From the Department of Medicine (Cardiac Catheterization Laboratory and Coronary Care Unit, Brigham and Womens Hospital), Harvard Medical School, Boston, Mass (J.M.G., E.R.E., C.R.); the Harvard-MIT Division of Health Sciences and Technology, Massachusetts Institute of Technology, Cambridge, Mass (E.R.E., J.C.S., P.S., C.R.); and Medtronic/Arterial Vascular Engineering, Santa Rosa, Calif (M.S.W.).
Correspondence to Joseph Garasic, MD, Brigham and Womens Hospital, 75 Francis St, Boston, MA 02115. E-mail jmgarasic{at}bics.bwh.harvard.edu
BackgroundClinical trials show that larger immediate postdeployment stent diameters provide greater ultimate luminal size, whereas animal data show that arterial injury and stent design determine late neointimal thickening. At deployment, a stent stretches a vessel, imposing a cross-sectional polygonal luminal shape that depends on the stent design, with each strut serving as a vertex. We asked whether this design-dependent postdeployment luminal geometry affects late neointimal thickening independently of the extent of strut-induced injury.
Methods and ResultsStainless steel stents of 3 different configurations were implanted in rabbit iliac arteries for 3 or 28 days. Stents designed with 12 struts per cross section had 50% to 60% less mural thrombus and 2-fold less neointimal area than identical stents with only 8 struts per cross section. Sequential histological sectioning of individual stents showed that immediate postdeployment luminal geometry and subsequent neointimal area varied along the course of each stent subunit. Mathematical modeling of the shape imposed by the stent on the artery predicted late neointimal area, based on the re-creation of a circular vessel lumen within the confines of the initial stent-imposed polygonal luminal shape.
ConclusionsImmediate postdeployment luminal geometry, dictated by stent design, determines neointimal thickness independently of arterial injury and may be useful for predicting patterns of intimal growth for novel stent designs.
Key Words: stents restenosis angioplasty pathology
This article has been cited by other articles:
![]() |
R. S. Schwartz, E. Edelman, R. Virmani, A. Carter, J. F. Granada, G. L. Kaluza, N. A.F. Chronos, K. A. Robinson, R. Waksman, J. Weinberger, et al. Drug-Eluting Stents in Preclinical Studies: Updated Consensus Recommendations for Preclinical Evaluation Circ Cardiovasc Interv, October 1, 2008; 1(2): 143 - 153. [Abstract] [Full Text] [PDF] |
||||
![]() |
C J Dean, A C Morton, N D Arnold, D R Hose, D C Crossman, and J Gunn Relative importance of the components of stent geometry to stretch induced in-stent neointima formation Heart, December 1, 2005; 91(12): 1603 - 1604. [Full Text] [PDF] |
||||
![]() |
S. Windecker, R. Simon, M. Lins, V. Klauss, F. R. Eberli, M. Roffi, G. Pedrazzini, T. Moccetti, P. Wenaweser, M. Togni, et al. Randomized Comparison of a Titanium-Nitride-Oxide-Coated Stent With a Stainless Steel Stent for Coronary Revascularization: The TiNOX Trial Circulation, May 24, 2005; 111(20): 2617 - 2622. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. F. LaDisa Jr., Lars. E. Olson, R. C. Molthen, D. A. Hettrick, P. F. Pratt, M. D. Hardel, J. R. Kersten, D. C. Warltier, and P. S. Pagel Alterations in wall shear stress predict sites of neointimal hyperplasia after stent implantation in rabbit iliac arteries Am J Physiol Heart Circ Physiol, May 1, 2005; 288(5): H2465 - H2475. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. F. LaDisa Jr., L. E. Olson, I. Guler, D. A. Hettrick, J. R. Kersten, D. C. Warltier, and P. S. Pagel Circumferential vascular deformation after stent implantation alters wall shear stress evaluated with time-dependent 3D computational fluid dynamics models J Appl Physiol, March 1, 2005; 98(3): 947 - 957. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. F. LaDisa Jr., L. E. Olson, I. Guler, D. A. Hettrick, S. H. Audi, J. R. Kersten, D. C. Warltier, and P. S. Pagel Stent design properties and deployment ratio influence indexes of wall shear stress: a three-dimensional computational fluid dynamics investigation within a normal artery J Appl Physiol, July 1, 2004; 97(1): 424 - 430. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. F. LaDisa Jr., D. A. Hettrick, L. E. Olson, I. Guler, E. R. Gross, T. T. Kress, J. R. Kersten, D. C. Warltier, and P. S. Pagel Stent implantation alters coronary artery hemodynamics and wall shear stress during maximal vasodilation J Appl Physiol, December 1, 2002; 93(6): 1939 - 1946. [Abstract] [Full Text] [PDF] |
||||
![]() |
J Gunn, N Arnold, K H Chan, L Shepherd, D C Cumberland, and D C Crossman Coronary artery stretch versus deep injury in the development of in-stent neointima Heart, October 1, 2002; 88(4): 401 - 405. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. J Alp, N. E.J West, N. Arnold, J. Gunn, A. P Banning, and K. M Channon Increased intimal hyperplasia in experimental vein graft stenting compared to arterial stenting: comparisons in a new rabbit model of stent injury Cardiovasc Res, October 1, 2002; 56(1): 164 - 172. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. S. Schwartz, E. R. Edelman, A. Carter, N. Chronos, C. Rogers, K. A. Robinson, R. Waksman, J. Weinberger, R. L. Wilensky, D. N. Jensen, et al. Drug-Eluting Stents in Preclinical Studies: Recommended Evaluation From a Consensus Group Circulation, October 1, 2002; 106(14): 1867 - 1873. [Full Text] [PDF] |
||||
![]() |
C. Briguori, C. Sarais, P. Pagnotta, F. Liistro, M. Montorfano, A. Chieffo, F. Sgura, N. Corvaja, R. Albiero, G. Stankovic, et al. In-stent restenosis in small coronary arteries: Impact of strut thickness J. Am. Coll. Cardiol., August 7, 2002; 40(3): 403 - 409. [Abstract] [Full Text] [PDF] |
||||
![]() |
C.-W. Hwang and E. R. Edelman Arterial Ultrastructure Influences Transport of Locally Delivered Drugs Circ. Res., April 19, 2002; 90(7): 826 - 832. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Windecker and B. Meier All stents are not alike or is the difference in the eye of the observer only? Eur. Heart J., November 1, 2001; 22(21): 1973 - 1977. [PDF] |
||||
![]() |
R. Hoffmann, C. Jansen, A. Konig, P.K. Haager, G. Kerckhoff, J. Vom Dahl, V. Klauss, P. Hanrath, and H. Mudra Stent design related neointimal tissue proliferation in human coronary arteries; an intravascular ultrasound study Eur. Heart J., November 1, 2001; 22(21): 2007 - 2014. [Abstract] [PDF] |
||||
![]() |
P. Tozzi, J. O. Solem, D. Boumzebra, A. Mucciolo, C. Y. Genton, P. Chaubert, and L. K. von Segesser Is the GraftConnector a valid alternative to running suture in end-to-side coronary arteries anastomoses? Ann. Thorac. Surg., September 1, 2001; 72(3): S999 - 1003. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Kastrati, J. Mehilli, J. Dirschinger, F. Dotzer, H. Schuhlen, F.-J. Neumann, M. Fleckenstein, C. Pfafferott, M. Seyfarth, and A. Schomig Intracoronary Stenting and Angiographic Results : Strut Thickness Effect on Restenosis Outcome (ISAR-STEREO) Trial Circulation, June 12, 2001; 103(23): 2816 - 2821. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. R. Edelman, P. Seifert, A. Groothuis, A. Morss, D. Bornstein, and C. Rogers Gold-Coated NIR Stents in Porcine Coronary Arteries Circulation, January 23, 2001; 103(3): 429 - 434. [Abstract] [Full Text] [PDF] |
||||
![]() |
C.-W. Hwang and E. R. Edelman Arterial Ultrastructure Influences Transport of Locally Delivered Drugs Circ. Res., April 19, 2002; 90(7): 826 - 832. [Abstract] [Full Text] [PDF] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2000 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |