| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
From the Department of Medicine/Division of Cardiovascular Diseases, The
University of Alabama at Birmingham Medical Center.
Methods and ResultsElectrodes were positioned in the right
atrial appendage (RA), left subclavian vein (LSV), proximal
coronary sinus (CSos), and distal coronary sinus (DCS)
in 14 patients with chronic atrial fibrillation (170±185 days). Using
a step-up protocol, we compared ADFTs for a single-current pathway
(RA
ConclusionsFor internal atrial defibrillation in humans,
sequential biphasic waveforms delivered over dual-current pathways
resulted in a markedly reduced (>50% reduction) ADFT compared with a
single shock over a single-current pathway.
In both human17 18 19 20 and
animal21 22 23 24 25 studies, sequential shocks delivered
over dual-current pathways have been shown to reduce the
ventricular defibrillation threshold.
We26 have recently shown that sequential shocks
delivered over dual-current pathways with transvenous electrodes can
markedly reduce the atrial defibrillation threshold (ADFT) in a sheep
model of acutely induced atrial fibrillation. The effect of
dual-current defibrillation pathways with sequential shocks on ADFTs in
humans is unknown. The purpose of the present study was to
investigate the effect of sequential shocks delivered through single-
and dual-current pathways on internal ADFTs in humans with atrial
fibrillation who are candidates for cardioversion. In addition, the
effect of defibrillator capacitance on the dual-current configuration
threshold was investigated.
Defibrillation Electrodes
Defibrillator and Waveforms
For the single-current pathway system, a biphasic shock from a
single-capacitor defibrillator was emulated, with the leading-edge
voltage (VLPh2) of the second phase set equal to
the trailing-edge voltage (VTPh1) of the first
phase (Figure 2A
The ADFT for one single-current pathway with a single 7.5/2.5-ms
biphasic waveform (Figure 2A
Defibrillation Protocol
For the single-current pathway, a step-up protocol was used starting
with a shock strength of 150 V and increasing or decreasing by 50-V
steps up to a maximum of 500 V. If a shock failed, the next shock
strength was increased by 50 V; if the shock succeeded, the next shock
was decreased by 50 V. For dual-current shocks, a step-up protocol was
used starting with a shock strength of 150 V and increasing by 50-V
steps until a successful result was achieved. The lowest step was 150 V
because of the minimal 50-V output of the Ventritex defibrillator. With
a tissue/electrode interface impedance of 50
For each delivered shock or shocks, the leading-edge voltage and the
shock waveform phase durations were programmed in the HVS-02
defibrillator or defibrillators. The actual current waveform delivered
to each pair of electrodes was recorded across a 0.25-
Statistical Methods
All patients were in atrial fibrillation at the time of the protocol,
with a mean duration of 165±187 days (range, 34 to 740 days). The
duration of atrial fibrillation was determined from the time of the
first ECG evidence of atrial fibrillation without any intercurrent
therapy to restore sinus rhythm to the day of the protocol. Six
patients were receiving antiarrhythmic therapy at the time of the
internal procedure. Nine patients had undergone recent
transthoracic cardioversion, which failed to convert seven
of them to sinus rhythm. The other two patients had short-term
successes with transthoracic cardioversion.
Internal cardioversion was successful in 12 of the 14 patients (Table 2
Atrial Defibrillation Threshold
The effect of randomization on the mean leading-edge voltage at ADFT is
shown in Figure 4
Comparison With Previous Defibrillation Studies
It has been argued that electrodes for ventricular
defibrillation should encompass as much of the fibrillating tissue as
possible so that the current pathway for the shock traverses both
ventricles as well as the intraventricular
septum.39 These findings and theories have been
used to develop more efficient lead systems for implantable
ventricular defibrillators.40 One way
to potentially increase the minimum gradient throughout the heart and
to minimize the gradient near the electrodes is by giving two smaller
shocks separated spatially and temporally instead of one large
shock.24 41 42
We4 previously demonstrated in a sheep model of
atrial fibrillation that certain biphasic waveforms were superior to
monophasic waveforms with a single defibrillation current pathway.
Also, we showed that the optimal single-current pathway used electrodes
that encompassed both atria (right atrial appendage and distal
coronary sinus).4 Furthermore, we have
shown in this same sheep model that the use of a dual-current
defibrillation pathway and sequential biphasic shocks, with the first
shock delivered between the right atrial appendage and the distal
coronary sinus and the second shock delivered between the
proximal coronary sinus and the left pulmonary artery,
demonstrated >70% reduction in energy requirements compared with the
optimum single-current pathway of the right atrial appendage to distal
coronary sinus.26 In this sheep
study,26 selection of the electrode
configurations was guided by a mapping study in sheep with acutely
induced atrial fibrillation.43 That mapping study
demonstrated that the earliest sites of atrial activation after
unsuccessful single-current pathway shocks were dependent on electrode
configuration and appeared to occur in regions in which the potential
gradient field produced by the shock would be predicted to be low. The
dual-electrode configurations used in this animal study26
were designed to use a second shock and current pathway that
encompassed the areas in which the previous mapping study had shown the
earliest activations to occur.
Several clinical studies5 6 7 8 have demonstrated
the efficacy and safety of internal atrial defibrillation that uses a
single shock with a single-current pathway. Cooper et
al8 compared multiple monophasic and
biphasic waveforms for internal atrial defibrillation in patients with
induced atrial fibrillation. They demonstrated that certain biphasic
waveforms (7.5/2.5 ms) were more efficacious than symmetrical biphasic
waveforms (5/5 ms) with a single-pathway right atrial appendage to
distal coronary sinus electrode configuration. The ADFT for the
best biphasic waveform (7.5/2.5 ms) was 1 to 2 J, which is still in the
range that is considered painful to most
patients.6 9 13 14 16
The present study represents the first published data for
the use of sequential shocks and dual-current pathways for atrial
defibrillation in humans. The 7.5/2.5-ms biphasic waveforms used in the
present study were chosen on the basis of previous work by Cooper
and coworkers.8 In that clinical study, the use
of a dual-current defibrillation pathway and sequential biphasic
shocks, with the first shock delivered between the right atrial
appendage and the distal coronary sinus and the second shock
delivered between the proximal coronary sinus and the left
subclavian vein, demonstrated >50% reduction in energy requirements
compared with the single-current pathway of the right atrial appendage
to distal coronary sinus. The dual-electrode configurations
used in the present clinical study were designed to use a second shock
and current pathway that encompassed areas of the atria in which the
potential gradient field was predicted to be low after a single-current
shock delivered from the right atrial appendage to distal
coronary sinus. Thus, these findings in humans are similar to
the animal models of internal atrial defibrillation that used
dual-current pathways and sequential biphasic shocks. In both the
animal and human studies, the dual-current sequential-shock
configurations demonstrated a significant reduction in ADFT compared
with a single-current, single-shock configuration. The differences in
actual threshold values between animals and humans can be explained by
differences in species as well as by type of atrial fibrillation
(induced versus chronic). Furthermore, the ADFT for the single-current
pathway configuration was at least one step above the minimal output of
the cardioverter/defibrillator used. The ADFTs for several patients
with both dual-current systems were at the minimal output of the
cardioverter/defibrillator used, and therefore the real ADFTs could
have been even lower. Thus, the actual difference between the single-
and dual-current systems might have been underestimated by the
defibrillation protocol used in the present study.
Effect of Capacitance
Safety
Tolerability and Clinical Implications
Murgatroyd and coworkers,6 using a right atrial
appendage to coronary sinus single-current vector and a 3/3-ms
biphasic waveform, evaluated the tolerability of transvenous atrial
defibrillation shocks in 19 patients. Shock strengths were increased
until defibrillation was achieved, and shocks that required sedation
were scored as causing "severe discomfort." They found that the
range of tolerable shock strengths not requiring sedation was 0.1 to
1.2 J; however, the mean defibrillation threshold was 2.16±1.02 J
(range, 0.7 to 4.4 J). Thus, the majority of the patients required
sedation to achieve successful atrial defibrillation, with only 2 of
the 19 patients being successfully defibrillated without sedation.
Another study48 evaluated the discomfort of
internal single-current pathway biphasic shocks for atrial
defibrillation in 11 nonsedated patients with a history of paroxysmal
or chronic atrial fibrillation. Shocks were delivered in 0.5-J
increments between right atrial and coronary sinus electrodes.
The majority (73%) of the patients indicated that shock strengths
Tomassoni et al16 compared the effect of
different capacitances on ADFT and pain perception in humans. They
found that a higher-capacitance waveform (120 µF) was associated with
a higher ADFT energy requirement but a decreased perception of pain in
6 of 10 patients compared with a lower-capacitance waveform (50 µF).
They found that the ADFT for 120-µF shocks was 2.0±1.0 J and that
even shocks of <0.5 J were considered painful by most patients.
Ammer and coworkers15 evaluated the differences
in pain threshold between different defibrillation waveforms in humans
undergoing internal atrial cardioversion. They compared two biphasic
waveforms and found that a 6/6-ms biphasic waveform had significantly
lower leading-edge voltage requirements at ADFT and was associated with
a decrease in pain perception, with patients requiring less sedation
than for a 3/3-ms biphasic waveform. The mean ADFT energy requirement
for the 6/6-ms biphasic waveform was 6.8±2.8 J. From these studies, it
appears that discomfort caused by the shock is related to the strength
of the shock and that even with optimization of electrode and waveform
configurations, the mean ADFTs determined in previous clinical studies
were still >1 J, which level has been shown to cause severe discomfort
in most patients.6 9 12 13 14 15
The present study did not measure discomfort of the shocks because
deep analgesia/sedation was used in all patients. For the two
dual-current defibrillation pathways, there was at least a slight
decrease in ADFT for each individual patient compared with the
single-current pathway configuration. However, the mean ADFT of
2.0±0.4 J for the best configuration (150-µF dual current) was still
in a range that other studies would predict to be painful to most
patients. The dual-current pathways did result in a markedly reduced
ADFT; however, it remains to be determined if this reduction in
defibrillation voltage and energy decreases discomfort sufficiently to
justify the added complexity of additional electrodes. Use of an active
left pectoral can electrode to replace the left subclavian vein
electrode used in the present study would mean that three internal
electrodes would be needed with a dual-current pathway system. However,
the proximal and distal coronary sinus electrodes could be on
the same catheter. This configuration may be desirable in patients with
a high ADFT and may enable more patients to be candidates for an
implantable device.
Summary
Received September 3, 1997;
revision received February 13, 1998;
accepted February 20, 1998.
2.
Wolf PA, Dawber TR, Thomas E Jr, Kannel WB.
Epidemiologic assessment of chronic atrial fibrillation and risk of
stroke: the Framingham study. Neurology. 1978;28:973977.
3.
Pai S, Torres V. Atrial fibrillation: new management
strategies. Curr Probl Cardiol. 1993;18:233300.
4.
Cooper RAS, Alferness CA, Smith WM, Ideker RE.
Internal cardioversion of atrial fibrillation in sheep.
Circulation. 1993;87:16731686.
5.
Johnson EE, Yarger MD, Wharton JM. Monophasic and
biphasic waveforms for low energy internal cardioversion of atrial
fibrillation in humans. Circulation. 1993;88(suppl I):I-592.
Abstract.
6.
Murgatroyd FD, Slade AKB, Sopher SM, Rowland E, Ward
DE, Camm AJ. Efficacy and tolerability of transvenous low energy
cardioversion of paroxysmal atrial fibrillation in humans. J
Am Coll Cardiol. 1995;25:13471353.[Abstract]
7.
Murgatroyd FD, Johnson EE, Cooper RA, Lau C-P, Alt E,
Kappenburger LJ, Smith JM, Camm AJ, Wharton JM. Safety of low energy
transvenous atrial defibrillation: world experience.
Circulation. 1994;90(suppl I):I-14. Abstract.
8.
Cooper RAS, Johnson EE, Wharton JM. Internal atrial
defibrillation in humans: the improved efficacy of biphasic waveforms
and the importance of phase duration. Circulation. 1997;95:14871496.
9.
Levy S, Ricard P, Gueunoun M, Yapo F, Trigano J,
Mansouri C, Paganelli F. Low-energy cardioversion of spontaneous atrial
fibrillation: immediate and long-term results. Circulation. 1997;96:253259.
10.
Levy S, Ricard P, Lau CP, Lok NS, Camm AJ, Murgatroyd
F, Jordaens LJ, Kappenberger LJ, Brugada P, Ripley KL. Multicenter low
energy transvenous atrial defibrillation (XAD) trial results in
different subsets of atrial fibrillation. J Am Coll
Cardiol. 1997;29:750755.[Abstract]
11.
Lau CP, Tse HF, Lok NS, Ho DSW, Soffer MB, Murgatroyd
F, Camm AJ. Initial clinical experience of an implantable human
atrial defibrillator. Pacing Clin Electrophysiol. 1996;20:220225.
12.
Nathan AW, Bexton RS, Spurrell RAJ, Camm AJ.
Internal transvenous low energy cardioversion for the treatment of
cardiac arrhythmias. Br Heart J. 1984;52:377384.
13.
Jung W, Tebbenjohanns J, Wopert C, Luderitz B. Safety,
efficacy, and pain perception of internal atrial defibrillation in
humans. Circulation. 1995;92(suppl I):I-472. Abstract.
14.
Steinhaus DM, Cardinal D, Mongeon L, Mattson L, Waters
M, Foley L, Corrigan S. Atrial defibrillation: are low energy shocks
acceptable to patients? Pacing Clin Electrophysiol. 1996;19:625. Abstract.
15.
Ammer R, Alt E, Ayers G, Schmitt C, Pasquantonio J,
Schmidt M, Putter K, Schomig A. Pain threshold for low energy
intracardiac cardioversion of atrial fibrillation with low or no
sedation. Pacing Clin Electrophysiol. 1997;20:230236.[Medline]
[Order article via Infotrieve]
16.
Tomassoni G, Newby KH, Kearney MM, Brandon MJ, Barold
H, Natale A. Testing different biphasic waveforms and capacitances:
effect on atrial defibrillation threshold and pain perception.
J Am Coll Cardiol. 1996;28:695699.[Abstract]
17.
Bardy GH, Ivey TD, Allen MD, Johnson G, Greene HL.
Prospective comparison of sequential pulse and single pulse
defibrillation with use of two different clinically available systems.
J Am Coll Cardiol. 1989;14:165171.[Abstract]
18.
Yee R, Jones DL, Klein GJ, Sharma AD, Kallok MJ.
Sequential pulse countershock between two transvenous catheters:
feasibility, safety, and efficacy. Pacing Clin
Electrophysiol. 1989;12:18691877.[Medline]
[Order article via Infotrieve]
19.
Jones DL, Klein GJ, Guiraudon GM, Sharma AD, Kallok MJ,
Bourland JD, Tacker WA Jr. Internal cardiac defibrillation in man:
pronounced improvement with sequential pulse delivery to two different
lead orientations. Circulation. 1986;73:484491.
20.
Jones DL, Klein GJ, Guiraudon GM, Sharma AD. Sequential
pulse defibrillation in humans: orthogonal sequential pulse
defibrillation with epicardial electrodes. J Am Coll
Cardiol. 1988;11:590596.[Abstract]
21.
Jones DL, Klein GJ, Kallok MJ. Improved internal
defibrillation with twin pulse sequential energy delivery to different
lead orientations in pigs. Am J Cardiol. 1985;55:821825.[Medline]
[Order article via Infotrieve]
22.
Jones DL, Natale A, Kim YH, Klein GJ, Wood GK, Robarts
JP. New waveforms and defibrillation in pigs: biphasic, sequential and
biphasic-sequential. Proc Am Assoc Med Inst. 1990;25:46.
Abstract.
23.
Guse PA, Walcott GP, Rollins DL, Smith WM, Ideker
RE. Defibrillation electrode configurations developed from cardiac
mapping that combine biphasic shocks with sequential timing. Am
Heart J. 1992;124:14911500.[Medline]
[Order article via Infotrieve]
24.
Johnson EE, Alferness CA, Wolf PD, Smith WM, Ideker RE.
Effect of pulse separation between two successive biphasic shocks given
over different lead configurations on ventricular
defibrillation efficacy. Circulation. 1992;85:22672274.
25.
Guse PA, Rollins DR, Krassowska W, Wolf PD, Smith WM,
Ideker RE. Effective defibrillation in pigs using interleaved and
common phase sequential biphasic shocks. Pacing Clin
Electrophysiol. 1993;16:17191734.[Medline]
[Order article via Infotrieve]
26.
Cooper RAS, Smith WM, Ideker RE. Internal cardioversion
of atrial fibrillation: marked reduction in defibrillation threshold
with dual current pathways. Circulation. 1997;96:26932700.
27.
Keane D. Impact of pulse characteristics on atrial
defibrillation energy requirements. Pacing Clin
Electrophysiol. 1994;17:10481057.[Medline]
[Order article via Infotrieve]
28.
Krassowska W, Frazier DW, Pilkington TC, Ideker RE.
Potential distribution in three-dimensional periodic
myocardium, II: application to extracellular stimulation.
IEEE Trans Biomed Eng. 1990;37:267284.[Medline]
[Order article via Infotrieve]
29.
Lepeschkin E, Jones JL, Rush S, Jones RE. Local
potential gradients as a unifying measure for thresholds of
stimulation, standstill, tachyarrhythmia and
fibrillation appearing after strong capacitor discharges. Adv
Cardiol. 1978;21:268278.[Medline]
[Order article via Infotrieve]
30.
Frazier DW, Krassowska W, Chen P-S, Wolf PD, Dixon EG,
Smith WM, Ideker RE. Extracellular field required for excitation in
three-dimensional anisotropic canine myocardium. Circ
Res. 1988;63:147164.
31.
Chen P-S, Wolf PD, Claydon FJ III, Dixon EG, Vidaillet
HJ Jr, Danieley ND, Pilkington TC, Ideker RE. The potential gradient
field created by epicardial defibrillation electrodes in dogs.
Circulation. 1986;74:626636.
32.
Witkowski FX, Penkoske PA, Plonsey R. Mechanism of
cardiac defibrillation in open-chest dogs with unipolar DC-coupled
simultaneous activation and shock potential
recordings. Circulation. 1990;82:244260.
33.
Wharton JM, Wolf PD, Smith WM, Chen P-S, Frazier DW,
Yabe S, Danieley N, Ideker RE. Cardiac potential and potential gradient
fields generated by single, combined, and sequential shocks during
ventricular defibrillation. Circulation. 1992;85:15101523.
34.
Lesigne C, Levy B, Saumont R, Birkui P, Bardou A, Rubin
B. An energy-time analysis of ventricular
fibrillation and defibrillation thresholds with internal electrodes.
Med Biol Eng. 1976;14:617622.[Medline]
[Order article via Infotrieve]
35.
Yabe S, Smith WM, Daubert JP, Wolf PD, Rollins DL,
Ideker RE. Conduction disturbances caused by high current
density electric fields. Circ Res. 1990;66:11901203.
36.
Jones JL, Lepeschkin E, Jones RE, Rush S. Response of
cultured myocardial cells to countershock-type electric field
stimulation. Am J Physiol. 1978;235:H214H222.
37.
Pansegrau DG, Abboud FM. Hemodynamic
effects of ventricular defibrillation. J Clin
Invest. 1970;49:282297.
38.
Dahl CF, Ewy GA, Warner ED, Thomas ED. Myocardial
necrosis from direct current countershock: effect of paddle size and
time interval between discharge. Circulation. 1974;50:956961.
39.
Ideker RE, Wolf PD, Alferness CA, Krassowska W, Smith
WM. Current concepts for selecting the location, size, and shape of
defibrillation electrodes. Pacing Clin Electrophysiol. 1991;14:227240.[Medline]
[Order article via Infotrieve]
40.
Dixon EG, Tang ASL, Wolf PD, Meador JT, Fine MJ, Calfee
RV, Ideker RE. Improved defibrillation thresholds with large contoured
epicardial electrodes and biphasic waveforms. Circulation. 1987;76:11761184.
41.
Cooper RAS, Wallenius ST, Smith WM, Ideker RE. The
effect of phase separation on biphasic waveform defibrillation.
Pacing Clin Electrophysiol. 1993;16:471482.[Medline]
[Order article via Infotrieve]
42.
Sweeney RJ, Gill RM, Reid PR. Characteristics of
multiple-shock defibrillation. J Cardiovasc
Electrophysiol. 1995;6:89102.[Medline]
[Order article via Infotrieve]
43.
Cooper RAS, Smith WM, Ideker RE. Early activation sites
after unsuccessful internal atrial defibrillation shocks: the effect of
electrode configuration. Pacing Clin Electrophysiol. 1996;19:706. Abstract.
44.
Leonelli FM, Kroll MW, Brewer JE. Defibrillation
thresholds are lower with smaller storage capacitors. Pacing Clin
Electrophysiol. 1995;18:16611665.[Medline]
[Order article via Infotrieve]
45.
Swerdlow CD, Kass RM, Chen PS, Hwang C, Raissi S.
Effect of capacitor size and pathway resistance on defibrillation
threshold for implantable defibrillators. Circulation. 1994;90:18401846.
46.
Natale A, Newby KH, Brandon J, Kearney M, Beheiry S,
Henderson PH, Wolverton A, Fanelli R, Pisano E. Defibrillation with
different energy levels and during sleep: impact on pain perception.
J Am Coll Cardiol. 1997;29:195A. Abstract.
47.
Saksena S, Prakash A, Mangeon L, Varanasi S, Kolettis
T, Mathew P, de Groot P, Mehra R, Krol RB. Clinical efficacy and safety
of atrial defibrillation using biphasic shocks and current
nonthoracotomy endocardial lead configurations. Am J
Cardiol. 1995;76:913921.[Medline]
[Order article via Infotrieve]
48.
Jung W, Pfeiffer D, Wolpert C, Pizzulli L, Fehske W,
Schumacher B, Lewalter T, Omran H, Korte T, Luderitz D. Which patients
do benefit from an implantable atrial defibrillator? J Am
Coll Cardiol. 1996;27:301.
© 1998 American Heart Association, Inc.
Clinical Investigation and Reports
Marked Reduction in Internal Atrial Defibrillation Thresholds With Dual-Current Pathways and Sequential Shocks in Humans
![]()
Abstract
Top
Abstract
Introduction
Methods
Results
Discussion
References
BackgroundThis study tested the
ability of sequential shocks delivered through dual-current pathways to
lower the atrial defibrillation threshold (ADFT) compared with a
biphasic shock through a standard single-current pathway.
DCS) that used a single 7.5/2.5-ms biphasic shock from a 150-µF
capacitor with those for a dual-current pathway system (RA
DCS
followed by CSos
LSV) using sequential 7.5/2.5-ms biphasic shocks
with capacitor discharge waveforms for 150-µF and 600-µF
capacitors. Both dual-current pathway configurations (2.0±0.4 J for
150-µF capacitance, 2.4±0.5 J for 600-µF capacitance) had a
significantly lower ADFT than the single-current pathway (5.1±1.8 J).
Whereas the dual-current pathway with 150-µF capacitor shocks had a
significantly lower energy threshold, there was no statistical
difference in terms of leading-edge voltage compared with the
dual-current pathway with 600-µF capacitance shocks. There were no
ventricular arrhythmias induced with appropriately
synchronized shocks.
Key Words: defibrillation atrium fibrillation
![]()
Introduction
Top
Abstract
Introduction
Methods
Results
Discussion
References
Atrial fibrillation
is the most common arrhythmia encountered in clinical medicine
and often requires pharmacological and/or electrical therapy to restore
sinus rhythm.1 2 3 Recently, biphasic waveform
shocks delivered via transvenous defibrillation electrodes have been
shown to be an effective and feasible technique in
animals4 and humans5 6 7 8 9 for
the termination of atrial fibrillation. Clinical evaluation of an
implantable atrial defibrillator is currently in
progress.10 This device delivers biphasic shocks
through a single-current pathway with electrodes in the lateral
coronary sinus and right atrium.11
However, the present energy requirements for successful atrial
defibrillation with this lead configuration remain sufficiently high
that they are usually painful to the
patient.6 9 12 13 14 15 Some clinical studies have
demonstrated that the pain associated with internal atrial
defibrillation shocks is related to the intensity of the
shock.6 9 12 13 15 16 Current research efforts
focus on methods to lower the defibrillation threshold in hope of
lowering the shock intensity and decreasing the pain associated with
internal atrial defibrillation.
![]()
Methods
Top
Abstract
Introduction
Methods
Results
Discussion
References
Patients
Fourteen patients who were referred to the University of Alabama
Medical Center or Birmingham Veterans Administration Medical Center
Electrophysiological Services were originally
screened and enrolled. All patients were seen in consult or referred
for cardioversion of atrial fibrillation. Informed consent was obtained
in all cases, and the study protocol was approved by the Institutional
Review Board of the University of Alabama Medical Center and Birmingham
Veterans Administration Medical Center. The patients were brought to
the clinical electrophysiology laboratory in a fasting, nonsedated
state. All patients were in atrial fibrillation on arrival at the
electrophysiology laboratory. The patients were given
intravenous sedation/analgesia, and local
anesthesia was used at the sites of catheter insertion.
Synchronization of the shocks to ventricular activation was
accomplished with the use of the filtered and amplified signal of the
surface ECG lead II or aVF from a standard
electrophysiological mapping system
(Cardiolab, Prucka Engineering Inc). External pacing and defibrillation
patches (PD2200, Zoll Medical Corp) were positioned in the
anterior-posterior position for back-up external defibrillation and
postshock external bradycardia pacing if necessary via a standard
external defibrillator/external pacemaker (PD 2000 Zoll Medical
Corp).
Two 6F catheters with 10 rings and one 6F catheter with 30 rings
(Electro-Catheter Corp) were used as the defibrillation electrodes.
Each ring consisted of a 5-mm electrode with 2-mm interelectrode
spacing. One 10-pole catheter was used as the anode and was advanced
from the right femoral vein so that the distal tip was positioned in
the anteromedial right atrial appendage, with the body of the
electrodes positioned along the posterolateral right atrium (Figure 1
). Poles 3 to 10 were combined to make a
defibrillation electrode with total surface area of
140
mm2, and poles 1 to 2 were used to pace or
record a bipolar electrogram from the high right atrium. The
cathode for the first current pathway consisted of the distal 10 poles
of the 30-pole catheter, which was advanced from the right internal
jugular vein until the distal tip was positioned as laterally as
possible in the coronary sinus (Figure 1
). Poles 11 to 12 of
the 30-pole catheter were used to pace or record a bipolar left
atrial (coronary sinus) electrogram. The anode for the second
current pathway consisted of poles 13 to 22 of the
30-pole coronary sinus catheter. These electrodes were usually
located at the ostium of the coronary sinus and along the low
posterior right atrium (Figure 1
). The cathode for the second current
pathway consisted of the second 10-pole catheter, which was advanced
from the right femoral jugular vein until the body of the electrodes
was positioned at the junction of the innominate and left subclavian
vein (Figure 1
). All 10 poles were combined to make the defibrillation
electrode. Of the 14 patients, 12 were receiving oral warfarin therapy
for at least 3 weeks before the procedure, and oral anticoagulation was
withheld the night before the procedure. Venous access was not
attempted unless the international normalized ratio was <1.7 the
morning of the procedure. The other 2 patients were receiving oral
warfarin therapy until admission to the hospital, at which time
warfarin was discontinued and intravenous heparin therapy
was started. The intravenous heparin was discontinued 4
hours before the procedure. Oral anticoagulation with warfarin was
resumed for all patients before discharge from the hospital.


View larger version (209K):
[in a new window]
Figure 1. Two pictures are shown for each panel. The lower
picture in each panel is centered over the cardiac silhouette, and the
upper picture in each panel is centered over the base of the heart to
include the left subclavian vein. Left, Fluoroscopic views of
defibrillation electrode positions in right anterior oblique view
(RAO). Right, Fluoroscopic views of defibrillation electrode positions
in left anterior oblique view (LAO). RAap indicates right atrial
appendage; DCS, distal coronary sinus; CSos, coronary
sinus os/proximal coronary sinus; and LSCV, left subclavian
vein.
The defibrillation waveforms were delivered from a Ventritex
HVS-02 programmable cardioverter/defibrillator (Ventritex Inc), which
has a capacitance of 150 µF. For the dual-current pathway
configuration, two defibrillators were used. Each device has two
programmable outputs, each capable of delivering a truncated
exponential monophasic waveform. The pulse widths and polarities of
both outputs are programmable. The shocks were synchronized to the QRS
by use of an electrophysiological
stimulator (DTU 201, Bloom Associates, Ltd). The analog output from the
electrophysiological mapping system
(surface limb lead II or aVF) was sensed by the stimulator as an
external S1, which then stimulated the dual
cardioverter/defibrillators via the S2 and
S3 outputs. The timing of the two shocks was
controlled with the stimulator by controlling the timing of the
S1-S2 and
S2-S3 intervals. Cross talk
between the output of the first defibrillator that could trigger the
second defibrillator prematurely was controlled by use of
custom-designed optical isolation units for the trigger pulse inputs
for each defibrillator.
). For the dual-current
pathway system, a single-capacitor system was also emulated (Figure 2B
and 2C
). The leading-edge voltage of the first phase of the second
current pathway shock (VLPh3) was set equal to
the trailing-edge voltage of the second phase of the first current
pathway shock (VTPh2). Trailing-edge voltages for
each phase VT were calculated by use of the
following equation:
where VL is the leading-edge voltage of
the previous phase, t is the duration of the previous phase, R is the
resistance estimated by the resistance from the previous shock, and C
is the capacitance. The polarity of each phase was opposite to the
preceding phase, with an interphase delay of 0.2 ms.


View larger version (21K):
[in a new window]
Figure 2. Defibrillation waveforms used. A, Single-current
pathway (SP) biphasic waveform. This waveform emulated the output of a
single 150-µF capacitor discharge in that the trailing-edge voltage
of phase 1 (VTPh1) was equal (±10 V) to the leading-edge
voltage of phase 2 (VLPh2). Also, phase 2 of the biphasic
waveform was opposite in polarity to phase 1. B, Two biphasic waveforms
delivered sequentially through dual-current pathways (DP). All four
phases of the two biphasic waveforms emulated a single 150-µF
capacitor discharge in that the trailing-edge voltage of phase 1
(VTPh1) was equal (±10 V) to the leading-edge voltage of
phase 2 (VLPh2), the trailing-edge voltage of phase 2
(VTPh2) was equal (±10 V) to the leading-edge voltage of
phase 3 (VLPh3), and the trailing-edge voltage of phase 3
(VTPh3) was equal (±10 V) to the leading-edge voltage of
phase 4 (VLPh4). Also, each phase was opposite in polarity
to its preceding phase. C, Two biphasic waveforms delivered
sequentially through dual-current pathways. All four phases of the two
biphasic waveforms emulated a single 600-µF capacitor
discharge.
) was compared with that for the
dual-current pathways with two sequential 7.5/2.5-ms biphasic waveforms
(Figure 2B
). The single-current pathway waveform emulated a single
150-µF capacitor waveform. With one of the tests for the dual-current
pathway configurations, all four phases of the two biphasic waveforms
emulated a single 150-µF capacitor discharge when the leading-edge
voltage of the first phase was >250 V. When leading-edge voltages of
the first phase were between 150 and 250 V for the 150-µF capacitor
dual-current configuration, the fourth-phase leading-edge voltage was
set at 50 V, which was the lowest possible output of the HVS-02
programmable cardioverter/defibrillator. The four phases of the other
dual-current pathway test configuration emulated a single 600-µF
capacitor discharge (Figure 2C
). The 600-µF capacitor discharge was
created by having a variable resistor (0 to 600
) in series with
the defibrillation electrode/heart circuit; enough resistance was added
to emulate a 600-µF capacitor discharge. The amount of added
resistance was set with the use of a standard ohmmeter.
On arrival at the
electrophysiological laboratory, all
patients were in sustained atrial fibrillation. We determined the order
of testing the configurations by drawing chits. The chits were not
replaced until after 3 patients had been studied successfully, so that
each configuration was tested as the first configuration four times,
second four times, and third four times in 12 of the 14 patients
successfully studied. This was to help control for the effect of
duration of atrial fibrillation before each shock. After a successful
defibrillation, atrial fibrillation was reinduced with rapid atrial
pacing from the right atrial appendage or coronary sinus and
allowed to persist for
1 minute.
, setting the leading
trailing-edge of the first phase to 150 V allowed for adequate
programming of the other phases of the two biphasic waveforms. The ADFT
for each configuration was defined as the lowest-strength shock that
successfully defibrillated the atria.
resistor
in series with the electrodes, and the actual voltage waveform was
recorded with a 200:1, 100-M
resistor divider in parallel with
the electrodes. These data were digitized at 20 000 Hz and
recorded on a waveform analyzer (DATA6100, Data Precision).
The impedance between the electrodes and the total delivered energy of
each phase of the two biphasic waveforms were computed from the current
and voltage waveforms.
Results are expressed as mean±SD unless otherwise specified.
For all statistical tests, a P value
0.05 was considered
significant. Student's t test was used to compare the ADFT
leading-edge voltages, leading-edge currents, and total delivered
energies for different shock configurations. Individual differences in
ADFT for different pathways were tested, controlling for multiple
comparisons, with the Student-Newman-Keuls test. The overall effect of
current-pathway configuration was tested by use of repeated-measures
ANOVA. The correlation between duration of atrial fibrillation and ADFT
was measured by standard parametric coefficient methods and by
Spearman's rank nonparametric correlation coefficient
methods.
![]()
Results
Top
Abstract
Introduction
Methods
Results
Discussion
References
Patient Characteristics
Clinical characteristics of the study population are summarized in
Table 1
. Left atrial size and ejection
fraction were estimated by transthoracic
echocardiography. Echocardiographic
data were obtained within the 2- to 3-month period before
performance of the defibrillation protocol. Table 2
shows the duration of atrial
fibrillation for each patient and the antiarrhythmic medications that
the patients were taking at the time of the protocol. Also, patients
who had undergone transthoracic cardioversion within the 3
months preceding the protocol are shown in Table 2
along with the
electrode location, number of attempted shocks, maximum
transthoracic energy delivered, and internal cardioversion
outcome.
View this table:
[in a new window]
Table 1. Patient Clinical Characteristics
View this table:
[in a new window]
Table 2. Patient Electrophysiological and Cardioversion
Characteristics
). In 2 patients, internal cardioversion was unsuccessful. In 1
patient (patient 6), cardioversion could not be achieved by any of the
three configurations despite a maximum voltage of 500 V. This patient
was successfully cardioverted with a 360-J transthoracic
countershock at the time of the protocol and maintained in sinus rhythm
for 15 hours. In the other unsuccessfully treated patient (patient 7),
the coronary sinus could not be engaged, and thus the
coronary sinus catheter could not be positioned. Internal
cardioversion was not attempted, and the patient was successfully
cardioverted with a 200-J transthoracic countershock and
maintained sinus rhythm for at least 24 hours, when the patient was
discharged from the hospital.
Table 3
shows the ADFT data for
leading-edge voltage, leading-edge current, mean impedance for the
first current pathway, mean impedance for the second current pathway,
and total delivered energy for each electrode configuration. The
single-current pathway had a significantly higher ADFT in terms of
leading-edge voltage, leading-edge current, and total delivered energy
than both dual-current pathways. Also, the ADFT in terms of total
delivered energy was lower for both dual-current pathways than for the
single-current pathway for each individual patient (Figure 3
). When we compared the two dual-current
pathway configurations, there was no significant difference in ADFT in
terms of leading-edge voltage or leading-edge current; however, the
600-µF dual-current pathway had a significantly higher ADFT in terms
of delivered energy than the 150-µF dual-current pathway
configuration. There was no significant difference in mean impedance
among the three electrode configurations for each current pathway.
View this table:
[in a new window]
Table 3. Atrial Defibrillation Threshold Data

View larger version (18K):
[in a new window]
Figure 3. ADFT for each patient. Graph shows relationship of
total delivered energy at ADFT for each shock configuration for each
patient.
. These data are from
the 12 patients who could be successfully defibrillated with an
internal method. Although there appeared to be a trend toward a higher
ADFT for the single-current pathway when it was given first, it was not
statistically significant. Thus, randomization so that each
configuration was the first one given on four separate occasions
controlled for the effect of first shock and for the duration of atrial
fibrillation before the shock. There was no relationship between ADFT
and duration of atrial fibrillation with the randomization scheme used
for these 12 patients. If we used an estimated ADFT of 15 J and
included data for patient 6, who could not be successfully
defibrillated and who had been in atrial fibrillation for >2 years,
standard correlation coefficients comparing the duration of atrial
fibrillation and ADFT total energy were 0.862 for the 150-µF single
pathway, 0.863 for the 150-µF dual-current pathway, and 0.0869 for
the 600-µF dual-current pathway, which were all statistically
significant. However, Spearman's rank correlation coefficients were
0.55 for the 150-µF single pathway, 0.066 for the 150-µF
dual-current pathway, and 0.033 for the 600-µF dual-current pathway,
which were not statistically significant.

View larger version (44K):
[in a new window]
Figure 4. Effect of shock order. Graph shows relationship of
mean±SD of leading-edge voltages at ADFT for each shock configuration
in relation to the order of each shock after randomization. Each
configuration was given first four times. SP 150 indicates
single-pathway 150-µF biphasic shock; DP 150, dual-pathway 150-µF
biphasic shocks; and DP 600, dual-pathway 600-µF biphasic
shocks.
![]()
Discussion
Top
Abstract
Introduction
Methods
Results
Discussion
References
Background
In this study, we compared the cardioversion efficacies of a
single biphasic shock delivered via a single-current defibrillation
pathway with that of sequential shocks delivered via a dual-current
electrode system in humans with atrial fibrillation. We also studied
the effect of capacitance on the sequential-shock, dual-pathway
electrode system with the hope that an electrode and waveform
configuration could be found that halted atrial fibrillation with shock
strengths of low voltage and current. We chose to study these
configurations for several reasons. First, internal ADFTs when a
single-current pathway is used in humans have been shown to be in the
1- to 5-J range for patients with a history of paroxysmal atrial
fibrillation5 6 7 8 9 ; however, these shock strengths
are still in a range that is associated with significant
discomfort.6 12 13 14 15 Second, studies in animals
with acutely induced atrial fibrillation26 have
shown a significant decrease in defibrillation threshold requirements
when dual-current pathways with sequential shocks are used compared
with single-current pathways with single shocks. Third, some clinical
studies have demonstrated that the pain associated with internal atrial
defibrillation shocks is related to the intensity of the
shock,6 9 12 13 15 and it is hypothesized that
pain is more closely related to the leading-edge voltage and current of
the shock than to the shock energy.15 27 By
increasing the capacitance of the sequential shocks, we hypothesized
that the leading-edge voltage and current of the shock would be
reduced, possibly decreasing the pain associated with internal atrial
defibrillation shocks.
Animal studies28 29 30 31 32 33 have demonstrated a
close relationship between the extracellular potential gradient
distribution produced throughout the ventricles by a shock and whether
or not the shock successfully terminates ventricular
fibrillation. For defibrillation to occur, it is thought that a minimal
potential gradient must be generated by the shock throughout most or
all of the ventricular
myocardium.31 32 33 The distribution of
potential gradients in the heart after a shock is very uneven for
electrodes located on or in the heart. There are high-gradient areas
near the electrodes and low-gradient areas farther away from the
electrodes.31 33 These areas of low gradient are
the regions in which earliest activation originates after unsuccessful
defibrillation shocks.31 33 Furthermore,
high-potential gradients can have detrimental effects on the heart,
including postshock arrhythmias,34
conduction disturbances,35 36 myocardial
dysfunction,37 and myocardial
necrosis.38 The potential gradient distribution
created by a shock in the heart depends on several factors, including
electrode size and location.39 The optimal
electrode system for ventricular defibrillation minimizes
the high-gradient areas near the electrodes and raises the critical
amount of ventricular tissue above the minimum gradient to
achieve defibrillation.
Animal and human16 44 45 studies have
demonstrated that ventricular defibrillation energy
requirements are significantly lower for capacitances of <120 µF
than for higher capacitances. Peak voltage and current are greater for
the lower-capacitor waveforms; however, some clinical
studies6 9 12 13 15 16 but not
others14 46 have demonstrated that the pain
associated with internal atrial defibrillation shocks is related to the
intensity of the shock. It has been hypothesized that the pain
associated with an internal shock is in part related to the
leading-edge voltage and/or current instead of the
energy.15 27 Therefore, defibrillation energy
might be minimized with a lower capacitance, but the pain associated
with the shock might be minimized with a higher capacitance. In the
present study, increasing the capacitance of the delivered
sequential biphasic waveforms by a factor of four did not result in a
difference in defibrillation threshold in terms of leading-edge voltage
or leading-edge current. Because of their higher tilt, the
lower-capacitance sequential biphasic shocks defibrillated with a lower
total delivered energy than the higher-capacitance sequential biphasic
shocks.
The safety of internal atrial defibrillation has been addressed in
several previous animal and human studies.4 5 6 7 8
All of those studies demonstrated that as long as the shocks were
synchronized to the ventricular activation, almost no
ventricular arrhythmias were seen. In the
present study, when the shocks were appropriately synchronized to
the R wave, no ventricular arrhythmias were
observed in >100 single- and dual-current pathway shocks. The only
serious complication was induction of ventricular
fibrillation due to inappropriate synchronization from equipment error
and delivery of the shock during the ventricular vulnerable
period. This emphasizes the critical importance of ensuring accurate
synchronization of the shock to the R wave to avoid potentially lethal
postshock ventricular arrhythmias in implantable
devices; if this is not done, back-up ventricular
defibrillation will be needed. There were no long-term complications
and temporary pacing was not required in any of the patients for
postshock conduction delays. No permanent sinus or AV nodal dysfunction
occurred as a complication of the defibrillation protocol.
Several clinical studies have shown the efficacy of biphasic
waveforms through a single-current pathway for internal atrial
fibrillation.5 6 8 9 15 16 27 47 ADFTs, in terms
of mean energy, have varied between these studies from as low as
1.9±1.6 J8 to as high as 9.6±6.6
J.47
1
J were painful, and all patients required sedation for shock strengths
>2 J. The mean ADFT was 9.6±6.6 J (range, 0.5 to 20 J).
Sequential shocks delivered through dual-current pathways resulted
in a marked reduction in ADFT energy requirements.
Increased-capacitance/lower-tilt sequential biphasic shocks did not
result in a significantly lower leading-edge voltage or current than
lower-capacitance/higher-tilt sequential biphasic shocks. Although the
sequential-shock configurations evaluated in the present study were
more efficient for internal atrial defibrillation, threshold levels
were still in the range in which most patients probably feel
significant discomfort without sedation. It is hoped that continued
research into this area will result in even more efficient
defibrillation waveforms and lead systems to help further minimize
discomfort to the patient. A dual-current defibrillation system could
make more patients eligible for an implantable device.
![]()
Footnotes
Reprint requests to Randolph A.S. Cooper, MD, Room B-140 Volker Hall, 1670 University Blvd, University of Alabama at Birmingham, Birmingham, AL 35294-0019.
![]()
References
Top
Abstract
Introduction
Methods
Results
Discussion
References
1.
Kannel WB, Abbott RD, Savage DD, McNamara PM.
Epidemiologic features of chronic atrial fibrillation. N
Engl J Med. 1982;306:10181022.[Abstract]
This article has been cited by other articles:
![]() |
S. J. Walsh, G. Manoharan, O. J. Escalona, J. Santos, N. Evans, J. McC. Anderson, M. Stevenson, J. D. Allen, and A.A. J. Adgey Novel rectangular biphasic and monophasic waveforms delivered by a radiofrequency-powered defibrillator compared with conventional capacitor-based waveforms in transvenous cardioversion of atrial fibrillation. Europace, October 1, 2006; 8(10): 873 - 880. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. R. J. Mitchell, P. A. R. Spurrell, K. Kamalvand, M. Higson, R. Shanmuganathan, N. R. Patel, and N. Sulke What is the optimal electrode configuration for atrial defibrillators in man? Europace, January 1, 2002; 4(1): 41 - 44. [Full Text] [PDF] |
||||
![]() |
P.E. Vardas, E.G. Manios, E.M. Kanoupakis, D.N. Dermitzaki, H.E. Mavrakis, and E.M. Kallergis Atrial defibrillation threshold in humans minutes after atrial fibrillation induction. 'a stitch in time saves nine' Eur. Heart J., September 1, 2001; 22(17): 1613 - 1617. [Abstract] [PDF] |
||||
![]() |
X. Zheng, M. E. Benser, G. P. Walcott, S. D. Girouard, D. L. Rollins, W. M. Smith, and R. E. Ideker Reduction of Atrial Defibrillation Threshold With an Interatrial Septal Electrode Circulation, November 21, 2000; 102(21): 2659 - 2664. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Boriani, M. Biffi, V. Cervi, G. Bronzetti, G. Magagnoli, R. Zannoli, and A. Branzi Evaluation of Myocardial Injury Following Repeated Internal Atrial Shocks by Monitoring Serum Cardiac Troponin I Levels Chest, August 1, 2000; 118(2): 342 - 347. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Santini, C. Pandozi, F. Colivicchi, F. Ammirati, M. Carmela Scianaro, A. Castro, F. Lamberti, and G. Gentilucci Transoesophageal low-energy cardioversion of atrial fibrillation. Results with the oesophageal-right atrial lead configuration Eur. Heart J., May 2, 2000; 21(10): 848 - 855. [Abstract] [PDF] |
||||
![]() |
C. W. Hogue Jr and M. L. Hyder Atrial fibrillation after cardiac operation: risks, mechanisms, and treatment Ann. Thorac. Surg., January 1, 2000; 69(1): 300 - 306. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Timmermans, A. Nabar, L.-M. Rodriguez, G. Ayers, and H. J. J. Wellens Use of Sedation During Cardioversion With the Implantable Atrial Defibrillator Circulation, October 5, 1999; 100(14): 1499 - 1501. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Cooklin, M. R. Olsovsky, R. G. Brockman, S. R. Shorofsky, and M. R. Gold Atrial defibrillation with a transvenous lead: A randomized comparison of active can shocking pathways J. Am. Coll. Cardiol., August 1, 1999; 34(2): 358 - 362. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Timmermans, L.-M. Rodriguez, G. M. Ayers, H. Lambert, J. L. R. M. Smeets, J. W. S. Vlaeyen, A. Albert, and H. J. J. Wellens Effect of Butorphanol Tartrate on Shock-Related Discomfort During Internal Atrial Defibrillation Circulation, April 13, 1999; 99(14): 1837 - 1842. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1998 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |