(Circulation. 1999;99:2815-2818.)
© 1999 American Heart Association, Inc.
Basic Science Reports |
From the Department of Cardiothoracic Surgery, Allegheny General Hospital, and Cardiothoracic Surgery Research, Cardiovascular and Pulmonary Research Institute, Pittsburgh, Pa.
Correspondence to Dennis R. Trumble, MS, Allegheny General Hospital, Cardiothoracic Surgery Research, 9th Floor, South Tower, 320 E North Ave, Pittsburgh, PA 15212. E-mail trumble{at}pgh.auhs.edu
| Abstract |
|---|
|
|
|---|
Methods and ResultsInitial in vivo tests were performed on 6 pigs (weight, 41±4 kg). RV wall motion and stroke volume were monitored via transesophageal echocardiography. Acute RV failure was created by graded right coronary ligation, which yielded a 63% reduction in RV stroke volume (39.9±8.2 to 14.7±1.9 mL; P<0.002). We secured the balloon over the RV free wall by attaching it to the edges of the opened pericardium. The sternum was then reapproximated, and data were collected with the device off and on (every beat). Device placement had no deleterious effect on RV function. Balloon activation returned RV stroke volumes to normal (37.8±9.2 mL) and increased mean pulmonary artery pressures from 13±2 to 16±3 mm Hg (P<0.01). RV compression did not induce or exacerbate tricuspid regurgitation. Mean aortic pressure improved from postinfarction levels but did not return to normal.
ConclusionsWe conclude that the pulmonary circulation can be supported in the short term via cardiac compression and that balloon copulsation techniques for short-term RV failure should be tested in long-term models.
Key Words: heart-assist device hemodynamics balloon heart failure ventricles
| Introduction |
|---|
|
|
|---|
When pharmacological agents are unable to improve RV function, surgeons must rely on mechanical means to restore blood flow to the pulmonary circulation and LV. Current options for mechanical assistance include centrifugal pumps, positive displacement pumps, and right-sided diastolic counterpulsation. The drawback common to all these mechanisms is that they require invasive, time-consuming procedures to secure requisite cannulae and/or anastomoses to the pulmonary artery, unlike the intra-aortic balloon pump, which can be positioned and retracted through a femoral artery cutdown. Hence, development of a quick and simple means to support the pulmonary circulation during periods of transient RV dysfunction is needed.
This report describes preliminary trials of an RV copulsation balloon (RVCB) designed to effect right-heart assistance via direct cardiac compression. The primary purpose of this study is to determine whether significant RV support can be achieved via pneumatic actuation of a balloon placed between the sternum and the RV. The ultimate objective is to develop a simplified RV assist device that can be inserted and removed as readily as an intra-aortic balloon pump.
| Methods |
|---|
|
|
|---|
|
Unstressed, the balloon measures roughly 5 cm in diameter and 2 cm in height and occupies a volume of 40 mL. Because of its high compliance, however, balloon volumes can be readily expanded to twice this value without damage. Two lateral convolutions allow the balloon to lay flat under negative pressures, provide a smooth surface for RV compression, and maximize active diastolic filling.
Pneumatic Drive System
Device actuation is controlled via a commercial IABP pump
console (System 83, Datascope Corp) modified to deliver a predetermined
volume of helium gas to the balloon during cardiac systole. To meet RV
copulsation requirements, a second safety chamber was connected in
parallel with the original (Figure 2
) to
double driver volume capacity to 65 mL. Cardiac synchronization is
achieved by setting the driver inflation delay switch to the "out"
position, thereby disabling the automatic delay function and allowing
balloon inflation to occur immediately after detection of the QRS
complex. Apart from these 2 alterations, drive-console operating
conditions are identical for both RV copulsation and aortic
counterpulsation.
|
Experimental Procedure
The operative procedures described below were performed in
compliance with the "Guide for the Care and Use of Laboratory
Animals" prepared by the National Academy of Sciences and published
by the National Institutes of Health (NIH publication 86-23, revised
1985). In addition, this project was approved by the
Allegheny-Singer Institutional Animal Care and Use Committee.
Six female pigs (weight, 35 to 45 kg) were used to determine the feasibility of supporting RV function via pneumatic actuation of a balloon placed between the RV and the sternum. Each animal was sedated with intramuscular injections of ketamine (615 mg) and xylazine hydrochloride (80 mg) before induction of general anesthesia. Anesthesia was maintained with 1% to 2% isoflurane delivered through an endotracheal tube. A Swan-Ganz catheter and arterial pressure line were placed via the external jugular vein and carotid artery, respectively. The chest was entered through a median sternotomy, and 3 additional pressure lines were placed directly into the left atrium and both ventricles. RV stroke volume was calculated based on Doppler velocity profiles measured across the pulmonary valve via transesophageal echocardiography (TEE).
Acute RV failure was induced by graded right coronary artery ligation. RV dysfunction was considered significant when a 50% reduction in stroke volume was observed. Bretylium tosylate (50 mg), procainamide hydrochloride (600 mg), and lidocaine (160 mg/h) were administered to prevent arrhythmias. The RVCB was subsequently placed over the heart so that the (empty) balloon was sandwiched between the RV free wall and Teflon backing. The device was secured in place by suturing the Teflon backing to the edges of the opened pericardium. The sternum was then reapproximated in preparation for RVCB activation. Balloon inflation was timed to begin on detection of the QRS complex, whereas the deflation point was adjusted manually to coincide with cardiac diastole (based on RV and balloon pressure waveforms).
Hemodynamic data were collected under 4 conditions: before RV infarction (baseline); after RV infarction and before balloon placement (postinfarction); after balloon placement with the device inactive (balloon off); and after balloon placement with the device inflated with every heartbeat (balloon on). RVCB driveline pressures were also recorded during intervals of cardiac assist. RV copulsation was continued for periods up to 30 minutes. On completion of the study, all animals were euthanized with 20 mEq of KCl while they remained under general anesthesia.
Data Collection and Statistical Analysis
Data were collected under closed-chest conditions to mimic
physiological conditions anticipated during
clinical application of a catheter-based RVCB. Pressure and ECG
waveforms were digitized at a rate of 100 samples/s for periods of 1 to
2 minutes and stored in an IBM PS/2 personal computer (data acquisition
package: CODAS, Dataq Instruments). Data sets were collected on
establishment of steady-state hemodynamics for each
condition tested (defined as stable arterial pressures
maintained for
30 seconds). These data were processed with XANALYZE,
a comprehensive cardiovascular waveform
analysis program.6 Data sets comprising 10
contiguous heartbeats were isolated and the waveforms averaged to
minimize spurious results. Repeated-measures ANOVA was performed (TRUE
EPISTAT, Epistat Services) to determine the significance of differences
between treatment groups. A 2-sided P value of <0.05 was
considered statistically significant. All summary data are expressed as
mean±SD.
| Results |
|---|
|
|
|---|
|
Impact of Device Placement on Cardiac Function
Placement of the RVCB over the RV free wall and
reapproximation of the sternum had no significant effect on cardiac
function, as evidenced by hemodynamic measurements
taken after infarction (before device insertion) and after insertion
with the balloon deflated (Table 1
). Furthermore, TEE revealed
no evidence of cardiac compression after device fixation.
Hemodynamic Response to Device Activation
RVCB activation produced marked increases in right-heart pressures
and flows, restoring these parameters to near-normal
(preinfarction) levels (Table 1
). RV and pulmonary
artery systolic pressures increased 69% and 56%,
respectively, over postinfarction baseline, whereas RV
diastolic pressures were lowered by 30% (Figure 3
). RV stroke volumes increased
dramatically from 14.7±1.9 to 37.8±9.2 mL (P<0.005), and
mean pulmonary artery pressures improved from 12.7±2.4 to
15.8±2.9 mm Hg (P<0.01). Systolic
arterial pressures were also markedly improved with balloon
activation, increasing from 66.3±11.6 to 78.8±8.7 mm Hg
(P<0.02), but did not return to preinfarction levels
(103.7±10.5 mm Hg). Heart rate remained unchanged from the
postinfarction baseline rate of 76 bpm, and no significant cardiac
arrhythmias were noted.
|
Substantial displacement of the RV free wall toward the
ventricular septum was observed via TEE during balloon
inflation. Doppler measurements across the pulmonary
outflow tract showed increased blood flow velocity during RV ejection
(Table 2
). Similar measurements made
across the tricuspid valve revealed that RV compression did not induce
or exacerbate regurgitation.
|
| Discussion |
|---|
|
|
|---|
To date, there are no reports in the literature regarding deployment of a copulsating balloon in the anterior mediastinum to assist RV function. Several groups, however, have proposed the use of dynamic cardiac compression as a possible means to aid the failing heart and have tested these techniques in animals.11 12 These studies suggest that external forces can be safely applied to the epicardial surface of the heart to improve contractile efficiency and cardiac output. Reports published by Anstadt et al13 14 describe a method of direct mechanical ventricular actuation (DMVA) that involves placement of a contoured cup over both ventricles. The cup is implanted through a left thoracotomy, is held in place by a slight vacuum introduced between the heart and rigid housing, and uses positive and negative pneumatic pressures to expand and deflate a flexible diaphragm that encircles the heart. DMVA has proven quite effective in that total cardiac support has been achieved during ventricular fibrillation for periods up to 7 days.15 Still, despite significant advantages over conventional assist schemes (eg, no cannulation or blood-contacting parts), this technique is limited to those cases in which extremely invasive measures and extensive surgical resources can be brought to bear.
The ultimate goal of this research is to develop an inexpensive, catheter-based RV assist device that can be deployed and retrieved with the ease of an intra-aortic balloon pump. This article describes our first step toward this objective: that is, a pilot study designed to determine whether significant RV support can be achieved via timed inflation and deflation of a balloon placed between the sternum and the heart.
Our data show that balloon copulsation can significantly increase pulmonary artery pressures and restore RV stroke volume to normal levels during periods of severe RV dysfunction. In these experiments, however, postinfarction bradycardia served to limit total cardiac output to 78% of baseline under assist conditions (versus 30% without balloon augmentation). The use of temporary atrioventricular sequential pacing may therefore be warranted in bradycardic patients to normalize heart rate and further improve cardiac output.
Benefits to LV function were also observed as a result of RV copulsation, most likely owing to increased blood flow to the LV combined with displacement of the septum toward the LV during balloon inflation. Placement of the balloon between the RV free wall and sternum had no significant impact on cardiac function, and gross examination yielded no evidence of damage to the epicardial surface after brief (30-minute) periods of support.
Experimental Limitations
The use of Doppler echocardiography to
calculate RV stroke volume can be considered a drawback owing to
potential inaccuracies caused by beam misalignment, errors in vessel
cross-sectional area estimates, and assumptions of laminar flow,
constant flow area, and uniform flow velocity across the
vessel.16 17 Baseline calculations nevertheless yielded
mean stroke volumes (40 mL) and cardiac outputs (3.65 L/min)
commensurate with values expected in 41-kg pigs, which suggests that
these measurements were indeed reasonably accurate. Moreover,
Doppler echocardiography is considered a very
reliable technique for assessing percent changes in stroke volume owing
to uniform measuring conditions within each experiment. Given the
uniformity and consistency of these pulmonary flow
measurements, it is reasonable to conclude that these stroke volume
calculations reflect actual changes due to RV ischemia and
copulsation assistance.
These data were collected under short-term circulatory assist conditions and therefore do not provide information regarding long-term use. Rather, this study was designed to determine the feasibility of RV balloon copulsation from a biomechanical perspective. Results from these preliminary trials indicate that this technique is a mechanically viable means of supporting the pulmonary circulation. Whether balloon copulsation can be successfully used to effect long-term RV support has yet to be determined.
Conclusions
We conclude that this method of cardiac copulsation is an
effective means of providing short-term RV assistance and that
additional studies are warranted. Future work will involve the
development and manufacture of a catheter-mounted copulsation balloon
and custom-designed drive unit suitable for human use. In vivo testing
will be performed in a model of short-term RV infarction and will mimic
the proposed clinical method of device placement. The duration of
support will be significantly extended, and
histological examinations of the RV will be performed
to determine whether prolonged cardiac compression leads to significant
myocardial damage. These trials, if successful, would confirm the
efficacy of minimally invasive RV copulsation and bring this technology
closer to clinical implementation.
| Acknowledgments |
|---|
Received December 16, 1998; revision received February 16, 1999; accepted February 16, 1999.
| References |
|---|
|
|
|---|
2. Romand JA, Donald FA, Suter PM. Acute right ventricular failure, pathophysiology and treatment. Monaldi Arch Chest Dis. 1995;50:129133.[Medline] [Order article via Infotrieve]
3.
Santamore WP, Gray LA Jr. Left ventricular
contributions to right ventricular systolic
function during LVAD support. Ann Thorac Surg. 1996;61:350356.
4.
Cooper GJ, Loisance DY, Miyama M, Abe Y, Deleuze H.
Direct mechanical assistance of the right ventricle with the hemopump
in a porcine model. Ann Thorac Surg. 1995;59:443447.
5.
Pavie A, Leger P. Physiology of
univentricular versus biventricular support.
Ann Thorac Surg. 1996;61:347349.
6. Stewart S, Liang I, Flack J, Clark R. A comprehensive cardiovascular waveform analysis program for IBM-compatible personal computers. Biomed Instrum Technol. 1992;26:3947.[Medline] [Order article via Infotrieve]
7. Doyle AR, Dhir AK, Moors AH, Latimer AR. Treatment of perioperative low cardiac output syndrome. Ann Thorac Surg. 1995;59:S3S11.
8. Rowles JR, Mortimer BJ, Olsen DB. Ventricular assist and total artificial heart devices for clinical use in 1993. ASAIO J. 1993;39:840855.[Medline] [Order article via Infotrieve]
9. Argenziano M, Oz MC, Rose EA. The continuing evolution of mechanical ventricular assistance. Curr Probl Sur. 1997;34:321385.
10. Votapka TV, Pennington DG. Mechanical circulatory support of the failing heart. Curr Opin Cardiol. 1994;9:231236.[Medline] [Order article via Infotrieve]
11.
Kawaguchi O, Goto Y, Ohgoshi Y, Yaku H, Murase M, Suga
H. Dynamic cardiac compression improves contractile efficiency of the
heart. J Thorac Cardiovasc Surg. 1997;113:923931.
12. Mitsumaru A, Yozu R, Inoue Y, Tanaka S, Yoshito H, Kanda K, Tsutsui Y, Tsutsui N, Kawada S. Experimental study of combination of extraaortic balloon counterpulsation and ventricular assist cup to acute heart failure in dogs. ASAIO J. 1997;43:187192.[Medline] [Order article via Infotrieve]
13. Anstadt MP, Anstadt GL, Lowe JE. Direct mechanical ventricular actuation: a review. Resuscitation. 1991;21:723.[Medline] [Order article via Infotrieve]
14. Anstadt MP, Perez-Tamayo RA, Banit DM, Walthall HP, Cothran RL Jr, Abdel-Aleem S, Anstadt GL, Jones PL, Lowe JE. Myocardial tolerance to mechanical actuation is affected by biomaterial characteristics. ASAIO J. 1994;40:M329M334.[Medline] [Order article via Infotrieve]
15. Perez-Tamayo RA, Anstadt MP, Cothran RL Jr, Reisinger RJ, Schenkman DI, Hulette C, Reimer KA, Anstadt GL, Lowe JE. Prolonged total circulatory support using direct mechanical ventricular actuation. ASAIO J. 1995;41:M512M517.[Medline] [Order article via Infotrieve]
16. Goldberg SJ, Allen HD, Marx GR, Donnerstein RL. Flow computation. In: Doppler Echocardiography. 2nd ed. Philadelphia, Pa: Lea & Febiger; 1988:153186.
17. Feigenbaum H. Hemodynamic information derived from echocardiography. In: Echocardiography. 5th ed. Philadelphia, Pa: Lea & Febiger; 1994:181215.
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|