Donate Help Contact The AHA Sign In Home
American Heart Association
Circulation
Search: search_blue_button Advanced Search
Circulation. 1996;93:683-690

This Article
Right arrow Abstract Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Nishioka, T.
Right arrow Articles by Siegel, R. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Nishioka, T.
Right arrow Articles by Siegel, R. J.

(Circulation. 1996;93:683-690.)
© 1996 American Heart Association, Inc.


Articles

Absence of Focal Compensatory Enlargement or Constriction in Diseased Human Coronary Saphenous Vein Bypass Grafts

An Intravascular Ultrasound Study

Presented in part at the 67th Scientific Sessions of the American Heart Association, Dallas, Tex, November 14-17, 1994.

Toshihiko Nishioka, MD; Huai Luo, MD; Hans Berglund, MD; Neal L. Eigler, MD; Chong-Jin Kim, MD; Steven W. Tabak, MD; Robert J. Siegel, MD

From the Division of Cardiology, Cedars-Sinai Medical Center, Los Angeles, Calif.

Correspondence to Robert J. Siegel, MD, Division of Cardiology, Room 5335, Cedars-Sinai Medical Center, 8700 Beverly Blvd, Los Angeles, CA 90048.


*    Abstract
up arrowTop
*Abstract
down arrowIntroduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Background No in vivo data are available on the occurrence of compensatory enlargement or vessel constriction in diseased human coronary saphenous vein bypass grafts (SVBGs). The aim of this intravascular ultrasound (IVUS) study was to examine to what extent lumen reduction is accompanied by (1) vessel wall thickening and (2) arterial wall constriction in SVBGs.

Methods and Results We used IVUS to examine 43 SVBGs from 42 patients (32 men, 10 women; mean age, 72±5 years) 8 to 23 (11±4) years after SVBG. IVUS images were obtained with a 3.5F monorail ultrasound catheter with a 30-MHz frequency and were analyzed at the lesion site, the reference site, and an intermediate site. The lumen area was significantly (P<.01) decreased; the vessel wall area (SVBG cross-sectional area minus lumen area) and the plaque area (area within the external elastic lamina minus lumen area) were significantly (P<.01) increased from the reference site through the lesion site. However, SVBG cross-sectional area was the same at these three sites (24.0±8.1 versus 24.4±8.6 versus 24.5±8.6 mm2, P=NS), and the external elastic lamina area was also quite constant in each vessel (17.8±6.0 versus 17.7±6.4 versus 17.6±6.2 mm2, P=NS).

Conclusions These in vivo IVUS data from human coronary SVBGs demonstrate that (1) no focal compensatory enlargement or vessel constriction occurred in stenotic segments compared with the reference segments and that (2) the absence of focal compensatory enlargement appears to be a potentially important factor in the progression of stenoses in coronary SVBGs.


Key Words: veins • bypass • grafting • ultrasonics • remodeling


*    Introduction
up arrowTop
up arrowAbstract
*Introduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Compensatory enlargement of human atherosclerotic vessels was first described by Glagov et al1 in a postmortem histopathological study of left main coronary arteries. They showed that human coronary arteries enlarge during the development of arterial atherosclerotic plaque and that the lumen area is preserved until the plaque development exceeds the compensatory mechanisms of the blood vessel. This finding has been validated in vivo in studies of human coronary arteries by use of high-frequency epicardial echocardiography2 and IVUS imaging3 4 and reconfirmed by the pathological study of the left anterior descending coronary arteries from humans and nonhuman primates.5 Compensatory enlargement is also observed in human carotid arteries by high-resolution duplex scanning6 and in superficial femoral arteries studied by IVUS.7 These observations suggest that compensatory enlargement is a common phenomenon in all human atherosclerotic arteries. Moreover, a recent study8 of human femoral arteries using IVUS imaging and histopathological techniques demonstrated that arterial wall constriction or focal failure of compensatory enlargement may be a paradoxical mechanism for the development of severe arterial luminal narrowing. However, no data exist as to whether compensatory enlargement or vessel constriction occurs in stenotic SVBGs.

Accordingly, the aim of this IVUS study was to examine to what extent lumen reduction is accompanied by (1) vessel wall thickening and (2) failed or incomplete compensatory enlargement or constriction of the stenotic vessel segment in SVBG.


*    Methods
up arrowTop
up arrowAbstract
up arrowIntroduction
*Methods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Patients and Vessels Studied
Forty-eight SVBGs from 47 consecutive patients (36 men, 11 women; mean age, 71±6 years) who had not undergone previous catheter intervention were studied by IVUS 8 to 23 years (mean, 12 years) after coronary SVBG. Informed consent was obtained from all patients before the IVUS procedure. Excluded from the study were 1 vessel in which we failed to introduce the imaging catheter into the SVBG and 4 vessels in which IVUS images were suboptimal for quantitative measurements because of heavy intimal calcification of the SVBGs or a technical problem with the IVUS system. In total, images from 43 SVBGs from 42 patients were included and analyzed in this study.

IVUS System and Imaging Procedure
To ensure reproducibility of measurements, only one IVUS system was used in this study. The IVUS imaging system consisted of an imaging catheter (Sonicath, Boston Scientific Corp) and a SONOS Intravascular System imaging console (Hewlett-Packard). The imaging catheter has a 30-MHz single piezoelectric crystal transducer mechanically rotating at 1800 rpm within a 3.5F monorail over-the-wire catheter sheath.

The right or left femoral artery was punctured by use of the Seldinger technique, an 8F or 9F arterial introducer sheath was advanced retrogradely over a guide wire, and the sheath was placed in the femoral artery. After the completion of angiography of native coronary arteries and SVBGs, the imaging catheter was introduced into the SVBG through an 8F to 9F coronary guiding catheter over a 0.014- or 0.018-in guide wire. To obtain maximum vasodilation, 100 to 200 µg of nitroglycerin was administered into the SVBG before or during the IVUS catheter imaging. After the imaging catheter was advanced across the lesion to the distal portion of the vessel under fluoroscopic guidance, IVUS imaging was performed during the slow pullback (1 mm/s) of the imaging catheter. The two-dimensional images of the SVBG were displayed on a SONOS Intravascular System imaging console and recorded on a 0.5-in Super-VHS videotape for subsequent playback, review, and quantitative analysis.

Image Analysis
IVUS images were analyzed off-line with a SONOS Intravascular System. In the SVBG examined, the ostial portion and the site of anastomosis to the native coronary artery were excluded from the measurements to eliminate any distortion due to sutures at these sites.

In each SVBG, with digital angiographic images as a road map, three sites were selected by IVUS for quantitative analysis: the lesion site with the smallest lumen area, the reference site with the largest lumen area, and an intermediate site with a lumen area of intermediate size between the lesion and reference sites.

For these three sites in each SVBG, vessel lumen area was measured by planimetry. To assess the vessel lumen area (in square millimeters), the lumen-intimal border was traced and the area within this border was measured with a planimeter. Contrast medium was injected to enhance the ultrasound definition of the lumen in cases in which the lumen-intimal border was ambiguous.

In 36 vessels of 43 SVBGs, the entire vessel cross-sectional area (in square millimeters) was clearly delineated and measured by tracing the outer border of the whole vessel, as shown in Fig 1Down. The vessel wall area (in square millimeters) was calculated by subtracting the lumen area from the vessel cross-sectional area. Lumen area ratio, vessel wall area ratio, and vessel cross-sectional area ratio were also defined and calculated according to the following formulas numbered (1), (2), and (3) compared with the vessel cross-sectional area at the reference site: (1) lumen area ratio=lumen area at the lesion, intermediate, or reference site/vessel cross-sectional area at the reference site; (2) vessel wall area ratio=vessel wall area at the lesion, intermediate, or reference site/vessel cross-sectional area at the reference site; and (3) vessel cross-sectional area ratio=vessel cross-sectional area at the lesion, intermediate, or reference site/vessel cross-sectional area at the reference site.



View larger version (92K):
[in this window]
[in a new window]
 
Figure 1. IVUS images of the SVBG to the left anterior descending coronary artery 9 years after CABG. The patient had recurrent exertional angina pectoris, and a significant stenosis was observed at the midportion of this vein graft by both angiography and IVUS imaging. A, B, and C, IVUS images with delineation of the lumen-intimal borders (white dotted lines) and the outer borders of the whole vessel (white dotted lines with arrowheads). D, E, and F, Same IVUS images as in A, B, and C, respectively, but without the dotted lines delineating the vessel borders. A, At the reference site of the SVBG, the lumen area was 16.6 mm2 and the vessel cross-sectional area was 24.3 mm2. B, At the intermediate site, the lumen area was 10.5 mm2 and the vessel cross-sectional area was 26.5 mm2. C, At the lesion site, the lumen area was 6.3 mm2 and the vessel cross-sectional area was 26.0 mm2. Despite the progression of disease severity from A to C, the vessel cross-sectional areas were similar, and no compensatory enlargement or constriction of the SVBG was observed.

In 25 vessels of 43 SVBGs, a sonolucent zone that has been reported to represent media9 10 was clearly visualized by IVUS, as shown in Fig 2Down. The EEL of the vessel was defined as the outer border of the sonolucent zone, and the area within the EEL (the EEL area) was measured by planimetry. The plaque area (in square millimeters) was defined and calculated by subtracting the lumen area from the EEL area. Lumen area ratio, plaque area ratio, and EEL area ratio were defined and calculated according to the following formulas numbered (4), (5), and (6) compared with the EEL area at the reference site: (4) lumen area ratio=lumen area at the lesion, intermediate, or reference site/EEL area at the reference site; (5) plaque area ratio=plaque area at the lesion, intermediate, or reference site/EEL area at the reference site; and (6) EEL area ratio=EEL area at the lesion, intermediate, or reference site/EEL area at the reference site.



View larger version (42K):
[in this window]
[in a new window]
 
Figure 2. IVUS images of the SVBG to the right coronary artery 11 years after CABG. The patient had recurrent exertional angina pectoris, and a significant stenosis was observed at the proximal portion of this vein graft by both angiography and IVUS imaging. The white dotted lines show the lumen-intimal borders, and the white dotted lines accompanied by white arrowheads indicate the EEL. A, At the reference site of the SVBG, the lumen area was 6.6 mm2 and the external EEL area was 12.3 mm2. B, At the intermediate site, the lumen area was 4.2 mm2 and the EEL area was 13.2 mm2. C, At the lesion site, the lumen area was 2.8 mm2 and the EEL area was 12.9 mm2. Despite the progression of disease severity from A to C, the EEL areas were similar, and no compensatory enlargement or constriction of the SVBG was observed.

In Vitro Saphenous Vein Study
In 18 SVBGs in which only vessel cross-sectional area was used as a marker of vessel remodeling, the plaque area could not be calculated, and consequently the disease severity at the reference site of SVBG was uncertain. Therefore, 7 fresh saphenous veins were harvested from 5 patients undergoing CABG, imaged by IVUS under physiological pressures, and compared with the measurements at the reference sites of SVBGs. The extra saphenous vein segments that were prepared for CABG with all their side branches tied were obtained and stored in normal saline solution at 3°C. The IVUS images were obtained within 6 hours of harvesting. A three-way stopcock was attached to one end of the vessel to close the lumen, and a Touey-Borst connector was attached to the other end. The saline solution, attached to a pressure bag, was connected to a Touey-Borst connector. The vessel was distended with a nonpulsatile pressure of 100 mm Hg.11 All in vitro IVUS imaging was performed in a saline bath. The IVUS system used in this in vitro study was exactly the same system as used for the in vivo study. The IVUS catheter was introduced through the Touey-Borst connector, and two-dimensional cross-sectional images of the vessel were recorded on videotape during the slow pullback (1 mm/s) of the imaging catheter. The IVUS images were analyzed with the same imaging console as the in vivo study, and the same kinds of measurements and calculated variables were derived for 35 segments selected from 7 veins in random fashion.

In fresh saphenous veins, the lumen area and the entire vessel cross-sectional area were measured without difficulty in every case, and these measurements and the calculated data of fresh saphenous veins were compared with those of 36 SVBGs in which the vessel cross-sectional area was measured.

Reproducibility Study
Intraobserver and interobserver variabilities of the measurements of the lumen area and the vessel cross-sectional area were calculated on 10 segments of SVBG selected at random from 36 SVBGs in which the vessel cross-sectional area was measured. Intraobserver and interobserver variabilities of the EEL area were calculated on 10 segments of SVBG selected at random from 25 SVBGs in which the EEL area was measured. Two independent observers measured the lumen area, the vessel cross-sectional area, and the EEL area of these segments on two occasions separated by a minimum of 7 days. These data were analyzed, and the mean difference between two observations and the range of the differences were derived.

Statistical Analysis
All data were expressed as mean±SD. Measured and calculated areas at the lesion site, the reference site, and the intermediate site were compared by repeated-measures ANOVA with the Student-Newman-Keuls test as the post hoc test. For comparison of data between the reference segments in SVBGs and the fresh saphenous veins, an unpaired Student's t test was performed. A value of P<.05 was considered statistically significant.


*    Results
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
*Results
down arrowDiscussion
down arrowReferences
 
The IVUS studies of in vivo SVBGs were completed without any vascular complications. Of 43 SVBGs in which quantitative analysis was performed, reproducible measurements could be made for the vessel cross-sectional area in 36 SVBGs, for the EEL area in 25 SVBGs, and for both the vessel cross-sectional area and the EEL area in 18 SVBGs.

Reproducibility
For the measurement of the lumen area, the mean difference between two observations by the same observer was 3.1%, with a range of 0% to 9.0%. Between two observers, the mean difference was 3.8%, with a range of 1.0% to 5.6%.

For the measurement of the vessel cross-sectional area, intraobserver variability was calculated, with a mean difference of 1.6% and a range of 0% to 2.7%. Interobserver variability was also determined, with a mean difference of 2.7% and a range of 0.4% to 5.7%.

For the measurement of the EEL area, the mean difference between two observations by the same observer was 1.7%, with a range of 0% to 3.2%. Between two observers, the mean difference was 2.7%, with a range of 0.8% to 4.5%.

Fresh Saphenous Veins and SVBGs in Which the Vessel Cross-sectional Area Was Measured
The measured and calculated data of in vitro fresh saphenous veins are summarized in Table 1Down and Fig 3Down. The lumen-intimal border and the outer border of the entire vessel were clearly delineated without difficulty in every case, but the sonolucent zone was not observed or was ambiguous in most cases. Consequently, only lumen area and vessel cross-sectional area were measured, and lumen area ratio, vessel wall area ratio, and vessel cross-sectional area ratio were calculated. Even fresh saphenous veins had a substantial vessel wall area of 7.4±2.1 mm2, which occupied {approx}30% of the total vessel cross-sectional area.


View this table:
[in this window]
[in a new window]
 
Table 1. Summary of IVUS Measurements and Calculated Data on In Vitro Fresh Saphenous Veins and In Vivo SVBGs in Which the Entire Vessel CSA Was Measured



View larger version (37K):
[in this window]
[in a new window]
 
Figure 3. Bar graph shows comparison of lumen area, vessel wall area, and vessel cross-sectional area in fresh saphenous veins and SVBGs at the reference site, intermediate site, and lesion site. The IVUS measurements at the reference sites of the SVBGs were similar to those of the fresh saphenous veins. In SVBGs, lumen area gradually decreased and vessel wall area increased as the lumen area became smaller. However, SVBG cross-sectional area was not different at the lesion site, intermediate site, and reference site. CSA indicates cross-sectional area. *P<.01, **P<.001.

Fig 1Up shows IVUS images of one of the 36 SVBGs in which the outer border of the entire vessel was clearly visualized. In 53% (19/36) of SVBGs, as shown in Fig 1Up, the vessel was surrounded by minimal tissue or separated from the surrounding tissue, and the outer border of the entire vessel was traced without difficulty. Even when the whole circumference of the SVBG was surrounded by connective tissue, the outer border was differentiated from the adjacent tissue under careful observation of the different movements and the different echo intensities of the vessel itself and the surrounding tissue. Despite the disease progression from Fig 1AUp through 1C, the vessel cross-sectional areas were almost identical.

The measured and calculated data of in vivo SVBGs in which the vessel cross-sectional area was measured are summarized with those of fresh saphenous veins in Table 1Up and Fig 3Up. All values at the reference sites of the SVBGs were similar to those of the fresh saphenous veins. In the SVBGs, the lumen area was greatest at the reference site (15.8±5.0 mm2), moderate at the intermediate site (9.4±3.4 mm2), and least at the lesion site (5.5±2.2 mm2), as expected from the definition of these three sites. The vessel wall area was greatest at the lesion site (19.0±7.4 mm2), moderate at the intermediate site (15.0±6.2 mm2), and least at the reference site (8.4±3.9 mm2). All differences of measurements among these sites were statistically significant (P<.01). However, the SVBG cross-sectional areas were quite constant in each vessel, measuring 24.0±8.1 mm2 at the reference site, 24.4±8.6 mm2 at the intermediate site, and 24.5±8.6 mm2 at the lesion site.

Fig 4ADown and 4BDown demonstrates the relation between lumen area ratio and vessel cross-sectional area ratio (A) and between lumen area ratio and vessel wall area ratio (B). In Fig 4ADown, if focal compensatory enlargement of SVBG occurs at the stenotic segments, the vessel cross-sectional area ratio should become larger as the vessel lumen area ratio decreases, and if failed or incomplete compensatory enlargement, or constriction of the stenotic vessel segment occurs, the vessel cross-sectional area ratio should decrease, accompanying the lumen area ratio reduction. However, the vessel cross-sectional area ratio was near one and constant for the intermediate and stenotic lesion sites compared with the reference sites. Moreover, the variability of the SVBG cross-sectional area (mean±SD, 1.009±0.052; range, 0.825 to 1.136) shown in this graph was almost identical to that of the fresh saphenous veins (mean±SD, 1.013±0.076; range, 0.868 to 1.240). In Fig 4BDown, the plotted points of the vessel wall area ratios of the stenotic vessel segments (intermediate and lesion sites) indicate that lumen area reduction is due to the increase in vessel wall area.



View larger version (17K):
[in this window]
[in a new window]
 
Figure 4. Two graphs show the relation between lumen area ratio and vessel cross-sectional area ratio (A) and between lumen area ratio and vessel wall area ratio (B). A, Vessel cross-sectional area ratio was near one and constant for the intermediate and stenotic lesion sites compared with the reference sites. B, Vessel wall area ratios at the stenotic vessel segments were almost on the dotted line, which indicates that lumen area reduction is due to the increase in vessel wall area.

SVBGs in Which the EEL Area Was Measured
Fig 2Up shows IVUS images of one of the 25 SVBGs in which the EEL area was measured. No focal compensatory enlargement was observed at the intermediate site and the stenotic lesion site compared with the reference site, despite the disease progression from Fig 2AUp through 2C.

Table 2Down and Fig 5Down summarize the measured and calculated data of SVBGs in which the EEL area was measured. The lumen area significantly (P<.001) decreased and the plaque area significantly (P<.01) increased from the reference site through the lesion site. However, the EEL areas were quite constant in each vessel, measuring 17.8±6.0 mm2 at the reference site, 17.7±6.4 mm2 at the intermediate site, and 17.6±6.2 mm2 at the lesion site.


View this table:
[in this window]
[in a new window]
 
Table 2. Summary of IVUS Measurements and Calculated Data on SVBGs In Vivo in Which the EEL Area Was Measured



View larger version (31K):
[in this window]
[in a new window]
 
Figure 5. Bar graph shows comparison of lumen area, plaque area, and EEL area in SVBGs at the reference site, intermediate site, and lesion site. In SVBGs, lumen area gradually decreased and plaque area increased as the lumen area became smaller. However, SVBG EEL area was similar at the lesion site, intermediate site, and reference site. *P<.01, **P<.001.

Fig 6ADown and 6BDown demonstrates the relation between lumen area ratio and EEL area ratio (A) and between lumen area ratio and plaque area ratio (B). In Fig 6ADown, the EEL area ratio was near one and constant (mean±SD, 1.014±0.075; range, 0.863 to 1.194) for the intermediate and stenotic lesion sites compared with the reference sites. In Fig 6BDown, the plotted points of plaque area ratios of the stenotic vessel segments (intermediate and lesion sites) indicate that lumen area reduction is due to the increase in plaque area.



View larger version (16K):
[in this window]
[in a new window]
 
Figure 6. Two graphs show the relation between lumen area ratio and EEL area ratio (A) and between lumen area ratio and plaque area ratio (B). A, EEL area ratio was near one and constant for the intermediate and stenotic lesion sites compared with the reference sites. B, Plaque area ratios at the stenotic vessel segments were almost on the dotted line, which indicates that lumen area reduction is due to the increase in plaque area.

SVBGs in Which Both the EEL Area and the Vessel Cross-sectional Area Were Measured
The measured and calculated data of 18 SVBGs in which both the EEL area and the vessel cross-sectional area were measured are summarized in Table 3Down. The lumen area was significantly (P<.001) decreased and the plaque area and the EEL area were significantly (P<.01) increased from the reference site through the lesion site. However, the EEL areas and the vessel cross-sectional areas were quite constant, and the EEL area ratio and the vessel cross-sectional area ratio were near one.


View this table:
[in this window]
[in a new window]
 
Table 3. Summary of IVUS Measurements and Calculated Data in SVBGs in Which Both the EEL Area and the Vessel CSA of the Vessels Were Measured


*    Discussion
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
*Discussion
down arrowReferences
 
This is the first study to demonstrate an absence of focal compensatory enlargement or vessel constriction (shrinkage) in diseased human coronary SVBGs. Our in vivo IVUS study of three different segments in SVBGs using the total vessel cross-sectional area or the area within the EEL as an index of vessel remodeling found that no focal compensatory enlargement or vessel constriction occurs in the stenotic segments compared with the reference segments with minimal disease. The SVBG behaves like a prosthetic conduit without any difference in the vessel cross-sectional area and the EEL area among the lesion site, intermediate site, and reference site. These unique findings in SVBGs are different from previous reports demonstrating the presence of compensatory enlargement in native coronary arteries by postmortem histopathological examination,1 5 epicardial echocardiography,2 and IVUS imaging3 4 or from the paradoxical constriction or the focal failure of compensatory enlargement reported by Pasterkamp et al8 in femoral arteries using IVUS imaging and histopathological examination. The absence of focal compensatory enlargement in stenotic SVBGs appears to be a potentially important contributing factor associated with the progression of stenoses in coronary SVBGs.

Technical Concerns
To investigate the vascular remodeling in human coronary arteries, three conventional quantitative techniques are used: pathological study, epicardial ultrasound imaging during cardiac surgery, and IVUS imaging during coronary catheterization. In pathological studies, perfusion fixation is needed to prevent shrinkage of the specimen. However, even this fixation method and pathological preparation alter quantitative measurements and thus are not ideal compared with in vivo studies.12 Epicardial ultrasound imaging is performed under general anesthesia, in an open-chest state, and the ultrasound probe is placed directly on the exposed epicardial coronary artery; thus, measurement of dimension and area of the vessel may not represent the normal physiological state. The accuracy of IVUS measurements has been validated.13 14 However, IVUS imaging, which requires selective SVBG catheterization and insertion of the imaging catheter into the vessel, has the potential to cause vasospasm in up to 3% of patients.15 16 To prevent this potential side effect, 100 to 200 µg of nitroglycerin was injected directly into the SVBG before or during the imaging with the IVUS catheter. No clinically or angiographically apparent vasospasm was detected in this study.

In studies of native coronary arteries using IVUS, lumen area and area within the internal elastic lamina or EEL are usually measured because the sonolucent zone reported to represent the media9 10 is often relatively clear and the border between the adventitia and surrounding connective tissue is usually difficult to differentiate. However, in the study of SVBGs, it is unclear which portion of the vessel wall, intima, media, or adventitia results in remodeling of the whole vessel. Therefore, in this study, we used both the EEL area and the vessel cross-sectional area to examine the focal compensatory enlargement or vessel constriction in SVBGs. We found that of 43 SVBGs, the outer border of the whole vessel was clearly demonstrated in 36 vessels (84%) and a circumferential sonolucent zone was identified in 25 vessels (58%). The difference between SVBGs and native coronary arteries in the ability to measure the vessel cross-sectional area and the EEL area may reflect the pathological changes that occur in SVBGs after implantation as arterial conduits. The media and the adventitia are reported to be replaced with fibrous tissue, and the internal and external elastic laminae may become obscure even under microscopic examination.17

In cases in which the sonolucent zone of the SVBG was not identified and only the outer border of the vessel was delineated, the plaque area could not be calculated, and therefore, the disease severity at the reference site in SVBG was uncertain. Therefore, to compare the disease severity between the reference site in SVBGs and that of fresh saphenous veins, we measured in vitro the lumen area and the vessel cross-sectional area of fresh saphenous veins that were prepared for CABG but not used. The measured and calculated variables of the reference sites in SVBGs were very similar to those of fresh saphenous veins. These data suggest that the reference sites of the SVBGs are minimally diseased and can be used as true references in this study.

Intraobserver and interobserver variabilities for measurements of lumen area, vessel cross-sectional area, and EEL area in SVBGs were quite small, with mean differences of <4%. Thus, the measurements are considered to be reproducible and reliable.

Compensatory Enlargement in Native Coronary Arteries and Other Arteries
Compensatory arterial enlargement in response to plaque formation was first described in coronary arteries18 and peripheral arteries19 of monkeys and was illustrated in human left main coronary arteries by postmortem histological examination.1 Glagov et al1 showed that human left main coronary arteries enlarge in relation to the development of atherosclerotic plaque and that the lumen area is preserved until the plaque development exceeds the compensatory mechanisms of the vessel, usually when plaque size is {approx}40% of the arterial cross-sectional area. Although the mechanisms of compensatory arterial enlargement have not been clarified, some hypotheses have been proposed20 : (1) The local increase in wall shear stress caused by plaque development may stimulate endothelium-dependent arterial dilatation or (2) the development of plaque may lead to degradation of the media and adventitia, resulting in passive bulging of the plaque. This concept of compensatory enlargement suggests and is consistent with the findings that angiographically normal or nearly normal segments of the artery have substantial plaque, and further, that the lesion severity can frequently be underestimated by angiography, which visualizes only the lumen of the vessel. This finding has been validated in in vivo human coronary arteries by IVUS imaging3 4 and also confirmed by the pathological study of the left anterior descending coronary arteries from humans and nonhuman primates.5

As noted by Losordo et al,7 however, many of the conclusions of pathological and IVUS studies1 3 4 5 are based on the correlation between plaque area size and vessel area size pooled from single measurements of different vessels of different individuals, and no comparison was made within the same vessel. If lumen area size, plaque area size, and vessel area size of different individuals are analyzed together, the interpretation of the data becomes very complex because the vessel size is reported to vary according to age, body weight, height, heart size, sex, blood pressure level, and plasma lipid concentrations.1 5 21 In studies of native human coronary arteries, only McPherson et al2 have previously compared coronary arterial size at proximal and distal coronary segments by epicardial echocardiography. They showed that in patients with coronary arterial lesions, total arterial area increased from the proximal reference site to the lesion site of the artery, confirming the concept of compensatory enlargement in native coronary arteries. In our study, the factors that make the measurements at the lesion site, intermediate site, and reference site comparable are that each of the segments are a part of the same vessel, that each segment of the SVBG was exposed to the same systemic factors over time, and that the SVBGs have no side branches, which have been reported to affect vessel size.22

Our findings do not exclude the possibility that uniform vessel enlargement or constriction was occurring in the whole SVBGs, because no control measurements were made just after the implantation of the SVBGs. There are inherent limitations to using in vitro fresh saphenous veins as a reference standard, even when imaged, as in this study, under physiological pressures. Namely, the size of the vein graft itself and the lack of physiological conditions all may create some degree of artifact. Although the vessel cross-sectional area of in vitro fresh saphenous vein was identical to that demonstrated in patients, it is still possible that this is artifactual and that the SVBG either dilated or constricted in its entirety from its condition immediately after implantation at the time of surgery. Nonetheless, it is clear from our data that there was no difference in the degree of vessel remodeling among the three different sites (reference, intermediate, and lesion site), and we believe it is unlikely that the same degree of vessel remodeling would occur at each site with different disease severity.

Development of Stenosis and Vessel Remodeling in SVBG
The long-term patency of SVBGs is reported to be 40% to 60% for grafts >=10 years old after CABG.23 However, in most patients who need multiple CABGs, the two internal mammary arteries are not sufficient, and SVBGs are needed for complete revascularization. Thus, the autologous saphenous vein still remains the most commonly used coronary artery bypass conduit.

The closure of SVBGs soon after surgery is reported to be caused by thrombosis, whereas late closure of SVBG is due to fibromuscular intimal hyperplasia and/or atherosclerosis, with some contribution from mural thrombi.24 Many mechanisms resulting in development of stenosis in SVBG are postulated, including functional differences in venous and arterial endothelium, mechanical and functional endothelial damage, secondary activation of platelet and vasoactive substances, and vessel wall ischemia caused by the loss of vasa vasorum.17 24 In addition to these possible mechanisms, our data suggest that the absence of focal compensatory enlargement could be one of the mechanisms resulting in the development of stenoses in SVBGs that have been implanted for >=8 years. One possible explanation for the absence of focal compensatory enlargement in SVBGs may be related to fibrous tissue proliferation secondary to intimal damage and vessel wall ischemia. Besides fibromuscular intimal hyperplasia, it has been reported that the smooth muscle fibers of the media are replaced, in part or in whole, by fibrous tissue and collagen, and in the adventitia there is a marked increase in fibrous tissue, with severe disruption or complete replacement of elastic fibers.17 The previously described mechanisms of compensatory arterial enlargement may not be operant under the circumstances in which much of the normal vessel wall has been replaced by fibrous tissue.

Conclusions
These in vivo IVUS data from human coronary SVBGs 8 to 23 years after implantation demonstrate (1) that no focal compensatory enlargement or vessel constriction occurred in stenotic segments compared with the reference segments and (2) that the absence of compensatory enlargement appears to be a potentially important factor in the progression of stenoses in coronary SVBGs.


*    Selected Abbreviations and Acronyms
 
CABG = coronary artery bypass graft surgery
EEL = external elastic lamina
IVUS = intravascular ultrasound
SVBG = saphenous vein bypass graft


*    Acknowledgments
 
This work was supported in part by the Self Defense Forces Central Hospital, Japan; the Swedish Medical Association; St Paul's Hospital, South Korea; the Lee E. Siegel, MD, Memorial Fund; and the Herbert Stein, MD, Research Fund. The authors thank Dr Michael C. Fishbein for his critical review of this manuscript.

Received June 20, 1995; revision received August 23, 1995; accepted September 25, 1995.


*    References
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
up arrowDiscussion
*References
 
1. Glagov S, Weisenberg E, Zarins CK, Stankunavicius R, Kolettis GJ. Compensatory enlargement of human atherosclerotic coronary arteries. N Engl J Med. 1987;316:1371-1375. [Abstract]

2. McPherson DD, Sirna SJ, Hiratzka LF, Thorpe L, Armstrong ML, Marcus ML, Kerber RE. Coronary arterial remodeling studied by high-frequency epicardial echocardiography: an early compensatory mechanism in patients with obstructive coronary atherosclerosis. J Am Coll Cardiol. 1991;17:79-86. [Abstract]

3. Gerber TC, Erbel R, Görge G, Ge J, Rupprecht H-J, Meyer J. Extent of atherosclerosis and remodeling of the left main coronary artery determined by intravascular ultrasound. Am J Cardiol. 1994;73:666-671. [Medline] [Order article via Infotrieve]

4. Hermiller JB, Tenaglia AN, Kisslo KB, Phillips HR, Bashore TM, Stack RS, Davidson CJ. In vivo validation of compensatory enlargement of atherosclerotic coronary arteries. Am J Cardiol. 1993;71:665-668.[Medline] [Order article via Infotrieve]

5. Clarkson TB, Prichard RW, Morgan TM, Petrick GS, Klein KP. Remodeling of coronary arteries in human and nonhuman primates. JAMA. 1994;271:289-294. [Abstract/Free Full Text]

6. Steinke W, Els T, Hennerici M. Compensatory carotid artery dilatation in early atherosclerosis. Circulation. 1994;89:2578-2581. [Abstract/Free Full Text]

7. Losordo DW, Rosenfield K, Kaufman J, Pieczek A, Isner JM. Focal compensatory enlargement of human arteries in response to progressive atherosclerosis: in vivo documentation using intravascular ultrasound. Circulation. 1994;89:2570-2577. [Abstract/Free Full Text]

8. Pasterkamp G, Wensing PJW, Post MJ, Hillen B, Mali WPTM, Borst C. Paradoxical arterial wall shrinkage may contribute to luminal narrowing of human atherosclerotic femoral arteries. Circulation. 1995;91:1444-1449. [Abstract/Free Full Text]

9. Gussenhoven EJ, Essed CE, Lancée CT, Mastic F, Frietman P, Van Egmond FC, Reiber J, Bosch H, Van Urk H, Roelandt J, Bom N. Arterial wall characteristics determined by intravascular ultrasound imaging: an in vitro study. J Am Coll Cardiol. 1989;14:947-952. [Abstract]

10. Nishimura RA, Edwards WD, Warnes CA, Reeder GS, Holmes DR Jr, Tajil AJ, Yock PG. Intravascular ultrasound imaging: in vitro validation and pathologic correlation. J Am Coll Cardiol. 1990;16:145-154. [Abstract]

11. Coy KM, Park JC, Fishbein MC, Laas T, Diamond GA, Adler L, Maurer G, Siegel RJ. In vitro validation of three-dimensional intravascular ultrasound for the evaluation of arterial injury after balloon angioplasty. J Am Coll Cardiol. 1992;20:692-700. [Abstract]

12. Siegel RJ, Swan K, Edwalds G, Fishbein MC. Limitations of postmortem assessment of human coronary artery size and luminal narrowing: differential effects of tissue fixation and processing on vessels with different degrees of atherosclerosis. J Am Coll Cardiol. 1985;5:342-346. [Abstract]

13. Pandian NG, Kreis A, Brockway B, Isner JM, Sacharoff A, Boleza E, Caro R, Muller D. Ultrasound angioscopy: real-time, two-dimensional, intraluminal ultrasound imaging of blood vessels. Am J Cardiol. 1988;62:493-494. [Medline] [Order article via Infotrieve]

14. Potkin BN, Bartorelli AL, Gessert JM, Neville RF, Almagor Y, Roberts WC, Leon MB. Coronary artery imaging with intravascular high-frequency ultrasound. Circulation. 1990;81:1575-1585. [Abstract/Free Full Text]

15. Nishioka T, Luo H, Eigler NL, Tabak SW, Lepor N, Forrester JS, Siegel RJ. The evolving utility of intracoronary ultrasound. Am J Cardiol. 1995;75:539-541. [Medline] [Order article via Infotrieve]

16. Hausmann D, Erbel R, Alibelli-Chemarin M-J, Boksch W, Caracciolo E, Cohn JM, Culp SC, Daniel WG, De Scheerder I, DiMario C, Ferguson JJ III, Fitzgerald PJ, Friedrich G, Ge J, Görge G, Hanrath P, Hodgson JM, Isner JM, Jain S, Maier-Rudolph W, Mooney M, Moses JW, Mudra H, Pinto FJ, Smalling RW, Talley JD, Tobis JM, Walter PD, Weidinger F, Werner GS, Yeung AC, Yock PG. The safety of intracoronary ultrasound: a multicenter survey of 2207 examinations. Circulation. 1995;91:623-630. [Abstract/Free Full Text]

17. Spray TL, Roberts WC. Changes in saphenous veins used as aortocoronary bypass grafts. Am Heart J. 1977;94:500-516. [Medline] [Order article via Infotrieve]

18. Bond MG, Adams MR, Bullock BC. Complicating factors in evaluating coronary artery atherosclerosis. Artery. 1981;9:21-29. [Medline] [Order article via Infotrieve]

19. Armstrong ML, Heistad DD, Marcus ML, Megan MB, Piegors DJ. Structural and hemodynamic responses of peripheral arteries of macaque monkeys to atherogenic diet. Arteriosclerosis. 1985;5:336-346. [Abstract/Free Full Text]

20. Zarins CK, Weisenberg E, Kolettis G, Stankunavicius R, Glagov S. Differential enlargement of artery segments in response to enlarging atherosclerotic plaques. J Vasc Surg. 1988;7:386-394. [Medline] [Order article via Infotrieve]

21. Hort W, Lichti H, Kalbfleisch H, Kohler F, Frenzel H, Milzner-Schwarz U. The size of human coronary arteries depending on the physiological and pathological growth of the heart, the age, the size of the supplying areas and the degree of coronary stenosis: a postmortem study. Virchows Arch. 1982;397:37-59.

22. Hutchins GM, Miner MM, Boitnott JK. Vessel caliber and branch-angle of human coronary artery branch-points. Circ Res. 1976;38:572-576. [Abstract/Free Full Text]

23. Campeau L, Enjalbert M, Lespérance J, Vaislic C, Grondin CM, Bourassa MG. Atherosclerosis and late closure of aortocoronary saphenous vein grafts: sequential angiographic studies at 2 weeks, 1 year, 5 to 7 years and 10 to 12 years after surgery. Circulation. 1983;68(suppl II):II-1-II-7.

24. Angelini GD, Newby AC. The future of saphenous vein as a coronary artery bypass conduit. Eur Heart J. 1989;10:273-280.[Abstract/Free Full Text]




This article has been cited by other articles:


Home page
J Am Coll CardiolHome page
Y. J. Hong, G. S. Mintz, S. W. Kim, S. Y. Lee, S. Y. Kim, T. Okabe, A. D. Pichard, L. F. Satler, R. Waksman, K. M. Kent, et al.
Disease progression in nonintervened saphenous vein graft segments a serial intravascular ultrasound analysis.
J. Am. Coll. Cardiol., April 14, 2009; 53(15): 1257 - 1264.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
G. T. Lau, L. J. Ridley, P. G. Bannon, L. A. Wong, J. Trieu, D. B. Brieger, H. C. Lowe, B. S. Freedman, and L. Kritharides
Lumen Loss in the First Year in Saphenous Vein Grafts Is Predominantly a Result of Negative Remodeling of the Whole Vessel Rather Than a Result of Changes in Wall Thickness
Circulation, July 4, 2006; 114(1_suppl): I-435 - I-440.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
J. Pregowski, P. Tyczynski, G. S. Mintz, S.-W. Kim, A. Witkowski, R. Waksman, A. Pichard, L. Satler, K. Kent, M. Kruk, et al.
Incidence and Clinical Correlates of Ruptured Plaques in Saphenous Vein Grafts: An Intravascular Ultrasound Study
J. Am. Coll. Cardiol., June 21, 2005; 45(12): 1974 - 1979.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
P. Schoenhagen, K. M. Ziada, D. G. Vince, S. E. Nissen, and E. M. Tuzcu
Arterial remodeling and coronary artery disease: the concept of "dilated" versus "obstructive" coronary atherosclerosis
J. Am. Coll. Cardiol., August 1, 2001; 38(2): 297 - 306.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
J. D. Mannion, M. L. Ormont, M. G. Magno, J. E. O'Brien, Y. Shi, and A. Zalewski
Sustained reduction of neointima with c-myc antisense oligonucleotides in saphenous vein grafts
Ann. Thorac. Surg., December 1, 1998; 66(6): 1948 - 1952.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
J. G. Motwani and E. J. Topol
Aortocoronary Saphenous Vein Graft Disease : Pathogenesis, Predisposition, and Prevention
Circulation, March 10, 1998; 97(9): 916 - 931.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
J. D. Mannion, M. L. Ormont, Y. Shi, J. E. O'Brien Jr.,, W. Chung, F. Roque, and A. Zalewski
Saphenous vein graft protection: Effects of c-myc antisense
J. Thorac. Cardiovasc. Surg., January 1, 1998; 115(1): 152 - 161.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
J. E. O'Brien Jr., Y. Shi, A. Fard, T. Bauer, A. Zalewski, and J. D. Mannion
WOUND HEALING AROUND AND WITHIN SAPHENOUS VEIN BYPASS GRAFTS
J. Thorac. Cardiovasc. Surg., July 1, 1997; 114(1): 38 - 45.
[Abstract] [Full Text]


Home page
CirculationHome page
Y. Shi, J. E. O'Brien Jr, J. D. Mannion, R. C. Morrison, W. Chung, A. Fard, and A. Zalewski
Remodeling of Autologous Saphenous Vein Grafts : The Role of Perivascular Myofibroblasts
Circulation, June 17, 1997; 95(12): 2684 - 2693.
[Abstract] [Full Text]


Home page
Arterioscler. Thromb. Vasc. Bio.Home page
A. Zalewski and Y. Shi
Vascular Myofibroblasts : Lessons From Coronary Repair and Remodeling
Arterioscler Thromb Vasc Biol, March 1, 1997; 17(3): 417 - 422.
[Full Text]


Home page
Arterioscler. Thromb. Vasc. Bio.Home page
H. Luo, T. Nishioka, N. L. Eigler, J. S. Forrester, M. C. Fishbein, H. Berglund, and R. J. Siegel
Coronary Artery Restenosis After Balloon Angioplasty in Humans Is Associated With Circumferential Coronary Constriction
Arterioscler Thromb Vasc Biol, November 1, 1996; 16(11): 1393 - 1398.
[Abstract] [Full Text]


This Article
Right arrow Abstract Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Nishioka, T.
Right arrow Articles by Siegel, R. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Nishioka, T.
Right arrow Articles by Siegel, R. J.