(Circulation. 1997;96:801-808.)
© 1997 American Heart Association, Inc.
Articles |
From the Department of Medicine (M.L.K., P.D.G., K.L.H., J.V.T.), The University of Calgary, The Foothills Hospital, Calgary, Alberta, Canada, and The Technion-Israel Institute of Technology (R.B.), Haifa, Israel.
Correspondence to Rafael Beyar, MD, DSc, Professor and Head, Heart System Research Center, Department of Biomedical Engineering, Technion-IIT, Haifa 32000, Israel. E-mail rafael{at}biomed.technion.ac.il
| Abstract |
|---|
|
|
|---|
Methods and Results Among patients with no previous infarction, five had no collaterals (group A) and six had angiographically visible collaterals (group B). Previous anterior infarction was present in nine patients (group C). Data were acquired with the LAD angioplasty wire passed beyond the apex using a view aligned with the LV long axis. Frame-by-frame dynamics of apex rotation were measured from the angular movement of the portion of the wire that traversed the apex. Aortic pressure recordings allowed precise temporal definition of the cardiac cycle. Dynamics of apex rotation were measured at fixed intervals until 60 seconds of occlusion and up to 60 seconds of reperfusion. In group A, counterclockwise apex rotation (twist) during ejection of -22.0±1.7° (mean±SEE) was followed by rapid clockwise rotation (untwist) during isovolumic relaxation. During 60 seconds of ischemia, maximum apex rotation decreased to -8.2±2.0° (P<.001 versus baseline). In group B, baseline apex rotation was similar (-26.2±6.9°) to that in group A, but ischemia had less effect, with apex rotation values of -17.7±3.4° (P<.05 versus group A values). Group C was characterized by reduced baseline apex rotation values (-9.7±3.1°, P<.05 versus group A values), with little change observed during ischemia (-8.1±2.6°).
Conclusions Apex rotation, an index of ventricular twist, is sensitive to acute ischemia in patients without previous myocardial infarction. Visible collaterals to the ischemic region attenuate the acute ischemic response at 60 seconds. Previous myocardial infarction causes abnormalities in the baseline twist pattern with no further deterioration at 60 seconds of ischemia.
Key Words: angioplasty myocardial infarction ischemia collateral circulation
| Introduction |
|---|
|
|
|---|
Studies have been performed to examine and quantify LV twist in both dogs and humans. Lower9 was the first to observe that the muscle fibers of the LV inner wall ran opposite to those of the outer wall and that contraction could be compared "...with the wringing of a linen cloth to squeeze out the water." More than 300 years later, other studies confirmed this early observation. Using multiple implanted markers and biplane cine angiography in transplanted hearts, Ingels et al2 10 and Hansen et al7 11 calculated torsional deformation from the three-dimensional motion of the markers throughout systole. Using similar techniques, Waldman et al12 measured regional transmural deformation in the LV wall of instrumented dogs. Arts et al4 studied torsional deformation in the canine LV using two-dimensional echocardiography. By using echo-derived short-axis images of the LV at the levels of the mitral valve and the papillary muscles, twist was calculated as the difference between the angles of rotation at these two levels. Later, MRI and end-diastolic tagging were used to measure ventricular twist in patients.8
Beyar et al5 used radiopaque markers and biplane cine angiography to follow the dynamic twist-radial shortening relationship through the entire cardiac cycle and demonstrated rapid untwisting during the isovolumic relaxation phase. The fact that most of the untwisting occurs before the end of isovolumic relaxation has been confirmed by MRI tagging.13 The method described by Beyar et al5 was recently extended to hearts with multiple radiographic markers in patients undergoing heart transplantation, which showed that changes in the pattern of rapid untwisting can be used as a highly sensitive indicator of impending rejection14 and that twist is a reliable measurement of LV function.
Previous studies using MRI tagging8 have shown that slice rotation increases gradually from base to apex and that the base of the heart rotates only minimally. Based on that observation and on direct measurement of base rotation, we have recently shown that apex rotation alone, studied by an optical method in dogs, can provide a reliable index of LV twist,15 which is markedly sensitive to acute ischemia in an experimental model.16
The effects of acute ischemia on LV twist have not been studied
in human patients. It is hypothesized that ischemic dysfunction
will result in marked alteration of twist amplitude and dynamics,
similar to our observation in dogs.16 Measurement of apex
rotation is feasible during coronary angiography through
observation of the apical segment of the LAD by use of an apical
long-axis view.17 Similarly, during angioplasty to the
LAD, it is possible to measure the dynamics of apex rotation from this
view by observing the motion of an angioplasty wire positioned in the
apical segment of the LAD artery (Fig 1
). Using this
approach, we measured the dynamics of apex rotation in patients
undergoing angioplasty to the LAD as a human model of acute
ischemia. The aims of the study were (1) to describe the
dynamic pattern of apex rotation in patients, (2) to show the time
course of the development and resolution of ischemic
dysfunction as reflected by the measurement of apex rotation, (3) to
show the relation between abnormalities in the dynamic pattern of twist
and the presence of collateral flow, and (4) to study whether previous
anterior MI affects the baseline twist and its response to acute
ischemia.
|
| Methods |
|---|
|
|
|---|
|
Protocol
The protocol was approved by the Conjoint Research Ethics Board
of The University of Calgary and The Foothills Hospital, and all
patients gave written informed consent. The ECG and aortic pressure
were measured and recorded (VR16, Electronics for Medicine;
Honeywell). An 8F sheath was introduced via a femoral
arterial puncture, and the routine over-the-wire
angioplasty procedure was followed. With the use of a
diagnostic catheter, images of the LAD lesion were obtained
in the most appropriate radiographic views. In addition,
the apical view aligned along the long axis of the LV was selected and
recorded. This view was selected so that the apical segment of the
LAD appeared in the middle of the LV base, as defined by the course of
the coronary sinus (visualized during the venous phase of the
injection) and the position of the septum (visualized by the
arterial blushing) (Fig 1
). An 8F angioplasty catheter was
advanced and engaged in the left main coronary artery, and the
angioplasty wire was advanced until it traversed the apex. A lateral
view (Figs 1
and 2
) was used to verify that the wire
extended beyond the apex. Subsequently, a baseline measurement of apex
rotation was obtained, and the angioplasty procedure proceeded. After a
satisfactory result had been obtained (as determined by the PTCA
operator), timed angiographic (30 frames/s) and
hemodynamic measurements of cardiac cycles were
obtained for baseline and at 20 and 60 seconds of ischemia
(induced by balloon inflation to a maximum of 14 atm). After 60 seconds
of occlusion, the balloon was deflated, and measurements were taken at
20 and 60 seconds of reperfusion.
|
Measurements of Apex Rotation
Images were digitized frame by frame into a computer by use of a
frame grabber (Bioscan Snapshot Plus; INFRASCAN, Inc) and stored for
further analysis. The phases of the cardiac cycle were defined
from the ECG, and central aortic pressure was corrected for
transmission delay.
Timing
ED was defined as the peak of the QRS deflection in lead II, and
the beginning of ejection was defined by the rapid increase in aortic
pressure. The maximum value of systolic twist was measured, and
the time at which it occurred was recorded (the number of
milliseconds before the end of ejection). ES was defined by the
dicrotic notch. The end of isovolumic relaxation was approximated as
the third frame after the dicrotic notch. This approximation was based
on the calculation of the time from ES (frame with the smallest volume)
to mitral valve opening, which was obtained using the LV angiogram
taken before the PTCA.
Angle Selection
As shown schematically in Fig 1
, this view was selected so that
the apical segment of the LAD appeared in the middle of the LV base, as
defined by the course of the coronary sinus (visualized during
the venous phase of the injection) and the position of the septum
(visualized by the arterial blushing). An angioplasty
catheter was advanced and engaged in the left main coronary
artery, and the angioplasty wire was advanced until it traversed the
apex. A right anterior oblique view (Figs 1A
and 2A
) was used to
confirm that the wire position was over the LV apex. The apical view
that is used for the analysis is shown in Fig 2B
. The segment
of the apical wire to be measured when positioned at the LV apex was
determined by use of both the lateral and apical views. A portion of
the wire that traversed the apex but was not bent by the tortuosity of
the LAD was selected.
Angular Rotation
From the apical view described above, the angle of the wire was
measured for each frame throughout the cardiac cycle. In each frame,
two points at the apical location of the wire were marked, and the
angle of the wire was calculated. The procedure was repeated frame by
frame until at least two complete cycles were digitized. After
obtaining absolute angular measurements, the value of the apex rotation
at ED was arbitrarily defined as 0°, and analysis of rotation
throughout the cycle in relation to the end-diastolic
angular position was obtained. Intraobserver variability was
ascertained by having the same operator trace the same sequence of
images days apart. The variability was very small, with very few
variations between different digitization sessions. Similarly,
interobserver variability was determined by having a research nurse
blindly repeat the digitization process. Interobserver variability was
also of minor value.
Statistics
Averages, SDs, and SEEs were calculated for each group.
Comparisons between groups were performed using the Student
t test, and P<.05 was taken to be significant.
Multivariate ANOVA was applied to take into account
correlations in the data arising from repeated measurements of
individuals. To improve power, attention was focused on the examination
of particular linear combinations, in particular the average twist
value, which is simply the average of values: maximum systolic
twist, ES, and the end of the isovolumic relaxation period. Although
this is somewhat simplistic, detailed examination of the more complex
patterns over the cycle would require more subjects to provide adequate
power. Within- and between-group comparisons were tested against the
relevant components of the pooled error variance/covariance
matrix. For multiple comparisons, the Tukey procedure was used where
feasible; otherwise, Bonferroni corrections were applied. When
available, "exact" probability values are reported; otherwise,
statistical significance (P<.05) was noted. Statistical
analysis was performed with the use of the SAS system, version
6.09 (SAS Institute Inc), on a Sun Sparc2 workstation.
| Results |
|---|
|
|
|---|
|
|
|
After 20 seconds of occlusion, twisting during ejection was markedly attenuated (the maximum value was -9.1±2.7°), and no significant untwisting occurred during the isovolumic relaxation period. Initial twisting during ejection was followed by untwisting at mid ejection and then further twisting during the isovolumic relaxation period, in contrast to the simpler "sawtooth" pattern observed during the control state.
After 60 seconds of occlusion, twisting during ejection was further
attenuated (-8.2±2.0°) and, in some individuals (see Fig 3B
), was
followed by a short twisting motion during the isovolumic relaxation
period. This fully dysfunctional pattern did not change as
ischemia continued (not shown). On release of balloon
occlusion, as little as 20 seconds was required for almost complete
recovery of the systolic twist pattern, with some abnormalities
in the isovolumic relaxation pattern. Overall, these differences
between conditions were statistically significant (P<.005).
A significant difference was present between occlusion and baseline
(P<.001) and occlusion and reperfusion
(P<.05).
Group B (Visible Collateral Vessels, No Previous MI)
The baseline twist, ischemia, and reperfusion results are
presented for a typical case subject in Fig 5
.
Note that the baseline twist pattern is similar to the baseline pattern
presented in Fig 3A
. However, 20 seconds of ischemia
diminished systolic twisting only minimally. The average
results for group B are shown in Fig 6
and given in
Table 2
. Note again that the magnitude and pattern of twist at baseline
are similar to those of group A; however, twist at 60 seconds of
occlusion showed a smaller decrease and was greater than at the
corresponding time in group A. There was a trend toward a decrease in
twist with occlusion for this group (P=.08). Therefore, the
protective effect of the collateral circulation on myocardial function
is clearly manifested by the altered twist response during acute
ischemia.
|
|
Group C (Previous MI)
This group is heterogeneous in terms of the extent of
the MI and the interval that elapsed before angioplasty (Table 1
), as
well as in terms of the presence or absence of collaterals. Examples of
two cases from this group are shown in Fig 7
. Note that
baseline abnormalities in the twist pattern characterize both patients.
In one patient (Fig 7A
), occlusion had a small effect on the twist
amplitude and waveform. In the other example (Fig 7B
), 20 seconds of
ischemia produced a dyskinetic pattern that disappeared after
60 seconds of ischemia with development of complete anterior
wall dysfunction. The average results for group C are presented
in Fig 8
and Table 2
. Note that the baseline twist
pattern is abnormal. Reduced twist as well as no untwisting before the
end of isovolumic relaxation is apparent. Overall, there were no
significant effects of occlusion or reperfusion on the twist magnitude
(P=.24).
|
|
Comparison of Systolic Twist for the Three Groups
The end-systolic twist values for the three groups are
summarized in Table 2
. There were significant differences among the
three groups (P=.038 by Wilks' lambda) under the three
conditions (baseline, 60-second occlusion, and 60-second reperfusion).
Individual examination of the three conditions revealed significant
differences between groups at baseline (P=.034, F test for
1-way ANOVA), with a significant difference between the group with
previous MI (group C) and the group with collaterals (group B) (Tukey
multiple comparisons test). Similar results were obtained under
occlusion (P=.04 for the F test and group C versus group B,
different by the Tukey test). Examination of the groups after
reperfusion revealed no statistically significant difference
(P=.13), although because the comparisons lacked power, it
would be erroneous to conclude that the groups were the same under this
condition. Therefore, the statistical analysis supports the
concept that under conditions of no collateral protection,
ischemia causes a marked change in twist whereas collaterals
provide partial protection, reflected in twist. In addition, baseline
abnormalities exist in patients with previous MI.
| Discussion |
|---|
|
|
|---|
Our findings of twist dynamics for the normally functioning LV are consistent with our earlier reports using apex rotation measurements15 18 as well as with our previous measurements by multiple markers,5 all of which demonstrated rapid untwisting during early diastole. Similar findings have been reported with the use of tagged MRI in patients13 and with the use of angiographic markers in the human transplanted heart.14 19 The magnitude of twist reported in the present study is also consistent with previous results (ie, 13°18 and 15°15 ).
The assumption behind the current method of twist measurement is that the rotation of the base is minimal and can be ignored. This assumption has been validated in our previous work in open-chest, open-pericardium dogs15 and is also substantiated by the results of a previous study that used MRI tagging techniques.8 Because the base rotates opposite to the apex to some extent,5 15 our measurements of apex rotation are a slight underestimation of the true rotation of the apex relative to the base. In contrast, the current method measures twist at its maximum point, ie, the apex, whereas previous methods did not usually measure true apical twist. Interestingly, our apex rotation measurements are similar to previously reported twist values measured by rotation of the apex relative to the base. Using multiple markers and biplane cineangiography, Hansen et al11 reported a value of 13.3±6.0°, and using MRI tagging, Buchalter et al8 reported a value of 11.2±1.3°. Therefore, we conclude that the error introduced by not accounting for base rotation is minimal.
The effects of acute ischemia on twist in patients have not
been evaluated before. We show here that apex rotation is highly
sensitive to ischemia and becomes abnormal within a few seconds
after balloon occlusion.20 Early abnormalities include
twist attenuation with a reversal of twist during mid systole, with
untwisting during the second part of ejection and, subsequently,
twisting during isovolumic relaxation (Figs 3B
and 7B
). This pattern,
which we have also observed during acute ischemia in the dog
(using our optical device to measure apex
rotation),15 16 21 is comparable to an early dyskinetic
motion observed by Tyberg et al in hypoxic papillary
muscles22 and in an experimental model of regional
ischemia.23 This pattern probably results due to
late contraction of the ischemic region as
ventricular pressure is decreasing. When ischemia
is fully developed, usually within 60 seconds of the occlusion, the
dominant pattern is one of limited or absent twisting during ejection
and paradoxical twisting during isovolumic relaxation (Fig 3B
).
Obviously, our present observations are limited to acute occlusion
of the LAD, which results in localized anterior ischemia. How
ischemia of different territories would affect apical rotation
has not been evaluated in patients. However, our observations in dogs
suggest that circumflex occlusion produces significant, perhaps
comparable effects on apical twist.16 21 They suggest that
apex rotation is a parameter that integrates the
performance of all regions of the LV. Nonetheless, it seems
clear that during the procedure, the current method can provide the
operator with important information related both to the amount of
dysfunction caused by balloon occlusion and particularly to the
recovery of function after deflation of the balloon.
Because LV twist has been proven as an index of ventricular systolic function as well as an index sensitive to diastolic dysfunction, it is important to compare the results of the current study with previous studies assessing systolic and diastolic function during angioplasty. Various studies have shown a reduction in global systolic function during angioplasty. Using various techniques, it has been shown that ejection fraction decreases reversibly.24 25 26 27 In most of these studies, LV function was measured at between 30 and 50 seconds of ischemia, and after such ischemic intervals, ischemic effects resolved completely within 5 to 15 minutes of reperfusion. Regional function has been measured by left ventriculography at 30 to 50 seconds of occlusion, and clear dyskinesis of the anterior and apical segments has been demonstrated.27 Transthoracic echocardiography has shown rapid changes in regional function at 16 to 20 seconds of ischemia and resolution within 10 to 20 seconds of reperfusion.28 Transesophageal echocardiography has shown that regional dysfunction develops with 10 seconds of coronary artery occlusion.29 These observations are consistent with the results of the present study showing that at 20 seconds of ischemia, severe dysfunction is present in most cases, and no further deterioration in function is seen at 60 seconds of ischemia. In the present study, recovery of systolic function was indeed immediate, with almost complete recovery being observed after 20 seconds of reperfusion. This time sequence of events is also supported by investigations that used Doppler indexes of aortic flow and showed decreased contractile function during occlusion and complete recovery after reperfusion.30
Diastolic dysfunction has also been reported after angioplasty. An increase in the time constant of relaxation was consistently observed during balloon occlusion.27 Diastolic abnormalities have also been documented by measuring a shift in the end-diastolic pressure-volume relationship, which, in contrast to systolic indexes, persisted 12 minutes after reperfusion.31 Additional evidence for delayed normalization of the diastolic abnormalities was provided by Doppler mitral-inflow measurements,32 which showed persistent diastolic dysfunction at 60 seconds of reperfusion. Other studies using radionuclide techniques have shown reduced LV filling that persisted at 5 minutes after reperfusion. Therefore, it seems that diastolic dysfunction is typical for acute ischemia and may persist longer than systolic dysfunction. Indeed, in the current study, we have shown that an abnormal relaxation pattern was still present at 20 seconds of reperfusion in some patients, although systolic function had recovered completely, consistent with previous observations.
During angioplasty, the effects of collaterals on ventricular function have been studied by use of Doppler-measured aortic-root acceleration.33 It was shown that in the presence of visible collaterals, no decrease in contractile function was measured compared with a group of patients without collaterals. This finding is consistent with our observations that the effect of balloon occlusion on ventricular twist is markedly attenuated in the presence of visible collaterals. An important observation is that the presence of collaterals markedly attenuates the ischemia-induced decrease in twisting during balloon angioplasty. The presence of a normal baseline twist pattern that is minimally attenuated during balloon occlusion indicates that the collateral circulation is sufficient to limit regional dysfunction during the short ischemic periods studied here.
Previous MI had a marked effect on the averaged baseline twist pattern, with only a small additional effect being observed during acute occlusion. Interestingly, some patients showed a pattern of no twist at all during ejection, with no effect of occlusion, whereas other patients with reduced baseline twist demonstrated additional ischemic effects on balloon inflation. In the entire group of patients with previous MI, a tendency toward increased postreperfusion twist may indicate that reflex mechanisms may play a role in increasing regional function after reperfusion.
The present study emphasizes the tool as a research modality allowing measurements of dynamic systolic parameters of ventricular function in patients during rapid ischemic maneuvers. This is the first study of its kind and can form the basis for studies of various interventions on twist dynamics. Validations for the use of apex rotation for twist measurements are based on our previous animal experiments.15 16 The method can also be used in the catheterization laboratory as a quick way to assess ventricular function during LAD angioplasty or even during left coronary injections of contrast material. However, assessing the value of such a method as a clinical tool requires further investigation and feasibility studies.
In summary, a method has been demonstrated whereby the systolic and diastolic dynamics of LV twist can be evaluated during angioplasty from the angular motion of the apical portion of the angioplasty wire. We have shown a typical baseline pattern of ejection twisting and rapid isovolumic-relaxation untwisting, consistent with earlier observations. This twist pattern is highly sensitive to ischemia, with characteristic changes appearing a few seconds after balloon occlusion and disappearing immediately after reperfusion. In addition, collateral protection has been shown to reduce twist abnormalities during ischemia. In the presence of previous MI, baseline twist abnormalities reflect the permanent insult to the myocardium, but an increase in twist after reperfusion may reflect a secondary postischemic inotropic response of viable myocardium in that territory. Therefore, measurements of apex rotation during angioplasty provide a new window for observations of pathophysiological twist dynamics in patients and help us to understand the effects of ischemia and regional dysfunction on twist.
| Selected Abbreviations and Acronyms |
|---|
|
| Acknowledgments |
|---|
Received December 9, 1996; revision received February 18, 1997; accepted February 24, 1997.
| References |
|---|
|
|
|---|
2. Ingels NB, Daughters GT, Stinson EB, Alderman EL. Measurements of midwall myocardial dynamics in intact man by radiography of surgically implanted markers. Circulation. 1975;52:859-867.
3. Arts T, Reneman RS. Measurement of deformation of canine epicardium in vivo during cardiac cycle. Am J Physiol.. 1980;239:H432-H437.
4. Arts T, Meerbaum S, Reneman RS, Corday E. Torsion of the left ventricle during the ejection phase in the intact dog. Cardiovasc Res. 1984;18:183-193.
5. Beyar R, Yin FCP, Hausknecht M, Weisfeldt ML, Kass DA. Dependence of left ventricular twist-radial shortening relations on cardiac cycle phase. Am J Physiol. 1989;257:H1119-H1126.
6. Azhari H, Buchalter M, Sideman S, Shapiro E, Beyar R. A conical model to describe the nonuniformity of the left ventricular twisting motion. Ann Biomed Eng. 1992;20:149-165.
7. Hansen DE, Daughters GT, Alderman EL, Stinson EB, Baldwin JC, Miller DC. Effect of acute human cardiac allograft rejection on left ventricular systolic torsion and diastolic recoil measured by intramyocardial markers. Circulation. 1987;76:998-1008.
8. Buchalter MB, Weiss JL, Rogers WJ, Zerhouni EA, Weisfeldt ML, Beyar R, Shapiro EP. Noninvasive quantification of left ventricular rotational deformation in normal humans using magnetic resonance imaging myocardial tagging. Circulation. 1990;81:1236-1244.
9. Lower R. Tractatus de Corde, London, 1669. London, UK: Oxford University Press; 1932.
10. Ingels NB, Hansen DE, Daughters GT, Stinson EB, Alderman EL, Miller DC. Relation between longitudinal, circumferential, and oblique shortening and torsional deformation in the left ventricle of the transplanted human heart. Circ Res. 1989;64:915-927.
11. Hansen DE, Daughters GT, Alderman EL, Ingels NB, Miller DC. Torsional deformation of the left ventricular midwall in human hearts with intramyocardial markers: regional heterogeneity and sensitivity to the inotropic effects of abrupt rate changes. Circ Res. 1988;62:941-952.
12. Waldman LK, Fung YC, Covell JW. Transmural myocardial deformation in the canine left ventricle: normal in vivo three-dimensional finite strains. Circ Res. 1985;57:152-163.
13. Rademakers FE, Buchalter MB, Rogers WJ, Zerhouni EA, Weisfeldt ML, Weiss JL, Shapiro EP. Dissociation between left ventricular untwisting and filling: accentuation by catecholamines. Circulation. 1992;85:1572-1581.
14. Yun KL, Niczyporuk MA, Daughters GT III, Ingels NB Jr, Stinson EB, Alderman EL, Hansen DE, Miller DC. Alterations in left ventricular diastolic twist mechanics during acute human cardiac allograft rejection. Circulation. 1991;83:962-973.
15. Kroeker CAG, ter Keurs HEDJ, Knudtson ML, Tyberg JV, Beyar R. An optical device to measure dynamics of apex rotation of the left ventricle. Am J Physiol. 1993;265:H1444-H1449.
16. Kroeker CAG, Tyberg JV, Beyar R. Effects of ischemia on left ventricular apex rotation: an experimental study in anesthetized dogs. Circulation. 1995;92:3539-3548.
17. Beyar R, Yin FCP, Hausknecht M, Kass D. Studies of left ventricular twist in dogs and patients. In: Sideman S, Beyar R, eds. Imaging, Analysis, and Simulation of the Cardiac System: Proceedings of the 5th Henry Goldberg Workshop. London, UK: Freund Publishers; 1990:137-151.
18. Kroeker CAG, Tyberg JV, Beyar R. The effects of load manipulations, heart rate, and contractility on apical rotation: an experimental study in anesthetized dogs. Circulation. 1995;92:130-141.
19. Moon MR, Ingels NB Jr, Daughters GT II, Stinson EB, Hansen DE, Miller DC. Alterations in left ventricular twist mechanics with inotropic stimulation and volume loading in human subjects. Circulation. 1994;89:142-150.
20. Tennant R, Wiggers CJ. The effect of coronary occlusion on myocardial contraction. Am J Physiol. 1935;112:351-361.
21. Kroeker CAG. The Dynamics of Left Ventricular Torsion Throughout the Cardiac Cycle. Calgary, Canada: The University of Calgary; 1995. Dissertation.
22. Tyberg JV, Yeatman LA, Parmley WW, Urschel CW, Sonnenblick EH. Effects of hypoxia on mechanics of cardiac contraction. Am J Physiol. 1970;218:1780-1788.
23. Tyberg JV, Forrester JS, Wyatt HL, Goldner SJ, Parmley WW, Swan HJC. An analysis of segmental ischemic dysfunction utilizing the pressure-length loop. Circulation. 1974;49:748-754.
24. Takano H, Glantz SA. Gadolinium attenuates the upward shift of the left ventricular diastolic pressure-volume relation during pacing-induced ischemia in dogs. Circulation. 1995;91:1575-1587.
25. Kitazume H, Kubo I, Iwama T, Ageishi Y, Suzuki A. Left ventricular function during transient coronary occlusion: digital subtraction left ventriculograms during coronary angioplasty. Clin Cardiol. 1991;14:665-670.
26. Kayden DS, Remetz MS, Cabin HS, Deckelbaum LI, Cleman MW, Wackers FJ, Zaret BL. Validation of continuous radionuclide left ventricular function monitoring in detecting silent myocardial ischemia during balloon angioplasty of the left anterior descending coronary artery. Am J Cardiol. 1991;67:1339-1343.
27. Bertrand ME, Lablanche JM, Fourrier JL, Traisnel G, Mirsky I. Left ventricular systolic and diastolic function during acute coronary artery balloon occlusion in humans. J Am Coll Cardiol. 1988;12:341-347.
28. Alam M, Khaja F, Brymer J, Marzelli M, Goldstein S. Echocardiographic evaluation of left ventricular function during coronary artery angioplasty. Am J Cardiol. 1986;57:20-25.
29. Koolen JJ, Visser CA, David GK, Hoedemaker G, Bot H, vanWezel HB, Dunning AJ. Transesophageal echocardiographic assessment of systolic and diastolic dysfunction during percutaneous transluminal coronary angioplasty. J Am Soc Echocardiogr. 1990;3:374-383.
30. Hunt AC, Chow SC, Escaned J, Perry RA, Seth A, Shiu MF. Changes in Doppler indices of cardiac function during and after percutaneous transluminal coronary angioplasty. Br Heart J. 1991;66:346-350.
31. Wijns W, Serruys PW, Slager CJ, Grimm J, Krayenbuehl HP, Hugenholtz PG, Hess OM. Effect of coronary occlusion during percutaneous transluminal angioplasty in humans on left ventricular chamber stiffness and regional diastolic pressure-radius relations. J Am Coll Cardiol. 1986;7:455-463.
32. Lange HW, Schluter M, Hamm C, Schofer J, Kupper W, Mathey DG, Bleifeld W. Left ventricular diastolic function in PTCA: a Doppler echocardiographic study. Z Kardiol. 1987;76:525-529.
33. Khaja F, Sabbah HG, Brymer JF, Stein PD. Influence of coronary collaterals on left ventricular function in patients undergoing coronary angioplasty. Am Heart J. 1988;116:1174-1180.
This article has been cited by other articles:
![]() |
V. Ferferieva, P. Claus, K. Vermeulen, C. Missant, M. Szulik, F. Rademakers, and J. D'hooge Echocardiographic assessment of left ventricular untwist rate: comparison of tissue Doppler and speckle tracking methodologies Eur J Echocardiogr, July 1, 2009; 10(5): 683 - 690. [Abstract] [Full Text] [PDF] |
||||
![]() |
W.-J. Kim, B. H. Lee, Y. J. Kim, J. H. Kang, Y. J. Jung, J.-M. Song, D.-H. Kang, and J.-K. Song Apical Rotation Assessed by Speckle-Tracking Echocardiography as an Index of Global Left Ventricular Contractility Circ Cardiovasc Imaging, March 1, 2009; 2(2): 123 - 131. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Buckberg, J. I.E. Hoffman, A. Mahajan, S. Saleh, and C. Coghlan Cardiac Mechanics Revisited: The Relationship of Cardiac Architecture to Ventricular Function Circulation, December 9, 2008; 118(24): 2571 - 2587. [Abstract] [Full Text] [PDF] |
||||
![]() |
A T Burns, I G McDonald, J D Thomas, A MacIsaac, and D Prior Doin' the twist: new tools for an old concept of myocardial function Heart, August 1, 2008; 94(8): 978 - 983. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. N. Kirkpatrick, M. A. Vannan, J. Narula, and R. M. Lang Echocardiography in Heart Failure: Applications, Utility, and New Horizons J. Am. Coll. Cardiol., July 31, 2007; 50(5): 381 - 396. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Helle-Valle, J. Crosby, T. Edvardsen, E. Lyseggen, B. H. Amundsen, H.-J. Smith, B. D. Rosen, J. A.C. Lima, H. Torp, H. Ihlen, et al. New Noninvasive Method for Assessment of Left Ventricular Rotation: Speckle Tracking Echocardiography Circulation, November 15, 2005; 112(20): 3149 - 3156. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. B Popovic, J. P. Sun, H. Yamada, J. Drinko, K. Mauer, N. L Greenberg, Y. Cheng, C. S Moravec, M. S Penn, T. N Mazgalev, et al. Differences in left ventricular long-axis function from mice to humans follow allometric scaling to ventricular size J. Physiol., October 1, 2005; 568(1): 255 - 265. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Paetsch, D. Foll, A. Kaluza, R. Luechinger, M. Stuber, A. Bornstedt, A. Wahl, E. Fleck, and E. Nagel Magnetic resonance stress tagging in ischemic heart disease Am J Physiol Heart Circ Physiol, June 1, 2005; 288(6): H2708 - H2714. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Fuchs, M. F. Muller, H. Oswald, H. Thony, P. Mohacsi, and O. M. Hess Cardiac rotation and relaxation in patients with chronic heart failure Eur J Heart Fail, October 1, 2004; 6(6): 715 - 722. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Garot, O. Pascal, B. Diebold, G. Derumeaux, B. L. Gerber, J.-L. Dubois-Rande, J. A. C. Lima, and P. Gueret Alterations of systolic left ventricular twist after acute myocardial infarction Am J Physiol Heart Circ Physiol, January 1, 2002; 282(1): H357 - H362. [Abstract] [Full Text] [PDF] |
||||
![]() |
S.-J. Dong, P. S. Hees, C. O. Siu, J. L. Weiss, and E. P. Shapiro MRI assessment of LV relaxation by untwisting rate: a new isovolumic phase measure of tau Am J Physiol Heart Circ Physiol, November 1, 2001; 281(5): H2002 - H2009. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1997 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |