(Circulation. 1997;96:4286-4297.)
© 1997 American Heart Association, Inc.
Articles |
From the Department of Cardiology (T.B., P.H., R.E.), University of Essen (Germany); the Second Department of Medicine, Cardiology, Angiology, General Hospital St. Elisabeth (W.T., H.J.N.), Linz, Austria; and EFMT Research and Development Center for Micro Therapy (L.U.W.), Bochum, Germany.
Correspondence to Raimund Erbel, MD, Department of Cardiology, University of Essen, Hufelandstrasse 55, 45122 Essen, Germany. E-mail rerbel{at}uni-essen.de
| Abstract |
|---|
|
|
|---|
Methods and Results In 23 patients with chronic stable LV aneurysms, LV end-systolic and end-diastolic volumes (LVEDV, LVESV) and ejection fraction (LVEF) by tomographic 3DE were compared with results from 3D magnetic resonance tomography (3DMRT) as an independent reference as well as with the conventional techniques of single plane and biplane 2D echocardiography and biplane cineventriculography. Dynamic 3DE image data sets were obtained from a transthoracic apical view with the use of a rotating probe with acquisition gated to control for ECG and respiration (Echoscan, TomTec). Volumes were calculated from the 3D data sets by summating the volumes of multiple parallel disks. 3DE results correlated and agreed well with those by 3DMRT, with better correlation and agreement than provided by other techniques for LVEDV (3DE: r=.97, SEE=14.7 mL, SD of differences from 3DMRT=14.5 mL; other techniques: r=.84 to .93, SEE=30.7 to 41.6 mL [P<.001 versus 3DE by F test], SD of differences=31.5 to 40.7 mL [P<.001 versus 3DE by F test]). The same also pertained to LVESV (3DE: r=.97, SEE=12.4 mL, SD of differences=12.9 mL; other techniques: r=.81 to .90, SEE=24.7 to 37.2 mL [P<.001], SD of differences=27.6 to 36.8 mL [P<.005]) and LVEF (3DE: r=.74, SEE=5.6%, SD of differences=6.7%; other techniques: r=.14 to .59, SEE=9.5% to 10.1% [P<.01], SD of differences=9.5% to 12.6% [P<.05]). Compared with 3DMRT, 3DE was less time consuming and patient discomfort was less.
Conclusions Tomographic 3DE is an accurate noninvasive technique for calculating LV volumes and systolic function in patients with LV aneurysm. Unlike current 2D methods, tomographic 3DE requires no geometric assumptions that limit accuracy.
Key Words: echocardiography ventricle aneurysm magnetic resonance imaging
| Introduction |
|---|
|
|
|---|
To determine LV volumes and LVEF, as the important
parameters of global systolic LV function, LVEDV
and LVESV must be measured accurately. Until now, asymmetric LV volumes
have been approximated by different algorithms calculating symmetric
rotation ellipsoids by the use of single or biplane
echocardiographic or radiographic
imaging.1,2 The following methods have been
developed: the ellipsoid method,8 the area-length
method,9 and the disk
method,10 for single and biplane calculation,
respectively. Erbel et al11 correlated all these
methods by echocardiographic and
radiographic determination of volumes in asymmetric model
hearts. Only for the biplane disk method was linearity at ANCOVA found.
Nonetheless, by summing up elliptic disks from the two diameters of two
orthogonal planes (Fig 1
), the biplane
disk method can create all approximations close to the true dimensions
as well as overestimation, but most often underestimation is observed
(Fig 2
).11
Erbel et al11 therefore suggested that more than
two planes should be used to determine LV dimensions more accurately.
Three-dimensional echocardiographic image acquisition,
with the use of multiple echocardiographic image
planes, has recently been validated for LV volume determination but was
limited in the acquisition of equidistant cross sections for using the
disk method, as suggested by Weyman,12 as well as
in the acquisition of heart ratetriggered dynamic data sets for
determination of LV function.1317
|
|
The technical requirements for realizing this idea by means of
echocardiographic tomographic imaging are available
today. Wollschlaeger et al18 first described a
technique for the acquisition of a complete dynamic three-dimensional
echocardiographic data set, which allows an unlimited
choice of views for all cardiac structures and spaces but also a
division in equidistant tomographic slices in a parallel order, as in
magnetic resonance or computer tomographic imaging (Fig 1
). To test the
accuracy of this method for quantifying LV volumes in the presence of
aneurysm, our group performed an experimental prestudy with
asymmetric model hearts. We compared the volumes calculated from
tomographic 3DE to true dimensions obtained by direct
measurements.19 This in vitro study revealed good
agreement and correlation between 3DE measurements and direct
measurements of the true model heart volume without significant
underestimation or overestimation but poor agreement and correlation
between bp2DE and both 3DE and direct measurements, with strong
overestimation of 2D measurements.
After these experiments the purpose of the following study was to evaluate for the first time the accuracy of tomographic 3DE as a clinical application for quantifying global systolic function in a series of patients with LV aneurysm after myocardial infarction. 3DE results were compared with results from 3DMRT as an independent reference that has been validated for determining LV volumes accurately20,21 as well as with the conventional techniques of sp2DE, bp2DE, and bpCVG.
| Methods |
|---|
|
|
|---|
Three-dimensional Echocardiography
For dynamic 3DE, image data acquisition was performed by a
commercially available phased-array sector scanner (model SSH 140 A;
Toshiba Corp) and a 3.75-MHz transducer from a
transthoracic apical view. 3D image data sets were obtained
by rotating the transducer with a special rotation motor device
(TomTec) in 2 degree steps and digitizing 90 ultrasound images by the
Echoscan system (TomTec) automatically.19 During
the examination, patients were placed in a left decubital supine
position, and the rotation device was mounted on a stand device, fixing
the transducer at the chest wall during rotation. For dynamic 3DE
reconstruction, rotation was performed automatically by ECG and
respiratory triggering. At every step of rotation 2D images were
acquired at every ECG interval of 40 ms and at end expiration,
respectively. After digital storage, the 3D data set underwent a
postprocessing procedure composed of a gap filling between the acquired
images (interpolation) and a conversion of data points (pixels) ordered
in a cylindrical coordinate system into a right-angled (Cartesian)
coordinate system. For interpolation, average image information from
the two images bordering the gap was used as infill. Volumes were
calculated from the 3D data sets by summating the volumes of multiple
parallel disks (Figs 1
, 3
, and 4
).19 The distance
between the tomographic cross-sectional images, identical to the disk
thickness, was 2.9 mm.
|
|
Tomographic Magnetic Resonance Imaging
Images were obtained with a commercially available 1.5-T
superconducting magnetic resonance imaging (MRI) system (Siemens
Magnetom SP 63). ECG-gated localizing spin-echo sequences were used to
identify the intrinsic long axes of the heart. For all patients,
short-axis MRI images were started at the mitral valve plane. Long axes
were subdivided in the middle, and the two levels at the same position
of each half (1st/1st, 2nd/2nd, etc) were scanned
simultaneously with a slice thickness of 8.0 mm. The
left ventricle was covered with five or six of those short-axis
series.
Biplane Cineventriculography
BpCVG was performed in a digitally equipped
catheterization laboratory (HICOR, Siemens). For bpCVG,
35 mL of a nonionic contrast agent (Ultravist, Schering) was injected
into the left ventricle with a flow rate of 12 mL/s with the use of a
5F or 7F pigtail catheter. The biplane projections were
recorded in 30 degree right anterior oblique (RAO) and 60 degree
left anterior oblique (LAO) projections. The afterward image
calibration was performed with the use of a metal ball with a diameter
of 5.0 cm with identical positions of the x-ray tubes. This was
possible with an automatic readjustment to the correct positions using
the system memory function.
Monoplane and Biplane Echocardiography
Conventional transthoracic
echocardiographic imaging of the left ventricle was
performed in the apical RAO equivalent view and for biplane studies the
apical RAO and apical two-chamber view.11 Both 2D
studies were performed immediately after scanning for the 3D
echocardiographic examination with the same
phased-array sector scanner (model SSH 140 A; Toshiba Corp) with a
3.75-MHz transducer. Patients remained in the same left decubital
supine position as for the 3D examination before. The resolution of the
system has been tested previously and demonstrated high accuracy for LV
volume and EF determination in patients with LV
aneurysm.22
Quantitative Measurements
After image acquisition for all of the described methods, LVEDV
and LVESV were determined and resulting LVEF was calculated. For both
3DE and 3DMRT, LVEDV and LVESV were calculated with the use of the 3D
disk method (Fig 1
).19 Biplane volume
calculations for bpCVG and echocardiographic image data
were performed by the disk method with an external system (ECHO-COM 2.0
for Windows; Individual Software GmbH). Therefore, two biplane images
of one-phase end diastole or end systole were displayed
side by side on one screen with the use of a special copy function
integrated into the echocardiographic unit (Hewlett
Packard Sonos 1500). Monoplane echocardiographic volume
determination was performed with the use of the area-length method
because this is the method most often applied in routine
echocardiographic diagnostics.
| Statistical Analysis |
|---|
|
|
|---|
In a second analysis, linear regression was calculated, specifying the correlation coefficient (r), intercept (a), slope (b), standard error of estimation (SEE), and 95% confidence interval of the regression line. As a reference technique for statistical analysis, independent variables (x-axis) were measured by 3DMRT and dependent variables (y-axis) by 3DE and 2D methods. A two-tailed F test at the 5% significance level was used to analyze the equality of SEEs between the correlation of 3DE to 3DMRT and the correlation of 3DMRT to the other techniques.
| Results |
|---|
|
|
|---|
|
Three-dimensional Echocardiography Versus
Three-dimensional Magnetic Resonance Imaging
The results for LVEDV and LVESV followed a normal distribution in
the 3DE measurements but not in the 3DMRT measurements; the results for
LVEF followed a normal distribution in both the 3DE and 3DMRT
measurements. For LVEDV, LVESV, and LVEF, values obtained by 3DE ranged
from 78 to 297 mL (mean, 173±62 mL), 53 to 252 mL (mean, 117±53 mL),
and 19% to 50% (mean, 33±8%) and ranged from 82 to 300 mL (mean,
183 mL), 56 to 276 mL (mean, 121 mL), and 19% to 54% (mean, 36±10%)
for 3DMRT. The differences between the two methods ranged for LVEDV,
LVESV, and LVEF from -47 to 11 mL (mean, -10.7±14.5 mL), -41 to 10
mL (mean, -3.4±12.9 mL), and -20.7% to 9.7% (mean, -2.5±6.7%)
with SEE of ±3.0 mL, ±2.7 mL, and ±1.4% and 95% confidence
interval between -16.6 and -4,7 mL, -8.3 and 1.9 mL, and -5.2% and
0.3%, respectively. Limits of agreement amounted to -39.7 and 18.4
mL, -29.1 and 22.4 mL, and -15.8% and 10.9%, respectively. With
P=.002, P=.22, and P=.091, a
significant difference of the mean difference from 0 was found for
LVEDV as an underestimation of 3DE compared with 3DMRT measurements;
however, no significant difference of the mean difference from 0 was
found for LVESV and LVEF.
Linear regression analysis was used to correlate LVEDV, LVESV, and LVEF measurements by the two methods. The correlations were described by y=0.97x-5.6 mL, r=.97, SEE=±14.7 mL; y=0.92x+6.2 mL, r=.97, SEE=±12.4 mL; and y=0.61x+11.7%, r=.74, SEE=±5.6%.
The plot in Fig 5a
depicts the finding
that the mean difference of LVEDV measurements is near zero, indicating
that there was no relevant systematic overestimation or
underestimation. Nevertheless, statistical analysis detected a
statistically significant underestimation by 3DE compared with 3DMRT.
However, the limits of agreement were small, indicating good agreement
between the two methods. For LVEF, Fig 6a
shows the mean difference to be near zero with small limits of
agreement. Fig 7a
documents the finding
of a good correlation between the two methods for LVEDV, which might
underline the good agreement between the two methods. For LVEF, Fig 8a
documents low correlation but with
acceptable limits of the 95% confidence interval and an acceptably
small SEE.
|
|
|
|
Biplane Cineventriculography and Biplane
and Monoplane Echocardiography Versus Magnetic
Resonance Imaging
On 2D measurements, only the results of LVEF by all methods and of
LVESV by bpCVG were normally distributed. On measurement of LVEDV
values for bpCVG, bp2DE and sp2DE ranged from 105 to 429 mL (mean, 223
mL), 95 to 446 mL (mean, 175 mL), and 88 to 397 mL (mean, 162 mL). The
mean differences compared with 3DMRT were 49.0±31.5 mL, -8.0±40.7
mL, and -20.9±37.6 mL, with limits of agreement between -14.0 and
111.9 mL, -89.4 and 73.4 mL, and -96.2 and 54.4 mL and 95%
confidence intervals between 33.5 and 64.4 mL, -24.6 and 8.7 mL, and
-36.3 and -5.5 mL, respectively. SEE was ±7.9 mL, ±8.5 mL, and
±7.9 mL. With P<.001, P=.007, and
P=.014 by t test, there was a statistically
significant overestimation for LVEDV measured by bpCVG and a
statistically significant underestimation by bp2DE and sp2DE. With
P<.001, P<.001, and P<.001 by
F test, the standard deviations of the differences were
significant larger compared with the standard deviation of the
differences between 3DE to 3DMRT.
On measurement of LVESV values for bpCVG, bp2DE and sp2DE ranged from 72 to 262 mL (mean, 141±54 mL), 52 to 291 mL (mean, 110 mL), and 42 to 323 mL (mean, 105 mL). The mean differences compared with 3DMRT were 25.3±24.4 mL, -11.2±27.6 mL, and -15.4±36.8 mL, with limits of agreement between -23.4 and 74.0 mL, -66.4 and 44.0 mL, and -89.0 and 58.2 mL and 95% confidence intervals between 13.3 and 37.2 mL, -22.5 and 0.1 mL, and -30.4 and 0.4 mL, respectively. SEE was ±5.1 mL, ±5.8 mL, and ±7.7 mL. With P<.001, P=.064, and P=.057 by t test, there was a statistically significant overestimation for LVEDV measured by bpCVG but no statistically significant overestimation or underestimation for bp2DE and sp2DE measurements compared with results from 3DMRT. With P<.005, P<.001, and P<.001 by F test, the standard deviations of the differences were statistically significant larger compared with the standard deviation of the differences between 3DE to 3DMRT.
On LVEF measurement we found ranges of 9% to 52% (mean, 37±12%), 18% to 60% (mean, 39±9%), and 18% to 57% (mean, 38±11%), respectively. The mean differences compared with results from 3DMRT were 1.5±9.5%, 2.7±12.6%, and 1.8±11.6%, with limits of agreement between -17.5% and 20.4%, -22.4% and 27.9%, and -21.3% and 24.9% and 95% confidence intervals between -3.2% and 6.1%, -2.4% and 7.9%, and -2.9% and 6.5%, respectively. SEE was ±2.4%, ±2.6%, and ±2.4%. With P=.546, P=.303, and P=.466 by t test, no statistically significant overestimation or underestimation was found. With P<.05, P<.001, and P<.005 by F test, the standard deviations of the differences were statistically significant larger compared with the standard deviation of the differences between 3DE to 3DMRT. In bpCVG measurements there was, however, a clear outlier, which was probably due to a very uncommon ventricular shape and was omitted for statistical analysis as a statistically accepted procedure.23
Compared with 3DMRT, the correlation for the LVEDV was described for bpCVG by y=1.17x+19.5 mL, r=.93, and SEE=±30.7 mL; for bp2DE by y=1.00x-8.8 mL, r=.84, and SEE=±41.6 mL; and for sp2DE by y=0.95x-11.3 mL, r=.84, and SEE=±38.4 mL. With P<.001, P<.001, and P<.001 by F test, SEEs were statistically significantly larger compared with the correlation of 3DE to 3DMRT. For LVESV, the correlation was described for bpCVG by y=0.91x+35.7 mL, r=.90, and SEE=±24.7 mL; for bp2DE by y=0.88x+3.5 mL, r=.88, and SEE=±27.4 mL; and for sp2DE by y=0.90x-3.2 mL, r=.81, and SEE=±37.2 mL. With P<.001, P<.001, and P<.001 by F test, SEEs were statistically significantly larger compared with the correlation of 3DE to 3DMRT. For LVEF, the correlation was described for bpCVG by y=0.83x+7.5%, r=.59, and SEE=±9.7%; for bp2DE by y=0.13x+34.0%, r=.14, and SEE=±9.5%; and for sp2DE by y=0.39x+23.7%, r=.36, and SEE=±10.1%. With P<.005, P<.01, and P<.005 by F test, SEEs were statistically significant larger compared with the correlation of 3DE to 3DMRT.
Fig 5b
and d document that neither the mean differences between bpCVG
and 3DMRT nor sp2DE and 3DMRT are close enough to zero, indicating a
remarkable systematic overestimation for LVEDV measurements by bpCVG
and a slight underestimation of sp2DE measurements. Fig 5c
shows that
there was no relevant overestimation or underestimation for bp2DE
measurements. Only for the 3DE measurements of LVEDV were no outliers
found. The bpCVG outlier of a remarkable overestimation of that
technique, which was omitted from this analysis, was obtained
from the acromegaly patient having an extreme asymmetric left ventricle
with a very large apical aneurysm but a normal, tall
ventricular base. Thus by bpCVG using the 60 degree LAO
projection, the large aneurysm was prominent at the second
plane for biplane calculation and its large protrusion was calculated
over the whole ventricular shape, but not only for the
apex. We also found overestimation for the sp2DE and bp2DE measurements
in the same patient, but to a significantly smaller extent, so we
decided not to exclude the 2DE outliers. However, it must be taken into
account that inclusion of the cineventriculographic outlier would yield
much greater limits of agreement and the exclusion of the 2DE outliers
much smaller limits of agreement. Figs 6b
, 6c
, and 6d
document the fact
that for LVEF the mean differences between bpCVG, bp2DE and sp2DE
compared with 3DMRT are close enough to zero, indicating that there was
no relevant overestimation or underestimation. In all three comparisons
for LVEDV (Figs 5b
, 5c
, and 5d
) and LVEF (Figs 6b
, 6c
, and 6d
),
however, the limits of agreement were wide, indicating unacceptable
agreement between the three methods compared with 3DMRT. Fig 7b
documents for bpCVG an acceptable correlation coefficient of
r=.93, but the overestimation and an SEE of ±30.7 mL
indicate poor correlation compared with 3DMRT. In comparison to Fig 7b
, Figs 7c
and 7d show slightly better correlation for the 2DE methods
than for bpCVG, but nevertheless the limits of the 95% confidence
interval are wide and SEEs are large, also indicating poor correlation.
Figs 8b
, 8c
, and 8d
show the correlations for LVEF, which are poorer
for all three 2D methods as approved to 3DMRT.
Accuracy of Repeat Measurements and Interobserver
Comparison
For repeat measurements of LVEDV, LVESV, and LVEF by 3DE, we found
a mean difference of -0.3±6.2 mL, -0.4±3.9 mL, and 0.4±2.7%, SEE
of ±1.3 mL, ±0.8 mL, and ±0.6%, limits of agreement of ±12.4 mL,
±7.8 mL, and ±5.4%, and a 95% confidence interval of ±2.5 mL,
±1.6 mL, and ±1.1%, respectively. For the interobserver measurements
of LVEDV, LVESV, and LVEF by 3DE, we found a mean difference of
1.3±7.5 mL, 2.0±5.1 mL, and -0.5±4.8%, SEE of ±1.6 mL, ±1.1 mL,
and ±1.0%, limits of agreement of ±9.6 mL, ±10.1 mL, and ±9.4%,
and a 95% confidence interval of ±3.1 mL, ±2.1 mL, and ±2.0%,
respectively.
These results for LVEDV and LVEF are plotted in Figs 9a
, 9b
, 9c
, and 9d
and reveal good
reproducibility and a small interobserver variability of 3DE
measurements, which indicates furthermore that tomographic 3DE provides
reliable and accurate measurements in clinical studies.
|
| Discussion |
|---|
|
|
|---|
On the basis of the experience obtained in the in vitro study,19 in this in vivo study tomographic determination of global systolic LV function by 3DE was compared with 3DMRT, sp2DE and bp2DE, and bpCVG in 23 patients with chronic LV aneurysm. We found good agreement only between tomographic 3DE and the 3DMRT measurements of LV volumes and LVEF, and similar to the in vitro results, tomographic 3DE measurements were highly reproducible combined with low interobserver variability. In contrast to previous 2DE studies10,24,25 we found a far smaller but nevertheless significant underestimation for the determination of LVEDV by 3DE but no significant underestimation or overestimation of LVESV and LVEF. For sp2DE, bp2DE, and bpCVG, agreement and correlation with 3DMRT of both LV volumes and LVEF were poor, with significant underestimation of sp2DE and bp2DE measurements of LVEDV and remarkable overestimation of LVEDV and LVESV measured by bpCVG. However, as a more relevant finding than systematic errors for all measurements-LV volumes and LVEF-the standard deviations of the differences between 3DMRT and either bpCVG, bp2DE, or sp2DE were statistically significantly larger compared with the standard deviation of the differences between tomographic 3DE to 3DMRT.
The rotational scanning approach used in this study revealed 3D data
sets with high resolution. Compared with 2DE imaging, echo dropouts
were less relevant in the 3D reconstruction because of the compensatory
effect of information from adjacent wall portions and interpolation.
Limitations of the rotational approach for clinical studies include the
apical position of the transducer, which tends to be superior to the
anatomic apex of the left ventricle,30 and the
cone-shaped reconstruction, both potentially limiting the detection of
apical aneurysms in some cases, leading to underestimation of
LV volumes. The volume underestimation observed for the bp2DE
measurements10,28 is in addition caused by the
aneurysm shapes deforming the calculated ellipse (Fig 2
).
Similarly, this relationship leads to the underestimation found for
sp2DE volume measurements.
All in all, there is little information in the literature on determination of systolic function with the use of imaging techniques in patients with LV aneurysm, except for use of the cineventriculographic technique as the standard technique over a long period of time. Although previous studies comparing 2D and 3D imaging techniques with cineventriculography have shown good correlation and agreement for volume measurements in normal-shaped left ventricles,30 studies comparing 2D and 3D imaging techniques with cineventriculography have documented the limitation of even bpCVG for determining asymmetric ventricular volumes11,31 which is confirmed by our study.
Comparison to Previous Studies
The tomographic 3D echocardiographic technique
used here provides a highly accurate estimate of LV volumes without
geometric assumptions with the use of complete 3DE image information on
the heart with the ability to calculate 3DE image reconstructions with
a realistic gray-scale surface texture. Using this technique, in both
normal and pathologically shaped left ventricles, Kupferwasser et
al31 found an improved correlation between 3DE
and bpCVG measurements of LVEDV and LVESV
(y=0.95x-5.9, r=.99;
y=0.93x+1.9, r=.99, respectively) as
compared with bp2DE and bpCVG measurements
(y=0.84x-0.35, r=.98;
y=0.86x-1.64, r=.97, respectively).
In normal-shaped left ventricles, Nosir et al32
documented excellent correlation of 3DE with radionuclide angiography
for the calculation of LVEF (y=0.9x+3.7,
r=.99), with very small limits of agreement (-0.39% and
0.32%, bias=-0.035%).
Several quantitative evaluations of LV volumes by three-dimensional echocardiography have been performed by two different techniques, but both depend on a geometric assumption, although a 3D data set is acquired from multiple image planes. The first technique is polyhedral surface reconstruction to estimate ventricular surface and a large number of tetrahedrons reconstructed between image planes for volume computation.14,16 Sapin et al16 also documented accurate volume measurements in asymmetric left ventricles for this technique by using excised porcine hearts. This group also showed better correlation of this 3D technique with single-plane cineventriculography for LV volume estimation compared with bp2DE in patients with normal-shaped left ventricles. In their study they found equivalent measurements for bias and standard deviation of LVEF for 3DE (bias=6.6±9.8%) and bp2DE (bias=7.5±10.7%) compared with cineventriculography, which is different from the findings in the study presented here on patients with LV aneurysm.33 Applying this technique, systolic LV function was also evaluated by Gopal et al34 in unselected patients with suspected heart disease. The results for LVEF in that study comparing 3DE with equilibrium radionuclide angiography are close to the results in our own study comparing tomographic 3DE with 3DMRT. Recently, in young patients with pulmonary hypertension and compressed left ventricles, Apfel et al35 found results similar to those in our study with a slight underestimation of LVEDV and LVESV by 3DE (r=.94 and 0.87, bias=-6.9±6.9 mL and -16±11.2 mL, respectively) compared with 3DMRT, also with a poorer correlation for 2DE. LVEF by 3DE also had closer agreement with 3DMRT (bias=1.1±7.7%) than 2DE (bias=4.4±13.9%). As in our study, the bias results for measurements of LVEF were close to zero for 3DE as well as for 2DE, but the standard deviation as the more critical parameter was significantly larger for 2DE. The second technique combines the digitized video images with their spatial locations obtained by the spark gap triangulation system.15,17 Jiang et al36 validated the spark-gap technique for the determination of LV volumes in the presence of aneurysm and found good agreement of calculated in vitro volumes with actual values. In the same study in an in vivo animal model, Jiang et al showed good correlation of this 3D technique with actual volume values (LVEDV: r=.99, SEE=4.3 cm3; LVESV: r=.99, SEE=3.5 cm3) and LVEF (r=.99, SEE=2%) of ventricles with aneurysm. A slight underestimation was found for LVEDV, with a correlation of y=1.04x-4.2, similar to the study presented (y=0.97x-5.6).36
Limitations of the Study
No relevant differences in the image quality of the
cross-sectional images between tomographic 3DE and 3DMRT were found in
our study, providing manual planimetry with the same accuracy. However,
3DMRT imaging in this special study population was limited by reduced
blood pool detection in the aneurysms caused by blood stasis
and therefore limited delineation of the endocardial border in this
region.
Even though the acquisition of a 3DE data set provides accurate representation of the LV cavity shape, especially in asymmetric ventricles, there are few limitations that lead to minor underestimation. For the tomographic 3DE technique used in our study, the reason for this underestimation was mainly the cone-shaped data set with limited detection of apical aneurysms. For 3DE techniques using multiple image planes of the left ventricle from different views with the creation of wire-frame models of the ventricle,15,33 the reason for the underestimation is mainly the need for volume computation using geometric assumptions with the tendency to round off sharply protruding regions, as discussed by Pearlman37 previously as a serious limitation.
Manual endocardial border tracing in both tomographic techniques was a serious limitation because of the time factor. Integration of an automated border detection technique is a necessary prerequisite for the application of volume calculations by tomographic 3DE in clinical routine diagnoses. Another limitation was the image frequency of 25 images/s, as provided by the 3D system used in this study. This system does not allow selection of the image representing the exact end-diastolic and end-systolic points of time. Thus both 2D echocardiographic systems and 3D reconstruction systems need to be able to process higher image frequencies.
Clinical Implications
The tomographic 3DE technique used in this study for the first
time provides noninvasive clinical information on patients with LV
aneurysm before or after aneurysmectomy: accurate
quantification of LV volumes and systolic function as well as
dynamic 3D visualization of the real anatomy of the ventricle
and the aneurysm. Previous studies have described the
relationship of residual myocardium to the aneurysm
area in standard 2DE projections, but accurate determination of
heart size, ventricular function, and aneurysm size
has not been available.5 In comparison, with 3DE
imaging it should be possible also to determine size, shape,
localization, and dynamic characteristics of LV aneurysms, as
shown in the first preliminary results in human
subjects.38 This would provide preoperative
information for the surgeon about patch size and make it possible to
calculate the resulting ventricular size, shape, and
function after aneurysmectomy. Furthermore, with the use of 2D
techniques, the number of prognostic parameters before
aneurysmectomy is too small for reliable preoperative patient
selection, on the basis of noninvasive data.6
After the clinical validation presented, the next step in
clinical studies is to determine whether tomographic 3DE will improve
patient selection and decision making before aneurysmectomy,
intraoperative guiding, perioperative monitoring, and
postoperative control, as well as the prediction of outcome. To monitor
a patient for these criteria would require many repeat examinations.
Therefore it would be more ideal to use tomographic 3DE than 3DMRT
because it is less time consuming, less costly, and patient discomfort
is less.
Conclusions
The purpose of this study was to validate the clinical application
of tomographic 3DE by the routine transthoracic approach to
determine LV volumes and systolic function without geometric
assumptions in patients with LV aneurysms. Results from
tomographic 3DE correlated and agreed well with those by 3DMRT, with
better correlation and agreement than provided by current 2D
techniques. Unlike current 2D techniques, 3DE required no geometric
assumptions that limit accuracy. Tomographic 3DE was shown to be an
accurate technique that can be used in clinical routine because it is
noninvasive and less time consuming than other tomographic techniques.
This is the basis for further clinical studies and will probably point
to new procedures for preoperative selection and
perioperative monitoring of patients for
aneurysmectomy as for follow-up studies after myocardial
infarction to detect LV remodeling. Moreover, future developments to
improve the clinical applicability of 3DE volume quantification to
routine practice must include automated contour detection for faster
volume detection.
| Selected Abbreviations and Acronyms |
|---|
|
|
| Acknowledgments |
|---|
Received May 30, 1997; revision received August 19, 1997; accepted September 7, 1997.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
D. D. Glower and J. E. Lowe Left Ventricular Aneurysm Card. Surg. Adult, January 1, 2008; 3(2008): 803 - 822. [Full Text] |
||||
![]() |
H. Alkadhi, L. Desbiolles, L. Husmann, A. Plass, S. Leschka, H. Scheffel, R. Vachenauer, T. Schepis, O. Gaemperli, T. G. Flohr, et al. Aortic Regurgitation: Assessment with 64-Section CT Radiology, October 1, 2007; 245(1): 111 - 121. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Dewey, M. Muller, S. Eddicks, D. Schnapauff, F. Teige, W. Rutsch, A. C. Borges, and B. Hamm Evaluation of Global and Regional Left Ventricular Function With 16-Slice Computed Tomography, Biplane Cineventriculography, and Two-Dimensional Transthoracic Echocardiography: Comparison With Magnetic Resonance Imaging J. Am. Coll. Cardiol., November 21, 2006; 48(10): 2034 - 2044. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. M. Lang, V. Mor-Avi, L. Sugeng, P. S. Nieman, and D. J. Sahn Three-Dimensional Echocardiography: The Benefits of the Additional Dimension J. Am. Coll. Cardiol., November 21, 2006; 48(10): 2053 - 2069. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. C. Houck, J. E. Cooke, and E. A. Gill Live 3D Echocardiography: A Replacement for Traditional 2D Echocardiography? Am. J. Roentgenol., October 1, 2006; 187(4): 1092 - 1106. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Schaefer, G. P. Meyer, M. Fuchs, G. Klein, M. Kaplan, K. C. Wollert, and H. Drexler Impact of intracoronary bone marrow cell transfer on diastolic function in patients after acute myocardial infarction: results from the BOOST trial Eur. Heart J., April 2, 2006; 27(8): 929 - 935. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. M. Lang, M. Bierig, R. B. Devereux, F. A. Flachskampf, E. Foster, P. A. Pellikka, M. H. Picard, M. J. Roman, J. Seward, J. Shanewise, et al. Recommendations for chamber quantification Eur J Echocardiogr, March 1, 2006; 7(2): 79 - 108. [Abstract] [Full Text] [PDF] |
||||
![]() |
|