| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
(Circulation. 1997;96:4400-4407.)
© 1997 American Heart Association, Inc.
Articles |
From the Department of Cardiology, Cleveland Clinic Foundation, Cleveland, Ohio.
Correspondence to Bruce L. Wilkoff, MD, Director, Cardiac Pacing and Tachyarrhythmia Devices, Department of Cardiology/F15, Cleveland Clinic Foundation, 9500 Euclid Ave, Cleveland, OH 44195. E-mail wilkofb{at}cesmtp.ccf.org
| Abstract |
|---|
|
|
|---|
Methods and Results Three interrelated studies were performed. In group 1, the importance of hot can location was investigated by pairing a right ventricular lead to five different hot can placement sites in seven pigs. The defibrillation energies for right pectoral, left pectoral, left subaxillary, and right and left abdominal hot can sites were 20.3±2.7,* 15.9±3.8, 14.9±2.5, 32.0±3.4,* and 30.0±3.4 J,* respectively (*P<.005 versus left pectoral and left subaxillary sites). In group 2, the value of a three-electrode configuration with an abdominal hot can placement was investigated by adding a subclavian vein lead to the pectoral or abdominal hot can configurations in seven pigs. The defibrillation energies for left pectoral and abdominal sites were 18.6±4.2 and 29.0±5.8 J (P=.0001), respectively. The addition of a right or left subclavian vein lead with an abdominal hot can reduced the threshold to 19.3±4.2* or 18.8±3.2,* respectively (*P=.0001 versus abdominal site). In group 3, the contribution of the abdominal hot can electrode to the three-electrode configuration was tested by a comparison with two purely transvenous two-electrode configurations in six pigs. The defibrillation energy (19.9±3.2 J) for the abdominal hot can with a subclavian vein lead was lower than the transvenous lead configurations with a subclavian vein (29.0±2.5 J, P=.0001) or a superior vena cava lead (30.7±3.7 J, P=.0001). The right ventricular lead was the sole cathode during the first phase of the biphasic shock in all experiments.
Conclusions Defibrillation energy depends on the hot can placement site. The addition of a subclavian vein lead with an abdominal hot can improves defibrillation efficacy to the level of the pectoral placement and is better than a purely transvenous lead configuration.
Key Words: defibrillation ventricles electrical stimulation death, sudden
| Introduction |
|---|
|
|
|---|
The purpose of this study was (1) in group 1, to evaluate the best geometry for defibrillation with a hot can system using a single RV lead; (2) in group 2, to determine whether the addition of a transvenous lead located in the subclavian vein improves DFTs when an abdominal hot can implantation site is used; and (3) in group 3, to measure the contribution of the abdominal hot can electrode to the reduction of DFTs in a three-electrode configuration.
| Methods |
|---|
|
|
|---|
The external waveform amplifier and defibrillator (Angeion Research Defibrillation System, ARD-9000, Angeion Corp) was used. The ARD-9000 operates as a high-voltage, linear amplifier using software-generated waveforms. The device samples the current and voltage every 0.1 ms and adjusts the waveform voltage to mimic a capacitive discharge. The continuous voltage adjustment responds to the impedance changes that occur as a function of voltage changing during a discharge.20 Preliminary tests in a saline bath demonstrated that the voltage waveforms generated by this device have <3% variation from waveforms generated by a true capacitor discharge in the range of 100 to 750 V. The ARD-9000 continuously calculates the instantaneous impedance by dividing the delivered voltage by the current. The median of all these values is then used as the overall shock impedance. The ARD-9000 measures the current and voltage every 0.1 ms. The average current was calculated as the sum of all sampled current values divided by the number of samples.
VF was induced with 60-Hz alternating current (15 V) for 4 seconds through the RV defibrillation lead. After sustained VF lasting 10 seconds from initiation of the alternating current, a biphasic defibrillation waveform (135-µF capacitance, 65% tilt in each phase) was delivered. If VF was not terminated by this waveform, a rescue shock (monophasic 5-ms square waveform, 20 to 50 J) was delivered to terminate the VF. A recovery period of at least 3 minutes was allowed between each episode of VF. VF was not reinitiated until the heart rate and blood pressure stabilized and returned to preshock levels.
Determination of DFT
Threshold testing randomized the sequence of the implantation
sites (group 1) and defibrillation lead configurations (groups 2 and
3). The location of the RV lead was fixed after the initial lead
configuration was validated. This required a single shock success at
35 J for the abdominal hot cantoRV lead configuration or the
purely transvenous lead configuration and at
25 J for all other lead
configurations. After validation, DFT was determined by a "down-up,
down-up" technique21,22 until three
reversals of defibrillation success were completed. After the initial
successful shock, the stored energy was decremented by 1 J after each
successful defibrillation. When a shock failed to defibrillate, the
next trial was performed by incrementing the stored energy by 1 J. This
process was repeated until three reversals in decrement or increment
occurred. The final shock was always a success after the last failure.
The DFT was defined as the average of the shock energy, voltage, or
ampere values obtained with all trials starting from the successful
shock before the first defibrillation failure until the last successful
defibrillation.
After determination of the DFT for the lead configurations, preparation stability was ensured by repeating the DFT measurements for the first of the randomized lead configurations tested in that animal. If the stored energy at DFT was within 2 J of the first DFT, then the preparation was deemed stable, and the experiment was included in the study. If there was a change of >2 J between the first DFT energy and this last DFT energy, the preparation was not considered stable, and the data were rejected.
The use of experimental animals in this study was approved by the Animal Research Committee of the Cleveland Clinic Foundation and conformed to the position of the American Heart Association on Research Animal Use.
Statistical Analysis
The mean and SD of all parameters were calculated.
Repeated-measures one-way ANOVA was used to compare stored energy,
delivered energy, peak voltage, peak current, average current, median
impedance, and pulse width among the lead systems. Pairwise comparisons
of each defibrillation parameter were performed by the
method of contrasts (Fisher's least significant difference
test)23 for the post hoc analysis.
The null hypothesis was rejected for P<.05.
| Results |
|---|
|
|
|---|
|
Group 1 Results
Complete DFT data sets were obtained from seven pigs (34±3 kg).
The DFT parameters and pulse width for each hot can
implantation site are given in detail in Table 1
.
|
Fig 2
shows the stored energy at DFT in
five different hot can implantation sites. The stored energies at DFT
for the abdominal implantation sites were significantly higher than for
the pectoral implantation sites (P=.0001 for each pectoral
site versus each abdominal site). There was no difference in stored
energy between the right abdominal and left abdominal sites
(P=.1588). Conversely, the stored energies at DFT for both
the left pectoral and left subaxillary sites were
25% lower than
for the right pectoral site and were statistically significant
(P=.0042 and P=.0007 versus right site,
respectively). There was no difference in stored energy between the
left pectoral and left subaxillary sites (P=.4711). Fig 3
shows the individual experimental
stored energy data at DFT for the right and left pectoral sites. Except
for one experiment, the stored energy at DFT for the left pectoral site
was lower than for the right pectoral site.
|
|
The impedance at DFT for the abdominal implantation sites was significantly higher than for the pectoral implantation sites. In the abdominal implantation sites, there were no differences in the impedance between the right and left sites (P=.4376). Conversely, for the pectoral implantation sites, the impedance at DFT for the left pectoral site was lower than for the left subaxillary site (P=.0041).
Group 2 Defibrillation Protocol
This protocol tested the impact on DFTs of the addition of a lead
in the right or left subclavian vein when an abdominal hot can
electrode (Angeion Corp) implantation site was used. The hot can
electrode (see group 1 for a complete description) was implanted in the
left pectoral and the left upper abdominal sites. A unipolar
defibrillation lead (model 497, Intermedics Inc) with a 5-cm-long
electrode and an 8.3F diameter (surface area of 4.4
cm2) was advanced to either the right or left
subclavian vein under fluoroscopy (Fig 4
). The RV apex, either the right or left
subclavian vein lead, and the hot can electrode were connected to the
ARD-9000. DFT parameters were evaluated with four different
defibrillation lead configurations, which included control experiments
with pectoral or abdominal hot can electrodes and the abdominal hot can
electrode with a right or left subclavian vein lead in random order in
each pig (detailed in Fig 4
).
|
Group 2 Results
Complete DFT data sets were obtained from seven pigs (26±2 kg).
The DFT parameters and pulse width for each electrode
configuration are given in detail in Table 2
.
|
The stored energy at DFT for the abdominal control configuration was
55% higher than the pectoral control configuration
(P=.0001). However, the addition of a subclavian vein lead
to the abdominal control configuration reduced the stored energy at DFT
and performed as well as the pectoral control configuration (Fig 5
). There was no difference in stored
energy between the right and left subclavian vein experiment
configurations (P=.6784). Fig 6
shows the individual experimental
stored energy data at DFT for the pectoral control and left subclavian
vein experiment configurations. There was no significant difference in
stored energy between the two lead configurations (P=.887).
Fig 7
shows the stored energy data at DFT
in individual experiments for the abdominal control and left subclavian
vein experiment configurations. An additional left subclavian vein lead
with the abdominal placement of the hot can electrode significantly
reduced the stored energy at DFT in all experiments
(P=.0001).
|
|
|
Although the impedance at DFT for the abdominal control was significantly higher than for the pectoral control (P=.0001), an additional subclavian vein lead with the abdominal control configuration reduced the impedance. There was no difference in impedance between the right subclavian vein experiment and left subclavian vein experiment configurations (P=.9842).
Group 3 Defibrillation Protocol
This protocol tested the value of using a hot can electrode for
abdominal implantation. The hot can electrode (see group 1 for a
complete description) was implanted in the left upper abdominal sites.
A unipolar defibrillation lead (model 497, Intermedics Inc) with a
5-cm-long electrode and an 8.3F diameter (surface area of 4.4
cm2) was advanced to either the superior vena
cava or left subclavian vein under fluoroscopy (Fig 8
). The RV apex and either the superior
vena cava or left subclavian vein lead or the abdominal hot can
electrode were connected to the ARD-9000. DFT parameters
were evaluated at three different defibrillation lead configurations,
which included the abdominal hot can electrode with a left subclavian
vein lead configuration (three-electrode abdominal hot can) and two
purely transvenous lead configurations (subclavian vein to RV and
superior vena cava to RV) in random order in each pig (detailed in Fig 8
).
|
Group 3 Results
Complete DFT data sets were obtained from six pigs (38±6
kg). The DFT parameters and pulse width for each electrode
configuration are given in detail in Table 3
.
|
Fig 9
shows the mean stored energy
at DFT in each defibrillation lead configuration. The stored energy at
DFT for the three-electrode abdominal hot can configuration was 31%
and 35% lower than the subclavian veintoRV lead
(P=.0001) and the superior vena cavatoRV lead
configurations (P=.0001). There was no significant
difference in stored energy between the two purely transvenous lead
configurations (P=.2475). An additional subclavian vein lead
with the abdominal placement of the hot can electrode configuration
(three-electrode abdominal hot can) significantly reduced the stored
energy at DFT in all experiments compared with both purely transvenous
lead configurations ("cold can").
|
The impedance at DFT for both cold can configurations using either the
subclavian vein or superior vena cava was 66.6 or 53.4
,
respectively. However, the abdominal hot can three-electrode
configuration reduced the impedance to 44.7
(P=.0001).
| Discussion |
|---|
|
|
|---|
Implantation Site
Previous studies using two epicardial patches have reported that
the patch position that directed current through a large portion of the
LV myocardium, particularly the
interventricular septum, was predictive of defibrillation
efficacy.2427 In addition, with implantation of
two transvenous leads as the defibrillation lead configuration,
positioning the proximal defibrillation lead in the subclavian
innominate vein decreased DFT energy requirements compared with
positioning it in the superior vena cava.28
Similarly, with a single RV lead unipolar system, the location of the
hot can electrode affects defibrillation
efficacy.18,19 Thus, the electrode location plays
an important role and influences defibrillation efficacy.
Although the preferred implantation site of the hot can electrode is the left pectoral area, it may be impossible or inconvenient to implant the hot can electrode in such an area in some cases. Schofield et al17 reported that DFT energies were satisfactory with a hot can electrode implanted in the right pectoral side even though the vector between the cathode and anode passed predominantly through the RV. In our study, the DFT energy for the right pectoral implantation site was 30% higher than the left pectoral site. In a preliminary study, Seidl et al18 examined the differences in biphasic shock DFT energy achieved at hot can ICD implantation using four different left pectoral locations, including high pectoral medial position, high pectoral lateral position, low pectoral medial position, and low pectoral lateral position. Their results suggested that the low pectoral medial position was best. In another preliminary study, Min et al19 investigated the effect of hot can position on DFT energy in a human thorax model by finite-element analysis. This model indicated that locating the hot can electrode beneath the clavicle in a submuscular position would lower DFT energy. In our study, there was no significant difference in DFT energy between the left pectoral and left subaxillary sites. However, three of seven pigs had lower DFT energies for the left subaxillary site than for the left pectoral site. In a nonthoracotomy subcutaneous patch electrode system, Saksena et al29 reported that the use of an axillary patch generated lower DFT energy compared with pectoral and apical locations. The mechanism by which the axillary site produces a lower DFT energy may be related to the axillary site vector, which is more perpendicular to the ventricular septum, thus passing more current through the LV septum.25
In addition, our study indicated that the DFT energy for the abdominal hot can sites was approximately twice that of the pectoral hot can sites. The abdominal hot cantoRV lead configuration may reduce the current delivery through the myocardium because of an increased impedance and altered geometry. Thus, changing the hot can electrode position may change the DFT energy by altering the impedance and field distribution.
Tripolar Configuration
A left pectoral hot can electrode system incorporating a single
transvenous RV lead is clinically acceptable in most
patients.1316 In a few reports, despite
lowering pathway resistance, the addition of a third defibrillation
lead in the superior vena cava30 or
coronary sinus31 position did not
generally affect the DFT energy of an RV leadtoleft pectoral hot
can electrode configuration. In contrast, a few clinical reports showed
that the addition of a superior vena cava lead to a hot canRV lead
configuration improved defibrillation efficacy with lower
impedance.32,33 This was supported by
finite-element analysis.33 Thus,
previous studies investigating the defibrillation effect of an
additional lead with the left pectoral implantation site of a hot can
electrode have reported mixed findings.
In our study, however, the addition of a defibrillation lead in the
right or left subclavian vein improved defibrillation efficacy for
abdominal hot can implants. In fact, this additional lead decreased the
stored energy at DFT by one third (Fig 5
) and lowered impedance by 28%
compared with the hot can electrode alone. Furthermore, the DFT energy
for an addition of a subclavian vein lead with an abdominal hot
cantoRV lead system was lower than those for a left pectoral hot
can electrodetoRV lead system in three (43%) of the seven pigs
(Fig 6
). This increase in defibrillation efficacy with the addition of
a subclavian vein lead is probably due to both an improved current
vector through the LV and a reduction of shock impedance.
The impedance of the defibrillation pathway is an important
determination of defibrillation efficacy.3437
Reduction of impedance directly results in higher peak and average
current emanating from the RV electrode at the same voltage. Shocks
into high-impedance systems result in weaker current density fields
than shocks into low-impedance systems for a constant voltage source.
If the intramyocardial electrical field strength produced by the shock
into the high-impedance system is above a critical minimum
value,38,39 defibrillation may occur. Thus, the
intramyocardial current field strength produced by the shock appears to
be a critical parameter that determines whether a
particular shock fails. Therefore, reductions in impedance should
correspond to reductions in energy requirements for defibrillation.
However, not all reductions of impedance necessarily result in reduced
DFT. For example, impedance is reduced 20% when the left subclavian
vein electrode is moved to the superior vena cava (see Table 3
).
However, DFT energy is not affected. Thus, there are some methods of
reducing impedance that help and others that do not help with
defibrillation. If the heart can be viewed as a resistor, then
reduction of impedance in series to the heart would improve the voltage
gradient with the heart. However, reductions in impedance that are
parallel to the heart would shunt current away from the heart and
reduce the voltage gradient in the heart.
Clinical Implications
In patients in whom left pectoral implantation is impossible, it
will often be possible to implant a hot can electrode in the right
pectoral site. However, even though the size of defibrillators
continues to decrease, it still precludes pectoral implantation for
some patients. In these cases, the device must be implanted in an
abdominal site. Therefore, the reduction in DFT energy with an
additional lead may be important in such cases. Other reasons, such as
previous infection, pacemakers, and physician choice, may require an
increase in safety margins for patients undergoing abdominal
implantation of a hot can electrode ICD.
Limitations
The pig heart is similar anatomically to the human heart. Hence,
the pig is commonly used by many investigators as a VF model. However,
the anatomy of the abdomen and thorax in the pig is not exactly
the same as that of the human. Therefore, the findings of this study
may not be completely consistent with the clinical
situation.
Although the addition of a subclavian vein lead with the abdominal hot can electrode placement can significantly increase defibrillation efficacy, the long lead tunneling procedure from the thoracic venous entry site to the abdominal hot can electrode pocket may be an aggravating factor for infection and lead complications.40
Conclusions
The major findings in this study are as follows. (1) The hot can
electrode implantation site plays a important role in the determination
of defibrillation efficacy. The DFT energies for the left pectoral and
left subaxillary implantation sites were virtually the same. However,
the DFT energy increased 30% for the right pectoral implantation and
100% for the abdominal implantation. (2) The addition of a subclavian
vein lead with the abdominal hot can implantation improves
defibrillation efficacy, making DFT energy levels comparable to that of
the left pectoral implants. (3) Furthermore, this defibrillation lead
configuration reduces DFT energy by 30% compared with a purely
transvenous lead configuration (cold can).
| Selected Abbreviations and Acronyms |
|---|
|
| Acknowledgments |
|---|
Received May 27, 1997; revision received August 25, 1997; accepted September 12, 1997.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
Y. Yamanouchi, J. E. Brewer, K. F. Olson, K. A. Mowrey, T. N. Mazgalev, B. L. Wilkoff, and P. J. Tchou Fully Discharging Phases : A New Approach to Biphasic Waveforms for External Defibrillation Circulation, August 24, 1999; 100(8): 826 - 831. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Yamanouchi, J. E. Brewer, K. A. Mowrey, A. M. Donohoo, B. L. Wilkoff, and P. J. Tchou Optimal Small-Capacitor Biphasic Waveform for External Defibrillation : Influence of Phase-1 Tilt and Phase-2 Voltage Circulation, December 1, 1998; 98(22): 2487 - 2493. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1997 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |