(Circulation. 1995;92:3464-3472.)
© 1995 American Heart Association, Inc.
Articles |
From the Cardiovascular Imaging Center, Departments of Cardiology (P.M.V., M.P., J.D.T.), Biomedical Engineering (N.L.G., K.A.P.), and Thoracic and Cardiovascular Surgery (D.M.C.), Cleveland Clinic Foundation, Cleveland, Ohio.
Correspondence to Pieter M. Vandervoort, MD, Department of Cardiology, Desk F15, Cleveland Clinic Foundation, 9500 Euclid Ave, Cleveland, OH.
| Abstract |
|---|
|
|
|---|
Methods and Results Maximum, transvalvular, and net pressure gradients are calculated and compared with Doppler-derived gradients in an in vitro model. Pressure recovery and pressure loss coefficients are calculated. Simultaneous Doppler and catheter gradients are obtained intraoperatively in five patients undergoing mitral valve replacement. Centerline Doppler gradients correspond closely with maximum catheter gradients but are higher than transvalvular and net pressure gradients. Thirty-six percent of the initial pressure drop is recovered between the valve leaflets and is independent of valve size or configuration. A variable amount of postvalvular pressure recovery is observed depending on aortic or mitral configuration. Side orifice velocities are 85±4% of the centerline velocities. Incorporation of the pressure loss coefficient in the simplified Bernoulli equation shows close agreement between centerline Doppler gradients and transvalvular gradients (r=.99, y=1.11x-0.19).
Conclusions Gradients across the St Jude valve measured by Doppler ultrasound are higher than transvalvular or net catheter gradients due to downstream pressure recovery. This is more marked for Doppler gradients based on centerline velocities than side orifice velocities and is more pronounced for valves in an aortic than a mitral configuration. Therefore, to be comparable with invasive transvalvular catheter gradients, either Doppler gradients should be calculated based on side orifice velocity measurements or the Doppler gradient calculation should include the pressure loss coefficient when based on central orifice velocities.
Key Words: prothesis valves pressure
| Introduction |
|---|
|
|
|---|
Recently, the presence of pressure recovery downstream from the prosthesis has been proposed as a potential cause for the apparent discrepancies between the Doppler and catheter gradients.8 9 10 11 This has been evaluated in vitro only for valves mounted in the aortic position. Because pressure recovery depends critically on the outflow geometry,12 we hypothesized that pressure recovery would be different in central and side orifices and for valves in a mitral and aortic configuration. In the present study, we investigated the velocity and pressure distribution in St Jude heart valves using computational fluid dynamics and in vitro flow modeling. The aim of the study was to assess pressure recovery in central and side orifices of bileaflet prosthetic valves, determine the relative discrepancy between Doppler and catheter gradients for valves in the aortic versus mitral position, and explore ways to adjust for these differences, by either altering the Doppler examination of the valve or modifying the observed data.
| Theoretical Background |
|---|
|
|
|---|
![]() |
where p is given in Pascals, v is given in
m/s, and blood density
is 1050 kg/m3. This
simplification is possible because (1) the effect of blood viscosity
can be neglected as boundary layer formation is suppressed except in
the immediate vicinity of the walls; (2) the inertial term is
negligible for flow through a restrictive orifice (such as a
prosthetic heart valve) since the mass of blood being
accelerated across the valve at any instant is small; and (3) proximal
to the orifice, most energy is present as potential energy and the
proximal velocity can be omitted. This simplified Bernoulli equation is
used in daily echocardiographic practice to calculate
pressure gradients across stenotic valves or estimate
intraventricular pressures from mitral or tricuspid
regurgitant velocity spectra.14
The use of this simplified equation is based on an additional assumption: that all potential energy converted into kinetic energy at the level of the stenosis is completely lost downstream in turbulent friction, vortex formation, and heat. Although this is true for the abrupt stenoses found in most native and prosthetic valves, it is not a physical necessity. If, for example, a stenosis gradually flairs from its narrowest point, then streamlines of flow may remain attached to the walls, allowing smooth deceleration and partial conversion of kinetic energy back into potential energy or pressure.15 The slight outward flair of the two leaflets of the St Jude valve operates in a similar fashion, allowing gradual deceleration of flow with partial reattachment of streamlines and pressure recovery downstream.
The actual recovery of pressure for flow through a diffuser (or the St Jude valve) can be specified by the experimentally determined static pressure recovery coefficient Cp15 :
![]() |
where pmin and pdownstr are the static
pressures (in Pascals) at the orifice and downstream from the orifice,
respectively;
is the constant fluid density, and v is the flow
velocity at the orifice. For uniform entrance and exit flows and no
frictional losses, the ideal pressure recovery coefficient is the
following15 :
![]() |
where Ai and Ao are the entrance and exit flow areas, respectively. The second substitution is only possible for a two-dimensional diffuser (as is the case for the central orifice of the St Jude valve) where areas can be replaced by width of the inlet (Wi) and outlet (Wo) since the height of inlet and outlet are the same. The "effectiveness" of a diffuser can be defined by comparing the actual pressure recovery coefficient with the theoretical ideal15 :
![]() |
In general, Cp will fall below the theoretical ideal
and
will be less than 1 since viscous dissipation and velocity
peaking in the outlet stream will reduce pressure recovery.
The net pressure loss across a diffuser (which is more clinically relevant for prosthetic heart valves) is specified by the pressure loss coefficient K, defined by the change in total pressure through the diffuser15 :
![]() |
For a diffuser discharging into a large reservoir (relative to the diffuser exhaust), the pressure loss coefficient K and the pressure recovery coefficient Cp are related by15 :
![]() |
For a perfect diffuser with complete pressure recovery, Cp=1 and K=0; in the absence of any pressure recovery downstream from the flow obstruction, Cp=0 and K=1.
| Methods |
|---|
|
|
|---|
500
iterations). This numerical simulation provided pressure information,
velocity magnitude, and the axial and radial components of velocity at
each point of the flow field. Velocity and pressure profiles through
central and side orifices were reconstructed for comparison with the
experimental data.
In Vitro Experiment
Prosthetic bileaflet heart valves (sizes
No. 19 through
29, St Jude Medical Inc) were mounted in an in vitro flow model
producing steady flow between 215 and 405 mL/s. The model consisted of
two plexiglas chambers: a proximal chamber (8x18x30 cm) that is
sealed and can be pressurized and a distal larger discharge chamber
(100x18x30 cm) at atmospheric pressure. Flow entered the proximal
chamber by gravity from an upper reservoir under controlled pressure
and left the distal chamber via an overflow outlet, which ensured a
steady pressure difference between both chambers. The fluid (1%
aqueous solution of cornstarch, which provides acoustic reflectors) was
pumped into the upper reservoir to maintain its level, and flow rate
was varied by altering the pressure difference between proximal and
distal chambers. The prosthetic heart valves were mounted with
the leaflets in a vertical position in the partition between the two
chambers. To simulate a prosthetic valve in a mitral position
(distal chamber large compared with the valve), flow discharged
directly into the distal chamber. For valves Nos. 19 through 25, an
aortic position was also simulated. Circular tubings (8 cm in length)
with diameters of 26 and 31 mm were fixed distal to the valve to
simulate an artificial ascending aorta.
Micrometer measurements of the height and width of central orifice inlet and outlet were obtained for valve sizes Nos. 19 through 29.
Data acquisition. Pressure measurements were obtained using
high-fidelity pressure transducers (Millar Inc). The catheter was
connected to a microprocessor controlled infusion pump (Harvard
Apparatus) and slowly (0.8 mm/s) pulled back through the
central and side orifices of the St Jude heart valves. The pressure
catheter (7F) was stabilized with an 8F introducer sheath to ensure
stable and straight pullbacks. The pressure transducer located at the
side of the catheter tip was rotated upward to avoid interference with
the leaflets when the catheter was pulled back through the valve. The
pressure waveforms were digitized at 1000 Hz over 100 seconds (8 cm
distance) and stored to disc (Macintosh Quadra) using
LABVIEW (National Instruments).16 Pressure
measurements started 1.5 cm proximal to the valve and were acquired
continuously until 6.5 cm downstream from the valve.
Simultaneous Doppler velocity measurements were
obtained using a Sonos 1500 echocardiograph
(Hewlett-Packard) equipped with digital storage and retrieval
capabilities. A 2.5-MHz ultrasound transducer was positioned proximal
to the orifice along the centerline of the flow. By directing the
continuous wave ultrasound beam, velocities in the central and side
orifices could be interrogated separately. Pulsed Doppler
velocities were measured
1.5 cm proximal to the valve. These
proximal velocities ranged between 38 and 59 cm/s and are neglected in
the simplified Bernoulli equation for gradient calculations.
Data
analysis. The proximal pressure
(Pprox, measured approximately 1.5 cm proximal to
the valve), the minimal pressure (Pmin), and the distal
pressure (Pdist, measured approximately 6.5 cm
downstream from the valve) were read out from both pressure tracings
through central and side orifices. For the central pressure tracing
only, an additional pressure measurement (Psh) was read out
at the "shoulder" of the pressure curve occurring at the distal
edge of the leaflets (Fig 1
).
|
From these pressure
measurements, the maximum and net pressure
gradients across central and side orifices were calculated. Also, for
the central orifice, a transvalvular gradient was
calculated (Fig 1
). The maximum catheter pressure gradient was
calculated as the difference between the proximal and the minimum
pressure measurements (Pprox-Pmin). The net
pressure drop was given by the difference between the proximal and the
distal pressure measurements (Pprox-Pdist).
The transvalvular pressure gradient across the central
orifice was calculated as Pprox-Psh. The
percentage of maximal pressure drop in central and side orifices
recovered downstream was calculated as
100 · (Pdist-Pmin)/(Pprox-Pmin).
For flow through the central orifices, the pressure recovery
consistently showed two phases. Phase 1 was the initial more
rapid pressure recovery from Pmin to
Psh, occurring between the valve leaflets (early or
valvular recovery); the percent pressure recovery is calculated
as
100 · (Psh-Pmin)/(Pprox-Pmin).
Phase 2 represented the pressure recovery occurring distal
from the valve (late or postvalvular recovery) and was
calculated as
100 · (Pdist-Psh)/(Pprox-Pmin)
(Fig 1
). For each valve, the pressure recovery coefficient
Cp and the total pressure loss coefficient K were
calculated.
From the continuous wave Doppler velocity measurements through central and side orifices, pressure gradients were calculated with the simplified Bernoulli equation.
Clinical Study
We studied five patients undergoing mitral
valve replacement
with a St Jude prosthetic valve intraoperatively after coming
off cardiopulmonary bypass. All patients were in a regular
sinus rhythm or atrioventricular paced rhythm and
hemodynamically stable during the data acquisition. A
transesophageal echoprobe was inserted as part of the
routine diagnostic and monitoring procedure in patients
undergoing valve surgery. The Omniplane transducer was rotated to
clearly visualize the two valve leaflets with separation of central and
side orifices of the prosthetic valve (Fig 2
).
It was possible to position the continuous wave Doppler cursor
selectively across the central and the side orifices of the valve in
all patients. Doppler velocity spectra were obtained with the use
of a Sonos 1500 echocardiograph (Hewlett-Packard) and
recorded on
-in videotape. A left atrial
fluid-filled
pressure catheter was inserted through the left atrial appendage, and
the left ventricular pressure was measured with a
fluid-filled line through a direct needle stick of the left
ventricle near the left ventricular apex. Pressure
measurements of the left atrium and the left ventricle were acquired
simultaneously. The pressure waveforms were amplified
(module 300, Marquette Tram) and subsequently digitized at 1000 Hz with
a National Instruments data acquisition board (AT-MIO-16) and stored to
disc (Gateway 2000 486 Dx 2/50) using custom software written in
LABVIEW III (National Instruments).
Simultaneously with left atrial and left
ventricular pressures, we measured Doppler velocities
through the central and subsequently through the side orifices. Only
cardiac cycles with high quality simultaneous left atrial
and left ventricular pressure and Doppler velocity data
were used for analysis. Ectopic ventricular or
supraventricular beats were excluded. The peak
instantaneous pressure gradient during early filling (E wave) was
measured with a customized analysis package written in
LABVIEW. Peak transvalvular Doppler
gradients were calculated off-line with the calculation package
incorporated in the echocardiograph.
|
Statistical Analysis
Pressure profiles obtained from
numerical simulations were
compared with the experimental data, and a correlation coefficient was
calculated. Pressure measurements for mitral and aortic configurations
were compared with paired Student's t tests.
Doppler pressure gradients were compared with catheter gradients using linear regression analysis, and a correlation coefficient was calculated. The difference between Doppler and catheter gradients (Doppler minus catheter) were calculated and expressed as mean±SD.
The impact of valve size and outlet configuration were compared with ANOVA.
| Results |
|---|
|
|
|---|
|
|
In Vitro Experiments
Pressure Measurements
Central versus side orifices. The pressure profile
through the central orifice shows a deep pressure well with gradual
pressure recovery downstream. This distal pressure recovery through the
central orifice occurs in two phases: a first rapid increase in
pressure following the minimal pressure and a second phase of further
but slower increase in downstream pressure (Fig 5
). This
first phase corresponds to the pressure recovery occurring between the
two leaflets before flow leaves the valve; the second phase reflects
the additional pressure recovery downstream from the level of the
valve.
|
In the side orifices, the localized low pressure zone is less
pronounced and occurs more distally. The maximum pressure gradients
measured in the side orifices are 24±10% (mean±SD,
P<.001) lower than the maximum pressure gradients measured
across the central orifices. The pressure recovery through the side
orifice does not systematically show the biphasic pattern seen in the
central orifice. The percent of maximum pressure drop recovered in the
side orifices is significantly lower than that in the central orifices
for all valves (Table 1
).
|
Aortic versus mitral
configuration. In aortic valves with
the smallest aorta (26 mm), as much as 58.1±5.0% of the initial
pressure drop in the central orifice is recovered downstream;
36.7±3.9% of this total pressure is recovered during deceleration of
flow within the first phase (valvular), and an additional
21.3±6.0% is recovered in the second phase
(postvalvular). Increasing the size of the aorta decreases
the total pressure recovered in the central orifice significantly to
49.8±4.7%, P<.001. This decrease is entirely due to less
pressure recovery during phase 2 as 37.2±4.7% is recovered between
the leaflets (phase 1) and only 12.6±0.7% of pressure recovery occurs
downstream from the valve in the aorta (phase 2). For valves in the
mitral configuration, total pressure recovery through the central
orifice is only 37.3±6.4%, P<.001, occurring almost
entirely at the level of the valve (phase 1, 35.6±4.2%) with
virtually no further pressure recovery downstream (phase 2,
1.7±3.3%). Fig 6
shows the difference in centerline
velocity profile for a valve in the mitral and aortic (26-mm aorta)
positions.
|
Overall, maximum pressure drop and percent pressure recovery are significantly lower (P<.001) in the side orifices than in the central orifices. The percent pressure recovery in the side orifice of the aortic valves with the smallest (26 mm) aorta is 43.9±9.9%. The percent pressure recovery decreases to 33.9±7.9% (P<.001) for the aortic valves with the bigger aorta (31 mm) and is only 17.9±4.7% for the side orifices of the mitral position.
Doppler Velocity Measurements
Continuous
wave Doppler velocities through the central
orifice range from 175 to 430 cm/s, and velocities through the side
orifices range from 149 to 350 cm/s and are 85±4% (mean±SD,
P<.001) of the central orifices. The ratio between maximum
velocities in central and side orifices is independent of valve size
and configuration.
Maximum versus net pressure gradients. The pressure
gradients (y) calculated from the continuous wave
Doppler velocity measurements in central and side orifices with the
simplified Bernoulli equation correlate closely with the maximum
catheter gradients (x) measured in central and side
orifices, respectively, with r=.99,
y=1.04x+0.22, and
p (catheter minus
Doppler)=1.0±1.8 mm Hg. As expected, these maximal Doppler
gradients at the level of the valve are significantly higher than the
net catheter gradients. This pressure difference is significantly
higher for gradients based on centerline velocity measurements
(114.8±52.2%,
p=13.4±12.1 mm Hg) than for gradients
calculated
from side orifice velocities (54.2±36.4%,
p=5.7±4.4
mm Hg)
(P<.001) (Table 2
). The discrepancies
between Doppler gradients across the central orifice and the net
pressure drop in aortic valves with the 26-mm aorta (167.1±51.8%) are
significantly larger than in aortic valves with the 31-mm aorta
(123.3±41.0%, P<.001) and larger than in the mitral
valves (74.2±15.4%, P<.001). The difference between
Doppler gradients across the side orifices and the net invasive
pressure drop is significantly less, in particular, for the mitral
configuration (Table 2
).
|
Transvalvular gradients.
Maximum Doppler
gradients (y) obtained through the central orifices
correlate closely (r=.99) but are significantly higher than
the transvalvular pressure gradients (x) with
y=1.73x-0.29 and
p=9.9±9.6
mm Hg
(73.6±18.6%).
Pressure Recovery Coefficient
(Cp) and Pressure Loss
Coefficient (K)
Micrometer measurements of inlet (Wi) and
outlet width (Wo) of the central orifice show a
fixed relation of Wi/Wo=0.65±0.02
(range, 0.63 to 0.68) for all valve sizes studied (Nos. 19 to 29).
Based on these measurements, a theoretical pressure recovery
coefficient for ideal uniform flow through the St Jude valve is
calculated as
Cp(ideal)=1-0.652=0.58.
The
actual pressure recovery coefficient Cp for the central
orifice of all valves is 0.36±0.04 (mean±SD), and the total
pressure
loss coefficient K calculated for all valves was 0.64±0.04. ANOVA
shows no effect of valve size or configuration. The diffuser
effectiveness (
) is calculated by comparing the actual with the
theoretical pressure recovery coefficient:
=0.36/0.58=0.62.
When Doppler-derived gradients
across the central orifice are
multiplied by the pressure loss coefficient K=0.64, the
transvalvular gradients (x) are closely
approximated with r=.99,
y=1.11x-0.19, and
p=1.4±2.1
mm Hg
(11.1±11.9%) as illustrated in Fig 7
. Introduction of
this pressure loss coefficient also reduces the difference between
Doppler and net pressure gradients (including pressure recovery
downstream from the valve) by almost 80% with
p=3.9±4.6
mm Hg
(Table 3
). Residual discrepancy was present only for
valves in the aortic position; for St Jude valves in a mitral
configuration, Doppler velocity measurements through the central
orifices showed close agreement with the net transvalvular
gradient (
p=0.8±0.8 mm Hg).
|
|
Clinical Study
Fig 2
shows a transesophageal
ultrasound image of
a St Jude valve prosthesis in the mitral position illustrating
the two valve leaflets with the central and side orifices. Fig
8
shows a Doppler velocity tracing (top) with the
simultaneously acquired left atrial, left
ventricular, and aortic pressures (bottom) for three
consecutive cardiac cycles. The first Doppler tracing on the left
was obtained through the central orifice showing the highest
velocities. With slight rotation of the echoprobe, the Doppler
velocity profile through the side orifice was obtained (third tracing
on the right) showing a markedly lower peak velocity. On the middle
tracing, a double-velocity contour can be seen when the
echotransducer is in an intermediate position, sampling velocities from
both the central and the side orifices. Although there is a marked
difference in the calculated Doppler gradients between the first
and the third cardiac cycles, the hemodynamic
conditions are unchanged, as illustrated by the
simultaneously acquired invasive pressure measurements.
|
A total of 42
cardiac cycles with high-quality
simultaneous Doppler and pressure data were
analyzed. The invasive peak pressure gradients during early
filling range from 4 to 9 mm Hg (6.2±1.2 mm Hg). The corresponding
Doppler gradients measured across the central orifice range from 7
to 11 mm Hg (9.1±1.1 mm Hg) and are significantly higher than the
invasive pressure gradients with a difference (Doppler minus
catheter gradient) of 47±20% (2.9±0.9 mm Hg) (Fig
9a
). When the pressure loss coefficient (K=0.64) is
incorporated in the simplified Bernoulli equation, the Doppler
gradients calculated across the central orifice range from 5.2 to 7
mm Hg (5.9±0.7 mm Hg) and show very good agreement with invasive
measurements with a difference (Doppler minus catheter gradient) of
-6±13% (-0.4±0.8 mm Hg) (Fig 9b
).
Doppler gradients measured
across the side orifices range from 5.4 to 9.1 mm Hg (6.9±1.1 mm Hg)
and also agree very well with the invasive pressure gradients with a
difference (Doppler minus catheter gradient) of 15±12% (0.8±0.5
mm Hg) (Fig 9a
).
|
| Discussion |
|---|
|
|
|---|
Results of Present Study
The numerical simulation shows that
a localized low-pressure
zone is present at the entrance of the central orifice. The
pressure profile through the central orifice illustrates this
deep-pressure well with gradual increase in pressure further
downstream. A similar low-pressure well occurs in the side orifices
but is less pronounced and is located more downstream at the distal
margins of the valve leaflets. These low-pressure zones correspond
to local areas of high velocities as illustrated by the numerical
simulations.
Our observations in the in vitro experiments are very
similar to the
numerical predictions. In the vitro experiments, we find very close
agreement between Doppler gradients and maximum catheter pressure
gradients measured in the low-pressure well occurring between the
prosthetic valve leaflets. In the clinical setting, this
maximum pressure gradient cannot be measured invasively because it is
not possible to position a catheter across or between mechanical
prosthetic valve leaflets in patients. Doppler gradients
across the central orifices are significantly higher than the
transvalvular and net catheter pressure gradients measured
across the valve. These differences are due to downstream pressure
recovery. This downstream pressure recovery through the central
orifices shows a biphasic pattern. The first part has a rapid increase
in pressure and occurs predominantly between the leaflets of the valve.
This phase of the pressure recovery critically depends on the gradual
deceleration of flow with reattachment of streamlines between the
slightly outward-directed valve leaflets. Because this phase of
pressure recovery occurs between the valve leaflets, it would not be
detected in the clinical situation by catheter measurements. We find
that a fixed percentage of
35% of the maximum pressure drop is
recovered between the two leaflets of the valve. This percentage is
independent of valve size or valve configuration. A second phase of
pressure recovery occurs further downstream beyond the level of the
valve. The in vitro experiments showed that during the second phase, a
smaller portion of the initial pressure drop is recovered and is highly
dependent on the distal geometry. This second phase of pressure
recovery is most pronounced in the aortic configuration with the
smallest aorta, where an additional 21.3±6.0% of the initial pressure
drop is recovered. In the mitral configuration, virtually no further
pressure recovery occurs beyond the level of the valve. Because
pressure recovery critically depends on downstream geometry and
catheter position, direct comparison with previous studies is difficult
when in vitro setup and testing conditions are not exactly similar.
Baumgartner et al have studied pressure recovery in St Jude valves in
the aortic position (32-mm aorta) under pulsatile flow conditions in
vitro. They reported 41.8±9.1% pressure recovery at 3 cm downstream
from the valve with the distal pressure measured at the wall of the
aorta. In the present study, we found 49.8±4.7% pressure recovery
for a St Jude valve with a 31-mm aorta and the distal pressure measured
in the center of the aorta at 6.5 cm from the valve. These findings are
very similar; however, in the study by Baumgartner et al,
valvular and postvalvular pressure recoveries were
not analyzed separately, and the authors did not evaluate
pressure recovery with a different-size aorta or mitral
configurations.
Distal pressure recovery is significantly less through the side orifices and does not show consistently the biphasic pattern observed in the central orifice. Because overall pressure recovery through the side orifices is less pronounced, Doppler gradients across the side orifices approximate the transvalvular and net pressure drop closer than Doppler gradients across the central orifice. In the central orifice, flow decelerates with reattachment of streamlines on both leaflets, allowing for less flow separation and more pressure recovery. For flow through the side orifices, reattachment of streamlines is only possible on one side and therefore less pressure recovery occurs. However, in the presence of an aorta, streamlines can also reattach downstream to the aortic wall, resulting in increased pressure recovery through the side orifices for valves in the aortic position compared with valves in the mitral configuration.
The observation of fixed percentage of pressure recovery through the central orifice is also reflected by the calculated pressure recovery coefficient Cp of 0.36±0.04. The clinically more relevant transvalvular pressure drop is reflected by the pressure loss coefficient K=0.64±0.04. For all St Jude prosthetic valves tested (Nos. 19 to 29), the total pressure loss coefficient K is constant and independent of valve size or downstream geometry. This constant K reflects the proportion of the total kinetic energy present at the orifice that is lost downstream in friction, vortex formation, and heat and therefore reflects the true net static pressure drop across a stenosis.
Implications for Measurements of Doppler
Gradients
Application of the simplified Bernoulli equation to Doppler
velocity measurements across the central orifices of St Jude valves
will yield significantly higher calculated pressure gradients across
the valve than the pressure gradient measured during cardiac
catheterization. However, a thorough understanding of
the hydrodynamic profile of this valve allows us to interpret this
discrepancy. The results of this study have important implications for
the noninvasive assessment of St Jude prosthetic valve function
using Doppler ultrasound. Incorporation of the experimentally
determined pressure loss coefficient K for the St Jude valve in the
simplified Bernoulli equation is necessary for a comparison of
noninvasive Doppler gradients with the invasive gradients measured
during cardiac catheterization. We find that this
pressure loss coefficient K is similar for all valve sizes, allowing us
to closely approximate the invasively measured
transvalvular gradient for all St Jude valves independent
of valve size or position. When the Doppler gradient
(
v2) across the central orifice
was multiplied
by the pressure loss coefficient K=0.64, an excellent agreement was
observed between the noninvasive and invasive transvalvular
pressure gradients.
The findings of the present study also show that pressure recovery through the side orifices is less pronounced, and Doppler gradients through the side orifices closely approximate the transvalvular and net pressure gradients, especially for mitral valves. This suggests that alteration of the Doppler examination of the St Jude valve and measurement of transvalvular velocities selectively through the side orifices provide more accurate Doppler-derived pressure gradients. Although selective sampling of central and side orifices may be difficult for St Jude valves in the aortic position, it should be possible for valves in the mitral position. In particular, the transesophageal imaging window enables us to image the mitral prosthesis in the near field and selectively sample velocities through central and side orifices.
We studied five patients undergoing mitral valve replacement with a St Jude heart valve and found close agreement between Doppler gradients measured across the side orifices of the valve and the invasive pressure gradients, with the Doppler gradients being only slightly higher. Similar to the numerical and in vitro observations, Doppler gradients measured across the central orifice are significantly higher than the invasive measurements. Incorporation of the pressure loss coefficient K into the simplified Bernoulli equation across the central orifice also provides excellent agreement between noninvasive and invasive pressure gradient measurements.
Study Limitations
The purpose of the numerical simulation was
to illustrate the
spatial velocity and pressure distribution within the central and side
orifices of a St Jude valve. The results of any numerical simulation
are limited by its inherent assumptions. Our model is a simplified
two-dimensional cross section of the prosthetic valve and
therefore is limited in reflecting what is happening in a complex
three-dimensional structure such as the St Jude valve. Although it
was not the purpose of the present study to validate the results of
the numerical model, the pressure profiles predicted by the numerical
simulation closely match our in vitro findings, lending credibility to
results of the computer simulation.
The phenomenon of pressure recovery has been well described under steady10 15 and pulsatile8 11 flow conditions. We selected a steady flow model because this setup allows for continuous pressure pull-back measurements within the valve orifices. This is necessary to locate precisely the minimal pressure, measure the amount of pressure recovery occurring within the valve orifices, and calculate pressure loss and pressure recovery coefficients within the central orifice of the St Jude valve. In the presence of pulsatile flow, the acceleration phase tends to stabilize streamlines and may reenforce pressure recovery, whereas deceleration tends to destabilize flow and become more turbulent, therefore reducing pressure recovery. The effect of acceleration and deceleration on the instantaneous pressure recovery was beyond the aim of the study; however, this question merits further study in the future.
The apparent differences between invasive and noninvasive transvalvular pressure gradients is due not only to pressure recovery at the level of the St Jude valve. As the present study illustrates, pressure recovery can also occur beyond the level of the valve, depends critically on the downstream geometry, and could be more than 20% of the maximum pressure drop in the presence of a small ascending aorta. Depending on whether the valve discharges into the aorta or into an open chamber as is the left ventricle, the exact position of the catheter downstream, and whether downstream flow is uniform throughout the cross section of the distal chamber, a variable discrepancy may remain. In our in vitro study, for the mitral position the distal chamber is large compared with the size of the left ventricle. In the clinical setting, in particular in the presence of left ventricular hypertrophy with a small left ventricular cavity, it is possible that downstream pressure recovery occurs within the ventricle, and therefore differences between mitral and aortic position may be less than those found in our in vitro study. The use of fluid-filled catheters21 as well as differences in sampling location in left atrium and left ventricle22 may introduce some inaccuracies into the pressure measurements obtained in our clinical study; however, overall the clinical observations closely match the numerical and in vitro findings.
Pressure recovery critically depends on the gradual deceleration of flow with reattachment of streamlines. In the present study, we evaluated normal functioning valve prostheses. In the presence of thrombus or pannus overgrowth interfering with normal leaflet excursion, the hydrodynamic profile of the valve may be significantly altered, and the amount of pressure recovery in the central orifice may be reduced or completely dissappear, as was suggested by the results of an in vitro study.23 If prosthetic valve dysfunction is suspected, careful examination of leaflet excursion by transthoracic or transesophageal imaging is necessary. When abnormal leaflet excursion is present (or suspected), (1) the Doppler gradients across the central orifice should be calculated without incorporating the pressure loss coefficient into the simplified Bernoulli equation and/or (2) the Doppler gradients should be measured across the side orifices, where pressure recovery is minimal and the simplified Bernoulli equation is adequate.
Other possible reasons for gradient overestimation by Doppler under physiological conditions is the known difference between instantaneous peak gradients (measured by Doppler) and the peak-to-peak gradients typically measured by catheter. Doppler gradients may be erraneous under conditions where the simplified Bernoulli equation is no longer valid, eg, when the proximal velocity is not negligible or for flow through nonrestrictive orifices where the inertial term in the Bernoulli equation cannot be neglected. Doppler gradients can underestimate true gradients if the peak velocity could not be resolved or if the direction of the flow is oblique to the ultrasound beam and therefore only its component parallel to the Doppler beam is recorded.
Although maximum, transvalvular, and net pressure gradients are all different gradients that physically exist, it is currently not known which of these gradients is most relevant to reflect the work load imposed on the heart.
Conclusions
We explored and explained the previously reported
differences
between Doppler and catheter gradients across the St Jude valve. We
demonstrated that the discrepancies between Doppler and catheter
gradients are not due to errors in either Doppler or catheter
techniques of measuring transvalvular gradients but; rather
that Doppler ultrasound and catheters measure different gradients
that both exist physically. To be able to compare catheter and
Doppler gradients obtained across a St Jude valve
prosthesis, we suggest either multiplying the Doppler
gradient (obtained across the central orifice) by the experimentally
defined pressure loss constant K=0.64 or, in particular for valves in
the mitral position, calculating transvalvular gradients
from Doppler velocity measurements obtained by selectively
interrogating the side orifices of the prosthetic heart
valve.
Received March 15, 1995; revision received August 7, 1995; accepted August 8, 1995.
| References |
|---|
|
|
|---|
2.
Hegrenaes L, Hatle L. Aortic stenosis
in adults: noninvasive estimation of pressure differences by continuous
wave Doppler echocardiography. Br
Heart J. 1985;54:396-404.
3.
Currie P, Seward JB, Reeder GS, Vliestra RE, Bresnahan
DR, Bresnahan JF, Smith HC, Hagler GJ, Tajik AJ. Continuous wave
Doppler echocardiographic assessment of severity of
calcific aortic stenosis: a simultaneous
Doppler-catheter correlative study in 100 adult
patients. Circulation. 1985;71:1162-1169.
4. Williams GA, Labovitz AJ. Doppler hemodynamic evaluation of prosthetic (Starr-Edwards and Bjork-Shiley) and bioprosthetic (Hancock and Carpentier-Edwards) cardiac valves. Am J Cardiol. 1985;56:325-332. [Medline] [Order article via Infotrieve]
5.
Wilkins GT, Gillam LD, Kritzer GL, Levine RA, Palacios
IF, Weyman AE. Validation of continuous wave Doppler
echocardiographic measurements of mitral and tricuspid
prosthetic valve gradients: a simultaneous
Doppler-catheter study.
Circulation. 1986;74:786-795.
6.
Burstow DJ, Nishimura RA, Bailey KR, Reeder GS, Holmes
DR, Seward JB, Tajik AJ. Continuous wave Doppler
echocardiographic measurements of prosthetic
valve gradients: a simultaneous Doppler-catheter
correlative study. Circulation. 1989;80:504-514.
7. Rothbart RM, Smucker ML, Gibson RS. Overestimation by Doppler echocardiography of pressure gradients across Starr-Edwards prosthetic valves in the aortic position. Am J Cardiol. 1988;61:475-476. [Medline] [Order article via Infotrieve]
8.
Baumgartner H, Khan S, DeRobertis M, Czer L, Maurer G.
Discrepancies between Doppler and catheter gradients in
aortic prosthetic valves in vitro: a manifestation of localized
pressure gradients and pressure recovery.
Circulation. 1990;82:1467-1475.
9. Levine RA, Jimoh A, Cape EG, McMillan S, Yoganathan AP, Weyman AE. Pressure recovery distal to a stenosis: potential cause of gradient `overestimation' by Doppler echocardiography. J Am Coll Cardiol. 1989;13:706-713. [Abstract]
10. Clark C. The fluid mechanics of aortic stenosis, I: theory and steady flow experiments. J Biomechan. 1976;9:521-528. [Medline] [Order article via Infotrieve]
11. Voelker W, Ruel H, Stelzer T, Schmidt A, Karsch KR. Pressure recovery in aortic stenosis: an in vitro study in a pulsatile flow model. J Am Coll Cardiol. 1992;20:1585-1593. [Abstract]
12. Baumgartner H, Schima H, Tulzer G, Kuhn P. Effect of stenosis geometry on the Doppler-catheter gradient relation in vitro: a manifestation of pressure recovery. J Am Coll Cardiol. 1993;21:1018-1025. [Abstract]
13. Yoganathan AP, Cape EG, Sung H, Williams FP, Timoh A. Review of hydrodynamic principles for the cardiologist: applications to the study of blood flow and jets by imaging techniques. J Am Coll Cardiol. 1988;12:1344-1353. [Abstract]
14. Hatle L, Angelsen B. Doppler Ultrasound in Cardiology: Physical Principles and Clinical Applications. 2nd Ed. Philadelphia, Pa: Lea & Febiger; 1986:24.
15. Blevins RB. Applied Fluid Dynamics Handbook. New York, NY: Van Nostrand Reinhold Company Inc; 1984:144-158.
16. Greenberg NL, Vandervoort PM, Stewart WJ, Savage RM, McCarthy PM, Thomas JD. An integrated system for simultaneous acquisition and processing of physiologic and ultrasound data. Comp Cardiolo. 1994:615-618.
17.
Veyrat C, Witchitz S, Lessana A, Ameur A, Abitbol G,
Kalmanson D. Valvar prosthetic dysfunction: localisation
and evaluation of dysfunction using the Doppler technique.
Br Heart J. 1985;54:273-284.
18. Pandis IP, Ross J, Mintz GS. Normal and abnormal prosthetic valve function as assessed by Doppler echocardiography. J Am Coll Cardiol. 1987;8:317-326.
19.
Teirstein PS, Yock PG, Popp RL. The accuracy of
Doppler ultrasound measurements of pressure gradients across
irregular, dual, and tunnellike obstructions to blood flow.
Circulation. 1985;72:577-584.
20. Baumgartner H, Khan S, DeRobertis M, Czer L, Maurer G. Effect of prosthetic aortic valve design on the Doppler-catheter gradient correlation: an in vitro study of normal St Jude, Medtronic-Hall, Starr-Edwards and Hancock valves. J Am Coll Cardiol. 1992;19:324-332. [Abstract]
21. Grossman W. Pressure measurement. In: Grossman W, ed. Cardiac Catheterization and Angiography. 2nd Ed. 1980:103-114.
22.
Courtois M, Kovacs SJ, Ludbrook PA. Transmitral
pressure-flow velocity relation. Importance of regional pressure
gradients in the left ventricle during diastole.
Circulation. 1988;78:661-671.
23. Silber H, Khan S, DeRobertis M, Gray R, Matloff J, Maurer G. Effect of prosthetic valve obstruction on the Doppler-catheter gradient correlation for St Jude valves: in vitro studies. Circulation. 1992;86(suppl I):I-806. Abstract.
This article has been cited by other articles:
![]() |
J. Price, H. Lapierre, L. Ressler, B.-K. Lam, T. G. Mesana, and M. Ruel Prosthesis-patient mismatch is less frequent and more clinically indolent in patients operated for aortic insufficiency J. Thorac. Cardiovasc. Surg., September 1, 2009; 138(3): 639 - 645. [Abstract] [Full Text] [PDF] |
||||
![]() |
J Mascherbauer and H Baumgartner The authors' reply: Heart, April 1, 2009; 95(7): 592 - 593. [Full Text] [PDF] |
||||
![]() |
D. Tanne, L. Kadem, R. Rieu, and P. Pibarot Hemodynamic impact of mitral prosthesis-patient mismatch on pulmonary hypertension: an in silico study J Appl Physiol, December 1, 2008; 105(6): 1916 - 1926. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. Thomas and Z. B. Popovic Assessment of Left Ventricular Function by Cardiac Ultrasound J. Am. Coll. Cardiol., November 21, 2006; 48(10): 2012 - 2025. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Ruel, H. Al-Faleh, A. Kulik, K. L. Chan, T. G. Mesana, and I. G. Burwash Prosthesis-patient mismatch after aortic valve replacement predominantly affects patients with preexisting left ventricular dysfunction: Effect on survival, freedom from heart failure, and left ventricular mass regression J. Thorac. Cardiovasc. Surg., May 1, 2006; 131(5): 1036 - 1044. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Amarelli, A. Della Corte, G. Romano, G. Iasevoli, G. Dialetto, L. S. De Santo, M. De Feo, M. Torella, M. Scardone, and M. Cotrufo Left ventricular mass regression after aortic valve replacement with 17-mm St Jude Medical mechanical prostheses in isolated aortic stenosis J. Thorac. Cardiovasc. Surg., March 1, 2005; 129(3): 512 - 517. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. R. Rezaian, K. Aghasadeghi, and J. Kojuri Evaluation of the Hemodynamic Performance of St.Jude Mitral Prostheses: A Pilot Study by Dobutamine-Stress Doppler Echocardiography Angiology, January 1, 2005; 56(1): 81 - 86. [Abstract] [PDF] |
||||
![]() |
K. E. Richards, D. Deserranno, E. Donal, N. L. Greenberg, J. D. Thomas, and M. J. Garcia Influence of structural geometry on the severity of bicuspid aortic stenosis Am J Physiol Heart Circ Physiol, September 1, 2004; 287(3): H1410 - H1416. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Albes, M. Hartrumpf, V. Rudolph, T. Krempl, E. Huttemann, R. Vollandt, and T. Wahlers Are mechanical valves with enhanced inner diameter advantageous in the small sized aortic annulus? Ann. Thorac. Surg., November 1, 2003; 76(5): 1564 - 1570. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Teijeira Omniscience valve results: letter 2 Ann. Thorac. Surg., August 1, 2003; 76(2): 659 - 659. [Full Text] [PDF] |
||||
![]() |
M. A. Vannan and K. Sarkar Assessment of mechanical aortic valve prosthesis by means of Doppler echocardiography: what to measure and why? J. Thorac. Cardiovasc. Surg., August 1, 2003; 126(2): 317 - 320. [Full Text] [PDF] |
||||
![]() |
D. Garcia, J. G. Dumesnil, L.-G. Durand, L. Kadem, and P. Pibarot Discrepancies between catheter and Doppler estimates of valve effective orifice area can be predicted from the pressure recovery phenomenon: practical implications with regard to quantification of aortic stenosis severity J. Am. Coll. Cardiol., February 5, 2003; 41(3): 435 - 442. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A. Levine and E. Schwammenthal Stenosis is in the eye of the observer: impact of pressure recovery on assessing aortic valve area J. Am. Coll. Cardiol., February 5, 2003; 41(3): 443 - 445. [Full Text] [PDF] |
||||
![]() |
D. Gilon, E. G. Cape, M. D. Handschumacher, J.-K. Song, J. Solheim, M. VanAuker, M. E. E. King, and R. A. Levine Effect of three-dimensional valve shape on the hemodynamics of aortic stenosis: Three-dimensional echocardiographic stereolithography and patient studies J. Am. Coll. Cardiol., October 16, 2002; 40(8): 1479 - 1486. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Bech-Hanssen, K. Caidahl, I. Wallentin, P. Ask, and B. Wranne Assessment of effective orifice area of prosthetic aortic valves with Doppler echocardiography: An in vivo and in vitro study J. Thorac. Cardiovasc. Surg., August 1, 2001; 122(2): 287 - 295. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. S. Firstenberg, P. M. Vandervoort, N. L. Greenberg, N. G. Smedira, P. M. McCarthy, M. J. Garcia, and J. D. Thomas Noninvasive estimation of transmitral pressure drop across the normal mitral valve in humans: importance of convective and inertial forces during left ventricular filling J. Am. Coll. Cardiol., November 15, 2000; 36(6): 1942 - 1949. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Pibarot and J. G. Dumesnil Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention J. Am. Coll. Cardiol., October 1, 2000; 36(4): 1131 - 1141. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Shandas, J. Kwon, and L. Valdes-Cruz A Method for Determining the Reference Effective Flow Areas for Mechanical Heart Valve Prostheses : In Vitro Validation Studies Circulation, April 25, 2000; 101(16): 1953 - 1959. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Y. Leung, J. Wong, L. Rodriguez, M. Pu, P. M. Vandervoort, and J. D. Thomas Application of Color Doppler Flow Mapping to Calculate Orifice Area of St Jude Mitral Valve Circulation, September 22, 1998; 98(12): 1205 - 1211. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. H. Marcus, R. S. Heinrich, J. Bednarz, S. Lupovitch, J. Abruzzo, R. Borok, B. Vandenberg, R. E. Kerber, W. Piccione, A. P. Yoganathan, et al. Assessment of Small-Diameter Aortic Mechanical Prostheses : Physiological Relevance of the Doppler Gradient, Utility of Flow Augmentation, and Limitations of Orifice Area Estimation Circulation, September 1, 1998; 98(9): 866 - 872. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1995 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |