From the Cardiovascular Section/Department of Medicine, University of
Oklahoma Health Sciences Center and the Department of Veterans Affairs Medical
Center, Oklahoma City; and the Department of Cardiology, Heart Lung Institute,
University Hospital Utrecht, Netherlands (F.H.M.W.).
Correspondence to Hiroshi Nakagawa, MD, PhD, Department of Medicine/Cardiovascular Section, University of Oklahoma Health Sciences Center, 920 S.L. Young Blvd, WP3120, Oklahoma City, OK 73104. E-mail hiroshi-nakagawa{at}ouhsc.edu
Methods and ResultsIn 11 anesthetized dogs, the thigh
muscle was exposed and bathed with heparinized canine blood. A 7F
ablation catheter with a 2- or 5-mm irrigated tip electrode was
positioned perpendicular or parallel to the thigh muscle.
Radiofrequency current was delivered at constant voltage (50 V) for 30
seconds during saline irrigation (20 mL/min) to 148 sites. Tissue
temperature at depths of 3.5 and 7 mm and lesion size were
measured. In the perpendicular electrode-tissue orientation,
radiofrequency applications at 50 V with the 2-mm electrode compared
with the 5-mm electrode resulted in lower power at 50 V (26 versus 36
W) but higher tissue temperatures, larger lesion depth (8.0 versus
5.4 mm), and greater diameter (12.4 mm versus 8.4 mm).
Also, in the parallel orientation, overall power was lower with the
2-mm electrode (25 versus 33 W), but tissue temperatures were higher
and lesions were deeper (7.3 versus 6.9 mm). Lesion diameter was
similar (11.1 versus 11.3 mm) for both electrodes.
ConclusionsThe smaller electrode resulted in transmission of a
greater fraction of the radiofrequency power to the tissue and resulted
in higher tissue temperature, larger lesions, and lower dependency of
lesion size on the electrode orientation.
This study tested the hypothesis that because convective cooling by the
blood stream is not required with a saline-irrigated electrode, a
smaller electrode can be used without sacrificing ablation efficacy.
Potential advantages of a smaller electrode include higher electrogram
resolution (improving mapping accuracy to increase ablation efficacy
and decrease the number of RF
applications),37 38 39 increased ablation catheter
flexibility and mobility, and the ability to design small ablation
electrodes for use in small children and in small anatomic spaces, such
as in the coronary veins. To test the hypothesis, we compared
in vivo tissue temperature and lesion size between a 2-mm and a 5-mm
ablation electrode oriented perpendicular and parallel to the tissue
interface, respectively, during RF ablation with saline irrigation in a
canine thigh muscle preparation.31
The 2- or 5-mm tip electrode was positioned perpendicular or parallel
to the thigh muscle in different experiments (Figure 2
The ablation electrodes were irrigated through the catheter lumen with
room-temperature (20°C to 22°C), heparinized (2 U/mL) normal saline
at 20 mL/min. Saline irrigation was started 3 to 5 seconds before the
onset of RF application and was maintained until 5 seconds after the
completion of RF delivery. RF energy (500 kHz) was delivered at
constant voltage (Radionics, model 3D-J) between the ablation electrode
and an adhesive electrosurgical dispersive pad (skin patch) applied to
the shaven skin of the opposite thigh. During each RF application,
current and impedance were monitored and recorded along with the
electrode and tissue temperatures.
Ablation Protocol
After Ablation
Statistical Analysis
Group 1: Perpendicular Electrode-Tissue Orientation
Group 2: Parallel Electrode-Tissue Orientation
Body Impedance
Heating Efficacy
In the parallel orientation, the tissue voltage is still smaller with
the 5-mm electrode (30.3 versus 35.1 V with the 5-mm and 2-mm
electrodes, Figure 8
Because of the difference between the length and diameter in the 5-mm
electrode, the contact area differed greatly between the perpendicular
and parallel orientations, confirmed by the insulated electrode
impedance measurements with deionized water. In the 2-mm electrode,
electrode length and diameter are similar, making tissue temperatures
and lesion sizes relatively independent of the electrode-tissue
orientation.
Constant-Voltage Versus Constant-Power Delivery
Tissue Superheating
Study Limitations
Because the intramural blood flow may be less in the resting thigh
muscle than in the myocardium, the thigh muscle may provide
less of a heat sink and therefore may result in greater lesion size.
However, a previous study suggested that the tissue temperature profile
during RF ablation may be independent of intramyocardial
perfusion.10 In addition, the lesion size in the
present study is similar to lesions described in other studies in
which RF current was delivered to the ventricular
myocardium of a beating canine
heart.28 29 32 33 36 More importantly, the degree
of difference in lesion size between the 2- and 5-mm electrodes would
be expected to be similar in the human heart, because the same basic
electrical principles apply.
The 2- and 5-mm electrodes were compared in the perpendicular
orientation in group 1 dogs and in the parallel orientation in group 2
dogs. Because perpendicular and parallel orientations were studied in 2
different groups of dogs, the tissue temperatures and lesion sizes for
the 2 orientations were compared only qualitatively and not
statistically. However, body weight (23.8±3.0 versus 23.2±2.5 kg) and
body impedance (26±2 versus 26±1
Clinical Implications
The use of small electrodes will require a significant reduction in RF
power, because of a reduction in the amount of RF power lost to the
blood pool and thus a greater percentage of the overall power delivered
to the tissue. When required, lesion size and depth can be increased
without tissue superheating and the occurrence of a steam pop either by
a prolonged RF application at lower power43 or,
preferably, by higher RF power delivered in the pulsed mode, in which
the off periods prevent tissue superheating in the hottest, shallow
sites.47
Another potential advantage of a small irrigated electrode would be the
ability to make a small, discrete, accurately placed RF lesion with a
very low RF power application, such as for ablation of an anteroseptal
or midseptal accessory pathway or ablation in a small child.
Received October 28, 1997;
revision received February 2, 1998;
accepted February 13, 1998.
2.
Borggrefe M, Budde T, Podczeck A, Breithardt G. High
frequency alternating current ablation of an accessory pathway in
humans. J Am Coll Cardiol. 1987;10:576582.[Abstract]
3.
Jackman WM, Kuck KH, Naccarelli GV, Carmen L, Pitha J.
Radiofrequency current directed across the mitral anulus with a bipolar
epicardial-endocardial catheter electrode configuration in dogs.
Circulation. 1988;78:12881298.
4.
Wittkampf FHM, Hauer RNW, Robles de Medina EO.
Control of radiofrequency lesion size by power regulation.
Circulation. 1989;80:962968.
5.
Ring ME, Huang SKS, Gorman G, Graham AR.
Determinants of impedance rise during catheter ablation of bovine
myocardium with radiofrequency energy. Pacing Clin
Electrophysiol. 1989;12:170176.[Medline]
[Order article via Infotrieve]
6.
Haverkamp W, Hindricks G, Gulker H, Rissel U,
Pfennings W, Borggrefe M, Breithardt G. Coagulation of
ventricular myocardium using radiofrequency
alternating current: biophysical aspects and experimental findings.
Pacing Clin Electrophysiol. 1989;12:187195.[Medline]
[Order article via Infotrieve]
7.
Haines DE, Verow AF. Observation on electrode-tissue
interface temperature and effect on electrical impedance during
radiofrequency ablation of ventricular
myocardium. Circulation. 1990;82:10341038.
8.
Langberg JJ, Calkins H, El-Atassi R, Borganelli M,
Leon A, Kalbfleisch SJ, Morady F. Temperature monitoring during
radiofrequency catheter ablation of accessory pathways.
Circulation. 1992;86:14691474.
9.
Haines DE, Watson DD, Verow AF. Electrode radius
predicts lesion radius during radiofrequency energy heating: validation
of a proposed thermodynamic model. Circ Res. 1990;67:124129.
10.
Haines DE, Watson DD. Tissue heating during
radiofrequency catheter ablation: a thermodynamic model and
observations in isolated perfused and superfused canine right
ventricular free wall. Pacing Clin
Electrophysiol. 1989;12:962976.[Medline]
[Order article via Infotrieve]
11.
Jackman WM, Wang X, Friday KJ, Fitzgerald DM, Roman C,
Moultion K, Margolis PD, Bowman AJ, Kuck KH, Naccarelli GV, Pitha JV,
Dyer J, Lazzara R. Catheter ablation of
atrioventricular junction using radiofrequency current
in 17 patients: comparison of standard and large-tip catheter
electrodes. Circulation. 1991;83:15621576.
12.
Langberg JJ, Lee MA, Chin MC, Rosenqvist M.
Radiofrequency catheter ablation: the effects of electrode size on
lesion volume in vivo. Pacing Clin Electrophysiol. 1990;13:12421248.[Medline]
[Order article via Infotrieve]
13.
Jackman WM, Wang X, Friday KJ, Roman C, Moulton KP,
Beckman KJ, McClelland JH, Twidale N, Hazlitt HA, Prior MI, Margolis
PD, Calame JD, Overholt ED, Lazzara R. Catheter ablation of accessory
AV pathways (Wolff-Parkinson-White syndrome) by radiofrequency current.
N Engl J Med. 1991;324:16051611.[Abstract]
14.
Calkins H, Sousa J, El-Atassi R, Rosenheck S, de
Buitleir M, Kou WH, Kadish AH, Langberg JJ, Morady F. Diagnosis and
cure of the Wolff-Parkinson-White syndrome or paroxysmal
supraventricular tachycardias during a single
electrophysiological test. N
Engl J Med. 1991;324:16121618.[Abstract]
15.
Kuck KH, Schluter M, Geiger M, Siebels J, Duckeck W.
Radiofrequency current catheter ablation of accessory
atrioventricular pathways. Lancet. 1991;33:15571561.
16.
Haissaguerre M, Dartigues JF, Warin JF, Le Metayer P,
Montserrat P, Salamon R. Electrogram patterns predictive of successful
catheter ablation of accessory pathway. Circulation. 1991;84:188202.
17.
Lesh MD, Van Hare GF, Schamp DJ, Chein W, Lee M,
Griffin JC, Langberg JJ, Cohen TJ, Lurie KG, Scheinman MM. Curative
percutaneous catheter ablation using radiofrequency
energy for accessory pathways in all locations: results in 100
consecutive patients. J Am Coll Cardiol. 1992;19:13031309.[Abstract]
18.
Jackman WM, Beckman KJ, McClelland JH, Wang X, Friday
KJ, Roman CA, Moulton KP, Twidale N, Hazlitt HA, Prior MI, Oren J,
Overholt ED, Lazzara R. Treatment of supraventricular
tachycardia due to atrioventricular nodal
reentry by radiofrequency catheter ablation of slow-pathway conduction.
N Engl J Med. 1992;327:313318.[Abstract]
19.
Kay GN, Epstein AE, Dailey SD, Plumb VJ. Selective
radiofrequency ablation of the slow pathway for the treatment of
atrioventricular nodal reentrant
tachycardia: evidence for involvement of perinodal
myocardium within the reentrant circuit.
Circulation. 1992;85:16751688.
20.
Jazayeri MR, Hempe SL, Sra JS, Dhala AA, Blanck Z,
Deshpande SS, Avitall B, Krum DP, Gilbert CJ, Akhtar M. Selective
transcatheter ablation of the fast and slow pathways using
radiofrequency energy in patients with atrioventricular
nodal reentrant tachycardia. Circulation. 1992;85:13181328.
21.
Klein LS, Shih H, Hackett FK, Zipes DP, Miles WM.
Radiofrequency catheter ablation of ventricular
tachycardia in patients without structural heart disease.
Circulation. 1992;85:16661674.
22.
Nakagawa H, Beckman KJ, McClelland JH, Wang X, Arruda
M, Santoro I, Hazlitt A, Abdalla I, Singh A, Gossinger H, Sweidan R,
Hiaro K, Widman L, Pitha JV, Lazzara R, Jackman WM. Radiofrequency
catheter ablation of idiopathic left ventricular
tachycardia guided by a Purkinje potential.
Circulation. 1993;88:26072617.
23.
Morady F, Harvey M, Kalbfleisch SJ, El-Atassi R,
Calkins H, Langberg JJ. Radiofrequency catheter ablation of
ventricular tachycardia in patients with
coronary artery disease. Circulation. 1993;87:363372.
24.
Stevenson WG, Khan H, Sager P, Saxon LA, Middlekauff
HR, Natterson PD, Wiener I. Identification of reentry circuit sites
during catheter mapping and radiofrequency ablation of
ventricular tachycardia late after myocardial
infarction. Circulation. 1993;88:16471670.
25.
Wilber DJ, Kopp DE, Glascock DN, Kinder CA, Kall JG.
Catheter ablation of the mitral isthmus for ventricular
tachycardia associated with inferior
infarction. Circulation. 1995;92:34813489.
26.
Langberg JJ, Gallagher M, Strickberger SA, Amirana O.
Temperature-guided radiofrequency catheter ablation with very large
distal electrode. Circulation. 1993;88:245249.
27.
Nath S, Dimarco JP, Haines DE. Basic aspects of
radiofrequency catheter ablation. J Cardiovasc
Electrophysiol. 1994;5:863876.[Medline]
[Order article via Infotrieve]
28.
Wittkampf FHM, Hauer RN, Robles de Medina EO.
Radiofrequency ablation with a cooled porus electrode catheter.
J Am Coll Cardiol. 1988;11:17. Abstract.
29.
Budde T, Borggrefe M, Podczeck A, Jacob B, Langwasser
J, Frenzel H, Breithardt G. Radiofrequency ablation: an improvement of
ablation techniques in comparison to direct-current delivery? In:
Breithardt G, Borggrefe M, Zipes DP, eds. Nonpharmacological
Therapy of Tachyarrhythmias. Mount Kisco, NY:
Futura Publishing Co; 1987:221241.
30.
Huang SKS, Cuenoud H, Tan-de-Guzman W. Increase
in lesion size and decrease in rise with a saline infusion electrode
catheter for radiofrequency catheter ablation. Circulation.
1989;80(suppl II):II-324. Abstract.
31.
Nakagawa H, Yamanashi SW, Pitha JV, Arruda M,
Wang X, Ohtomo K, Beckman KJ, McClelland JH, Lazzara R, Jackman WM.
Comparison of in vivo tissue temperature profile and lesion geometry
for radiofrequency ablation with a saline-irrigated electrode versus
temperature control in a canine thigh muscle preparation.
Circulation. 1995;91:22642273.
32.
Nakagawa H, Yamanashi WS, Pitha JV, Yong KC, Arruda M,
Rome M, Wang X, Ohtomo K, Lazzara R, Jackman WM. Comparison of
radiofrequency lesions in the canine left ventricle using a saline
irrigated electrode versus temperature control. J Am Coll
Cardiol. 1995;26(special issue):42A. Abstract.
33.
Johnson SB, Packer DJ. Characterization of the
tissue sequelae of the impedance rise occurring with ablation with an
open, irrigation tip catheter. Circulation. 1996;94(suppl
I):I-494. Abstract.
34.
Rothman SA, Hsia HH, Chmielewski IL, Cossu SF, Vogel
RL, Buxton AE, Miller JM. Standard vs saline irrigated-tip
radiofrequency ablation: lesion size in normal vs scar tissue.
J Am Coll Cardiol. 1997;29:374A. Abstract.
35.
Nakagawa H, Khastir T, Arruda M, Beckman KJ, McClelland
JH, Widman L, Gonzalez M, Lazzara R, Jackman WM. Radiofrequency
catheter ablation using a saline irrigated electrode in patients with
prior failed accessory pathway ablation. Pacing Clin
Electrophysiol. 1995;18:832. Abstract.
36.
Ruffy R, Imran MA, Santel DJ, Wharton M. Radiofrequency
delivery through a cooled catheter tip allows the creation of larger
endomyocardial lesions in the ovine heart.
J Cardiovasc Electrophysiol. 1995;6:10891096.[Medline]
[Order article via Infotrieve]
37.
Jackman WM, Friday KJ, Yeung-Lai-Wah JA, Fitzgerald DM,
Beck B, Bowman AJ, Stelzer P, Harrison L, Lazzara R. New catheter
technique for recording left free-wall accessory
atrioventricular pathway activation: identification of
pathway fiber orientation. Circulation. 1988;78:598611.
38.
Jackman WM, Friday KJ, Fitzgerald DM, Bowman AJ,
Yeung-Lai-Wai JA, Lazzara R. Localization of left free-wall and
posteroseptal accessory atrioventricular
pathways by direct recording of accessory pathway activation.
Pacing Clin Electrophysiol. 1989;12:204214.[Medline]
[Order article via Infotrieve]
39.
Wittkampf FHM, Simmers T, Wever E, Wilde A, Hauer R.
The bulls eye: a novel ablation electrode for recording of high
local bipolar electrograms. Pacing Clin Electrophysiol. 1995;18:833. Abstract.
40.
Wittkampf FHM, Simmers TA, Hauer RNW, Robles de Medina
EO. Myocardial temperature response during radiofrequency catheter
ablation. Pacing Clin Electrophysiol. 1995;18:307317.[Medline]
[Order article via Infotrieve]
41.
Wittkampf FHM, Nakagawa H, Yamanashi WS, Imai S,
Jackman WM. Thermal latency in radiofrequency ablation.
Circulation. 1996;93:10831086.
42.
Avital B, Morgan M, Hare J, Khan M, Lessila C.
Intracardiac explosions during radiofrequency ablation: histopathology
in the acute and chronic dog model. Circulation.
1992;86(suppl I):I-191. Abstract.
43.
Imai S, Nakagawa H, Yamanashi WS, Wittkampf FHM, Pitha
JV, Campbell BW, McClelland JH, Beckman KJ, Arruda M, Lazzara R,
Jackman WM. Avoidance of steam "pop" during radiofrequency
ablation. Circulation. 1996;94(suppl I):I-494. Abstract.
44.
Nakagawa H, Imai S, Arruda M, Schleinkofer D, Freihoff
F, Shah N, Beckman KJ, McClelland JH, Gonzalez M, Otomo K, Lazzara R,
Jackman WM. Prevention of pericardial tamponade during radiofrequency
ablation of accessory pathways using a saline irrigated electrode.
Pacing Clin Electrophysiol. 1997;20:1068. Abstract.
45.
Imai S, Nakagawa H, Yamanashi WS, Wittkampf FHM, Pitha
J, Campbell B, Arruda M, Lazzara R, Jackman WM. Effects of blood flow
on lesion size during radiofrequency ablation in temperature control
mode. Pacing Clin Electrophysiol. 1997;20:1123. Abstract.
46.
Otomo K, Wang X, Yamanashi WS, Nakagawa H, Scherlag BJ,
Lazzara R, Jackman WM. Electrode sliding increases radiofrequency
lesion depth. Circulation. 1994;90(suppl I):I-126. Abstract.
47.
Nakagawa H, Wittkampf FHM, Imai S, Yamanashi WS,
Holden M, Arruda M, Pitha JV, Lazzara R, Jackman WM. Pulsed current
delivery combined with saline irrigation produces deeper radiofrequency
lesions without steam "pop." J Am Coll Cardiol.
1997;29(suppl): 374A. Abstract.
48.
Nakagawa H, Yamanashi WS, Imai S, Wittkampf FHM, Pitha
J, Holden M, Arruda M, Lazzara R, Jackman WM. Long linear atrial
lesions by radiofrequency ablation using temperature control:
differences of lesion size by ablation during sinus rhythm and atrial
fibrillation. Pacing Clin Electrophysiol. 1997;20:1124.
Abstract.
© 1998 American Heart Association, Inc.
Basic Science Reports
Inverse Relationship Between Electrode Size and Lesion Size During Radiofrequency Ablation With Active Electrode Cooling
![]()
Abstract
Top
Abstract
Introduction
Methods
Results
Discussion
References
BackgroundClinical efficacy has
driven the use of larger electrodes (7F, length
4 mm) for
radiofrequency ablation, which reduces electrogram resolution and
causes variability in tissue contact depending on electrode
orientation. With active cooling, ablation electrode size may be
reduced. The purpose of this study was to examine the effect of
electrode length on tissue temperature and lesion size with saline
irrigation used for active cooling.
Key Words: catheter ablation tachyarrhythmias
![]()
Introduction
Top
Abstract
Introduction
Methods
Results
Discussion
References
When
radiofrequency (RF) current was initially explored as an energy source
for catheter ablation, these procedures were performed with
conventional 6F, 2-mm electrodes.1 2 3 4 5 Clinical
efficacy was limited by coagulum formation and impedance rise at
relatively low power levels due to limited convective cooling by
circulating endocavitary blood.1 2 4 5 6 7 8 9 10 11 The
introduction of larger (7F to 8F, 4- to 8-mm) tip electrodes allowed
for RF delivery at higher power levels because of better electrode
cooling due to exposure of the larger noncontact electrode surface to
the circulating endocavitary blood, which markedly improved clinical
efficacy.11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 However, there are multiple
inherent disadvantages of larger electrodes: (1) a reduction in
electrogram resolution, which makes it more difficult to identify the
optimal ablation site; (2) greater variability in electrode-tissue
coupling, depending on catheter tip orientation relative to the
endocardium (parallel versus perpendicular); and (3) a reduction in
flexibility and mobility of the catheter, which may impair positioning
of the ablation electrode. Recently, active electrode cooling by saline
irrigation has been introduced to prevent an impedance rise, allowing
the use of higher RF power to produce significantly larger and deeper
lesions.28 29 30 31 32 33 34 35 36
![]()
Methods
Top
Abstract
Introduction
Methods
Results
Discussion
References
The experimental protocol was approved by the University of
Oklahoma Institutional Animal Care and Use Committee. Eleven mongrel
dogs weighing 20 to 26 kg were anesthetized with sodium
pentobarbital (25 mg/kg) and mechanically ventilated with room air.
General anesthesia was maintained with supplemental doses
of sodium pentobarbital. The right carotid artery was cannulated for
continuous monitoring of atrial pressure. The thigh muscle preparation
was used as previously described (Figure 1
).31 The dog was
initially placed on its right side, and a 20-cm skin incision was made
over the left thigh muscle. The skin, connective tissue, and thin
superficial muscle were gently dissected, exposing the surface of the
thicker underlying muscle. The edges of the skin were raised to form a
cradle that was filled with heparinized canine blood from the same dog,
maintained at 36°C to 37°C. The blood in the cradle was exchanged
at a rate of 350 mL/min. A custom 7F catheter with a central lumen and
either a 2-mm or 5-mm tip electrode with 6 irrigation holes (0.4-mm
diameter) located radially around the electrode 1 mm from the tip
electrode (Cordis Webster, Inc) was used in this study (Figure 2
). The tip electrodes also contained a
thermocouple for measurement of the tip electrode temperature
(accuracy, ±2°C between 20°C and 120°C).

View larger version (36K):
[in a new window]
Figure 1. Schematic of canine thigh muscle preparation. Skin
overlying thigh muscle was incised and raised to form a cradle. Cradle
was filled with heparinized canine blood at 36°C to 37°C, which was
exchanged at 350 mL/min. A 7F ablation electrode catheter with either a
2- or 5-mm tip electrode was positioned against thigh muscle with a
constant weight of 10 g.

View larger version (78K):
[in a new window]
Figure 2. Schematic of ablation catheter and temperature
probes in canine thigh muscle preparation. 7F ablation catheter has a
central lumen and 2- or 5-mm tip electrode with 6 irrigation holes
(0.4-mm diameter) located circumferentially around electrode, 1.0
mm from tip. Electrode has an internal thermocouple for measuring
electrode temperature. A, Perpendicular electrode-tissue orientation.
In group 1 experiment, ablation catheter was positioned perpendicular
to thigh muscle and held with a contact weight of 10 g. Electrode
tip was irrigated through central lumen with heparinized saline at 20
mL/min during RF application. Two fluoroptic thermal sensor probes were
inserted 3.5 and 7.0 mm into tissue directly adjacent to electrode
to record tissue temperature. B, Parallel electrode-tissue
orientation. In group 2 experiment, ablation electrode was positioned
parallel to thigh muscle with a contact weight of 10 g. Two
fluoroptic thermal sensor probes were inserted 3.5 and 7.0 mm into
tissue adjacent to middle of electrode to record tissue
temperature.
). In both
electrode-tissue orientations, a constant weight of 10 g was
applied to the tip electrode. Tissue temperatures were measured with 2
fluoroptic thermal sensor probes (Luxtron Inc, model 3000-4;
measurement range, 0°C to 125°C; accuracy, ±0.2°C) bundled
together with shrink tubing. The probe extended 3.5 and 7.0 mm
from the end of the shrink tubing. The probes were situated 3.5 and
7.0 mm below the surface directly adjacent to the ablation
electrode, as illustrated in Figure 2
.
The dogs were divided into 2 groups: perpendicular
electrode-tissue orientation (group 1, 5 dogs) and parallel
electrode-tissue orientation (group 2, 6 dogs). Five to 8 RF
applications were delivered at constant voltage (50 V) to the 2-mm and
5-mm electrodes at separate sites on the left thigh muscle. The skin
incision was closed, the dog was turned into its left side, and 5 to 8
RF applications were delivered to the right thigh muscle. Before each
RF application, the skin cradle was depleted of blood and the
electrode-tissue contact area was flushed with deionized water,
creating a high-resistance barrier around the electrode except at the
electrode-tissue contact site. A short (1-second), noninjurious (20-V)
RF test application was used to measure the insulated electrode
impedance to estimate electrode-tissue interface impedance. The
temperature probes were then inserted, and the cradle was filled with
heparinized canine blood. RF energy was delivered at 50 V for 30
seconds. The impedance was recorded at the onset of RF application.
The RF application was terminated immediately in the event of an
impedance rise >10
. The occurrence of an audible "pop" was
recorded, but the application of RF energy was not terminated if
the pop was associated with an impedance rise of
10
. After each
RF application, the electrode was examined for coagulum formation. At
the end of each experiment, the impedance was measured between a skin
patch (18x7.5 cm) placed on the thigh muscle and the skin patch on the
opposite thigh, as an approximate measurement of the body component of
the ablation impedance (body impedance).
Two hours after the completion of the ablation protocol, 30 mL
of 2% triphenyl tetrazolium chloride was administered
intravenously. This dye stains intracellular dehydrogenase,
which distinguishes viable and necrotic tissue. The dogs were
euthanized, and the thigh muscles were excised and fixed in 10%
formalin. The thigh muscles were sectioned, and the maximal depth,
maximal diameter, depth at the maximal diameter, and lesion surface
diameter were measured.31
The values are expressed as mean±SD. The electrical
parameters of RF delivery (voltage, current, impedance, and
power), electrode and tissue temperatures, and lesion dimensions were
compared between the 2- and 5-mm electrodes in each electrode
orientation by use of a two-tailed t test for unpaired
variables. A
2 test was used to determine
the significance of the difference in occurrence of audible pop between
the 2 electrodes in each electrode orientation. A value of
P=0.05 was used as the level of statistical
significance.
![]()
Results
Top
Abstract
Introduction
Methods
Results
Discussion
References
A total of 148 RF lesions were produced in the thigh muscles of
the 11 dogs; 31 and 32 RF applications were delivered in the
perpendicular orientation with the 2-mm and 5-mm electrodes,
respectively, and 42 and 43 RF applications in the parallel orientation
with the 2-mm and 5-mm electrodes, respectively. The electrical RF
parameters, temperatures, and lesion dimensions obtained in
these 2 groups are shown in the Table
and Figures 3 through 5![]()
![]()
. Because of active electrode
cooling, neither an impedance rise >10
nor coagulum formation
occurred in any RF application. Because of saline irrigation, the
electrode temperature remained below 53°C during all RF
applications.
View this table:
[in a new window]
Table 1. Radiofrequency Parameters and Temperatures During
Ablation

View larger version (19K):
[in a new window]
Figure 3. Electrical parameters and temperatures
(temp) recorded during RF application at 50 V through 2-mm
electrode with saline irrigation (Irrig, 20 mL/min) in perpendicular
electrode-tissue orientation. Saline irrigation was initiated 5 seconds
before onset of RF application and resulted in a decrease in ablation
electrode temperature from 36°C to 29°C. During RF application, tip
electrode temperature remained 38°C to 42°C. Increase in
temperature resulted in decrease in impedance from 96 to 86
, with
an increase in current from 0.52 to 0.58 A. Tissue temperature
increased throughout 30-second application and reached 101°C at depth
of 3.5 mm and 67°C at depth of 7.0 mm.

View larger version (21K):
[in a new window]
Figure 4. Electrical parameters and
temperatures recorded during RF application at 50 V through 5-mm
electrode with saline irrigation (20 mL/min) in perpendicular
electrode-tissue orientation. Format and animal same as Figure 3
. Note,
longer (5-mm) electrode with larger electrode-blood contact area was
associated with lower overall impedance (67 to 71
) and higher
overall current (0.70 to 0.75 A), but tissue temperatures reached only
64°C at depth of 3.5 mm and only 44°C at 7.0-mm depth.

View larger version (91K):
[in a new window]
Figure 5. Diagram of lesion dimensions for 2 groups studied.
Values are millimeters (mean±SD). (A) indicates maximal lesion depth;
(B), maximal lesion diameter; (C), depth at maximal lesion diameter;
and (D), lesion surface diameter. *P<0.05 between 2-mm
and 5-mm ablation electrodes within same electrode-tissue
orientation.
The insulated electrode impedance, measured with deionized water,
was not significantly different for the 2- and 5-mm electrodes (229±15
and 228±13
, respectively, Table 1
). During RF ablation (cradle
filled with blood), the overall impedance with the 2-mm electrode was
higher than with the 5-mm electrode (98±8 versus 70±9
,
P<0.01). The delivered RF power was lower with the 2-mm
electrode (26.0±2.1 versus 36.4±5.0 W, P<0.01), but
tissue temperatures at the depths of 3.5 and 7 mm were
significantly higher with the 2-mm electrode (Figures 3
and 4
and Table 1
).
Consequently, lesions created with the 2-mm electrode were
significantly wider (maximal diameter, 12.4±1.4 versus 8.4±0.9
mm, P<0.01) and deeper (maximal depth, 8.0±1.0 versus
5.4±0.9 mm, P<0.01, Figure 5
). Small, sharp increases in impedance
(<10
) usually coinciding with an audible pop occurred in 5 of 31
RF applications (after 15.2±3.6 seconds) with the 2-mm electrode and
in 0 of 32 RF applications with the 5-mm electrode
(P<0.05).
In the parallel orientation, the insulated electrode impedance,
measured with deionized water, was significantly different for the 2-mm
and 5-mm electrodes (238±22 and 187±22
, P<0.01, Table 1
). During RF ablation, the overall impedance was higher with the 2-mm
electrode (101±7 versus 76±5
, P<0.01), resulting in
lower RF power (25.0±1.7 versus 33.2±2.2 W, P<0.01), but
tissue temperatures were significantly higher (Table 1
). Although
lesions obtained with the 2-mm electrode were significantly deeper than
with the 5-mm electrode (7.3±0.6 versus 6.9±0.5 mm,
P<0.01, Figure 5
), there was no significant difference in
the maximum lesion diameter (11.1±1.1 versus 11.3±1.1 mm,
P=NS, Figure 5
). Audible pops occurred in 7 of 42 RF
applications (after 18.1±8.8 seconds) with the 2-mm electrode but none
of the 43 RF applications with the 5-mm electrode
(P<0.05).
The body impedance, measured between a skin patch placed on the
thigh muscle and the skin patch on the opposite thigh, was 26±2
in
group 1 and 26±1
in group 2.
![]()
Discussion
Top
Abstract
Introduction
Methods
Results
Discussion
References
In this study, we investigated the effects of electrode size on
lesion formation when the requirement for a large electrode size (to
maintain a low electrode-tissue interface temperature) is eliminated by
active cooling with saline irrigation. With equal RF voltage, the
impedance was higher and the power was lower with a 2-mm electrode than
a 5-mm electrode. However, tissue temperatures and lesion sizes were
significantly larger with the smaller (2-mm) electrode, especially in
the perpendicular orientation.
The unexpected finding of greater tissue heating with a smaller
electrode at lower overall RF power may be explained on the basis of a
more detailed analysis of the various components of the
ablation circuit. The overall ablation circuit can be approximated as
illustrated in Figure 6
.
RRemote represents the impedance of the
connecting cables, catheter, skin patch, and animal body (all impedance
except for the ablation electrode interface with the tissue and blood).
The impedance of all of the cables, connected in series, was measured
at <1
. The impedance of the ablation catheter was measured at 3.5
. The average body impedance was 26
. Therefore,
RRemote can be estimated at 30
. The ablation
electrode impedance (electrode-tissue interface
[RTissue] and electrode-blood interface
[RBlood] connected in parallel) can be
estimated by subtracting 30
(RRemote) from
the overall impedance measured during RF ablation. In the perpendicular
electrode orientation, the 2-mm ablation electrode impedance can be
estimated at 98
-30
=68
(Figure 7A
). At 50 V, the voltage drop across the
electrode-tissue interface (tissue voltage) is 50 Vx(68
/98
)=34.7 V, and the voltage drop across RRemote
is 50 Vx30
/98
=15.3 V. The insulated electrode impedance was
measured with deionized water at 229
. The electrode-tissue
interface impedance (RTissue) can be estimated at
199
by subtracting RRemote (30
) from the
insulated electrode impedance. The tissue power, which is effective
heating power, is (tissue
voltage)2/RTissue=(34.7
V)2/199
=6.1 W. For the 5-mm electrode in the
perpendicular orientation, the overall impedance was 70
. By the
same calculations (Figure 7B
), the 5-mm electrode had a tissue voltage
of 28.6 V and tissue power of only 4.1 W. Therefore, the 2-mm electrode
delivered 49% more heating power to the tissue, resulting a marked
increase in tissue temperatures and lesion size (Table 1
and Figures 3 through 5![]()
![]()
). The larger electrode resulted in greater current shunting
to the blood, increasing overall current. The ineffective current
shunted through the blood pool also has to pass through the ineffective
impedance (RRemote). Consequently, a larger
proportion of the RF voltage is lost to RRemote,
decreasing the voltage available for tissue heating (28.6 versus 34.7 V
with the 5-mm and 2-mm electrodes, respectively).

View larger version (14K):
[in a new window]
Figure 6. Circuit for RF ablation can be considered to have
overall impedance consisting of nonablation electrode impedance
(RRemote) produced by cables, skin patch, and body, which
is in series with impedance of ablation electrode consisting of
electrode-tissue interface impedance (RTissue) and
electrode-blood interface impedance (RBlood) connected in
parallel.

View larger version (35K):
[in a new window]
Figure 7. Estimation of tissue voltage and tissue power for
2- and 5-mm electrodes in perpendicular electrode-tissue orientation
(see text for details).
), but the lower
electrode-tissue interface impedance (due to a larger electrode-tissue
contact area) leads to a tissue heating power similar to the 2-mm
electrode (5.8 versus 5.9 W, Figure 8
), resulting in similar lesion
diameters (Figure 5
). However, the similar tissue power over a smaller
area with the 2-mm electrode resulted in greater tissue temperatures at
the 3.5- and 7.0-mm depths and greater lesion depth.

View larger version (30K):
[in a new window]
Figure 8. Estimation of tissue voltage and tissue power for
2- and 5-mm electrodes in parallel electrode-tissue orientation (see
text for details).
In this study, RF energy was delivered at constant voltage,
whereas most previous experimental and clinical studies used constant
RF power or variable power based on electrode temperature
("temperature control").4 5 6 7 8 9 10 21 23 24 25 26 27 The
use of constant power (instead of constant voltage) would have
magnified the difference in tissue voltage and tissue power between the
2- and 5-mm electrodes and between the perpendicular and parallel
orientations, because voltage is highly dependent on impedance when
constant power is used. For example, at 25 W with the 2- and 5-mm
electrodes in the perpendicular orientation, the overall voltages would
be
49.5 and 41.8 V, the tissue voltages would be 34.3 and 23.9 V,
and tissue power would be 5.9 and 2.9 W with the 2-mm and 5-mm
electrodes, respectively. Therefore, at 25 W, the tissue power would be
103% higher for the 2-mm electrode than the 5-mm electrode, compared
with a 49% increase in tissue power for a constant 50-V application.
For the 5-mm electrode, the tissue power would be 52% greater in the
parallel orientation than the perpendicular orientation at 25 W (4.4
versus 2.9 W) compared with 41% greater at 50 V (5.8 versus 4.1 W).
Therefore, the use of constant-voltage applications should decrease the
variability in lesion size with differences in electrode size and
orientation and may be preferable to constant-power applications
because of a greater predictability in lesion size.
Small, sharp increases in impedance coinciding with audible
pops occurred only with the 2-mm electrode. The selective occurrence of
a steam pop with the smaller electrode was most likely due to the
higher tissue temperatures, resulting from greater tissue power at the
same RF voltage setting, because similar pops were observed with the
5-mm electrode in another study that used 66 V.31
The steam pop is thought to be caused by the sudden release of steam
from below the surface of the tissue.31 33 40 41 42 43 44
Steam formation occurs when the tissue temperature several millimeters
below the surface reaches 100°C. This requires some degree of surface
cooling to prevent the surface temperature from reaching 100°C (with
an impedance rise) before deeper tissue temperatures reach 100°C to
produce steam. Sufficient surface cooling can occur without irrigation
when the ablation electrode is exposed to high blood
flow45 or when the electrode is sliding with each
cardiac contraction.46 Therefore, steam pops
occur relatively often with conventional ablation electrodes, even in
the temperature control mode, and can be recognized by a brief, sharp
5- to 10-
increase in impedance.31 33 42 43 47
The occurrence of pericardial tamponade after RF applications has been
linked to the occurrence of a steam pop.44
Pericardial tamponade may occur when the steam vents through the
epicardium rather than the endocardium. This has occurred more
frequently with an irrigated electrode than with a conventional
electrode,44 probably as a result of steam
formation closer to the atrial epicardium due to greater endocardial
cooling and higher sustained RF power applications. Preliminary studies
indicate that tissue superheating can be prevented by pulsing the RF
application with on and off cycles, such as 5 seconds on and 5 seconds
off.47 The off periods allow the hottest region,
relatively shallow sites, to cool, which prevents superheating and
steam formation. The deeper areas cool less during off periods,
allowing a progressive increase in deep tissue temperature, resulting
in greater lesion depth.47 Alternatively, tissue
superheating can be prevented while deep lesions are obtained by
application of RF current at lower power for a prolonged
period.43
This study used the canine thigh muscle preparation instead of a
beating heart to control electrode-tissue orientation and contact
pressure and to allow the measurement of tissue temperature at various
depths beneath the electrode. The flat surface of the thigh muscle also
allows accurate determination of lesion size and geometry. In contrast,
delivering RF current to the trabeculated endocardium,
especially in a beating heart, often results in an irregular lesion
shape and significantly greater variation in lesion size than was found
in this study, which would make it more difficult to quantify the
effects of changing any parameter (such as electrode
length) on RF lesion size.
) were not significantly
different between the group 1 and group 2 dogs. Therefore, except for
catheter orientation, there were no procedural differences between the
studies in the 2 groups, suggesting that a quantitative comparison may
be reasonable.
This study confirms earlier findings that active electrode cooling
allows for relatively deep lesions to be created with small electrodes
at relatively low power levels.28 29 30 This study
importantly demonstrates a relatively large improvement in heating
efficiency with decreasing electrode size. Therefore, with active
electrode cooling, the ablation electrode size can be decreased with
multiple potential benefits, including (1) an improvement in
electrogram resolution, which should increase mapping accuracy and
decrease the number of RF applications3739; (2)
a decrease in catheter tip stiffness, which may improve catheter
flexibility and mobility to facilitate reaching the ablation site; and
(3) the ability to make very small ablation electrodes for use in small
children or small ablation spaces, such as the middle cardiac vein and
other coronary veins. With active electrode cooling, the
electrode length could ideally be reduced to the electrode diameter,
resulting in the same electrode-tissue contact area regardless of
electrode-tissue orientation. This would make lesion size independent
of electrode orientation. Because electrode cooling is provided by
irrigation, RF voltage or power can be chosen and maintained
independent of the local blood flow ("extrinsic cooling"), leading
to a more consistent and predictable lesion
size.31 32 33 34 35 36 In contrast, the power delivered (and
lesion depth) in the temperature control mode (without irrigation)
varies greatly with local blood flow. In low-flow areas, such as in a
dilated atrium during atrial fibrillation, the RF power is markedly
reduced, producing small and nontransmural
lesions.48
![]()
Acknowledgments
This study was supported by a grant (R01-HL-39670) from the
National Institutes of Health and a grant (HRC-RRP-A028) from the
Oklahoma Center for the Advancement of Science and Technology.
![]()
Footnotes
Presented in part at the 16th Annual Scientific Sessions of the North American Society of Pacing and Electrophysiology, Boston, Mass, May 36, 1995, and previously published in abstract form (Pacing Clin Electrophysiol. 1995;18:917).
![]()
References
Top
Abstract
Introduction
Methods
Results
Discussion
References
1.
Hoyt RH, Huang SK, Marcus FI, Roger S. Factors
influencing trans-catheter radiofrequency ablation of the
myocardium. J Appl Cardiol. 1986;1:469486.
This article has been cited by other articles:
![]() |
M. Linhart, H. Mollnau, A. Bitzen, S. Wurtz, J. W. Schrickel, R. Andrie, F. Stockigt, C. Weiss, G. Nickenig, L. M. Lickfett, et al. In vitro comparison of platinum-iridium and gold tip electrodes: lesion depth in 4 mm, 8 mm, and irrigated-tip radiofrequency ablation catheters Europace, May 1, 2009; 11(5): 565 - 570. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Yokoyama, H. Nakagawa, D. C. Shah, H. Lambert, G. Leo, N. Aeby, A. Ikeda, J. V. Pitha, T. Sharma, R. Lazzara, et al. Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and Thrombus Circ Arrhythmia Electrophysiol, December 1, 2008; 1(5): 354 - 362. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. K. Voeller, R. B. Schuessler, and R. J. Damiano Jr. Surgical Treatment of Atrial Fibrillation Card. Surg. Adult, January 1, 2008; 3(2008): 1375 - 1394. [Full Text] |
||||
![]() |
A. Da Costa, C. Romeyer-Bouchard, V. Dauphinot, D. Lipp, L. Abdellaoui, M. Messier, J. Thevenin, J.-C. Barthelemy, and K. Isaaz Cavotricuspid isthmus angiography predicts atrial flutter ablation efficacy in 281 patients randomized between 8 mm- and externally irrigated-tip catheter Eur. Heart J., August 1, 2006; 27(15): 1833 - 1840. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Nakagawa, K. Yokoyama, R. Lazzara, W. M. Jackman, F. H.M. Wittkampf, and J. V. Pitha Response to Letter Regarding Article "Comparison of Electrode Cooling Between Internal and Open Irrigation in Radiofrequency Ablation Lesion Depth and Incidence of Thrombus and Steam Pop" Circulation, June 27, 2006; 113(25): e937 - e937. [Full Text] [PDF] |
||||
![]() |
M J Earley, D J R Abrams, A D Staniforth, S C Sporton, and R J Schilling Catheter ablation of permanent atrial fibrillation: medium term results Heart, February 1, 2006; 92(2): 233 - 238. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Yokoyama, H. Nakagawa, F. H.M. Wittkampf, J. V. Pitha, R. Lazzara, and W. M. Jackman Comparison of Electrode Cooling Between Internal and Open Irrigation in Radiofrequency Ablation Lesion Depth and Incidence of Thrombus and Steam Pop Circulation, January 3, 2006; 113(1): 11 - 19. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Deneke, K. Khargi, K.-M. Muller, B. Lemke, A. Mugge, A. Laczkovics, A. E. Becker, and P. H. Grewe Histopathology of intraoperatively induced linear radiofrequency ablation lesions in patients with chronic atrial fibrillation Eur. Heart J., September 1, 2005; 26(17): 1797 - 1803. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. K. Bruce, T. J. Bunch, M. A. Milton, A. Sarabanda, S. B. Johnson, and D. L. Packer Discrepancies Between Catheter Tip and Tissue Temperature in Cooled-Tip Ablation: Relevance to Guiding Left Atrial Ablation Circulation, August 16, 2005; 112(7): 954 - 960. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. J. Bunch, G. K. Bruce, S. B. Johnson, A. Sarabanda, M. A. Milton, and D. L. Packer Analysis of Catheter-Tip (8-mm) and Actual Tissue Temperatures Achieved During Radiofrequency Ablation at the Orifice of the Pulmonary Vein Circulation, November 9, 2004; 110(19): 2988 - 2995. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Da Costa, E. Faure, J. Thevenin, M. Messier, S. Bernard, K. Abdel, C. Robin, C. Romeyer, and K. Isaaz Effect of Isthmus Anatomy and Ablation Catheter on Radiofrequency Catheter Ablation of the Cavotricuspid Isthmus Circulation, August 31, 2004; 110(9): 1030 - 1035. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. d'Avila, C. Houghtaling, P. Gutierrez, O. Vragovic, J. N. Ruskin, M. E. Josephson, and V. Y. Reddy Catheter Ablation of Ventricular Epicardial Tissue: A Comparison of Standard and Cooled-Tip Radiofrequency Energy Circulation, May 18, 2004; 109(19): 2363 - 2369. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. G. Spitzer, L. Karolyi, C. Rammler, and T. Otto Primary closed cooled tip ablation of typical atrial flutter in comparison to conventional radiofrequency ablation Europace, January 1, 2002; 4(3): 265 - 271. [Abstract] [PDF] |
||||
![]() |
E. Sosa, M. Scanavacca, A. d'Avila, F. Oliveira, and J. A. F. Ramires Nonsurgical transthoracic epicardial catheter ablation to treat recurrent ventricular tachycardia occurring late after myocardial infarction J. Am. Coll. Cardiol., May 1, 2000; 35(6): 1442 - 1449. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1998 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |