(Circulation. 1995;92:2210-2219.)
© 1995 American Heart Association, Inc.
Articles |
From Hippokration Hospital (C. Stefanadis, C. Stratos, C.V., S.M., I.K., E.T., K.T., L.S., P.T.), Department of Cardiology, University of Athens, Greece; and Department of Cardiology (H.B.), Ohio State University, Columbus.
Correspondence to Christodoulos Stefanadis, MD, 9 Tepeleniou St, Paleo Psychico, Athens 154 52, Greece.
| Abstract |
|---|
|
|
|---|
Methods and Results With this method, instantaneous diameter of the thoracic aorta was acquired by a newly designed intravascular catheter developed in our institution that incorporates an ultrasonic displacement meter at its distal end. Instantaneous aortic pressure was acquired simultaneously at the same aortic level with a catheter-tip micromanometer. Aortic pressure-diameter loops were derived from computer analysis of data. After in vitro and animal testing, elastic properties of the aorta were investigated in coronary artery disease (CAD) patients (n=15) and compared with those of control subjects (n=10). Aortic distensibility was less in the CAD group than in the control group (1.73±0.33 versus 3.95±1.09x10-6xcm2xdyne-1, P<.001). Compared with control subjects, the mean value of the slope of the pressure-diameter loops was significantly greater in the CAD group (38.89±8.75 versus 19.62±5.46 mm Hg · mm-1, P<.001), whereas the mean value of the intercept was lower in this latter group of patients (-785.60±177.55 versus -313.43±126.41 mm Hg, P<.001). An excellent correlation was found between the slope of pressure-diameter loop and age in the group of control subjects (r=.827). Ninety-three percent of the patients with CAD had values above the upper 95% confidence limits of the control subjects (P<.001). In a third group of patients (n=16) in whom assessment of pulse wave velocity was also included in the study of the elastic properties of the aorta, pulse wave velocity had a strong inverse correlation with aortic distensibility (r=-.95) and a strong positive correlation with the slope of the pressure-diameter loop (r=.97).
Conclusions This new method of determination of pressure-diameter of the aorta enables an accurate and reliable evaluation of the elastic properties of the aorta in conscious humans and may be useful for a profound study of human aorta mechanics.
Key Words: elasticity pressure aorta waves catheters
| Introduction |
|---|
|
|
|---|
The feasibility of a new method for the direct estimation of the elastic properties of the aorta in conscious humans with simultaneous acquisition of instantaneous aortic pressure and diameter was assessed in the present study. With this method, aortic diameters were acquired with a newly designed intravascular catheter developed in our institution that incorporates an ultrasonic displacement meter. Aortic pressures were acquired at the same aortic level with a catheter-tip micromanometer. By this method, elastic properties of the aorta were investigated in patients with coronary artery disease (CAD) and compared with those of control subjects.
| Methods |
|---|
|
|
|---|
|
|
The principle of the ultrasonic dimension technique has been described previously.7 In brief, the transit time of acoustic impulses, traveling at the sonic velocity of approximately 1.5x106 mm/s from the transmitting piezoelectric crystal to the opposing receiver crystal, is measured. A voltage proportional to transit time, and thus instantaneous dimension, is continuously recorded. Crystal alignment is verified after fabrication of the catheter with an oscilloscope (model 2120, B&K Precision) displaying the received ultrasonic signal. Similarly, continuous verification is obtained during the procedure.
The technical characteristics of the device were determined with the following performance tests.
The resolution for assessment of changes in diameter of the 5-MHz ultrasonic dimension crystals used in the present study is 10 µ, as determined by the manufacturer.
Frequency response testing of the diameter device up to 40 Hz showed a flat response.
There was no measurable phase lag between forced oscillations of the device and the signal in the frequency response range.
For evaluation of the possible mechanical impedance that the device exerts on the aorta, we measured the force that is exerted on the aortic wall by the catheter-tip arms. This force depends on the angle of the arms of the Y-shaped end of the catheter. For the model that was used in the present study, which is suitable for study of large vessels such as the aorta, the maximum force exerted is 0.45 g per arm when the distance between the arms is 1 cm. This value is less than those reported to be acceptable for the study of aortic mechanics,29 and it is thus concluded that the device does not offer significant mechanical impedance to the motions of the aorta.
Aortic pressures
Aortic pressures are simultaneously obtained with a
catheter-tip micromanometer (model SPC-330,
Millar Instruments). This high-frequency response pressure gauge
allows excellent reproduction of pressure waveforms and
avoids time delay and motion artifacts associated with fluid-filled
pressure catheters.
Procedure
For insertion of the
diameter device, a long (50 cm) 8F guiding
sheath was inserted through a 9F introducer placed through a puncture
in the right femoral artery and advanced to the level of the proximal
descending aorta. Then, the catheter-tip (with the wires collapsed)
is inserted into the guiding sheath and advanced to the descending
aorta. Once the catheter-tip is in position, the guiding sheath is
withdrawn to expose completely the Y-shaped
end of the catheter, which allows the arms to spread apart until they
touch the aortic wall and follow freely its movements during the
cardiac cycle (Fig 2
). All of these manipulations for insertion
and
positioning of the catheter are performed under fluoroscopic
observation.
The catheter-tip micromanometer (3F) is
inserted through a 5F introductory sheath placed through a puncture
into the left femoral artery. To permit recording of pressure
and dimension at the same site, the catheter-tip
micromanometer is advanced retrogradely under
fluoroscopic control, and the tip is located minimally below the exact
level of the pair of crystals to avoid distortion in the received
ultrasonic signal by presenting an obstruction in the path of the
propagated ultrasound.7 10 Fig 3
shows
the
tips of the Y-shaped catheter and the tip
of the catheter-tip micromanometer within the
thoracic descending aorta of a patient, on a fluoroscopic image (Fig
3A
) and on an image obtained with transesophageal
echocardiography (Fig 3B
).
|
Data Acquisition
and Processing
A VF-1 mainframe (Crystal Biotech) was fitted with
appropriate
modules for acquisition of aortic diameters, aortic pressures, and ECG.
Ultrasonic crystals, pressure micromanometer, and
ECG leads were connected to a 5-MHz length-gauge module (LG-510), a
dual-pressure module (BP-1), and an ECG module, respectively.
Signals of aortic pressure, aortic diameter, and ECG collected with
VF-1 mainframe are simultaneously displayed in
real-time mode on a personal computer (IBM 486 DX) with a
multichannel 12-bit analog-to-digital converter (Data
Translation Inc) and commercially available data acquisition software
(Dataflow, Crystal Biotech). Signals are digitized every 5 ms. The
sensitivity of the LG-510 length-gauge module, to which the
crystals are connected, is 100 mV/mm.
The digitized data are stored and later processed with the use of commercially available software (Microsoft Excel for Windows).
For aortic pressure and diameter measurements and subsequent calculations, approximately 10 consecutive beats are averaged.
Fig
4
shows in a schematic view the dimension and
pressure gauges positioned at the level of the proximal descending
aorta.
|
Experimental Testing
To test the feasibility, efficacy, and
safety of the
diameter-measuring device, the new method was applied in six
experimental animals (pigs of either sex; weight, 15 to 23 kg). The
animals were premedicated, anesthetized, and mechanically
ventilated as previously described.13 The investigation
conforms with the Guide for Care and Use of Laboratory
Animals published by the National Institutes of Health (NIH
Publication 85-23, revised 1985). Before insertion of the diameter
device and the catheter-tip micromanometer,
bolus heparin (100 U/kg) was administered; incremental heparin was
administered during the procedure. The diameter and pressure gauges
were inserted through surgically exposed iliac arteries and positioned
in the descending thoracic aorta as described. Recordings were
started when hemodynamic indexes (heart rate, aortic
pressure, and aortic diameter) had reached a completely steady state
after instrumentation.
The device-based method of measurement of aortic diameters was validated by comparison with the standard ultrasonic dimension-gauge method in three of the study pigs in the following manner. After baseline measurements were taken, the diameter device and the pressure micromanometer were removed. After midline laparotomy, the abdominal aorta was exposed, and a pair of free crystals identical to those used for fabrication of the device were implanted in the wall of the abdominal aorta with minimal dissection of the adventitia. Crystals were placed so that they opposed each other, and their lead wires were connected to the VF-1 mainframe for diameter measurements. Then, under fluoroscopic control, the diameter device was reinserted and positioned at a level just below that of the free crystals. When a completely steady state of hemodynamic indexes was reached, aortic diameter was measured with both the pair of free crystals and the diameter device.
On completion of measurements, the animals were killed while in deep anesthesia by administration of potassium chloride. The aorta was perfused in situ under constant pressure of 100 mm Hg with Ringer's lactate for 5 minutes and subsequently with Karnovsky's fixative fluid for 30 minutes. Transverse blocks of the thoracic aortic segments in which the catheter was placed were taken to include the entire circumference. Samples were placed in Karnovsky's fixative fluid, where they remained for 24 hours. The blocks of tissues were then cut longitudinally and separated into two parts. One half was further processed for light microscopy and the other half for scanning electron microscopy as previously described.13 30 The tissues for light microscopy were stained with hematoxylin and eosin and Masson's trichrome stains.
Clinical Study
One hundred twenty-seven consecutive male
patients who
underwent diagnostic cardiac
catheterization for evaluation of chest pain were
selected as potential subjects for the study. Patients with
arterial hypertension (systolic
arterial pressure
140 mm Hg and/or diastolic
arterial pressure
90 mm Hg), valvular heart
disease, history of previous myocardial infarction, congenital heart
disease, dilated cardiomyopathy, left
ventricular dysfunction, chronic obstructive
pulmonary disease, history of cerebrovascular accident, or
diabetes mellitus were excluded before entry into the study. With these
criteria, 25 patients were selected and divided into two groups
according to the angiographic result. Fifteen patients who had CAD
(luminal stenosis
50% in diameter) in at least one of the
major coronary arteries were included in the CAD group. Ten
patients with angiographically normal coronary arteries were
used as controls (patients with plaque disease, ie, patients with
coronary atherosclerotic lesions causing luminal
stenosis <50% in diameter, were also excluded). Treatment
with all medications except aspirin was discontinued at least five
half-lives before the study. All subjects had normal serum
electrolytes, as well as normal renal and hepatic functions.
An additional 16 patients (4 women and 12 men; mean age, 45±15 years; age range, 25 to 75 years) who underwent diagnostic catheterization and did not necessarily meet the criteria were selected regardless of the angiographic result. In this third group of patients, pulse wave velocity was determined in addition to pressure-diameter relation studies.
Approval for the study was obtained from the institutional ethical committee, and written informed consent was obtained from each patient after each was provided with a detailed description of the protocol.
Diagnostic cardiac catheterization and studies of the elastic properties of the aorta were performed during the same catheterization session. Studies were performed in the morning after an overnight fast, without premedication, in a catheterization laboratory at a controlled room temperature of approximately 22°C. First, routine coronary arteriography and left ventriculography with nonionic contrast medium (Ultravist 370, Schering AG) were performed. Next, the 7F introducer in the right femoral artery was exchanged for a 9F introducer. In addition, a 5F introducer (except for studies of pulse wave velocity, where an 8F introducer was used for combined insertion of two pressure micromanometers) was inserted through a puncture in the left femoral artery. Before insertion of the diameter device and the pressure micromanometer, the patient received an intravenous bolus injection of 100 U/kg heparin and, during the procedure, continuous infusion of heparin to maintain activated clotting time of >300 seconds. Finally, the diameter device and the pressure micromanometer were inserted and advanced to the same level of the thoracic aorta (see "Procedure").
To allow patients to relax after diagnostic catheterization and instrumentation and thus to allow hemodynamic parameters to stabilize, as well as to exclude any effect of contrast medium on the elastic properties of the aorta, baseline measurements of all hemodynamic indexes were obtained approximately 20 minutes after instrumentation and at least 30 minutes after the last infusion of contrast medium.31
In the additional 16 patients in whom determination of pulse wave velocity was also included in the study of the elastic properties of the aorta, a second catheter-tip micromanometer inserted through the same 8F introducer was advanced to the thoracic aorta, and the pressure sensor was located 10 cm below the level of the dimension crystals. After baseline measurements for the determination of pressure-diameter relation, the first catheter-tip micromanometer positioned initially to the level of the dimension crystals was advanced, and its tip was positioned 10 cm above the level of the crystals, so that the distance between the two pressure sensors was 20 cm. Then, aortic pressures were recorded simultaneously with the use of the two micromanometers on photographic paper at a speed of 150 mm · sec-1 with an Electronics for Medicine/Honeywell VR-12 device.
The possible alterations in smooth
muscle cell tone that might be
caused by the continuous contact of the arms of the catheter in
prolonged studies was tested in the following manner. To produce a wide
range of aortic pressures, three consecutive handgrip isometric tests
(with a 30-minute interval between each exercise) were performed by
four control subjects after the resting period that followed the
instrumentation. To avoid variations in heart rate, exercises were
performed during atrial pacing at a rate of 20 beats per minute above
the individual baseline heart rate of each patient. A dynamometer was
used for the isometric tests. The subjects were suitably instructed to
avoid performing the Valsalva maneuver and carefully observed during
the tests. During the 3-minute period of each consecutive isometric
exercise at 50% of maximal voluntary contraction, aortic pressure and
aortic diameter were recorded. Pulse-by-pulse values of
peak systolic aortic pressures were plotted versus peak
systolic diameters for each handgrip exercise. Likewise,
pulse-by-pulse values of diastolic aortic pressures
were plotted versus diastolic diameters (Fig 5A
through
5C). Then, for each individual, the slopes and elevations of the three
regression lines for both systolic and diastolic
values were tested for significant differences (Fig
5D
).32
|
Determination of Pressure-Diameter Relation
Stored digitized
data were processed with commercially available
computer software (Excel for Windows). Aortic pressure-diameter
relation was obtained by plotting the pressure (ordinate)- and diameter
(abscissa)-digitized data.
Calculation of Aortic Distensibility
The distensibility of
the aorta was calculated with the
following
formula11 13 15 16 17 18 21 33 :
![]() |
where
d is diastolic aortic diameter,
d is
systolic-diastolic aortic diameter (pulsatile
change in aortic diameter), and
P is
systolic-diastolic aortic pressure (pulse
pressure).
Determination of Pulse Wave Velocity
Theoretical pulse wave
velocity was calculated from the
following formula2 :
![]() |
where EP is Peterson's pressure-strain elastic modulus, given by the following equation:
![]() |
where
g is gravitational constant and
is blood density.
Measured pulse wave velocity was calculated as the ratio of the distance between the two pressure sensors to the delay time between the pressure waves recorded simultaneously with the two micromanometers as previously described.34
Data and Statistical Analyses
Data values are expressed as
mean±SD. All variables were
tested for normal distribution with the Kolmogorov-Smirnov
one-sample test. For comparison of patient characteristics between
the two groups, the unpaired t test was used. Bivariate
correlation coefficients were calculated with Pearson's
product-moment method (continuous versus continuous
variables) or with Spearman's rank method (continuous versus
discrete variables) where appropriate. Linear regression
analysis of pressure versus diameter was performed in each
patient separately to determine the slope (ie, the regression
coefficient b of the regression equation) and the intercept
(ie, the a of the regression equation) of the regression
line. These parameters, characterizing the
pressure-diameter relation and the elastic properties of the aorta,
were tested for differences between control subjects and CAD patients.
The independent relations of the above slope of the
pressure-diameter loop to its potential predictors were
analyzed with stepwise multiple linear regression, both in the
total study population (potential predictors: systolic
pressure, diastolic pressure, age, diastolic
diameter, presence or absence of CAD, and heart rate) and in the
control subjects (potential predictors: systolic pressure,
diastolic pressure, age, diastolic diameter,
and heart rate). Moreover, to test for significant differences among
the simple linear regression lines in the study for alterations in
smooth muscle tone, an overall test for coincidental regressions was
performed for the systolic and diastolic
measurements separately.32 Linear regression
analysis and the Bland-Altman method35 were used
to compare externally and internally measured aortic wall displacements
for the validation of the diameter measurement technique. Nonlinear
regression analysis was used to assess the relation between
pulse wave velocity and aortic distensibility and between pulse wave
velocity and slope of the pressure-diameter loop. A value of
P<.05 was considered statistically significant.
| Results |
|---|
|
|
|---|
In gross inspection after each procedure, no thrombi were observed on the catheter. No embolic events were observed in any of the experimental pigs. Moreover, no thrombi were observed on the endothelial surface at the site of catheter placement.
Although not superimposable, the diameter tracings with the
free
crystals (external measurement) and the diameter device (internal
measurement) were exactly similar, as shown in Fig 6A
.
There was an excellent correlation (r=1) between aortic wall
displacements measured with the two pairs of crystals (Fig 6B
).
Moreover, there was a high degree of agreement between the external and
internal aortic wall displacement (mean difference 0.0002 mm with a
95% confidence interval of -0.009 to 0.009 mm) with no obvious
relation between the difference of the two methods for measuring wall
displacement and their mean value (r=-.046; Fig
6C
).
|
Scanning electron microscopy of the endothelial surface at the site of the catheter placement disclosed neither thrombus formation nor endothelial denudation. Light microscopy revealed no internal elastic lamina or deep wall damage.
Clinical Study
The diameter-measuring device could be
inserted and positioned
easily in all patients studied. There were neither technical
difficulties nor complications. In all subjects, aortic pressure and
diameter signals of excellent quality were obtained (Fig 7
),
and clockwise pressure-diameter loops were
derived from analysis of data (Fig 8
).
|
|
The effect of
handgrip exercise on aortic pressures and diameters is
shown in Table 1
. In each of the four patients, there
were no significant differences among the three regression lines for
both systolic and diastolic values of the
consecutive handgrip tests (Fig 5D
).
|
For CAD patients
and control subjects, age, body surface area, and
heart rate were similar (Table 2
). Pulse
pressure was greater in the CAD group (P<.05) than in
the control group, whereas mean systolic and
diastolic pressures were similar. Both mean
systolic and diastolic aortic diameters were
greater in the CAD group (P<.05 and P<.01,
respectively), whereas mean pulsatile change in aortic diameter was
greater in the control group (P<.001). Aortic
distensibility was less in the CAD group compared with control subjects
(P<.001, Table 2
).
|
Representative examples
of pressure and diameter
waveforms in a control subject and a patient with CAD are depicted in
Fig 7
, whereas respective pressure-diameter relations are shown
in
Fig 8
. In Fig 7
, a reduced aortic diameter
pulsatility is observed in
the CAD patient despite the increased pulse pressure, denoting lower
distensibilty. The same association of CAD with aortic elastic
properties is denoted by the steeper slope of the pressure-diameter
loop (Fig 8
).
Compared with control subjects, the mean
value of the slope of the
pressure-diameter loops was significantly greater in the CAD group
(P<.001), whereas mean value of the intercept was lower in
this latter group of patients (P<.001, Table 2
).
Both
findings denote reduced elastic properties in the CAD patients.
In the
group of control subjects, multiple regression analysis
revealed that age was the only factor predictive of the slope of the
pressure-diameter loop (F=17.267,
R2=.683, P<.01), whereas in the
total study population, CAD and age were found to be the most
significant predictive factors (F=24.708,
R2=.832, P<.001; Table
3
).
Therefore, to interpret the values of the slope of the
pressure-diameter loop in patients with CAD, we constructed a
nomogram of this slope to age. The 95% confidence limits of the
derived regression line (slope of pressure-diameter loop versus
age, r=.827) allowed a definition of statistical range for
the values of the slope of control subjects. As shown in the
scatterplot of Fig 9
, 93% of the patients with CAD had
values above the upper 95% confidence limits of the control subjects
(P<.001).
|
|
There was no difference between the theoretical
and the measured
pulse wave velocity (7.7±3.5 versus 7.4±3.7 m/s,
P=NS).
Moreover, there was an excellent correlation between the theoretical
and the measured pulse wave velocity (r=.98,
P<.001). The measured pulse wave velocity had a strong
inverse correlation with aortic distensibility (r=-.95,
Fig 10A
). In addition, this index was strongly correlated
with the slope of the pressure-diameter linear regression line
(r=.97, Fig 10B
).
|
| Discussion |
|---|
|
|
|---|
Elastic Properties of the Aorta: Clinical
Significance
Aortic elasticity is an important component of left
ventricular afterload, constituting a major determinant of
left ventricular power output.36 37 Moreover,
aorta, by virtue of its viscoelastic properties, dampens the
intermittently generated hydraulic energy of the left ventricle along
the arterial tree and provides a continuous flow, which is
fundamental for proper metabolic exchanges of tissues. In
addition, aortic elasticity strongly influences coronary blood
flow.12 25
Determination of the Elastic Properties of the Aorta:
Methodology
Several methods have been used for the evaluation of the
regional
elastic properties of the aorta in humans and involve determination of
changes of aortic dimensions in relation to changes of aortic
pressure.15 16 17 18 19 20 21 22 23 24 25 26
However, when changes in aortic pressure
occur, it may be difficult to distinguish whether the changes of the
aortic elasticity indexes used to estimate the elastic properties are
secondary to changes of blood pressure, changes of the intrinsic
elastic properties of the aorta, or both.
When the pressure-diameter relation is studied, changes of aortic elasticity related to pressure alone, within certain limits, can be defined. Changes of aortic elasticity due to changes in aortic pressure alone after an intervention will result in sliding of the aortic pressure-diameter loop upward or downward along the same hypothetical sigmoid curve of elasticity, whereas changes of the intrinsic elastic properties of the vessel will result in shifting of the loop either to the left or to the right. In addition, the rate of change of the aortic diameter during systole or diastole in health and disease states and after therapeutic interventions can be studied. Moreover, study of the pressure-diameter relation of the aorta provides insight into the viscoelasticity of the vessel. Pressure-diameter (or volume) relation of the aorta has been studied in experimental animals,1 2 3 5 6 8 9 10 whereas human data have been acquired postmortem38 and during open heart surgery.15 Owing to methodological limitations, data regarding the study of pressure-diameter relation of the human aorta in conscious humans are limited.27 28 These limitations are mainly related to the fact that recordings must be instantaneous and simultaneous, leading in most of the cases to an invasive acquisition of pressure and diameter tracings. With most of the devices used until now, it has been difficult to avoid laparotomy or thoracotomy for measurement of the pressure-diameter relation of the aorta. However, it is of importance that anesthesia, thoracotomy, recent surgery, and acute manipulation of the vessel all modify responses of arterial smooth muscle to a variety of interventions,4 7 39 inducing a systematic error in the measurements and thus imposing the need for acquiring data in conscious humans. Another important factor for the reliable determination of the pressure-diameter relation is the simultaneous acquisition of data (pressure and diameter) at the same level of the aorta.
Imura et al27 developed an apparatus for the determination of the elastic properties of the aorta that consists of an ultrasonic displacement meter for transcutaneous diameter measurement and a catheter-tip micromanometer for pressure measurement. Reported results are interesting; however, the limited approachability to all levels of the aorta with the ultrasonic displacement meter is an obvious limitation of the technique.
Recently, Lang et al28 reported a method for the determination of regional elastic properties of the human aorta that involves the combination of transesophageal echocardiography with automated border detection and calibrated subclavian pulse tracings. The technique is very promising and may provide new insights into the physiopathology of disease states of the aorta. However, not all sites of the aorta are accessible with transesophageal echocardiography. Moreover, acquisition of aortic pressure at a different level than that of aortic diameter may affect the pressure-diameter relation.1 40 Finally, in the study of Lang et al, a remarkable percentage of patients was excluded owing to either inadequate detection of the aortic area or poor quality of the subclavian tracing.
Diameter-Measuring Device
Performance and experimental
testing determined the
accuracy of the diameter device, whereas experimental and clinical
studies proved the feasibility, effectiveness, and safety of the
apparatus for the determination of pressure-diameter
relation. In addition, it was demonstrated that indexes of aortic
elasticity measured with this apparatus correlate well with
established indexes such as pulse wave velocity. With this method,
estimation of the elastic properties of the aorta can be obtained in
conscious humans during an ordinary heart
catheterization, thus avoiding the influence of factors
such as anesthesia, thoracotomy, acute manipulation of the
aorta, and so on. The most important aspect of our method is that
aortic diameters and pressures are measured simultaneously
and accurately at the same point. When the distance between the
pressure and diameter measuring points exceeds a certain limit, the
hysteresis in the loop may be reduced or even abolished.1
Besides, it has been reported that there is a change of pressure wave
amplitude between central and peripheral arteries during
drug studies, or when changes occur in heart rate or in relative
content of ascending aorta pressure wave harmonics.40
Therefore, although measuring diameter and pressure at different levels
may not be a problem in screening studies, it is apparent that it may
lead to errors when the detection of subtle changes is required. The
diameter device is also characterized by high sensitivity, which is
essential in measuring pulsatile diameter change in blood vessel
because the change is very small. Moreover, the device can be used for
estimation of the aortic elastic properties at different levels to
obtain a "mapping" of the distensibility of the aorta.
Association of CAD and Age With the Elastic Properties of the
Aorta
With the use of our new method, it was demonstrated that CAD is
a
potent independent factor associated with aortic elastic
performance. This is in accordance with previous studies from
our laboratory16 17 18 22 and
other
laboratories20 25 26 in which it was
demonstrated that
elasticity of the aorta is unfavorably affected in the presence of CAD.
Deterioration of aortic elastic properties can be attributed to
mechanical effect of atherosclerotic lesions and/or abnormal nutrition
of the aortic
wall.11 13 16 17 22
Moreover, the results of the present study showed that age was an independent factor determining aortic elastic performance, both in control subjects and in the total study population. These findings are in agreement with previous studies showing that distensibility of the aorta decreases with age.14 26 Structural and macroscopic alterations of the aortic wall that are observed with advancing age34 41 may be responsible for the deterioration of the elastic performance of the vessel.
Specific Comments: Study Limitations
The possible effect of
constant contact by the arms on smooth
muscle tone in prolonged studies was studied during repeated handgrip
exercises in the same individual. Our study indicates that there is no
smooth muscle response to the prolonged contact of the aortic wall by
the arms of the device.
A factor limiting the wide application of the technique is its invasive nature. However, this method is best suited to the detection of minute changes of the elastic properties of the aorta and not to the screening of patients, for which other noninvasive methods are more suitable.
Insertion of a relatively large introductory sheath into the femoral artery may lead to some to question the safety of the diameter-measuring device regarding arterial trauma. However, these sheaths are generally considered by interventional cardiologists to be safe. Also, in our total study population, no complications were encountered, thus confirming the safety of the diameter-measuring device. However, refinement of the device will allow for the use of smaller sheaths to minimize the risk of arterial damage.
Perspectives
More than 100 subjects were studied in several
experimental
protocols. The applicability of this device in the study of the elastic
properties of the aorta appears to be wide. The dynamic elastic modulus
may describe the elastic properties of the vessel; however, it contains
no information about the viscous, or time-dependent, behavior of
the vessel wall. Information of this nature is provided by estimation
of the components of the complex modulus, which in turn can be obtained
by analysis of instantaneous aortic pressures and diameters
recorded simultaneously and at the same
point.42 Thus, a more profound study of aortic elastic
properties can be obtained by the newly developed apparatus
in conditions in which elastic properties of the aorta have been found
to be affected, such as advanced
age14 23 26 or essential
hypertension,21 23 as well as in many others in which
these properties are expected to be altered, such as diabetes mellitus,
diseases of the connective tissue, and so on. Moreover, the effect of
several pharmacological agents can be studied, whereas insights can be
gained into the potential mechanisms by which these drugs act. The high
sensitivity of the device makes it ideal for the study of subtle
changes in aortic elastic properties before the underlying disease
becomes clinically evident. Moreover, further refinement of the
diameter device will allow its application to smaller caliber vessels
with a view to a more extensive study of the elastic properties of the
arterial tree.
In conclusion, the newly developed diameter-measuring device enables an accurate and reliable determination of the elastic properties of the aorta in conscious humans and may serve as a useful means for a profound study of human aorta mechanics.
| Acknowledgments |
|---|
Received January 30, 1995; revision received April 18, 1995; accepted May 3, 1995.
| References |
|---|
|
|
|---|
2.
Patel DJ, DeFreitas FM, Greenfield JC Jr, Frey DL.
Relationship of radius to pressure along the aorta in living
dogs. J Appl Physiol. 1963;18:1111-1117.
3. Pieper HP, Paul LT. Responses of aortic smooth muscle studied in intact dogs. Am J Physiol. 1969;217:154-160.
4.
Aars H. Effects of altered smooth muscle tone
on aortic diameter and aortic baroreceptor activity in
anesthetized rabbits. Circ Res. 1971;28:254-262.
5. Nicolosi GR, Pieper HP. Aortic smooth muscle responses to changes in venous return studied in intact dogs. Am J Physiol. 1971;221:1209-1216.
6. Pagani M, Schwartz PJ, Bishop VS, Malliani A. Reflex sympathetic changes in aortic diastolic pressure-diameter relationship. Am J Physiol. 1975;229:286-290.
7. Pagani M, Baig H, Sherman A, Manders T, Quinn P, Patrick T, Franklin D, Vatner S. Measurement of multiple simultaneous small dimensions and study of arterial-pressure dimension relations in conscious animals. Am J Physiol. 1978;235:H610-H617.
8. Dujardin JP, Paul LT, Pieper HP. Effect of acute volume loading of aortic smooth muscle activity in intact dogs. Am J Physiol. 1980;238:379-383.
9. Yano M, Kumada T, Matsuzaki M, Kohno M, Hiro T, Kohtoku S, Miura T, Katayama K, Ozaki M, Kusukawa R. Effect of diltiazem on aortic pressure-diameter relationship in dogs. Am J Physiol. 1989;25:H1580-H1587.
10. Cabrera Fischer EI, Armentano RL, Levenson J, Barra JG, Morales MC, Breitbard GJ, Pichel RH, Simon A. Paradoxically decreased aortic wall stiffness in response to vitamin D3-induced calcinosis: a biphasic analysis of segmental elastic properties in conscious dogs. Circ Res. 1991;68:1594-1559.
11.
Stefanadis C, Karayannacos P, Boudoulas H, Stratos C,
Vlachopoulos C, Dontas I, Toutouzas P. Medial necrosis and acute
alterations in aortic distensibility following removal of the vasa
vasorum of canine ascending aorta. Cardiovasc Res. 1993;27:951-956.
12. Ohtsuka S, Kakihana M, Watanabe H, Sigishita Y. Chronically decreased aortic distensibility causes deterioration of coronary perfusion during increased left ventricular contraction. J Am Coll Cardiol. 1994;24:1406-1414. [Abstract]
13.
Stefanadis C, Vlachopoulos C, Karayannacos P, Boudoulas
H, Stratos C, Fillipides T, Agapitos M, Toutouzas P. Effect of
vasa vasorum flow on function and structure of the aorta.
Circulation. 1995;91:2669-2678.
14.
Avolio AP, Chen S, Wang R, Zhang C, Li M, O'Rourke MF.
Effects of aging on changing arterial compliance and
left ventricular load in a northern Chinese urban
community. Circulation. 1983;68:50-58.
15.
Greenfield JC, Patel DJ. Relation between
pressure and diameter in the ascending aorta of man.
Circ Res. 1962;10:778-781.
16. Stefanadis C, Wooley CF, Bush CA, Kolibash AJ, Boudoulas H. Aortic distensibility abnormalities in coronary artery disease. Am J Cardiol. 1987;59:1300-1304. [Medline] [Order article via Infotrieve]
17. Stefanadis C, Wooley CF, Bush CA, Kolibash AJ, Boudoulas H. Aortic distensibility in post-stenotic aortic dilatation: the effect of co-existing coronary artery disease. J Cardiol. 1988;18:189-195. [Medline] [Order article via Infotrieve]
18.
Stefanadis C, Stratos C, Boudoulas H, Kourouklis C,
Toutouzas P. Distensibility of the ascending aorta: comparison
of invasive and noninvasive techniques in healthy men and in men with
coronary artery disease. Eur Heart J. 1990;11:990-996.
19. Hirata K, Triposkiadis F, Sparks E, Bowen J, Wooley CF, Boudoulas H. The Marfan syndrome: abnormal elastic properties. J Am Coll Cardiol. 1991;18:57-63. [Abstract]
20. Dart AM, Lacombe F, Yeoh JK, Cameron JD, Jennings GL, Laufer E, Esmore DS. Aortic distensibility in patients with isolated hypercolesterolaemia, coronary artery disease, or cardiac transplant. Lancet. 1991;338:270-273. [Medline] [Order article via Infotrieve]
21. Stratos C, Stefanadis C, Kallikazaros I, Boudoulas H, Toutouzas P. Ascending aorta distensibility abnormalities in hypertensive patients and response to nifedipine administration. Am J Med. 1992;93:505-512. [Medline] [Order article via Infotrieve]
22.
Stefanadis C, Stratos C, Boudoulas C, Vlachopoulos C,
Kallikazaros I, Toutouzas P. Distensibility of the ascending
aorta in coronary artery disease and changes after
nifedipine administration. Chest. 1994;104:1017-1023.
23. Isnard RN, Pannier BM, Laurent S, London GM, Diebold B, Safar M. Pulsatile diameter and elastic modulus of the aortic arch in essential hypertension: a noninvasive study. J Am Coll Cardiol. 1989;13:399-405. [Abstract]
24. Pasiersky TJ, Binkley PF, Pearson A. Evaluation of aortic distensibility with transesophageal echocardiography. Am Heart J. 1992;123:1288-1292. [Medline] [Order article via Infotrieve]
25. Bogren HG, Mohiaddin RH, Klipstein RK, Firmin DN, Underwood RS, Rees SR, Longmore DB. The function of the aorta in ischemic heart disease: a magnetic resonance and angiographic study of aortic compliance and blood flow patterns. Am Heart J. 1989;118:234-237. [Medline] [Order article via Infotrieve]
26.
Mohiaddin RH, Underwood SR, Bogren HG, Firmin DN,
Klipstein RH, Rees RSO, Longmore DB. Regional aortic compliance
studied by magnetic resonance imaging: the effects of age, training,
and coronary artery disease. Br Heart J. 1989;62:90-96.
27.
Imura T, Yamamoto K, Satoh T, Kanamori K, Mikami T,
Yasuda H. In vivo viscoelastic behavior of the human
aorta. Circ Res. 1990;66:1413-1419.
28.
Lang RM, Cholley BP, Korcarz C, Marcus RH, Shroff SG.
Measurement of regional elastic properties of the human aorta: a
new application of transesophageal
echocardiography with automated border detection
and calibrated subclavian pulse tracings.
Circulation. 1994;90:1875-1882.
29.
Mallos AJ. An electrical caliper for continuous
measurement of relative displacement. J Appl
Physiol. 1962;17:131-134.
30. Hayat MA. Introduction to Biological Scanning Electron Microscopy. Baltimore, Md: University Park Press; 1978:91-166, 207-238.
31. Baron MG. Radiological and angiographic examination of the heart. In: Braunwald E, ed. Heart Disease. Philadelphia, Pa: WB Saunders; 1988:140-179.
32. Zar JH. Biostatistical Analysis. Englewood Cliffs, NJ: Prentice-Hall; 1984:304.
33. Caro CG, Pedley TJ, Scroten RC, Seed WA. The Mechanics of the Circulation. Oxford, UK: Oxford University Press; 1978:86-105, 243-349.
34. Nichols WW, O'Rourke MF. Properties of the arterial wall. In: Nichols WW, O'Rourke MF, eds. McDonald's Blood Flow in Arteries. London, UK: Edward Arnold; 1990:77-124.
35. Bland J, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307-310. [Medline] [Order article via Infotrieve]
36. Urschel CW, Covell JW, Sonnenblick EH, Ross J, Braunwald E. Effects of decreased aortic compliance on performance of the left ventricle. Am J Physiol. 1968;214:298-304.
37. Binkley PF, Boudoulas H. Measurement of myocardial inotropy. In: Leier CV, ed. Cardiotonic Drugs: A Clinical Review. New York, NY: Marcel Dekker; 1991:37-41.
38. Hallock P, Benson IC. Studies in the elastic properties of human isolated aorta. J Clin Invest. 1937;16:595-602.
39. Vatner SF, Braunwald E. Cardiovascular control mechanisms in the conscious state. N Engl J Med. 1975;293:970-976. [Medline] [Order article via Infotrieve]
40.
Karamanoglu M, O'Rourke MF, Avolio AP, Kelly
RP. An analysis of the relationship between central
aortic and peripheral upper limp pressure waves in
man. Eur Heart J. 1993;14:160-167.
41. Schlatmann TJ, Becker AE. Histologic changes in the normal aging aorta: implication for dissecting aneurysms. Am J Cardiol. 1977;39:13-20. [Medline] [Order article via Infotrieve]
42. Milnor WR. Hemodynamics. Baltimore, Md: Williams and Wilkins; 1989:58-101.
This article has been cited by other articles:
![]() |
C. Vlachopoulos, F. Kosmopoulou, N. Alexopoulos, N. Ioakeimidis, G. Siasos, and C. Stefanadis Acute mental stress has a prolonged unfavorable effect on arterial stiffness and wave reflections. Psychosom Med, March 1, 2006; 68(2): 231 - 237. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. R. Heerman, P. Segers, C. D. Roosens, F. Gasthuys, P. R. Verdonck, and J. I. Poelaert Echocardiographic assessment of aortic elastic properties with automated border detection in an ICU: in vivo application of the arctangent Langewouters model Am J Physiol Heart Circ Physiol, May 1, 2005; 288(5): H2504 - H2511. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. B. Santana, J. G. Barra, J. C. Grignola, F. F. Gines, and R. L. Armentano Pulmonary artery smooth muscle activation attenuates arterial dysfunction during acute pulmonary hypertension J Appl Physiol, February 1, 2005; 98(2): 605 - 613. [Abstract] [Full Text] [PDF] |
||||
![]() |
G J Nollen, B E Westerhof, M Groenink, A Osnabrugge, E E van der Wall, and B J M Mulder Aortic pressure-area relation in Marfan patients with and without {beta} blocking agents: a new non-invasive approach Heart, March 1, 2004; 90(3): 314 - 318. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Vlachopoulos, K. Hirata, and M. F O'Rourke Effect of sildenafil on arterial stiffness and wave reflection Vascular Medicine, November 1, 2003; 8(4): 243 - 248. [Abstract] [PDF] |
||||
![]() |
J. J. Oliver and D. J. Webb Noninvasive Assessment of Arterial Stiffness and Risk of Atherosclerotic Events Arterioscler. Thromb. Vasc. Biol., April 1, 2003; 23(4): 554 - 566. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Hager, H. Kaemmerer, U. Rapp-Bernhardt, S. Blucher, K. Rapp, T. M. Bernhardt, M. Galanski, and J. Hess Diameters of the thoracic aorta throughout life as measured with helical computed tomography J. Thorac. Cardiovasc. Surg., June 1, 2002; 123(6): 1060 - 1066. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. P. Cholley, R. M. Lang, C. E. Korcarz, and S. G. Shroff Smooth muscle relaxation and local hydraulic impedance properties of the aorta J Appl Physiol, June 1, 2001; 90(6): 2427 - 2438. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Stefanadis, C. Chrysochoou, D. Markou, K Petraki, D. B. Panagiotakos, C. Fasoulakis, A. Kyriakidis, C. Papadimitriou, and P. K. Toutouzas Increased Temperature of Malignant Urinary Bladder Tumors In Vivo: The Application of a New Method Based on a Catheter Technique J. Clin. Oncol., February 1, 2001; 19(3): 676 - 681. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Katz, D. Payne, and S. Pauker Early Surgery versus Conservative Management of Dissecting Aneurysms of the Descending Thoracic Aorta Med Decis Making, October 1, 2000; 20(4): 377 - 390. [Abstract] [PDF] |
||||
![]() |
C. Stefanadis, J. Dernellis, E. Tsiamis, L. Diamantopoulos, A. Michaelides, and P. Toutouzas Assessment of Aortic Line of Elasticity Using Polynomial Regression Analysis Circulation, April 18, 2000; 101(15): 1819 - 1825. [Abstract] [Full Text] [PDF] |
||||
![]() |
C Stefanadis, J Dernellis, E Tsiamis, C Stratos, L Diamantopoulos, A Michaelides, and P Toutouzas Aortic stiffness as a risk factor for recurrent acute coronary events in patients with ischaemic heart disease Eur. Heart J., March 1, 2000; 21(5): 390 - 396. [Abstract] [PDF] |
||||
![]() |
H. Smulyan and M. E. Safar The Diastolic Blood Pressure in Systolic Hypertension Ann Intern Med, February 1, 2000; 132(3): 233 - 237. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Stefanadis, L. Diamantopoulos, C. Vlachopoulos, E. Tsiamis, J. Dernellis, K. Toutouzas, E. Stefanadi, and P. Toutouzas Thermal Heterogeneity Within Human Atherosclerotic Coronary Arteries Detected In Vivo : A New Method of Detection by Application of a Special Thermography Catheter Circulation, April 20, 1999; 99(15): 1965 - 1971. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. E. Hardt, A. Just, R. Bekeredjian, W. Kubler, H. R. Kirchheim, and H. F. Kuecherer Aortic pressure-diameter relationship assessed by intravascular ultrasound: experimental validation in dogs Am J Physiol Heart Circ Physiol, March 1, 1999; 276(3): H1078 - H1085. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Stefanadis, E. Tsiamis, J. Dernellis, and P. Toutouzas Effect of estrogen on aortic function in postmenopausal women Am J Physiol Heart Circ Physiol, February 1, 1999; 276(2): H658 - H662. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. Khan, R. W. Millard, M. Gabel, R. A. Walsh, and B. D. Hoit Effect of congestive heart failure on in vivo canine aortic elastic properties J. Am. Coll. Cardiol., January 1, 1999; 33(1): 267 - 272. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Stefanadis, J. Dernellis, E. Tsiamis, C. Stratos, I. Kallikazaros, and P. Toutouzas AORTIC FUNCTION IN PATIENTS DURING INTRA-AORTIC BALLOON PUMPING DETERMINED BY THE PRESSURE-DIAMETER RELATION J. Thorac. Cardiovasc. Surg., December 1, 1998; 116(6): 1052 - 1059. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Stefanadis, J. Dernellis, M. Vavuranakis, E. Tsiamis, C. Vlachopoulos, Konstantinos Toutouzas, L. Diamandopoulos, C. Pitsavos, and P. Toutouzas Effects of ventricular pacing-induced tachycardia on aortic mechanics in man Cardiovasc Res, August 1, 1998; 39(2): 506 - 514. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Stefanadis, J. Dernellis, C. Stratos, E. Tsiamis, C. Vlachopoulos, K. Toutouzas, S. Lambrou, C. Pitsavos, and P. Toutouzas Effects of balloon mitral valvuloplasty on left atrial function in mitral stenosis as assessed by pressure-area relation J. Am. Coll. Cardiol., July 1, 1998; 32(1): 159 - 168. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Stefanadis, J. Dernellis, C. Vlachopoulos, C. Tsioufis, E. Tsiamis, K. Toutouzas, C. Pitsavos, and P. Toutouzas Aortic Function in Arterial Hypertension Determined by Pressure-Diameter Relation : Effects of Diltiazem Circulation, September 16, 1997; 96(6): 1853 - 1858. [Abstract] [Full Text] |
||||
![]() |
C. Stefanadis, E. Tsiamis, C. Vlachopoulos, C. Stratos, K. Toutouzas, C. Pitsavos, S. Marakas, H. Boudoulas, and P. Toutouzas Unfavorable Effect of Smoking on the Elastic Properties of the Human Aorta Circulation, January 7, 1997; 95(1): 31 - 38. [Abstract] [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1995 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |