Treatment of Arrhythmias: ICD/Pacemaker-Excluding CRT I

Abstract 662: Electromagnetic Interference (EMI) of Implanted Cardiac Devices by MP3 Player Headphones

Sinjin Lee1; Benjamin Ransford2; Kevin Fu2; Kohno Tadayoshi3; William H Maisel4

1 Beth Israel Deaconess Med Cntr, Boston, MA
2 Univ of Massachusetts, Amherst, MA
3 Univ of Washington, Seattle, WA
4 Beth Israel Deaconess Med Cntr, Boston, MA

Background: Potential MP3 player (iPod) interactions with pacemakers (PM) and implantable defibrillators (ICD) have received much attention. Yet, little is known about the effects of MP3 player headphones, most of which contain the magnetic substance, neodymium. A magnetic field strength of ≥ 5–10 gauss at the PM or ICD has the potential to interact with the device.

Methods: PM or ICD function was assessed in 60 pts during exposure to various headphones and iPods. Pts were monitored continuously via ECG and the device programmer. Abnormal device behavior was defined as oversensing, undersensing, inappropriate pacing, and reprogramming. Headphone magnetic field strength was also measured in vitro.

Results: No significant EMI from MP3 players was noted. However, device interference by headphones was documented in 14/60 (23%) pts (TABLE). Inappropriate (asynchronous) pacing was observed in 4/27 (15%) PM pts while inhibition of ICD detection occurred in 10/33 (30%) ICD pts. One PM pt experienced device reprogramming (1/27, 4%). Headphones with a measured magnetic field strength of ≥10 gauss at 2 cm were more likely to interact than those with lower magnetic strength (17/60, 28% vs. 1/60, 2%, P<0.0001). EMI was not observed when headphones were placed ≥3 cm from the skin surface.

Conclusion: Clinically significant EMI can occur when headphones are placed in close proximity to implanted PMs and ICDs. In vitro measurements of headphone magnetic strength at 2 cm accurately predict clinical interaction with implanted devices. The magnetic strength of headphones varies depending on the model and design. Device pts should be advised to keep headphones at least 3 cm from their device.


MP3 Player Headphone Clinical Interactions with PM-ICD

Table 1





This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Lee, S.
Right arrow Articles by Maisel, W. H
PubMed
Right arrow Articles by Lee, S.
Right arrow Articles by Maisel, W. H