(Circulation. 1997;96:267-273.)
© 1997 American Heart Association, Inc.
Articles |
From the Division of Cardiology (D.N., A.B., M.G., M.H., P.D.), St. Michael's Hospital, University of Toronto, Telectronics Pacing Systems Inc (S.T.), Denver, Colo, COR Medical (S.T.), Toronto, Canada, and the Lahey-Hitchcock Medical Center (D.M.), Burlington, Mass.
Correspondence to Dr David Newman, MD, St. Michael's Hospital, Division of Cardiology, 30 Bond St, Toronto, ON, Canada, M5B 1W8. E-mail newmand{at}SMH.Toronto.on.ca
| Abstract |
|---|
|
|
|---|
Methods and Results One hundred twenty patients were prospectively studied with a single-model nonthoracotomy implantable cardioverter defibrillator (ICD) system at the time of implant and at 3 months. The pooled data of all shocks delivered to all patients were fitted to a logistic function to construct a defibrillation voltage/energy dose-response relationship. The crude logit curve was weighted in quartiles according to the average shock energy delivered per patient. Shocks at implant (n=802; 6.6±2.5 shocks/patient) and follow-up (n=292; 2.4±1.2 shocks/patient) were analyzed. The modeled voltage/energy required for 50% successful defibrillation (95% CI) in the pooled data was 367 V (273, 461) and 9.8 J (6.7, 12.9) at implant and 338 V (264, 412) and 10.5 J (8, 13.0) at follow-up. The conventional measure of lowest successful voltage/energy (95% CI) was 430 V (411, 449) and 12.1 J (11, 13.2) at implant and 415 V (391, 439) and 11.3 J (10, 12.6) at follow-up. There were no statistically significant differences between implant and follow-up energy requirements with either method.
Conclusions The nonthoracotomy lead system used in this study demonstrated stability of defibrillation energy requirements at implant and 3-month follow-up. A new technique for the estimation of the defibrillation energy dose-response relationship was derived by using a weighted logistic regression analysis.
Key Words: defibrillation modeling, mathematical electrophysiology arrhythmia defibrillator, implantable
| Introduction |
|---|
|
|
|---|
The compromise in general practice is to assess the minimum energy that successfully terminates induced VF.2 Test shocks of increasing or decreasing voltage and energy are delivered until at least one failed shock is preceded or followed by a success. This method gives a value commonly referred to as the defibrillation threshold, which corresponds in animal studies to the energy that would be successful in 50% to 60% of delivered defibrillating shocks.1 2 3 Conventional "single-point" defibrillation threshold measures exclude potentially useful data available from an analysis of all shocks delivered and cannot quantify data when all delivered shocks are successful. Furthermore, conventional defibrillation threshold measurement requires strict adherence to a uniform defibrillation protocol among participating centers. As a result, it is possible that a single-point measure of a defibrillation "threshold" may provide significant error in the assessment of the effects of new therapies or technological advances for ICD recipients. An alternative approach, used in this study, examines the defibrillation dose-response relationship by means of pooling the data from a population of patients. Such a method uses all data obtained from all shocks delivered to all patients and was used to assess the stability of a new transvenous defibrillation system in the same group of patients over a 3-month period.
Newer nonthoracotomy lead systems are now commonly available in which the shock pathway is between a combination of intrathoracic or extrathoracic electrodes including endovascular, subcutaneous sites, or the ICD pulse generator itself. Frequently, these systems use leads placed but not anchored in endovascular positions in the high right atrialsuperior vena cava position or more proximally at the innominate veinsuperior vena cava junction.4 The stability of defibrillation energy requirements for these systems in humans has had limited study. Some studies have found no change over time,5 6 7 whereas others have found a rise in energy requirements over a 6- to 12-week period.8 9
We hypothesized that the transvenous-lead system under study would demonstrate stability of defibrillation energy requirements over time. We further evaluated a new technique for summating all shocks delivered to allow a comparison of the energy and voltage requirements in the same population of patients over time and to compare these measures with the conventional LSE estimate of defibrillation threshold.
| Methods |
|---|
|
|
|---|
550 V) from induced VF. All
intraoperative testing used a model 4510 implant support device
(Telectronics Pacing Systems) with delivery of biphasic shocks at a
60:40 phase ratio (6-ms pulse duration) from a 150-microfarad
capacitor. The implanted defibrillators delivered shocks of identical
configuration. All shocks were initially started with a right
ventricular anodetoright atrial cathode configuration.
If unsuccessful, polarity was reversed prior to changing to a
subcutaneous lead.
|
A follow-up study was performed at 3 months in all patients. This study
used the same shock pathway and configuration that were employed at the
time of implant. At the follow-up study, the initial shock used was the
LSV at the time of the implant. If successful, defibrillation was again
attempted at this LSV minus 50 V. If unsuccessful, defibrillation was
attempted at 100 V higher (Fig 1B
).
In all cases, failed shocks were followed by a rescue shock at the maximal output of the implant support device (750 V) or the implanted defibrillator (650 V). All VF inductions were induced with 50-Hz stimuli available through the implant support device at implant or via the device's VF induction mode at follow-up. All testing was performed under general anesthesia with a balanced technique that included narcotics and inhalation anesthetics for implantation and propofol anesthesia at follow-up studies. Impedance to shock delivery was recorded as the impedance measured from a 500-V shock using the final implant configuration. This was compared with the impedance from a 500-V test shock at the follow-up study.
Data Analysis
Clinical variables assessed included patient demographics
and antiarrhythmic drug use. The shock parameters
recorded were stored voltage, calculated delivered energy based on
delivered voltage and measured impedance, and success or failure of the
defibrillation shock. The analysis used success or failure of
all delivered shocks (test or rescue) as the dependent variable;
delivered voltage or energy and time (implant versus 3 months) were the
covariates.
The data regarding shock voltage and energies delivered were fitted
with a logistic function for the construction of a defibrillation
voltage or energy dose-response relationship. By design, patients who
received lower energy shocks were specially selected by virtue of their
prior successful defibrillation at higher shock energies. As a result,
the crude overall logit curve should be relatively shallow and skewed
by the higher success rates for lower energy shocks. A new overall
logit curve was derived by weighing a separate logit function for four
equally sized, more homogenous patient groups that were divided in
quartiles according to the average shock energy (in volts or joules)
delivered per patient. The logit curve for each quartile gives a
different estimate for the probability of shock defibrillation success
for all energies given. In this manner, each curve describes how the
entire population would behave if all patients acted according to a
particular quartile. Each point on these four curves is then averaged
to produce a weighted logit curve according to the equation
![]() | (1) |
i is the predicted weighted
success rate,
and ß are constants, and
i refers to
an additive error term of discrepancies between the real data and the
model. The goodness of fit for the weighted logit curve to the data was
assessed with the average proportion of variation explained and the
variance of the regression "success rate" on the energy level.
(See "Appendix" for details of the statistical procedures
used.) From the weighted logit curve, the energies (or voltage) associated with a 50% or 80% probability of successful defibrillation (E50 or E80 in joules or V50 or V80 in volts) at implant and follow-up were obtained. These estimates were also compared with the conventional LSV at implant (LSVimplant) when available (ie, a failure preceded by a success pair available). The LSV at follow-up (LSVfollow-up) was limited to the two test defibrillation shocks used. A pair of test shocks could determine LSVfollow-up as either LSVimplant-50 V, LSVimplant, or LSVimplant+100 V, depending on the response to the abbreviated up-down protocol at follow-up. All descriptive data are presented as mean±SD. Comparison of continuous data was made by using the Student's t test.
| Results |
|---|
|
|
|---|
A total of 93 (77%) patients at implant and 84 (70%) at the 3-month follow-up did not take antiarrhythmic medications. A total of 6 (5%) and 10 (8%) patients were on class Ia drugs, and 22 (18%) and 26 (22%) were on class III drugs at implant and 3-month follow-up, respectively. Sixteen patients (13%) had a change of drug usage between implant and follow-up. These changes were evenly spread in all possible directions (4 patients from class Ia to class III, 4 patients from class III to Ia, 4 patients from class Ia to no therapy, and 4 patients from class III therapy to no therapy).
Defibrillation Voltage and Energy
A total of 1094 shocks were available for analysis. Of
these, 864 (79%) were successful and 230 (21%) were not successful.
At implant, 802 shocks were delivered (6.6±2.5 shocks per patient),
and 292 shocks were delivered during follow-up studies (2.4±1.2 shocks
per patient). The distribution of shock energies used, in volts or
joules, at implant and follow-up, are shown in Fig 2
.
The median voltage of all shocks was 500 V at implant and 450 V at the
3-month follow-up (Fig 2A
and 2B
). The corresponding median energies
for shock delivery were 16.4 and 13.4 J at implant and follow-up,
respectively (Fig 2C
and 2D
). The distribution of delivered shock
energies were divided into four equal-sized groups for the calculation
of the weighted logit curve.
|
The logistic regression curves, based on all patients, are illustrated
in Figs 3
and 4
. The effects of weighting
the logit curve according to the distribution of shock energies used
are illustrated in Fig 3B
and 3C
. All data in Fig 3
refer to the
population of patients studied at the time of implant, with the
delivered shock measured in volts. The accompanying mean defibrillation
success data points for the population at these delivered voltages are
superimposed on the logit curve. Note that the raw overall logit curve
in Fig 3A
is relatively shallow. As expected, this is
consistent with the generally higher proportion of successful
defibrillation results among the selected patients who received low
energies because of their prior successful defibrillation at higher
energies. Fig 3B
illustrates the four different logit curves obtained
when the predicted success rate over the entire energy range was
calculated based only on the results from the relatively more
homogenous subgroup that received a different quartile of net delivered
voltages, ie, from 0% to 25% (group 1), 25% to 50% (group 2), 50%
to 75% (group 3), and 75% to 100% (group 4) of all delivered
voltages. These four curves were then averaged to produce the weighted
logit curve for the overall population that is shown in Fig 3C
. Note
that the weighting increased the shape and shifted the calculated curve
downward into the domain of the independent variable. From this
final weighted logit curve, the V50 and V80
were calculated as summarized in the Table
.
|
|
|
By the same techniques, a weighted logit curve was constructed for
voltage requirements at the 3-month follow-up and for energy (in
joules) at implant and the 3-month follow-up. These results are
illustrated in Fig 4A
(for volts) and 4B (for joules). There is no
statistical difference between these two curves, ie, the covariate
"time" did not statistically alter the logit model, unlike the
parameters of voltage (P<.01) and joules
(P<.01). The goodness of fit, as assessed by the average
proportion of variation explained by the predicted logit function, was
31.8% for the weighted logit curve when energy was expressed in volts
and 23.9% in joules at baseline and 35.7% (volts) and 37.7% (joules)
at the 3-month follow-up.
Conventional point estimates of defibrillation threshold, ie, LSV (or
corresponding energy) are summarized in the Table
. The mean±SD
defibrillation threshold defined as LSV or joules was 430±81 V and
12.0±4.9 J at implant and 416±90 V and 11.3±4.8 J at the 3-month
follow-up, respectively. At implant, successful defibrillation at all
test shocks delivered occurred in 26 of 120 patients, precluding
measurement of the LSVimplant in these patients. In 8 of
these 26 patients, a minimal voltage of 250 V was reached. Similarly,
an LSVfollow-up value could not be obtained in 46 of 120
patients at follow-up, largely because of physician preference to
discontinue the protocol due to failure of a first test shock (n=5),
after delivery of at least two shocks (n=14), or with only one test
shock delivered at LSVimplant (n=13) or higher (n=14).
There were no instances of a failure of test shock at
LSVimplant followed by failure at
LSVimplant+100 V.
The Table
shows the E50, E80, V50,
and V80 and the 95% CIs derived from the weighted logit
curve, in joules and volts, at baseline and 3 months, respectively.
There was no significant change between any of these values. Also in
the Table
are the available conventional single-point measures of LSE,
which are within the 95% CI for E50 at baseline and the
3-month follow-up. There was no effect of age, sex, ejection fraction,
or drug use on defibrillation energy requirements.
The impedance for a 500-V shock was 48.2±7.8
at implant and
58.5±9.6
at the 3-month follow-up (P<.05).
| Discussion |
|---|
|
|
|---|
It should also be noted that 18% of the patients in this series
required additional implantation of a subcutaneous patch. This may be
related to the relatively stringent criteria used for establishing
minimal implant success (550 V or
13 J) or to the positioning of the
shock cathode on the atrial lead. Nonetheless, since the data were
compared in a pairwise fashion, the stability of the measures obtained,
using whatever configuration was ultimately required, remains the
same.
A further result of this study is the derivation of a weighted logistical regression model to assess the measurement of defibrillation energy requirements in a clinical population. Such a population-based method to assess energy needs may have certain advantages over commonly used methods. Most importantly, it allows the use of all shocks given to all patients in the assessment of therapy, even in situations in which relatively few shocks per patient are delivered or detailed defibrillation testing protocols have been incomplete or violated. This analysis pools all shocks delivered to a population independently of the type of defibrillation protocol used. By allowing the measurement of the predicted energy for 50% shock success rather than the use of an LSV, this method may also be more reliable for the study of effects of new therapies or interventions in ICD patients. Furthermore, it is possible that information based on the "shape" of the derived curve fitted to the data (ie, the calculated parameters for energies other than E50) may have relevance to clinical or basic research. The logistic function used in this study has been found to be a valid model to fit defibrillation energy requirements in both porcine2 and canine1 data. It is a function common to many biological models and, because of its sigmoidal shape, can model data in the low range of energy deliveries, unlike an exponential or other mathematical function.2
In contrast, all previous studies of defibrillation in humans have relied on the use of a single LSE, or "defibrillation threshold," as the measure of defibrillation energy requirements. Because not enough shocks are delivered in individual patients to assess the energydefibrillation success relationship, pooling data from all shocks in all patients allows useful estimates of this relation in the population under study. Importantly, published studies of defibrillation threshold, by design, exclude patients in whom defibrillation thresholds could not be obtained. The effects of a systematic bias in favor of those in whom a failure-success reversal is obtainable are unknown, but such a bias could lead to underestimation or overestimation of the utility of new therapies. The pooling of population data for modeling defibrillation efficacy has been attempted in animal models only.3 10 11 None of these attempts tried to weigh the effects of homogenous subgroups of the population as used in the present study, and none provided details of the statistical models used.
The use of a derived logit function for the pooled assessment of all shocks delivered in a patient population requires the assumption that interpatient variability is largely or entirely mediated through effects of shock magnitude on defibrillation success and not by any other factor. Similar to other studies, we found no influence of sex or left ventricular function on our results.12 13 14 Furthermore, all patients were analyzed twice, such that the entire data set represents paired-data information. Antiarrhythmic drugs could have had some confounding effects on the results; however, the relative magnitude of drug usage was low, and changes in drug use were evenly divided among all possible combinations.
The impedance measure differed at the time of implant and at the 3-month follow-up. The mechanism of this difference is not known and may be related to either time-dependent variabilities in the defibrillation lead-tissue interface, the defibrillation threshold, or an artifact related to differences in anesthetic agents or the surgical setting in this patient population. By whatever mechanism, the changes in impedance over time support the utility of assessing defibrillation energy requirements with respect to the actual voltage delivered by a device rather than a calculated energy parameter that is in turn dependent on the impedance to shock delivery.
Limitations
There are some important limitations in this study. There may
still be unidentified sources of variation within a
heterogeneous patient population, despite attempts to
mitigate the effects of interpatient heterogeneity by
grouping patients into quartiles of delivered energies. The comparator
used in this trial was that of the conventional LSE. Others have
suggested that single-point measures of LSE are reproducible and
accurate if performed in triplicate.2 This could not be
addressed in the present study; however, triplicate measures of LSE
have not been routinely performed in published human studies.
Conclusion
Transvenous defibrillation thresholds are stable over time
using the defibrillator system described. An analysis using a
summation of all shocks delivered in the aggregate may be a useful
alternative to the defibrillation "threshold" for estimating
defibrillation energy requirements. Such a method allows a large number
of data points to be fitted to a weighted logistic function, allowing
for the construction of a defibrillation energy dose-response
relationship to human data. This method allows estimates of the
V50 and E50 of clinically relevant delivered
energies.
| Selected Abbreviations and Acronyms |
|---|
|
| Acknowledgments |
|---|
| Appendix 1 |
|---|
|
|
|---|
i's (shock intensity in
volts or joules) distribution. The mean values
i of the independent
variable xij were calculated. Based on the
quartiles of
i's
distribution, the samples were divided into four groups as follows
1. the samples are ranked in terms of the
i
2. if
i
Q25,
then i
G1
3. if
Q25<
i
Med,
then i
G2
4. if
Med<
i
Q75,
then i
G3
5. if
i
Q75,
then i
G4
6. where Gi L=1,2,3,4 are the four groups.
Within each group the success rate of the x
variable was modeled with a logistic regression, where
![]() | (2) |
![]() | (3) |
![]() | (4) |
![]() | (5) |
i is the predicted weighted
success rate and
i refers to an additive
error term of discrepancies between the real data and the model.
The weighted predicted PWG,i describes the
success rate for the whole group. The
i
designed by linearizing the logit transformation enables the
development of an overall logistic curve for the whole data. The curve
fitting between gi and xi
indicated in all cases a coefficient of multiple determination, ie,
R2
1. The first step, partitioning of
samples into homogenous groups, allowed a better description of the
data. The second step, which involves the combination of
heterogeneous groups, resulted in a better fit to the
success rate for the whole population by significantly decreasing the
variance of the regression "success rate" on the variable
x.
The estimated value of the independent variable x for
any level of success rate is obtained as follows
![]() | (6) |
/ß
The goodness of fit of the model was assessed by using the
average proportion of variation explained (AVPE) derived by Gordon et
al15 and reviewed by Hosmer and
Lemenshow16
![]() | (7) |
=

pi,
and
=1-
, and by the unbiased
estimator of the variance of the regression "success rate" on the
independent variable x measured by
![]() | (8) |
![]() | (9) |
The 95% CIs for xp are calculated with the
t distribution with n-2 degrees of freedom as
follows
![]() | (10) |
Received October 7, 1996; revision received December 17, 1996; accepted January 9, 1997.
| References |
|---|
|
|
|---|
2. Jones DL, Irish WD, Klein GJ. Defibrillation efficacy: comparison of defibrillation threshold versus dose-response curve determination. Circ Res. 1991;60:45-51.
3.
Schuder JC, Stoeckle H, West JA, Keskar PY, Gold
JH, Denniston RH. Ventricular defibrillation in
the dog with a bielectrode intravascular catheter. Arch
Intern Med. 1973;132:286-290.
4. Stajduhar KC, Ott GY, Kron J, McAnulty JH, Oliver RP, Reynolds BT, Adler SW, Halperin BD. Optimal electrode position for transvenous defibrillation: a prospective randomized study. J Am Coll Cardiol. 1996;27:90-94.[Abstract]
5. Hsia H, Mitra RL, Flores BT, Marchlinski FE. Early postoperative increase in defibrillation threshold with nonthoracotomy system in humans. PACE Pacing Clin Electrophysiol. 1994;17:1166-1173.[Medline] [Order article via Infotrieve]
6. Hsia HH, Rothman SA, Thome LM, Adelizzi NM, Whitley DM, Buston AE, Miller JM. Early postoperative stability in transvenous single-lead defibrillation system in man. Circulation. 1994;90(suppl I):I-122. Abstract.
7. Schwartzman D, Callans DJ, Gottlieb CD, Marchlinski FE. Biphasic shock attenuates the early rise in defibrillation threshold after implantation of a nonthoracotomy lead system. Circulation. 1994;90(suppl I):I-122. Abstract.
8.
Venditti FJ, Martin DT, Vassolas G, Bowen S.
Rise in chronic defibrillation thresholds in nonthoracotomy implantable
defibrillator. Circulation. 1994;89:216-223.
9. Martin DT, John R, Venditti FJ. Increase in defibrillation threshold in non-thoracotomy implantable defibrillators using a biphasic waveform. Am J Cardiol. 1995;76:263-266.[Medline] [Order article via Infotrieve]
10. Ewy GA, Horan W. Electrode catheter for transvenous defibrillation. Med Instrument. 1976;10:155-158.
11.
Schuder JC, Rahmoeller GA, Stoeckle H.
Transthoracic ventricular defibrillation with
triangular and trapezoidal waveforms. Circ Res. 1966;19:689-694.
12. Raitt MH, Johnson G, Lee G, Poole JE, Kudenchuk PJ, Bardy GH. Clinical predictors of the defibrillation threshold with the unipolar implantable defibrillation system. J Am Coll Cardiol. 1995;25:1576-1583.[Abstract]
13. Strickberger SA, Brownstein SL, Wilkoff BL, Zinner AJ. Clinical predictors of defibrillation energy requirements in patients treated with a nonthoracotomy defibrillator system. Am Heart J. 1996;131:257-260.[Medline] [Order article via Infotrieve]
14. Dorian P, Connolly S, Yusuf S. The impact of left ventricular dysfunction on outcomes with the implantable defibrillator. Am Heart J. 1994;127:1159-1163.[Medline] [Order article via Infotrieve]
15. Gordon T, Kannel WB, Halperin M. Prediction of coronary heart disease. J Chron Dis. 1979;31:427-440.
16.
Lemenshow S, Hosmer DW. A review of goodness of
fit statistics for use in the development of logistic regression
models. Am J Epidemiol. 1982;115:92-108.
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1997 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |