(Circulation. 2000;102:1484.)
© 2000 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Thoraxcenter (K.K., M.A.C., M.S., I.K., P.S., J.M.R.L., P.W.S.), University Hospital Rotterdam Dijkzigt, the Netherlands; and Daniel den Hoed Cancer Center (J.P.A.M., V.L.M.A.C., P.C.L.), Rotterdam, the Netherlands. Correspondence to P.W. Serruys, MD, PhD, FACC, Department of Interventional Cardiology, University Hospital Rotterdam Dijkzigt, Thoraxcenter Bd408, Dr Molewaterplein 40, 3015 GD Rotterdam, the Netherlands.
BackgroundThe "edge effect," late lumen loss at the margins of the treated segment, has become an important issue in the field of coronary brachytherapy. The aim of the present study was to assess the edge effect in noninjured margins adjacent to the irradiated segments after catheter-based intracoronary ß-irradiation.
Methods and ResultsFifty-three vessels were assessed by means of 3-dimensional intravascular ultrasound after the procedure and at 6- to 8-month follow-up. Fourteen vessels (placebo group) did not receive radiation (sham source), whereas 39 vessels were irradiated. In the irradiated group, 48 edges (5 mm in length) were identified as noninjured, whereas 18 noninjured edges were selected in the placebo group. We compared the volumetric intravascular ultrasound measurements of the noninjured edges of the irradiated vessels with the fully irradiated nonstented segments (IRS, n=27) (26-mm segments received the prescribed 100% isodose) and the noninjured edges of the vessels of the placebo patients. The lumen decreased (6 mm3) in the noninjured edges of the irradiated vessels at follow-up (P=0.001). We observed a similar increase in plaque volume in all segments: noninjured edges of the irradiated group (19.6%), noninjured edges of the placebo group (21.5%), and IRS (21.0%). The total vessel volume increased in the IRS in the 3 groups. No edge segment was subject to repeat revascularization.
ConclusionsThe edge effect occurs in the noninjured margins of radiation source train in both irradiated and placebo patients. Thus, low-dose radiation may not play an important role in this phenomenon, whereas nonmeasurable device injury may be considered a plausible alternative explanation.
Key Words: brachytherapy angioplasty ultrasonics
This article has been cited by other articles:
![]() |
P. Jimenez-Quevedo, M. Sabate, D. J. Angiolillo, M. A. Costa, F. Alfonso, J. A. Gomez-Hospital, R. Hernandez-Antolin, C. Banuelos, J. Goicolea, F. Fernandez-Aviles, et al. Vascular Effects of Sirolimus-Eluting Versus Bare-Metal Stents in Diabetic Patients: Three-Dimensional Ultrasound Results of the Diabetes and Sirolimus-Eluting Stent (DIABETES) Trial J. Am. Coll. Cardiol., June 6, 2006; 47(11): 2172 - 2179. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. W. Serruys, W. Wijns, G. Sianos, I. de Scheerder, P. A. van den Heuvel, W. Rutsch, H. D. Glogar, C. Macaya, P. H. Materne, S. Veldhof, et al. Direct stenting versus direct stenting followed by centered beta-radiation with intravascular ultrasound-guided dosimetry and long-term anti-platelet treatment: Results of a randomized trial: Beta-radiation Investigation with Direct stenting and Galileo in Europe (BRIDGE) J. Am. Coll. Cardiol., August 4, 2004; 44(3): 528 - 537. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Tanabe, P. W. Serruys, M. Degertekin, G. Guagliumi, E. Grube, C. Chan, T. Munzel, J. Belardi, W. Ruzyllo, L. Bilodeau, et al. Chronic Arterial Responses to Polymer-Controlled Paclitaxel-Eluting Stents: Comparison With Bare Metal Stents by Serial Intravascular Ultrasound Analyses: Data From the Randomized TAXUS-II Trial Circulation, January 20, 2004; 109(2): 196 - 200. [Abstract] [Full Text] [PDF] |
||||
![]() |
P.W. Serruys, G. Sianos, W. van der Giessen, H.J.R.M. Bonnier, P. Urban, W. Wijns, E. Benit, M. Vandormael, R. Dorr, C. Disco, et al. Intracoronary {beta}-radiation to reduce restenosis after balloon angioplasty and stenting. The Beta Radiation In Europe (BRIE) study Eur. Heart J., September 1, 2002; 23(17): 1351 - 1359. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Hehrlein, J. J. DeVries, A. Arab, S. D. Haller, K. Schloesser, F. O. Tio, and T. A. Fischell Failure of a Novel Balloon-Expandable {gamma}-Emitting (103Pd) Stent to Prevent Edge Effects Circulation, November 6, 2001; 104(19): 2358 - 2362. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Sianos, I. P. Kay, M. A. Costa, E. Regar, K. Kozuma, P. J. de Feyter, E. Boersma, C. Disco, and P. W. Serruys Geographical miss during catheter-based intracoronary beta-radiation: incidence and implications in the BRIE study J. Am. Coll. Cardiol., August 1, 2001; 38(2): 415 - 420. [Abstract] [Full Text] [PDF] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2000 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |