(Circulation. 1996;94:286-297.)
© 1996 American Heart Association, Inc.
Articles |
the Cardiology Department, University of Essen (M.H., G.C., D.B., R.E.), and the Second Medical Clinic, Johannes Gutenberg University, Mainz (R.B., J.M.), Germany.
Correspondence to Michael Haude, MD, Cardiology Department, University of Essen, Hufelandstr 55, 45122 Essen, Federal Republic of Germany.
| Abstract |
|---|
|
|
|---|
Methods and Results In 25 patients with postangioplasty restenosis of the left anterior descending artery, ECG-gated digital subtraction coronary angiograms were recorded at baseline and during hyperemia (12 mg papaverine IC) before and after balloon predilatation (PTCA), after implantation of a Palmaz-Schatz stent, and after 6 months. Densitometric evaluation revealed different time and density parameters to calculate two definitions of myocardial perfusion reserve (MPR1 and MPR2) and maximum flow ratio (MaxFR). Poststenotic MPR1 increased from 1.57±0.14 to 2.59±0.86 after PTCA and to 3.10±0.41 after stenting, with 2.90±0.65 at follow-up (ANOVA, P<.05), while reference MPR1 remained unchanged at 3.10±0.40. Poststenotic MPR2 increased from 1.36±0.28 to 2.50±1.20 and to 3.40±0.58, respectively, with 3.20±0.92 at follow-up (ANOVA, P<.05), while reference MPR2 remained unchanged at 3.40±0.60. MaxFR was 2.13±0.53 after PTCA, 2.83±0.35 after stenting, and 2.73±0.58 at follow-up (ANOVA, P<.05). A good correlation was found between minimal stenotic luminal diameter and MPR1 or MPR2 (r=.87 and r=.94) and between luminal gain and MaxFR (r=.75). A negative correlation was measured between recoil and MPR1, MPR2, and MaxFR (r=-.80, r=-.86, and r=-.83). At follow-up, a steeper correlation was found between MPR and minimal stenosis diameter (MPR1: slope, 0.52 versus 0.91; MPR2: slope, 1.48 versus 1.95) and between MaxFR and net lumen gain (slope, 0.78 versus 1.27).
Conclusions Coronary stent implantation in patients with postangioplasty restenosis normalized poststenotic myocardial perfusion immediately as a result of a larger postprocedural lumen and a more pronounced inhibition of elastic recoil. After 6 months this benefit was sustained despite progressive lumen loss.
Key Words: perfusion angioplasty stents elasticity
| Introduction |
|---|
|
|
|---|
Densitometric evaluation of ECG-gated digital subtraction coronary angiograms derives different parameters that provide information about coronary blood flow and myocardial perfusion.12 13 14 15 16 17 Coronary flow reserve and MPR can be calculated as the ratio of densitometric results obtained from measurements during baseline and hyperemia.14 18 19 Extensive work was performed to improve this measurement technique and the reliability of the results by introducing on-line measurements, improvement of temporal resolution, reduction of background scatter, application of quality control to the time-density curves, and semiautomatic definition of ROIs.16 17 20 21 22
The purpose of this study was to determine the effect of adjunct coronary stenting in comparison with coronary balloon angioplasty on MPR in patients with single coronary artery disease of the left anterior descending artery by x-ray densitometric techniques.
| Methods |
|---|
|
|
|---|
|
Palmaz-Schatz Stent
All patients received implantation of a single second-generation Palmaz-Schatz stent 15 mm long mounted on a stent delivery system (SDS, Johnson & Johnson). The mode of implantation was described elsewhere.23 The balloon size of the SDS was identical to the balloon size used for predilatation of the target lesion. Implantation pressure was 6 atm for 30 seconds followed by a high-pressure stent dilatation at 12 atm for 30 seconds with the same noncompliant high-pressure balloon as used for predilatation of the target lesion.
Catheterization Procedure
All patients were pretreated with aspirin and nitrates. At the start of catheterization, a heparin bolus (10 000 IU), heparin infusion (1200 IU/h), and nitroglycerin infusion (3 mg/h) were administered.
Coronary angiography was performed via the femoral approach. After intracoronary injection of 0.2 mg nitroglycerin, angiograms of the right and left coronary arteries were recorded in different orthogonal projections to document coronary anatomy and the target lesion in the left anterior descending artery. Then, a 5-minute break followed to compensate for contrast mediuminduced changes in coronary flow. Afterward, baseline digital angiography of the left coronary artery was performed by ECG-triggered power injection of 6 mL contrast medium (Iopromid, Ultravist 370, Schering; 4 mL/s) with a temporal delay of four cycles after start of x-ray exposure via an 8F guiding catheter. Recording time was 20 seconds. After baseline recording and after another 5-minute break, digital angiography of the left coronary artery was repeated under hyperemic conditions induced by intracoronary injection of 12 mg papaverine. Then, PTCA of the target lesion was performed with a 3.0-mm noncompliant balloon catheter in 7 patients, a 3.5-mm balloon catheter in 14, and a 4.0-mm balloon catheter in 4. In each case, insufflation pressure was 12 atm for 30 seconds, which was tolerated by all patients. The angioplasty result was documented in the same orthogonal projections as before intervention, followed by an obligatory 5-minute break. Then, digital angiograms were recorded at baseline and during hyperemia as described previously. Subsequently, stents were implanted with balloon sizes similar to those during the angioplasty procedure, followed by stent dilatation as described earlier. Finally, control angiograms were recorded after intracoronary injection of 0.2 mg nitroglycerin in the same projections as before. With a temporal delay of 5 minutes after the last angiogram, digital angiograms were recorded at baseline and during hyperemia as described above.
During 6-month angiographic follow-up, cineangiograms of the target vessel were recorded in the same projections as during the interventional procedure. Additionally, digital angiograms were recorded at baseline and during hyperemia according to the previous protocol.
Quantitative Coronary Angiography
After 2.8-fold optical magnification, selected end-diastolic cine frames with illustration of the target lesion before and after balloon angioplasty, after stent implantation, and at 6-month follow-up were digitized for quantitative evaluation by the Cardiovascular Measurement System (CMS, Medis). The matrix size of this system is 512x512x8 bits. Analysis was based on the worst-view projection. The system provides automated vessel-edge detection and calculated the following parameters using the empty catheter for calibration24 : proximal, distal, and interpolated reference diameters; proximal, distal, and interpolated reference areas assuming a circular cross-sectional area; absolute and relative MLDs; absolute and relative minimal cross-sectional areas assuming a circular cross section; lesion length; plaque area and volume; and eccentricity.
On the basis of these results, the following derived parameters were calculated: acute gain (mm) as postprocedural MLD minus preprocedural MLD, acute gain (mm2) as postprocedural minimal cross-sectional area minus preprocedural minimal lumen area, relative gain index as the gain (mm) corrected for the reference diameter (mm), relative gain index as the gain (mm2) corrected for the reference cross-sectional area (mm2), late loss (mm) as postprocedural MLD (mm) minus MLD (mm) at follow-up, late loss (mm2) as postprocedural minimal cross-sectional area (mm2) minus minimal cross-sectional area (mm2) at follow-up, relative loss index as late loss (mm) corrected for the reference diameter (mm), relative loss index as late loss (mm2) corrected for the reference cross-sectional area (mm2), net gain (mm) as MLD (mm) at follow-up minus preprocedural MLD (mm), and net gain (mm2) as minimal cross-sectional area (mm2) at follow-up minus preprocedural minimal cross-sectional area (mm2).
Digital Image Processing System
Densitometric perfusion measurements were performed by a digital image processing system controlled by a DEC PDP 11/73 control and processing computer. Data selection, acquisition, digitization, subtraction mode, interactive definition of ROIs, calculation of quality control indexes, and the determination of the densitometric parameters were performed with this computer. Live images and parametric images could be displayed simultaneously on up to three video monitors. After real-time digitization, a total of 192 frames at a matrix size of 256x256x8 bits were stored in a 12-MB acquisition buffer (VTE Picturecom). Two 90-MB magnetic disks with high transfer rates allowed digitization of several angiographic scenes per patient for repetitive measurements during interventions.
Program for Real-time ECG-Triggered Frame Digitization
To improve temporal resolution, we developed a program for ECG-triggered digitization of six frames per heart cycle. Continuous digitization of the patient's ECG allows the calculation of the mean RR time interval of the 10 heartbeats preceding the start of digitization of cineangiograms, representing the phase of trigger acquisition.16 This mean RR time interval was divided into seven equal time segments starting at the R wave. At the beginning of the first six time segments, a trigger signal is coded, while the seventh time segment is excluded as a compensation for changes in heart rate after the injection of contrast medium. After real-time digitization of the angiograms is started, six heart phasegated frames per cycle are digitized according to the coded trigger signals, beginning at the time of each R wave. Digitization is followed by qualitative visual control using the subtraction mode to detect abnormalities caused by arrhythmia or changes in respiration.
Interactive Definition of ROIs
For the definition of ROIs, the user can define the outer contour via mouse-supported videometry to provide myocardium tangentially in the ROI and to exclude larger epicardial vessels within the ROI. After the number of ROIs and their width are defined, the program automatically computes the ROIs using a center-point method and displays them on a TV monitor superposed on a digital image depicting the coronary artery tree.20 25 In a similar way, corresponding background ROIs not superposed on myocardium were defined. In this study, we selected four poststenotic ROIs in the perfusion bed of the left anterior descending artery and four reference ROIs in the perfusion bed of the left circumflex artery.
Quality Control of the Six Heart PhaseGated Densitograms for Each ROI
For each of the myocardial and corresponding background ROIs, six densitograms were generated according to the six phases throughout the cardiac cycle. In a next step, the heart phasegated densitograms of the background ROIs were subtracted from the densitograms of the myocardial ROIs to compensate for changes in background noise throughout the acquisition time. Then the heart phasegated mean density value of the four cycles before contrast medium injection was subtracted from the subsequent values. Thereby, a total of six heart phasegated densitograms per myocardial ROI were generated from digital subtraction angiograms and after correction for background noise.
Each of these six heart phasegated densitograms was subsequently checked for its quality and received a quality index, Qp, which represents the product of the subindexes Q1, Q2, Q3: Qp(n)=Q1xQ2xQ3, with n=1, 2, 3, 4, 5, 6. The quality subindex Q1 compares the shape of the densitogram with a prototype densitogram, which was defined by the function t
txe1-t.20 25 After standardization on maximum intensity, two thresholds of tolerance were fixed on each side of the prototype densitogram, which were closer along the ascending part and wider along the descending part of the densitogram. The first threshold of tolerance includes the area of ±1.5 gray-scale levels along the prototype densitogram, which represents the mean value of scatter due to noise. Each measurement point of the actual densitogram within this first threshold of tolerance received a value q=1. The second threshold of tolerance represented an empirically generated area along the prototype densitogram.20 25 An actual densitogram value within this second threshold of tolerance received a quality value q with 0<q<1, with linear interpolation of the quality levels q between the borders of this second threshold of tolerance. The second lowest quality value q was assigned the quality level q1 and thereby allowed compensation for a single outlier.
Quality level Q2 checks for variation undulations in each of the six heart phasegated densitograms.20 25 A densitogram with a single discrete maximum, defined as an intensity value larger than the two following values, received a quality value of Q2=1. More than one local maximum in the densitogram resulted in a Q2 value according to the following equation:
![]() |
Quality level Q3 checks the level of intensity of each of the six heart phasegated densitograms per ROI and avoids densitograms with intensities next to the threshold of noise being assigned high quality.20 25 Densitograms within the threshold received a quality level Q3=0.
Each of the six heart phasegated densitograms per ROI then received a weighting coefficient (Wn) according to the equation
![]() |
Finally, these six adjusted densitograms per ROI were summarized to a single new densitogram that was subsequently used for calculation of the densitometric parameters.
Definition of Densitometric Parameters
The following parameters were derived from the quality controlled densitograms.
1. According to the suggestions of Vogel et al,13 14 local mean contrast medium appearance time (MCAT), was defined as the time from the start of contrast medium injection to the time when a certain threshold of intensity is passed. This time parameter replaced mean transit time.
2. Maximum intensity of the densitogram, which was used as a substitute for vascular volume, was as outlined by Vogel and coworkers.13 14
MPR was defined as MPR=(baseline MCAT/hyperemic MCAT)x(hyperemic Imax/baseline Imax), where Imax is maximum intensity.
3. Haude et al16 26 defined rise time (RT) as the time from the start of local contrast mediuminduced myocardial opacification to its maximum to replace mean transit time. Since maximum intensity was shown to poorly represent vascular volume and showed a poor reproducibility in human studies, they defined MPR to be solely related to the ratio of MPR
baseline RT/hyperemic RT.
4. Pijls and coworkers17 defined mean transit time (MTT) as
![]() |
Statistical Analysis
The individual data were used to calculate mean and SD. ANOVA was performed when normal distribution of data was given. Otherwise, the Kruskal-Wallis test was used to test for differences between means of several groups. If significant differences were found, the Bonferroni t test or Dunn's test was applied for multiple comparisons when appropriate. A value of P<.05 was considered statistically significant. Correlation coefficients, lines of regression, and SEEs were computed to document correlation between variables. In terms of reproducibility, we applied the technique of Bland and Altman27 to check for the agreement of repeated measurements by calculating the mean of the differences of data pairs and the corresponding SDs.
| Results |
|---|
|
|
|---|
Angiographic Results
A significant increase of minimal stenosis diameter was measured, from 0.99±0.49 to 2.04±0.33 mm after PTCA and to 3.15±0.36 mm after adjunct stenting, accounting for initial lumen gains of 1.04±0.22 and 1.11±0.32 mm, respectively (Table 2
). Acute recoil was 0.96±0.15 mm after PTCA versus 0.07±0.17 mm after adjunct stenting (P<.001). At follow-up, a significant late loss of 0.97±0.23 mm was measured, with a residual MLD of 2.18±0.57 mm. As a result, the net luminal gain accounted for 1.18±0.13 mm. A close linear correlation was found between acute luminal gain and late luminal loss, which improved when data were corrected for vessel size (Fig 1
). Four patients presented restenosis according to the dichotomous definition of a >50% diameter stenosis at follow-up (54%, 58%, 62%, and 64%, respectively). Table 2
summarizes the results of quantitative coronary angiography.
|
|
Densitometric Results
Densitograms of acceptable quality throughout the individual measurement sequence were obtained for 92 of 100 poststenotic ROIs in comparison to 99 of 100 reference ROIs. The average quality index of the six heart phasegated densitograms was 0.65±0.21 for poststenotic ROIs and 0.83±0.12 for reference ROIs (P<.05).
During the sequence of densitometric measurements, mean heart rate and blood pressure did not change significantly (Table 3
).
|
The results for the different densitometric parameters derived at baseline and during hyperemia before and after intervention, after adjunct stenting, and at follow-up are listed in Table 4
. Baseline densitometric results for poststenotic ROIs were similar, whereas hyperemic results changed significantly in comparison to preinterventional results.
|
Poststenotic MPR, defined as the ratio of baseline rise time divided by hyperemic rise time, increased from 1.57±0.14 before intervention to 2.59±0.86 after PTCA and to 3.10±0.41 after stenting (P<.05, Fig 2
). At follow-up, poststenotic MPR was 2.90±0.65 (P<.05 versus before intervention). MPR in reference ROIs did not change significantly throughout the measurement sequences.
|
Similar results were measured when MPR was defined as the ratio of baseline mean contrast medium appear-ance time divided by hyperemic mean contrast medium appearance time multiplied by the ratio of hyperemic maximal intensity divided by baseline maximal intensity. Poststenotic MPR increased from 1.36±0.28 before intervention to 2.50±1.2 after PTCA and to 3.40±0.87 after stenting, with a nonsignificant drop to 3.20±0.92 at follow-up (Fig 2
). Again, MPR did not change significantly for reference ROIs throughout the measurements (Fig 2
).
Poststenotic maximal flow ratio, defined as the ratio of hyperemic mean transit time before and after intervention, was 2.13±0.53 after PTCA, 2.83±0.35 after stenting, and 2.73±0.58 at follow-up (Fig 2
). Results for reference ROIs did not change significantly (Fig 2
).
Reproducibility of Densitometric Measurements
A good reproducibility was found for both definitions of MPR and for maximal flow ratio (Fig 3
). Accuracy and precision were better for measurements after stenting compared with measurements before and after PTCA and for measurements in reference ROIs compared with poststenotic ROIs (Table 5
).
|
|
Correlation Between Densitometric Perfusion Measurements and Results of Quantitative Coronary Angiography
A linear correlation was documented between MPR and minimal stenosis diameter independent of the definition of MPR and between maximal flow ratio and luminal gain measured before and after PTCA and after stenting (Fig 4
). A curvilinear correlation was found when results of densitometric perfusion measurements were compared with area stenosis measurements (Fig 4
). A negative correlation was documented between MPR or maximal flow ratio and recoil (Fig 5
).
|
|
At follow-up, the close correlation between residual stenosis dimensions and MPR or maximal flow ratio persisted, but the slopes of the regression lines were steeper in all measurements compared with the acute results (Fig 6
).
|
| Discussion |
|---|
|
|
|---|
The results of this study document that poststenotic MPR can be restored immediately to normal reference values of >3.0 on an average only after stent implantation but not after balloon predilatation with the same balloon size in carefully selected patients with a single postangioplasty coronary restenosis excluding other factors that affect vascular capacity, such as angiographically discernible collaterals, myocardial infarction of the target area, hypertension, diabetes, or myocardial hypertrophy. This supports previously published data obtained in a small number of unselected patients with stent implantation by use of densitometric or intracoronary Doppler techniques.29 30 Since baseline perfusion conditions remained constant throughout the measurement sequences, as could be documented by the different densitometric parameters, the subsequent improvement of poststenotic MPR after balloon angioplasty and after stenting is induced primarily by an improvement of hyperemic perfusion. The intraindividual comparison between poststenotic and reference MPRs at each step is of major importance for the interpretation of the functional improvement after balloon angioplasty and adjunct stenting. Since diffuse, nonstenotic atherosclerosis of the reference vessel may be present even in patients with single coronary artery disease, which could be documented angiographically and even more clearly by intravascular ultrasound,31 this may have an impact on the intraindividual target goal in improving poststenotic MPR. Therefore, we selected patients with a single stenosis of the left anterior descending coronary artery, allowing for simultaneous estimation of the reference MPR in the perfusion bed of the left circumflex artery during the same injection.
Metabolic (lactate), humoral (endothelin, thromboxane), or myogenic (vasodilatation, spasm) factors have been shown to affect flow conditions in complicated angioplasty procedures with prolonged ischemia, but it seems unlikely that they affect the additional improvement of MPR after stenting compared with balloon angioplasty in the elective and uncomplicated procedures reported here, although they cannot be ruled out completely.32 33 34 Since the study protocol fixed the duration of balloon insufflation to 30 seconds with sufficient time to recover from ischemia, a preconditioning effect during repeat balloon inflations, as is known from longer insufflations, cannot be assumed as an explanation for the improvement of MPR after stenting compared with balloon angio-plasty.35 36 37 Hemodynamic parameters such as blood pressure and heart rate were within the range of 10% changes throughout the individual densitometric measurements.
There was a close relationship between angiographically measured stenosis dimensions and related poststenotic MPR, supporting the important role of residual stenosis in the functional result. The measured stenosis dimensions before and after balloon predilatation and after adjunct stent implantation and postinterventional elastic recoil were comparable to those reported by others.8 38 39 40
Long-term follow-up documented a substantial lumen loss at the stented vessel site, but MPR of the supplied area remained in the range of the normal reference MPR, as was shown by a steeper relationship between morpho-metric and functional results, suggesting some long-term recovery of microvascular capacity. Similar results were reported by Zijlstra et al,28 who documented, in contrast to the immediate postangioplasty result, a long-term recovery of poststenotic MPR to the normal range 5 months after the procedure. They found larger epicardial dimensions in the target vessel at follow-up compared with the immediate postangioplasty result in 10 of 15 patients, which was interpreted as a contribution to the functional improvement. Johnson et al41 reported a late increase in cross-sectional obstruction area after balloon angioplasty in one third of their patients. Nevertheless, in our study we could not document a late improvement of residual stenosis after stenting. Therefore, recovery of the microvascular capacity contributes more to the persisting normalization of MPR after 6 months.
Densitometric evaluation of digital subtraction angiocardiograms provides information about coronary flow by analyzing the contrast medium propagation through epicardial vessels, the capillary bed, and the coronary venous system and by generating time-density curves similar to indicator dilution curves.12 13 14 15 16 17 It was shown that intra-coronary injections of papaverine or adenosine induced maximum hyperemia most reliably and, because of their shorter half-life compared with dipyridamole, allowed for a repetition of measurements during cardiac catheterization.42 43 Maximum dilatation of the vascular bed is achieved
25 to 60 seconds after intracoronary injection of 8 to 12 mg papaverine. This time period is sufficient to provide steady-state hyperemic conditions during the acquisition of the time-density curves.16 17
Alternative techniques that provide information about the functional effect of coronary interventions on poststenotic flow or perfusion include visual estimation of contrast medium propagation according to the TIMI classification,44 which is popular but not very precise. Coronary sinus thermodilution techniques were used to estimate coronary blood flow45 but never gained wide acceptance in clinical practice, primarily because of difficulties encountered with instrumentation of the coronary sinus. Second, measurements reflect flow mainly within the myocardial territory supplied by the left anterior descending artery. Doppler techniques via intracoronary catheters or more recently available Doppler wires allow the recording of epicardial flow velocity.46 47 48 Kirkeeide et al49 and Gould et al50 defined stenosis flow reserve as a single measure of severity derived from all integrated stenosis geometry by fluid dynamic equations using quantitative coronary angiography. Nevertheless, these techniques do not reflect myocardial perfusion within the precapillaries and capillaries.
Limitations of the Study
In general, the results of this study are obtained from a carefully selected patient population with a single postangioplasty restenosis of the left anterior descending artery, excluding other factors that could influence vascular capacity as described before. Therefore, these results should not be extrapolated to an unselected patient population, as was outlined by others, who documented an improvement in poststenotic MPR after balloon angioplasty in selected but not in unselected patients.51 Assuming the exclusion of other factors influencing microvascular capacity, the results obtained in patients with postangioplasty restenosis are likely to be expanded to individuals undergoing primary angioplasty and stenting. This is supported by quantitative angiographic studies, which have shown similar results on acute gain and elastic recoil after balloon angioplasty for primary and restenotic lesions.8 40 An average 30% loss of the maximum achievable diameter and an average 50% loss of the maximum achievable cross-sectional area immediately after balloon deflation is reported for both primary and restenotic lesions.8 39 40
Furthermore, this study is not intended to directly compare balloon angioplasty and stenting with respect to their impact on long-term myocardial perfusion reserve. The importance of early improvement of myocardial perfusion reserve after stenting to normal reference values on long-term outcome has not yet been established in comparison to balloon angioplasty. Therefore, a randomized trial is mandatory comparing the results on myocardial perfusion reserve in two groups of patients with balloon angioplasty or predilatation and stent implantation.
Densitometric Measurements
Densitometric evaluation of digital subtraction coronary angiograms requires excellent angiographic recordings free of motion artifacts and arrhythmia. This could be obtained in the majority of our patients after careful instruction to withhold breath in deep inspiration for about 20 seconds. In two patients, extrasystoles throughout the recordings required repetition of the digital subtraction coronary angiograms. Atrial fibrillation will not allow these kinds of measurements.
Contrast medium was used as the indicator, an agent that itself has hyperemic properties that might influence the time-density curves, especially during the second part representing contrast medium washout. Furthermore, the requirement of complete substitution of blood by the contrast medium bolus cannot be fulfilled in every situation under clinical conditions in the catheterization laboratory. Nevertheless, reproducibility measurements, performed in a subset of patients under baseline and hyperemic conditions before and after balloon angioplasty and after stenting showed good results. A general limitation of the applied densitometric approach is the poor temporal resolution of the time-density curves resulting from the R wavetriggered image acquisition. In vivo studies pointed out that up to 40% changes of flow reserve using the technique proposed by Vogel et al13 14 account for inaccuracy related to the poor temporal resolution.19 Furthermore, the quality of the densitograms is rarely taken into account. Pijls et al17 fitted the time-density curve to a gamma function and calculated a relative fitting error, which should be <10% to accept the fit and to calculate mean transit time. In our series, we improved temporal resolution by generating not only one but six time-density curves per ROI throughout the cardiac cycle. In addition, we applied sophisticated quality controls to these time-density curves, which have been validated previously.20 25 Thereby, we were able to improve accuracy and precision of the densitometric parameters and of the derived MPR and maximum flow ratio.20 25
There is an ongoing discussion as to which densitometric parameter best represents mean transit time and vascular volume. Several authors pointed out that maximum intensity of the time-density curves does not precisely reflect vascular volume.17 19 25 Therefore, Pijls et al17 calculated maximum flow ratio by comparing hyperemic mean transit time before and after coronary interventions, ruling out the problem of differences in vascular volume at baseline and during hyperemia, since vascular volume is uniform during maximum hyperemia. Therefore, we applied two definitions of MPR. One was defined by Vogel et al,13 14 applying mean contrast medium appearance time and maximum intensity to substitute for mean transit time and vascular volume. Another description of MPR, defined by our group, uses only the ratio of baseline and hyperemic rise time and does not use maximum intensity for reasons listed above.16 Rise time is reported to have a better reproducibility and representation of epicardial electromagnetic flow compared with mean contrast medium appearance time.17 Furthermore, we applied maximum flow ratio as defined by Pijls et al.17 All these definitions came up with the same results, that adjunct stenting improves MPR and maximum flow ratio in comparison to balloon angioplasty.
Quantitative Coronary Angiography
Despite use of an extensively validated quantitative coronary angiography system, there are some general problems with the detection of luminal dimensions from subsequent angiograms before and after balloon angioplasty, after stenting, and at late follow-up. Especially in the postangioplasty setting, a substantial overestimation of the vascular lumen can occur, since postangioplasty tears, slits, and dissections can be filled by contrast medium. Intravascular ultrasound imaging provides a better delineation in this situation. Changes in epicardial vasomotor tone were eliminated by repeat intracoronary injection of nitroglycerin. As a result, reference vessel dimensions were similar throughout the measurements. Nevertheless, quantitative angiographic results of this study are comparable to those of other studies.6 7 8 23 39 In particular, incomplete dilatation as the reason for the postangioplasty angiographic and subsequent functional result can be ruled out, since the average acute gain of 1.04 mm after balloon angioplasty was comparable to that reported for PTCA arms of large-scale, randomized trials, with an acute gain ranging from 0.79 to 1.24 mm.6 7 52 53 54 55
In conclusion, coronary stenting is able to restore MPR within the normal range immediately compared with predilatation using the same balloon size and inflation pressure in selected patients with a single postangioplasty restenosis of the left anterior descending artery. A pronounced inhibition of elastic recoil resulting in a smaller residual stenosis after stenting seems to be primarily responsible. At 6 months, this improvement was sustained despite progressive luminal renarrowing at the stented vessel site, suggesting recovery of microvascular capacity.
| Acknowledgment |
|---|
|
|
|---|
| Selected Abbreviations and Acronyms |
|---|
|
Received October 19, 1995; revision received December 27, 1995; accepted January 4, 1996.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
K. Leineweber, D. Bose, M. Vogelsang, M. Haude, R. Erbel, and G. Heusch Intense Vasoconstriction in Response to Aspirate From Stented Saphenous Vein Aortocoronary Bypass Grafts J. Am. Coll. Cardiol., March 7, 2006; 47(5): 981 - 986. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. E. Langerak, H. W. Vliegen, J. W. Jukema, A. H. Zwinderman, H. J. Lamb, A. de Roos, and E. E. van der Wall Vein Graft Function Improvement after Percutaneous Intervention: Evaluation with MR Flow Mapping Radiology, September 1, 2003; 228(3): 834 - 841. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Gregorini, J. Marco, B. Farah, M. Bernies, C. Palombo, M. Kozakova, I. M. Bossi, B. Cassagneau, J. Fajadet, C. Di Mario, et al. Effects of Selective {alpha}1- and {alpha}2-Adrenergic Blockade on Coronary Flow Reserve After Coronary Stenting Circulation, |