(Circulation. 1997;95:2162-2168.)
© 1997 American Heart Association, Inc.
Articles |
From the Departments of Radiology (M.J.A.S., X.G., M.U., J.J.C.-C., K.A.) and Pathology-Laboratory Medicine (J.L.T.), University of Minnesota Hospital and Clinic (Minneapolis), and The Jesse E. Edwards Registry of Cardiovascular Disease (J.L.T.), United Hospital, St Paul, Minn.
Correspondence to Kurt Amplatz, MD, Section of Cardiovascular and Interventional Radiology, Department of Radiology, University of Minnesota Hospital and Clinic, Box 292-Mayo, 420 Delaware St, SE, Minneapolis, MN 55455. E-mail sharafuddinm{at}mirlink.wustl.edu
| Abstract |
|---|
|
|
|---|
Methods and Results Percutaneous closure of surgically created fossa ovalis ASD was attempted in 15 minipigs. The mean balloon-stretched ASD diameter was 12.3±2.3 mm (range, 10 to 16 mm). The self-expanding prosthesis was braided from 0.005-in Nitinol wires in the shape of two flat buttons with a short connecting waist with a diameter corresponding to that of the defect to be closed. Polyester filling was added to enhance thrombogenicity. Pulmonary arteriography with levo-phase was obtained before placement; immediately after placement; and at 1-week, 1-month, and 3-month follow-ups. Four animals were killed at 1 week, 1 month, and 3 months for histopathological correlation. Three deaths resulted from ventricular fibrillation (one during anesthesia and two during the placement procedure). Successful placement of the prosthesis was achieved in the remaining 12 animals. Overall immediate ASD closure on angiography occurred in 7 of 12 animals (all polyester-filled prostheses). Absent or trace shunt by angiography was present in 11 of 12 devices at 1 week, with the remaining one demonstrating a small shunt. All septal defects were completely closed at 1 month with the exception of one case in which delayed partial dislodgment of an undersized prosthesis into the right atrium had developed. Closure rate at 3 months was 100%. Neoendothelialization and fibrous incorporation of the prosthesis were completed within 1 to 3 months.
Conclusions Effective and permanent occlusion of secundum ASDs is feasible with a device that offers the advantages of easy placement, self-centering, and repositionability.
Key Words: heart septal defects heart defects, congenital pediatrics shunts
| Introduction |
|---|
|
|
|---|
|
|
| Methods |
|---|
|
|
|---|
|
|
For introduction, the prosthesis was collapsed and advanced through a simple delivery system consisting of a 6F or 7F thin-walled, nontapered, kink-resistant Teflon introducing sheath and a 0.038-in delivery cable. Occluders of >10 mm were woven from 0.005-in wires and required a 7F introducing sheath, whereas smaller prostheses were woven from 0.004-in wires and could be introduced through a 6F sheath. Before deployment, the prosthesis was attached to the delivery cable with a microscrew connection and withdrawn into a loader for introduction into the delivery catheter. The device was then pushed through the delivery catheter.
Animal Model
All animals were treated according to the "Principles of
Laboratory Animal Care" of the National Society for Medical Research
and the "Guide for the Care and Use of Laboratory Animals" (NIH
publication No. 80-23, revised 1985). The study protocol was
preapproved by the institutional animal care committee. After sedation
of the animals with an intramuscular injection of Telazol (1:1
tiletamine/zolazepam; 8 mg/kg), general anesthesia was
induced through mask inhalation of 1% isoflurane and maintained
through endotracheal inhalation of 1% isoflurane during mechanical
endotracheal ventilation. All chronic experimental procedures were
performed under sterile conditions with continuous ECG monitoring.
A surgical model for the creation of ASD is preferred over
percutaneous transcatheter dilatation of
the foramen ovale to ensure closer resemblance to defects in
humans.13 The choice of the animal species is also very
important to ensure a suitable atrial septal anatomy for the
creation of a secundum-type defect. After a brief initial experience
with surgically created defects in dogs, we abandoned this animal model
because of the poorly developed fossa ovalis and small size of the
atrial septum in the canine. Consequently, a swine model was adopted
because of the well-developed fossa ovalis and comparable atrial septal
anatomy to that of humans, allowing the creation of defects
that closely mimic human secundum ASDs in both size and location.
Nineteen Yucatan minipigs, of either sex and weighing 25 to 30 kg,
underwent surgery for ASD creation. With the animals under general
anesthesia and with the use of sterile conditions, a
transverse left thoracotomy was performed that exposed the heart, and
through a vertical pericardiotomy, clamps were placed across the left
and right atrial appendages; both appendages were entered through
purse-string sutures. The left index finger was inserted into the right
atrium to palpate the fossa ovalis while a sharp punch instrument was
introduced from the left atrial appendage through the purse-string
suture. After being guided to the fossa ovalis with the opposing index
finger, the instrument was thrust through the septum, which was
captured in the instrument and cut with a sliding sharp knife. A 10-mm
cutter was used in all animals, resulting in a mean balloon-stretched
defect diameter of 12.3±2.3 mm (range, 10 to 16 mm). After
creation of the ASD, the chest was closed, and the animal was allowed
to recover for
1 week before the percutaneous closure
procedure.
Occlusion Technique
With the animal under general anesthesia and with
the use of sterile conditions, transvenous vascular access was
established via cutdown and vascular sheath placement into the femoral
or jugular vein (10 and 2 animals, respectively). Baseline blood gas
and oxygen saturation levels were obtained from the left atrium, main
pulmonary artery, and superior vena cava. The ASD was
visualized through levo-phase pulmonary angiography or direct
left atrial contrast injection. The defect was crossed with an
angiographic catheter, and a J-tipped guide wire was introduced. A
volume-diametercalibrated 7F Berman balloon catheter was exchanged
over the wire and positioned in the left atrium. The balloon catheter
was inflated with air and pulled back through the defect under
fluoroscopic observation. Slight deformity of the balloon contour
during pullback established the stretched diameter of the septal
defect. The balloon catheter was removed, reinflated with the same
amount of air, and passed through various holes in a sizing plate to
determine the stretched diameter of the defect. Appropriate
prosthesis size was then chosen according to this
balloon-stretched diameter.24 25 The guiding sheath was
introduced through the septal defect over the guide wire. The
prosthesis was attached to the delivery cable and withdrawn
into an adapter tube (loader). The prosthesis was then loaded
into the guiding sheath and advanced through pushing of the delivery
cable. With fluoroscopic guidance, the left atrial button was released
into the left atrium, with care taken to avoid the left atrial
appendage. While constant tension was maintained on the delivery cable,
the right atrial button was then deployed through withdrawal of the
delivery sheath. Proper placement was verified fluoroscopically and
manually, through pushing and pulling on the delivery cable, before the
prosthesis was detached by turning the delivery cable in a
counterclockwise direction with a vice. Misplacement of both buttons
into the left atrium occurred occasionally, which was easily corrected
by recapturing the left atrial button into the guide catheter and
redeploying the device. Pulmonary arteriography and blood gas
measurements were performed after deployment to check for residual
shunt.
Follow-up Studies
Sequential angiographic studies and blood gas measurements were
performed at 1-week, 1-month, and 3-month intervals. With the animal
under general anesthesia, the pulmonary artery was
catheterized via transfemoral or transjugular venous cutdown, and the
atrial septum was imaged in a shallow (15° to 20°) left anterior
oblique projection. Residual shunts were angiographically graded
subjectively by two independent observers and classified as trace
(barely perceptible opacification), small (faint right atrial
opacification with or without faint jet), moderate (obvious right
atrial opacification less than the left atrium with or without a
clearly visible jet), or large (bright right atrial opacification).
Comparisons of shunt closure rates after device placement and of mean
Qp/Qs ratios before closure, immediately after
closure, and on follow-up were performed using a two-tailed nonpaired
Student's t test. Transthoracic
echocardiography with color Doppler was also
used in a number of animals during placement and on follow-up
examinations.
Histopathological Correlation
Animals were killed after 1 week, 1 month, or 3 months (four
animals per time interval). The heart and great vessels were explanted
and fixed in a buffered physiological solution
containing 10% formaldehyde or 2% glutaraldehyde for
histopathology or scanning electron microscopy.
| Results |
|---|
|
|
|---|
Technical Success
Percutaneous closure of surgically created defects
of the fossa ovalis was attempted in 15 minipigs. As mentioned, 3
animals died as a result of ventricular fibrillation, with
one death occurring during anesthesia. In the remaining 12
animals, placement of the prosthesis was successful, resulting
in an overall success rate of 85% (12/14) and a technical success rate
of 100%. The mean balloon-stretched ASD diameter was 12.3±2.3 mm
(range, 10 to 16 mm), whereas the mean device waist diameter was
11±2.3 mm (range, 8 to 16 mm). The addition of Dacron
filling slightly increased the resistance to the initial advancement of
the device through the introducing sheath but did not appreciably alter
the ease of deployment. Successful deployment was independent of the
approach (transjugular or transfemoral) or the angle of attack. Once
the ASD was crossed, placement of the device could be accomplished in
<1 minute.
Preclosure angiography demonstrated a large left-to-right atrial shunt
in all animals, with a corresponding mean Qp/Qs
ratio of 1.7±0.55. Immediately after the placement of the
prosthesis, complete closure of the defect on angiography was
noted in 7 of the 12 animals (58%) (Fig 4
), whereas the
remaining animals demonstrated only small shunts. Significant lowering
of the mean Qp/Qs ratio occurred immediately
after device placement (1.0±0.15, P<.02). Significantly
higher immediate closure rates were achieved with polyester-filled
prostheses compared with rates for nonfilled prostheses (86% versus
0%; P=.018). A tendency toward a higher rate of immediate
closure was also noted when the prosthesis waist diameter
closely fit the defect (87%) compared with prostheses that were
undersized by >3 mm (33%). However, the difference did not reach
statistical significance (P=.09), which is probably the
result of the small sample size.
|
Angiographic Follow-up
Follow-up angiography at 1 week demonstrated complete angiographic
shunt closure in 8 of 12 animals (67%), whereas persistent trivial
shunts were demonstrated in 3, and persistent small shunts were
demonstrated in the remaining animal. Angiographic follow-up of 8
animals, in which follow-up was available at 1 month after device
placement, showed complete shunt closure in all except 1 animal (88%
closure rate), in which angiography showed the prosthesis to
have become partially dislodged into the right atrium with a large
residual left-to-right atrial shunt. In that animal, a 10-mm-diameter
device had been placed across a 16-mm defect because of the
unavailability of a larger device at that time. Three-month
angiographic follow-up, which was available in 4 animals, showed
complete closure in all (100% closure rate). The mean
Qp/Qs ratio on the various follow-up intervals
showed no significant difference from the immediate postplacement value
(mean difference, 0.02 to 0.2; P>.1).
Pathology
The device occupied the region of the fossa ovalis in all animals,
with the exception of one device found at 1 month to be partially
dislodged into the right atrium. In all cases, there was no compromise
of the orifices of the left pulmonary veins, coronary
sinus, or other vital atrial structures by the retention buttons of the
device.
Four specimens were examined at 1 week after device placement. Three of the devices used in these specimens were not filled with polyester fibers. The wire mesh of the polyester-filled prosthesis was completely invested by a layer of organizing fibrin. Complete or near-complete coverage of the wire mesh with organizing fibrin thrombus occurred in two nonfilled devices, whereas the remaining one was only partially covered. Two of the nonfilled prostheses were patent to manual probing.
Postmortem gross and microscopic examination of seven devices at 1 to 3
months after placement showed complete or near-complete neoendocardium
coverage of both the right and left atrial discs. On visual inspection,
a smooth, shiny, glistening pseudoendocardium was found to cover both
aspects of the prosthesis (Fig 5
). One device,
found at 1 month to be partially dislodged, demonstrated only partial
endothelialization and a firm fibrous connection with
the atrial septum along its inferior margin. Examination
with scanning electron microscopy in one specimen showed both surfaces
of the device to be completely covered with flat
neoendothelial cells, continuous with the endocardium
of the surrounding atrial septal rim.
|
The lungs did not have emboli or other abnormalities on gross morphological inspection in any of the cases.
| Discussion |
|---|
|
|
|---|
50% of secundum ASDs were found to be amenable to closure with
the clamshell device.26 The presented
prosthesis requires the presence of a smaller septal rim around
the defect, which is <70% of that required with the clamshell
occluder. Current ASD closure devices primarily accomplish closure
through the application of a patch to the defect, which may not
completely adapt to the variable thickness of the septal rim,
particularly with devices in which the retention flanges are widely
separated24 25 (Fig 1BDevice dislodgment can occur if the size of the defect greatly exceeds the waist diameter of the device or approaches the diameter of the retention buttons, as did occur in 1 animal in our study. On the other hand, placement of a disproportionately large device may result in mushrooming of the retention buttons and weakening of the cross-clamping forces against the septal rim, which increases the risk of blood flow behind the discs and may result in incomplete endothelialization.
Follow-up studies of the clamshell occlusion device reported a
delayed rate of metal fatigue fractures of one or more arms of
30%.3 However, unlike acute device failure that occurs
as a result of frame or strut fractures,16 delayed device
fracture usually poses little clinical significance provided that no
fragmentation or embolization ensues. Fatigue fractures that occur as a
result of repetitive bending (
100 000 cardiac contractions per day)
are less likely to occur with the presented device because of
the low profile and round configuration of the retention discs, which
do not extend to the free atrial wall. This configuration should also
reduce the risk of atrial perforations compared with designs with
struts or sharp corners, which can come in contact with the atrial wall
during cardiac contraction. Another important factor preventing
mechanical failure with the presented device is the exceptional
strength and high energy-absorbing capacity of Nitinol,29
which enable it to dissipate four times more strain compared with steel
alloy.30 Furthermore, the densely intermeshed wire network
should reduce the risk of embolization in case of wire breakage.
Although a primarily angiographic follow-up was used in our study, echocardiography with color Doppler may be better suited in clinical applications, both during device placement as an adjunctive cross-sectional imaging modality to assess the relation of the prosthesis to vital atrial structures and for postplacement follow-up of residual shunts and device position.4 11
In conclusion, the small introduction system, simple and reliable placement technique, and favorable initial experimental success demonstrate the promising potential of this device for the percutaneous closure of secundum ASDs in all age groups. The device is also being modified to allow treatment of muscular ventricular septal defects. Anticoagulation was not used in our experiments. However, heparinization is advocated in clinical use to lower the risk of catastrophic systemic embolization.27 31 At present, the main drawbacks of the presented device include the requirement of accurate defect sizing, small number of animals used in the present study, lack of adequate long-term follow-up, and limited clinical experience.
|
| Acknowledgments |
|---|
| Footnotes |
|---|
Received June 24, 1996; revision received November 18, 1996; accepted November 21, 1996.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
J. Ritzema, I. C. Melton, A. M. Richards, I. G. Crozier, C. Frampton, R. N. Doughty, J. Whiting, S. Kar, N. Eigler, H. Krum, et al. Direct Left Atrial Pressure Monitoring in Ambulatory Heart Failure Patients: Initial Experience With a New Permanent Implantable Device Circulation, December 18, 2007; 116(25): 2952 - 2959. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Giardini and A. Donti Enhanced Thrombogenesis is Physiologic After Transcatheter Closure of Patent Foramen Ovale Stroke, July 1, 2007; 38(7): e55 - e55. [Full Text] [PDF] |
||||
![]() |
M. Sigler and C. Jux Biocompatibility of septal defect closure devices Heart, April 1, 2007; 93(4): 444 - 449. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Jux, H. Bertram, P. Wohlsein, M. Bruegmann, and T. Paul Interventional Atrial Septal Defect Closure Using a Totally Bioresorbable Occluder Matrix: Development and Preclinical Evaluation of the BioSTAR Device J. Am. Coll. Cardiol., July 4, 2006; 48(1): 161 - 169. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Schwerzmann and O. Salehian Hazards of percutaneous PFO closure Eur J Echocardiogr, December 1, 2005; 6(6): 393 - 395. [Full Text] [PDF] |
||||
![]() |
B Meier Closure of patent foramen ovale: technique, pitfalls, complications, and follow up Heart, April 1, 2005; 91(4): 444 - 448. [Full Text] [PDF] |
||||
![]() |
J. Masura, P. Gavora, and T. Podnar Long-term outcome of transcatheter secundum-type atrial septal defect closure using Amplatzer septal occluders J. Am. Coll. Cardiol., February 15, 2005; 45(4): 505 - 507. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Massimo, B. Gianfranco, and C. Mario Risk of thrombus formation on devices used to close transcatheter atrial septal defect and patent foramen ovale J. Am. Coll. Cardiol., October 19, 2004; 44(8): 1712 - 1712. [Full Text] [PDF] |
||||
![]() |
U. Krumsdorf, S. Ostermayer, K. Billinger, T. Trepels, E. Zadan, K. Horvath, and H. Sievert Incidence and clinical course of thrombus formation on atrial septal defect and patient foramen ovale closure devices in 1,000 consecutive patients J. Am. Coll. Cardiol., January 21, 2004; 43(2): 302 - 309. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. W. Moore and D. S. Levi Transcatheter closure of atrial shunts: Focus on a lingering issue J. Am. Coll. Cardiol., January 21, 2004; 43(2): 310 - 312. [Full Text] [PDF] |
||||
![]() |
M. Zanchetta, E. Onorato, G. Rigatelli, L. Pedon, M. Zennaro, A. Carrozza, and P. Maiolino Intracardiac echocardiography-guided transcatheter closure of secundum atrial septal defect: A new efficient device selection method J. Am. Coll. Cardiol., November 5, 2003; 42(9): 1677 - 1682. [Abstract] [Full Text] [PDF] |
||||
![]() |
G Fischer, J Stieh, A Uebing, U Hoffmann, G Morf, and H H Kramer Experience with transcatheter closure of secundum atrial septal defects using the Amplatzer septal occluder: a single centre study in 236 consecutive patients Heart, February 1, 2003; 89(2): 199 - 204. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Rickers, M. Jerosch-Herold, X. Hu, N. Murthy, X. Wang, H. Kong, R. T. Seethamraju, J. Weil, and N. M. Wilke Magnetic Resonance Image-Guided Transcatheter Closure of Atrial Septal Defects Circulation, January 7, 2003; 107(1): 132 - 138. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. K. Verma, S. K. S. Thingnam, A. Sharma, J. S. Taneja, J. S. Varma, and A. Grover Delayed Embolization of Amplatzer Septal Occluder Device: An Unknown Entity: A Case Report Angiology, January 1, 2003; 54(1): 115 - 118. [Abstract] [PDF] |
||||
![]() |
Z.-D. Du, Z. M. Hijazi, C. S. Kleinman, N. H. Silverman, K. Larntz, and Amplatzer Investigators Comparison between transcatheter and surgical closure of secundum atrial septal defect in children and adults: Results of a multicenter nonrandomized trial J. Am. Coll. Cardiol., June 5, 2002; 39(11): 1836 - 1844. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Jux, H. Bertram, P. Wohlsein, M. Brugmann, P. Wuboldt, C. Fink, T. Paul, and G. Hausdorf Experimental ASD closure using autologous cell-seeded interventional closure devices Cardiovasc Res, January 1, 2002; 53(1): 181 - 191. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. G Alekyan, V. A Boukharin, V. P Podzolkov, and M. G Poursanov Residual Aortopulmonary Window Closure With Amplatzer Septal Occluder Asian Cardiovasc Thorac Ann, September 1, 2001; 9(3): 229 - 231. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. S. Rao, F. Berger, C. Rey, J. Haddad, B. Meier, K. P. Walsh, J. S. Chandar, T. R. Lloyd, J. Suarez de Lezo, R. Zamora, et al. Results of transvenous occlusion of secundum atrial septal defects with the fourth generation buttoned device: comparison with first, second and third generation devices J. Am. Coll. Cardiol., August 1, 2000; 36(2): 583 - 592. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y V Maeno, L N Benson, P R McLaughlin, and C Boutin Dynamic morphology of the secundum atrial septal defect evaluated by three dimensional transoesophageal echocardiography Heart, June 1, 2000; 83(6): 673 - 677. [Abstract] [Full Text] |
||||
![]() |
K C Chan, M J Godman, K Walsh, N Wilson, A Redington, and J L Gibbs Transcatheter closure of atrial septal defect and interatrial communications with a new self expanding nitinol double disc device (Amplatzer septal occluder): multicentre UK experience Heart, September 1, 1999; 82(3): 300 - 306. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. Amin, X. Gu, J. M. Berry, J. L. Bass, J. L. Titus, M. Urness, Y.-M. Han, and K. Amplatz New Device for Closure of Muscular Ventricular Septal Defects in a Canine Model Circulation, July 20, 1999; 100(3): 320 - 328. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. Amin, X. Gu, J. M. Berry, J. L. Titus, S. S. Gidding, and A. P. Rocchini Periventricular closure of ventricular septal defects without cardiopulmonary bypass Ann. Thorac. Surg., July 1, 1999; 68(1): 149 - 153. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. Amin, A. P Rocchini, and D. F Wax Closure of Residual Shunts After Fontan Operation With Amplatzer Occluder Asian Cardiovasc Thorac Ann, June 1, 1999; 7(2): 121 - 123. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Fischer, H.H. Kramer, J. Stieh, P. Harding, and O. Jung Transcatheter closure of secundum atrial septal defects with the new self-centering Amplatzer Septal Occluder Eur. Heart J., April 1, 1999; 20(7): 541 - 549. [Abstract] [PDF] |
||||
![]() |
B. D. Thanopoulos, G. S. Tsa |