binaghi, and S. Garancini "False-Positive" Myocardial Perfusion Scintigraphy Findings in Patients with Angiographically Normal Coronary Arteries: Insights from Intravascular Sonography Studies J. Nucl. Med., December 1, 2000; 41(12): 1935 - 1940. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. W. Frey, J. M. Hodgson, C. Muller, H.-P. Bestehorn, and H. Roskamm Ultrasound-Guided Strategy for Provisional Stenting With Focal Balloon Combination Catheter : Results From the Randomized Strategy for Intracoronary Ultrasound-Guided PTCA and Stenting (SIPS) Trial Circulation, November 14, 2000; 102(20): 2497 - 2502. [Abstract] [Full Text] [PDF] |
||||
![]() |
P.W Serruys, J.N Hamburger, J Fajadet, M Haude, H Klues, R Seabra-Gomes, T Corcos, C Hamm, L Pizzuli, B Meier, et al. Total occlusion trial with angioplasty by using laser guidewire. The TOTAL trial Eur. Heart J., November 1, 2000; 21(21): 1797 - 1805. [Abstract] [PDF] |
||||
![]() |
N. Al-Saadi, E. Nagel, M. Gross, B. Schnackenburg, I. Paetsch, C. Klein, and E. Fleck Improvement of myocardial perfusion reserve early after coronary intervention: assessment with cardiac magnetic resonance imaging J. Am. Coll. Cardiol., November 1, 2000; 36(5): 1557 - 1564. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. J. Cantor, E. D. Peterson, J. J. Popma, J. P. Zidar, M. H. Sketch Jr., J. E. Tcheng, and E. M. Ohman Provisional stenting strategies: systematic overview and implications for clinical decision-making J. Am. Coll. Cardiol., October 1, 2000; 36(4): 1142 - 1151. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. R. Ward, G. Pasterkamp, A. C. Yeung, and C. Borst Arterial Remodeling : Mechanisms and Clinical Implications Circulation, September 5, 2000; 102(10): 1186 - 1191. [Full Text] [PDF] |
||||
![]() |
A. Lafont, J. L. Dubois-Rande, P. G. Steg, P. Dupouy, D. Carrie, P. Coste, A. Furber, F. Beygui, L. J. Feldman, S. Rahal, et al. The French randomized optimal stenting trial: a prospective evaluation of provisional stenting guided by coronary velocity reserve and quantitative coronary angiography J. Am. Coll. Cardiol., August 1, 2000; 36(2): 404 - 409. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Wieneke, M. Haude, J. Ge, C. Altmann, S. Kaiser, D. Baumgart, C. von Birgelen, D. Welge, and R. Erbel Corrected coronary flow velocity reserve: a new concept for assessing coronary perfusion J. Am. Coll. Cardiol., June 1, 2000; 35(7): 1713 - 1720. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. G. Hundley, L. D. Hillis, C. A. Hamilton, R. J. Applegate, D. M. Herrington, G. D. Clarke, G. A. Braden, M. S. Thomas, R. A. Lange, R. M. Peshock, et al. Assessment of Coronary Arterial Restenosis With Phase-Contrast Magnetic Resonance Imaging Measurements of Coronary Flow Reserve Circulation, May 23, 2000; 101(20): 2375 - 2381. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Kern Coronary Physiology Revisited : Practical Insights From the Cardiac Catheterization Laboratory Circulation, March 21, 2000; 101(11): 1344 - 1351. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. P. Faxon Predicting Restenosis : Bigger Is Better but Not Best Circulation, March 7, 2000; 101(9): 946 - 947. [Full Text] [PDF] |
||||
![]() |
J. A. Suwaidi, S. Hamasaki, S. T. Higano, R. A. Nishimura, D. R. Holmes Jr, and A. Lerman Long-Term Follow-Up of Patients With Mild Coronary Artery Disease and Endothelial Dysfunction Circulation, March 7, 2000; 101(9): 948 - 954. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Stankovic, A. Manginas, V. Voudris, G. Pavlides, G. Athanassopoulos, M. Ostojic, and D. V. Cokkinos Prediction of Restenosis After Coronary Angioplasty by Use of a New Index : TIMI Frame Count/Minimal Luminal Diameter Ratio Circulation, March 7, 2000; 101(9): 962 - 968. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Seabra-Gomes Functional assessment of coronary stenosis: it does make sense, but why don't I do it more often? Eur. Heart J., March 2, 2000; 21(6): 422 - 424. [PDF] |
||||
![]() |
J.J Piek, E Boersma, C di Mario, E Schroeder, C Vrints, P Probst, B de Bruyne, C Hanet, E Fleck, M Haude, et al. Angiographical and Doppler flow-derived parameters for assessment of coronary lesion severity and its relation to the result of exercise electrocardiography Eur. Heart J., March 2, 2000; 21(6): 466 - 474. [Abstract] [PDF] |
||||
![]() |
C LE FEUVRE Angioplasty and stenting in patients with renal disease Heart, January 1, 2000; 83(1): 7 - 8. [Full Text] [PDF] |
||||
![]() |
M. J. Kern, S. Puri, R. G. Bach, T. J. Donohue, P. Dupouy, E. A. Caracciolo, W. R. Craig, F. Aguirre, E. Aptecar, T. L. Wolford, et al. Abnormal Coronary Flow Velocity Reserve After Coronary Artery Stenting in Patients : Role of Relative Coronary Reserve to Assess Potential Mechanisms Circulation, December 21, 1999; 100(25): 2491 - 2498. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A. M. van Liebergen, J. J. Piek, K. T. Koch, R. J. G. Peters, R. J. de Winter, C. E. Schotborgh, and K. I. Lie Hyperemic coronary flow after optimized intravascular ultrasound-guided balloon angioplasty and stent implantation J. Am. Coll. Cardiol., December 1, 1999; 34(7): 1899 - 1906. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Colombo and C. Briguori Optimal balloon angioplasty versus elective stent implantation J. Am. Coll. Cardiol., December 1, 1999; 34(7): 1907 - 1909. [Full Text] [PDF] |
||||
![]() |
A. Betriu, M. Masotti, A. Serra, J. Alonso, F. Fernandez-Aviles, F. Gimeno, T. Colman, J. Zueco, J. L. Delcan, E. Garcia, et al. Randomized comparison of coronary stent implantation and balloon angioplasty in the treatment of de novo coronary artery lesions (START): A four-year follow-up J. Am. Coll. Cardiol., November 1, 1999; 34(5): 1498 - 1506. [Abstract] [Full Text] [PDF] |
||||
![]() |
C J Vrints, M J Claeys, J Bosmans, V Conraads, and J P Snoeck Effect of stenting on coronary flow velocity reserve: comparison of coil and tubular stents Heart, October 1, 1999; 82(4): 465 - 470. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Takagi, Y. Tsurumi, Y. Ishii, K. Suzuki, M. Kawana, and H. Kasanuki Clinical Potential of Intravascular Ultrasound for Physiological Assessment of Coronary Stenosis : Relationship Between Quantitative Ultrasound Tomography and Pressure-Derived Fractional Flow Reserve Circulation, July 20, 1999; 100(3): 250 - 255. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. J. W. Bech, N. H. J. Pijls, B. De Bruyne, K. H. Peels, H. R. Michels, H. J. R. M. Bonnier, and J. J. Koolen Usefulness of Fractional Flow Reserve to Predict Clinical Outcome After Balloon Angioplasty Circulation, February 23, 1999; 99(7): 883 - 888. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Ferrari, B.a. Schnell, G. S. Werner, and H. R. Figulla Safety of deferring angioplasty in patients with normal coronary flow velocity reserve J. Am. Coll. Cardiol., January 1, 1999; 33(1): 82 - 87. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A. M. van Liebergen, J. J. Piek, K. T. Koch, R. J. de Winter, and K. I. Lie Immediate and Long-Term Effect of Balloon Angioplasty or Stent Implantation on the Absolute and Relative Coronary Blood Flow Velocity Reserve Circulation, November 17, 1998; 98(20): 2133 - 2140. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Rodriguez, F. Ayala, V. Bernardi, O. Santaera, E. Marchand, C. Pardinas, C. Mauvecin, D. Vogel, L. C. Harrell, I. F. Palacios, et al. Optimal coronary balloon angioplasty with provisional stenting versus primary stent (OCBAS): Immediate and long-term follow-up results J. Am. Coll. Cardiol., November 1, 1998; 32(5): 1351 - 1357. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. R. Holmes Jr., J. Hirshfeld Jr., D. Faxon, R. Vlietstra, A. Jacobs, S. B. King III, T. M. Bashore, N. D. Bridges, C. B. Higgins, L. F. Hiratzka, et al. ACC expert consensus document on coronary artery stents: Document of the American College of Cardiology J. Am. Coll. Cardiol., November 1, 1998; 32(5): 1471 - 1482. [Full Text] [PDF] |
||||
![]() |
M. Albertal, M. Voskuil, J.J. Piek, B. de Bruyne, G. Van Langenhove, P.I. Kay, M.A. Costa, E. Boersma, T. Beijsterveldt, J.E. Sousa, et al. Coronary Flow Velocity Reserve After Percutaneous Interventions Is Predictive of Periprocedural Outcome Circulation, April 2, 2002; 105(13): 1573 - 1578. [Abstract] [Full Text] [PDF] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1997 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |