(Circulation. 1997;95:1320-1327.)
© 1997 American Heart Association, Inc.
Articles |
the Department of Cardiovascular and Thoracic Surgery and the Division of Cardiovascular Medicine (A.F.B.), Stanford (Calif) University School of Medicine; Research Institute of the Palo Alto (Calif) Medical Foundation (G.T.D., S.D.N., N.B.I.); and Cardiac Surgery and Cardiology Sections (M.R.M., A.D., M.K., D.C.M.), Department of Veterans Affairs Medical Center, Palo Alto, Calif.
Correspondence to D. Craig Miller, MD, Department of Cardiovascular Surgery, Falk Cardiovascular Research Center, Stanford University Medical School, Stanford, CA 94305.
| Abstract |
|---|
|
|
|---|
Methods and Results To evaluate the effects of LV volume and pressure unloading on septal geometry and function, LV preload was abruptly reduced by clamping left atrial pressure between 0 and -2 mm Hg in seven open-chest, anesthetized dogs by use of a pressure-control servomechanism to withdraw blood from the left atrium. With left atrial pressure clamping, maximal LV pressure decreased 30±12% (mean±SD) (P<.0001) and LV end-diastolic cross-sectional area (determined by two-dimensional echocardiography) decreased by 53±16% (P<.0001). This caused the septum to shift toward the left (RV septal free-wall dimension increased; P<.004) and flatten (radius of curvature increased; P<.0002), while LV septal free-wall dimension fell (P<.0001). Septal end-diastolic thickness increased 23±15% (P<.0005), reflecting a decline in septal preload. Systolic septal thickening decreased (P<.002), while systolic septal output (Septal Output=Septal ThickeningxHeart Rate) fell from 30±17 to 15±22 cm/min (P<.002). This was associated with movement along the septal Frank-Starling equivalent (septal output versus end-diastolic septal thickness [preload] relation) to a less productive portion of the curve.
Conclusions LV unloading not only altered interventricular septal geometry but also reduced septal systolic thickening and output, all of which may contribute to impaired RV contractility during mechanical LV support.
Key Words: echocardiography heart-assist device ventricles mechanics
| Introduction |
|---|
|
|
|---|
LV volume and pressure unloading (preload reduction) has been shown to alter RV systolic mechanics in both closed-chest animals and open-chest humans.5 6 7 8 Mechanical LV support impairs global RV contractility, but in normal hearts or those with isolated RV free-wall ischemia, overall RV pump performance and mechanical efficiency do not fall because of a simultaneous and offsetting decrease in RV afterload.6 9 Previous studies from our laboratory and others4 5 6 10 also showed that the interventricular septum flattens and shifts toward the left with LV unloading, and the transseptal pressure gradient falls; however, the inability to measure septal thickness and geometry simultaneously in these previous studies has hindered evaluation of septal mechanics during LV support. Damiano and associates11 showed that RV function is highly dependent on contraction of the left ventricle and septum, with LV systolic forces responsible for >60% of RV contractile force. As the septum shifts away from the right ventricle, one would predict a decline in RV contractile force solely on the basis of systolic ventricular interdependence, a direct interaction due to anatomic coupling of the ventricles. Increasing levels of mechanical LV support cause the interventricular septum to move farther to the left, which further impairs RV contractility,6 but the effects of LV unloading on intrinsic septal function remain unclear. Does LV unloading depress systolic thickening of the interventricular septum, or is septal systolic function normal and its contribution to RV contractile force diminished merely by anatomic shifting consequent to changes in transseptal pressure? In the present study, the left ventricle was rapidly volume unloaded using a large-bore cannula inserted into the left atrium to reduce LAP to 0 or -2 mm Hg; this pressure was maintained at the desired target level with continued servo-controlled blood withdrawal. Using 2D epicardial echocardiography, we were able to quantify RV and LV size and shape and evaluate changes in septal thickening, septal geometry, and septal systolic function in response to rapid LV unloading.
| Methods |
|---|
|
|
|---|
The LA cannula was connected to a computer-driven, pressure-control servomechanism as previously described.12 The computer software used to drive the pump was designed to allow continuous feedbackcontrolled suction through the LA cannula to maintain LAP at a predetermined level. After two baseline beats were recorded, the servo-pump mechanism was initiated during the third beat, and LAP was clamped at the desired level. The servo pump continuously withdrew blood from or infused blood into the left atrium as necessary to maintain LAP at the desired level for a minimum of three cardiac cycles.
Before initiation of data collection, UL-FS49, a highly selective and specific dromotropic agent that affects the sinoatrial node but does not significantly change QT interval, inotropic state, or systolic or diastolic blood pressure was administered as necessary to reduce the heart rate to 80 to 100 min-1 (0.25 to 0.5 mg/kg IV; Boehringer-Ingelheim). To minimize the effects of intrathoracic pressure variation, the respirator was temporarily interrupted at end expiration during data collection for
10 seconds. Simultaneous 2D epicardial echocardiographic images and hemodynamic data were obtained at LAP clamping levels of 0 and -2 mm Hg for all animals; data-acquisition runs containing premature ventricular contractions were excluded and repeated. Although data were acquired for all animals at both levels of LAP clamping, the observations reported herein include five of seven animals at LAP clamping of 0 mm Hg and six of seven animals at LAP clamping of -2 mm Hg. Arrhythmias were quite common with the rapid withdrawal of blood through the LA cannula; despite attempts to repeat runs with irregular beats, ectopy-free LAP control was not possible in all cases. Therefore, to exclude erroneous data, runs with atrial or ventricular premature beats or inadequate echocardiographic images were excluded from this analysis.
At the conclusion of the experiment, the dogs were killed by use of B-euthanasia (0.1 mL/lb IV; Schering-Plough Animal Health Corp), and proper positioning of the catheters and cannula was confirmed. All animals received humane care in compliance with the "Principles of Laboratory Animal Care" formulated by the National Society for Medical Research and the Guide for the Care and Use of Laboratory Animals prepared by the National Academy of Sciences and published by the National Institutes of Health (DHEW [NIH] publication 85-23, revised 1985). The study was approved by the Stanford Medical Center Laboratory Research Animal Review Committee and conducted according to Stanford University policy.
Data Acquisition
2D direct epicardial echocardiography was performed with the use of a 5.0-MHz transducer (77020 AC; Hewlett-Packard) to evaluate geometric changes of the left ventricle, right ventricle, and interventricular septum. Cross-sectional images were obtained of both the right and left ventricles at the LV midpapillary muscle level (parasternal short-axis equivalent). A surface ECG signal and LAP tracing were simultaneously recorded on the echocardiographic images to allow proper temporal registration with the hemodynamic data at 30 Hz (every 33.3 ms) on .5-in videotape for later stop-frame analysis. All echocardiographic data were reviewed in real time, and segments of interest were transferred to a microcomputer (VAX 11/750; Digital Equipment Corp) through a time base corrector (AD 51, Hotronic, Inc) to synchronize video images with the image-processing system (IP8500, Gould Inc). Echocardiographic frames of interest were analyzed with software developed for our laboratory, and the system was calibrated with the use of the internally generated markers displayed by the echocardiogram.
During each data-acquisition run, seven channels of analog data were recorded (LVP, LAP, PA pressure, RVP, aortic flow, PA flow, and surface lead ECG) and digitized simultaneously at 240 Hz with the use of two 486-based microcomputers (486-33 and 486-50; JDR Microdevices Inc) with high-speed data-acquisition cards (DT 3831-G; Data Translation Inc) controlled by commercially available software (Labtech Control 3.2.0; Laboratory Technology Corp). This software system also controlled the servo-pump mechanism throughout the experiment.
Data Analysis
Echocardiographic images were analyzed at both ED and ES. To coordinate the 240-Hz hemodynamic data and 30-Hz echocardiographic data, ED was defined as the time of peak ECG R wave, and ES was defined as the frame or time (number of frames beyond peak ECG R wave multiplied by 33.3 ms) of minimum LV echocardiographic area. All echocardiographic RV and LV endocardial tracings eliminated trabecular irregularities and cut through the base of the papillary muscles. Cross-sectional LV or RV area was then calculated as the area within the endocardial tracings for each ventricle. SFW dimensions were measured from the midpoint of the LV and RV endocardial free-wall perimeters to their respective endocardial septal surfaces, perpendicular to the septum; septal thickness was also measured along this line. Echocardiographic measurements included: RV and LV AreaED and AreaES, RV and LV SFWED and SFWES, and SeptED and SeptES. RV and LV APEDs were also measured, and RV and LV short-axis end-diastolic eccentricities (eED) were calculated to quantify ventricular shape: eED=(APED2-SFWED2)1/2/APED. Stroke area was computed for each ventricle as Stroke Area=AreaED-AreaES, and Sept-ST was calculated as Sept-ST=SeptES-SeptED (positive value represents thickening during systole).
Septal ROC at ED was quantified with the use of techniques described by Silverman et al13 and Hutchins et al.14 The LV endocardial surface was traced with the image-processing system, and the middle portion of the septal perimeter (excluding the anterior and posterior regions of insertion into the RV or LV free walls) was best fit by matching curves of known ROC etched on a transparent sheet. Septal curvature was measured independently in duplicate by two individuals to maximize accuracy. To decrease the influence of LV intracavitary volume on measurement of changes in ROC from loaded to unloaded beats, ROC was normalized using the ROCi as previously described.10 15 Assuming the LV cavity cross section to be perfectly circular, ROCi=(LV AreaED/
)1/2 and n-ROC=ROC/ROCi.
LVP, RVP, and TSP (LVP-RVP) were determined at ED (LVPED, RVPED, and TSPED), and mean developed (ejection) LVP, RVP, and TSP were calculated where the beginning of ejection was defined as the time of rapid upstroke of the aortic or PA flow signal and the end of ejection as the time of minimum LV echocardiographic area to allow matching of the hemodynamic and echocardiographic data. Maximal LVP and RVP were also determined (LVPmax, RVPmax). RV afterload was quantified with the use of pulmonary vascular impedance (input impedance plus characteristic impedance), calculated by use of Fourier analysis as previously reported.6 9 16 Input impedance describes the ratio of oscillatory pressure to oscillatory flow and depends on the cross-sectional area of the pulmonary bed and LAP. Characteristic impedance represents the pulsatile component of pulmonary flow and is influenced by vascular wall compliance and blood inertia. Effective Ea of the pulmonary vascular bed (RV Ea), which is a function of maximal RVP and SV, was also calculated to estimate RV afterload as RV Ea=RVPmax/SV.8 17 RV Ea represents a steady-state approximation of maximal pulmonary vascular impedance.
Global LV and RV SVs were determined by integrating aortic and PA flow, respectively, throughout systole; global SW was calculated for each ventricle as SW=SVxMean Developed Pressure (Mean Ejection Pressure-ED Pressure). To quantify Sept-Output (septal displacement per minute), Sept-ST was multiplied by heart rate: Sept-Output=Sept-STx(RR Intervalx60/1000). Sept-Output was then plotted against end-diastolic thickness (septal preload) to examine the septal Frank-Starling equivalent.
Statistical Analysis
All data are reported as mean±SD. The two normal "unloaded" baseline beats before LAP clamping were averaged (beats 1 and 2 in Figs 1
and 2) and then compared with the initial "unloaded" beat after LAP clamping (beat 4 in Figs 1 and 2![]()
) with the use of Student's t test for paired comparisons and Hotelling's T2 test for multivariate analysis of related variables (Sept-ST and Sept-Output). Differences were considered significant at a level of P<.05.
|
|
| Results |
|---|
|
|
|---|
|
Typical echocardiographic images from the same animal illustrated in Figs 1 and 2![]()
are shown in Fig 3A and 3B![]()
. In this animal with LAP clamping of -2 mm Hg, LV AreaED fell from 17.1 to 7.5 cm2, while RV AreaED increased from 11.5 to 13.4 cm2. For all animals with LAP clamping, LV preload (ie, LV AreaED) fell by 52% (P<.0001); LV stroke area and SV fell by 59% (P<.0004) and 73% (P<.0001), respectively; and global LV SW decreased 77% (P<.0006) (Table 2
). These findings all demonstrate substantial volume unloading of the left ventricle with LAP clamping. On the right side of the heart, RV preload (ie, RV AreaED) increased 21% (P>.03) during LAP clamping, but RV afterload decreased significantly. Although there was no change in PA characteristic impedance (281±68 versus 276±132 dyne s/cm5; P>.91), PA input impedance declined significantly from 809±330 to 606±274 dyne s/cm5 (P<.02). In addition, RV Ea, another reflection of RV afterload, dropped significantly from 1.8±0.6 to 1.6±0.5 mm Hg/mL (P<.02). RV stroke area increased fivefold with LAP clamping (P<.007). Although there was a trend for RV SV and SW to rise, these changes did not reach statistical significance (RV SV, P>.07; RV SW, P>.06) (Table 2
). The end-diastolic geometric configuration of both ventricles changed markedly (Fig 3A and 3B![]()
). Although RV SFWED increased (P<.004), RV APED did not change (P>.90), thereby resulting in the RV cross-sectional shape becoming less elliptical (ie, RV eED decreased; P<.003) (Table 2
). On the other hand, although both LV SFWED and APED decreased (P<.0001 for both), LV SFWED fell to a relatively greater extent, causing the LV shape to become more elliptical in the short-axis plane (eg, LV eED increased; P<.005).
|
|
With respect to septal geometry, a significant leftward shift occurred (increase in RV SFWED and decrease in LV SFWED) associated with the fall in TSPED from 2.6±1.5 to -3.8±1.9 mm Hg (P<.0001). Along with this leftward septal shift, the septum itself flattened; septal ROC more than doubled with LV unloading (P<.0002) (Table 3
), and n-ROC rose by greater than threefold (P<.0001). Fig 3
(A and B) illustrates how the septum both flattened and shifted to the left; in this animal, RV SFWED increased from 2.2 to 2.9 cm, while LV SFWED decreased from 4.1 to 2.5 cm, resulting in the marked increase in septal ROC (septal flattening). Significant changes in septal mechanics also occurred in all animals (Table 3
). SeptED increased by 23% with LAP clamping (P<.0005). Thickening of the septum represents a decrease in fiber length or a reduction in septal preload. Furthermore, Sept-ST significantly declined (P<.002) and Sept-Output fell by 96% (P<.002). In addition, developed LVP dropped from 77±17 to 54±19 mm Hg (P<.0001) while developed RVP did not change (23±4 versus 23±4 mm Hg; P>.40), resulting in a significant decline in mean developed TSP (52±15 versus 33±15 mm Hg; P<.0001). Fig 4
demonstrates the changes in the Sept-Output versus SeptED relation (Frank-Starling equivalent) during LAP clamping in all animals. While SeptED increased (a decrease in septal preload; see above), Sept-Output decreased; in most cases, the septum was working on a less-productive region of its Frank-Starling curve.
|
|
| Discussion |
|---|
|
|
|---|
Morphologically, the septum can be considered a part of the left ventricle, but it contributes importantly to the function of both ventricles in normal hearts. Banka and coworkers1 examined septal motion in humans using simultaneous RV and LV angiography and found that the septum thickened equally in both the RV and LV directions and also shortened along its apex-to-base longitudinal axis. Li and Santamore26 observed that injecting glutaraldehyde into the septum of isolated rabbit hearts depressed RV developed pressure by 37%, and Agarwal and colleagues27 demonstrated that isolated ischemia of the interventricular septum was associated with RV but not LV systolic dysfunction. Thirty minutes after ligation of the septal coronary artery in dogs, Agarwal et al27 observed a shift in the RV Frank-Starling relation (RV SW versus RVPED) downward and to the right (impaired RV systolic performance), while the LV relation did not change. This occurred in the presence of normal systolic shortening in the RV free wall but increased shortening (16%) in the LV free wall (an apparently compensatory response) as measured with the use of ultrasonic segment length crystals. They concluded that while septal ischemia did not provoke LV dysfunction, possibly due to "hyperfunction" of the nonischemic zones, the high dependency of RV function on normal septal contractility resulted in significant impairment of RV output during septal ischemia.27
More recently, Woodard et al,21 in an elegant study, isolated the systolic effects of LVP on RV mechanics by rapidly withdrawing blood from the left ventricle immediately after diastole via a large-bore LV apical cannula. With this preparation, they were able to quantify LV-to-RV systolic interaction independently of diastolic interference; the LV was not unloaded until systole had begun, and thus end-diastolic conditions were equal for both loaded and unloaded ejections. This becomes important when one notes that diastolic and systolic interactions tend to counteract each other during steady-state changes in LV or RV load. With rapid systolic unloading, mean developed LVP dropped by 63%, which caused a fall in mean developed RVP (14%), RV SV (18%), and RV SW (27%). This was associated with a relative leftward shift of the septum as LV SFW distance decreased and RV SFW distance increased at ES. Whether septal position was the cause or the effect of this systolic "cross talk" between the ventricles or whether septal function per se was impaired owing to the resultant change in TSP was not clear. Determination of the role of the septum in modulating the RV response to LV support would require not only an evaluation of septal contractility and load but a complete evaluation of the 3D geometric septal changes induced by LV unloading, which undoubtedly affects septal afterload.
Using myocardial markers to simultaneously assess 3D RV, LV, and septal geometry, we previously studied the effects of TSP on 3D septal position during LV unloading with an LVAD in closed-chest, conscious dogs.6 LV unloading was not rapid because the animals were allowed to achieve a steady hemodynamic state after the initiation of maximal LV support (100% LVAD flow). TSP fell from 4±4 to -5±4 mm Hg with LV unloading, causing movement of the interventricular septum to the left (14% decrease in LV SFW dimension, 11% increase in RV SFW dimension). The leftward septal shift was associated with a significant decline (-26%) in RV end-systolic elastance (slope of the end-systolic pressure-volume relation), a 15% fall in preload recruitable SW, and a 47% reduction in the slope of the dP/dtmaxend-diastolic volume relationship. Our inability to measure septal thickness in this preparation, however, did not allow us to determine if the decrease in global RV systolic mechanics was due to impairment of septal contractility (systolic thickening) or whether septal function was normal and its contribution to RV contraction was merely diminished by virtue of its leftward shift and flattening.
In the present study, peak LVP fell 30% with rapid LV unloading, which consequently decreased RV afterload (23% decline in PA input impedance, 8% fall in pulmonary Ea) and augmented RV preload (21% rise in RV end-diastolic cross-sectional area). Global RV output, which is a function of RV preload, afterload, and contractility, did not change significantly; although there was a tendency for RV SV and SW to increase, these changes were not consistent and did not reach statistical significance. Although we could not directly measure RV contractility in the present study, one can safely predict on the basis of earlier studies that because afterload fell, preload increased, and output did not significantly change, RV contractility did not improve and most likely declined in these normal canine hearts. This is consistent with previous findings of a 15% to 47% fall in RV contractility with mechanical LV support.4 6
This investigation focused on the geometric and intrinsic functional septal changes induced by LV unloading in an attempt to determine how this influenced RV systolic mechanics. Santamore and coworkers28 showed that RV contraction is the result of a decrease in the surface area of the RV free wall plus a reduction of RV SFW distance during systole. We found that as TSP fell, the interventricular septum moved toward the left and flattened, and the right ventricle developed a less elliptical cross-sectional shape as the RV SFWED increased with respect to the RV APED (RV short-axis eccentricity decreased; see Fig 3A and 3B![]()
). This configuration alone could be responsible for impaired RV contractility, because the septum is less able to contribute to RV ejection; moreover, we found that Sept-ST and Sept-Output were diminished. The septum normally thickens during systole, helping to expel blood from both ventricles, and the ability of the septum to thicken is determined in part by the contractile state of its muscle fibers. With LV unloading, Sept-ST decreased by 94% and Sept-Output fell by 96% in the setting of reduced septal preload. The combination of decreased septal output with an increase in septal end-diastolic thickness is predicted on the basis of the Frank-Starling relation (Fig 4
); the sarcomeres appeared to be working on a less productive portion of the Frank-Starling curve. Therefore, not only did LV unloading induce septal geometric changes that were disadvantageous to RV ejection, LV unloading also decreased septal systolic output. All of these changes may contribute to the reduction in RV pump function that is seen clinically during mechanical LV support.
Recently, using transesophageal 2D echocardiography, Kawai and associates29 analyzed regional RV function in 14 patients by artificially dividing the RV cross-sectional area into free-wall and septal portions on the basis of grids drawn from the RV center of gravity (at ED) to the junctions of the free wall and septum. Regional fractional area change was then determined before and after LVAD insertion. Fractional area change of the RV free-wall portion did not change, but fractional area change of the RV septal portion was 50% lower. Although this technique does not directly measure septal contractility, it implies a decrease in septal contractile force consistent with the findings of the current study.
Development of RV failure after the initiation of mechanical LV support is not a rare clinical phenomenon. Although RV systolic dysfunction has not been shown to occur in normal hearts or in hearts with isolated RV free-wall ischemia,6 8 9 it has been suggested that coexistent ischemia of the interventricular septum may precipitate RV systolic dysfunction in patients dependent on mechanical LV assistance. Daly and colleagues30 found that isolated septal ischemia during bypass of the left side of the heart in open-chest calves decreased RV developed pressure by 29% and RV fractional area change (measured by use of 2D epicardial echocardiography) by 66%. With reperfusion of the septal arteries, these load-dependent measures of RV pump performance returned toward normal. In addition, Sept-ST fell with LV assistance and decreased further after septal ischemia but returned to its preischemic level with septal reperfusion. Daly et al30 concluded that because the septum plays a major role in RV ejection, further reduction of septal contraction due to ischemia could potentially provoke clinical RV failure during LVAD support.
In our experimental preparation, all animals were able to generate negative LVPs (or LV suction) with an intact, normal mitral-valve apparatus and LAP clamping
0. During early diastole when LAP clamping was 0 mm Hg, minimum LVP decreased from a baseline value of 3.3±1.7 to 1.2±1.2 mm Hg in the initial clamped beat and fell to -1.3±1.8 mm Hg during the subsequent beat. With LAP clamping at -2 mm Hg, minimum LVP fell from a baseline value of 3.9±1.3 to 0.4±1.1 mm Hg in the initial clamped beat and to -1.8±1.9 mm Hg during the subsequent beat. LV filling occurs for the most part due to the AV pressure gradient, and if LAP is 0, LVP must be negative to produce filling. It is becoming increasingly clear that LV suction plays a significant role in diastolic filling in times of stress and possibly in cardiac disease states.12 31 32 33 34 35 In our preparation, when LAP was decreased to
0, the normal LA-LV pressure gradient was abolished; however, the left ventricle generated negative pressures that maintained LV filling. Negative transmural LVPs as low as -30 mm Hg have been reported with various techniques used to occlude the mitral annulus rapidly at precise times (eg, ES) to prevent LV filling (or LV volume clamping).31 32 33 Although these studies unequivocally demonstrate the presence of LV suction, the myocardial mechanisms underlying this process remain incompletely characterized.
As LVP fell below zero with LAP pressure-clamping, the interventricular septum both flattened and thickened throughout diastole, creating a half-moon shape with respect to the left ventricle at ED in lieu of its more normal circular configuration. Similar motion of the septum during early diastole has been observed clinically by Weyman and associates36 in patients with severe mitral stenosis. Using transthoracic echocardiography, they noted that the septum flattened and moved to the left during early diastole, but then, presumably after the initial rapid-filling phase, the septum moved rightward to regain its normal circular configuration before ejection. Weyman et al36 speculated that this early leftward shift was due to unequal filling of the ventricles, with the right side filling much faster than the left owing to the substantial obstruction to flow across the stenotic mitral valve. Patients with mitral stenosis were subsequently found to develop negative LVPs (suction) at rest,34 but the role of the septum in the generation of LV suction remains unclear. During diastolic filling, the ventricular wall usually becomes thinner; however, in the current study, we found that the septum flattened and thickened with LAP clamping. Negative LVPs were produced in this preparation while septal thickening appeared to impede LV filling. Could this apparent dysfunctional septal movement actually be contributing to LV filling, or do negative LVPs develop despite abnormal septal motion? It is likely that the thick, flat, nonfunctioning septum associated with LV unloading observed in this preparation did not contribute substantially to the development of negative LVPs, which would suggest that LV suction is not dependent on the interventricular septum. Additional investigation with the use of more sophisticated techniques of evaluating 3D LV wall geometry will be necessary to define the mechanical basis of LV suction.
Study Limitations
This experiment evaluated anesthetized open-chest dogs with presumably normal cardiac function, and although attempts were made to minimize the negative inotropic and arrhythmogenic effects of general anesthesia through the use of fentanyl and benzodiazepines instead of inhalational agents, this preparation is not completely analogous to the intact, closed-chest, conscious state. However, all conditions (open chest, open pericardium, level of anesthesia, etc) were identical during the baseline and the LAP clamped beats, except for the intended perturbation.
Throughout this study, RV endocardial echocardiographic cross-sectional area was used as an estimate of RV volume. Although this is not as accurate as measuring 3D RV volume with myocardial markers,6 Morita et al8 demonstrated that RV area correlated well (r=.96±0.03) with known intracavitary latex balloon volumes in excised canine hearts and was independent of changes in LV balloon volume.
After implantation of the LA cannula and multiple intracardiac monitoring devices, it was not possible to close the pericardium in a physiological manner. Although the absolute magnitude of the pericardial influence on intracardiac pressures and ventricular function remains incompletely defined, it is widely accepted that the pericardium significantly influences diastolic ventricular interaction and also modulates systolic interaction.19 20 23 The clinical application in which we were ultimately interested, however, was that of LVAD use in humans, in whom it is neither possible nor desirable to close the pericardium with all the cannulas in place. Therefore, the open-pericardium preparation used in the current experiment is representative of the clinical scenario. In this preparation, we found that rapid LV volume unloading not only caused the interventricular septum to shift to the left and flatten but also decreased Sept-ST and Sept-Output, all of which may contribute to impaired RV function during mechanical LV support.
| Selected Abbreviations and Acronyms |
|---|
|
| Acknowledgments |
|---|
Received July 29, 1996; revision received October 16, 1996; accepted October 27, 1996.
| References |
|---|
|
|
|---|
2. Richenbacher WE, Pierce WS. Right ventricular failure following implantation of a left ventricular assist device. Curr Surg. 1983;40:274-277.[Medline] [Order article via Infotrieve]
3. Farrar DJ, Compton PG, Hershon JJ, Fonger JD, Hill JD. Right heart interaction with the mechanically assisted left heart. World J Surg. 1985;9:89-102.[Medline] [Order article via Infotrieve]
4.
Chow E, Farrar DJ. Effects of left ventricular pressure reductions on right ventricular systolic performance. Am J Physiol. 1989;257:H1878-H1885.
5. Elbeery JR, Owen CH, Savitt MA, Davis JW, Feneley MP, Rankin JS, VanTrigt P. Effects of the left ventricular assist device on right ventricular function. J Thorac Cardiovasc Surg. 1990;99:809-816.[Abstract]
6. Moon MR, Castro LJ, DeAnda A, Tomizawa Y, Daughters GT, Ingels NB, Miller DC. Right ventricular dynamics during left ventricular assistance in closed-chest dogs. Ann Thorac Surg. 1993;56:54-67.[Abstract]
7.
Farrar DJ, Compton PG, Hershon JJ, Hill JD. Right ventricular function in an operating room model of mechanical left ventricular assistance and its effects in patients with depressed left ventricular function. Circulation. 1985;72:1279-1285.
8. Morita S, Kormos RL, Madarino WA, Eishi K, Kawai A, Gasior TA, Deneault LG, Armitage JM, Hardesty RL, Griffith BP. Right ventricular/arterial coupling in the patient with left ventricular assistance. Circulation. 1992;86(suppl II):II-316-II-325.
9. Moon MR, Castro LJ, DeAnda A, Daughters GT, Ingels NB, Miller DC. Effects of left ventricular support on right ventricular mechanics during experimental right ventricular ischemia. Circulation. 1994;90(part 2):II-92-II-101.
10.
Dong S, Smith ER, Tyberg JV. Changes in the radius of curvature of the ventricular septum at end diastole during pulmonary arterial and aortic constrictions in the dog. Circulation. 1992;86:1280-1290.
11.
Damiano RJ, LaFollette P, Cox JL, Lowe JE, Santamore WP. Significant left ventricular contribution to right ventricular systolic function. Am J Physiol. 1991;261:H1514-H1524.
12.
Ingels NB, Daughters GT, Nikolic SD, DeAnda A, Moon MR, Bolger AF, Komeda M, Derby GC, Yellin EL, Miller DC. Left atrial pressure-clamp servomechanism demonstrates LV suction in canine hearts with normal mitral valves. Am J Physiol. 1994;267:H354-H362.
13. Silverman KJ, Hutchins GM, Weiss JL, Moore GW. Catenoidal shape of the interventricular septum in idiopathic hypertrophic subaortic stenosis: two dimensional echocardiographic confirmation. Am J Cardiol. 1982;49:27-32.[Medline] [Order article via Infotrieve]
14. Hutchins GM, Moore GW, Skoog DK. Catenoidal shape of the interventricular septum acquired secondary to myocardial infarction. Am J Clin Pathol. 1985;83:53-58.[Medline] [Order article via Infotrieve]
15.
King ME, Braun H, Goldblatt A, Liberthson R, Weyman AE. Interventricular septal configuration as a predictor of right ventricular systolic hypertension in children: a cross-sectional echocardiographic study. Circulation. 1983;68:68-75.
16.
Calvin JE, Baer RW, Glantz SA. Pulmonary artery constriction produces a greater right ventricular dynamic afterload than lung microvascular injury in the open chest dog. Circ Res. 1985;56:40-56.
17.
Kelly RP, Ting C, Yang T, Liu C, Maughan WL, Chang M, Kass DA. Effective arterial elastance as index of arterial vascular load in humans. Circulation. 1992;86:513-521.
18. Henderson Y, Prince AL. The relative systolic discharges of the right and left ventricles and their bearing on pulmonary congestion and depletion. Heart. 1914;5:217-226.
19. Elzinga G, van Grondelle R, Westerhof N, van den Bos GC. Ventricular interference. Am J Physiol. 1974;226:941-947.
20. Bove AA, Santamore WP. Ventricular interdependence. Prog Cardiovasc Dis. 1981;23:365-388.[Medline] [Order article via Infotrieve]
21.
Woodard JC, Chow E, Farrar DJ. Isolated ventricular systolic interaction during transient reductions in left ventricular pressure. Circ Res. 1992;70:944-951.
22. Taylor RR, Covell JW, Sonnenblick EH, Ross J. Dependence of ventricular distensibility on filling of the opposite ventricle. Am J Physiol. 1967;213:711-718.
23.
Janicki JS, Weber KT. The pericardium and ventricular interaction, distensibility, and function. Am J Physiol. 1980;238:H494-H503.
24. Starr I, Jeffers WA, Meade RH Jr. The absence of conspicuous increments of venous pressure after severe damage to the right ventricle of the dog, with a discussion of the relation between clinical congestive failure and heart disease. Am Heart J. 1943;26:291-301.
25. Sawatani S, Mandell G, Kusaba E, Schraut W, Cascade P, Wajszczuk WJ, Kantrowitz A. Ventricular performance following ablation and prosthetic replacement of right ventricular myocardium. Trans Am Soc Artif Intern Organs. 1974;20:629-636.
26.
Li KS, Santamore WP. Contribution of each wall to biventricular function. Cardiovasc Res. 1993;27:792-800.
27. Agarwal JB, Yamazaki H, Bodenheimer MM, Banka VS, Helfant RH. Effects of isolated interventricular septal ischemia on global and segmental function of the canine right and left ventricle. Am Heart J. 1981;102:654-658.[Medline] [Order article via Infotrieve]
28. Santamore WP, Meier GD, Bove AA. Effects of hemodynamic alterations on wall motion in the canine right ventricle. Am J Physiol. 1979;236:H254-H262.
29. Kawai A, Kormos RL, Mandarino WA, Morita S, Deneault LG, Gasior TA, Armitage JM, Griffith BP. Differential regional function of the right ventricle with the use of a left ventricular assist device. ASAIO J. 1997;38:M676-M679.
30. Daly RC, Chandrasekaran K, Cavarocchi NC, Tajik AJ, Schaff HV. Ischemia of the interventricular septum: a mechanism of right ventricular failure during mechanical left ventricular assist. J Thorac Cardiovasc Surg. 1992;103:1186-1191.[Abstract]
31. Hori M, Yellin EL, Sonnenblick EH. Left ventricular diastolic suction as a mechanism of ventricular filling. Jpn Circ J. 1982;46:124-129.[Medline] [Order article via Infotrieve]
32.
Nikolic S, Yellin EL, Tamura K, Vetter H, Tamura T, Meisner JS, Frater RWM. Passive properties of canine left ventricle: diastolic stiffness and restoring forces. Circ Res. 1988;62:1210-1222.
33.
Ohtani M, Nikolic SD, Glantz SA. A new approach to in situ left ventricular volume clamping in dogs. Am J Physiol. 1991;261:H1335-H1343.
34. Sabbah HN, Anbe DT, Stein PD. Negative intraventricular diastolic pressure in patients with mitral stenosis: evidence of left ventricular diastolic suction. Am J Cardiol. 1980;45:562-566.[Medline] [Order article via Infotrieve]
35.
Udelson JE, Bacharach SL, Cannon RO, Bonow RO. Minimum left ventricular pressure during ß-adrenergic stimulation in human subjects. Circulation. 1990;82:1174-1182.
36. Weyman AE, Heger JJ, Kronik G, Wann LS, Dillon JC, Feigenbaum H. Mechanism of paradoxical early diastolic septal motion in patients with mitral stenosis: a cross-sectional echocardiographic study. Am J Cardiol. 1977;40:691-699.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
K. B. Leissner, F. Mahmood, J. R. Aragam, A. Amouzgar, and R. Ortega Catecholamine-Induced Cardiomyopathy and Pheochromocytoma Anesth. Analg., August 1, 2008; 107(2): 410 - 412. [Full Text] [PDF] |
||||
![]() |
N. Hristov, O. J. Liakopoulos, G. D. Buckberg, and G. Trummer Septal structure and function relationships parallel the left ventricular free wall ascending and descending segments of the helical heart Eur. J. Cardiothorac. Surg., April 1, 2006; 29(Suppl_1): S115 - S125. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Saleh, O. J. Liakopoulos, and G. D. Buckberg The septal motor of biventricular function Eur. J. Cardiothorac. Surg., April 1, 2006; 29(Suppl_1): S126 - S138. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. L. Gaynor, H. S. Maniar, S. M. Prasad, P. Steendijk, and M. R. Moon Reservoir and conduit function of right atrium: impact on right ventricular filling and cardiac output Am J Physiol Heart Circ Physiol, May 1, 2005; 288(5): H2140 - H2145. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. S. Poston and B. P. Griffith Heart Transplantation J Intensive Care Med, January 1, 2004; 19(1): 3 - 12. [Abstract] [PDF] |
||||
![]() |
H. S. Maniar, S. M. Prasad, S. L. Gaynor, C. M. Chu, P. Steendijk, and M. R. Moon Impact of pericardial restraint on right atrial mechanics during acute right ventricular pressure load Am J Physiol Heart Circ Physiol, January 1, 2003; 284(1): H350 - H357. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. P. Klima, M.-Y. Lee, J. L. Guerrero, P. J. LaRaia, R. A. Levine, and G. J. Vlahakes Determinants of maximal right ventricular function: Role of septal shift J. Thorac. Cardiovasc. Surg., January 1, 2002; 123(1): 72 - 80. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Ochiai, S. Morita, Y. Tanoue, Y. Kawachi, R. Tominaga, and H. Yasui Effects of amrinone, a phosphodiesterase inhibitor, on right ventricular/arterial coupling immediately after cardiac operations J. Thorac. Cardiovasc. Surg., July 1, 1998; 116(1): 139 - 147. [Abstract] [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1997 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |