(Circulation. 1997;96:2016-2021.)
© 1997 American Heart Association, Inc.
Articles |
From the Cardiovascular System Laboratory, The Bruce Rappaport Faculty of Medicine, Technion-Israel Institute of Technology, Haifa, Israel.
Correspondence to Shlomo A. Ben-Haim, MD, DSc, Cardiovascular System Laboratory, The Bruce Rappaport Faculty of Medicine, Technion-Israel Institute of Technology, Efron St, PO Box 9649, 31096 Haifa, Israel. E-mail sol{at}biomed.technion.ac.il
| Abstract |
|---|
|
|
|---|
Methods and Results The NFM system uses an ultralow magnetic field to measure the real-time three-dimensional (3D) location of the tip of the locatable catheter. While in stable contact with the endocardium, between 30 and 40 consecutive tip locations were sampled and used for the 3D reconstruction of the RA geometry. The location of the catheter tip was presented in real time, superimposed over the RA geometry. We selected a point on the 3D reconstruction and delivered RF energy to that site via the tip of the locatable catheter. The catheter was then completely withdrawn and renavigated twice to the same point, at which RF energy was delivered again. At autopsy, the distance between the centers of the three ablation points (mean±SEM) was 2.3±0.5 mm (n=27). Similarly, we used the NFM system to guide the generation of linear lesions. The measured length of the linear lesions on the NFM 3D view was close to the actual lesion length measured at autopsy (correlation coefficient, .96; P=.002; n=6). Furthermore, the location, shape, and continuity of the linear lesions corresponded to the autopsy findings.
Conclusions We conclude that the NFM system can guide the application of RF energy without the use of fluoroscopy in a highly accurate and reproducible manner.
Key Words: electrophysiology arrhythmia catheter ablation mapping
| Introduction |
|---|
|
|
|---|
RF energy is delivered to target sites on the endocardial surface. Selection of target sites is based on a combination of anatomic and electrical criteria. These criteria are different for each type of arrhythmia. Common to all cardiac RF ablation is the use of fluoroscopy to guide the navigation of the ablation catheter to the target site. The use of fluoroscopy for this purpose is problematic for the following reasons. (1) The patient and medical team are exposed to ionizing radiation. (2) The endocardial surface is invisible on the radiograph; therefore, the ablation targets can only be approximated by the relationship between the ablation catheter tip image and that of nearby structures (such as ribs or blood vessels). (3) Sites to which RF energy was previously delivered cannot be recognized later by fluoroscopy; therefore, operators can repeatedly apply RF energy to the same site without noticing. (4) Fluoroscopy generates a two-dimensional representation of cardiac anatomy; therefore, the navigation of the ablation catheter has to rely on repeated multiple-plane views to assess the correct 3D position of the catheter.
Recently, RF ablation has been applied to treatment of atrial flutter and fibrillation.8 9 However, the ablation lesions in these cases are usually more complex than the regular point ablations for other arrhythmias. Linear ablations matching the maze surgical procedure have already been used with some success in eliminating atrial fibrillation in humans10 and dogs.11 12 The difficulties in generating complex lesions stress even more strongly the shortcomings of the current method for guiding RF energy application.
In the present study, we tested the value of a new magnetic 3D NFM system for cardiac ablation. The system uses ultralow magnetic fields to locate a sensor positioned near the tip of a regular mapping and ablation catheter. Recorded data of the catheter location and intracardiac electrogram are reconstructed in real time and presented as a 3D geometric map color-coded with the electrophysiological information. We evaluated the accuracy and reproducibility of the NFM system in guiding both point and elongated lesion generation.
| Methods |
|---|
|
|
|---|
System Components
The navigation and mapping system is composed of a miniature
passive magnetic field sensor, an external ultralow magnetic field
emitter (location pad), and a processing unit (CARTO, Biosense).
The locatable catheter (NAVI-STAR, Cordis-Webster) is similar to a regular electrophysiological 8F deflectable catheter. The tip of the catheter is mounted on the distal end of the shaft and includes the tip and several more proximal electrodes that allow recording of unipolar or bipolar signals. Just proximal to the tip electrode lies the magnetic sensor, totally embedded within the catheter. Magnetic fields emitted from the location pad are received by the sensor and are transmitted along the catheter shaft to the main processing unit.
The locator pad is located beneath the operating table and generates ultralow magnetic fields (5x10-6 to 5x10-5 T) that code the mapping space around the animal's chest with both temporal and spatial distinguishing characteristics. These fields contain the information necessary to resolve the location and orientation of the sensor.
Catheter Mapping and Ablation
Catheter Mapping
Under fluoroscopic guidance, two locatable 8F catheters were
introduced through the appropriate vascular access and positioned
inside the heart chambers. The first catheter was introduced into the
coronary sinus or the right ventricular apex and
served as a reference catheter. A mapping catheter was introduced into
the mapped chamber. The mapping system determined the location and
orientation of both the mapping and reference catheters. The location
of the mapping catheter was gated to a fiducial point in the cardiac
cycle and recorded relative to the location of the fixed reference
catheter at that time, thus compensating for both animal and cardiac
motion. While moving the catheter inside the heart, the system
analyzed its location and presented it to the user,
thus allowing navigation without the use of fluoroscopy. The mapping
catheter was dragged over the endocardium, sequentially acquiring the
location of its tip together with its electrogram while in stable
contact with the endocardium.
The LAT at each site was determined from the unipolar intracardiac
electrogram. We collected unipolar recordings filtered at 0.5
to 240 Hz. The LAT at each site was calculated as the interval between
a fiducial point on the body-surface ECG or intracardiac electrogram
and the steepest negative intrinsic deflection from the unipolar
recording. The stability of the catheter-wall contact was
evaluated at each site by examination of the end-diastolic
stability (the distance in millimeters between two successive
end-diastolic locations) and the LAT stability (measured as
the difference in milliseconds between two successive LATs). A point
was added to the map only if the end-diastolic location
stability was <2 mm and the LAT stability was <2
ms.13 The electrophysiological
information was color-coded and superimposed on the 3D chamber geometry
(Fig 1
). The reconstruction was updated
in real time with acquisition of a new site. The map was
presented as a 3D object in which interpolation of activation
between adjacent sites was allowed only for nearby sites. Whenever the
distance between adjacent sites was higher than a predetermined
threshold (triangle-fill threshold), no interpolation was made and the
sites were connected by a "wire" frame (Fig 1
). We used a
threshold of 25 mm for the figures displayed in this article.
|
Ablation Energy
RF energy was delivered from the distal electrode of the
locatable mapping catheter (4.0 mm), whereas the anode was a
larger patch electrode placed on the animal's back. RF ablation was
performed with a 500-kHz RF generator (RFG-3C; Radionics) in a
temperature-controlled mode of 80°C for 60 seconds. These
corresponded to energy of 20 to 35 W. Impedance measurements served to
determine constant contact with the endocardium and coagulum formation
on the catheter tip (impedance rise >150
).
Protocols
Protocol 1. The aim of this set of experiments was to
perform three distinct point ablations at a single anatomic site.
Subsequent to executing the electroanatomic map of the RA, we selected
an arbitrary site for the point ablations to be performed. We delivered
RF energy to that site three times, and each ablation point was tagged
(Fig 2
). For each RF application, we
navigated the catheter to the target from a site outside the RA. Nine
swine were used in this protocol. The heart was excised, and the
distance between the centers of the ablation sites, as well as the
total area of the combined lesion, was measured.
|
Protocol 2. The objective of this set of experiments was to
perform a line of ablation in the RA. We tested our capability to form
a continuous line consisting of 6 to 12 separate ablation points guided
by the NFM system. Subsequent to executing the electroanatomic map of
the RA, we chose the path on which we planned to deliver RF energy.
Each ablation point was tagged and added to the map as described in the
previous protocol. The catheter was removed after each point ablation
and navigated back to a nearby site on the path, where another burst of
RF energy was applied (Fig 2B
). Six swine were used in this protocol.
We compared the length of the lesion as measured on the NFM system with
that measured at autopsy, as well as the shape and continuity.
Tissue Preparation
Animals were killed and hearts were removed. Special care was
taken to avoid damaging the ablated regions. Distances and areas were
measured on preserved specimens. Tissues were preserved in 4%
formaldehyde for further analysis and micrograph
preparations.
Statistics
Results are reported as mean±SEM. Correlation coefficients were
calculated with the Pearson test. A value of P<.05 was
considered statistically significant.
| Results |
|---|
|
|
|---|
Single Ablation Points
The distances between the centers of the three ablation
points and the area of the combined ablation lesion were measured from
the tissue specimens. Fig 3
presents
the result of a representative point ablation
experiment, the left panel showing the electroanatomic map with the
tagged ablation sites and the right panel the corresponding sites on
the heart tissue. A single RF energy application generated a circular
lesion with an average diameter of 5.0 mm and an area of 19.6
mm2. Table 1
summarizes the
results of nine experiments.
|
|
Linear Ablations
We compared the length of a longitudinal lesion performed on the
heart with the length of the lesion as viewed on the NFM map. Fig 4
presents a typical trial of linear
lesion, the left panel showing the electroanatomic map with a line
composed of several tagged ablation points and the right panel the same
ablation line on the heart tissue. Table 2
summarizes the comparison between the
linear ablations performed with the NFM system on the reconstructed
maps and the actual lines performed on the heart tissue. The
correlation coefficient between the length of intended and actually
performed lines was .96, with P=.002 (n=6).
|
|
| Discussion |
|---|
|
|
|---|
In previous studies, we evaluated the accuracy of mapping using the NFM system.13 In those studies, we showed that the location system used in the NFM system has a static in vitro accuracy of 0.2 mm and an in vivo accuracy of 0.9 mm. Similar studies have been performed in humans, with similar results.14 The reconstruction procedure used by the NFM system to calculate the heart chamber volumes has been shown to be accurate.15
Mapping Procedure
The mapping procedure was relatively short and created the
electroanatomic "road map" necessary to guide the ablation
procedure described in this article. Nevertheless, this mapping
procedure was sequential and required that the heart beat with the same
rate and rhythm throughout the mapping procedure.
We evaluated the reproducibility of navigating the ablation catheter to a site on the endocardium using the NFM system. We found that a site identified by the electroanatomic map of this system corresponded to a site on the heart, as indicated by the relatively small distances between the centers of the repeated ablation points. The system also assisted in the guidance of RF energy delivery for longitudinal lesions to the RA. We showed that we could design a line of ablation on the viewing station and deliver the RF energy accordingly to form the desired lesion on the endocardium. The more experienced we became, the shorter the procedure became, together with an overall improvement of accuracy. No fluoroscopy was used to guide the mapping or ablation procedure except for that required for acquiring the first three boundary points.
Although not directly comparable, the average reported fluoroscopy time for RA mapping and ablation procedures is generally between 15.2 and 63 minutes.16 In the present study, we clearly showed that no significant x-ray exposure was needed either to map or to ablate the RA.
Ablation in the RA
Navigation of the catheter to the site of origin of a focal
arrhythmia or to a certain anatomic location at which the RF
energy should be delivered is easily accomplished with the NFM system.
We found that the distances between the centers of the three individual
ablation points were very small, reaching almost half the length of the
ablation catheter-tip electrode. In addition, the ability to visualize
the catheter tip in real time on the monitor screen (Fig 2
) allowed
selection of the desired tip orientation during RF energy delivery. The
orientation of the catheter tip is important for precise determination
of the lesion size and shape. Catheter tip orientation is also
important for determining temperature rise during RF delivery. In their
study, McRury et al17 recorded up to 12°C divergence
from the attempted temperature, provided that the catheter orientation
was at an angle of
45°.
There are no published data on the accuracy of RF ablation procedures using fluoroscopic guidance. The only published studies of the accuracy of RF ablation used intracardiac echocardiography.18 19 However, one may judge the value of the NFM system to guide the delivery of RF energy by reviewing the reported number of RF applications per cardiac ablation procedure. Usually, between 3 and 20 RF applications are delivered during a standard RF ablation of a single supraventricular site.16 There are several reasons for repeated RF applications, including inappropriate detection of the target and insufficient energy delivery. The NFM system's systematic "bookkeeping" of all electrograms and previous ablation sites may decrease the number of RF energy applications.
Clinical Relevance
For AV nodal reentry tachycardias, AV reentry
tachycardias, and atrial tachycardias, ablation
therapy is the treatment of choice.20 Recent studies
report a high success rate (>90%) when the current methods and
devices for ablation treatment of these arrhythmias are
used.6 More recently, ablation treatments for atrial
flutter and fibrillation are being developed. Currently, the success
rate for ablation of these indications is <90%. One of the
differences between the ablation procedures for treatment of atrial
flutter and fibrillation is the need to induce a continuous line of
block. The NFM system can guide these catheter ablation procedures
using the real-time navigation on top of an electroanatomic map of the
chamber and tagging sites at which RF energy was delivered.
Limitations of the NFM System
The NFM mapping concept is based on a beat-by-beat approach. This
approach, although highly accurate, is time-consuming, because the
mapping catheter has to be dragged to each new site and positioned for
acquiring the information of a new point. To accurately perform this
process, one must confirm that the heart is beating at the same rate
and rhythm before including the information from each new site.
Furthermore, if the patient develops a new arrhythmia, the
entire electroanatomic map should be acquired, because both the exact
chamber anatomy and its activation sequence are changed. This
limitation is partially overcome by using an "auto-map" mode, in
which the mapping system automatically accepts information from new
sites when certain stability criteria are met.
As reported in previous studies, the mapping system has an error in determining the location of the catheter tip.13 This error may result from respiratory movements not fully corrected by the reference catheter as well as from animal body movements. The overall effect of these errors on repositioning the catheter was found in the current study to be <2 mm. This error corresponds with <50% of the single-ablation lesion size.
Conclusions
Our studies performed in the swine RA show that the NFM system
produces reconstruction of the atrium that allows highly accurate
guidance for catheter ablation procedures. The short electroanatomic
mapping procedure, the precise 3D navigation, and the reproducibility
of catheter-tip repositioning suggest that NFM guidance for RF ablation
may be better than fluoroscopy. Further clinical studies are needed to
assess the value of the NFM system in treating arrhythmia
patients.
| Selected Abbreviations and Acronyms |
|---|
|
| Acknowledgments |
|---|
Received November 13, 1996; revision received March 12, 1997; accepted March 30, 1997.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
C. M. Tracy, M. Akhtar, J. P. DiMarco, D. L. Packer, H. H. Weitz, M. A. Creager, D. R. Holmes Jr, G. Merli, G. P. Rodgers, C. M. Tracy, et al. American College of Cardiology/American Heart Association 2006 Update of the Clinical Competence Statement on Invasive Electrophysiology Studies, Catheter Ablation, and Cardioversion: A Report of the American College of Cardiology/American Heart Association/American College of Physicians Task Force on Clinical Competence and Training Developed in Collaboration With the Heart Rhythm Society J. Am. Coll. Cardiol., October 3, 2006; 48(7): 1503 - 1517. [Full Text] [PDF] |
||||
![]() |
F. Grothues, O. Wolfram, C. Fantoni, H. Boenigk, A. Götte, C. Tempelmann, H.U. Klein, and A. Auricchio Volume measurement by CARTOTM compared with cardiac magnetic resonance. Europace, January 1, 2006; 8(1): 37 - 41. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. Nielsen, H. Kottkamp, C. Piorkowski, J.-H. Gerds-Li, H. Tanner, and G. Hindricks Radiofrequency ablation in children and adolescents: results in 154 consecutive patients. Europace, January 1, 2006; 8(5): 323 - 329. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Gaita, R. Riccardi, D. Caponi, D. Shah, L. Garberoglio, L. Vivalda, A. Dulio, A. Chiecchio, E. Manasse, and R. Gallotti Linear Cryoablation of the Left Atrium Versus Pulmonary Vein Cryoisolation in Patients With Permanent Atrial Fibrillation and Valvular Heart Disease: Correlation of Electroanatomic Mapping and Long-Term Clinical Results Circulation, January 18, 2005; 111(2): 136 - 142. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Arenal, S. del Castillo, E. Gonzalez-Torrecilla, F. Atienza, M. Ortiz, J. Jimenez, A. Puchol, J. Garcia, and J. Almendral Tachycardia-Related Channel in the Scar Tissue in Patients With Sustained Monomorphic Ventricular Tachycardias: Influence of the Voltage Scar Definition Circulation, October 26, 2004; 110(17): 2568 - 2574. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Nademanee, J. McKenzie, E. Kosar, M. Schwab, B. Sunsaneewitayakul, T. Vasavakul, C. Khunnawat, and T. Ngarmukos A new approach for catheter ablation of atrial fibrillation: mapping of the electrophysiologic substrate J. Am. Coll. Cardiol., June 2, 2004; 43(11): 2044 - 2053. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Dickfeld, H. Calkins, M. Zviman, R. Kato, G. Meininger, L. Lickfett, R. Berger, H. Halperin, and S. B. Solomon Anatomic Stereotactic Catheter Ablation on Three-Dimensional Magnetic Resonance Images in Real Time Circulation, November 11, 2003; 108(19): 2407 - 2413. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Ernst, F. Ouyang, F. Lober, M. Antz, and K.-H. Kuck Catheter-induced linear lesions in theleft atrium in patients with atrial fibrillation: An electroanatomic study J. Am. Coll. Cardiol., October 1, 2003; 42(7): 1271 - 1282. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Arenal, E. Glez-Torrecilla, M. Ortiz, J. Villacastin, J. Fdez-Portales, E. Sousa, S. del Castillo, L. Perez de Isla, J. Jimenez, and J. Almendral Ablation of electrograms with an isolated, delayed component as treatment of unmappable monomorphic ventricular tachycardias in patients with structural heart disease J. Am. Coll. Cardiol., January 1, 2003; 41(1): 81 - 92. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Kornowski Left ventricular electromechanical mapping for determination of myocardial function and viability J. Am. Coll. Cardiol., September 18, 2002; 40(6): 1075 - 1078. [Full Text] [PDF] |
||||
![]() |
J. Lessick, G. Hayam, A. Zaretsky, S. A. Reisner, Y. Schwartz, and S. A. Ben-Haim Evaluation of inotropic changes in ventricular function by NOGA mapping: comparison with echocardiography J Appl Physiol, August 1, 2002; 93(2): 418 - 426. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Ricard, M. Imianitoff, K. Yaici, J. M. Coutelour, M. Bergonzi, J. P. Rinaldi, and N. Saoudi Atypical atrial flutters Europace, January 1, 2002; 4(3): 229 - 239. [Abstract] [PDF] |
||||
![]() |
P. Adragao, D. Cavaco, C. Aguiar, J. Palos, F. Morgado, R. Ribeiras, M. Abecasis, J. Neves, D. Bonhorst, and R. Seabra-Gomes Ablation of pulmonary vein foci for the treatment of atrial fibrillation. Percutaneous electroanatomical guided approach Europace, January 1, 2002; 4(4): 391 - 399. [Abstract] [PDF] |
||||
![]() |
N Saoudi, F Cosio, A Waldo, S.A Chen, Y Iesaka, M Lesh, S Saksena, J Salerno, and W Schoels A classification of atrial flutter and regular atrial tachycardia according to electrophysiological mechanisms and anatomical bases. A Statement from a Joint Expert Group from the Working Group of Arrhythmias of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology Eur. Heart J., July 2, 2001; 22(14): 1162 - 1182. [PDF] |
||||
![]() |
C. Reithmann, E. Hoffmann, U. Dorwarth, T. Remp, and G. Steinbeck Electroanatomical mapping for visualization of atrial activation in patients with incisional atrial tachycardias Eur. Heart J., February 1, 2001; 22(3): 237 - 246. [Abstract] [PDF] |
||||
![]() |
C. M. Tracy, M. Akhtar, J. P. DiMarco, D. L. Packer, H. H. Weitz, W. L. Winters, J. L. Achord, A. W. Boone, J. W. Hirshfeld Jr, B. H. Lorell, et al. American College of Cardiology/American Heart Association Clinical Competence Statement on invasive electrophysiology studies, catheter ablation, and cardioversion: A report of the american college of cardiology/american heart association/american college of physicians-american society of internal medicine task force on clinical competence J. Am. Coll. Cardiol., November 1, 2000; 36(5): 1725 - 1736. [Full Text] [PDF] |
||||
![]() |
C. M. Tracy, M. Akhtar, J. P. DiMarco, D. L. Packer, H. H. Weitz, W. L. Winters, J. L. Achord, A. W. Boone, J. W. Hirshfeld Jr, B. H. Lorell, et al. American College of Cardiology/American Heart Association Clinical Competence Statement on Invasive Electrophysiology Studies, Catheter Ablation, and Cardioversion : A Report of the American College of Cardiology/American Heart Association/American College of Physicians-American Society of Internal Medicine Task Force on Clinical Competence Circulation, October 31, 2000; 102(18): 2309 - 2320. [Full Text] [PDF] |
||||
![]() |
F. E. Marchlinski, D. J. Callans, C. D. Gottlieb, and E. Zado Linear Ablation Lesions for Control of Unmappable Ventricular Tachycardia in Patients With Ischemic and Nonischemic Cardiomyopathy Circulation, March 21, 2000; 101(11): 1288 - 1296. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. A. Strickberger, B. P. Knight, G. F. Michaud, F. Pelosi, and F. Morady Mapping and ablation of ventricular tachycardia guided by virtual electrograms using a noncontact, computerized mapping system J. Am. Coll. Cardiol., February 1, 2000; 35(2): 414 - 421. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Bhargava and R. Kornowski Biosense Left Ventricular Electromechanical Mapping Asian Cardiovasc Thorac Ann, December 1, 1999; 7(4): 345 - 348. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Pappone, G. Oreto, F. Lamberti, G. Vicedomini, M. L. Loricchio, S. Shpun, M. Rillo, M. P. Calabro, A. Conversano, S. A. Ben-Haim, et al. Catheter Ablation of Paroxysmal Atrial Fibrillation Using a 3D Mapping System Circulation, September 14, 1999; 100(11): 1203 - 1208. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Gepstein, G. Hayam, S. Shpun, D. Cohen, and S. A. Ben-Haim Atrial Linear Ablations in Pigs : Chronic Effects on Atrial Electrophysiology and Pathology Circulation, July 27, 1999; 100(4): 419 - 426. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Gepstein, A. Goldin, J. Lessick, G. Hayam, S. Shpun, Y. Schwartz, G. Hakim, R. Shofty, A. Turgeman, D. Kirshenbaum, et al. Electromechanical Characterization of Chronic Myocardial Infarction in the Canine Coronary Occlusion Model Circulation, November 10, 1998; 98(19): 2055 - 2064. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Gepstein, G. Hayam, and S. A. Ben-Haim Activation-Repolarization Coupling in the Normal Swine Endocardium Circulation, December 2, 1997; 96(11): 4036 - 4043. [Abstract] [Full Text] |
||||
![]() |
C. de Chillou, D. Lacroix, D. Klug, I. Magnin-Poull, C. Marquie, M. Messier, M. Andronache, C. Kouakam, N. Sadoul, J. Chen, et al. Isthmus Characteristics of Reentrant Ventricular Tachycardia After Myocardial Infarction Circulation, February 12, 2002; 105(6): 726 - 731. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||