(Circulation. 1996;94:3311-3317.)
© 1996 American Heart Association, Inc.
Articles |
the Division of Cardiology, Faculty of Medicine, University of Leicester, UK (R.K.A., D.C.I., D.P. de B., A.H.G.); the Division of Cardiology, University of Connecticut, Farmington (M.A.A.); and the Section of Cardiovascular Medicine, Yale University, New Haven, Conn (M.D.E.).
Correspondence to Dr R.K. Aggarwal, Department of Cardiology, Regional Cardiothoracic Centre, Freeman Hospital, Newcastle upon Tyne NE7 7DN, UK.
Background Monoclonal anti-rabbit platelet glycoprotein (GP) IIb/IIIa antibody (AZ1) was adsorbed onto cellulose polymercoated intracoronary stents to enhance their thromboresistance. We evaluated the antithrombotic efficacy of AZ1 antibodyeluting stents.
Methods and Results Twenty-three polymer-coated stents with AZ1 antibody bound by passive adsorption (AZ1-eluting) were compared with 23 control polymer-coated stents adsorbed with either no antibody (base-polymer, n=12) or isotype-matched irrelevant antibody (anti-CMVeluting, n=11) by implantation into balloon-damaged, flow-reduced iliac arteries of New Zealand White rabbits. In 13 animals (acute group), flow measurements were made with transit-time flow probes and platelet adhesion was ascertained by use of 111In-labeled autologous platelets. In the other 10 animals (chronic group), stent occlusion was assessed macroscopically after they were killed 28 days after stenting. Arteries with AZ1-eluting stents had significantly less platelet deposition (15.8±4.5x107) than either base-polymer (32.1±4.3x107) or anti-CMVeluting (35.2±8.8x107) controls (ANOVA, P<.0001). Compared with base-polymer or anti-CMVeluting controls, arteries with AZ1-eluting stents showed a marked reduction in cyclic blood flow variation (P<.0001) and a significantly greater mean blood flow 2 hours after stent deployment (P<.0001). There was a significant improvement in the patency rate of AZ1-eluting stents compared with controls at both 2 hours (92% versus 46%, P=.034) and 28 days (100% versus 40%, P=.015).
Conclusions Platelet GP IIb/IIIa antibody eluting from polymer-coated stents reduces platelet deposition, improves blood flow, virtually abolishes cyclic flow variation, and improves patency rates after stent implantation in a rabbit iliac artery model. Its potential for reducing stent-related thrombosis in humans warrants further evaluation.
Key Words: stents thrombosis platelet aggregation inhibitors
This article has been cited by other articles:
![]() |
J. E. Sousa, P. W. Serruys, and M. A. Costa New Frontiers in Cardiology: Drug-Eluting Stents: Part II Circulation, May 13, 2003; 107(18): 2383 - 2389. [Full Text] [PDF] |
||||
![]() |
J. H. Baron, A. H. Gershlick, K. Hogrefe, J. Armstrong, C. M. Holt, R. K. Aggarwal, M. Azrin, M. Ezekowitz, and D. P. de Bono In vitro evaluation of c7E3-Fab (ReoProTM) eluting polymer-coated coronary stents Cardiovasc Res, June 1, 2000; 46(3): 585 - 594. [Abstract] [Full Text] [PDF] |
||||
![]() |
D M Whelan, W J van der Giessen, S C Krabbendam, E A van Vliet, P D Verdouw, P W Serruys, and H M M van Beusekom Biocompatibility of phosphorylcholine coated stents in normal porcine coronary arteries Heart, March 1, 2000; 83(3): 338 - 345. [Abstract] [Full Text] |
||||
![]() |
J Gunn and D Cumberland Stent coatings and local drug delivery. State of the art Eur. Heart J., December 1, 1999; 20(23): 1693 - 1700. [PDF] |
||||
![]() |
R. R. Azar, R. G. McKay, P. D. Thompson, J. A. Hirst, J. F. Mitchell, D. B. Fram, D. D. Waters, and F. J. Kiernan Abciximab in primary coronary angioplasty for acute myocardial infarction improves short- and medium-term outcomes J. Am. Coll. Cardiol., December 1, 1998; 32(7): 1996 - 2002. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. S. Deitch, J. K. Williams, M. R. Adams, C. A. Fly, D. M. Herrington, R. E. Jordan, M. T. Nakada, J. A. Jakubowski, and R. L. Geary Effects of ß3-Integrin Blockade (c7E3) on the Response to Angioplasty and Intra-Arterial Stenting in Atherosclerotic Nonhuman Primates Arterioscler Thromb Vasc Biol, November 1, 1998; 18(11): 1730 - 1737. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. J. Topol and P. W. Serruys Frontiers in Interventional Cardiology Circulation, October 27, 1998; 98(17): 1802 - 1820. [Full Text] [PDF] |
||||
![]() |
O. F. Bertrand, R. Sipehia, R. Mongrain, J. Rodes, J.-C. Tardif, L. Bilodeau, G. Cote, and M. G. Bourassa Biocompatibility aspects of new stent technology J. Am. Coll. Cardiol., September 1, 1998; 32(3): 562 - 571. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Yamawaki, H. Shimokawa, T. Kozai, K. Miyata, T. Higo, E. Tanaka, K. Egashira, T. Shiraishi, H. Tamai, K. Igaki, et al. Intramural delivery of a specific tyrosine kinase inhibitor with biodegradable stent suppresses the restenotic changes of the coronary artery in pigs in vivo J. Am. Coll. Cardiol., September 1, 1998; 32(3): 780 - 786. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Gershlick Endovascular manipulation to restrict restenosis Vascular Medicine, August 1, 1998; 3(3): 177 - 188. [Abstract] [PDF] |
||||
![]() |
A H Gershlick and J Baron Dealing with in-stent restenosis Heart, April 1, 1998; 79(4): 319 - 323. [Full Text] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1996 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |