(Circulation. 1995;92:1634-1643.)
© 1995 American Heart Association, Inc.
Articles |
From the Department of Medicine, University of Washington, Seattle, and the Philadelphia Heart Institute (S.M.D.), Presbyterian Medical Center, Philadelphia, Pa.
Correspondence to Bradford E. Gliner, c/o Gust H. Bardy, MD, Electrophysiology Section, Mail Stop RG-22, University Hospital, Seattle, WA 98195.
| Abstract |
|---|
|
|
|---|
Methods and Results Three interrelated studies were performed in 19 swine to establish the relative transthoracic defibrillation efficacy of biphasic shock waveforms. In study 1, we measured voltage (V50) and energy (E50) strength-duration curves for monophasic and biphasic truncated exponential waveforms. We then independently examined the effects of phase duration and tilt on biphasic waveform defibrillation with a total waveform duration from study 1 that provided the minimum V50 (study 2) and the minimum E50 (study 3). At each pulse duration tested in study 1, biphasic waveforms defibrillated with significantly less voltage and energy than monophasic waveforms. At a duration of 12 ms, there was a voltage minimum for biphasic waveform defibrillation. At this duration, V50 was 1378±505 V for the biphasic waveform compared with 2185±361 V for the monophasic waveform, P=.01. For both monophasic and biphasic waveforms, E50 increased with pulse duration. With a total pulse duration of 12 ms, E50 was 169±101 J for the biphasic waveform compared with 414±114 J for the monophasic waveform, P=.003. In study 2, optimization of phase duration and total tilt reduced the defibrillation requirements of the 12-ms "minimum voltage" biphasic waveform to 1284±187 V and 129±36 J. In study 3, the 8-ms "minimum energy" biphasic waveform had an E50 of 115±35 J that was 11% less than the 12-ms biphasic waveform, P=.11; however, voltage requirements of 1476±239 V were 15% higher, P=.005.
Conclusions This study demonstrates the superiority of truncated biphasic waveforms over truncated monophasic waveforms for transthoracic defibrillation of swine. Biphasic waveforms should prove as advantageous at reducing voltage and energy requirements for transthoracic defibrillation as they have for internal defibrillation.
Key Words: fibrillation defibrillation waves death, sudden
| Introduction |
|---|
|
|
|---|
Just as studies of defibrillation waveforms using internal electrodes have promoted advancements in internal defibrillator therapy, eg, lower defibrillation requirements and higher defibrillation safety margins, we anticipate that the results of this study will allow similar advances to be made for external defibrillators. To determine whether this is possible, we compared the performance of biphasic truncated exponential waveforms with that of monophasic truncated exponential waveforms. We also tested a clinically used monophasic damped sine waveform. This waveform gave us the basis for the prediction that two of the biphasic waveforms tested, one chosen for minimum voltage and the other for minimum energy, might be practically applied to human transthoracic defibrillation.
| Methods |
|---|
|
|
|---|
Defibrillation efficacy was compared through determination of V50 and E50 for each of the waveforms. V50 and E50 are the peak shock voltage and total delivered energy, respectively, required to produce a 50% likelihood that the shock will defibrillate. In study 1, we measured V50 and E50 strength-duration curves for monophasic and biphasic truncated exponential waveforms. The next two studies were explorations of the relative defibrillation efficacies of two different biphasic waveforms that, on the basis of their total duration, were found in study 1 to require either a minimum V50 or a minimum E50. In study 2, the "minimum voltage" biphasic waveform was examined, whereas study 3 tested the "minimum energy" biphasic waveform. In both of these latter studies, the total biphasic waveform duration was held fixed while the total waveform tilt and the phase duration were independently varied in a randomized and interleaved manner.
Sample Size
A minimum of six animals were used in each study.
With six
animals, an approximately 25-J difference in E50 (130-V
difference in V50) could be detected with 80% power
(ß=.20) and 95% confidence (
=.05,
two-sided).12
Animal Preparations
We studied 19 healthy domestic swine
weighing 57±6 kg (range,
49 to 68 kg). We chose the swine model because the thorax and
coronary arteries better approximate human anatomy than
do canine models.13 14 15 General
anesthesia was
induced with pentobarbital sodium (15 mg/kg IV to effect) after
preanesthesia by ketamine (20 mg/kg IM) and
xylazine (2 mg/kg IM) injection. We maintained the animals in a
surgical plane of anesthesia throughout the study using a
combination of ketamine (2 to 3 mg/kg), diazepam (0.1 mg/kg),
and pentobarbital sodium (2 to 4 mg/kg) every 20 minutes. The carotid
artery was catheterized, and arterial gases and systemic
blood pressure were monitored throughout the study. The swine were
intubated with a cuffed endotracheal tube and ventilated with oxygen by
a respiration pump (model 607, Harvard Apparatus) adjusted
as needed to maintain normal arterial blood gases. A
jugular vein was cannulated for administration of anesthetic and
placement of electrodes for the induction of ventricular
fibrillation (VF). Arterial blood pressure and surface ECG
lead II were monitored and allowed to return to baseline values between
inductions of VF. Normal saline was infused as needed to support
baseline arterial blood pressure. Body temperature was
maintained at 39°C by a water blanket (model K-20, Baxter Healthcare
Corp, Pharmaseal Division). The animals were paralyzed with
succinylcholine before induction of VF.
Fibrillation-Defibrillation Episodes
We applied self-adhesive
Fast-Patch defibrillation pads with
a surface area of 82 cm2 (Physio-Control Corp) to the
shaved thorax in an anterior-anterior position similar to the
position used clinically on human beings. The lower left anterior
electrode was always the anode for monophasic defibrillation waveforms
and for the first phase of biphasic waveforms. The ECG and blood
pressure were allowed to stabilize, and these parameters
were taken as baseline values. VF was induced by a 60-Hz current
applied for
1 to 2 seconds through a bipolar catheter located in the
right ventricle. After induction of VF, ventilation was discontinued at
end expiration and fibrillation continued for 15 seconds from the
beginning of induction before defibrillation was attempted.
Defibrillation waveforms were generated by a custom defibrillator
(Bradford E. Gliner, Seattle, Wash) designed to deliver monophasic and
biphasic truncated exponential waveforms. If the first defibrillation
attempt was not successful, we delivered rescue shocks (see below)
until defibrillation occurred. Ventilation was then resumed. We
included only the first defibrillation attempt per fibrillation episode
in the waveform efficacy analysis.
Defibrillation Waveforms
The monophasic and biphasic
truncated waveforms used in this
study are similar to those used by others in animal and clinical
studies of internal
defibrillation1 2 3 4 5 6 7 8 16 17 18 19 20 21 22
and are
represented in Fig 1A
and 1B
. In addition,
we also evaluated, to a limited extent, the defibrillation efficacy of
a damped-sine-wave shock (Fig 1C
) in an effort to provide a
foundation for future clinical studies. Damped-sine-wave shocks
are the most commonly used waveforms in transthoracic
defibrillation. This waveform was examined in reference to the clinical
standard energy setting of 200 J for monophasic waveforms as
recommended by the American Heart Association.23
|
Fig
1
plots the time course of the shock voltage for typical
monophasic
and biphasic truncated exponential waveforms and the damped sine
waveform used in study 1. These waveforms were generated by the
controlled discharge of a single capacitor. The duration of the
monophasic waveform in this example is 8 ms. Another important
parameter of interest is the waveform tilt, which for the
monophasic waveform is given by the following equation:
![]() |
The tilt quantifies how steeply the monophasic waveform discharges and also gives an indication of how much of the stored capacitor charge is delivered. The higher the waveform tilt, the greater the electrical efficiency of the shock-generating mechanism, because more of the stored charge is used for defibrillation. However, as we (see "Results") and others16 17 18 24 25 have shown, a high waveform tilt results in greater delivered voltage and energy for defibrillation. The waveform tilt is dependent on the product of the device capacitance and the impedance of the transthoracic discharge pathway. Because the impedance will vary from animal to animal and from shock to shock, we adjusted the device capacitance to achieve the desired waveform tilt for each shock.
The voltage time course of a typical
biphasic waveform is plotted in
Fig 1B
. Like the monophasic waveform, it is generated by the
discharge
of a single capacitor. The total tilt of the biphasic waveform is given
by the following equation:
![]() |
The
duration of this biphasic waveform is given by a convention
adopted in this study, which ignores the dead time caused by switching
between the shock phases. The biphasic waveform duration in Fig
1B
is
thus 8 ms, with 4 ms per phase, even though a total duration of 8.5 ms
elapses from beginning to end of the waveform. We do not think that the
0.5-ms delay between biphasic shock phases affected these experiments,
because a study of shock phase separation for internal defibrillation
showed no effect for delays <8 to 10 ms.19 20 An
additional parameter, phase duration, is needed to
characterize the biphasic waveform. Phase duration is the percentage of
the total waveform duration occupied by the first shock phase. For
example, the biphasic waveform in Fig 1B
has a phase duration
of 50%,
because the duration of the first phase is 4 ms and the total duration
of the waveform is 8 ms, as measured in this report.
We also tested a
monophasic damped sine waveform to reproduce a
waveform widely used in clinical transthoracic
defibrillation. Fig 1C
shows an example of a typical
damped-sine-wave shock used in this study. This waveform was
generated by the discharge of a 50-µF capacitor through a 30-mH,
10-
inductor. As in standard defibrillation, only the shock voltage
of this waveform was changed, with other parameters that
affected its wave shape held constant. However, peak delivered voltage,
delivered energy, and waveform duration changed with load impedance
because of the passive nature of the wave-shaping elements. No
commercially available damped-sine-wave
transthoracic defibrillator is available to vary the shock
strength according to the protocol required in this study.
Consequently, the custom device built for this study was used to test
defibrillation efficacy of the monophasic damped sine waveform as well.
However, the charge voltage of this device was limited to 3200 V (200 J
delivered to a 50-
load) because of the circuit design requirements
needed to generate truncated waveforms.
The shock strengths recorded in this study were given in terms of both the peak applied voltage and the delivered energy. To do this, the voltage and current waveforms of the applied shock were recorded at 10 000 samples per second (model Lab-NB I/O board, National Instruments). The peak shock voltage was measured from the recording, and the total energy delivered was computed on the basis of the peak delivered voltage and the capacitance used to generate the shock. The energy computation was done separately for each of the two phases of the biphasic waveform. The defibrillation impedance was also reported as the mean of the ratio of the shock voltage to the current for voltage and current computed point by point throughout the discharge.
V50 and E50 Estimation
For each of the
waveforms used in studies 1, 2, and 3, we
estimated V50 using an up-and-down protocol with
three reversals.26 27 28 29
In this protocol, only the results
of the first shocks were considered for the V50 estimate;
subsequent rescue shocks delivered to defibrillate the animal were not
included in the analysis. The protocol began by attempting
defibrillation with a shock strength estimated to be equal to
V50 for the waveform under study. Subsequent waveforms
studied used the cumulative average V50 value from
preceding experiments for the first defibrillation attempt. The
succeeding shock voltages were incremented or decremented by 20%,
depending on failure or success of the previous episode, until a
reversal from failure to success or from success to failure occurred.
We then changed the voltage by 10% until we obtained a total of three
reversals for each waveform. We estimated V50 by taking the
average of the voltages delivered, excluding the first attempt and
including the next step after the third reversal. We defined this
average as the V50 if it was within 20% of the voltage
used for the first attempt; otherwise, we defined V50 as
the average of the voltages at each reversal. The first definition is
more accurate when the initial estimate of V50 is close to
the true V50, whereas the alternative definition was
used to reduce bias when the initial estimate was not close. We also
expressed defibrillation requirements in terms of E50.
We allowed respiration, arterial pressure, and surface ECG to return to baseline values, with a minimum of 2 minutes between the end and the start of successive episodes. Defibrillation was defined as conversion from ventricular fibrillation within 3 seconds after the shock without any additional intervention. Studies 1 through 3 were conducted independently with different animals. Also, all of the tested waveforms were randomly interleaved during the defibrillation threshold determination within each study to mitigate the effects of spontaneous shifts in defibrillation threshold.
Experimental Protocols
The experimental protocols were
designed to accomplish two
goals: first, to determine whether transthoracic
defibrillation could be accomplished by use of a biphasic shock
waveform with less voltage and energy than with monophasic waveforms
and second, to explore the sensitivity of biphasic defibrillation
efficacy to variations in waveform tilt and phase duration. The first
goal was pursued not only to determine whether biphasic waveform
defibrillation is as efficacious when applied externally as when
applied internally but also to compare biphasic waveform defibrillation
with the performance of a standard clinical damped sine
waveform. These goals were achieved by use of the protocols outlined in
study 1 (below).
The second goal of this work was also twofold: to
further characterize
transthoracic biphasic waveform defibrillation and to
identify which combination of waveform parameters resulted
in either a minimum voltage or a minimum energy defibrillation
waveform. Studies 2 and 3 (below) accomplish the second goal by testing
the shock durations yielding the minimum voltage and minimum energy
biphasic waveforms found in study 1, respectively. Table 1
summarizes the waveform parameters
investigated in each study.
|
Study 1
The first study
determined the V50 and
E50 strength-duration curves of truncated monophasic
and biphasic waveforms. The total tilt of both waveforms was set to
80%, and the phase duration of the biphasic waveforms was set to 50%.
The phase duration was chosen on the basis of reports indicating that
the first phase should be at least 50% of the total
duration.4 21 The total tilt of the biphasic and
monophasic waveforms was selected to achieve electrical efficiency, ie,
utilization of stored electrical charge, without incurring
low-voltage tails at the end of the waveform, which have been
reported to decrease defibrillation
efficacy.16 17 18 19 24 25
If we required the shock generation process to be at least 95%
efficient, then an 80% tilt waveform would be an appropriate choice,
since it results in 96% of the capacitor charge being delivered. We
examined total waveform durations of 5.3, 8, 12, 18, and 27 ms (each
duration increased by 50%). Fig 2
illustrates these
monophasic and biphasic waveforms in which, for the sake of clarity,
the switching delay between phases has been eliminated. The damped sine
waveform (see Fig 1C
) was also tested in this series of
experiments.
The shocks were randomly selected and interleaved. A 2200-V, 8-ms
biphasic waveform was typically used for rescue defibrillation. Six
swine were studied in this portion of the investigation.
|
Study 2
In this study, two series of experiments
were conducted with the
total biphasic waveform duration fixed while either the phase duration
(part A) or the total tilt (part B) was changed. The waveform duration
was set to 12 ms, which required the least voltage for defibrillation
in study 1 (see "Results"). In part A, the phase duration was
varied from 30% to 70% in 10% increments while the total tilt was
kept at 80%. In part B, the total tilt was varied as 60%, 70%, 80%,
90%, 95%, and 99% while phase duration was kept at 50%. These
waveforms are illustrated in Fig 3
. The shocks for parts
A and B were randomly selected and interleaved. Rescue shocks in study
2 were typically 12-ms, 1800-V biphasic waveforms. Six swine were
studied in this portion of the investigation.
|
Study 3
As in study 2, in this study the total biphasic waveform
duration was fixed while either the phase duration (part A) or the
total tilt (part B) was changed. The waveform duration was set to 8 ms,
which required the minimum energy for defibrillation in study 1 (see
"Results"). In part A, phase duration was varied from 40% to
70% in 10% increments while the total tilt was kept at 80%. In part
B, the total tilt was varied as 70%, 80%, 90%, and 95% while phase
duration was kept at 50%. These waveforms are also illustrated in Fig
3
, the difference from those used in study 2 being the shorter
total
duration. Also included in study 3 was the minimum voltage 12-ms
biphasic waveform having a 60% phase duration and an 80% total tilt.
The results from study 2 indicated that this waveform will defibrillate
with minimum delivered voltage (see "Results"). Defibrillation by
a monophasic damped sine waveform, the same as that used in study 1,
was also examined. This allowed an overall comparison of the minimum
voltage and minimum energy biphasic defibrillation waveforms with this
traditional defibrillation waveform to be made. Again, shocks from
parts A and B were randomly selected and interleaved. Rescue shocks in
study 3 were typically 12-ms, 1800-V biphasic waveforms. Seven swine
were studied in this portion of the investigation.
Statistical Analysis
Values are expressed as mean±SD.
Statistical comparisons were
made with ANOVA and paired t tests with JMP
3.0.2 software (SAS Institute). Comparisons having values of
P<.05 were considered statistically significant.
| Results |
|---|
|
|
|---|
12 ms. With a total duration of 12 ms,
V50 was 1378±505 V for the biphasic waveform compared with
2185±361 V for the monophasic waveform (P=.01) and
>2500 V
for the damped sine waveform (P=.0003).
|
Fig
4B
shows that biphasic waveforms defibrillated with
significantly
less energy than monophasic waveforms at each duration
(P=.0007). Because we were not able to defibrillate with
many of the monophasic waveforms at maximum device voltage, we reported
the maximum energy tested as E50. The maximum energy tested
for the damped sine waveform was 180 J, which was not high enough for
an E50 estimation. Thus, the E50 for some
monophasic waveforms and the damped sine waveform was underestimated in
this study. The defibrillation requirements for swine proved higher
than anticipated. We were not able to predict the higher requirements
from the literature because previous relevant work has been done only
in dogs or small pigs. For both monophasic and biphasic waveforms,
E50 increased with duration. With a total duration of 12
ms, E50 was 169±101 J for the biphasic waveform compared
with 414±114 J for the monophasic waveform (P=.003) and
>180 J for the damped sine waveform. Mean defibrillation impedance was
42±6
for study 1 animals.
Study 2
Figs 5
and 6
show the
characteristic
curves for the 12-ms biphasic waveform. Fig 5
shows a voltage
minimum
and energy minimum with a phase duration of 60%. However,
defibrillation voltage and energy requirements with a phase duration of
50% were not statistically different from the requirements with a
phase duration of 60% (P=.3 for V50,
P=.4 for E50). Defibrillation requirements with
a 40% phase duration were significantly higher than those with a 60%
phase duration (P=.005 for V50,
P=.005 for E50).
|
|
As total tilt increased
from 60% to 90%, defibrillation voltage
requirements increased gradually, as Fig 6A
shows. For tilts
>90%,
V50 increased sharply. A total tilt of 95% had
significantly higher voltage requirements than 90% total tilt
(P=.02). We were not able to estimate V50 for a
99% tilt waveform with a capacitor voltage limit of 3.2 kV. Because of
the smaller capacitor values, as total tilt increases from 60% to
90%, defibrillation energy requirements decreased gradually, as Fig
6B
shows. For tilts >90%, E50 increased because of the
sharply increased voltage requirements; however, the energy difference
for 90% and 95% tilt did not reach statistical significance
(P=.3). Mean defibrillation impedance was 41±3
for
study 2 animals.
Study 3
Figs 7
and 8
show the
characteristic
curves for the 8-ms biphasic waveform. Although Fig 4B
shows
the 5.2-ms
waveform to be the minimum energy biphasic waveform from study 1, the
8-ms waveform was chosen for this study for practical reasons. This is
because voltage requirements of the 5.2-ms waveform were higher than
deemed suitable for a clinically usable biphasic
transthoracic defibrillator. The effect of phase duration
on the voltage and energy was not as great for the 8-ms waveform as it
was for the 12-ms waveform, as Fig 7
shows. However, there
appears to
be a trend for a voltage and energy minimum at a phase duration of 50%
(V50, P=.11 for 40% versus 50%,
P=.08 for 50% versus 60% phase duration;
E50, P=.09 for 40% versus 50%,
P=.06 for 50% versus 60% phase duration).
|
|
As total tilt
increased from 70% to 95%, defibrillation voltage
requirements of the 8-ms biphasic waveform increased sharply, as Fig
8A
shows. Because of the smaller capacitor values, as total tilt increased
from 60% to 90%, defibrillation energy requirements did not change
(Fig 8B
). For 95% total tilt, energy requirements increased
significantly because of the sharply increased voltage requirements
(P=.01 versus 90% total tilt). Mean defibrillation
impedance was 44±3
for study 3 animals.
| Discussion |
|---|
|
|
|---|
14%
higher. For monophasic truncated waveforms, increasing total duration
reduced V50 slightly but increased E50
significantly.
|
From study 3, the 8-ms minimum energy biphasic waveform
with 50% phase
duration and 80% total tilt required 15% more voltage and 11% less
energy than the 12-ms minimum voltage biphasic waveform with 60% phase
duration and 80% total tilt (see Table 3
for
comparison). The minimum voltage biphasic waveform
required <60% of the capacitor voltage of the monophasic damped sine
waveform, and the minimum energy biphasic waveform required <50% of
the stored energy of the monophasic damped sine waveform.
|
Other Biphasic Waveform Transthoracic
Defibrillation Studies
The first study of transthoracic defibrillation
with a
biphasic waveform was conducted on dogs by Gurvich and
Markarychev.11 They showed that lower peak currents were
needed for defibrillation with a biphasic damped sine waveform than
with a monophasic damped sine waveform. The investigators also
demonstrated a higher therapeutic index for biphasic over monophasic
damped sine waveforms by showing a greater ratio between the energy
required to produce cardiac damage and the energy required for
defibrillation. Schuder et al30 31 first compared
monophasic and biphasic rectangular waveforms for
transthoracic defibrillation. Consistent with our
findings, they found that in calves, biphasic rectangular waveforms
defibrillated with greater success and at lower energies than
monophasic rectangular waveforms. Biphasic waveforms were also
associated with a more rapid return to a normal perfusing rhythm after
the shock than monophasic waveforms.
Two different animal studies failed to find an advantage of biphasic shocks over monophasic shocks for use in transthoracic defibrillation. Kerber et al32 found that dual-pathway sequential (biphasic) shocks did not improve shock success or reduce energy requirements compared with standard single-shock transthoracic defibrillation. Their study differs from ours in the animal model, the defibrillation pad position, and the type of defibrillation waveforms, which may affect the results. The conical shape of the canine thorax compared with the rounder swine thorax alters the current pathway and may affect the results. They used three or four defibrillation pads to deliver sequential shocks; thus, as the authors suggested, a more bidirectional approach may be more effective. Also, as acknowledged by the authors, the long intershock separation of 100 ms may not be optimum, especially compared with the 0.5-ms shock separation used in this study.19 20
Also in contrast to our results, Scott et al33 did not find an improvement in defibrillation success or lower defibrillation energy requirements with biphasic transthoracic shocks in dogs. Differences from our study again include the animal model, the defibrillation pad location, and the delivered waveform. The conical shape of the canine thorax compared with the rounder swine thorax and a left chest and right chest lateral defibrillation pad position compared with a lower left and upper right pad position may affect the results. In addition, we controlled the total tilt of the waveforms by adapting capacitance, whereas they used two 150-µF capacitors and controlled the duration of the pulses. It is well known that effective biphasic defibrillation can be achieved with the second-phase leading-edge voltage set to only a fraction of that of the first phase.6 7
Walcott et al34 compared monophasic and biphasic truncated waveforms and a monophasic damped sine waveform for external defibrillation of beagles. Like our study, theirs also found that the biphasic waveforms required less peak voltage and total energy than monophasic waveforms. However, in contrast to our study, they found a relatively low delivered energy requirement for the damped sine waveform. The size and anatomic geometry of the animal models may have contributed to the differences observed. The ratio of defibrillation pad area to animal size is much greater for the Walcott study and may affect the defibrillation current pathway. Also, the capacitance and inductance of the damped sine waveform were not reported by Walcott. Differences in these circuit parameters may produce damped sine waveforms of different efficacies.35 36
The only clinical comparative study of transthoracic
monophasic and biphasic waveforms was conducted by Echt et
al.37 They compared the efficacy of a 200-J hybrid
biphasic damped sine waveform (Gurvich waveform) with that of a 200-J
monophasic damped sine waveform (Edmark waveform) for cardioversion and
defibrillation of induced ventricular arrhythmias
during electrophysiology studies. They demonstrated a higher
probability of cardioversion/defibrillation with the biphasic Gurvich
waveform (98.3% versus 85.5%, P<.02). Although the
biphasic Gurvich waveform had a higher efficacy, the capacitor voltage
(
5900 V) and stored energy requirements (
280 J) are high, and the
waveform circuit requires a large wave-shaping inductor (80 mH)
that is energy-inefficient; ie,
28% of the stored energy is
lost in the wave-shaping. However, the greater efficacy of the
biphasic Gurvich waveform over the monophasic Edmark waveform in human
transthoracic defibrillation suggests that biphasic
truncated exponential waveforms might also be applied beneficially in
the clinical setting.
Limitations
In study 1, which demonstrated the superiority of
biphasic
waveforms, we compared a single type of biphasic shock waveform with a
single type of monophasic shock waveform. The tested monophasic
waveforms were chosen to have the same duration, energy content, and
electrical efficiency as the comparable biphasic waveform. Although
scientifically rigorous, this experimental design did not ensure that
we used the best possible monophasic waveform for testing against our
candidate biphasic waveforms. For example, there is reason to believe
that modest (about 13%) reductions in delivered voltage thresholds for
the monophasic waveforms could be attained. This indication comes from
the work of Bourland et al,38 who empirically derived an
equation describing the relation between peak shock current and the
shock duration and tilt for externally applied monophasic shocks. This
equation predicts that truncating the 12-ms monophasic waveform early,
after 6 ms, would lead to a 13.2% reduction in the delivered voltage
threshold. Alternatively, a 13% reduction could be predicted to occur
if a larger capacitor were used to produce a 55% tilt, 12-ms
monophasic waveform. Even against these predictions, however, the 12-ms
biphasic waveform would still require significantly less voltage,
27% less than the monophasic.
Another consideration is that our exploration of biphasic waveform parameter sensitivities independently tested the effects of phase duration, total tilt, and total duration. Independent parameter variation allowed us to characterize first-order effects in a practical, yet comprehensive, study. A more exhaustive study would be required to determine the interdependence of the waveform parameters (second-order effects). However, because changes in the waveform parameters gradually affected defibrillation efficacy, second-order effects are probably negligible.
Hypothetical Biphasic Waveforms for Clinical
Transthoracic Defibrillation
One of the benefits of this study is that
our comparison of
biphasic and damped sine waveforms makes it possible to estimate the
defibrillation energy requirements of these biphasic waveforms in
humans. Furthermore, the results from our exploration of waveform
parameter sensitivity could be used to tailor biphasic
waveforms for use in the clinical setting.
The results of study 1 show
that the minimum voltage and energy
biphasic waveforms could, at first approximation, be selected on the
basis of shock duration. The choices for waveform tilt can be made on
the basis of a pragmatic consideration, the transthoracic
impedance, which can be expected to vary from patient to patient and
from shock to shock. The typical patient impedance is 75 to 80
,
with 50
at the 5th percentile and 120
at the 95th
percentile.39 40 41 Although this study
compensated for the
variations in transthoracic impedance by capacitor
selection to maintain a given tilt, a clinical device will use a single
capacitance; therefore, the effects of tilt variation must be
anticipated.
There are some practical guidelines for choosing the
nominal tilt
of the hypothetical biphasic waveform. Our study suggests that we do
not want a total tilt >90%, because defibrillation voltage
requirements increased sharply. If we are constrained to a maximum tilt
of 90%, then accommodating a relatively low patient impedance of 50
would require a 104-µF capacitor for the minimum voltage biphasic
waveform. For the typical patient impedance of 80
, a nominal 76%
tilt waveform would result; the high-impedance patient of 120
would receive a 62% tilt waveform. A 69-µF capacitor would be
appropriate for the 8-ms minimum energy biphasic waveform. In this
example of a waveform of fixed capacitance and duration, tilt is
typically 76% and varies between 62% and 90%, depending on patient
impedance.
Another interesting consequence of differing transthoracic
impedance is that peak current, in addition to tilt, varies with
patient impedance. For the low-impedance patient of 50
, peak
current would increase by 60% over the typical 80-
patient. This
might be of additional benefit and compensate for any loss in efficacy
due to a higher tilt. At the other extreme, the high-impedance
patient of 120
would have a peak current 33% less than the typical
80-
patient; however, the total tilt would be low at 62%, which
might compensate for the lower peak current. The net effect may be that
there is peak current compensation for the low-impedance patient
and total tilt compensation for the high-impedance patient.
The characteristics of the defibrillation waveform may also be actively adjusted in response to a real-time measurement (ie, during delivery of the waveform) of a patient-dependent electrical parameter. For example, on the basis of the impedance of each shock, the total duration, phase duration, and total tilt of the waveform may be adjusted for each shock delivered. For example, biphasic waveforms with relatively longer first phases may have better conversion properties than waveforms with equal or shorter first phases, provided that the total duration exceeds a critical minimum. Therefore, in the case of high-impedance patients, it may be desirable to increase not only the total duration but also the phase duration of the biphasic waveform to increase the overall defibrillation efficacy. Because impedance varies among patients and from shock to shock, such a patient-specific technique may have a further advantage of being able to deliver the most efficacious waveform for a given capacitor size charged to a predetermined voltage.
Several caveats must be considered in the estimation of human
defibrillation requirements from the experimental data obtained in this
study. First, the swine in our study required more energy to
defibrillate than people typically require. For example, with 10 to 15
seconds of VF, a 200-J monophasic damped sine waveform has a
probability of defibrillation of
86% in people37 and
<50% in our swine. The reason for the higher energy requirement is
not known. Although there is evidence that the efficacy of the
monophasic damped sine waveform decreases at higher
impedances,39 40 41 the mean
defibrillation impedance of the
animals was 42
compared with a typical human defibrillation
impedance of 80
. If impedance were the cause of the difference,
defibrillation efficacy would be higher in the animals than in human
beings, which is not the case. Perhaps energy differences can be
attributed to physiological and
pathophysiological differences of the heart
itself or to differences in the interface between defibrillation
electrode and skin.
Finally, we used V50 and E50 to describe defibrillation requirements, which, while useful in comparing many different waveforms, is not representative of the high probability of defibrillation required for clinical application. The scaling of the V50 and E50 to the high probabilities needed for human transthoracic defibrillation remains to be studied. However, estimates can be made from our data with the following equation, which assumes that the probabilities of defibrillation for damped sine and biphasic waveforms scale equally as a function of energy.
![]() |
This equation estimates the energy required for clinical application of a biphasic waveform, Eclin,bi, by scaling the recommended first-shock energy of the monophasic damped sine waveform, Eclin,ds, by the ratio of E50 estimates for the biphasic waveform, Eswin,bi, and the monophasic damped sine waveform, Eswine,ds, from our animal model. In our study, we were not able to estimate E50 for Eswine,ds because of the high voltage requirements; however, we know that it is greater than the maximum energy tested, 180 J. We conservatively use 200 J for Eswine,ds. Because the clinically recommended first-shock energy is 200 J, Eswine,ds and Eclin,ds cancel, and we are left with the following identity.
![]() |
This identity is valid only within the context of this study, ie, using relatively large swine and estimating the energy of a waveform that has the same clinical efficacy as a 200-J damped sine waveform.
The above equation, in conjunction with the data presented in
Table 3
and Figs 5 through
8![]()
![]()
![]()
, enables us to
estimate the clinical
defibrillation voltage and energy requirements of the hypothetical
minimum voltage and minimum energy transthoracic biphasic
shock waveforms. These values, along with the waveform
parameters, are given in Table 4
. Note that
both of these waveforms are predicted to defibrillate with markedly
less energy than the 200 J recommended for human defibrillation with
the monophasic damped sine waveform and that the peak delivered
voltages are much less than those developed by the damped sine
waveform. These differences in delivered energy and voltage not only
would be expected to benefit the patient but also portend the
possibility of constructing small, lightweight external defibrillators.
This goal is important because an essential element to increasing
resuscitation rates for victims of sudden cardiac arrest will be the
institution of a wider network of defibrillator-equipped first
responders capable of reaching victims within the first few minutes
after collapse. The miniaturization and cost reduction of automatic
external defibrillators are critical to realizing this goal of reducing
response times to <6 minutes from time of collapse, which is typically
required for
survival.42 43 44 45 46 47
If biphasic waveforms prove to
be as advantageous at reducing energy and voltage requirements in
transthoracic defibrillation as they have been in internal
defibrillation, the size, weight, and cost of present
transthoracic defibrillators could be reduced
substantially. If automatic external defibrillators can be made small,
lightweight, and less costly, they will become a more appropriate
lifesaving tool for a much broader group of emergency first
responders.
|
| Acknowledgments |
|---|
| Footnotes |
|---|
Received January 18, 1995; accepted March 17, 1995.
| References |
|---|
|
|
|---|
2. Winkle RA, Mead RH, Ruder MA, Gaudiani V, Buch WS, Pless B, Sweeney M, Schmidt P. Improved low energy defibrillation efficacy in man with the use of a biphasic truncated exponential waveform. Am Heart J. 1989;117:122-127. [Medline] [Order article via Infotrieve]
3. Fain ES, Sweeney MB, Franz MR. Improved internal defibrillation efficacy with a biphasic waveform. Am Heart J. 1989;117:358-364. [Medline] [Order article via Infotrieve]
4.
Swartz JF, Fletcher RD, Karasik PE.
Optimization of biphasic waveforms for human nonthoracotomy
defibrillation. Circulation. 1993;88:2646-2654.
5.
Dixon EG, Tang ASL, Wolf PD, Meador JT, Fine MJ,
Calfee RV, Ideker RE. Improved defibrillation thresholds with
large contoured epicardial electrodes and biphasic waveforms.
Circulation. 1987;76:1176-1184.
6. Chapman PD, Vetter JW, Souza JJ, Wetherbee JN, Troup PJ. Comparison of monophasic with single and dual capacitor biphasic waveforms for nonthoracotomy canine internal defibrillation. J Am Coll Cardiol. 1989;14:242-245. [Abstract]
7. Kavanagh KM, Tang ASL, Rollins DL, Smith WM, Ideker RE. Comparison of the internal defibrillation thresholds for monophasic and double and single capacitor biphasic waveforms. J Am Coll Cardiol. 1989;14:1343-1349. [Abstract]
8. Feeser SA, Tang ASL, Kavanagh KM, Rollins DL, Smith WM, Wolf PD, Ideker RE. Strength-duration and probability of success curves for defibrillation with biphasic waveforms. Circulation. 1990;82:2129-2141.
9. Jones JL, Jones RE. Improved defibrillator safety factor with biphasic waveforms. Am J Physiol. 1983;245:H60-H65.
10. Jones JL, Jones RE. Decreased defibrillator-induced dysfunction with biphasic rectangular waveforms. Am J Physiol. 1984;247:H792-H796.
11. Gurvich NL, Makarychev VA. Defibrillation of the heart with biphasic electrical impulses. Kardiologia. 1967;7:109-112.
12. Armitage P, Berry G. Statistical Methods in Medical Research. Oxford, England: Blackwell Scientific Publications; 1987:179-185.
13. Howe BB, Fehn PA, Pensinger PR. Comparative anatomical studies of the coronary arteries of canine and porcine hearts. Acta Anat. 1968;71:13-21. [Medline] [Order article via Infotrieve]
14. Schaper W, Jageneau A, Xhonneux R. The development of collateral circulation in the pig and dog heart. Cardiologia. 1967;51:321-335.
15. Bowman TA, Hughes HC. Swine as an in vivo model for electrophysiologic evaluation of cardiac pacing parameters. PACE Pacing Clin Electrophysiol. 1984;7:187-194. [Medline] [Order article via Infotrieve]
16. Bardy GH, Dolack GL, Poole JE, Kudenchuk PJ, Johnson G, Raitt MH, Mehra R, DeGroot P, Hofer BO. A prospective randomized comparison in humans of 50% vs 65% tilt biphasic pulse defibrillation using the unipolar pectoral transvenous defibrillation system. Circulation. 1993;88(suppl I):I-113. Abstract.
17. Li HG, Yee R, Mehra R, DeGroot P, Zardini M, Morillo CA, Thakur RK, Klein GJ. The effect of biphasic shock waveform tilt upon defibrillation efficacy. Circulation. 1993;88(suppl I):I-113. Abstract.
18. Natale A, Van Hout W, DeGroot P, Deshpande S, Axtell K, Krum D, Sheth M, Sra J. Effects of tilt on biphasic defibrillation efficacy in patients. Circulation. 1993;88(suppl I):I-114. Abstract.
19. Cooper RAS, Wallenius ST, Smith WM, Ideker RE. The effect of phase separation on biphasic waveform defibrillation. PACE Pacing Clin Electrophysiol. 1993;16:471-482. [Medline] [Order article via Infotrieve]
20. Cooper RAS, Wallenius ST, Smith WM, Ideker RE. The effect of temporal separation of phases on biphasic waveform defibrillation efficacy. Proc IEEE-EMBS. 1991;13:766-767.
21. Tang ASL, Yabe S, Wharton M, Dolker M, Smith WM, Ideker RE. Ventricular defibrillation using biphasic waveforms: the importance of phasic duration. J Am Coll Cardiol. 1989;13:207-214. [Abstract]
22.
Feeser SA, Tang ASL, Kavanagh KM, Rollins DL, Smith WM,
Wolf PD, Ideker RE. Strength-duration and probability of
success curves for defibrillation with biphasic waveforms.
Circulation. 1990;82:2128-2141.
23.
American Heart Association, Emergency Cardiac Care
Committee. Guidelines for cardiopulmonary
resuscitation and emergency cardiac care. JAMA. 1992;268:2199-2241.
24.
Schuder JC, Rahmoeller GA, Stoeckle H.
Transthoracic ventricular defibrillation
with triangular and trapezoidal waveforms. Circ
Res. 1966;19:689-694.
25. Schuder JC, Stoeckle H, West JA, Keskar PY. Transthoracic ventricular defibrillation in the dog with truncated and untruncated exponential stimuli. IEEE Trans Biomed Eng. 1971;18:410-415. [Medline] [Order article via Infotrieve]
26. McDaniel WC, Schuder JC. The cardiac ventricular defibrillation threshold: inherent limitations in its application and interpretation. Med Instrum. 1987;21:170-176. [Medline] [Order article via Infotrieve]
27. McDaniel WC, Schuder JC. An up-down algorithm for estimation of the cardiac ventricular defibrillation threshold. Med Instrum. 1988;22:286-292.
28. Dixon WJ. The up-and-down method for small samples. Am Stat Assoc J. 1965;60:967-978.
29. Brownlee KA, Hodges JL, Rosenblatt M. The up-and-down method with small samples. Am Stat Assoc J. 1953;48:262-277.
30. Schuder JC, Gold JH, Stoeckle H, McDaniel WC, Cheung KN. Transthoracic ventricular defibrillation in the 100 kg calf with symmetrical one cycle bidirectional rectangular wave stimuli. IEEE Trans Biomed Eng. 1983;30:415-422. [Medline] [Order article via Infotrieve]
31. Schuder JC, McDaniel WC, Stoeckle H. Defibrillation of 100 kg calves with asymmetrical, bidirectional rectangular pulses. Cardiovasc Res. 1984;18:419-426. [Medline] [Order article via Infotrieve]
32. Kerber RE, Bourland JE, Kallok MJ, Hite P, Pritchard B, Charbonnier F, Birkett C, Fox-Eastham K, Kieso R. Transthoracic defibrillation using sequential and simultaneous dual shock pathways: experimental studies. PACE Pacing Clin Electrophysiol. 1990;13:207-217. [Medline] [Order article via Infotrieve]
33. Scott BD, Kallok MJ, Birkett C, Keiso RA, Kerber RE. Transthoracic defibrillation: effect of dual-pathway sequential pulse shocks and single pathway biphasic pulse shock in a canine model. Am Heart J. 1993;125:99-109. [Medline] [Order article via Infotrieve]
34. Walcott GP, Hagler JA, Walker RG, Morgan C, Swanson DK, Smith WM, Ideker RE. Comparison of monophasic, biphasic, and the Edmark waveform for external defibrillation. PACE Pacing Clin Electrophysiol. 1992;15:563. Abstract.
35.
Peleska B. Optimal parameters of
electrical impulses for defibrillation by condenser discharges.
Circ Res. 1966;18:10-17.
36. Wilson CM, Bailey A, Alen JD, Anderson J, Adgey AAJ. Transthoracic defibrillation threshold of sine and trapezoidal waveforms in defibrillation. J Electrocardiol. 1989;22:241-247. [Medline] [Order article via Infotrieve]
37. Echt DS, Greene HL, Kudenchuk PJ, Dimarco JP, Tang AS, Reiter MJ, Scheinman MM, Chapman PD, Akhtar M, Gliner B. Biphasic waveform is more efficacious than monophasic waveform for transthoracic cardioversion. PACE Pacing Clin Electrophysiol. 1993;16:914A. Abstract.
38. Bourland JD, Tacker WA, Geddes LA. Strength-duration curves for trapezoidal waveforms of various tilts for transchest defibrillation in animals. Med Instrum. 1978;12:38-41. [Medline] [Order article via Infotrieve]
39.
Kerber RE, Kouba CK, Martins J, Kelly K, Low R, Hoyt R,
Ferguson D, Balley L, Bennett P, Charbonnier F. Advance
prediction of transthoracic impedance in human
defibrillation and cardioversion: importance of impedance in
determining the success of low-energy shocks.
Circulation. 1984;70:303-308.
40.
Kerber RE, Martins JB, Kienzle MG, Constantin L,
Olshansky B, Hopson R, Charbonnier F. Energy, current, and
success in defibrillation and cardioversion: clinical studies using an
automated impedance-based method of energy adjustment.
Circulation. 1988;77:1038-1046.
41. Dalzell GWN, Cunningham SR, Anderson J, Adgey AA. Electrode pad size transthoracic impedance and success of external ventricular defibrillation. Am J Cardiol. 1989;64:741-744. [Medline] [Order article via Infotrieve]
42. Eisenberg MS, Horwood BT, Cummins RO, Reynolds-Haertle R, Hearne TR. Cardiac arrest and resuscitation: A tale of 29 cities. Ann Emerg Med. 1990;19:179-186. [Medline] [Order article via Infotrieve]
43. Cobb LA, Weaver WD, Fahrenbruch CE, Hallstrom AP, Copass MK. Community-based interventions for sudden cardiac death: impact, limitations, and changes. Circulation. 1992;85(suppl I):I-98-I-102.
44. Weaver WD, Copass MK, Hill DL, Fahrenbruch C, Hallstrom AP, Cobb LA. Cardiac arrest treated with a new automatic external defibrillator by out-of-hospital first responders. Am J Cardiol. 1986;57:1017-1021. [Medline] [Order article via Infotrieve]
45.
Eisenberg MS, Bergner L, Hallstrom A. Cardiac
resuscitation in the community: importance of rapid provision and
implications for program planning. JAMA. 1979;241:1905-1907.
46. Roth R, Stewart RD, Rogers K, Cannon GM. Out-of-hospital cardiac arrest: factors associated with survival. Ann Emerg Med. 1984;13:237-243. [Medline] [Order article via Infotrieve]
47.
Eisenberg M, Bergner L, Hallstrom A. Paramedic
programs and out-of-hospital cardiac arrest: factors associated
with successful resuscitations. Am J Public Health. 1979;69:30-38.
This article has been cited by other articles:
![]() |
K Kajino, T Iwami, R A Berg, A Hiraide, Y Hayashi, H Yukioka, H Tanaka, T Shimazu, and H Sugimoto Comparison of neurological outcomes following witnessed out-of-hospital ventricular fibrillation defibrillated with either biphasic or monophasic automated external defibrillators Emerg. Med. J., July 1, 2009; 26(7): 492 - 496. [Abstract] [Full Text] [PDF] |
||||
![]() |
P Jones and N Lode Ventricular fibrillation and defibrillation Arch. Dis. Child., October 1, 2007; 92(10): 916 - 921. [Abstract] [Full Text] [PDF] |
||||
![]() |
Part 3: Defibrillation Circulation, November 29, 2005; 112(22_suppl): III-17 - III-24. [Full Text] [PDF] |
||||
![]() |
W. Tang, M. H. Weil, S. Sun, D. Jorgenson, C. Morgan, K. Klouche, and D. Snyder The effects of biphasic waveform design on post-resuscitation myocardial function J. Am. Coll. Cardiol., April 7, 2004; 43(7): 1228 - 1235. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Zhang, R. S. Ramabadran, K. A. Boddicker, I. Bawaney, L. R. Davies, M. B. Zimmerman, S. Wuthrich, J. L. Jones, and R. E. Kerber Triphasic waveforms are superior to biphasic waveforms for transthoracic defibrillation: Experimental studies J. Am. Coll. Cardiol., August 6, 2003; 42(3): 568 - 575. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Schneider, P. R. Martens, H. Paschen, M. Kuisma, B. Wolcke, B. E. Gliner, J. K. Russell, W. D. Weaver, L. Bossaert, and D. Chamberlain Multicenter, Randomized, Controlled Trial of 150-J Biphasic Shocks Compared With 200- to 360-J Monophasic Shocks in the Resuscitation of Out-of-Hospital Cardiac Arrest Victims Circulation, October 10, 2000; 102(15): 1780 - 1787. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. T. Niemann, D. Burian, D. Garner, and R. J. Lewis Monophasic versus biphasic transthoracic countershock after prolonged ventricular fibrillation in a swine model J. Am. Coll. Cardiol., September 1, 2000; 36(3): 932 - 938. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Tang, M. H. Weil, S. Sun, H. Yamaguchi, H. P. Povoas, A. M. Pernat, and J. Bisera The effects of biphasic and conventional monophasic defibrillation on postresuscitation myocardial function J. Am. Coll. Cardiol., September 1, 1999; 34(3): 815 - 822. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Yamanouchi, J. E. Brewer, K. F. Olson, K. A. Mowrey, T. N. Mazgalev, B. L. Wilkoff, and P. J. Tchou Fully Discharging Phases : A New Approach to Biphasic Waveforms for External Defibrillation Circulation, August 24, 1999; 100(8): 826 - 831. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. A. Pagan-Carlo, J. J. Allan, K. T. Spencer, C. L. Birkett, R. Myers, and R. E. Kerber Encircling overlapping multipulse shock waveforms for transthoracic defibrillation J. Am. Coll. Cardiol., December 1, 1998; 32(7): 2065 - 2071. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Yamanouchi, J. E. Brewer, K. A. Mowrey, A. M. Donohoo, B. L. Wilkoff, and P. J. Tchou Optimal Small-Capacitor Biphasic Waveform for External Defibrillation : Influence of Phase-1 Tilt and Phase-2 Voltage Circulation, December 1, 1998; 98(22): 2487 - 2493. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. P. Walcott, S. B. Melnick, F. W. Chapman, J. L. Jones, W. M. Smith, and R. E. Ideker Relative Efficacy of Monophasic and Biphasic Waveforms for Transthoracic Defibrillation After Short and Long Durations of Ventricular Fibrillation Circulation, November 17, 1998; 98(20): 2210 - 2215. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Schauerte, F. A. Schondube, M. Grossmann, H. Dorge, F. Stein, B. Dohmen, A. Moumen, K. Erena, B. J. Messmer, P. Hanrath, et al. Influence of Phase Duration of Biphasic Waveforms on Defibrillation Energy Requirements With a 70-µF Capacitance Circulation, May 26, 1998; 97(20): 2073 - 2078. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. O. Cummins, M. F. Hazinski, R. E. Kerber, P. Kudenchuk, L. Becker, G. Nichol, B. Malanga, T. P. Aufderheide, E. M. Stapleton, K. Kern, et al. Low-Energy Biphasic Waveform Defibrillation: Evidence-Based Review Applied to Emergency Cardiovascular Care Guidelines : A Statement for Healthcare Professionals From the American Heart Association Committee on Emergency Cardiovascular Care and the Subcommittees on Basic Life Support, Advanced Cardiac Life Support, and Pediatric Resuscitation Circulation, April 28, 1998; 97(16): 1654 - 1667. [Full Text] [PDF] |
||||
![]() |
G. H. Bardy, F. E. Marchlinski, A. D. Sharma, S. J. Worley, R. M. Luceri, R. Yee, B. D. Halperin, C. L. Fellows, T. S. Ahern, D. A. Chilson, et al. Multicenter Comparison of Truncated Biphasic Shocks and Standard Damped Sine Wave Monophasic Shocks for Transthoracic Ventricular Defibrillation Circulation, November 15, 1996; 94(10): 2507 - 2514. [Abstract] [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1995 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |