(Circulation. 1997;95:1611-1622.)
© 1997 American Heart Association, Inc.
Articles |
From the Cardiovascular System Laboratory, The Bruce Rappaport Faculty of Medicine, Technion-Israel Institute of Technology, Haifa.
Correspondence to Shlomo A. Ben-Haim, MD, DSc, Cardiovascular System Laboratory, The Bruce Rappaport Faculty of Medicine, Technion-Israel Institute of Technology, Efron St, POB 9649, 31096 Haifa, Israel. E-mail sol{at}biomed.technion.ac.il.
| Abstract |
|---|
|
|
|---|
Methods and Results The method is based on using a new locatable catheter connected to an endocardial mapping and navigating system. The system uses magnetic technology to accurately determine the location and orientation of the catheter and simultaneously records the local electrogram from its tip. By sampling a plurality of endocardial sites, the system reconstructs the three-dimensional geometry of the chamber, with the electrophysiological information color-coded and superimposed on the anatomy. The accuracy of the system was tested in both in vitro and in vivo studies and was found to be highly reproducible (SD, 0.16±0.02 [mean±SEM] and 0.74±0.13 mm) and accurate (mean errors, 0.42±0.05 and 0.73±0.03 mm). In further studies, electroanatomical mapping of the cardiac chambers was performed in 34 pigs. Both the geometry and activation sequence were repeatable in all pigs.
Conclusions The new mapping method is highly accurate and reproducible. The ability to combine electrophysiological and spatial information provides a unique tool for both research and clinical electrophysiology. Consequently, the main shortcomings of conventional mappingnamely, prolonged x-ray exposure, low spatial resolution, and the inability to accurately navigate to a predefined sitecan all be overcome with this new method.
Key Words: mapping catheter ablation arrhythmia electrophysiology
| Introduction |
|---|
|
|
|---|
Cardiac mapping is a broad term that covers several modes of mapping such as body-surface,8 endocardial,2 9 and epicardial10 mapping. Nevertheless, common to all cardiac mapping procedures is the registration of the electrical activation sequence by recording of extracellular electrograms. Electrogram acquisition can be simultaneous from many sites, usually during surgery, with the use of fixed-shape electrode arrays11 12 13 14 such as epicardial socks and endocardial balloons.9 15 More commonly, cardiac mapping is performed with catheters that are introduced percutaneously into the heart chambers and sequentially record the endocardial electrograms. These electrophysiological catheters are navigated and localized with the use of multiplanar fluoroscopy. A major pitfall in the currently used methods is the inability to accurately associate the intracardiac electrogram with a specific endocardial site. Thus, the localization of the recording sites with fluoroscopy is inaccurate, cumbersome, and associated with high x-ray exposure for both the patient and the physician.
We present here a new method for nonfluoroscopic catheter-based endocardial mapping that enables the generation of 3D electroanatomical maps of the heart chambers. This method is based on using a new locatable catheter connected to an endocardial mapping and navigation system. The system comprises a miniature passive magnetic field sensor located at the tip of the catheter, an external ultralow magnetic field emitter, and a processing unit. The system uses the magnetic technology to accurately determine the location and orientation of the catheter in six degrees of freedom (x, y, z, roll, pitch, and yaw) and simultaneously records the intracardiac local electrogram from its tip. The 3D geometry of the chamber is reconstructed in real time with the electrophysiological information, which is color-coded, superimposed on the electroanatomical map. The ability to localize the catheter tip in real time with reference to the electroanatomical map has special value for guiding catheter ablation procedures in which radiofrequency energy is delivered from the tip. For this application, it is essential that the navigation tools allow accurate and reproducible repositioning of the catheter tip on top of a previously defined target.
The present study describes the mapping technology and provides both in vitro and in vivo accuracy results of the navigation system. We also present typical electroanatomical maps of cardiac chambers of the swine during different rhythms. The possible contribution of this methodology to both clinical and research electrophysiology is discussed.
| Methods |
|---|
|
|
|---|
The locatable catheter is similar to a regular electrophysiological 8F
deflectable-tip catheter (Cordis-Webster; Fig 1A
). The
catheter tip is mounted on the distal end of the shaft and includes the
tip electrode and several additional proximal electrodes that enable
recording of unipolar or bipolar signals. Just proximal to the tip
electrode lies the location sensor, totally embedded within the
catheter. Signals received within the sensor are transmitted along the
catheter shaft to the main processing unit. The catheter is equipped
with radiofrequency delivery capabilities, including a 4-mm tip and a
thermocouple, and can be used with an ordinary radiofrequency
generator.
|
The locator pad is located beneath the operating table and generates
ultralow magnetic fields (5x10-6 to
5x10-5 T) that code the mapping space around
the patient's chest with both temporal and spatial distinguishing
characteristics. These fields contain the information necessary to
resolve the location and orientation of the sensor in six degrees of
freedom (x, y, z, roll, pitch, and
yaw). The locator pad includes three coils (Fig 1B
). Each coil
generates a magnetic field that decays as a function of the distance
from that coil. The sensor measures the strength of the magnetic field,
thus enabling determination of the distance from each of its sources.
These distances determine the radii of theoretical spheres around each
coil. The intersection of these three spheres determines the location
of the sensor in space.
Mapping Procedure
Two locatable 8F catheters were introduced, by use of
fluoroscopic guidance, through the appropriate vascular access and
positioned inside the heart chambers. The first catheter was introduced
and placed in the CS or the RVA and served as a reference catheter. The
mapping catheter was introduced and placed in the mapped chamber. The
mapping system determined the locations and orientations of both the
mapping and reference catheters. The location of the mapping catheter
was gated to a fiducial point in the cardiac cycle and recorded
relative to the location of the fixed reference catheter at that time,
thus compensating for both subject and cardiac motion. For example,
choosing the R wave as the fiducial point resulted in the determination
of location at end diastole. In this setting, the end-diastolic
location and orientation of the catheter were continuously shown on the
screen of the mapping computer. By moving the catheter inside the
heart, the system continuously analyzed its location and presented it
to the user, thus enabling navigation without the use of
fluoroscopy.
The mapping catheter was dragged over the endocardium, sequentially
acquiring its tip location and its electrogram while the catheter was
in stable contact with the endocardium. The LAT at
each site was determined from the intracardiac
electrogram. We report results that were collected from unipolar
recordings filtered at 0.5 to 400 Hz. The LAT at each site was
calculated as the interval between a fiducial point on the body-surface
ECG or a fixed intracardiac electrogram and the steepest negative
intrinsic deflection from the mapping-catheter unipolar recording. The
stability of the catheter-wall contact was evaluated at every site by
examination of the end-diastolic stability (the distance in millimeters
between two consecutive end-diastolic locations) and the LAT stability
(measured as the difference in milliseconds between two consecutive
LATs). A point was added to the map only if the end-diastolic location
stability was <2 mm and the LAT stability was <2 ms. The
electrophysiological information was color-coded (red being the
shortest LAT and purple being the longest) and superimposed on the 3D
chamber geometry. The reconstruction was updated in real time with the
acquisition of each new site. The map was presented as a 3D object in
which interpolation of activation between adjacent sites was allowed
only for nearby sites. Whenever the distance between adjacent sites was
higher than a predetermined threshold, no interpolation was made, and
the sites were connected by wire frames. Fig 2
illustrates an example of
reconstruction of the swine left ventricle.
|
Validation of Intracardiac Signal Recording
The intracardiac signal from the tip of the locatable catheter
was correlated with a standard nonlocatable electrophysiological
catheter (CardioRhythm). The two catheters were positioned as close as
possible with the use of multiple x-ray projections. Both signals were
simultaneously recorded, filtered, and gained identically and later
were analyzed by cross-correlation methods. Correlation between signals
was calculated as the cross-correlation:
![]() |
Validation of In Vitro Location Accuracy
A special test jig was used to test the in vitro accuracy of the
system and locatable-catheter capabilities (Fig 3
). The
test jig was made of a Plexiglas block with seven different holes with
known absolute locations and distances between each other. The catheter
was positioned at each site, and its location was determined 25 times
at each site for five different catheter rolls (protocol A). The SD and
the range of the repeated measurements at each site were calculated.
Next, the relative distances between each site and the six other sites
were calculated as the geometric distances and compared with the known
distances (protocol B).
|
In Vivo Validation of Location Accuracy in the Pig Studies
Studies were performed on healthy male pigs weighing 30 to 40
kg. The experimental protocol was approved by the Animal Use and Care
Committee of the Technion Faculty of Medicine. All pigs were
premedicated with ketamine HCl (20 mg/kg IM) and xylazine (2 mg/kg IM),
anesthetized with pentobarbital (30 mg/kg), intubated, and placed on a
Harvard large-animal mechanical respirator. Vascular access was
obtained with vascular cutdown of the jugular veins, carotid arteries,
and femoral vein and artery as needed.
Protocol 1
At this stage, the catheter was introduced into the left
ventricle. The catheter was positioned in stable contact with the
endocardium, and its location was determined 20 times. This procedure
was repeated at 15 sites. The standard, maximal, and average deviations
at each site were calculated.
Protocol 2
Six locatable catheters were marked with a fine-tip permanent
marker along their shafts at 10-mm intervals and were used as mapping
catheters. Under fluoroscopic guidance, a long 8F sheath was introduced
into the RVA from the right jugular vein. The reference catheter was
introduced through the left jugular vein and was positioned in the CS.
The mapping catheter was introduced through the long sheath all the way
down to the RVA (Fig 4
). At this stage, a marking on the
mapping catheter was aligned with the valve of the long sheath, and the
system was used to record the mapping-tip location. The mapping
catheter was then withdrawn until the next mark was aligned with the
valve, and the tip location was recorded again. This procedure was
repeated 10 times. The distance between two sequential locations was
calculated by the system as the geometric distance and was compared
with the actual distance between the marks on the respective catheters.
Altogether, the procedure was carried out 37 times with six different
catheters and a total of 333 measured distances.
|
Protocol 3
Mapping of the left ventricle. Seventy-seven
ventricular activation maps were acquired, 55 during midseptal right
ventricular pacing and 22 during sinus rhythm. The reference catheter
was positioned, with fluoroscopic guidance, at the CS or the RVA. Using
fluoroscopy, the mapping catheter was introduced into the left
ventricle through the carotid artery. The first three points (two at
the base, near the aortic and mitral valves, and one at the apex) were
acquired with the use of fluoroscopy. No guiding catheters were used
for the mapping procedure. Each activation map was reconstructed from
multiple sampling points ranging from 20 to 352 points. During the
pacing experiments, the catheter was navigated to the earliest site of
activation without fluoroscopy. The relative location of the mapping
catheter tip to the right ventricular pacing catheter was qualitatively
evaluated by viewing the distance between the two catheters from
multiple fluoroscopic angles. Because the two catheters were on
different chambers, the distance between them depended on the thickness
of the septum. The two catheters were considered to be adjacent if the
maximal distance between them in the different fluoroscopic views was
<12 mm (three times the length of the tip) and the mapping
catheter was pointing toward the position of the pacing catheter.
Mapping the right ventricle and right atrium. A total of 11 activation maps of the right ventricle and 32 activation maps of the right atrium during sinus rhythm were acquired. The mapping catheter was introduced through the jugular or femoral veins under fluoroscopic guidance and positioned in the desired chamber. We acquired the first three points with fluoroscopy (in the right ventricle: apex, outflow tract, and tricuspid annulus; in the right atrium: near the SVC, at the right atrial junction near the IVC, and at the ostium of the CS). The rest of the mapping procedure was achieved without the use of fluoroscopy.
Statistical Analysis
Results are reported as mean±SEM. Correlation between signals
was calculated as the cross-correlation between two signals aligned for
a window of 400 ms around the QRS complex.
| Results |
|---|
|
|
|---|
Validation of In Vitro Location Accuracy
The in vitro studies consisted of two different protocols.
In protocol A, the dispersion of repeated location measurements at
the same site was studied. As can be seen from the
Table
, the SD of the "local cloud" of the repeated
measurements ranged from 0.11 to 0.29 mm at the seven different
sites, with the average SD being 0.16±0.02 mm. Maximal range
extended from 0.35 to 0.89 mm, with the average maximal range
being 0.55±0.07 mm.
|
In protocol B, the geometric distances of each site from the other six sites were calculated by the system and compared with the known distances. The mean error ranged from 0.31 to 0.71, with the total mean error being 0.42±0.05 mm.
Validation of In Vivo Location Accuracy
Protocol 1
Similar to the in vitro measurements, we tested the
reproducibility of repeated location measurements of the catheter tip
while it was inside the beating heart. The SD of the local cloud was
0.74±0.13 mm during 20 consecutive location determinations at 15
different sites. The average maximal range and average mean error for
all sites were 1.26±0.08 and 0.54±0.05 mm, respectively.
Protocol 2
In this protocol, the relative distances as measured by the system
while sequentially redrawing the catheter at 10-mm intervals were
compared with the actual distances. The procedure was repeated several
times with six different catheters. The mean error ranged from
0.63±0.07 to 0.84±0.07 mm. The overall mean error for 333
different measurements was 0.73±0.03 mm.
Protocol 3
General information. From our initial experience, we
found that after acquiring the first three points under fluoroscopic
guidance, the rest of the mapping procedure could be achieved without
the use of fluoroscopy. We also noted that the mapping catheter
achieved stable contact with the endocardium usually after three to
four heartbeats and that nonfluoroscopic navigation and mapping in the
swine heart were achieved relatively fast without the use of guiding
sheaths.
Results of mapping the left ventricle. Seventy-seven left
ventricular activation maps were acquired, 55 during pacing of the
right ventricle and septum and 22 during sinus rhythm. The geometries
of all maps were similar, and all had the following characteristics:
apex pointing to the right side of the chest, base located superiorly,
and left atrium close to the left lateral aspect of the left ventricle
(Fig 5
). We noted two indentations of the geometry,
which corresponded anatomically to the position of the papillary
muscles.
|
The activation sequence was similar in all maps acquired during similar conditions. During ventricular pacing, the earliest site of activation was at the site of pacing. During sinus rhythm, the earliest site of activation was on the superior part of the septum; in some pigs, a second breakthrough was noted more posteriorly on the septum. Invariably, the latest site of activation in both rhythms was on the left lateral wall close to the mitral valve annulus. The total activation time of the left ventricle varied between 40 and 80 ms during sinus rhythm and between 57 and 87 ms during paced rhythm. During pacing, when the catheter was navigated back to the site of earliest activation, its relative location was compared with the location of the right ventricular pacing electrode. Using our semiquantitative protocol, we found that in all cases its location was considered to be adjacent to the pacing electrode.
Fig 5A
shows a right lateral view of a typical 3D activation map of the
left ventricle during right ventricular midseptum pacing. The map was
reconstructed using 60 sampling points. Note that the site of earliest
activation (red) was located in the inferomedial septum adjacent to the
pacing site. Activation then spread to the rest of the ventricle, with
the posterolateral areas (blue and purple) being activated last. The
rest of the ventricle had intermediate LATs (yellow and green).
Fig 5B
represents a typical activation map of the left ventricle during
sinus rhythm, reconstructed from 78 sampling points. The earliest
activation site (red) was located in the anterosuperior region in all
maps. Rapid conduction was noted from the site of earliest activation
to the apex. The activation then spread to the rest of the endocardium,
with the posterobasal wall and the AV ring being activated last (blue
and purple).
Results of mapping the right atrium. The right atrial
chamber geometry was similar in all cases (Fig 6
),
showing the following details. Most of the right atrial wall area was
located on the posterior aspect. The most prominent part of the
anterior wall of the right atrium was the tricuspid annulus. On the
posterior wall of the right atrium, there were two prominent
protrusions: the CS ostium and the IVC. The SVC was mapped on the upper
part of the right atrium. The earliest site of activation (SA node) was
located on the posterior wall, close to the SVC connection with the
right atrial wall. The activation then spread to the rest of the
atrium, with relatively fast conduction along the posterior wall toward
the IVC and the superior part of the right atrium toward the left
atrium. The latest activation was detected at the tricuspid annulus and
along the CS. Total activation time of the right atrium varied between
48 and 89 ms.
|
Fig 7
displays three sequential steps in
the propagation map of the swine right atrium, shown from a right
lateral viewing angle. The red areas represent areas that have been
activated. Note again the spread of the activation wave from the
superolateral aspect of the atrium (SA node region) to the rest of the
atrium.
|
Results of mapping the right ventricle. The maps of the
right ventricle disclosed a similar anatomy (Fig 8
). All
the maps had similar shapes, with the right ventricular outflow tract
protruding to the left side of the pig's chest. The base was located
superiorly. Maps were acquired during sinus rhythm, and the earliest
activation site was on the lower part of the septum, with fast
conduction toward the apex. The septum seemed to
bulge into the right ventricular chamber. The latest sites of
activation were the right lateral wall and the posterobasal portion of
the right ventricle.
|
| Discussion |
|---|
|
|
|---|
In Vitro Studies
The accuracy of location was verified by use of bench studies with
a custom-made high-precision test jig (20x20x20 cm). This mapping
volume was large enough to contain the human heart. First, the
reproducibility of location determination at each site was determined
during multiple acquisitions with different orientations and was found
to be highly reproducible (SD, 0.16±0.02 mm; range,
0.55±0.07 mm). Next, the geometric distances between the
locations at the different sites were calculated and compared with the
actual known distances. A high degree of precision was found, with the
average error being 0.42±0.05 mm.
In Vivo Studies
Correcting for Animal and Cardiac Motion
When trying to determine the mapping catheter locations, one must
deal with two major motion artifacts. First, the heart is constantly
beating, so the catheter location changes throughout the cardiac cycle.
Second, the heart may move inside the body (respiration), and the
animal itself may be moving. By gating the catheter location to a
fiducial point in the cardiac cycle, we can overcome the cardiac cycle
motion artifacts as long as the cycle is periodic and repeatable. To
compensate for the pig's movements and the cardiac motion within the
pig, we correct the location of the mapping catheter with the location
of a reference catheter placed in a fixed intracardiac site.
In Vivo Accuracy Studies
As in the in vivo studies, the reproducibility of location at a
single site and the accuracy of the relative distances between
different sites were tested. We found that catheter location could be
determined with a high degree of precision (mean relative distance
error, 0.73±0.03 mm) and good reproducibility (SD,
0.74±0.05 mm).
Significance of and Possible Reasons for Error
Several factors can influence the in vivo accuracy results. The
inherent noise of the location system contributes to the
overall error, together with the fact that in the
heart the location of two catheters is used, thereby increasing the
relative location noise. Other factors that can influence location
accuracy are the reproducibility of the fiducial point on the ECG, the
effect of respiration on chamber geometry, and the reproducibility of
cardiac mechanics on a beat-to-beat basis. Altogether, the results
presented in this study prove that this method is highly accurate and
reproducible.
In addition, currently used electrophysiological catheters have 2- or 4-mm-tip electrodes. Hence, the location error reported here, which is <1 mm, is not significant, because the tip electrode actually averages electrograms from a 2- or 4-mm area.
Comparison With Other Mapping Modalities
As described earlier, current mapping methods can be divided into
methods that use a large number of electrodes and acquire the
electrograms simultaneously from a plurality of sites and methods that
acquire electrograms sequentially.
Current techniques for mapping simultaneously from multiple sites present certain difficulties. For example, the endocardial multielectrode balloon requires open-heart surgery, heart-lung bypass, emptying of the blood from the heart, and inflation of the balloon so that the electrodes will be in direct contact with the endocardium. In addition to the invasive nature of the procedure and the risk involved, the arrhythmia often cannot be induced during surgery.
Other methods that use multielectrode arrays are the epicardial sock and the basket electrode. The spatial resolution of these methods is limited by the relatively large distances between the recording electrodes and by the fact that the absolute position of the target on the myocardium is not known once the electrodes have been removed. Furthermore, the quality of the contacts of all the electrodes with the endocardium or epicardium cannot be ensured (ie, there are sites that are not mapped). Another pitfall of these methods (especially with the basket electrode) is the relative movement between the constantly contracting heart and the electrodes. The interelectrode distances on each sidearm of the basket catheter are fixed, whereas the distances between the actual recording sites on the endocardium decrease during systole. This leads to relative movement between the recording electrode and the tissue, significantly limiting the accuracy of this mapping method.
More commonly, clinical electrophysiology is performed with catheters that are introduced percutaneously into the heart chambers. During endocardial catheter mapping of the heart, the physician sequentially positions the tip at several locations. Localization and navigation of the tip are commonly done with multiplanar fluoroscopy. The currently used method is lacking in several different aspects. Because of the need to use multiple x-ray planes, the procedure is limited as to the number of recording sites, is time-consuming, and because a single localization is made over several cardiac cycles, cannot account for beat-to-beat variability. Moreover, another major limiting factor is the inability to accurately associate (tag) a specific spatial position in the heart with its specific electrogram. We are unaware of literature quantifying the spatial resolution and accuracy of fluoroscopy-based mapping methods. A further limitation of the currently used technique is the inability to accurately position the in-dwelling catheter tip in a site that was previously mapped. Finally, the use of ionizing radiation, which in some procedures can reach high-exposure doses, is hazardous for both the patient and the physician.
Study Limitations
The limitations of the method described in this paper are common
to other methods that use sequential sampling of endocardial sites. The
basic assumption is that the activation pattern and chamber geometry
are constant from beat to beat. This limitation is partially addressed
by the stability criteria used before the information, specifically the
location and LAT stability tests, is acquired. The addition of on-line
body-surface ECG cross-correlation test of consecutive QRS morphologies
will increase the likelihood of proving that the propagation wave is
reproducible.
Clinical Use
The new mapping method described in this report
enables, for the first time, the association of electrical and spatial
endocardial information. This unique ability, together with the
capability to collect a large number of sites in a relatively short
period without the use of ionizing radiation, may offer several
advantages for both the mapping procedure and guiding ablation
treatment. First, the high spatial resolution, found in our studies to
be <1 mm, enables the construction of a very detailed activation
map. Second, the ability to accurately combine the 3D anatomy of the
relevant chamber with its electrical activities offers unique insight
into the possible role of the anatomy of the chamber and the genesis of
cardiac arrhythmias. This quality may play a role in guiding
anatomically based ablation procedures. Third, the ability to navigate
the catheter in three dimensions without the limitations of
two-dimensional fluoroscopy should reduce x-ray exposure and procedure
time. Fourth, the ability to relocate accurately in three dimensions
the tip of the mapping catheter with reference to the electroanatomical
map that was acquired brings a unique value and advantages for ablation
procedures. One may first perform the mapping procedure, next develop
ablation strategies and identify the target for ablation, and finally
return accurately to the desired site for the delivery of energy. This
unique feature, not possible with currently used image-guided
techniques, may also be used for tagging the site where the energy was
delivered. Finally, the unique capability of the new method to
associate relevant electrophysiological information with the
appropriate spatial location in the heart may play a significant role
in guiding ablation strategies, not only in focal arrhythmias but also
in reentrant arrhythmias. In the latter case, pacing maneuvers rather
than the site of earliest local activation guide the selection of the
target for ablation. The ability to record and store a wide spectrum of
relevant electrophysiological information for every sample site can
become a valuable tool in the selection of the most appropriate site
for ablation. Thus, relevant information such as the LAT, the unipolar
and bipolar electrograms, the results of entrainment, and pace mapping
can all be analyzed, stored, and associated with a specific
"address" on the endocardium. This will enable selection of the
best target site for ablation and then, as discussed earlier,
navigating back with precision to the predefined site to deliver the
therapeutic energy.
Nevertheless, although the theoretical clinical advantages of the new method for mapping and ablations are obvious, future prospective studies will have to prove its clinical utility. This study opens the way to investigate the potential impact of the new method on specific clinical parameters such as reduction of fluoroscopy and procedure times, the reduction of number of radiofrequency applications, and success rate.
In conclusion, the method described in this paper combines, for the first time, electrophysiological information with the 3D anatomy of the heart chambers. This will allow better understanding of the mechanisms involved in the genesis of arrhythmias and the influence of pharmacological and nonpharmacological therapies.
| Selected Abbreviations and Acronyms |
|---|
|
| Acknowledgments |
|---|
Received September 23, 1996; accepted November 18, 1996.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
B. John, M. K. Stiles, P. Kuklik, S. T. Chandy, G. D. Young, L. Mackenzie, L. Szumowski, G. Joseph, J. Jose, S. G. Worthley, et al. Electrical remodelling of the left and right atria due to rheumatic mitral stenosis Eur. Heart J., July 10, 2008; (2008) ehn329v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Ruiz-Granell, A. Ferrero, S. Morell-Cabedo, A. Martinez-Brotons, V. Bertomeu, A. Llacer, and R. Garcia-Civera Implantation of single-lead atrioventricular permanent pacemakers guided by electroanatomic navigation without the use of fluoroscopy Europace, June 3, 2008; (2008) eun139v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Okumura, B. D. Henz, S. B. Johnson, T. J. Bunch, C. J. O'Brien, D. O. Hodge, A. Altman, A. Govari, and D. L. Packer Three-Dimensional Ultrasound for Image-Guided Mapping and Intervention: Methods, Quantitative Validation, and Clinical Feasibility of a Novel Multimodality Image Mapping System Circ Arrhythmia Electrophysiol, June 1, 2008; 1(2): 110 - 119. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. S. Srivathsan, T. J. Bunch, S. J. Asirvatham, W. D. Edwards, P. A. Friedman, T. M. Munger, S. C. Hammill, Y.-M. Cha, P. A. Brady, A. Jahangir, et al. Mechanisms and Utility of Discrete Great Arterial Potentials in the Ablation of Outflow Tract Ventricular Arrhythmias Circ Arrhythmia Electrophysiol, April 1, 2008; 1(1): 30 - 38. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Rao, Y. Ling, R. He, A. L. Gilbert, N. G. Frangogiannis, J. Wang, S. F. Nagueh, and D. S. Khoury Integrated multimodal-catheter imaging unveils principal relationships among ventricular electrical activity, anatomy, and function Am J Physiol Heart Circ Physiol, February 1, 2008; 294(2): H1002 - H1009. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Carbucicchio, M. Santamaria, N. Trevisi, G. Maccabelli, F. Giraldi, G. Fassini, S. Riva, M. Moltrasio, M. Cireddu, F. Veglia, et al. Catheter Ablation for the Treatment of Electrical Storm in Patients With Implantable Cardioverter-Defibrillators: Short- and Long-Term Outcomes in a Prospective Single-Center Study Circulation, January 29, 2008; 117(4): 462 - 469. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. E. Eckart and L. M. Epstein Interventional Therapy for Atrial and Ventricular Arrhythmias Card. Surg. Adult, January 1, 2008; 3(2008): 1357 - 1374. [Full Text] |
||||
![]() |
J. E. Marine Catheter Ablation Therapy for Supraventricular Arrhythmias JAMA, December 19, 2007; 298(23): 2768 - 2778. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. F Tops, E. E van der Wall, M. J Schalij, and J. J Bax Multi-modality imaging to assess left atrial size, anatomy and function Heart, November 1, 2007; 93(11): 1461 - 1470. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Y. Reddy, P. Neuzil, Z. J. Malchano, R. Vijaykumar, R. Cury, S. Abbara, J. Weichet, C. D. McPherson, and J. N. Ruskin View-Synchronized Robotic Image-Guided Therapy for Atrial Fibrillation Ablation: Experimental Validation and Clinical Feasibility Circulation, May 29, 2007; 115(21): 2705 - 2714. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Abrams, M. J. Earley, S. C. Sporton, P. M. Kistler, M. A. Gatzoulis, M. J. Mullen, J. A. Till, S. Cullen, F. Walker, M. D. Lowe, et al. Comparison of Noncontact and Electroanatomic Mapping to Identify Scar and Arrhythmia Late After the Fontan Procedure Circulation, April 3, 2007; 115(13): 1738 - 1746. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Yla-Herttuala, T. T. Rissanen, I. Vajanto, and J. Hartikainen Vascular Endothelial Growth Factors: Biology and Current Status of Clinical Applications in Cardiovascular Medicine J. Am. Coll. Cardiol., March 13, 2007; 49(10): 1015 - 1026. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. M. Lang, V. Mor-Avi, L. Sugeng, P. S. Nieman, and D. J. Sahn Three-Dimensional Echocardiography: The Benefits of the Additional Dimension J. Am. Coll. Cardiol., November 21, 2006; 48(10): 2053 - 2069. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Tracy, M. Akhtar, J. P. DiMarco, D. L. Packer, H. H. Weitz, M. A. Creager, D. R. Holmes Jr, G. Merli, G. P. Rodgers, C. M. Tracy, et al. American College of Cardiology/American Heart Association 2006 Update of the Clinical Competence Statement on Invasive Electrophysiology Studies, Catheter Ablation, and Cardioversion: A Report of the American College of Cardiology/American Heart Association/American College of Physicians Task Force on Clinical Competence and Training Developed in Collaboration With the Heart Rhythm Society J. Am. Coll. Cardiol., October 3, 2006; 48(7): 1503 - 1517. [Full Text] [PDF] |
||||
![]() |
P. Kojodjojo, P. Kanagaratnam, O. R. Segal, W. Hussain, and N. S. Peters The Effects of Carbenoxolone on Human Myocardial Conduction: A Tool to Investigate the Role of Gap Junctional Uncoupling in Human Arrhythmogenesis J. Am. Coll. Cardiol., September 19, 2006; 48(6): 1242 - 1249. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Earley, D. J.R. Abrams, S. C. Sporton, and R. J. Schilling Validation of the Noncontact Mapping System in the Left Atrium During Permanent Atrial Fibrillation and Sinus Rhythm J. Am. Coll. Cardiol., August 1, 2006; 48(3): 485 - 491. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. V. Daniels, Y.-Y. Lu, J. B. Morton, P. A. Santucci, J. G. Akar, A. Green, and D. J. Wilber Idiopathic Epicardial Left Ventricular Tachycardia Originating Remote From the Sinus of Valsalva: Electrophysiological Characteristics, Catheter Ablation, and Identification From the 12-Lead Electrocardiogram Circulation, April 4, 2006; 113(13): 1659 - 1666. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Duytschaever, S. Y. Ho, D. Devos, and R. Tavernier The left hand as a model for the right atrium: a simple teaching tool. Europace, April 1, 2006; 8(4): 245 - 250. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Dong, H. Calkins, S. B. Solomon, S. Lai, D. Dalal, A. Lardo, E. Brem, A. Preiss, R. D. Berger, H. Halperin, et al. Integrated Electroanatomic Mapping With Three-Dimensional Computed Tomographic Images for Real-Time Guided Ablations Circulation, January 17, 2006; 113(2): 186 - 194. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Grothues, O. Wolfram, C. Fantoni, H. Boenigk, A. Götte, C. Tempelmann, H.U. Klein, and A. Auricchio Volume measurement by CARTOTM compared with cardiac magnetic resonance. Europace, January 1, 2006; 8(1): 37 - 41. [Abstract] [Full Text] [PDF] |
||||