(Circulation. 2003;107:1797.)
© 2003 American Heart Association, Inc.
Basic Science Reports |
From Virginia Commonwealth Universitys Medical College of Virginia, Richmond.
Correspondence to Mark A. Wood, MD, Box 980053, Richmond, VA 23298. E-mail mwoodmd{at}pol.net
Background Blood flow near a radiofrequency (RF) lesion can reduce lesion size by convective cooling. It is unknown whether blood flow through small vasculature within an RF lesion can prevent transmural lesion formation.
Methods and Results In 40 rabbit right ventricle preparations, 2 epicardial RF lesions were created straddling a selectively perfused (0 to 12 mL/min) marginal artery (diameter, 0.34±0.1 mm). RF lesions were created at either 60°C or 80°C and delivered either sequentially or simultaneously. Conduction through the lesion area was measured. The lesions were analyzed histologically. At a perfusion rate of 0 mL/min, all RF lesions were transmural and without conduction. As little as 1 mL of flow through the artery during RF delivery could prevent transmural lesion formation by preserving a cuff of tissue along the length of the vessel. High-energy delivery (45 W) and very high tissue temperatures (93°C) were needed to overcome the protective effect of vascular perfusion at 12 mL/min. The volume of preserved myocardium was related to arterial perfusion rate, artery diameter, and lesion temperature but not to the sequence of RF delivery (sequential versus simultaneous). Conduction persisted through the RF lesion in 20 experiments. Conduction through the lesion was related to the arterial perfusion rate and volume or cross-sectional area of preserved myocardium.
Conclusions Flow through even small intramyocardial vessels can prevent transmural lesion formation and preserve conduction through an RF lesion. These findings may represent an unrecognized obstacle to the creation of linear RF lesions in the clinical setting.
Key Words: catheter ablation vasculature conduction
This article has been cited by other articles:
![]() |
S. Takatsuki, F. Extramiana, M. Hayashi, A. Haggui, A. Messali, P. Milliez, A. Leenhardt, and B. Cauchemez High take-off left inferior pulmonary vein as an obstacle in creating a conduction block at the lateral mitral isthmus Europace, July 1, 2009; 11(7): 910 - 916. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Da Costa, C. Romeyer-Bouchard, V. Dauphinot, D. Lipp, L. Abdellaoui, M. Messier, J. Thevenin, J.-C. Barthelemy, and K. Isaaz Cavotricuspid isthmus angiography predicts atrial flutter ablation efficacy in 281 patients randomized between 8 mm- and externally irrigated-tip catheter Eur. Heart J., August 1, 2006; 27(15): 1833 - 1840. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. J. Perez, M. A. Wood, and C. M. Schubert Effects of Gap Geometry on Conduction Through Discontinuous Radiofrequency Lesions Circulation, April 11, 2006; 113(14): 1723 - 1729. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. V. Daniels, Y.-Y. Lu, J. B. Morton, P. A. Santucci, J. G. Akar, A. Green, and D. J. Wilber Idiopathic Epicardial Left Ventricular Tachycardia Originating Remote From the Sinus of Valsalva: Electrophysiological Characteristics, Catheter Ablation, and Identification From the 12-Lead Electrocardiogram Circulation, April 4, 2006; 113(13): 1659 - 1666. [Abstract] [Full Text] [PDF] |
||||
![]() |
G Haywood Can we ablate permanent atrial fibrillation? Heart, February 1, 2006; 92(2): 152 - 154. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. H.M. Wittkampf, M. F. van Oosterhout, P. Loh, R. Derksen, E.-j. Vonken, P. J. Slootweg, and S. Y. Ho Where to draw the mitral isthmus line in catheter ablation of atrial fibrillation: histological analysis Eur. Heart J., April 1, 2005; 26(7): 689 - 695. [Abstract] [Full Text] [PDF] |
||||
![]() |
D Sanchez-Quintana, J A Cabrera, J Farre, V Climent, R H Anderson, and S Y Ho Sinus node revisited in the era of electroanatomical mapping and catheter ablation Heart, February 1, 2005; 91(2): 189 - 194. [Abstract] [Full Text] [PDF] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2003 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |