(Circulation. 1995;92:2463-2472.)
© 1995 American Heart Association, Inc.
Articles |
From the Cardiac Catheterization Laboratory and Intracoronary Imaging Laboratory, Thoraxcenter, University Hospital Dijkzigt, Rotterdam, Netherlands.
Correspondence to P.W. Serruys, MD, PhD, Catheterization Laboratory, Thoraxcenter Room Bd 416, University Hospital Dijkzigt, Erasmus University Rotterdam, Dr Molewaterplein 40, 3015 GD Rotterdam, Netherlands.
| Abstract |
|---|
|
|
|---|
Methods and Results In 22 patients after balloon angioplasty, the
coil balloon was studied for (1) feasibility of local heparin delivery,
(2) symptoms and signs of ischemia during prolonged deployment
compared with angioplasty balloon occlusion, (3) coronary
pressure and flow distal to the inflated device, and (4) long-term
clinical and angiographic results. During prolonged
intracoronary deployment of the coil balloon (29±8
minutes), 5 of 22 patients developed mild chest pain versus 20 of 22
during angioplasty (275±283 seconds). Neither
hemodynamic nor vectorcardiographic signs of
ischemia were detected, in contrast to angioplasty balloon
occlusion. Baseline flow across the coil balloon was 44±31 mL/min,
increasing by a factor of 1.8±0.7 during pharmacologically induced
hyperemia. A mean volume of 14.2±6.1 mL containing 1416±608
IU of heparin was infused locally at a pressure of 122±54 mm Hg. At
7±1-month follow-up, 1 asymptomatic patient had
died, and of the remaining 21, 17 (81%) were
asymptomatic. Angiographic follow-up was obtained
in 15 of 21 patients (71%), including all 4 symptomatic
patients. Mean minimal luminal diameter after the procedure was
2.16±0.49 mm and at follow-up, 1.89±0.45 mm, which corresponds
to
a restenosis rate (diameter stenosis
50%) of 7%
(1/15).
Conclusions Intracoronary use of the coil balloon after balloon angioplasty proved to be feasible and subjectively as well as objectively well tolerated during prolonged deployment by virtue of its perfusion properties. High volumes of heparin solution can be infused locally at very low pressure. No unfavorable clinical or angiographic long-term effects were observed.
Key Words: angioplasty heparin restenosis hemodynamics perfusion
| Introduction |
|---|
|
|
|---|
A new local drug-infusion catheter (Dispatch Scimed Systems Inc) has been designed to overcome these limitations by combining infusion and perfusion characteristics.
The aim of this study was to assess in humans the following aspects of this new drug-delivery device: (1) acute feasibility of prolonged local deployment with simultaneous infusion of a heparin solution after balloon angioplasty, (2) signs of myocardial ischemia during prolonged intracoronary deployment compared with temporary balloon occlusion during angioplasty, (3) coronary hemodynamics distal to the inflated device, and (4) chronic effects of prolonged local coil balloon deployment and simultaneous infusion of a heparin solution after balloon angioplasty.
| Methods |
|---|
|
|
|---|
|
Study Population
The clinical and angiographic
characteristics of the 22 patients
included in the study are summarized in Table 1
. The
study period consisted of two consecutive evaluation phases, each of
which included 11 patients, each with a recruitment period of 25 days
(cohort 1) and 53 days (cohort 2), respectively. The selected study
populations represented 23% (11/48) and 11% (11/103),
respectively, of the population treated by balloon angioplasty over the
same time period at our institution. Inclusion criteria were stable or
unstable angina pectoris, one- or two-vessel disease, and normal or
hypokinetic left ventricular wall motion in the territory
supplied by the vessel to be treated, with a global ejection fraction
of >50%. In the second cohort, patients with a target vessel
supplying either the anterior wall (LAD, n=6) or the
inferior wall (RCA, n=3, and dominant circumflex artery,
n=2) were selected. Except for vessel size (reference diameter
2.5
mm), no specific lesion type precluded initial enrollment. Exclusion
criteria were age <18 years or >80 years, evolving myocardial
infarction in the territory supplied by the vessel to be treated (<7
days old), left bundle-branch block or bifascicular block, life
expectancy <6 months, and factors making follow-up difficult.
Written informed consent was obtained from all patients before
inclusion.
|
Balloon Angioplasty and Local Heparin Delivery
Procedure
After an intravenous bolus injection of heparin
10 000 IU, aspirin 250 mg, and diazepam 5 mg, a conventional balloon
angioplasty was performed. After an optimal angioplasty result was
achieved, as assessed visually, the nondilatational coil balloon was
inserted through an 8F guiding catheter and deployed at the angioplasty
site. A 30-minute-long deployment was attempted unless subjective
symptomatic intolerance or ECG signs (1 of 3 leads) of
ischemia occurred. A solution with commercially available
heparin (100 IU/mL in NaCl 0.9%) was infused at 12 to 36 mL/h. In the
first 11 patients, the infusion rate was increased in a stepwise manner
by 12 mL/h every 10 minutes up to 36 mL/h to detect potential
detrimental effects of augmenting volume administration. In the second
11 patients, the infusion rate was 36 mL/h from the onset.
Signs of Myocardial Ischemia: Angioplasty Balloon Versus
Coil Balloon
Symptoms. Angina was scored as described
earlier20 21 on a subjective scale from 1 to 10
during
angioplasty balloon occlusion and coil balloon deployment. Use of
analgesic was restricted as much as possible, and it was never
administered before drug delivery device insertion.
Left ventricular
hemodynamics.
An 8F pigtail high-fidelity tip manometer (Sentron, Cordis Europe
NV) was inserted via the contralateral femoral artery into the left
ventricle to monitor pressure changes throughout the procedure.
Pressure and derived indexes were calculated and displayed on-line
with an updated version of a system previously
described.22 23 24 25
Monitored left ventricular
hemodynamic changes included the following contraction
parameters: peak LVP, peak positive dP/dt, and maximal
velocity of contraction (Vmax), as well as
diastolic and relaxation parameters: EDP, peak
negative dP/dt, and time constant of relaxation (T). Time constant of
LVP fall, T (ms), was determined with the semilogarithmic model of
relaxation: P(t)=P0 e-t/T,
where P is pressure, t is time, and P0 is equivalent to
Pb (P begin) defined as pressure at the point at which
dP/dt is minimal (maximal negative dP/dt).24 In the
biexponential fitting model of pressure fall, T for the first 40 ms is
defined as T1 and that after 40 ms as T2.
Ischemia during coronary occlusion by conventional
balloon was evaluated by use of a baseline value
5 minutes before
angioplasty and a value at the end of PTCA. For evaluation of
ischemia during coil balloon deployment, a baseline value
5
minutes before inflation, a value every 5 minutes up to 30 minutes, the
last value before deflation, and the value at the end of coil balloon
inflation were selected. End values for conventional balloon and coil
balloon were defined as mean of the last value before deflation and
first value after deflation.
Vectorcardiography. Vectorcardiographic
analysis
(MIDA, Ortivus Medical AB) was performed continuously during the
procedure as previously reported.26 Vectorcardiographic
changes were assessed as changes in QRS VD as well as ST VM and were
calculated as described earlier27 and depicted as a
continuous trend. Clinically relevant QRS VD and ST VM changes were
defined as changes
15 µVs and
0.1 mV, respectively. Baseline
values were defined as the mean values over a 5-minute period before
conventional balloon inflation and coil balloon deployment,
respectively. Values during conventional balloon and coil balloon
deployment were represented as averaged value over the
whole inflation period. In cases of multiple inflations, the longest
inflation period was selected.
Ventriculography. Ventriculography (one projection at a 30° right anterior oblique angle at 25 frames per second) was performed before PTCA and during coil balloon deployment by injection of 0.75 mL/kg of nonionic contrast medium (Iopamiro 370, Bracco). Ventriculograms were assessed off-line by automated contour detection (CESAR system version 86) and analyzed according to the model of Slager et al28 : Segmental end-systolic and end-diastolic volume changes were calculated along a system of 20 coordinates: segmental contribution to total ejection fraction was used to quantify regional wall motion and regrouped in five regions of interest (anterobasal, anterolateral, posterobasal, posteroapical, and apical). According to the vessel treated, the corresponding regional wall motion change, defined as difference before PTCA versus during coil balloon deployment, was calculated.
Enzymes. CPK was determined 6 and 12 hours after the procedure. CPK-MB was measured whenever the values of CPK were elevated.
Coronary Hemodynamics
During coil balloon deployment, after
removal of the guide wire,
mean fluid-mediated pressure gradient across the device was
measured through the wire port. A 15-MHz 0.014-in. Doppler guide
wire (Flowire, Cardiometrics) was then inserted through the
guide-wire port. Coronary flow velocity was measured 1.5 to
2 cm distal to the device in basal conditions and after
intracoronary injection of papaverine (left
coronary artery, 12.5 mg; RCA, 8 mg) or adenosine (left
coronary artery, 18 µg; RCA, 12 µg).29 The
change in protocol from papaverine to adenosine was due to one
episode of ventricular fibrillation secondary to
intracoronary injection of papaverine. Angiography was
performed with the Doppler wire at the site of volume sampling.
Blood flow through the device was calculated as the product of
time-averaged peak flow velocity and cross-sectional area
determined by QCA at the site of Doppler volume sampling (Fig
2
). Flow measurements obtained by this method have been
validated with electromagnetic flow probes in open-chest
experimental animals.30
|
Inventory of Analyzed Ischemia and Perfusion
Parameters per Patient
Left ventricular tip manometry monitoring was
not
carried out in 4 of the 22 patients because of recent stroke, aortic
stenosis, or a severe peripheral arteriopathy.
Vectorcardiographic monitoring was performed in the first cohort of
patients (patients 1 through 11), whereas ventriculography before PTCA
as well as during coil balloon deployment was performed in the second
cohort of patients (10 of 11; patients 13 through 22). In one case,
ventriculography was not performed because of severe
peripheral arteriopathy.
Transcoil balloon pressure gradient measurement and Doppler study (baseline) were performed in 17 patients. In 5 patients, flow velocity evaluation distal to the deployed coil balloon was not performed because of a second significant lesion distal to the treated one (n=2), angiographically visible collaterals in a total occlusion (n=2), and technical failure (n=1).
Acute Feasibility and Long-term Evaluation
Acute feasibility
evaluation included technical, angiographic,
and clinical adverse events during and after coil balloon
procedure.
Long-term evaluation was scheduled after 6 months and consisted of an interview, a physical examination, and angiographic follow-up. In case of refusal of angiographic follow-up, a perfusion scintigram was requested.
Recorded clinical events were the following: death, cerebrovascular accident, myocardial infarction, bypass surgery, or a second percutaneous intervention involving the previously treated lesion between the time of the initial procedure and the follow-up angiography (7±1 months) as well as an unplanned stent implantation after balloon angioplasty. According to our institutional practice, stenting was planned after angioplasty if a dissection type B, C, or D was longer than 10 mm without compulsory flow impairment or in case of a dissection type E or F. All cerebrovascular accidents were considered, regardless of their cause. Myocardial infarction was diagnosed if there were new pathological Q waves according to the Minnesota Code31 or if there was an increase in serum creatine kinase to more than the normal value, together with a pathological increase in myocardial isoenzymes. Coronary artery bypass graft surgery was defined to include emergency or elective surgery involving the previously treated segment.
Angiographic Analysis
Before the procedure, the lesion type
was classified according
to the modified AHA classification.32 After PTCA as well
as after local heparin administration, the lesion was evaluated
qualitatively and by QCA. QCA was performed off-line with an
automated computer-assisted edge-detection system
(Cardiovascular Angiography Analysis System II
[CAAS II], Pie Medical).33 34
Acute and long-term angiographic results were evaluated with the same matched "working" projection before the procedure, after PTCA and coil balloon deployment, and at follow-up.35
Statistics
Continuous variables are expressed as
mean±SD.
Hemodynamic, vectorcardiographic, and ventriculographic
characteristics during conventional balloon and coil balloon deployment
were compared by repeated-measures ANOVA. Predefined comparisons
were performed by paired t test using Bonferroni correction.
Differences were considered significant if the null hypothesis could be
rejected at the .05 probability level.
Discrete variables are expressed as counts and percentages.
| Results |
|---|
|
|
|---|
Angiographically, in 22 patients, the MLD of the treated lesions was 1.07±0.40 mm at diagnostic angiography. After PTCA, with balloon nominal values ranging between 2.5 and 3.5 mm and with a mean inflation pressure of 8.6±2.9 atm (range, 4 to 14 atm) and a mean inflation time of 274±270 seconds (range, 60 to 1200 seconds), MLD increased to 1.92±0.36 mm. After coil balloon (balloon size ranging from 3.0 to 4.0 mm nominal size at mean inflation pressure of 6.9±1 atm and a mean inflation time of 29±8 minutes), MLD increased nonsignificantly to 1.98±0.41 mm. In 14 of 22 patients (64%), the diameter after coil balloon was greater than or equal to the diameter after PTCA. The diameter size of the drug-delivery device matched (n=19) or was 0.5 mm larger than (n=3) the largest balloon used for the conventional dilatation.
Dissections were observed in 7 patients after
PTCA (1 type A, 2 type B,
2 type C, 2 type D). Of these 7 dissections, 4 fulfilled the criteria
for stenting before local drug delivery (Fig 3
). One
type B dissection (8.2 mm long) was successfully treated by prolonged
deployment of the coil balloon device. Among the 6 remaining
dissections, a type A dissection after PTCA was evaluated as type B
(5.5 mm long) after coil balloon deployment. A type D dissection (34 mm
long) became a type F dissection after coil balloon deployment at two
sites for 15 and 18 minutes and was successfully treated by a bailout
stent implantation. Three further dissections (types B, C, and D) were
stented after coil balloon deployment according to the institutional
clinical practice (>10 mm long). Their dissection lengths after
heparin solution infusion did not change significantly (B, 10.5 versus
11.1 mm; C, 24.5 versus 19.4 mm; and D, 24.5 versus 26.1 mm). A last
type C dissection (<10 mm long) did not change significantly in length
after heparin infusion (5.6 versus 6.3 mm) (Fig 3
).
|
Signs of Ischemia: Angioplasty Balloon Versus Coil
Balloon
Symptoms. Of 22 patients, 18 (82%) suffered angina
grade 3 to 10 (mean, 8±3) during inflation of the conventional PTCA
balloon. Of 22 patients, 5 (23%) developed grade 2 to 7 (mean, 4±2)
symptoms during coil balloon deployment. Of these 5 patients, 3
described anginal symptoms of grade 2 to 3 and 2 patients, symptoms of
grade 6 to 7, one of whom had a lesion at a major bifurcation point
(coil balloon covering first marginal branch) and the second a
subocclusive dissection of the first diagonal branch occurring during
the introduction of the Doppler wire. After coil balloon removal
and intracoronary nitrate administration, flow was restored
and symptoms abated rapidly in both cases.
Left ventricular
hemodynamics.
Left ventricular hemodynamic
parameters during conventional balloon as well as coil
balloon inflation are summarized in Table 2
. No
significant hemodynamic changes were observed during
coil balloon deployment. During conventional balloon occlusion,
T1, T2, negative dP/dt, and EDP
changed significantly compared with the baseline value (Fig 4
).
|
|
Vectorcardiography. Vectorcardiographic monitoring did not
reveal any significant ST-VM changes during conventional balloon
inflation and coil balloon deployment compared with baseline values
(94.6±68.5 versus 125±84.4 mV and 91.5±59.2 versus
90.5±50.8 mV,
respectively). QRS VD showed an ischemia-induced increase
during conventional balloon angioplasty, but no changes were observed
during coil balloon deployment (6.6±4.6 versus 16.8±5.3 µVs
[P<.01] and 9.1±4.1 versus 9.0±2.8 µVs
[NS]) (Fig 5
).
|
Ventriculography. As summarized
in Table 3
,
during coil balloon deployment, ventricular function
remained unchanged with respect to ejection fraction (baseline value
during diagnostic procedure versus value during coil
balloon inflation). Similarly, regional left ventricular
wall motion in the targeted region seemed not to be affected by
prolonged coil balloon deployment.
|
CPK. No CPK rise was observed at 7±1 and 14±4 hours (71±10 and 40±16 IU/L [normal values up to 110 IU/L]), with the exception of one stented patient (CPK, 406 IU/L) at 7 hours after the procedure. In this patient, however, CPK-MB was <6% (6 IU/L), and CPK rise was attributed to soft-tissue compression secondary to the use of an external mechanical compression device for local hemostasis after sheath removal.
Coronary Hemodynamics
Coronary hemodynamic findings are
reported
with respect to the diameter of the device.
Mean pressure gradients
across the device were 32±13 (n=12), 22±13
(n=4), and 4 (n=1) mm Hg for device sizes of 3.0, 3.5, and 4.0
mm,
respectively. The influence of drug infusion flow (12 to 36 mL/h) on
pressure gradient was negligible. Flow velocities and flow measurements
are summarized in Table 4
.
|
Long-term Results
Clinical follow-up was available in all 22
patients. One
78-year-old asymptomatic patient with aortic
stenosis (gradient, 50 mm Hg) and suffering from a
reactivation of an ulcerative colitis died of sustained acute heart
failure 52 days after the procedure after two blood transfusions.
At 7±1 months, no patient presented with a myocardial infarction or a revascularization of the treated segment. Of 21 eligible for follow-up, 4 (19%) were symptomatic. Two complained of atypical angina and 2 of stable effort angina, CCS stage 2.36 Follow-up angiography was performed in 15 of 21 patients (71%) and included the 4 symptomatic patients. Of the 6 asymptomatic refusal patients, 3 had a thallium perfusion scintigraphy and 3 refused any further investigation.
Of the 4 symptomatic patients, the 2 suffering from stable effort angina presented with a progression of previously insignificant lesions. One of the 2 patients presenting with atypical angina had a negative exercise test and no significant lesions. The second presented with restenosis of a lesion dilated during the drug-delivery procedure but treated solely by conventional angioplasty.
The 3 asymptomatic patients who underwent a sesta-MIBI myocardial perfusion scintigraphy did not show any signs of redistribution, which presumably excluded a silent ischemia.
Postprocedure angiography (15 patients: coil balloon,
n=12; coil
balloon and stent, n=3) showed an MLD of 2.16±0.49 mm and at
follow-up an MLD of 1.89±0.45 mm, corresponding to an absolute
loss of 0.27±0.51 mm (Table 5
). If we compare the 12
lesions treated only with coil balloon and the 3 lesions also stented,
MLD after the procedure was 2.01±0.39 versus 2.78±0.39 mm, and
at
follow-up, 1.83±0.41 versus 2.13±0.62 mm, corresponding to a
loss
of 0.18±0.43 versus 0.65±0.71 ±13% (range, 17% to 59%).
One
patient of 15 (7%) had a diameter stenosis >50%.
|
| Discussion |
|---|
|
|
|---|
By design, the coil balloon is a nondilatational device and is not apt to develop a high mechanical force against the vessel wall (at nominal coil inflation pressure [6 atm], the pressure against the wall is <2 atm; Scimed data on file). However, it is able to exert a low uninterrupted pressure for a long time period because of its perfusion properties.
In our series, prolonged deployment (29±8 minutes; 30 minutes per protocol) of the coil balloon after angioplasty slightly increased the mean MLD (MLD, 1.98±0.41 versus 1.92±0.36 mm), and a case-by-case analysis showed an unchanged or improved MLD in 14 of 22 patients (64%) and deterioration in 8 of 22 (36%). No coronary spasms were observed after prolonged deployment.
Dissections
not longer than the device (20 mm) were not substantially
affected by local infusion of the heparinized solution (Fig 3
).
In one
patient with a type A dissection after PTCA, local infusion of heparin
revealed a dissection flap (type B dissection), presumably by
dissolving the mural thrombus and rendering the flap angiographically
visible or worsening a mural dissection (length, 5.5 mm). Of the 3
patients with a dissection length greater than the device after balloon
angioplasty, one patient (34 mm, type D dissection) developed an
occlusive dissection (type F), necessitating a subsequent bailout stent
implantation after two prolonged deployments of the coil balloon (first
proximal, then distal). The potential hazard of worsening a preexisting
dissection by fluid infusion has to be borne in mind when a dissection
cannot be entirely covered by the device.
The driving mechanisms of drug solution transfer by the coil balloon are pressure and flow. Pressure in the drug compartment can theoretically be modified in two different manners: (1) by regulation of flow rate and (2) by variation of inflation pressure in the compliant polyolefin copolymer coils.37 In our study, pressures inside the drug compartment did not vary significantly in respect to the infusion rate used (12 mL/h, 114±12 mm Hg [n=4]; 24 mL/h, 114±45 mm Hg [n=5]; and 36 mL/h, 127±66 mm Hg [n=13]). This may be due to infusion fluid leakage at the proximal and distal ends of the coil, maintaining infusion pressure within the above-mentioned limits, and provides the system by its design and 6-atm inflation pressure with an infusion pressure control. Variations of inflation pressure were not specifically studied because of the recommended nominal inflation pressure of 6 atm by the manufacturer. As used in this study, heparin was driven by low pressure (117±54 mm Hg=0.15±0.07 atm) and by flow (12 to 36 mL/h) at the site of delivery. A volume of 14.2±6 mL (corresponding to 140 to 2220 IU of heparin) was infused at the site of angioplasty.
Theoretical disadvantages related to the design of the device could be an uneven delivery of drug to the endothelial surface due to the contact of the spiral coil with the vessel wall and jet-stream lesions produced at the level of the infusion slits. At the infusion rates used, however, only weeping could be observed at the distal end of the coil balloon. In normal porcine coronary arteries, infusions of Evans blue dye using a coil-balloon catheter showed macroscopically a spiral-shaped distribution of the dye on the vessel surface. Histological examination revealed minimal endothelial damage, without jet-stream lesions.38
Signs of Ischemia
The perfusion properties of the coil
balloon catheter were
indirectly evaluated in the clinical setting by assessment of signs of
myocardial ischemia. The subjective tolerance to prolonged
deployment observed in these patients was encouraging.
Hemodynamic parameters of relaxation and of
diastolic function (T1 and
T2, negative dP/dt, and EDP; Table 2
and Fig
4
) did
not reveal any significant changes during prolonged coil balloon
deployment compared with respective baseline values. These
variables were the most discriminatory hemodynamic
parameters of ischemia, as demonstrated during
conventional balloon inflation, confirming earlier
reports.25 Vectorcardiographic analysis did not
show any significant changes during coil balloon deployment either.
Even QRS VD, which is more sensitive than ST VM in detecting
ischemia (Fig 5
),27 did not change significantly
from baseline. Ventriculographic analysis of regional wall
motion during intracoronary deployment of the coil balloon
(Table 3
) did not detect any significant changes compared with
diagnostic ventriculography using the "regional
contribution to ejection fraction" method.25 39
Limitations in Assessment of Signs of
Ischemia
Sequence of procedures. A limitation in the
methodological evaluation of signs of ischemia might be that
conventional balloon inflation systematically preceded the coil balloon
deployment and thereby enhanced collateral recruitment40
and preconditioned the myocardium to
ischemia.41 Of the 22 patients, 8 (36%) had
multiple conventional balloon inflations (range, 2 to 4) before coil
balloon insertion. To avoid this potential bias, a randomized sequence
of inflation might have been incorporated. However, the coil balloon
cannot be used directly for angioplasty because by design, this balloon
is not intended primarily for dilatations of stenotic
lesions.
In a previous study evaluating the Stack perfusion balloon,21 a final conventional balloon inflation was performed after perfusion balloon inflation, which was preceded by a conventional balloon inflation. In that study, although inflation time was 30% longer (107±55 versus 139±71 seconds) and pain score slightly lower (6.1±2.1 versus 5.2±3.1), ECG changes remained similar, indicating that the two preceding inflations (conventional balloon, 107±55 seconds and perfusion balloon, 513±333 seconds) did not alter objectively appreciable changes. These findings were consistent with previously reported data42 43 in which occlusion pressure measured distal to the stenosis during balloon inflation did not change after serial occlusions.
The time elapsed between the last conventional balloon inflation and coil balloon deployment was 16±6 minutes. Since metabolic disturbances after conventional balloon inflations are reported to be short-lasting44 and totally reversible within 5 minutes,41 it might be inferred that complete recovery had been achieved in our patients, allowing a return to metabolic baseline before coil balloon deployment.
Left ventricular hemodynamics. The alteration of left ventricular relaxation reflects an asynchrony of contraction-relaxation of the left ventricle. These parameters are maximally altered within 15 seconds of ischemia,25 and although they have a tendency to regress partially, synchrony remains altered throughout the entire duration of occlusion.45 Therefore, the absence of abnormalities throughout the total duration of coil deployment is indicative of an adequate coronary perfusion.
Ventriculography. It
could be argued that localized wall
motion abnormalities were missed because a single projection (right
anterior oblique 30°) was performed. This may be of particular
concern for the territories supplied by the LCx and, to a lesser
extent, for the RCA.46 However, even the
subanalysis of the 6 cases supplied by the LAD did not
reveal any left ventricular wall motion abnormality during
prolonged coil balloon deployment (Table 3
).
Perfusion Properties
Autoperfusion properties of the coil
balloon catheter measured
directly in vitro (38% glycerol at room temperature and 100 mm Hg
continuous perfusion pressure) have indicated that flow is a function
of device size (mean of 15 consecutive measurements: 54±3, 74±1,
and
80±2 mL/min for the 3.0-, 3.5-, and 4.0-mm devices, respectively; data
on file, Scimed). These data have been confirmed by our in vivo
Doppler guide wire studies (Table 4
) evaluating the 3.0- and
3.5-mm
devices (37±27 and 70±35 mL/min). For the 4.0-mm device, only
one set
of measurements in an RCA (segment 2) is available, precluding any
conclusive statement. These autoperfusion properties were also
confirmed by a preserved pharmacologically induced hyperemic
response during intracoronary deployment of the coil
balloon (Table 4
).
Previous data on the most extensively evaluated perfusion balloon (Stack perfusion catheter, Advanced Cardiovascular Systems) reported in vitro flows of 60 mL/min with glycerol 38% at room temperature and a continuous pressure gradient of 80 mm Hg, irrespective of the balloon inflation pressure.47 With fresh human citrate blood (43.6% at 37°C), measured flow was 55 mL/min with a driving pressure of 80 mm Hg.48 In the clinical setting, calculated flow was 55±23 mL/min (n=7), derived from flow-velocity measurements using a Doppler-tipped guide wire alongside the balloon and angiographic cross-sectional area.49 Despite apparently similar perfusion data (coil balloon versus Stack balloon), inflation times longer than 20 minutes were rarely performed in evaluations of the Stack balloon.21 50 51 Furthermore, Quigley et al21 reported indirect signs of ischemia as assessed by ECG in more than 30% of patients before reaching the target inflation time of 10 minutes, and maximal chest pain score was more than half the value of that sustained with conventional balloon inflation (3.2±3.5 versus 6.1±2.1). These data suggest that the previously reported perfusion rates through the Stack perfusion catheter may have been overestimated.
Long-term Results
One asymptomatic patient with aortic
stenosis died 52 days after PTCA of acute congestive heart
failure after blood transfusions for active ulcerative colitis.
Postmortem examination was not performed. Clinical outcome at 7±1
months of this first cohort of patients treated by local
low-pressure heparin infusion after balloon angioplasty (4 of 21
patients treated with coil balloon followed by stent deployment) showed
that 81% (17/21) of the patients were asymptomatic. In
this cohort of 21 patients, no myocardial infarction or
revascularization procedure occurred. Angiographic
follow-up was obtained in 15 of 21 (71%) of the patients.
Angiography did not show any aneurysmatic dilatation or an
excessive restenosis rate of the site treated with the coil
balloon. Restenosis rate, according to the categorical
criterion diameter stenosis >50%, was 7% (1/15) at
follow-up. The patient presenting an angiographic diameter
stenosis of 59% on the proximal LAD was clinically
asymptomatic, with a negative exercise test.
A cautious comparison to
an angioplasty series with similar "vessel
size" (reference diameter size at diagnostic
angiography) shows an expected angiographic restenosis rate
(% diameter stenosis >50%) at 6 months of 33.3% to 37.3%
compared with 7% in our series.52 The late loss observed
in the comparative series for a reference diameter of 2.75±0.04 and
2.92±0.06 mm was far superior to that observed in this series
(0.28±0.45 to 0.26±0.46 versus 0.18±0.43 mm) (Table
5
). This
preliminary result has prompted us to undertake a multicenter open
study to confirm whether this result can be achieved in a larger cohort
of patients.
Conclusions
In conclusion, the coil balloon does not induce
significant
ischemia during prolonged intracoronary deployment
by virtue of its perfusion properties. In situ infusion of high volumes
of heparinized solution after balloon angioplasty is feasible. The use
of the device seems, however, not recommendable in case of dissections
longer than 20 mm because of the potential risk of aggravating
dissections. Long-term effects of local heparin administration did
not show any deterioration of the treated site. On the basis of these
preliminary results, data collection in an open multicenter registry
appears indicated.
| Selected Abbreviations and Acronyms |
|---|
|
| Acknowledgments |
|---|
Received July 27, 1994; revision received June 5, 1995; accepted June 8, 1995.
| References |
|---|
|
|
|---|
2.
Serruys PW, Luijten HE, Beatt KJ, Geuskens R, de
Feyter PJ, van den Brand M, Reiber JHC, ten Katen HJ, van Es GA,
Hugenholtz PG. Incidence of restenosis after
successful coronary angioplasty: a quantitative angiographic
study in 342 consecutive patients at 1, 2, 3, and 4 months.
Circulation. 1988;77:361-371.
3.
Topol E. Promises and pitfalls of new devices
for coronary artery disease.
Circulation. 1991;83:689-694.
4.
Serruys PW, de Jaegere P, Kiemeneij F, Macaya C,
Rutsch W, Heyndrickx G, Emanuelsson H, Marco J, Legrand V, Materne P,
Belardi J, Sigwart U, Colombo A, Goy JJ, van den Heuvel P, Delcan J,
Morel MA for the Benestent Study Group. A comparison of balloon
expandable stent implantation with balloon angioplasty in patients with
coronary artery disease. N Engl J
Med. 1994;331:489-495.
5.
Fischman DL, Leon MB, Baim D, Schatz RA, Penn I, Detre
K, Savege MP, Veltri L, Ricci D, Nobuyoshi M, Cleman M, Heuser R,
Almond D, Teirstein P, Fish D, Colombo A, Brinker J, Moses J, Hirshfeld
J, Bailey S, Ellis S, Rake R, Goldberg S. A randomized
comparison of coronary stent placement and balloon angioplasty
in the treatment of coronary artery disease.
N Engl J Med. 1994;331:496-501.
6.
Thornton MA, Gruentzig AR, Hollman J, King SB III,
Douglas JS. Coumadin and aspirin in prevention of
recurrence after transluminal coronary angioplasty: a
randomized study. Circulation. 1984;69:721-727.
7. Ellis SG, Roubin GS, Wilentz J, Douglas JS Jr, King SB III. Effect of 18- to 24-hour heparin administration for prevention of restenosis after uncomplicated coronary angioplasty. Am Heart J. 1989;117:777-782. [Medline] [Order article via Infotrieve]
8. Herrman JPR, Hermans WRM, Vos J, Serruys PW. Pharmacological approaches to the prevention of restenosis following angioplasty: the search for the Holy Grail? (part I). Drugs. 1993;46:18-52. [Medline] [Order article via Infotrieve]
9. Herrman JPR, Hermans WRM, Vos J, Serruys PW. Pharmacological approaches to the prevention of restenosis following angioplasty: the search for the Holy Grail? (part II). Drugs. 1993;46:249-262.
10. Clowes AW, Reidy MA, Clowes MM. Mechanisms of stenosis after arterial injury. Lab Invest. 1983;49:208-215. [Medline] [Order article via Infotrieve]
11. Clowes AW, Reidy MA, Clowes MM. Kinetics of cellular proliferation after arterial injury, I: smooth muscle cell growth in the absence of endothelium. Lab Invest. 1983;49:327-333. [Medline] [Order article via Infotrieve]
12. Austin GE, Ratliff NB, Hollman J, Tabei S, Phillips DF. Intimal proliferation of smooth muscle cells as an explanation for recurrent coronary artery stenosis after percutaneous transluminal coronary angioplasty. J Am Coll Cardiol. 1985;6:369-375. [Abstract]
13. Wolinsky H, Thung SN. Use of a perforated balloon catheter to deliver concentrated heparin into the wall of the normal canine artery. J Am Coll Cardiol. 1990;15:475-481. [Abstract]
14. Lambert CR, Leone JE, Rowland SM. Local drug delivery catheters: functional comparison of porous and microporous designs. Coron Artery Dis. 1993;4:469-475. [Medline] [Order article via Infotrieve]
15. Santoian EC, Gravanis MB, Anderberg K, Scott NA, Karas SP, Schneider JE, King SB III. Use of a porous infusion balloon in swine coronary arteries: low pressure minimizes arterial damage. Circulation. 1991;84(suppl II):II-591. Abstract.
16. Hong MK, Wong SC, Popma JJ, Kent KM, Pichard AD, Satler LF, Mintz GS, Nikonow K, Leon M. A dual-purpose angioplastydrug infusion catheter for the treatment of intragraft thrombus. Cathet Cardiovasc Diagn. 1994;32:193-195. [Medline] [Order article via Infotrieve]
17. Hong MK, Wong SC, Farb A, Mehlman MD, Virmani R, Barry JJ, Leon MB. Feasibility and drug delivery efficiency of a new balloon angioplasty catheter capable of performing simultaneous local drug delivery. Coron Artery Dis. 1993;4:1023-1027. [Medline] [Order article via Infotrieve]
18. Kaplan AV, Kermode J, Grant G, Klein E, Vetter J, Hinohara T, Simpson JB. Intramural delivery of marker agent in ex vivo and in vivo models using a novel drug delivery sleeve. J Am Coll Cardiol. 1994;23:187A. Abstract.
19.
Fernandez-Ortiz A, Meyer BJ, Mailhac A, Chesebro JH,
Badimon L, Hassinger N, Owen WG, Fuster V, Badimon JJ.
Intravascular local delivery: an iontophoretic approach.
Circulation. 1994;89:1518-1522.
20.
Lehman KG, Atwood E, Snyder EL, Ellison RL.
Autologous blood perfusion for myocardial protection during
coronary angioplasty: a feasibility study.
Circulation. 1986;76:312-323.
21.
Quigley PJ, Hinohara T, Phillips HR, Peter RH, Behar
VS, Kong Y, Simonton CA, Perez JA, Stack RS. Myocardial
protection during coronary angioplasty with an autoperfusion
balloon catheter in humans. Circulation. 1988;78:1128-1134.
22. Meester GT, Zeelenberg C, Bernard N, Gorter S. Beat to beat analysis of cardiac catheterization data. In: Computers in Cardiology. Los Angeles, Calif: IEE Computer Society; 1974:63-65.
23. Meester GT, Bernard N, Zeelenberg C, Brower RW, Hugenholtz PG. A computer system for real time analysis of cardiac catheterization data. Cathet Cardiovasc Diagn. 1975;1:113-132. [Medline] [Order article via Infotrieve]
24. Brower RW, Meij S, Serruys PW. A model of asynchronous left ventricular relaxation predicting the bi-exponential pressure decay. Cardiovasc Res. 1983;17:482-488. [Medline] [Order article via Infotrieve]
25.
Serruys PW, Wijns W, van den Brand M, Meij S, Slager C,
Schuurbiers JCH, Hugenholz PG, Brower RW. Left
ventricular performance, regional blood flow, wall
motion, and lactate metabolism during transluminal
angioplasty. Circulation. 1984;70:25-36.
26. Dellborg M, Riha M, Swedberg K. Dynamic QRS- and ST-segment changes in myocardial infarction monitored by continuous on-line vectorcardiography. J Electrocardiol. 1990;23:11-19.
27.
Sederholm M, Grottum P, Erhardt L, Kjekshus J.
Quantitative assessment of myocardial ischemia and
necrosis by continuous vectorcardiography and measurement of creatine
kinase release in patients. Circulation. 1983;68:1006-1012.
28. Slager CJ, Hooghoudt TE, Serruys PW, Schuurbiers JCH, Reiber JHC, Meester GT, Verdouw PD, Hugenholtz PG. Quantitative assessment of regional left ventricular motion using endocardial landmarks. J Am Coll Cardiol. 1986;7:317-322. [Abstract]
29. Serruys PWS, Di Mario C, Kern M. Intracoronary Doppler. In: Topol E, ed. Textbook of Interventional Cardiology. Philadelphia, Pa: WB Saunders Co; 1994:1069-1121.
30.
Doucette JW, Corl PD, Payne HM, Flynn AE, Goto M, Nassi
M, Segal J. Validation of a Doppler guide wire for
intravascular measurement of coronary artery flow
velocity. Circulation. 1992;85:1899-1911.
31.
Blackburn H, Keys A, Simonson E, Rautaharju P, Punsar
S. The electrocardiogram in population studies:
a classification system. Circulation. 1960;21:1160-1175.
32.
Ellis SG, Vandormael MG, Cowley MJ, DiSciascio G,
Deligonul U, Topol E, Bulle TM, and the Multivessel Angioplasty
Prognosis Study Group. Coronary morphologic and clinical
determinants of procedural outcome with angioplasty for multivessel
coronary disease. Circulation. 1990;82:1193-1202.
33.
Reiber JHC, Serruys PW, Kooijman CJ, Wijns W, Slager
CJ, Gerbrands JJ, Schuurbiers JCH, Den Boer A, Hugenholtz PG.
Assessment of short-, medium-, and long-term variations in
arterial dimensions from computer-assisted
quantification of coronary cineangiograms.
Circulation. 1985;71:280-288.
34. Haase J, Escaned J, Montauban van Swiijndregt E, Ozaki Y, Gronenschild E, Slager CJ, Serruys PW. Experimental validation of geometric and densitometric coronary measurements on the new generation cardiovascular angiography analysis system (CAAS II). Cathet Cardiovasc Diagn. 1993;30:104-114. [Medline] [Order article via Infotrieve]
35. Lesperance J, Hudon G, White CW, Laurier J, Waters D. Comparison by quantitative angiographic assessment of coronary stenoses of one view showing the severest narrowing to two orthogonal views. Am J Cardiol. 1989;64:462-465. [Medline] [Order article via Infotrieve]
36. Campeau L. Grading of angina pectoris. Circulation. 1976;54:522-523. [Medline] [Order article via Infotrieve]
37. Gray WA, Emin S, Williams DO. Marked diameter enlargement in polyolefincopolymer angioplasty balloons after first inflation. Circulation. 1991;84(suppl II):II-591. Abstract.
38. Van der Giessen WJ, Violaris AG, van Beusekom HMM, Di Mario C, van Meegen JR, Serruys PW, Verdouw PD. Acute local delivery of drugs with incremental molecular size in normal coronary arteries of pigs. Eur Heart J. 1994;15:559. Abstract.
39. Arnold AE, Serruys PW, Rutsch W, Simoons ML, De Bono DP, Tijssen JGP, Lubsen J, Verstraete M, for the European Cooperative Study Group. Reasons for the lack of benefit of immediate angioplasty during recombinant tissue plasminogen activator therapy for acute myocardial infarction: a regional wall motion analysis. J Am Coll Cardiol. 1991;17:11-21. [Abstract]
40. Piek JJ, Koolen JJ, Metting van Rijn AC, Bot H, Hoedemaker G, David GK, Dunning AJ, Spaan JAE, Vissier CA. Spectral analysis of flow velocity in the contralateral artery during coronary angioplasty: a new method for assessing collateral flow. J Am Coll Cardiol. 1993;21:1574-1582. [Abstract]
41.
Deutsch E, Berger M, Kussmaul WG, Hirshfeld JW Jr,
Herrmann HC, Laskey WK. Adaptation to ischemia during
percutaneous transluminal coronary angioplasty:
clinical, hemodynamic, and metabolic
features. Circulation. 1990;82:2044-2051.
42. Probst P, Zangl W, Pachinger O. Relation of coronary arterial occlusion pressure during percutaneous transluminal coronary angioplasty to presence of collaterals. Am J Cardiol. 1985;55:1264-1269. [Medline] [Order article via Infotrieve]
43.
Meier B, Luethy P, Finci L, Steffenino GD, Rutishauser
W. Coronary wedge pressure in relation to spontaneously
visible and recruitable collaterals.
Circulation. 1987;75:906-913.
44. Serruys PW, Suryapranata H, Piscione F, Harmsen E, van den Brand M, de Feyter P, Hugenholtz PG, de Jong JW. Myocardial release of hypoxanthine and lactate during percutaneous transluminal coronary angioplasty. Am J Cardiol. 1989;63:45E-51E. [Medline] [Order article via Infotrieve]
45. De Scheerder IK, Tuccilli B, Strauss BH, de Feyter PJ, Serruys PW. Hemodynamic and metabolic observations associated with intracoronary stenting for acute closure following percutaneous coronary angioplasty. J Intervent Cardiol. 1991;4:35-39.
46. Holmes DR, Bove AA, Nishimura RA, Gehring DG, Chesebro JH, Owen RM, Smith HC. Comparison of monoplane and biplane assessment of regional left ventricular wall motion after thrombolytic therapy for acute myocardial infarction. Am J Cardiol. 1987;59:793-797. [Medline] [Order article via Infotrieve]
47. Stack RS, Quigley PJ, Collins G, Phillips HR III. Perfusion balloon catheter. Am J Cardiol. 1988;61:77G-80G. [Medline] [Order article via Infotrieve]
48. De Muinck ED, Angelini P, Dougherty K, Verkerke BJ, Rakhorst G, van Dijk RB, Lie KI. In vitro evaluation of blood flow through autoperfusion balloon catheters. Cathet Cardiovasc Diagn. 1993;30:58-62. [Medline] [Order article via Infotrieve]
49. Bach RC, Kern M, Bell C, Penick D, Wolford T. Quantitation of distal coronary flow provided by perfusion balloon catheters in patients during angioplasty. Circulation. 1992;86(suppl I):I-444. Abstract.
50. Leitschuh ML, Mills RM, Jacobs AK, Ruocco NA, LaRosa D, Faxon DP. Outcome after major dissections during coronary angioplasty using the perfusion balloon catheter. Am J Cardiol. 1992;67:1056-1060.
51. Muhlestein JB, Quigley PJ, Ohman EM, Bauman RP, Sketch MH, Tcheng JE, Davidson CJ, Peter RH, Behar VS, Krucoff MW, Kong Y, Phillips HR. Prospective analysis of possible myocardial damage or hemolysis occurring as a result of prolonged autoperfusion angioplasty. J Am Coll Cardiol. 1992;20:594-598. [Abstract]
52.
Foley DP, Melkert R, Serruys PW on behalf of the
CARPORT, MERCATOR, MARCATOR, and PARK Investigators. Influence
of coronary vessel size on renarrowing process and late
angiographic outcome after successful balloon angioplasty.
Circulation. 1994;90:1239-1251.
This article has been cited by other articles:
![]() |
A. C. Ferreira, A. A. Peter, T. A. Salerno, H. Bolooki, and E. de Marchena Clinical impact of drug-eluting stents in changing referral practices for coronary surgical revascularization in a tertiary care center Ann. Thorac. Surg., February 1, 2003; 75(2): 485 - 489. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Herdeg, M. Oberhoff, A. Baumbach, S. Schroeder, M. Leitritz, A. Blattner, D. I Siegel-Axel, C. Meisner, and K. R Karsch Effects of local all-trans-retinoic acid delivery on experimental atherosclerosis in the rabbit carotid artery Cardiovasc Res, February 1, 2003; 57(2): 544 - 553. [Abstract] [Full Text] [PDF] |
||||
![]() |
J Wohrle, E Al-Khayer, U Grotzinger, C Schindler, M Kochs, V Hombach, and M Hoher Comparison of the heparin coated vs the uncoated Jostent(R)--no influence on restenosis or clinical outcome Eur. Heart J., October 1, 2001; 22(19): 1808 - 1816. [Abstract] [PDF] |
||||
![]() |
N Meneveau, F Schiele, G Grollier, B Farah, J.M Lablanche, K Khalife, J Machecourt, N Danchin, J.E Wolf, M Simpson, et al. Local delivery of nadroparin for the prevention of neointimal hyperplasia following stent implantation: results of the IMPRESS Trial. A multicentre, randomized, clinical, angiographic and intravascular ultrasound study Eur. Heart J., November 1, 2000; 21(21): 1767 - 1775. [Abstract] [PDF] |
||||
![]() |
D. S Ettenson and E. R Edelman Local drug delivery: an emerging approach in the treatment of restenosis Vascular Medicine, May 1, 2000; 5(2): 97 - 102. [Abstract] [PDF] |
||||
![]() |
M. O Hiltunen, M. P Turunen, M. Laitinen, and S. Yla-Herttuala Insights into the molecular pathogenesis of atherosclerosis and therapeutic strategies using gene transfer Vascular Medicine, February 1, 2000; 5(1): 41 - 48. [Abstract] [PDF] |
||||
![]() |
A. Gershlick Endovascular manipulation to restrict restenosis Vascular Medicine, August 1, 1998; 3(3): 177 - 188. [Abstract] [PDF] |
||||
![]() |
M. Oberhoff, S. Novak, C. Herdeg, A. Baumbach, A. Kranzhofer, A. Bohnet, B. Horch, H. Hanke, K. K. Haase, and K. R. Karsch Local and systemic delivery of low molecular weight heparin stimulates the reendothelialization after balloon angioplasty Cardiovasc Res, June 1, 1998; 38(3): 751 - 762. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. J Feldman and G. Steg Optimal techniques for arterial gene transfer Cardiovasc Res, September 1, 1997; 35(3): 391 - 404. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Camenzind, W. H. Bakker, A. Reijs, I. M. van Geijlswijk, E. Boersma, M. J. B. Kutryk, E. P. Krenning, J. R. T. C. Roelandt, P. W. Serruys, and P. W. Serruys Site-Specific Intracoronary Heparin Delivery in Humans After Balloon Angioplasty : A Radioisotopic Assessment of Regional Pharmacokinetics Circulation, July 1, 1997; 96(1): 154 - 165. [Abstract] [Full Text] |
||||
![]() |
J. Gunn, C. M. Holt, S. E. Francis, L. Shepherd, M. Grohmann, C. M. H. Newman, D. C. Crossman, and D. C. Cumberland The Effect of Oligonucleotides to c-myb on Vascular Smooth Muscle Cell Proliferation and Neointima Formation After Porcine Coronary Angioplasty Circ. Res., April 19, 1997; 80(4): 520 - 531. [Abstract] [Full Text] |
||||
![]() |
O. Tahlil, M. Brami, L. J Feldman, D. Branellec, and Ph.G. Steg The DispatchTM catheter as a delivery tool for arterial gene transfer Cardiovasc Res, January 1, 1997; 33(1): 181 - 187. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1995 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |