From the Academic Hospital Maastricht, Maastricht, the Netherlands
(H.J.J.W., C.T.); Queen Mary Hospital, Hong Kong, China (C.-P.L., H.-F.T.);
University of Bonn, Bonn, Germany (B.L., W.J.); St Luke's Medical
Center, Milwaukee, Wis (M.A.); University Hospitals of Cleveland, Cleveland,
Ohio (A.L.W.); St George's Hospital, London, UK (A.J.C.); Academic
Hospital Gent, Gent, Belgium (L.J.); and InControl Inc, Redmond, Calif (G.A.).
Methods and ResultsThe study included 51 patients with recurrent
atrial fibrillation who had not responded to antiarrhythmic drugs, were
in New York Heart Association Heart failure class I or II, and were at
low risk for ventricular arrhythmias. The atrial
defibrillation threshold had to be
ConclusionsWith the Atrioverter, prompt and safe restoration of
sinus rhythm is possible in patients with recurrent atrial
fibrillation.
In the goat model,11 persistent AF results
in electrophysiological changes in the
atrium that favor maintenance of fibrillation and early
recurrence after successful conversion. Therefore, to shorten
the episode of AF as much as possible seems important to prevent
recurrence of AF after conversion to sinus rhythm. This article
describes the initial results with an implantable device (the METRIX
Atrioverter) that allows rapid and successful conversion of AF by
low-energy intracardiac shocks. The objective of our study was to
evaluate the safety and efficacy of this new therapy.
The Device
Study Protocol
At implantation, permanent leads were placed in the coronary
sinus, right atrium, and right ventricle. Right ventricular
pacing and sensing thresholds were obtained. AF detection and R-wave
synchronization were again assessed. Two of 4 successes at 240 V had to
be attained at the final lead location. Atrial defibrillation threshold
(ADFT) was estimated as follows. Starting with a 180-V shock, the shock
intensity was increased in 20-V steps until successful defibrillation
was achieved. After this initial success, AF was reinduced, and a shock
20 V less than the previously successful shock was given. Thereafter,
the shock intensity was decreased in 20-V steps until a shock intensity
was delivered that failed to convert the AF despite the delivery of 2
shocks at this intensity. Shocks were then delivered at 20-V steps of
increasing intensity until AF was successfully converted with the
delivery of 2 shocks at this intensity.
To implant the device, ADFT had to be
Postimplant evaluation was performed at predischarge; at 1, 3, and 6
months; and at 6-month intervals thereafter until the completion of the
study. AF detection and R-wave synchronization tests were performed at
predischarge and at 1- and 3-month follow-ups. The defibrillation level
was tested during the predischarge and at 1-month follow-up. Long-term
ADFT testing was repeated at the 3-month follow-up. Follow-up with
device interrogation was performed at 6 months and every 6 months
thereafter. At those times, AF was induced during sinus rhythm by
low-intensity shocks or by rapid atrial pacing by use of a separate
catheter. AF was permitted to last for
During preimplant, implant, and postimplant shock delivery, cardiac
rhythm was continuously recorded to document the possible induction
of ventricular arrhythmias. The decision to use
anticoagulant therapy in the preimplant, implant, and postimplant
periods was left to the investigator, with protocol suggesting the use
of anticoagulant therapy as if the patient did not have the device.
Treatment for Spontaneous Episodes
Safety
Efficacy
The device terminated 96% of the episodes, 1 episode spontaneously
converted before complete device therapy delivery, and 10 episodes (in
9 patients receiving a total of 48 shocks) could not be converted.
Seven of these episodes later spontaneously converted, 2 were
chemically cardioverted, and 1 was converted with external
cardioversion after antiarrhythmic drug pretreatment.
Early recurrence of AF (ERAF) (Figure 2
One patient with frequent episodes (1 to 2 per week) and a
drug-resistant rapid ventricular rate underwent His
bundle ablation and permanent ventricular pacing. One
patient had a failed conversion attempt before the 1-month follow-up
and underwent successful lead repositioning with the next spontaneous
episode and subsequent spontaneous episodes were successfully
terminated with the device. Three patients had successful conversions
early but experienced late failure. In 1 of these patients, spontaneous
episodes were successfully converted after repositioning of the right
atrial electrode. One patient had a high implant threshold, and 50% of
the episodes could be converted after implantation.
Ten patients did not receive Atrioverter treatment for spontaneous
episodes of AF. As shown by examination of the episodes log, 5 patients
had no episodes of AF (follow-up duration, 58 to 354 days; mean,
164±112 days). Five patients did have episodes of AF (1 patient had 1
episode, 1 had 2, 1 had 5, 1 had 6, and 1 had 15) during a follow-up of
52 to 382 days (mean, 177±124 days). Because they did not go to a
hospital, no Atrioverter treatment of these episodes was given. All 47
patients having their devices at the end of follow-up were in sinus
rhythm at that time.
AF Thresholds
Antiarrhythmic Medication After Implantation
Shock Tolerability
Complications
Four patients required repositioning of the right atrial lead: 1
patient because of lead dislocation and 3 because of an acute increase
in the ADFT. One patient had dislodgement of the right
ventricular lead that resulted in a change in signal
quality, inhibiting appropriate shock delivery. The lead was
successfully repositioned.
Efficacy of the device was evaluated by determining its ability to
recognize AF, to reject sinus rhythm, and to convert AF to sinus rhythm
by a low-energy shock. Arrhythmia recognition occurred with
high specificity. In contrast to the ventricular
defibrillator in which the life-threatening character of the
arrhythmia requires a short time from onset to therapy and
therefore a high sensitivity, the atrial defibrillator, which is
treating a nonlife-threatening arrhythmia, should be able to
recognize AF with a very high degree of specificity. The device was
effective in terminating AF, restoring 74% of episodes to sinus rhythm
without further intervention. With drug intervention for ERAF or failed
conversion, the overall efficacy of the device was 86%. As discussed
elsewhere,24 ERAF within minutes after a
successful cardioversion is an important problem. They were observed in
our patients in 27% of episodes and occurred at least once in
approximately half the patients. About half the time, it could be
controlled by the intravenous administration of
antiarrhythmic drugs, followed by repeated cardioversion. This might
present a problem when the device is used outside the hospital. Not
only control of ERAF but also the tolerance of the defibrillator shock
are important aspects of a long-term success of the implantable
Atrioverter. As shown by Ammer et al,25 whereas
shock energy determines the success of a shock, shock voltage
determines the pain perceived by the patient. By using waveforms that
deliver greater energy at lower peak voltages, internal cardioversion
of AF can be performed with less sedation and greater patient
tolerance. The discomfort from intracardiac low-energy shocks is still
poorly understood. Most investigators18 25 26 27
found marked individual variations in shock tolerability. The number of
shocks is important, with discomfort growing as the number of shocks
increases. This suggests that the first shock should be well above the
ADFT to make the shock successful. Lok et al26
showed that most patients can tolerate a shock level of 260 V without
sedation.
It is well known that longer-lasting AF is more difficult to convert to
sinus rhythm and has a higher incidence of
recurrences.23 28 29 It has been shown in
the animal model that electrophysiological
remodeling of the atria leads to marked shortening of the atrial
refractory period and a loss of the normal relation between heart rate
and the duration of the atrial refractory period favoring
ERAF.11
If this electrophysiological remodeling of
the atria can be prevented by defibrillation as early as possible, it
is likely that recurrences of AF will decrease and the number
of necessary shocks will diminish over time. Careful follow-up of the
number of episodes of AF after implantation of the Atrioverter should
answer that question.
The complications observed were mostly related to the use of
intracardiac catheters and the implantation of a device, such as the
necessity to reposition the catheter (6 patients) and infection (2
patients). In 1 patient, His bundle ablation and pacemaker implantation
were performed because of very frequent episodes of AF with a high
ventricular rate. A serious complication, atrial
perforation with cardiac tamponade, occurred in 1 patient. In
retrospect, inappropriate fixation of the atrial lead seems to be the
explanation of this complication. The relatively high complication rate
is probably related to the learning curve of using the device but may
also be caused by its use in 19 centers with differences in techniques,
management, and experiences. This resembles the early implantable
ventricular defibrillator
experience.30
Conclusions
Guest editor for this article was Mark E. Josephson, MD, Beth Israel Hospital, Boston, Mass.
Received April 1, 1998;
revision received June 1, 1998;
accepted June 22, 1998.
2.
Lown B, Amarasingham R, Neuman J. New method for
terminating cardiac arrhythmias: use of synchronized capacitor
discharge. JAMA. 1962;182:548555.
3.
Levy S, Lauribe P, Dolla E, Brémondy M, Calkins
H, Das S, Morady F. A randomized comparison of external and internal
cardioversion of chronic atrial fibrillation. Circulation. 1992;86:14151420.
4.
Scheinman M, Morady F, Hess D, Gonzalez R. Catheter
induced ablation of the atrioventricular junction to
control refractory supraventricular arrhythmias.
JAMA. 1982;248:851855.
5.
Gallagher J, Svenson R, Kassel J, German LD, Bardy GH,
Broughton A, Critelli G. Catheter technique for closed-chest ablation
of the atrioventricular conduction system: a
therapeutic alternative for the treatment of refractory
supraventricular tachycardia. N Engl
J Med. 1982;306:194200.[Abstract]
6.
Williamson BD, Man KC, Daoud E, Niebauer M,
Strickberger SA, Morady F. Radiofrequency catheter modification of
atrioventricular conduction to control the
ventricular rate during atrial fibrillation. N
Engl J Med. 1994;331:910917.
7.
Moe GK. On the multiple wavelet hypothesis of atrial
fibrillation. Arch Int Pharmacodyn Ther. 1962;140:183188.
8.
Allessie MA, Rensma PL, Brugada J, Smeets JRLM, Penn
O, Kirchhof CJHJ. Pathophysiology of atrial fibrillation. In:
Zipes DP, Jalife J, eds. Cardiac Electrophysiology: From Cell to
Bedside. Philadelphia, Pa: WB Saunders Co; 1990:548558.
9.
Cox J, Boineau J, Schuessler R. A review of surgery of
atrial fibrillation. J Cardiovasc Electrophysiol. 1991;2:541561.
10.
Haissaguerre M. Right and left atrial radiofrequency
catheter therapy of paroxysmal atrial fibrillation. J
Cardiovasc Electrophysiol. 1996;7:11321144.[Medline]
[Order article via Infotrieve]
11.
Wijffels MCEF, Kirchhof CJHJ, Dorland R, Allessie MA.
Atrial fibrillation begets atrial fibrillation: a study in awake
chronically instrumented goats. Circulation. 1995;92:19541968.
12.
Jain SC, Bhatnagar WM, Azaml RU, Awasthey P. Elective
countershock in atrial fibrillation with an intracardiac electrode: a
preliminary report. J Assoc Physicians India. 1970;18:821826.[Medline]
[Order article via Infotrieve]
13.
Powell AC, Garan H, McGovern BA, Fallon JT, Krishnan
SC, Ruskin JN. Low energy conversion of atrial fibrillation in the
sheep. J Am Coll Cardiol. 1992;20:707711.[Abstract]
14.
Cooper RAS, Alferness CA, Smith WM, Ideker R. Internal
cardioversion of atrial fibrillation in sheep. Circulation. 1993;87:16731686.
15.
Ayers GM, Alferness CA, Ilina M, Smith WM, Cooper RAS,
Ideker R. Ventricular proarrhythmic effects of
ventricular cycle length and shock strength in a sheep
model of transvenous atrial defibrillation. Circulation. 1994;89:413422.
16.
Sokoloski MC, Ayers GM, Kumagai K, Khrestian CM, Niwano
S, Waldo AL. Safety of transvenous atrial defibrillation: studies in
the canine sterile pericarditis model. Circulation. 1997;96:13431350.
17.
Keelan ET, Krum D, Hare J, Mughal K, Li H, Akhtar M,
Jazayeri MR. Safety of atrial defibrillation shocks synchronized to
narrow and wide QRS complexes during atrial pacing protocols simulating
atrial fibrillation in dogs. Circulation. 1997;96:20222030.
18.
Alt E, Schmitt C, Ammer R, Coenen M, Fotuhi P, Karch M,
Blasini R. Initial experience with intracardiac atrial defibrillation
in patients with chronic atrial fibrillation. PACE Pacing Clin
Electrophysiol. 1994;17:10671078.[Medline]
[Order article via Infotrieve]
19.
Saksena S, Prakash A, Mangeon L, Varanasi S, Kolettis
T, Mathew P, De Groot P, Mehra R, Krol RB. Clinical efficacy and safety
of atrial defibrillation using biphasic shocks and current
nonthoracotomy endocardial lead configurations. Am J
Cardiol. 1995;76:913921.[Medline]
[Order article via Infotrieve]
20.
Murgatroyd F, Slade AKB, Sopher M, Rowland E, Ward DE,
Camm AJ. Efficacy and tolerability of transvenous low energy
cardioversion of paroxysmal atrial fibrillation in humans. J
Am Coll Cardiol. 1995;25:13471353.[Abstract]
21.
Sra JS, Maglio C, Dhala A, Jazayeri MR, Akhtar M.
Feasibility of atrial fibrillation detection and use of a preceding
synchronization interval as a criterion for shock delivery in humans
with atrial fibrillation. J Am Coll Cardiol. 1996;28:15321538.[Abstract]
22.
Schmitt C, Alt E, Plewan A, Ammer R, Leibig M, Karch M,
Schömig A. Low energy intracardiac cardioversion after failed
conventional external cardioversion of atrial fibrillation.
J Am Coll Cardiol. 1996;28:994999.[Abstract]
23.
Lévy S, Ricard P, Lau CP, Lok NS, Camm AJ,
Murgutroyd FO, Jordaens LJ, Kappenberger LJ, Brugada P, Ripley KL.
Multicenter low energy transvenous atrial defibrillation (XAD) trial
results in different subsets of atrial fibrillation. J Am
Coll Cardiol. 1997;29:750755.[Abstract]
24.
Timmermans C, Rodriguez LM, Smeets JLRM, Wellens
HJJ. Immediate reinitiation of atrial fibrillation following
internal atrial defibrillation. J Cardiovasc
Electrophysiol. 1998;9:122128.[Medline]
[Order article via Infotrieve]
25.
Ammer R, Alt E, Ayers G, Schmitt C, Pasquantonio J,
Schmitdt M, Pütter K, Schömig A. Pain threshold for low
energy intracardiac cardioversion of atrial fibrillation with low or no
sedation. Pacing Clin Electrophysiol. 1997;20:230236.[Medline]
[Order article via Infotrieve]
26.
Lok NS, Lau CP, Tse HF, Ayers GM. Clinical shock
tolerability and effect of different right atrial electrode locations
on efficacy of low energy human transvenous atrial defibrillation using
an implantable lead system. J Am Coll Cardiol. 1997;30:13241330.[Abstract]
27.
Jung J, Heisel A, Fries R, Köllner V.
Tolerability of internal low-energy shock strengths currently needed
for endocardial atrial cardioversion. Am J Cardiol. 1997;80:14891490.[Medline]
[Order article via Infotrieve]
28.
Van Gelder IC, Crijns HJ, Van Gilst WH, Verwer R, Lie
KI. Prediction of uneventful cardioversion and maintenance of
sinus rhythm from direct-current electrical cardioversion of chronic
atrial fibrillation and flutter. Am J Cardiol. 1991;68:4146.[Medline]
[Order article via Infotrieve]
29.
Van Gelder IC, Crijns HJ, Tieleman RG, van Gilst WH,
Lie KI. Chronic atrial fibrillation: success of serial cardioversion
therapy and safety of oral anticoagulation. Arch Intern Med. 1996;156:25852593.
30.
Marchlinski FE, Flores BT, Buxton AH, Miller J,
Josephson ME. The automatic implantable cardioverter-defibrillator:
efficacy, complications and device failures. Ann Intern Med. 1986;104:481490.
© 1998 American Heart Association, Inc.
Clinical Investigation and Reports
Atrioverter: An Implantable Device for the Treatment of Atrial Fibrillation
![]()
Abstract
Top
Abstract
Introduction
Methods
Results
Discussion
Appendix 1
References
BackgroundDuring atrial
fibrillation, electrophysiological changes
occur in atrial tissue that favor the maintenance of the
arrhythmia and facilitate recurrence after conversion
to sinus rhythm. An implantable defibrillator connected to right atrial
and coronary sinus defibrillation leads allows prompt
restoration of sinus rhythm by a low-energy shock. The safety and
efficacy of this system, called the Atrioverter, were evaluated in a
prospective, multicenter study.
240 V during preimplant testing.
Atrial fibrillation detection, R-wave shock synchronization, and
defibrillation threshold were tested at implantation and during
follow-up. Shock termination of spontaneous episodes of atrial
fibrillation was performed under physician observation. Results are
given after a minimum of 3 months of follow-up. During a follow-up of
72 to 613 days (mean, 259±138 days), 96% of 227 spontaneous episodes
of atrial fibrillation in 41 patients were successfully converted to
sinus rhythm by the Atrioverter. In 27% of episodes, several shocks
were required because of early recurrence of atrial
fibrillation. Shocks did not induce ventricular
arrhythmias. Most patients received antiarrhythmic medication
during follow-up. In 4 patients, the Atrioverter was removed: in 1
because of infection, in 1 because of cardiac tamponade, and in 1
because of frequent episodes of atrial fibrillation requiring His
bundle ablation.
Key Words: atrial fibrillation atrial defibrillation atrium
![]()
Introduction
Top
Abstract
Introduction
Methods
Results
Discussion
Appendix 1
References
Atrial fibrillation (AF) is the most common sustained
cardiac arrhythmia1 with
hemodynamic, thromboembolic, and health and economic
consequences. Usually, restoration of sinus rhythm is attempted by
pharmacological or electrical therapy.2 3 In
chronic AF, therapy is directed toward maintenance of an
acceptable ventricular rate and prevention of
thromboembolic complications with anticoagulant therapy. When
ventricular rate control cannot be obtained by medication,
interruption or modification of AV conduction with or without
ventricular pacing can be
performed.4 5 6 The concept that multiple atrial
reentrant wavelets are present7 8 in AF
resulted in the approach to divide the atria into several segments,
either surgically (the MAZE operation) or by a
catheter.9 10 In most patients with AF, other
cardiovascular disease is present, affecting the
consequences of the arrhythmia and the selection of
therapy.
![]()
Methods
Top
Abstract
Introduction
Methods
Results
Discussion
Appendix 1
References
Patients
We enrolled 51 patients from 19 centers in 9 different countries
in this study. They had had prior episodes of AF that had spontaneously
terminated or been converted to normal sinus rhythm with intervals of
recurrence of AF episodes between 1 week and 3 months.
Treatment with at least 1 class I or III antiarrhythmic drug was
ineffective or only partially effective, resulting in
arrhythmia recurrences, or not tolerated because of
side effects. Exclusion criteria are given in Table 1
. The 51 patients underwent
clinical evaluation, including history and physical examination,
12-lead ECG, chest radiography,
echocardiography, and laboratory tests. The
protocol received approval from the ethics committee or institutional
review board of each participating center. Informed consent was
obtained from each patient.
View this table:
[in a new window]
Table 1. Exclusion Criteria for Atrial Defibrillator
Implant
The InControl METRIX Atrioverter system consists of an
implantable atrial defibrillator (model 3000 or 3020) connected to
right atrial (perimeter right atrial model 7205) and coronary
sinus (perimeter coronary sinus model 7109) defibrillation
leads and a bipolar endocardial ventricular pacing lead
(Figure 1
), a programmer, and a
defibrillation systems analyzer. The defibrillator is a
battery-powered, pectorally implanted device with a displacement of 53
cm3 and a weight of 79 g (model 3000) or 82
g (model 3020). With specific algorithms, the device detects AF and
delivers R-wave synchronous defibrillation shocks to convert AF to
sinus rhythm. It is also able to pace the ventricle after shock
delivery in case of bradycardia. Shocks can be delivered at a selected
voltage, with a maximal intensity of 300 V. The model 3000
defibrillator has an 80-µF capacitor and can deliver a maximal shock
of 3 J with a biphasic waveform of 3 ms/3 ms. The model 3020
defibrillator has a 160-µF capacitor with a maximal shock of 6 J with
a biphasic waveform of 6 ms/6 ms. The device stores the intracardiac
electrograms from the last 6 successfully terminated episodes of AF.
Real-time telemetry of 3 simultaneous intracardiac
electrograms can be obtained via the programmer. The device can be
programmed in an automatic mode with automatic, periodic activation of
the detection algorithm and shock delivery after a preset delay from
onset of AF, or it can be used in a patient- or
physician-activated mode in which the detection algorithm and
shock therapy are initiated by placing a magnet over the Atrioverter.
During this study, the device was programmed in the monitor mode, and
shock therapy was delivered in hospital by a physician.

View larger version (172K):
[in a new window]
Figure 1. Chest roentgenogram showing Atrioverter with leads
in right atrium, coronary sinus, and apex of right ventricle.
Right atrium and coronary sinus leads are used for
arrhythmia recognition and defibrillation. Right ventricle lead
is used for shock synchronization and, if needed,
ventricular pacing.
Preimplant, Implant, and Postimplant Procedures
During the patient inclusion testing procedure, either temporary
catheters or permanent leads were placed in the coronary sinus,
right atrium, and right ventricle. Patients were sedated according to
the normal clinical practice of each institution. Right
ventricular pacing and sensing thresholds were obtained.
All leads were connected to the defibrillation systems analyzer
or model 3020 device fixture. Atrial signals were examined by
performing the AF detection algorithm. After AF detection testing, the
synchronization algorithm was tested. Synchronization markers were
examined during sinus rhythm to ensure that the device was
synchronizing properly with the ventricular depolarization.
When proper synchronization was confirmed, the device was programmed to
deliver shocks synchronized to the R wave during sinus rhythm to induce
AF. If the device did not induce AF, rapid atrial pacing was used as an
alternative induction method. The AF detection and synchronization
tests were also performed during AF. To be eligible for implantation, 2
successes at 240 V had to be obtained during testing. This was later
revised to allow 1 success out of 3 attempts at 260 V for model 3000
and 240 V for model 3020.
240 V. Automatic mode
operation, which consisted of AF detection, capacitor charge, AF
redetection, synchronization, shock delivery, postshock pacing (if
needed), and AF detection, was also tested during the implant
procedure.
5 minutes before test shocks
were given.
When the patient sought treatment of a spontaneous episode, data
pertaining to shock effectiveness were recorded. Clinical factors,
such as antiarrhythmic medications and changes in system
performance in case of lead position changes, were also
recorded.
![]()
Results
Top
Abstract
Introduction
Methods
Results
Discussion
Appendix 1
References
After the first implant on October 31, 1995, 51 patients received
the Atrioverter, and all patients had
3 months of postimplant
follow-up as of July 4, 1997. These 51 patients were selected from 119
patients undergoing a screening testing procedure. Of 50 patients
screened, 17 received the model 3000, and 34 of 69 screened patients
received the model 3020. Of the failed screenings, 43 were due to high
ADFT. The remaining patients did not proceed to implant because of
either physician/patient decision or exclusions for enrollment. Patient
demographics are given in Table 2
. Two
patients had a history of PTCA, and 2 patients had had CABG. All
patients had been treated with
1 class I or III antiarrhythmic drug
(average, 3.9 drugs) that had been discontinued, was not tolerated, or
was only partially effective. The average duration of follow-up was
259±138 days (
8.6 months), with a median of 232 days (
7.7
months).
View this table:
[in a new window]
Table 2. Profile of 51 Patients
Studied
A total of 3719 shocks were delivered: 3049 during testing and 670
for spontaneous episodes of AF. All shocks for spontaneous episodes
were given during physician observation. There were no reported cases
of induction of ventricular arrhythmias or
inaccurately synchronized shocks during the study. Analysis of
the AF detection algorithm performance during observed
operations revealed a 100% specificity for the recognition of sinus
rhythm (as opposed to AF) and 92.3% sensitivity for the detection of
AF (as opposed to sinus rhythm). A total of 1062 tests of the detection
algorithm were performed in sinus rhythm and 1178 in AF.
Treatment of Spontaneous Episodes
Forty-one patients had 231 episodes of AF for which they sought
therapy (average, 5.6 episodes per patient; range, 1 to 26 episodes).
Four episodes of AF occurred immediately before the time of threshold
testing, and these episodes were used for clinical testing of the
device. Because these episodes were not treated as spontaneous episodes
but rather were used for this testing, they are excluded from further
analysis and discussion. We delivered 670 shocks for the
treatment of the remaining 227 episodes (median, 3 shocks per
episode).
) was present during the treatment
of 62 episodes (27% of all episodes) and seen in 21 of 41 patients
(51%). ERAF was defined as the resumption of AF within 1 minute after
a shock that resulted in sinus rhythm for
1 beat. This prevented
restoration of persistent sinus rhythm for 26 of the episodes. Six
episodes were chemically converted before additional shock delivery, 1
was converted with external defibrillation, and 19 were allowed to
convert spontaneously at a later time. For 22 episodes in which stable
sinus rhythm was eventually obtained, antiarrhythmic drugs were
injected intravenously followed by repeated cardioversion.
With ERAF taken into account, the overall clinical efficacy of the
device was 86.3%.

View larger version (115K):
[in a new window]
Figure 2. Example of early recurrence of AF after
successful shock-induced conversion to sinus rhythm.
ADFTs were measured at implantation and at the 3-month follow-up.
The data collected were the shock voltage and energy of 2 consecutive
successes and 2 consecutive failures; the average of these 2 values was
defined as the ADFT. As shown in Figure 3
, ADFTs were slightly higher at 3 months
for the 3000 model and slightly lower in the 3020 model, without
reaching statistical significance.

View larger version (22K):
[in a new window]
Figure 3. ADFTs at Atrioverter implant and 3 months later.
Shown are data expressed as mean with corresponding SD from models 3000
and 3020.
All but 3 the patients who entered the study were on
antiarrhythmic medication at the time of trial closure. At
implantation, 27 patients were on D-L sotalol (18 patients)
or amiodarone (9 patients). These numbers were 18 and 15
patients, respectively, at the end of the study.
Use of sedation preceding shock delivery was left to the treating
physician in conference with the patient. Marked variability in shock
tolerance between patients was reported by the investigators. The
interinstitutional variability in the level of sedation used made it
impossible to address the shock tolerance issue in detail.
Two patients with the device implanted had subclavian venous
thrombosis. One was treated successfully with urokinase. In the other
patient, it was considered to be a chronic problem, and no intervention
was performed. In 1 patient, a pericardial effusion developed a few
hours after testing during the 1-month follow-up study. During the
testing session, 19 shocks were delivered, 17 at low voltage for
induction of AF and 2 for defibrillation. When the patient returned to
hospital because of chest pain, shortness of breath, and hypotension,
the echocardiogram revealed cardiac tamponade. At operation,
perforation of the atrial wall by the right atrial electrode was found.
The right atrial and right ventricular leads were
extracted, and the right atrial perforation site was repaired. The
coronary sinus lead and the Atrioverter were removed a few days
later. The patient made an uneventful recovery. Post hoc review of the
serial chest x-rays from the 4-day postimplant period revealed that the
lead was floating in the atrium on postimplant day 1. The unstable
position of the right atrial lead probably resulted in atrial
perforation during repeated shock applications 1 month after
implantation. Two patients developed infection, requiring explantation
of the device and leads.
![]()
Discussion
Top
Abstract
Introduction
Methods
Results
Discussion
Appendix 1
References
Thirty-six years ago, Lown et al2 showed
that AF could be converted to sinus rhythm by a
transthoracic high-energy shock. In 1970, Jain et
al12 demonstrated that AF could be treated
successfully by giving shocks with 1 electrode in the right atrium and
the other over the chest wall. In 1992, Levy et
al3 reported that a high-energy shock given
within the atrium had a higher immediate success rate than
transthoracic defibrillation. More recently, it was shown
that the energy required to defibrillate could be markedly reduced by
the use of large-surface right atrial and coronary sinus
electrodes and biphasic shock waveforms.13 14 15 16 17 18 19 20 21 22 23
The present study was designed to determine the safety, efficacy,
and complications of an implanted device able to recognize AF and to
treat the arrhythmia by delivery of an appropriately timed
defibrillation voltage. To prevent a possible long period of
ventricular electrical standstill after successful
defibrillation, the device was equipped with an on-demand
ventricular pace function. A large number of defibrillation
shocks were given, both during the testing phase (3049 shocks) and
during attempts (670 shocks) to convert spontaneous episodes of AF
during a follow-up period of 72 to 613 days (mean, 259±138 days). In
all patients, the shocks were given under the observation of a
physician. No ventricular arrhythmias were induced
in any patient, indicating that the device was safe in the patient
population studied. This safety record, ie, the absence of
inadvertent precipitation of a ventricular
tachyarrhythmia, reflects the application of the
lessons learned from prior testing in animal
models13 14 15 16 17 18 and in
patients.18 19 20 21 This resulted in programming the
atrial shock synchronized to a QRS complex, which occurs
500
milliseconds after the preceding QRS complex, to avoid the T wave of
the preceding beat. Such an approach is possible because conversion of
AF is not an emergency.
The implantable Atrioverter recognizes AF with high specificity,
and in the patients studied, low-energy defibrillation shocks can be
given safely and effectively, resulting in prompt restoration of sinus
rhythm. Early recurrences of AF after a successful shock
occurred at least once in 21 of the 41 patients and usually required
the additional administration of antiarrhythmic drug therapy. Shock
tolerance varied markedly between patients. Lowering of the ADFT by
changing catheter characteristics and shock waveform and the
availability of a short-acting, nonaddictive analgetic/anxiolytic drug
should facilitate the acceptance of this device. The
patients studied fulfilled strict selection criteria, and the
performance of the device was observed in hospital. Several
pharmacological and nonpharmacological treatment options are currently
available to patients with AF. The proper place of the implantable
Atrioverter in the treatment of atrial fibrillation requires further
investigation.
![]()
Appendix 1
Top
Abstract
Introduction
Methods
Results
Discussion
Appendix 1
References
List of Participating Centers and Investigators
Academic Hospital Gent, Gent, Belgium (L. Jordaens, R.
Tavernier, F. Provenier); Hôpital Nord Marseille, Marseille,
France (S. Levy, V. Taramasco, E. Dolla); Hôpital Cardiologique,
Lille, France (S. Kacet, D. Lacroix, P. le Franc, D. Klug, C. Kouakam);
Hôpital "Hôtel Dieu," Rennes, France (C. Daubert, D.
Gras, P. Mabo, D. Pavin); University of Bonn, Bonn, Germany (B.
Lüderitz, W. Jung); Klinikum der Stadt Ludwigshafen,
Ludwigshafen, Germany (K. Seidl); Academic Hospital Maastricht,
Maastricht, Netherlands (H. Wellens, C. Timmermans, L.M. Rodriguez);
Academic Hospital Groningen, Groningen, Netherlands (H.
Crijns, I. van Gelder); Sahlgrenska Sjukhuset, Göteburg, Sweden
(N. Edvardsson, L. Lurje, M. Aunes, H. Walfridsson); Karolinska
Hospital, Stockholm, Sweden (M. Rosenqvist, C. Linde); St George's
Hospital, London, UK (A.J. Camm, E. Rowland, D. Ward, S.M. Sopher, M.
Gallagher, A. Slade, J. Waktare); The Queen Mary Hospital, University
of Hong Kong, China (C.P. Lau, H.F. Tse, N.S. Lok, K. Lee); University
of Calgary, Calgary, Canada (D.G. Wyse); Institute of
Cardiology, Montréal, Canada (B. Thibault, M.
Talajic); St Michael's Hospital, Toronto, Canada (P. Dorian,
D. Newman); St Luke's Medical Center, Milwaukee, Wis (J. Sra, M.
Akhtar, M. Miehl, Z. Blanck, S. Deshpande, A. Dhala, H. Jazayeri, D.
Kress); University Hospitals of Cleveland, Cleveland, Ohio (L. Biblo,
J. Mackall, A. Waldo, D. Rosenbaum, M. Carlson); University Hospital,
Ann Arbor, Mich (A. Strickberger, F. Morady, E. Daoud, R. Goyal, M.
Harvey, K. Ching Man); and University of Utah Medical Center, Salt Lake
City, Utah (R. Freedman, R. Klein, J. Mason, G. Muelheims, P. Spector,
S. Compton).
![]()
Footnotes
Reprint requests to Dr H.J.J. Wellens, Department of Cardiology, Academic Hospital Maastricht, PO Box 5800, 6202 AZ Maastricht, Netherlands.
![]()
References
Top
Abstract
Introduction
Methods
Results
Discussion
Appendix 1
References
1.
Kannel WB, Abbott RD, Savage DD, McNamara PM.
Epidemiologic features of atrial fibrillation: the Framingham study.
N Engl J Med. 1982;306:10181022.[Abstract]
This article has been cited by other articles:
![]() |
I. Cakulev, I. R. Efimov, and A. L. Waldo Cardioversion: Past, Present, and Future Circulation, October 20, 2009; 120(16): 1623 - 1632. [Full Text] [PDF] |
||||
![]() |
R. P. Ricci, A. Quesada, J. Almendral, F. Arribas, C. Wolpert, P. Adragao, M. Zoni-Berisso, X. Navarro, T. DeSanto, A. Grammatico, et al. Dual-chamber implantable cardioverter defibrillators reduce clinical adverse events related to atrial fibrillation when compared with single-chamber defibrillators: a subanalysis of the DATAS trial Europace, May 1, 2009; 11(5): 587 - 593. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. A. Stephenson and C. I. Berul Electrophysiological Interventions for Inherited Arrhythmia Syndromes Circulation, August 28, 2007; 116(9): 1062 - 1080. [Full Text] [PDF] |
||||
![]() |
C. M. Tracy, M. Akhtar, J. P. DiMarco, D. L. Packer, H. H. Weitz, M. A. Creager, D. R. Holmes Jr, G. Merli, G. P. Rodgers, C. M. Tracy, et al. American College of Cardiology/American Heart Association 2006 Update of the Clinical Competence Statement on Invasive Electrophysiology Studies, Catheter Ablation, and Cardioversion: A Report of the American College of Cardiology/American Heart Association/American College of Physicians Task Force on Clinical Competence and Training Developed in Collaboration With the Heart Rhythm Society J. Am. Coll. Cardiol., October 3, 2006; 48(7): 1503 - 1517. [Full Text] [PDF] |
||||
![]() |
A. Kadish and M. Mehra Heart Failure Devices: Implantable Cardioverter-Defibrillators and Biventricular Pacing Therapy Circulation, June 21, 2005; 111(24): 3327 - 3335. [Full Text] [PDF] |
||||
![]() |
D. Flammang, V. Loteanu, D. Hamani, M. Lambiez, and G. Flammang-Dorie Brady-tachy syndrome: Rapid atrial pacing efficacy in preventing atrial fibrillation recurrence assessed by reliable electrograms: the prefib pilot study Europace, January 1, 2005; 7(5): 482 - 489. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. R.J. Mitchell, A. E. Epstein, and For the AFFIRM Investigators Non-Pharmacological Methods of Maintaining Sinus Rhythm * Response Circulation, August 10, 2004; 110(6): e61 - e61. [Full Text] [PDF] |
||||
![]() |
A. Quesada, J.ús Almendral, F. Arribas, R. Ricci, C. Wolpert, P. Adragao, E. Cobo, X. Navarro, and on behalf DATAS investigators The DATAS rationale and design: a controlled, randomized trial to assess the clinical benefit of dual chamber (DDED) defibrillator Europace, January 1, 2004; 6(2): 142 - 150. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.C. Geller, S. Reek, C. Timmermans, T. Kayser, H.-F. Tse, C. Wolpert, W. Jung, A.J. Camm, C.-P. Lau, H. J.J. Wellens, et al. Treatment of atrial fibrillation with an implantable atrial defibrillator -- long term results Eur. Heart J., December 1, 2003; 24(23): 2083 - 2089. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. P. DiMarco Implantable Cardioverter-Defibrillators N. Engl. J. Med., November 6, 2003; 349(19): 1836 - 1847. [Full Text] [PDF] |
||||
![]() |
G. Manoharan, N. Evans, B. Kidwai, D. Allen, J. Anderson, and J. Adgey Novel Passive Implantable Atrial Defibrillator Using Transcutaneous Radiofrequency Energy Transmission Successfully Cardioverts Atrial Fibrillation Circulation, September 16, 2003; 108(11): 1382 - 1388. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Burashnikov and C. Antzelevitch Reinduction of Atrial Fibrillation Immediately After Termination of the Arrhythmia Is Mediated by Late Phase 3 Early Afterdepolarization-Induced Triggered Activity Circulation, May 13, 2003; 107(18): 2355 - 2360. [Abstract] [Full Text] [PDF] |
||||
![]() |
S K S Lairikyengbam, M H Anderson, and A G Davies Present treatment options for atrial fibrillation Postgrad. Med. J., February 1, 2003; 79(928): 67 - 73. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Santini and R. Ricci Atrial fibrillation and heart failure Europace, January 1, 2003; 5(s1): S55 - S59. [Full Text] [PDF] |
||||
![]() |
Y Blaauw, I C Van Gelder, and H J G M Crijns Treatment of atrial fibrillation Heart, October 1, 2002; 88(4): 432 - 437. [Full Text] [PDF] |
||||
![]() |
D. Schwartzman, S. K. Musley, C. Swerdlow, R. H. Hoyt, and E. N. Warman early recurrence of atrial fibrillation after ambulatory shock conversion J. Am. Coll. Cardiol., July 3, 2002; 40(1): 93 - 99. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Cooper, M. S. Katcher, and M. V. Orlov Implantable Devices for the Treatment of Atrial Fibrillation N. Engl. J. Med., June 27, 2002; 346(26): 2062 - 2068. [Full Text] [PDF] |
||||
![]() |
W. S. Aronow Management of the Older Person With Atrial Fibrillation J. Gerontol. A Biol. Sci. Med. Sci., June 1, 2002; 57(6): M352 - 363. [Abstract] [Full Text] |
||||
![]() |
A. Shimizu and O. A. Centurion Electrophysiological properties of the human atrium in atrial fibrillation Cardiovasc Res, May 1, 2002; 54(2): 302 - 314. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. Daubert and P. Mabo Implantable devices to treat atrial fibrillation: real prospects or just new gimmicks? Europace, January 1, 2002; 4(2): 161 - 164. [PDF] |
||||
![]() |
A.J. Camm Future role of electrical therapy for atrial fibrillation: reality for all? Eur. Heart J. Suppl., November 1, 2001; 3(suppl_P): P53 - P56. [Abstract] [PDF] |
||||
![]() |
P. A. Friedman, B. Dijkman, E. N. Warman, H. A. Xia, R. Mehra, M. S. Stanton, and S. C. Hammill Atrial Therapies Reduce Atrial Arrhythmia Burden in Defibrillator Patients Circulation, August 28, 2001; 104(9): 1023 - 1028. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. Zheng, M. E. Benser, G. P. Walcott, and R. E. Ideker Right Atrial Septal Electrode for Reducing the Atrial Defibrillation Threshold Circulation, August 28, 2001; 104(9): 1066 - 1070. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. W. Israel, J. R. Ehrlich, G. Gronefeld, A. Klesius, T. Lawo, B. Lemke, and S. H. Hohnloser Prevalence, characteristics and clinical implications of regular atrial tachyarrhythmias in patients with atrial fibrillation: insights from a study using a new implantable device J. Am. Coll. Cardiol., August 1, 2001; 38(2): 355 - 363. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. H. Everett IV, J. R. Moorman, L.-C. Kok, J. G. Akar, and D. E. Haines Assessment of Global Atrial Fibrillation Organization to Optimize Timing of Atrial Defibrillation Circulation, June 12, 2001; 103(23): 2857 - 2861. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. M. Scheinman and F. Morady Nonpharmacological Approaches to Atrial Fibrillation Circulation, April 24, 2001; 103(16): 2120 - 2125. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. H. Falk Atrial Fibrillation N. Engl. J. Med., April 5, 2001; 344(14): 1067 - 1078. [Full Text] [PDF] |
||||
![]() |
T. D. Blumenthal, T. T. Burnett, and C. D. Swerdlow Prepulses Reduce the Pain of Cutaneous Electrical Shocks Psychosom Med, March 1, 2001; 63(2): 275 - 281. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. Zheng, M. E. Benser, G. P. Walcott, S. D. Girouard, D. L. Rollins, W. M. Smith, and R. E. Ideker Reduction of Atrial Defibrillation Threshold With an Interatrial Septal Electrode Circulation, November 21, 2000; 102(21): 2659 - 2664. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Tracy, M. Akhtar, J. P. DiMarco, D. L. Packer, H. H. Weitz, W. L. Winters, J. L. Achord, A. W. Boone, J. W. Hirshfeld Jr, B. H. Lorell, et al. American College of Cardiology/American Heart Association Clinical Competence Statement on invasive electrophysiology studies, catheter ablation, and cardioversion: A report of the american college of cardiology/american heart association/american college of physicians-american society of internal medicine task force on clinical competence J. Am. Coll. Cardiol., November 1, 2000; 36(5): 1725 - 1736. [Full Text] [PDF] |
||||
![]() |
C. M. Tracy, M. Akhtar, J. P. DiMarco, D. L. Packer, H. H. Weitz, W. L. Winters, J. L. Achord, A. W. Boone, J. W. Hirshfeld Jr, B. H. Lorell, et al. American College of Cardiology/American Heart Association Clinical Competence Statement on Invasive Electrophysiology Studies, Catheter Ablation, and Cardioversion : A Report of the American College of Cardiology/American Heart Association/American College of Physicians-American Society of Internal Medicine Task Force on Clinical Competence Circulation, October 31, 2000; 102(18): 2309 - 2320. [Full Text] [PDF] |
||||
![]() |
E. G. Daoud, C. Timmermans, C. Fellows, R. Hoyt, R. Lemery, K. Dawson, and G. M. Ayers Initial Clinical Experience With Ambulatory Use of an Implantable Atrial Defibrillator for Conversion of Atrial Fibrillation Circulation, September 19, 2000; 102(12): 1407 - 1413. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Boriani, M. Biffi, V. Cervi, G. Bronzetti, G. Magagnoli, R. Zannoli, and A. Branzi Evaluation of Myocardial Injury Following Repeated Internal Atrial Shocks by Monitoring Serum Cardiac Troponin I Levels Chest, August 1, 2000; 118(2): 342 - 347. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-F. Tse, C.-P. Lau, and G.M. Ayers Atrial pacing for suppression of early reinitiation of atrial fibrillation after successful internal cardioversion Eur. Heart J., July 2, 2000; 21(14): 1167 - 1176. [Abstract] [PDF] |
||||
![]() |
C. Timmermans, S. Levy, G. M. Ayers, W. Jung, L. Jordaens, M. Rosenqvist, B. Thibault, J. Camm, L.-M. Rodriguez, H. J. J. Wellens, et al. Spontaneous episodes of atrial fibrillation after implantation of the Metrix Atrioverter: observations on treated and nontreated episodes J. Am. Coll. Cardiol., May 1, 2000; 35(6): 1428 - 1433. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. M. Scheinman Mechanisms of atrial fibrillation: is a cure at hand? J. Am. Coll. Cardiol., May 1, 2000; 35(6): 1687 - 1692. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. D. Swerdlow, W. Schsls, B. Dijkman, W. Jung, N. V. Sheth, W. H. Olson, B. D. Gunderson, and f. t. W. J. A. Investigators Detection of Atrial Fibrillation and Flutter by a Dual-Chamber Implantable Cardioverter-Defibrillator Circulation, February 29, 2000; 101(8): 878 - 885. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Timmermans, A. Nabar, L.-M. Rodriguez, G. Ayers, and H. J. J. Wellens Use of Sedation During Cardioversion With the Implantable Atrial Defibrillator Circulation, October 5, 1999; 100(14): 1499 - 1501. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Katz, R. J. Sweeney, R. M. Gill, P. R. Reid, and E. N. Prystowsky Relation of Atrial Refractoriness to Upper and Lower Limits of Vulnerability for Atrial Fibrillation/Flutter Following Implantable Ventricular Defibrillator Shocks Circulation, September 7, 1999; 100(10): 1125 - 1130. [Abstract] [Full Text] [PDF] |
||||
![]() |
H-F Tse, C-P Lau, and G M Ayers Incidence and modes of onset of early reinitiation of atrial fibrillation after successful internal cardioversion, and its prevention by intravenous sotalol Heart, September 1, 1999; 82(3): 319 - 324. [Abstract] [Full Text] [PDF] |
||||
![]() |
L.-M. Rodriguez, C. Timmermans, and H. J. J. Wellens Are Electrophysiological Changes Induced by Longer Lasting Atrial Fibrillation Reversible? : Observations Using the Atrial Defibrillator Circulation, July 13, 1999; 100(2): 113 - 116. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-P. Hammes, H. Hoerauf, A. Alt, E. Schleicher, J. T. Clausen, R. G. Bretzel, and H. Laqua N{varepsilon}(Carboxymethyl)Lysin and the AGE Receptor RAGE Colocalize in Age-Related Macular Degeneration Invest. Ophthalmol. Vis. Sci., July 1, 1999; 40(8): 1855 - 1859. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-F. Tse, C.-P. Lau, B. M. Yomtov, and G. M. Ayers Implantable atrial defibrillator with a single-pass dual-electrode lead J. Am. Coll. Cardiol., June 1, 1999; 33(7): 1974 - 1980. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Timmermans, L.-M. Rodriguez, G. M. Ayers, H. Lambert, J. L. R. M. Smeets, J. W. S. Vlaeyen, A. Albert, and H. J. J. Wellens Effect of Butorphanol Tartrate on Shock-Related Discomfort During Internal Atrial Defibrillation Circulation, April 13, 1999; 99(14): 1837 - 1842. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Nabar, L.-M. Rodriguez, C. Timmermans, A. van den Dool, J. L. R. M. Smeets, and H. J. J. Wellens Effect of Right Atrial Isthmus Ablation on the Occurrence of Atrial Fibrillation : Observations in Four Patient Groups Having Type I Atrial Flutter With or Without Associated Atrial Fibrillation Circulation, March 23, 1999; 99(11): 1441 - 1445. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-F. Tse, C.-P. Lau, J. S. Sra, H. J. G. M. Crijns, N. Edvardsson, S. Kacet, and D. G. Wyse Atrial Fibrillation Detection and R-Wave Synchronization by Metrix Implantable Atrial Defibrillator : Implications for Long-Term Efficacy and Safety Circulation, March 23, 1999; 99(11): 1446 - 1451. [Abstract] [Full Text] [PDF] |
||||
![]() |
A New Approach to AF Treatment: The Implantable Atrial Defibrillator Journal Watch Cardiology, December 18, 1998; 1998(1218): 7 - 7. [Full Text] |
||||
![]() |
A New Approach to AF Treatment: The Implantable Atrial Defibrillator. Journal Watch Cardiology, December 11, 1998; 1998(1211): 7 - 7. [Full Text] |
||||
![]() |
An Implantable Cardioverter for Atrial Fibrillation Journal Watch (General), October 27, 1998; 1998(1027): 5 - 5. [Full Text] |
||||
![]() |
M. E. Josephson New Approaches to the Management of Atrial Fibrillation : The Role of the Atrial Defibrillator Circulation, October 20, 1998; 98(16): 1594 - 1596. [Full Text] [PDF] |
||||
![]() |
W. Xu, H.-F. Tse, F. H.Y. Chan, P. C. W. Fung, K. L.-F. Lee, and C.-P. Lau New Bayesian Discriminator for Detection of Atrial Tachyarrhythmias Circulation, March 26, 2002; 105(12): 1472 - 1479. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1998 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |