Submitted on January 11, 2002
From the Cardiology Clinic (C.H., S.A.,
A.M.) and Radiation Oncology (D.B., M.K.) at St Josef Hospital, Ruhr
University, Bochum, Germany, and the Department of Pathology (P.G.,
K.M.M.) at Bergmannsheil-Kliniken, Ruhr-University, Bochum,
Germany. * To whom correspondence should be addressed. E-mail: andreas.muegge{at}ruhr-uni-bochum.de.
BackgroundIntravascular brachytherapy is being applied more and more in patients with coronary artery disease for the prevention of restenosis subsequent to balloon angioplasty, in particular after stent implantation. Several radiation sources (ß- and Methods and ResultsPostmortem calcified human coronary artery segments were stented and placed in an organ bath. Commercially available ConclusionsDose perturbation caused by calcified plaque and metallic stents is significant for ß-sources. This dosimetric difference between ß- and
Revised on March 22, 2002
Accepted on March 22, 2002
Dosimetric Measurements in Isolated Human Coronary Arteries. Comparison of Commercially Available
Iridium192 With
Strontium/Yttrium90
Emitters
Christoph Hanefeld MD,
-emitters) are available in clinical routine. It was the purpose of this study to compare the radiation doses at the level of the adventitia in diseased and stented human coronary arteries for 192Ir and 90Sr/Y emitters in routine use. In contrast to previously published work, we performed dosimetry instead of calculating depth-dose distribution by use of the Monte Carlo system.
-emitters (192Ir; Cordis Checkmate) and ß-emitters (90Sr/Y; Novoste Beta-Cath) were used. Relative dose distributions along the adventitia were measured by a specially designed scintillation detector system. Whereas dose perturbations caused by stents and calcified plaque were negligible for the 192Ir source, radiation from the beta source was significantly impaired (as much as 40%) at the level of the adventitia (3.0-mm vessel diameter). Dose perturbation was clearly dependent on the extent and severity of calcification, less affected by stent material.
-emitters in diseased coronary arteries should be considered when calculating doses in intravascular brachytherapy.
This article has been cited by other articles:
![]() |
P. G. Guerra, M. Talajic, B. Thibault, M. Dubuc, D. Roy, L. Macle, T. K. Leung, A. Arsenault, F. Harel, J. Gregoire, et al. {beta}-Radiation for the Creation of Linear Lesions in the Canine Atrium Circulation, August 24, 2004; 110(8): 911 - 914. [Abstract] [Full Text] [PDF] |
||||
![]() |
P H Grewe, T Deneke, C Hanefeld, and K-M Muller Human coronary morphology after {beta} radiation brachytherapy of in-stent restenosis Heart, June 1, 2004; 90(6): e32 - e32. [Abstract] [Full Text] [PDF] |
||||
|