Circulation, Vol 56, 679-683, Copyright © 1977 by American Heart Association
SS Barold, LS Ong, MD Falkoff and RA Heinle
This report describes inhibition of a normally functioning bipolar demand
pulse generator by diaphragmatic myopotentials. Transient pacemaker
suppression occurred repeatedly with deep respiration, straining, the
Valsalva maneuver, coughing, sneezing and laughing. When the magnet was
applied, none of these maneuvers inhibited the pacemaker. Extensive
investigations ruled out an intermittent electrode problem such as a wire
fracture or insulation break. Sensing of diaphragmatic myopotentials should
be considered in the differential diagnosis of unexplained pacemaker
pauses.
ARTICLES
Inhibition of bipolar demand pacemaker by diaphragmatic myopotentials
This article has been cited by other articles:
![]() |
K. R. Santos, P. Adragao, D. Cavaco, F. B. Morgado, R. Candeias, S. Lima, and J. A. Silva Diaphragmatic myopotential oversensing in pacemaker-dependent patients with CRT-D devices Europace, December 1, 2008; 10(12): 1381 - 1386. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Schulte, J. Sperzel, T. Schwarz, H.F. Pitschner, G. Strupp, and J. Neuzner Detection of ventricular fibrillation in implantable defibrillators with automatic gain control amplifiers: Effects of programming sensitivity Europace, January 1, 2000; 2(2): 160 - 162. [Abstract] [PDF] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1977 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |