(Circulation. 1997;95:2098-2107.)
© 1997 American Heart Association, Inc.
Articles |
From the Service de Cardiologie A (D.K., D.L., S.K., J.L.), Service d'Exploration Fonctionnelle (C.S., L.G.), and Service de Chirurgie Cardiovasculaire B (D.G., J.L.H.), Hôpital Cardiologique de Lille, France.
Correspondence to Dr Didier Klug, Service de Cardiologie A, Hôpital Cardiologique de Lille, Boulevard du Pr Leclercq, CHR-U 59037 Lille, France.
| Abstract |
|---|
|
|
|---|
Methods and Results Fifty-two patients were admitted for endocarditis related to PM-lead infection. The presentation was acute in 14 patients, with onset of symptoms in the first 6 weeks after the last procedure on the implant site, and chronic in 38 patients. Fever occurred in 86.5%. Clinical and/or radiological evidences of pulmonary involvement were observed in 38.4%. Pulmonary scintigraphy disclosed pulmonary infarcts in 31.2%. Local complications were found in 51.9%. Elevated C-reactive protein was found in 96.2%. A germ was isolated in 88.4% of patients and was a Staphylococcus in 93.5%. Transthoracic echocardiography demonstrated vegetations in only 23% of patients, whereas transesophageal echocardiography disclosed abnormal appearances on the PM lead in 94%. We systematically tried to ablate all the material. Two techniques were used: percutaneous ablation or surgical removal during extracorporeal circulation. All patients were treated with antibiotics after removal of the infected material. Two patients died before lead removal and 2 after surgical removal; the predischarge mortality was 7.6%, and the overall mortality was 26.9% after a follow-up of 20.1±13 months.
Conclusions The diagnosis of endocarditis related to PM-lead infection should be suspected in the presence of fever, complications, or pulmonary lesions after PM insertion. Transesophageal echocardiography should be performed to look for vegetations. Staphylococci are involved in the majority of these infections. The endocardial system must be entirely removed and appropriate antibiotic therapy pursued for 6 weeks.
Key Words: endocarditis pacing pacemakers
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
Definitions
The diagnosis of endocarditis on PM leads was considered with
the use of the Duke criteria11 (Tables 1
and 2![]()
).
|
|
Investigations
All patients underwent the following procedures:
1. Blood cultures on 3 consecutive days (three cultures from each patient with additional cultures if the temperature was >38.5°C or <36°C). Cultures at the site of battery implantation were performed when appropriate (wounds, local infection, or PM exteriorization). All cultures were performed before antibiotic therapy except in the case of patients referred from other institutions who were already receiving antibiotic therapy and in whom a germ had been isolated before admission.
2. Cultures of the leads and of the PM were done after ablation. In the majority of patients, the ablation was performed without interruption of the antimicrobial therapy. We did not look for rarer pathogens such as Legionella or Q fever, which have been found in endocarditis.
3. Repeated blood counts and measurements of creatinine, CRP, CICs, fibrinogen, and the erythrocyte sedimentation rate.
4. TTE and TEE performed by two different investigators. Serial echocardiograms for suspicion of an infected lead were performed by the same two investigators (except for four patients).
5. Chest roentgenogram.
6. Ventilation perfusion pulmonary scintigraphy before and after lead ablation in the majority of cases.
To exclude other sources of infection, the following were used:
1. Dental pantomogram, sinus radiographs, and specialist referral where appropriate.
2. Abdominal ultrasound.
3. Specialist referral and accessory investigations when warranted by the clinical findings.
Management
Ablation of the Material
We routinely tried to ablate all the material (PM and all
leads). Two techniques were used:
1. Percutaneous ablation. The leads were removed by simple traction. If the tip was fixed in the myocardium, a stylet wire (Cook pacemaker-leadextraction system) was used, and if the lead stopped in the sternoclavicular angle, special sheaths or a lasso introduced via the femoral vein was used to complete the ablation.
2. Surgical removal during extracorporeal circulation.
The choice of the technique of ablation (surgical versus
percutaneous) was based on the size of the vegetations
on TEE, the presence of an alteration of the tricuspid valve, and the
general condition of the patient. After the seventh patient, we adopted
a standardized policy for lead removal (Fig 1
): (1) if
TEE demonstrated vegetation >10 mm, the material was surgically
removed; (2) if TEE did not demonstrate vegetations or demonstrated
vegetation
10 mm, the material was
percutaneously removed.
|
Antimicrobial Therapy
All patients received intravenous antibiotics before
lead ablation. This therapy was adapted to the germ found at the blood
culture. If blood cultures were negative, empirical antistaphylococcal
therapy was given (usually vancomycin and amikacin, except for patients
referred from other institutions with alternative treatment
combinations that were clinically effective). Two antibiotics were
usually used, but 3 were administered in case of pulmonary
infection. Antibiotic therapy was continued intravenously
for 2 weeks after lead ablation and subsequently by mouth for 4 weeks;
the antibiotic was adapted to the germ found at lead culture.
Antibiotic therapy was discontinued after outpatient consultation to
verify the absence of recurrent symptoms and normalization of the
systemic parameters of inflammation.
Pacing After Ablation
An abdominal PM with permanent epicardial stimulation was
inserted after surgical or percutaneous removal in
patients requiring a permanent ventricular stimulation. In
the other cases, the indication for permanent pacing was reevaluated.
If necessary, a new transvenous permanent PM was inserted 1 to 2 months
after the percutaneous removal of the material.
Follow-up
An outpatient clinic was used for follow-up 4 weeks after
discharge. Further follow-up was obtained by phone contact with the
patients and their physicians.
Statistical Analysis
Data were analyzed by paired t tests or in
two-by-two tables using the
2 test and Fisher's
exact test. A value of P<.05 was accepted as
significant.
| Results |
|---|
|
|
|---|
The presentation was acute in 14 patients (6 women and 8 men aged 72.6±12.7 years), with onset of symptoms in the first 6 weeks after the last procedure on the implant site, and chronic in 38 patients (15 women and 23 men aged 61.9±18.6 years), with >6 weeks from the last procedure on the implant site to the onset of symptoms. Only 3 patients in the chronic group had the first symptom of infection 6 to 12 weeks after the last procedure.
In the acute group, 9 patients (64.3%) had a definite infective endocarditis on the basis of the Duke criteria, and 5 (35.7%) had a possible infective endocarditis with one major clinical criteria but fewer than three minor clinical criteria.
In the chronic group, 36 patients (94.7%) had a definite infective endocarditis on the basis of the Duke criteria, and 2 (5.3%) had a possible infective endocarditis with one major clinical criteria but fewer than three minor clinical criteria.
Thirteen patients (25%; 3 of 14 in the acute group and 10 of 38 in the chronic group) had impaired immunity due to diabetes mellitus in 3 patients, a neoplasm in 4, corticosteroid therapy in 2, alcohol abuse in 2, hemodialysis for chronic renal failure in 1, and an eosinophilic myocardiopathy in 1.
In the acute group, the last intervention at the PM site was a first
implantation in 11 patients (2 of whom required a new intervention due
to lead dysfunction) (Fig 2
). In the remaining 3
patients, it was a lead change or repositioning several months after
the first PM implantation.
|
Four patients had temporary endocardial pacing before implantation. In two patients, a hematoma occurred after the PM insertion that required evacuation in one case.
In the chronic group, the last intervention was a first implantation in
14 patients, elective PM replacement in 3, lead replacement in 8, and
PM exteriorization or local infection in the remaining 13 (Fig 2
). The
mean number of local interventions at the PM site was 2.5±1.7 (range,
1 to 8) with a mean lead number of 2.1 (range, 1 to 4). One patient had
temporary endocardial pacing after right femoral vein
catheterization, and 2 patients had an
electrophysiological study with a right
femoral vein catheterization just before or just after
the PM insertion. Three patients developed a hematoma after
implantation.
Symptoms (Fig 3
)
In the acute group, the time elapsed between the last intervention
and the first symptom was 4 days (range, 1 to 12 days). Fever was
present in 13 patients (92.8%). Local pain at the PM site was
found in 6 patients (42.8%), followed by a purulent discharge in 3 of
them. Pneumonia was observed in 2 patients (14.3%) and aseptic
pulmonary embolism in another patient (7.1%). Only 1 patient
had isolated local symptoms. Fever was isolated without
pulmonary or local symptoms in 6 patients. One patient died
from pulmonary embolism.
|
In the chronic group, the time elapsed between the last intervention and the first symptom was 25±28 months (range, 1 to >120 months), and the time elapsed between symptoms and the diagnosis was 8±12 months (range, 0 to 48 months).
Fever and chills were present in 32 patients (84.2%) with asthenia, and wasting was present in 10. Twenty-two (68.7%) of these patients described episodes of fever, with an episode of septic shock in 2 cases. Eleven patients had repeat episodes requiring several trials of antimicrobial therapy.
Arthralgia was present in five patients (13.2%). A history of spondylitis (8 to 12 months before the diagnosis) was found in two patients (5.3%); in one, the occurrence of repeated episodes of spondylitis led to the diagnosis.
Local symptoms occurred in 21 patients (55.3%): inflammation or discomfort in 3 patients, infection in 13, and exteriorization of the material in 5. Local infection or material exteriorization was isolated in 7 patients. Local symptoms were recurrent in 13 patients and were responsible for the great number of local interventions at the PM site before the diagnosis of PM-lead infection (3.86±2 local interventions versus 1.37±0.49 in the population without local infection or exteriorization; P<.0003 by t test). The time elapsed between the last intervention and the first symptom was 26.7±28.8 months when patients had local symptoms versus 23.5±29.0 months in patients without local symptoms (P=NS).
Pulmonary lesions were found in 17 patients (44.7%). A pleural effusion was present in 4 patients (pleural fluid cultures performed in 3 cases were negative). Pneumonia with pulmonary consolidation occurred in 6 patients and was recurrent in 2 cases; a germ was isolated in 4 of these 6 cases. One case of pneumonia evolved into a pulmonary abscess, and one was associated with septic shock. Interstitial pneumonia was found in 1 patient. Previous recurrent clinical pulmonary embolism, documented by pulmonary scintigraphy, was found in 3 patients. Five patients had a history of recurrent bronchitis with fever, cough, and expectoration. Eight of these 17 patients had undergone extensive investigations and had received antimicrobial therapy for these pulmonary symptoms; however, the diagnosis of infection related to the PM had not been specifically considered.
Blood Results (Fig 4
)
In the acute group, erythrocyte sedimentation rate and CRP were
increased in 13 patients (92.8%). CICs were found in 10 patients
(71.4%). Blood cell counts showed leukocytosis in 7 patients (50%),
in all cases due to an increase in polymorphonuclear cells.
|
In the chronic group, erythrocyte sedimentation rate and CRP were increased in 37 (97.4%) and 36 (94.7%) patients, respectively. CICs were observed in 75.9% of patients. Blood cell counts showed leukocytosis in 24 patients (63.2%), in all cases due to an increase in polymorphonuclear cells.
Echocardiography (TTE and TEE) (Fig 5A and
5B
)
In the acute group, TTE did not disclose abnormalities related to
the PM lead except in 1 patient in whom a large vegetation was visible
in the right ventricle. No further examination was performed in this
patient, who died suddenly with symptoms of pulmonary embolism.
TEE, performed in 12 patients, found abnormal echoes on the PM lead in
91.7% of patients. Vegetations were seen in 8 patients:
10 mm
in 1 patient (11 and 14 mm);
5 but <10 mm in 5 patients (4
in the atrium and 1 in the ventricle); and <5 mm in 2 patients (1
in the atrium associated with a vegetation on the tricuspid valve and 1
in the ventricle associated with a sleevelike prolongation onto the
lead). This sleevelike appearance was found in 4 patients (isolated in
2 and associated with a vegetation in 2).
|
In the chronic group, interpretable TTE was possible in 33 patients.
TTE disclosed vegetation on the PM lead in 10 (30.3%) of these
patients. TEE performed in 38 patients was normal in 2 and abnormal in
36 (94.7%); a vegetation was observed in 34 patients (
20 mm in
7 patients;
10 but <20 mm in 7 patients;
5 but <10 mm
in 17 patients; and <5 mm in 3 patients), and a sleevelike
appearance was visible in 10 patients (26.3%, associated with
vegetations on all but one occasion). A vegetation was detected on the
tricuspid valve in 6 patients (15.8%), and an additional perforation
was evidenced in 1 case. One patient, with a PM lead in the left
ventricle through a patent foramen ovale, had vegetations on the mitral
valve.
Ventilation Perfusion Pulmonary Scintigraphy
(Fig 6A
)
In the acute group, scintigraphy was not performed in
two patients (one with a bilateral pneumonia and one who died
suddenly). Pulmonary embolism occurred in three patients
(21.4%) before PM-lead ablation. Scintigraphy disclosed
embolism in two patients; the other patient died suddenly of
pulmonary embolism. One of the surviving patients had three
vegetations visible during TEE, one of them 6 mm long; the other
had two large vegetations (14 and 11 mm) associated with a
sleevelike appearance on the PM lead.
|
In the chronic group, pulmonary embolism was found before PM-lead ablation in 13 patients (34.2%). Five of these patients had no clinical history suggestive of pulmonary embolism, 3 described clinical pulmonary embolism, 4 had pneumonia (recurrent in 2 cases), and 1 had recurrent bronchitis. There was no significant statistical difference in vegetation size between patients with and without pulmonary embolism.
Bacteriology (Fig 7
)
In the acute group, a germ was isolated in 10 patients (71.4%) in
blood culture in 7 (50%) of 14 patients, local cultures in 3 (21.4%)
of 14 patients, and lead culture in 9 (69.2%) of 13 patients. The germ
was an S. aureus in 5 patients (50%) (3 MetiS, 2 MetiR), an
S. epidermidis in 3 patients (30%) (2 MetiS, 1 MetiR), an
S. species MetiR in 1 patient (10%), and a
Pseudomonas aeruginosa G in 1 patient (10%). Only 1 patient
had another potential source of infection (leg ulceration). All
patients received antibiotic therapy before lead ablation. The duration
of antibiotic therapy before lead ablation was 10.4±8.2 days in
patients with a positive lead culture versus 16.2±11.3 days in
patients with a negative lead culture (P=NS). No germ was
detected in 4 patients, but 2 of these had recurrent fever when
intravenous antibiotics were discontinued. TEE had
disclosed vegetations in the 5 patients with negative lead
cultures.
|
In the chronic group, a germ was detected in 36 patients (94.7%) in blood cultures in 24 (63.2%) of 38 patients, local cultures in 12 (33.3%) of 36, and lead cultures in 29 (80.6%) of 36. The germ was an S. aureus MetiS in 5 patients (13.9%), an S. epidermidis in 27 (75%; 20 MetiS, 3 MetiR, and 4 MetiR+MetiS), another S. in 4 (2 capiris, 1 specie, 1 cohnii) (11.1%; 3 MetiS, 1 MetiR), and a Gram-negative bacillus (1 Serratia, 1 Escherichia coli, and 1 Enterobacter cloacae) in 3 patients (8.3%). Two different germs were found in 7 patients, 4 with two different S. epidermidis MetiR and MetiS, 1 with S. aureus and S. epidermidis found in several blood cultures, and 2 with an S. associated with a Gram-negative bacillus. All but 1 patient received an antibiotic therapy before lead removal. Two antibiotics were used in 18 of 28 and 3 of 7 patients with a positive or a negative lead culture, respectively, and three antibiotics were used in 10 of 28 and 4 of 7 patients with a positive or a negative lead culture, respectively (P=NS by Fisher's exact test). The duration of antibiotic therapy before lead ablation was 9.7±6.1 days in patients with a positive lead culture versus 15.3±6.2 days in patients with a negative lead culture (P=.04).
The search for a possible source of infection, other than local contamination, was always negative, regardless of the germ found.
In our population, all but one of the cases of S. aureus in the acute group were associated with the presence of local symptoms. The time between the last intervention and the first symptom was 6.4±5.2 months when the responsible germ was an S. aureus versus 27.6±26.8 months when the germ was an S. epidermidis (P=.0010 by Welsh test).
Duke Criteria
Eight patients in the acute group and 27 in the chronic group met
the Duke pathological criteria (positive lead culture). The clinical
criteria (Table 2
) enabled us to classify these patients in the
definite endocarditis group before lead cultures were done for 2 (25%)
and 16 (59.3%) of the patients in the acute and chronic groups,
respectively. If we had considered local symptoms as a major clinical
criteria, 5 (62.5%) and 21 (77.8%) of the patients in the acute and
chronic groups, respectively, would have been classified as having
definite endocarditis by the clinical criteria alone. If we had
considered local symptoms as a minor clinical criteria, only 1 more
patient in the chronic group would have been classified as having
definite endocarditis by the clinical criteria alone. If we had
considered pulmonary symptoms as a major clinical criterion, 3
(37.5%) and 19 (70.4%) of the patients in the acute and chronic
groups, respectively, would have been classified as having definite
endocarditis by the clinical criteria alone. If we had considered
pulmonary symptoms as a minor clinical criterion, no more
patients would have been classified as having definite endocarditis. If
we had considered local and pulmonary symptoms as major
clinical criteria, 7 (87.5%) and 23 (85.2%) of the patients in the
acute and chronic groups, respectively, would have been classified as
having definite endocarditis by the clinical criteria alone.
Lead-Ablation Procedure
One patient with infection on PM leads located in the left
ventricle with vegetations on the mitral leaflets underwent surgical
ablation as primary therapy.
Two patients died before lead ablation. No ablation was attempted in an 81-year-old patient with a lead segment in the superior vena cava whose general condition precluded anesthesia and surgery.
Percutaneous removal was performed as primary therapy
in 38 patients with 78 leads (31 atrial and 47 ventricular
leads); the mean number of leads removed was 2.06 per patient, with a
maximum of 4 leads in a single patient. The vegetation size was
10 mm in all but 5 patients: the 3 initial patients in the
series with vegetations >10 mm and 2 elderly patients with
vegetation sizes of 14 and 16 mm who were in poor general
condition. Ablation failed in 2 patients (5.2% of patients) in whom
disruption occurred in the proximal part of the leads (1 atrial and 1
ventricular; 2.6% of leads). Disruption of the distal part
occurred in 9 ventricular leads (11.5% of all leads) with
recovery of all of the silicone; the wire-rope tip, however, was not
recovered. The location of these tips was extravascular in 2 patients,
at the right ventricular apex in 5, in the right atrium in
1, and in the pulmonary circulation (periprocedural
embolization) in 1. Two cases of tricuspid
regurgitation and 1 chordal disruption without
regurgitation appeared after ablation of 4 leads in 3
patients without previous clinical or echocardiographic evidence of
tricuspid endocarditis. Pericardial effusions were observed after 6
lead-extraction procedures in 3 patients, 1 of whom had right atrial
injury noted during surgical ablation. Pulmonary
scintigraphy was performed in 33 patients before and after
lead ablation, leading to the demonstration of vegetation migration
during ablation in 10 patients (30.3% of the patients; 10 of 66 leads,
or 15.1% of the leads). TEE demonstrated lead abnormalities in 7 of
these 10 patients (Fig 6B
).
Two (40%) of the 5 patients with vegetation size >10 mm had scintigraphic evidence of pulmonary embolism after lead removal, and pulmonary embolism was diagnosed by scintigraphy before lead removal in another patient. There was no significant statistical difference in vegetation size between patients with and without pulmonary migration. Only one patient (with a vegetation size of 40 mm) developed clinical evidence of septic pulmonary embolism with hemodynamic consequences. No deaths were observed after percutaneous ablation.
Surgical removal was performed as primary therapy in 10 patients with 21 leads (8 atrial and 13 ventricular) and after failure of percutaneous removal in 2 patients with 6 leads (3 atrial and 3 ventricular). The mean number of leads removed was 2.1 per patient, with a maximum of 4 leads in a single patient. The vegetation size was >10 mm except in 1 patient with a vegetation 8 mm in diameter who also had 2 retained lead segments in the superior vena cava. Two patients had retained lead segments in the superior vena cava; in 1 case, the lead was trapped within a thrombus that occluded the vena cava. Tricuspid valve resection was performed in 5 patients during the surgical procedure; TEE had shown abnormalities on the tricuspid valve in 3 of these 5 patients. Perioperative inspection confirmed the TEE in all but 2 cases, in which the TEE underestimated the size of the vegetations on the wire and did not detect small vegetations on the tricuspid valve. All the vegetations seen on the valves during surgery were at sites of contact with the leads. An intramyocardial infection was seen on a papillary muscle in 1 patient and at the site of insertion of the wire in another. Two patients developed pneumonia (1 patient with lung abscess after septic embolism) with collapse and died after surgery.
Comparison Between Acute and Chronic Groups
Comparisons of clinical, biological, and
echocardiographic data and results of
percutaneous lead removal in both groups are summarized
in Table 3
. Symptoms, blood results, and
echocardiographic data were not statistically different
between the two groups. The proportions of germs isolated and of
definite endocarditis according to the Duke criteria were higher in the
chronic than in the acute group. All the mechanical complications of
percutaneous lead removal occurred in the chronic
group.
|
Follow-up
The mean follow-up was 20.1±13 months (19.1±12.7 months in the
acute group, 20.5±13.2 months in the chronic group). No patients were
lost to follow-up.
Endocarditis recurred in a single patient after surgical removal. This patient, in whom two lead segments were left in the superior vena cava during the initial procedure, developed endocarditis on the tricuspid valve with the same germ (Serratia) 4 months after the surgical procedure. All the material was subsequently removed and a partial tricuspidectomy performed. Five months later, a contralateral PM was implanted, and 11 months later, septicemia recurred with the same germ. Extensive investigation to find a cause for the infection revealed a small intramyocardial segment of the previous PM lead. The material was removed surgically. The subsequent course was uneventful, with a follow-up of 14 months.
One patient in the acute group who underwent percutaneous removal without further subsequent permanent pacing presented 2 months later with endocarditis on the tricuspid valve that required tricuspidectomy.
Two patients died before lead ablation and 2 died after the surgical procedure. The overall predischarge mortality rate was 7.6%.
Ten patients died after hospitalization after a mean follow-up of 9.8±5.7 months, five in the acute group and five in the chronic group. Only three patients in the chronic group had symptoms of infection syndrome before death; interestingly, they were the patient without lead ablation and the patients with a wire rope not recovered in the right ventricle. The overall total mortality rate was 26.9%.
| Discussion |
|---|
|
|
|---|
Clinical Presentation
In the acute form, the short time elapsed between PM implantation
and the occurrence of infection facilitated the diagnosis. The vast
majority of patients had systemic symptoms as well as local signs of
infection. However, in one patient, isolated local symptoms occurred
without fever; biological markers of systemic inflammation led to
further investigations that revealed vegetations on the atrial lead on
TEE. In a second patient, pneumonia occurred several days after PM
implantation, in association with evidence of infection at the implant
site. However, delay in the diagnosis led to death from
pulmonary embolism. It therefore appears advisable to further
investigate or to closely observe patients with apparently isolated
local infection.
In the chronic form, the delay between the onset of symptoms and the diagnosis illustrates the difficulty in diagnosing PM-lead infection, as has been recognized previously.4 5 6 In our population, the delays in diagnosis were often related to the fact that PM-lead infection was not considered in the differential diagnosis. In other cases, possible clues to the diagnosis were ignored; for example, blood cultures positive for S. epidermidis were erroneously considered to be related to contamination of the specimens.12 13 Finally, the diagnosis was sometimes considered but wrongly excluded after the performance of inappropriate or insufficiently exhaustive tests; for example, negative results on TTE were sometimes considered by the treating physician as excluding the diagnosis of PM infection. The most common presentation was with fever or chills (84% of our population), often recurrent (68% of patients) and sometimes with associated symptoms suggestive of bronchitis. These episodes were generally not investigated further and were empirically treated by outpatient antibiotic therapy.
A history of local symptoms or signs related to the implant site was frequent (55.3% of our population) and often recurrent, and symptoms were erroneously considered to be related to local infection. The time between the last intervention and the first symptom was sometimes great (25±28 months; range, 1 to >120 months), and PM-lead infection was not considered in the differential diagnosis.
The diagnosis of systemic infection related to PM-lead infection must be systematically considered in the presence of chronic fever,6 13 14 recurrent bronchitis, or pulmonary infection15 or in case of recurrent or persistent evidence of infection at the implant site. In our population, there were no patients with clinical tricuspid regurgitation, but endocarditis of the right heart should be specifically excluded16 17 18 in cases with symptoms suggestive of pulmonary disease. Spondylitis was observed in only two (3.8% of all patients) of our patients and only in the chronic group; in one of these cases, recurrent spondylitis led to the diagnosis. Only a few cases of spondylitis as the presenting feature of PM-lead infection have been reported in the literature by Bonal et al19 and El Kohen et al.20
No patient in our series had symptoms of Osler's endocarditis. The criteria suggested by Von Reyn et al21 are not adaptable to the systemic infection related to PM-lead infection. We have used the Duke criteria to define systemic infection related to PM-lead infection. These criteria are adapted to this pathology, but some other clinical criteria could probably be added, such as local symptoms and pulmonary infections.
Echocardiography, TTE, and TEE
TTE showed vegetations in only 7% and 30% of patients in the
acute and chronic groups, respectively. This rate of detection is
markedly lower than previously reported rates18 22 23 24 that
suggested that TTE has an 80% sensitivity in the detection of
vegetations in the right heart. The explanation for this discrepancy
may relate to the fact that most of our patients had PM-leadrelated
endocarditis with intact tricuspid leaflets. TEE showed abnormalities
on the PM leads in 91.6% and 94.7% of patients in the acute and
chronic groups, respectively. The superiority of TEE over TTE has been
reported previously.25 The significance of the often-seen
sleevelike appearance of the PM lead is uncertain but rarely occurred
in isolation (3 of 14 patients). The major advantage of TEE related to
its ability to demonstrate tricuspid involvement, particularly in
patients whose poor echogenicity rendered TEE unreliable.
Bacteriology
Since 1979, six series have described patients with PM-related
sepsis; S. aureus and coagulase-negative staphylococcus
accounted for 80% of bacteremias. 6 26 27 This rate is in
accordance with the present study. The 30% and 19.4% rates of
negative lead cultures in the acute and chronic groups, respectively,
are surprising. However, it may be that some vegetations (particularly
small atrial vegetations) did not have an infectious
origin.28 In a rabbit model, it has been shown that
sterile vegetations may form on an intracardiac
catheter.29 Preoperative antibiotic therapy could account
for these negative lead cultures equally well. Finally, the amorphous
material (slime) that is universally found in staphylococcal
colonies30 31 32 could also be responsible for these negative
cultures by isolating the lead and the germ from the culture material.
In all these patients with negative cultures, the clinical evidence of
PM-lead infection was compelling. The majority of these patients had a
clinical history of infection, sometimes resistant to several
trials of antibiotics or with recurrent fever when antibiotics were
stopped. Furthermore, in all cases, ablation of the PM leads led to
resolution of the symptoms.
It is commonly accepted that the most common portal is the subcutaneous site of insertion of the pacing system. Extension along the lead into the vascular system is the usual explanation for the localization of the infection to the lead. Bacterial colonization of the lead during the course of bacteremia whose origin is not related to the pacing system has been less well documented. This mode of contamination represented 14% in the series reported by Kugener et al33 and Bryan et al.34 The presence of an important fibrous coating that isolates the lead from venous blood explains the rarity of hematogenous contamination. In our population, local contamination at the PM site was the most common portal.
Management
In our experience, as in the study by Camus et al,27
the high rate of uncontrolled infection or relapse among patients with
septicemia in relation to PM-material infection confirms the need for
immediate removal of the entire pacing system. This is obvious for
systemic infection related to PM-lead contamination but is also
important in obvious infection of the PM pocket or the subcutaneous
part of the lead. Because of the adherence of staphylococci to the
surfaces of intravenous leads and the poor bactericidal
activity of antibiotics on adherent bacteria, failure of antimicrobial
therapy is not unexpected; thus, immediate removal of the entire pacing
system should probably be performed in all cases. The incidence of
staphylococcal infection confirms the need for antistaphylococcal
therapy in all cases. Our protocol for antimicrobial therapy seems to
be reliable after removal of material with only one relapse. We think,
however, that the PM material should be removed immediately rather than
attempting prolonged antibiotic therapy alone.
The choice of the technique of ablation (surgical versus
percutaneous) was based on the size of the vegetations
assessed on TEE, the presence of morphological changes of the tricuspid
valve, and the general condition of the patient. This retrospective
study did not enable us to compare both methods. Septic
pulmonary embolism was a major concern after
percutaneous lead removal. At the beginning of our
experience, on the basis of the initial cases, we decided to remove the
material percutaneously only if TEE did not disclose
vegetations >10 mm. This policy was based on the observations by
Mugge et al,35 who found that embolism was more frequent
when vegetation size was >10 mm in endocarditis related to valve
infection, and by Robbins et al,36 who found more
complications when vegetation size was >10 mm in a series of 21
patients with tricuspid endocarditis. Percutaneous
ablation of infected leads was found to be a reliable technique in
patients with vegetations
10 mm. The two major complications
were vegetation migration without vital involvement and tricuspid
injury. Retrospectively, we did not find any significant relation
between pulmonary migration and vegetation size, but the number
of leads removed in patients with large vegetations was small. It must
be emphasized that pulmonary embolism with vital involvement
occurred in the patient with the largest vegetation size (40 mm).
Additionally, 2 patients (with vegetation sizes of 14 and 20 mm,
respectively) died after surgical removal. Conversely, no deaths
occurred after percutaneous ablation. Perhaps we should
expand the indications for percutaneous removal to
include patients with larger vegetation sizes. We did not perform a
surgical procedure when a wire rope was not completely recovered after
percutaneous ablation, but 2 of 9 such patients (2 of 7
patients with a wire rope in the circulation) died from infection
syndrome; these patients probably needed more aggressive
management.
Conclusions
The diagnosis of endocarditis related to PM-lead infection should
be systematically considered in patients with fever, a history of local
complications, or pulmonary pathology after PM insertion. We
describe two different clinical presentations: the acute
form, with an early presentation with sepsis, often in
conjunction with local signs of infection, and a chronic form beginning
several months later. The presentation may be atypical and
the presenting symptoms may occur late after the last intervention
at the PM site. CRP, CIC, and pulmonary
scintigraphy are of diagnostic value. TEE must
be performed in search of vegetations. Staphylococci are responsible
for the vast majority of these infections, especially S.
epidermidis in the chronic group and S. aureus in the
acute group. Immediate removal of the entire pacing system is
necessary, in addition to prolonged antimicrobial therapy including
antistaphylococcal agents. The technique of removal (surgical versus
percutaneous) depends on the size of the vegetations;
percutaneous removal seems to be safer when vegetation
size is
10 mm.
| Selected Abbreviations and Acronyms |
|---|
|
| Acknowledgments |
|---|
Received July 31, 1996; revision received November 4, 1996; accepted November 20, 1996.
| References |
|---|
|
|
|---|
2. Bluhm G. Pacemaker infection: a clinical study with special reference to prophylactic use of some isoxazolyl penicillins. Acta Med Scand Suppl. 1985;699:1-62. [Medline] [Order article via Infotrieve]
3. Block K, Russi E. Right heart endocarditis. Schweiz Med Wochenschr. 1989;47:1664-1672.
4. Glock Y, Sabatier J, Salvador-Mazencq M, Puel P. Les endocardites sur électrodes endocavitaires de stimulateurs cardiaques: a propos de 7 cas. Arch Mal Coeur. 1986;79:483-488.
5. Loffler S, Kasper J, Postulka J, Cornak V, Bohunicky F, Zelenay J, Treger J. Septic complications in patients with permanent pacemakers. Cor Vasa. 1988;30:400-404. [Medline] [Order article via Infotrieve]
6. Morgan G, Ginks W, Siddons QH, Letham A. Septicemia in patients with endocardial pacemaker. Am J Cardiol. 1979;44:221-224. [Medline] [Order article via Infotrieve]
7. Rubio-Alvarez J, Duran-Munoz D, Sierra-Quiroga J, Garcia-Bengochea JB. Right heart endocarditis and endocardial pacemakers. Ann Thorac Surg. 1989;48:147-151.
8. Choo MH, Holmes DR Jr, Gersh BJ, Maloney JD, Merideth J, Pluith JR. Permanent pacemaker infections: characterization and management. Am J Cardiol. 1981;48:559-564. [Medline] [Order article via Infotrieve]
9. Lewis AB, Hayes DL, Holmes DR Jr, Vlietstra RE, Pluth JR, Osborn MJ. Update on infections involving permanent pacemakers: characterization and management. J Thorac Cardiovasc Surg. 1985;89:758-763. [Abstract]
10. Hajrula A, Jarvinen A, Virtanen KS, Mattila S. Pacemaker infections: treatment with total or partial pacemaker system removal. Thorac Cardiovasc Surg. 1985;33:218-220. [Medline] [Order article via Infotrieve]
11. Bayer A, Ward J, Ginzton L, Shapiro S. Evaluation of new clinical criteria for the diagnosis of infective endocarditis. Am J Med. 1994;96:211-219. [Medline] [Order article via Infotrieve]
12. Hartstein AI, Jackson J, Gilbert DN. Prophylactic antibiotics and the insertion of permanent transvenous cardiac pacemakers. J Thorac Cardiovasc Surg. 1978;75:216-226.
13. Jacomet C, Cacoub P, Zuittere C, Gandjbakhch I, Godeau P. Septicemie associee à la présence d'un cardiostimulateur avec sonde endocavitaire. Rev Med Interne. 1992;13:296-298. [Medline] [Order article via Infotrieve]
14. Southall E, Pallister D. Meningitis and recurrent septicemia secondary to unsuspected pacemaker infection. Br Med J. 1982;284:1530-1537.
15.
Schwartz IS, Pervez N. Bacterial endocarditis
associated with a permanent transvenous cardiac pacemaker.
JAMA. 1971;218:736-737.
16. Arbulu A, Asfaw I. Tricuspid valvulectomy without prosthetic replacement: ten years of clinical experience. J Thorac Cardiovasc Surg. 1981;82:684-691. [Medline] [Order article via Infotrieve]
17. Chan P, Ogilby D, Segal B. Tricuspid valve endocarditis. Am Heart J. 1989;117:1140-1146. [Medline] [Order article via Infotrieve]
18. Panidis IP, Kotler MN, Mintz GS, Segal BL, Ross JJ. Right heart endocarditis: clinical and echocardiographic features. Am Heart J. 1984;107:759-764. [Medline] [Order article via Infotrieve]
19. Bonal J, Bouchiat C, De Jaureguiberry JP, Duval JL, Carli P. Spondylodiscites revelatrices d'endocardites sur électrodes endocavitaires de stimulateurs cardiaques. Arch Mal Coeur. 1994;87:1735-1738.
20. El Kohen M, Millaire A, Leroy O, Beuscart C, Vanjak D, Sivery B, Senneville E, Ducloux G. Endocardites infectieuses tricuspidiennes sur sonde de stimulation permanente endocavitaire. Arch Mal Coeur. 1990;83:1855-1857.
21. Von Reyn CF, Levy BS, Arbeit RD, Friedland G, Crumpacker CS. Infective endocarditis: an analysis based on strict case definitions. Ann Intern Med. 1981;94:505-517.
22. Ginzton LE, Siegel RJ, Criley JM. Natural history of tricuspid valve endocarditis: a two dimensional echocardiographic study. Am J Cardiol. 1982;49:1853-1859. [Medline] [Order article via Infotrieve]
23. Melvin ET, Berger M, Lutzker LG, Goldberg E, Mildvan D. Noninvasive methods for detection of valve vegetations in infective endocarditis. Am J Cardiol. 1981;66:523-525.
24. Daelemans R, Kersschot I, Van Den Branden F, Nagler J, D'Heer H, Parizel G, Brutsaert DL. Pacemaker endocarditis: contribution of two-dimensional echocardiography. Acta Cardiol. 1984;34:293-299.
25. Zehender M, Buchner C, Geibel A. Diagnosis of hidden pacemaker sepsis by transesophageal echocardiography and a new technique for lead extraction. Am Heart J. 1989;118:1050-1053. [Medline] [Order article via Infotrieve]
26. Mansour K, Kauten J, Hatcher C. Management of infected pacemaker: explantation, sterilization, and reimplantation. Ann Thorac Surg. 1985;40:617-619. [Abstract]
27. Camus C, Leport C, Raffi F, Michelet C, Cartier F, Vilde JL. Sustained bacteremia in 26 patients with a permanent endocardial pacemaker: assessment of wire removal. Clin Infect Dis. 1993;17:46-55. [Medline] [Order article via Infotrieve]
28. Lauck G. Adherent structures on implanted leads of cardiac pacemaker. Circulation. 1994;4(suppl I):I-70. Abstract.
29. Tanphaichita D, Ries K, Levison ME. Susceptibility to Streptococcus viridans endocarditis in rabbits with intracardiac pacemaker electrode or polyethylene tubing. J Lab Clin Med. 1974;84:726-731. [Medline] [Order article via Infotrieve]
30. Peters G, Locci R, Pulverer G. Adherence and growth of coagulase negative staphylococci on surfaces of intravenous catheters. J Infect Dis. 1982;146:4789-4793.
31. Peters G, Pulverer G. Microbiology of staphylococcal pacemaker lead infection. PACE Pacing Clin Electrophysiol. 1986;4:205-215.
32. Peters G, Saborowski F, Locci R. Investigations on staphylococcal infection on transvenous endocardial pacemaker electrodes. Am Heart J. 1984;108:358-365.
33. Kugener H, Tribouilloy C, Hermida JS, Jarry G, Avinee P, Maingourd Y. Endocardites infectieuses sur sondes de stimulation endocavitaire permanentes: interet de l'echocardiographie et revue de la litterature. Ann Cardiol Angeiol (Paris). 1993;42:331-338. [Medline] [Order article via Infotrieve]
34. Bryan C, Sutton J, Saunders D, Longaker D, Smith C. Endocarditis related to transvenous pacemakers: syndromes and surgical implications. J Thorac Cardiovasc Surg. 1978;75:758-762. [Abstract]
35. Mugge A, Daniel WG, Frank G, Lichtlen PR. Echocardiography in infective endocarditis: reassessment of prognostic implications of vegetation size determined by the transthoracic and the transesophageal approach. J Am Coll Cardiol. 1989;14:631-638. [Abstract]
36. Robbins MJ, Fratter R, Soeiro R, Frishman WH, Strom JA. Influence of vegetation size on clinical outcome of right sided infective endocarditis. Am J Med. 1986;80:165-171.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
C. Mihl, Z. Geyik, E.C. Cheriex, and J.M. van Opstal A challenging lead endocarditis Eur J Echocardiogr, November 3, 2009; (2009) jep166v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. M. Calvagna, R. Evola, G. Scardace, and S. Valsecchi Single-operator experience with a mechanical approach for removal of pacing and implantable defibrillator leads Europace, November 1, 2009; 11(11): 1505 - 1509. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Romeyer-Bouchard, A. Da Costa, V. Dauphinot, M. Messier, L. Bisch, B. Samuel, P. Lafond, P. Ricci, and K. Isaaz Prevalence and risk factors related to infections of cardiac resynchronization therapy devices Eur. Heart J., October 28, 2009; (2009) ehp421v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Kleemann, T. Becker, M. Strauss, N. Dyck, U. Weisse, W. Saggau, U. Burkhardt, and K. Seidl Prevalence of bacterial colonization of generator pockets in implantable cardioverter defibrillator patients without signs of infection undergoing generator replacement or lead revision Europace, October 27, 2009; (2009) eup334v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
Endorsed by the European Society of Clinical Micro, Authors/Task Force Members, G. Habib, B. Hoen, P. Tornos, F. Thuny, B. Prendergast, I. Vilacosta, P. Moreillon, M. de Jesus Antunes, et al. Guidelines on the prevention, diagnosis, and treatment of infective endocarditis (new version 2009): The Task Force on the Prevention, Diagnosis, and Treatment of Infective Endocarditis of the European Society of Cardiology (ESC) Eur. Heart J., October 1, 2009; 30(19): 2369 - 2413. [Full Text] [PDF] |
||||
![]() |
F. Franceschi, F. Thuny, R. Giorgi, I. Sanaa, E. Peyrouse, X. Assouan, S. Prevot, E. Bastard, G. Habib, and J.-C. Deharo Incidence, risk factors, and outcome of traumatic tricuspid regurgitation after percutaneous ventricular lead removal. J. Am. Coll. Cardiol., June 9, 2009; 53(23): 2168 - 2174. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. Gaca, B. Lima, C. A. Milano, S. S. Lin, R. D. Davis, J. E. Lowe, and P. K. Smith Laser-assisted extraction of pacemaker and defibrillator leads: the role of the cardiac surgeon. Ann. Thorac. Surg., May 1, 2009; 87(5): 1446 - 1450. [Abstract] [Full Text] [PDF] |
||||
![]() |
J C Lekkerkerker, C van Nieuwkoop, S A Trines, J G van der Bom, A Bernards, E T van de Velde, M Bootsma, K Zeppenfeld, J W Jukema, J-W Borleffs, et al. Risk factors and time delay associated with cardiac device infections: Leiden device registry Heart, May 1, 2009; 95(9): 715 - 720. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. S. Baman, S. K. Gupta, J. A. Valle, and E. Yamada Risk Factors for Mortality in Patients With Cardiac Device-Related Infection Circ Arrhythm Electrophysiol, April 1, 2009; 2(2): 129 - 134. [Abstract] [Full Text] [PDF] |
||||
![]() |
Efficacy of antibiotic prophylaxis before the implantation of pacemakers and cardioverter-defibrillators: results of a large, prospective, randomized, double-blinded, placebo-controlled trial. Circ Arrhythm Electrophysiol, February 1, 2009; 2(1): 29 - 34. |
||||
![]() |
J. Bis, K. Goscinska-Bis, M. Krejca, E. Zinka, J. Skarysz, L. Machej, and A. Bochenek Surgical extraction of cardiac resynchronization therapy system with concomitant implantation of a new system with the use of epicardial leads in a patient with endocarditis J. Thorac. Cardiovasc. Surg., September 1, 2008; 136(3): 786 - 787. [Full Text] [PDF] |
||||
![]() |
M. Pichlmaier, V. Marwitz, C. Kuhn, M. Niehaus, G. Klein, C. Bara, A. Haverich, and W.-R. Abraham High prevalence of asymptomatic bacterial colonization of rhythm management devices Europace, September 1, 2008; 10(9): 1067 - 1072. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. M. Khumri, J. B. Thibodeau, and M. L. Main Transesophageal echocardiographic diagnosis of left atrial appendage occluder device infection Eur J Echocardiogr, July 1, 2008; 9(4): 565 - 566. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Camboni, C. G. Wollmann, A. Loher, R. Gradaus, H. H. Scheld, and C. Schmid Explantation of Implantable Defibrillator Leads Using Open Heart Surgery or Percutaneous Techniques Ann. Thorac. Surg., January 1, 2008; 85(1): 50 - 55. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. R. Sohail, D. Z. Uslan, A. H. Khan, P. A. Friedman, D. L. Hayes, W. R. Wilson, J. M. Steckelberg, S. M. Jenkins, and L. M. Baddour Infective Endocarditis Complicating Permanent Pacemaker and Implantable Cardioverter-Defibrillator Infection Mayo Clin. Proc., January 1, 2008; 83(1): 46 - 53. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Klug, M. Balde, D. Pavin, F. Hidden-Lucet, J. Clementy, N. Sadoul, J. L. Rey, G. Lande, A. Lazarus, J. Victor, et al. Risk Factors Related to Infections of Implanted Pacemakers and Cardioverter-Defibrillators: Results of a Large Prospective Study Circulation, September 18, 2007; 116(12): 1349 - 1355. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. R. Sohail, D. Z. Uslan, A. H. Khan, P. A. Friedman, D. L. Hayes, W. R. Wilson, J. M. Steckelberg, S. Stoner, and L. M. Baddour Management and Outcome of Permanent Pacemaker and Implantable Cardioverter-Defibrillator Infections J. Am. Coll. Cardiol., May 8, 2007; 49(18): 1851 - 1859. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Z. Uslan, M. R. Sohail, J. L. St. Sauver, P. A. Friedman, D. L. Hayes, S. M. Stoner, W. R. Wilson, J. M. Steckelberg, and L. M. Baddour Permanent Pacemaker and Implantable Cardioverter Defibrillator Infection: A Population-Based Study Arch Intern Med, April 9, 2007; 167(7): 669 - 675. [Abstract] [Full Text] [PDF] |
||||
![]() |
K Matschke, S. Tugtekin, K Alexiou, M Knaut, J. Park, and M Schulze Malplacement of a Pacemaker Lead -- A Rare Cause for Aortic Valve Endocarditis Asian Cardiovasc Thorac Ann, February 1, 2007; 15(1): 64 - 65. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Marquie, G. De Geeter, D. Klug, C. Kouakam, F. Brigadeau, O. Jabourek, N. Trillot, D. Lacroix, and S. Kacet Post-operative use of heparin increases morbidity of pacemaker implantation. Europace, April 1, 2006; 8(4): 283 - 287. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Habib Management of infective endocarditis Heart, January 1, 2006; 92(1): 124 - 130. [Full Text] [PDF] |
||||
![]() |
R. J. Cook, R. W. Ashton, G. L. Aughenbaugh, and J. H. Ryu Septic Pulmonary Embolism: Presenting Features and Clinical Course of 14 Patients Chest, July 1, 2005; 128(1): 162 - 166. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. J. Rastan, N. Doll, T. Walther, and F. W. Mohr Pacemaker dependent patients with device infection--a modified approach Eur. J. Cardiothorac. Surg., June 1, 2005; 27(6): 1116 - 1118. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-P. Chang, M.-C. Chen, G. B.-F. Guo, and C.-L. Kao Less-Invasive Surgical Extraction of Problematic or Infected Permanent Transvenous Pacemaker System Ann. Thorac. Surg., April 1, 2005; 79(4): 1250 - 1254. [Abstract] [Full Text] [PDF] |
||||
![]() |
D Klug, F Wallet, D Lacroix, C Marquie, C Kouakam, S Kacet, and R Courcol Local symptoms at the site of pacemaker implantation indicate latent systemic infection Heart, August 1, 2004; 90(8): 882 - 886. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. O. Darouiche Treatment of Infections Associated with Surgical Implants N. Engl. J. Med., April 1, 2004; 350(14): 1422 - 1429. [Full Text] [PDF] |
||||
![]() |
R. J. Cook, T. A. Orszulak, V. T. Nkomo, J. A. Shuford, W. D. Edwards, and J. H. Ryu Aspergillus Infection of Implantable Cardioverter-Defibrillator Mayo Clin. Proc., April 1, 2004; 79(4): 549 - 552. [Abstract] [PDF] |
||||
![]() |
M.-R. Movahed, B. Kasravi, and C. S. Bryan Prophylactic Use of Vancomycin in Adult Cardiology and Cardiac Surgery Journal of Cardiovascular Pharmacology and Therapeutics, March 1, 2004; 9(1): 13 - 20. [Abstract] [PDF] |
||||
![]() |
Task Force Members, D. Horstkotte, F. Follath, E. Gutschik, M. Lengyel, A. Oto, A. Pavie, J. Soler-Soler, G. Thiene, A. von Graevenitz, et al. Guidelines on Prevention, Diagnosis and Treatment of Infective Endocarditis Executive Summary: The Task Force on Infective Endocarditis of the European Society of Cardiology Eur. Heart J., February 1, 2004; 25(3): 267 - 276. [Full Text] [PDF] |
||||
![]() |
F.A.L.E. Bracke, A Meijer, and L.M. van Gelder Lead extraction for device related infections: a single-centre experience Europace, January 1, 2004; 6(3): 243 - 247. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. M. Baddour, M. A. Bettmann, A. F. Bolger, A. E. Epstein, P. Ferrieri, M. A. Gerber, M. H. Gewitz, A. K. Jacobs, M. E. Levison, J. W. Newburger, et al. Nonvalvular Cardiovascular Device-Related Infections Circulation, October 21, 2003; 108(16): 2015 - 2031. [Full Text] [PDF] |
||||
![]() |
E. Dumont, C. Camus, F. Victor, C. de Place, D. Pavin, C. Alonso, P. Mabo, and J.C. Daubert Suspected pacemaker or defibrillator transvenous lead infection: Prospective assessment of a TEE-guided therapeutic strategy Eur. Heart J., October 1, 2003; 24(19): 1779 - 1787. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. del Rio, I. Anguera, J. M. Miro, L. Mont, V. G. Fowler Jr, M. Azqueta, and C. A. Mestres Surgical Treatment of Pacemaker and Defibrillator Lead Endocarditis: The Impact of Electrode Lead Extraction on Outcome Chest, October 1, 2003; 124(4): 1451 - 1459. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Di Salvo, F. Thuny, V. Rosenberg, V. Pergola, O. Belliard, G. Derumeaux, A. Cohen, D. Iarussi, R. Giorgi, J.-P. Casalta, et al. Endocarditis in the elderly: clinical, echocardiographic, and prognostic features Eur. Heart J., September 1, 2003; 24(17): 1576 - 1583. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Klug, F. Wallet, S. Kacet, and R. J. Courcol Involvement of Adherence and Adhesion Staphylococcus epidermidis Genes in Pacemaker Lead-Associated Infections J. Clin. Microbiol., July 1, 2003; 41(7): 3348 - 3350. [Abstract] [Full Text] [PDF] |
||||
![]() |
A.-Q. T. Nguyen, N. W. W. Choong, and P. C. Spittell 28-Year-Old Man With Recurrent Fever Mayo Clin. Proc., July 1, 2003; 78(7): 897 - 900. [PDF] |
||||
![]() |
M. R. Moon, C. J. Camillo, and M. J. Gleva Laser-assist during extraction of chronically implanted pacemaker and defibrillator leads Ann. Thorac. Surg., June 1, 2002; 73(6): 1893 - 1896. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A. Memon, J. Buckley, and A. O'Donnell Percutaneous Extraction of Entrapped Infective Transvenous Pacing Lead Asian Cardiovasc Thorac Ann, March 1, 2002; 10(1): 69 - 71. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Miralles, V. Moncada, H. Chevez, R. Rodriguez, J. Granados, and E. Castells Pacemaker endocarditis: approach for lead extraction in endocarditis with large vegetations Ann. Thorac. Surg., December 1, 2001; 72(6): 2130 - 2132. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. L. Chamis, G. E. Peterson, C. H. Cabell, G. R. Corey, R. A. Sorrentino, R. A. Greenfield, T. Ryan, L. B. Reller, and V. G. Fowler Jr Staphylococcus aureus Bacteremia in Patients With Permanent Pacemakers or Implantable Cardioverter-Defibrillators Circulation, August 28, 2001; 104(9): 1029 - 1033. [Abstract] [Full Text] [PDF] |
||||
![]() |
F A Bracke, A Meijer, and L M van Gelder Pacemaker lead complications: when is extraction appropriate and what can we learn from published data? Heart, March 1, 2001; 85(3): 254 - 259. [Full Text] |
||||
![]() |
E. Francia, P. Domingo, M. A. Sambeat, J. A. Montiel, R. Pericas, F. Sanchez, and M. Gurgui Pacemaker Infection by Brucella melitensis: A Rare Cause of Relapsing Brucellosis Arch Intern Med, November 27, 2000; 160(21): 3327 - 3328. [Full Text] [PDF] |
||||
![]() |
J. D. Chua, B. L. Wilkoff, I. Lee, N. Juratli, D. L. Longworth, and S. M. Gordon Diagnosis and Management of Infections Involving Implantable Electrophysiologic Cardiac Devices Ann Intern Med, October 17, 2000; 133(8): 604 - 608. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Robin, F. Tronc, C. Vedrinne, and G. Champsaur Video-assisted tricuspid valve surgery: a new surgical option in endocarditis on pacemaker Eur. J. Cardiothorac. Surg., August 1, 1999; 16(2): 243 - 245. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Habib, G. Derumeaux, J.-F. Avierinos, J.-P. Casalta, F. Jamal, F. Volot, M. Garcia, J. Lefevre, F. Biou, A. Maximovitch-Rodaminoff, et al. Value and limitations of the duke criteria for the diagnosis of infective endocarditis J. Am. Coll. Cardiol., June 1, 1999; 33(7): 2023 - 2029. [Abstract] [Full Text] [PDF] |
||||
![]() |
F Victor, C De Place, C Camus, H Le Breton, C Leclercq, D Pavin, P Mabo, and C Daubert Pacemaker lead infection: echocardiographic features, management, and outcome Heart, January 1, 1999; 81(1): 82 - 87. [Abstract] [Full Text] |
||||
![]() |
J G Voet, Y R Vandekerckhove, L L Muyldermans, L H Missault, and L J Matthys Pacemaker lead infection: report of three cases and review of the literature Heart, January 1, 1999; 81(1): 88 - 91. [Abstract] [Full Text] |
||||
![]() |
A. Da Costa, H. Lelievre, H. PhaD, G. Kirkorian, M. Celard, P. Chevalier, F. Vandenesch, J. Etienne, and P. Touboul Role of the Preaxillary Flora in Pacemaker Infections : A Prospective Study Circulation, May 19, 1998; 97(18): 1791 - 1795. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Da Costa, G. Kirkorian, M. Cucherat, F. Delahaye, P. Chevalier, A. Cerisier, K. Isaaz, and P. Touboul Antibiotic Prophylaxis for Permanent Pacemaker Implantation : A Meta-Analysis Circulation, May 19, 1998; 97(18): 1796 - 1801. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1997 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |