From the Division of Cardiovascular Disease, Department of Medicine
(A.E.E., G.N.K., V.J.P., S.M.D.), and the Department of Pathology (P.G.A.),
University of Alabama, Birmingham.
Correspondence to Andrew E. Epstein, MD, Division of Cardiovascular Disease, University of Alabama at Birmingham, University Station, THT 321L, Birmingham, AL 35294-0006. E-mail aepstein{at}uab.edu
Methods and ResultsThe hearts from 8 patients were examined. At
the time of ICD implantation, the patients' mean age was 47±11 years,
and the left ventricular ejection fraction was 0.24±0.10.
Four patients had ischemic heart disease, and 4 had dilated
cardiomyopathy. Five hearts were examined after
transplantation; 3, after death. The electrode-myocardial interfaces
were characterized by intense endocardial fibrosis and were remarkably
consistent. Each lead was encased by a ring of fibroelastic
tissue, and there was fibrosis of the right ventricular
myocardium adjacent to the leads. Fibrosis involved the
tricuspid valve in 5 patients, and 1 had perforation of the valve by
the lead. Microscopically, interstitial fibrosis was
adjacent to each lead in the current path of ICD shocks. Acute cell
injury was present only in the hearts that had received recent
shocks.
ConclusionsThe ICD electrode-myocardial interface is
characterized by intense fibrosis. The fibrosis associated with
endocardial ICD leads and the cumulative acute damage produced by
defibrillation discharges may explain changes in the defibrillation and
pacing thresholds and the difficulty of lead extraction that can be
encountered with transvenous ICD systems.
ICD Lead Systems
Statistical Analysis
ICD Lead System Data
Gross Morphology
In all 7 cases of chronic lead implantation, there was an
encircling band of fibroelastic tissue surrounding the lead (Figures 1
Microscopic Changes
The fibroelastic tissue scar associated with the leads was focal
and well circumscribed. Adjacent to this confluent scar were
variable degrees of interstitial fibrous connective
tissue insinuated between the myocardial fibers of the
interventricular septum. In many cases, there was an
interesting radial pattern of interstitial fibrosis, almost
suggesting lines of electrical injury from focal points on the ICD lead
(Figure 2C
In patient 2, who received shocks 7 days before death, and in patient
6, who had received shocks on the day of death, there was evidence of
acute cell injury (Figure 3
Finally, the microscopic findings observed in the hearts with
chronically implanted ICD leads were apparently related to the leads
themselves. Sarcoidosis was found unexpectedly in
patient 4 at pathological examination. The fibrosis associated with the
ICD lead was distinct from that associated with the sarcoid (Figure 4A
The inflammatory reaction incited by pacemaker leads has been well
described.18 19 20 21 A well-orchestrated response to
tissue injury begins with thrombus formation and activation of the
complement and fibrinolytic systems. Fluid, protein, and blood cells
enter tissue adjacent to the lead, producing an acute inflammatory
reaction mediated by neutrophils, macrophages, foreign body
giant cells, and fibroblasts. Granulation tissue forms that progresses
to a fibrous connective tissue scar.
There is ample evidence that transthoracic, epicardial, and
endocardial shocks can all cause myocardial damage. Dahl et
al30 showed that transthoracic shocks
caused myocardial necrosis in dogs and that shorter time intervals
between discharges and small paddle sizes led to greater necrosis.
Thus, repetitive shocks from small ICD endocardial leads might also
cause myocardial injury. Furthermore, Babbs et
al31 showed that the effective, damaging, and
lethal electrical doses of transthoracic shocks led to
injury and mortality in a dose-dependent manner. Van Vleet et
al32 concluded that single
transthoracic damped sinusoidal shocks were accompanied by
a large margin of safety over the defibrillation threshold, with a
12-fold suprathreshold shock required to produce death, a 6-fold
suprathreshold shock required to produce macroscopic damage, and a
3-fold suprathreshold shock necessary to induce microscopic damage.
Because ICDs usually operate at much lower energies, the chance of
tissue injury may be minimized. However, endocardial delivery may be
more hazardous. Doherty et al3 showed that the
threshold for significant injury was
The myocardial changes associated with endocardial defibrillator leads
are probably attributable to both shocks themselves and a foreign body
reaction. Van Vleet et al9 implanted leads in
dogs to which no shocks were given. At necropsy,
cardiovascular changes included formation of a fibrous
sheath over the lead along its course in the veins, right atrium, and
right ventricle; adhesion of the leads to adjacent venous and cardiac
structures, including the tricuspid valve; endocardial fibrosis; and
partial penetration of the myocardium at the apex of the
right ventricle by the lead tips. Many of these changes were observed
in our human counterparts.
Transvenous shocks in animals are associated with myocardial necrosis.
Barker-Voelz et al11 showed that necrosis was
concentrated at the distal electrode through which shocks were given.
Because all animals were killed 48 hours after shock delivery, the
development of fibrosis could not be assessed. However, Van Vleet et
al12 assessed cardiac damage in dogs with
chronically implanted defibrillation electrodes through which 4
episodes of multiple shocks were delivered. At 26 weeks, mechanical
injury from the lead on the right side of the heart was manifest by
endocardial fibrosis and fibrous sheath formation. Although the
fibrosis was most marked at contact points adjacent to portions of the
leads that were not adherent to the myocardium, flat areas
of endocardial fibrosis were seen at the electrode-myocardial
interface, and the leads were covered by an extensive fibrotic sheath
as in our patients. Although microscopic evidence of necrosis was
detected in 50% of the dogs, the severity was felt to be
insignificant.
Because of the injury produced, high-energy shocks have been used to
produce animal models of left ventricular
failure.33 34 Nevertheless, energies used are
much greater than those used clinically for defibrillation, virtually
always >100 J. However, because the work discussed above demonstrates
a dose-response relationship for myocardial injury, lower-energy shocks
may have important effects that could be difficult to recognize
clinically. For example, Perkins et al16
described changes in the hearts of 4 men who died as a result of
myocardial infarction and were treated for recurrent cardiac arrest
with 8 to 55 countershocks of 2.5 to 50 J delivered via a temporary
catheter. Myocardial necrosis secondary to the catheter was present
in 1 of the 4 hearts studied. Similarly, Avital et
al8 reported the absence of detectable myocardial
injury in patients receiving intraoperatively or spontaneous shocks
after operation.
Both interstitial and confluent (replacement) fibroses were
associated with ICD leads. Interstitial fibrosis itself is
not unique to patients with
cardiomyopathy.35 36 On the
other hand, confluent fibrosis is less frequent in nonischemic
cardiomyopathy35 but is the
norm in patients with scars resulting from myocardial
infarction.36 Although in both patients with
coronary artery disease and dilated
cardiomyopathy scaring is more prominent on the
left rather than the right side of the heart, our patients had
extensive interstitial fibrosis on the right side of the
heart adjacent to the lead, suggesting a relationship between the
presence of the lead and the scars themselves.
Clinical Implications
For nonsteroid-eluting pacemaker leads, the pacing threshold
increases in a time-dependent manner, usually within the first 2
months.19 In contrast, the pacing threshold for
transvenous ICD leads increases over time without an early
peak.22 This may be another consequence of
fibrosis at the electrode-tissue interface. With the introduction of
steroid-eluting transvenous ICD leads, fibrosis at the electrode-tissue
interface and other alterations, such as increases in the
defibrillation and pacing thresholds, may be attenuated. Serial pacing
threshold data are not available for our patients because the first 4
had devices that did not have pacing capability; of the 4 others, 1
died 8 days after ICD implantation; and 2 followed elsewhere had no
threshold data available.
Although "proarrhythmia" from ICDs is an infrequently
reported event, arrhythmia exacerbation by ICDs does
occur.37 38 39 When observed, it is usually a
consequence of the delivery of therapy, arrhythmia
acceleration, or new arrhythmias. On the other hand, the
myocardial fibrosis that develops at the electrode-tissue interface may
be responsible for new reentry circuits and possibly
arrhythmogenesis.40 41 42
As ICD leads age, lead failure will be recognized with greater
frequency.43 The fibrous sheath that develops
around pacemaker leads sometimes necessitates the use of dilating
sheaths and forceful traction for extraction. Given the more intense
inflammatory reaction and larger fibrous sheaths that encase ICD leads
and attach to intracardiac structures, such as the tricuspid valve,
more complicated extraction is to be expected. The contrasting degrees
of fibrosis associated with the pacemaker and ICD leads in patient 8
provide support for this speculation (Figure 4
Study Limitations
Areas for Future Study
Received February 25, 1998;
revision received May 27, 1998;
accepted June 9, 1998.
2.
Van Vleet JF, Tacker WA, Cechner PE, Bright RM, Greene
JA, Raffee MR, Geddes LA, Ferrans VJ. Effect of shock strength on
survival and acute cardiac damage induced by open-thorax defibrillation
of dogs. Am J Vet Med. 1978;39:981987.
3.
Doherty PW, McLaughlin PR, Billingham M, Kernoff R,
Goris ML, Harrison DC. Cardiac damage produced by direct current
countershock applied to the heart. Am J Cardiol. 1979;23:225232.
4.
Koning G, Veefkind AH, Schneider H. Cardiac damage
caused by direct application of defibrillator shocks to isolated
Langendorff-perfused rabbit heart. Am Heart J. 1980;100:473482.[Medline]
[Order article via Infotrieve]
5.
Fain ES, Billingham M, Winkle RA. Internal cardiac
defibrillation: histopathology and temporal stability of defibrillation
energy requirements. J Am Coll Cardiol. 1987;9:631638.[Abstract]
6.
Harman D, Matthews R, Gilli N, McCulloch R.
Differences in the pathological changes in dogs' hearts after
defibrillation with extrapericardial paddles and implanted
defibrillator electrodes. PACE Pacing Clin Electrophysiol. 1991;14:358361.[Medline]
[Order article via Infotrieve]
7.
Singer I, Hutchins GM, Mirowski M, Mower MM, Veltri
EP, Guarnieri T, Griffith LSC, Watkins L, Juanteguy J, Fisher S, Reid
PR, Weisfeldt ML. Pathologic findings related to the lead system and
repeated defibrillations in patients with the automatic implantable
cardioverter-defibrillator. J Am Coll Cardiol. 1987;10:382388.[Abstract]
8.
Avital B, Port S, Gal R, McKinnie J, Tchou P, Jazayeri
M, Troup P, Akhtar M. Automatic implantable cardioverter/defibrillator
discharges and acute myocardial injury. Circulation. 1990;81:14821487.
9.
Van Vleet JF, Schollmeyer MP, Engle WR, Tacker WA,
Bourland JD. Cardiovascular alterations induced by
chronic transvenous implantation of an automatic defibrillator
electrode in dogs. J Electrocardiol. 1981;14:6772.[Medline]
[Order article via Infotrieve]
10.
Van Vleet JF, Ferrans VJ, Barker MA, Tacker W, Bourland
JD, Schollmeyer MP. Ultrastructural alteration in the fibrous sheath,
endocardium, and myocardium of dogs shocked with
chronically implanted automatic defibrillator leads. Am J
Vet Res. 1982;43:909915.[Medline]
[Order article via Infotrieve]
11.
Barker-Voelz MA, Van Vleet JF, Tacker WA, Bourland JD,
Geddes LA, Schollmeyer MP. Alterations induced by a single
defibrillating shock applied through a chronically implanted catheter
electrode. J Electrocardiol. 1983;16:167179.[Medline]
[Order article via Infotrieve]
12.
Van Vleet JF, Tacker WA, Bourland JD, Kallok MJ,
Schollmeyer MP. Cardiac damage in dogs with chronically implanted
automatic defibrillator electrode catheters and given four episodes of
multiple shocks. Am Heart J. 1983;106:300307.[Medline]
[Order article via Infotrieve]
13.
Oyama K, Lee R, Ott GY, Haupt DW, Reynolds B, Halperin
BD. Cardiac histopathology associated with nonthoracotomy lead systems.
Circulation. 1994;90(suppl I):I-122. Abstract.
14.
Schirmer U, Hemmer W, Lindner KH, Anhäupl T,
Wieser T. Ultrastructural alterations in the right and left
ventricular myocardium following multiple low
energy endocardial countershocks in anesthetized dogs.
Pacing Clin Electrophysiol. 1997;20:7987.[Medline]
[Order article via Infotrieve]
15.
Epstein AE, Anderson PG, Kay GN, Dailey SM, Plumb VJ,
Shepard RB. Gross and microscopic changes associated with a
nonthoracotomy implantable cardioverter-defibrillator. PACE
Pacing Clin Electrophysiol. 1992;15:382386.[Medline]
[Order article via Infotrieve]
16.
Perkins DG, Klein GJ, Silver MD, Yee R, Jones DL.
Cardioversion and defibrillation using a catheter electrode: myocardial
damage assessed at autopsy. PACE Pacing Clin Electrophysiol. 1987;10:800804.[Medline]
[Order article via Infotrieve]
17.
Jones GK, Swerdlow C, Reichenbach DD, Lones M, Poole
JE, Dolack GL, Kudenchuk PJ, Bardy GH. Anatomical findings in patients
having had a chronically in-dwelling coronary sinus
defibrillation lead. Pacing Clin Electrophysiol. 1995;18:20622067.[Medline]
[Order article via Infotrieve]
18.
Stokes K, Bornzin G. The electrode-biointerface. In:
Barold SS, ed. Modern Cardiac Pacing. Mount Kisco, NY:
Futura Publishing Co; 1985:3377.
19.
Stokes K, Anderson J. Low threshold leads: the effect
of steroid elution. In: Antonioli GE, Aubert AE, Ector H, eds.
Pacemaker Leads 1991. Amsterdam, Netherlands: Elsevier
Science Publishers BV; 1991:537542.
20.
Mond HG, Stokes KB. The electrode tissue interface: the
revolutionary role of steroid elution. PACE Pacing Clin
Electrophysiol. 1992;15:95107.[Medline]
[Order article via Infotrieve]
21.
Anderson JM. Mechanisms of inflammation and infection
with implanted devices. Cardiovasc Pathol. 1993;2:33S41S.
22.
Epstein AE, Plumb VJ, Kirk KA, Kay GN. Pacing threshold
increase in nonthoracotomy implantable defibrillator leads:
implications for battery longevity and margin of safety. J
Interventional Cardiac Electrophysiol. 1997;1:131134.[Medline]
[Order article via Infotrieve]
23.
Venditti FJ, Martin DT, Vassolas G, Bowen S. Rise in
chronic defibrillation thresholds in nonthoracotomy implantable
defibrillator. Circulation. 1994;89:216223.
24.
Schwartzman D, Callans DJ, Gottlieb CD, Heo J,
Marchlinski FE. Early postoperative rise in defibrillation threshold in
patients with nonthoracotomy defibrillation lead systems: attenuation
with biphasic shock waveforms. J Cardiovasc
Electrophysiol. 1996;7:483493.[Medline]
[Order article via Infotrieve]
25.
Daoud EG, Man KC, Morady R, Strickberger SA. Rise in
chronic defibrillation energy requirements necessitating implantable
defibrillator lead system revision. Pacing Clin
Electrophysiol. 1997;20:714719.[Medline]
[Order article via Infotrieve]
26.
Schwartzman D, Hull ML, Callans DJ, Gottlieb CD,
Marchlinski FE. Serial defibrillation lead impedance in patients with
nonthoracotomy lead systems. J Cardiovasc
Electrophysiol. 1996;7:697703.[Medline]
[Order article via Infotrieve]
27.
Jung W, Manz M, Moosdorf R, Lüderitz B. Failure
of an implantable cardioverter-defibrillator to redetect
ventricular fibrillation in patients with a nonthoracotomy
lead system. Circulation. 1992;86:12171222.
28.
Brady PA, Friedman PA, Stanton MS. Effect of failed
defibrillation shocks on electrogram amplitude in a nonintegrated
transvenous defibrillation lead system. Am J Cardiol. 1995;76:580584.[Medline]
[Order article via Infotrieve]
29.
Callans DJ, Swarna US, Schwartzman D, Gottlieb CD,
Marchlinski FE. Postshock sensing performance in transvenous
defibrillation lead systems: analysis of detection and
redetection of ventricular fibrillation. J
Cardiovasc Electrophysiol. 1995;6:604612.[Medline]
[Order article via Infotrieve]
30.
Dahl CF, Ewy GA, Warner ED, Thomas ED. Myocardial
necrosis from direct current countershock: effect of paddle electrode
size and time interval between discharges. Circulation. 1974;50:956961.
31.
Babbs CF, Tacker WA, Van Vleet JF, Bourland JD, Geddes
LA. Therapeutic indices for transchest defibrillator shocks: effective,
damaging, and lethal electrical doses. Am Heart J. 1980;99:734738.[Medline]
[Order article via Infotrieve]
32.
Van Vleet JF, Tacker WA, Geddes LA, Ferrans VJ.
Sequential cardiac morphologic alterations induced in dogs by single
transthoracic damped sinusoidal waveform defibrillator
shocks. Am J Vet Res. 1978;39:271278.[Medline]
[Order article via Infotrieve]
33.
McDonald KM, Francis GS, Carlyle PF, Hauer K, Matthews
J, Hunter DW, Cohn JN. Hemodynamic, left
ventricular structural and hormonal changes after discrete
myocardial damage in the dog. J Am Coll Cardiol. 1992;19:460467.[Abstract]
34.
Geddes LA, Ragheb AO, Janas W. A selective,
non-ischemic, non-pharmacological left ventricular
failure animal model. PACE Pacing Clin Electrophysiol. 1994;17:312320.[Medline]
[Order article via Infotrieve]
35.
Roberts WC, Ferrans VJ, Buja LM. Pathologic aspects of
the idiopathic cardiomyopathies. Adv
Cardiol. 1974;13:349367.[Medline]
[Order article via Infotrieve]
36.
Unverferth DV, Baker PB, Swift SE, Chaffee R, Fetters
JK, Uretsky BF, Thompson ME, Leier CV. Extent of myocardial fibrosis
and cellular hypertrophy in dilated
cardiomyopathy. Am J Cardiol. 1986;57:816820.[Medline]
[Order article via Infotrieve]
37.
Hii JTY, Gillis AM, Wyse DG, Sheldon RS, Duff HJ,
Mitchell LB. Risks of developing supraventricular and
ventricular tachyarrhythmias after
implantation of a cardioverter-defibrillator, and timing the activation
of arrhythmia termination therapies. Am J
Cardiol. 1993;71:565568.[Medline]
[Order article via Infotrieve]
38.
Estes NAM, Haugh CJ, Wang PJ, Manolis AS.
Antitachycardia pacing and low-energy cardioversion for
ventricular tachycardia termination: a clinical
perspective. Am Heart J. 1994;127:10381046.[Medline]
[Order article via Infotrieve]
39.
Pinski SL, Fahy GJ. The proarrhythmic potential of
implantable cardioverter-defibrillators. Circulation. 1995;92:16511664.
40.
Fenoglio JJ, Pham TD, Harken AH, Horowitz LH, Josephson
ME, Witt AL. Recurrent sustained ventricular
tachycardia: structure and ultrastructure of subendocardial
regions in which tachycardia originates.
Circulation. 1983;68:518533.
41.
Bolick DR, Hackel DB, Reimer KA, Ideker RE.
Quantitative analysis of myocardial infarct structure in
patients with ventricular tachycardia.
Circulation. 1986;74:12661279.
42.
De Bakker JMT, van Capelle FJL, Janse MJ, Wilde AAM,
Coronel R, Becker AE, Dingemans KP, van Hemel NM, Hauer RNW. Reentry as
a cause of ventricular tachycardia in patients
with chronic ischemic heart disease: electrophysiologic and
anatomic correlation. Circulation. 1988;77:589606.
43.
Lawton JS, Ellenbogen KA, Wood MA, Stambler BS, Herre
JM, Nath S, Bernstein RC, DiMarco JP, Haines DE, Szentpetery S, Baker
LD, Damiano RJ. Sensing lead-related complications in patients with
transvenous implantable cardioverter-defibrillators. Am J
Cardiol. 1996;78:647651.The gross and microscopic changes
associated with nonthoracotomy implantable cardioverter-defibrillator
(ICD) leads are described. The electrode-myocardial interfaces were
characterized by intense endocardial fibrosis. Each lead was encased by
a ring of fibroelastic tissue, and there was fibrosis of the right
ventricle adjacent to the leads. Focal fibrosis involved the tricuspid
valve in 5 cases. Microscopically, interstitial fibrosis
was adjacent to each lead. Acute cell injury was present only in
hearts that had received recent shocks. The myocardial fibrosis
associated with endocardial ICD leads may explain changes in the
defibrillation threshold and pacing thresholds and the difficulty of
extraction that can be encountered with transvenous ICD systems.[Medline]
[Order article via Infotrieve]
© 1998 American Heart Association, Inc.
Clinical Investigation and Reports
Gross and Microscopic Pathological Changes Associated With Nonthoracotomy Implantable Defibrillator Leads
![]()
Abstract
Top
Abstract
Introduction
Methods
Results
Discussion
References
BackgroundAlthough the effects
of epicardial implantable cardioverter-defibrillator (ICD) leads on
underlying cardiac tissue have been reported, the gross and microscopic
changes associated with endocardial ICD leads are less well described.
This study describes the gross and microscopic changes associated with
endocardial ICD leads in humans.
Key Words: arrhythmia death, sudden defibrillation fibrillation
![]()
Introduction
Top
Abstract
Introduction
Methods
Results
Discussion
References
Although the effects of epicardial defibrillator leads on
underlying cardiac tissue have been reported in
animals1 2 3 4 5 6 and humans,7 8
the gross and microscopic changes associated with endocardial
implantable cardioverter-defibrillator (ICD) leads are less well
described,9 10 11 12 13 14 especially in
humans.15 16 17 Similarly, although the
electrode-endocardial interface for transvenous pacing leads is well
described,18 19 20 21 the interface for defibrillation
leads has received little attention.15 In the
case of endocardial pacing leads, there is a sequence of local events
that follow implantation consisting of injury, acute inflammation,
chronic inflammation, granulation tissue formation, foreign body
reaction, and fibrosis.18 19 20 21 In the case of
transvenous ICD leads, the electrical injury produced by shocks
confounds the previously described sequence of events associated with
the tissue response to an endocardial pacing lead. The morphological
changes associated with transvenous ICD leads and the tissue injury and
healing associated with ICD shocks may be responsible for the
sequential progression of electrical phenomena that have been observed
with these leads clinically, including elevation of the pacing
threshold22 and changes in the defibrillation
threshold,23 24 25 defibrillation lead system
impedance,26 and
electrogram.27 28 29 This article describes the
gross and microscopic changes associated with nonthoracotomy ICD leads
in humans.
![]()
Methods
Top
Abstract
Introduction
Methods
Results
Discussion
References
Pathology Protocol
The hearts of 8 patients with transvenous ICD leads were
examined in accordance with institutional guidelines. Photographs were
taken of the gross anatomy, and the hearts were then sectioned
to visualize the ICD lead and its interface with the endocardium.
Thereafter, the hearts were fixed in formalin, and tissue sections were
embedded in paraffin by use of standard histology techniques. These
serial 5-µm sections were examined microscopically with hematoxylin
and eosin stain, and Gomori's aldehyde fuchsin trichrome stain was
used to identify muscle, fibrous connective tissue, and elastic
tissue.
The ICD leads were manufactured by Cardiac Pacemakers, Inc
(Endotak series) in 7 cases and Ventritex, Inc (TVL series) in 1 case.
Each of these leads uses distal tip and spring electrodes in an
integrated bipolar system for sensing, pacing, and defibrillation. The
surface areas and lengths of the distal coils of these leads ranged
from 295 to 470 mm2 and 3.6 to 5.0 cm,
respectively. None of these leads were steroid eluting.
Quantitative data are expressed as mean±SD.
![]()
Results
Top
Abstract
Introduction
Methods
Results
Discussion
References
Study Patients
The hearts from 8 patients were available for evaluation. Their
demographic data are shown in Table 1
. Five specimens were obtained at
cardiac transplantation, and 3 were obtained after death. Of those who
died, the first expired after a self-inflicted gunshot wound (patient
1) and has been reported previously.15 The other
deaths occurred in patients awaiting cardiac transplantation, 1 of
progressive heart failure and cardiogenic shock (patient 3), and 1 of
cardiac arrest (patient 6).
View this table:
[in a new window]
Table 1. Demographics
Data regarding the ICD leads are presented in Table 2
. Those manufactured by Cardiac
Pacemakers, Inc had delivered monophasic shocks in 4 cases and biphasic
shocks in 3 cases, by both external testing equipment and the implanted
ICDs. The Ventritex lead had been used to deliver only biphasic shocks
via an external testing device and a biphasic ICD.
View this table:
[in a new window]
Table 2. Lead Data
The gross anatomical findings were consistent among all
cases. Each lead had been implanted at the right
ventricular apex and entwined in trabeculae
(Figure 1
). The tip of the lead was
embedded in the right side of the interventricular septum
in 5 cases but in the right ventricular free wall in 3
cases.

View larger version (106K):
[in a new window]
Figure 1. A, ICD lead (arrow) of patient 2 as it
appeared looking through right atrium (RA), across tricuspid valve
(TV), and into right ventricle (arrowhead). The distal spring was
embedded in right side of interventricular septum and
encased in dense fibroelastic tissue. B, Lead from patient 4 (arrow)
penetrating the tricuspid valve and embedded in endocardium of right
side of interventricular septum. This long-term implanted
lead was also encased in dense fibroelastic tissue (arrowhead). C and
D, Lead from patient 6, the only one not implanted long term. Where
lead (arrow) crosses tricuspid valve, there is accumulation of
thrombotic material connecting lead to valve leaflet. D, Thrombotic
material at distal portion of lead (arrow) but no fibrosis. E, Close-up
view of heart in A. Longitudinal incision made along the lead reveals
dense fibroelastic tissue that encases lead and attaches it to
tricuspid valve. F, Distal portion of lead from patient 3 and section
of interventricular septum (IVS). Despite application of
enough tension to separate wire coils of distal spring (arrow),
myocardium remained attached to lead.
and 2
). This band encased the lead up to
the level of the tricuspid valve, and in 5 cases, it actually attached
the lead to the valve itself (Figure 1
). In 1 case, the lead had
penetrated a valve leaflet (Figure 1B
). The lead of patient 6 was
implanted 8 days before death (Figure 1C
and 1D
). It was easily
extracted by the application of mild tension. In this case, there was
no fibrous tissue encasing the lead. Instead, there was thrombotic
material adherent to it and the immediately adjacent
myocardium. Traction was applied to the lead of patient 3
to explant it at autopsy. Even though enough tension was applied to
separate the wire coils of the lead, myocardium still
remained attached (Figure 1F
).

View larger version (111K):
[in a new window]
Figure 2. A, Serial sections of interventricular
septum starting near distal end of ICD lead and extending basally at 3-
to 5-mm intervals from patient 2, who underwent cardiac transplantation
676 days after ICD implantation and 7 days after last defibrillator
shock. Each section is oriented with right ventricular
surface of interventricular septum (R) at the top and left
ventricular surface (L) at bottom. Each heart with
long-term implanted leads displayed this characteristic fibroelastic
tissue that encircled the lead (curved arrow). B, Trichrome-stained
histological section of this fibroelastic tissue
(curved arrow) that encircled lead (L). Note fibrous connective tissue
(green) with some elastic tissue (purple) forming this encircling band.
C, Lower-power photomicrograph of trichrome-stained section of bottom
section of tissue from A. Band of fibroelastic tissue (curved arrow)
encircles lead, and beneath in myocardium lies fibrous
connective tissue (F). Area of confluent fibrous connective tissue
immediately adjacent to lead also extends into surrounding
myocardium, forming radial pattern of
interstitial fibrosis (arrows), suggesting that shocks had
caused lines of electrical injury.
After removal of the leads by traction and incision of the
surrounding fibroelastic band, the interventricular septum
was sectioned transversely, starting just below the lead and extending
basally at 3- to 5-mm intervals (Figure 2
). In all cases, there was a
dense fibroelastic reaction at the electrode-myocardial interface. This
reaction was most severe at the tip of the electrode and extended up
the lead, producing a ring of fibroelastic tissue surrounding the lead
and a fibrous scar in the adjacent myocardium (Figure 2
)
ranging from 250 to 1200 µm in thickness. The trichrome stain
(Figure 2B
and 2C
) demonstrates that this fibroelastic tissue contains
both fibrous connective tissue (green) and elastic tissue (purple).
).
). In these
patients, there were small areas of necrosis within the
interventricular septum near the leads and no evidence of
acute necrosis in any other areas of the heart. Patency of blood
vessels in the areas of fibrosis suggests that the changes were not due
to ischemia.

View larger version (119K):
[in a new window]
Figure 3. A and B, Acute cell injury and necrosis that
result, in addition to long-term changes, when recent shocks were
delivered. A, Higher-power photomicrograph of trichrome-stained tissue
section from C. In addition to chronic fibrosis (F), acute necrosis is
also present (arrows). B, High-power photomicrograph of one
area of acute cell injury with necrotic myocytes intermixed with
fibrosis (green) and normal myocytes (red). C, Low-power view of
a hematoxylin and eosinstained section from patient 6, who received
ICD system 8 days before death showing endocardial surface at point of
lead contact. Thrombotic material adheres to endocardial surface
immediately beneath lead (asterisk). Arrows outline multifocal areas of
acute myocyte necrosis in myocardium adjacent to lead. No
acute myocyte necrosis was evident in areas away from lead. D,
High-power view of myocyte necrosis (arrows) shown in C.
and 4B
). Patient 8 had both pacemaker
and ICD leads (Figure 4C
and 4D
). Minimal fibrosis associated with the
pacemaker lead contrasts with marked fibrosis associated with the ICD
lead. Furthermore, in this and all other patients, the
interstitial fibrosis was always in the septum along the
current pathway of ICD shocks. In contrast, myocardium not
in the current pathway, such as the free wall, was devoid of these
changes (Figure 4D
).

View larger version (108K):
[in a new window]
Figure 4. A, Gross photograph of
interventricular septum from patient 4 who had received no
shocks in the preceding 215 days before transplantation. Sarcoidosis
was unexpectedly demonstrated at pathological examination. As with
other long-term implanted leads, this one was encased in fibroelastic
tissue (curved white arrow) with discrete band of fibrous connective
tissue (arrows) adjacent to lead. Dense linear fibrotic band traversed
interventricular septum. Fibrotic tissue associated with
lead is not distinctly different than fibrous connective tissue
associated with sarcoid (curved black arrow) but clearly separated from
sarcoid reaction. B, Low-power trichrome-stained section of A.
Fibroelastic tissue surrounds lead (L); dense fibrotic tissue radiates
from lead and extends across interventricular septum
(arrows); fibrous tissue is associated with sarcoid reaction (curved
arrow). C, Endocardial surface of right ventricular free
wall of patient 8, who had both pacemaker (arrow) and ICD (curved
arrow) leads implanted. D, Hematoxylin and eosinstained section from
area in C. Rim of fibroelastic tissue encircles ICD lead and focus of
endocardial fibrosis at point of pacemaker lead contact. Fibrosis was
more pronounced in association with ICD lead and more attenuated with
smaller, smoother pacemaker lead. Because free-wall
myocardium was not in current pathway,
interstitial fibrosis was absent.
![]()
Discussion
Top
Abstract
Introduction
Methods
Results
Discussion
References
The electrode-myocardial interfaces of the hearts examined in this
study were characterized by intense endocardial fibrosis. Each lead was
encased by a ring of fibroelastic tissue and associated with fibrosis
of the interventricular septum and often the right
ventricular free wall. Fibrosis involved the tricuspid
valve in 5 patients, and the lead perforated the valve in 1 patient.
Microscopically, interstitial fibrosis was adjacent to each
lead, and acute cell injury was present only in patients who had
recent shocks. The observed pathological changes may explain changes in
pacing and defibrillation thresholds, defibrillation impedance, and
sensing performance that have been
reported.22 23 24 25 26 27 28 29 The fibrosis almost certainly
explains the difficulty that can be encountered when transvenous ICD
leads are explanted.
30 J in dogs receiving
countershocks applied directly to the heart.
First, there is an increase in the defibrillation threshold that
occurs with time in patients with nonthoracotomy ICD
systems.23 24 25 26 One possible cause is myocardial
fibrosis at the electrode-tissue interface. Because this threshold
increase is more common for ICD systems that deliver monophasic than
those that deliver biphasic shocks, it is possible that biphasic shocks
not only defibrillate more effectively but also cause less cellular
dysfunction than monophasic shocks. We saw no differences in the
histology of patients with monophasic and biphasic ICDs. Furthermore,
because ICD leads differ from standard bradycardia pacing leads
(stiffer, larger, exposed coil), factors other than shocks may be
important in the genesis of fibrosis.9 Because it
is the coils and shocks delivered through them that seem to incite the
fibrotic reaction, there is no reason to suspect that the reported
changes are specific to any one manufacturer or material.
).
Our discussion of the clinical implications of the findings
reported here is purely speculative. First, we do not have pacing or
defibrillation threshold data for the patients studied to correlate the
intensity of the fibrotic changes with alterations of these
parameters. Second, although we suggest that the
explantation of ICD leads is more difficult than of pacemaker leads,
comparative data are not available. Finally, because our series was
small and all patients either had died or had their hearts examined
after cardiac transplantation, this series may not be
representative of all patients with transvenous ICD
leads. Nevertheless, the consistency of the findings
suggests that these morphological changes may be common to all patients
with these lead systems.
With the advent of more efficacious waveforms and energy delivery
systems, defibrillation may be accomplished with less energy that in
turn could lead to less myocardial damage. Furthermore, steroid-eluting
leads may cause less myocardial fibrosis. These events could make
changes in the defibrillation and pacing thresholds less important
clinically. Whether the pathological changes described here will also
be of importance in atrial defibrillation is unknown. Further study is
required.
![]()
References
Top
Abstract
Introduction
Methods
Results
Discussion
References
1.
Schuder JC, Stoeckle H, Gold JH, West JA, Holland
JA. Ventricular defibrillation in the dog using implanted
and partially implanted electrode systems. Am J
Cardiol. 1994;33:243247.
This article has been cited by other articles:
![]() |
F. Franceschi, F. Thuny, R. Giorgi, I. Sanaa, E. Peyrouse, X. Assouan, S. Prevot, E. Bastard, G. Habib, and J.-C. Deharo Incidence, risk factors, and outcome of traumatic tricuspid regurgitation after percutaneous ventricular lead removal. J. Am. Coll. Cardiol., June 9, 2009; 53(23): 2168 - 2174. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Goette, F. Cantu, L. van Erven, P. Geelen, F. Halimi, J. L. Merino, J. M. Morgan, and on behalf of the Scientific Initiative Committee o Performance and survival of transvenous defibrillation leads: need for a European data registry Europace, January 1, 2009; 11(1): 31 - 34. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. P. Daubert, W. Zareba, D. S. Cannom, S. McNitt, S. Z. Rosero, P. Wang, C. Schuger, J. S. Steinberg, S. L. Higgins, D. J. Wilber, et al. Inappropriate Implantable Cardioverter-Defibrillator Shocks in MADIT II: Frequency, Mechanisms, Predictors, and Survival Impact J. Am. Coll. Cardiol., April 8, 2008; 51(14): 1357 - 1365. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Camboni, C. G. Wollmann, A. Loher, R. Gradaus, H. H. Scheld, and C. Schmid Explantation of Implantable Defibrillator Leads Using Open Heart Surgery or Percutaneous Techniques Ann. Thorac. Surg., January 1, 2008; 85(1): 50 - 55. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. A. Gatzoulis, G. K. Andrikopoulos, T. Apostolopoulos, E. Sotiropoulos, G. Zervopoulos, J. Antoniou, S. Brili, and C. I. Stefanadis Electrical storm is an independent predictor of adverse long-term outcome in the era of implantable defibrillator therapy Europace, January 1, 2005; 7(2): 184 - 192. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. A. L. E. Bracke, A. Meijer, and L. M. van Gelder Malfunction of endocardial defibrillator leads and lead extraction: where do they meet? Europace, January 1, 2002; 4(1): 19 - 24. [Full Text] [PDF] |
||||
![]() |
D. V. Exner, G. J. Klein, and E. N. Prystowsky Primary Prevention of Sudden Death With Implantable Defibrillator Therapy in Patients With Cardiac Disease: Can We Afford to Do It? (Can We Afford Not To?) Circulation, September 25, 2001; 104(13): 1564 - 1570. [Full Text] [PDF] |
||||
![]() |
R. E. Ideker, A. E. Epstein, and G. P. Walcott Is there an optimal capacitance for defibrillation? Europace, January 1, 2001; 3(4): 275 - 277. [PDF] |
||||
![]() |
D J Heaven, M Y Henein, and R Sutton Pacemaker lead related tricuspid stenosis: a report of two cases Heart, March 1, 2000; 83(3): 351 - 352. [Abstract] [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1998 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |