(Circulation. 1995;92:3408-3414.)
© 1995 American Heart Association, Inc.
Articles |
From the Intravascular Ultrasound Imaging and Cardiac Catheterization Laboratories, Washington (DC) Hospital Center.
| Abstract |
|---|
|
|
|---|
Methods and Results ELCA was used to treat 202 lesions in 190 patients. Forty-nine lesions in 48 patients were studied by use of sequential (before and after ELCA and after adjunctive device therapy) intravascular ultrasound (IVUS). External elastic membrane (EEM), lumen, and plaque+media (P+M=EEM-lumen) cross-sectional areas (CSAs) and lesion arcs of calcium were measured before and after ELCA and after adjunct device use. Lumen improvement after ELCA (1.4±0.5 to 2.7±0.8 mm2) was the result of both tissue ablation (decrease in P+M CSA from 16.8±7.1 to 15.9±6.7 mm2, P<.0001) and vessel expansion (increase in EEM CSA from 18.2±7.1 to 18.6±6.8 mm2, P=.0245), with no change in calcium. The decrease in P+M CSA was 39% of the CSA of the laser catheter used. Dissections were present in 39% of lesions, 84% within superficial calcium; fibrocalcific deposits developed a "fragmented" appearance.
Conclusions ELCA increased lumen CSA by both atheroablation and vessel expansion without calcium ablation. Superficial fibrocalcific deposits developed a characteristic fragmented appearance. These findings support both photoablation and forced vessel expansion as mechanisms of lumen enlargement and plaque dissection after ELCA.
Key Words: ultrasonics lasers angioplasty coronary disease
| Introduction |
|---|
|
|
|---|
IVUS allows transmural imaging of coronary arteries in humans in vivo, providing unique insight into the pathology of coronary artery disease by defining vessel wall geometry and the major components of the atherosclerotic plaque. Serial IVUS studies have been used to investigate the mechanisms of transcatheter therapy in humans in vivo.8 9 10 11 For example, several studies have delineated the mechanisms of balloon angioplasty,8 9 12 including the reasons for postangioplasty dissections.13 Other studies have used sequential (preintervention and postintervention) IVUS imaging to assess the contribution of tissue removal to lumen improvement after both DCA and rotational atherectomy.10 11
The purpose of this study was to use sequential (preintervention and postintervention) IVUS imaging to assess the mechanisms of lumen improvement after ELCA of coronary artery disease in humans in vivo.
| Methods |
|---|
|
|
|---|
The 37 men and 11 women had a mean age of 63±11 years. Twelve patients presented with stable angina pectoris; the rest had unstable angina, of whom 4 had postinfarction angina. The imaged and treated vessels were left main in 2, left anterior descending in 20, left circumflex in 9, and right coronary artery in 9 and saphenous vein grafts in 9 patients. Nine lesions were aortoostial in location.
All patients were treated and studied after giving informed consent. ELCA and IVUS imaging protocols have the ongoing approval of the Washington Hospital Center Institutional Review Board.
ELCA Procedure
The ELCA procedure (Spectranetics/Advanced
Interventional
Systems) was performed as described
elsewhere.1 2 3 14 15
The largest laser fiber catheter used was a 1.3-mm catheter in 6,
1.6-mm catheter in 6, 1.8-mm directional laser catheter in 32, 2.0-mm
catheter in 3, and 2.2-mm catheter in 2 lesions. Energy densities
ranged from 25 to 65 mJ/mm2 (mean, 57.9±5.4
mJ/mm2), and the number of pulses ranged from 30 to 1880
(mean, 339±352). A single laser pass technique was used in 69% of the
lesions. Adjunct devices were used in all lesions: PTCA in 31 and DCA
in 18.
Angiographic Analysis
Angiograms were reviewed by a core
angiographic laboratory that
was blinded to the ultrasound results. Standard qualitative
morphological criteria were recorded on the basis of their
identification in any unforeshortened view.16
Calcification was identified as readily apparent radiopacities within
the vascular wall at the site of the stenosis and was
classified as none to mild, moderate (radiopacities noted only during
the cardiac cycle before contrast injection), or severe (radiopacities
noted without cardiac motion before contrast injection, generally
compromising both sides of the arterial lumen). An
eccentric target lesion had one lumen edge in the outer one quarter of
the apparently normal lumen.
Target lesion location was designated as ostial, proximal, mid, and distal. Ostial lesions were those that began within 3 mm of a major coronary ostium.
Quantitative angiographic analysis was performed with a computer-assisted, automated edge detection algorithm (ImageComm).17 With the external diameter of the contrast-filled catheters used as the calibration standard, the minimal lumen diameter at end diastole before intervention was measured from multiple projections, and the results from the worst view were recorded. Lesion length was measured from shoulder to shoulder.
Intravascular Ultrasound Imaging
Intracoronary nitroglycerin
(100 to
200 µg) was administered before each IVUS imaging run. IVUS studies
were performed by use of one of two commercially available systems. In
each case, the ultrasound transducer was advanced beyond the target
lesion and withdrawn to the aortoostial junction in a single continuous
run without interruption. The first (Cardiovascular
Imaging Systems Inc/InterTherapy Inc) incorporated a single-element
25-MHz transducer and an angled mirror mounted on the tip of a flexible
shaft that was rotated at 1800 rpm within a 3.9F short monorail
polyethylene imaging sheath to form planar images in real time. The
second (Cardiovascular Imaging Systems Inc)
incorporated a single-element 30-MHz beveled transducer within
either a 2.9F long monorail imaging catheter having a common distal
lumen design (the distal lumen accommodates either the imaging core or
the guide wire but not both) or a 3.2F short monorail imaging catheter.
With both systems, the transducer was withdrawn automatically at 0.5
mm/s to perform the imaging sequence. IVUS studies were recorded on
1/2-in high-resolution super VHS tape for off-line
analysis.
Qualitative and Quantitative IVUS Analyses
The same anatomic
image slice was analyzed before
intervention, after ELCA, and after adjunct device use, and the
differences were compared. By use of one or more reproducible axial
landmarks (eg, the aortoostial junction and/or a large proximal or
distal side branch) and the known pullback speed of 0.5 mm/s, identical
cross-sectional slices on serial studies could be identified for
comparison. Qualitative (plaque morphology) and quantitative
(cross-sectional) analyses of the ultrasound images were
performed by a single individual blinded to the angiographic and
clinical results.
The in vitro validation of qualitative and quantitative IVUS analysis was reported previously.18 19 20 21 22 23 24 25 Single images were digitized, and the following lesion site measurements were made with computer planimetry: EEM CSA (in square millimeters); lumen CSA (in square millimeters); P+M CSA (in square millimeters), which is EEM minus lumen area; and minimum lumen diameter (in millimeters).
The EEM is shorthand for the media-adventitia border, which has been shown to be a reproducible measurement of total vessel CSA. When the atherosclerotic plaque encompassed the catheter, the lumen was assumed to be the size of the imaging catheter. Because medial thickness could not be measured accurately, P+M CSA was used as a measurement of the atherosclerotic plaque. Each border (EEM and lumen) is routinely traced two to four times, and the results are averaged.
In practice, the postintervention images are analyzed first to determine the image slice with the smallest final lumen CSA; if multiple image slices have the same lumen CSA, then the image slice with the largest plaque CSA and the most well-defined media-adventitia border is selected for analysis. (At a pullback speed of 0.5 mm/s, there are 60 video frames per millimeter of arterial length from which to select.) Second, the axial relation of this image slice to various longitudinal landmarks is studied, and the nearest longitudinal landmarks (well-defined side branches or peculiar and unique arcs of calcium or perivascular structures) that are repeatedly identifiable on multiple imaging runs are noted. The distances from the selected image slice to these axial landmarks are measured (from seconds of videotape where 2 seconds of videotape equals 1 mm of axial length, given a motorized transducer pullback speed of 0.5 mm/s). Next, this process is reversed to identify the preintervention image slice that corresponds to the selected postintervention image slice. (The axial landmarks are identified first, and the videotape is advanced or rewound the requisite number of seconds. The preintervention study is then analyzed frame by frame to identify the preintervention image that corresponds to the postintervention image.) Finally, the preintervention and postintervention image slices are compared visually to ensure that the image slices are indeed from the same anatomic location within the lesion. When the same lesions are studied at least 3 months apart, the intraclass correlation coefficient for each of these measurements is >.90. The intraclass correlation coefficient considers between-lesion and within-lesion variabilities and is widely used as a measure of interrater variability.26 27 Specifically, the intraclass correlation coefficient for repeated preintervention measurement of the EEM CSA is .99; of the lumen CSA, .96; and of the P+M CSA, .99. The intraclass correlation coefficient for repeated postintervention measurement of the EEM CSA is .99; of the lumen CSA, .92; and of P+M CSA, .98. In our laboratory, this protocol has been used to study acute and chronic transcatheter device effects in >2500 lesions. Some of these analyses have been reported previously.8 10 11 28 29 30 31 32
Because calcium produces bright echoes (brighter than the vessel adventitia) with acoustic shadowing of deeper structures, the measurement of the EEM CSA can sometimes be difficult. To circumvent this problem, two types of extrapolation can be used. Briefly, because the cross section of the coronary artery was more or less circular, extrapolation of the circumference of the EEM was possible, provided that each calcific deposit did not shadow >60° of the adventitial circumference. Also, real-time axial movement of the transducer just distal and proximal to a calcific deposit or to find the smallest circumferential arc of calcium within a large calcific deposit unmasked and filled in contiguous parts of the adventitia that were otherwise shadowed by that deposit. In this study, lesions that required circumferential extrapolation of >30° or axial transducer movement of >1 mm (2 seconds or 60 frames of videotape) were eliminated. Over this short length of coronary artery, there is a negligible change in EEM CSA when IVUS imaging is used in vivo.33 These techniques have been described previously.11 28 The reproducibility of the above analyses included lesions with target lesion calcification in which measurement of the EEM CSA required circumferential or axial extrapolation.
Plaque composition was assessed for the presence and extent of calcium. Calcium was identified as plaque that was brighter than the reference adventitia with acoustic shadowing of deeper arterial structures.34 Because of acoustic shadowing, the thickness of the calcific deposit could not be measured. Therefore, calcium was quantified by its total circumference (expressed as an arc, in degrees, measured with a protractor centered on the lumen). The location of target lesion calcium was defined as superficial (calcium at the intima-lumen interface or closer to the lumen than to the adventitia), deep (calcium at the media-adventitia border or closer to the adventitia than to the lumen), or both. The largest arc of superficial calcium within the lesion was also measured as above. To assess calcium removal, ultrasound images before and after ELCA, were compared quantitatively (to detect a change in the total or superficial arc of calcium) and qualitatively (ability to visualize deeper arterial structures that could not be seen before intervention).
Dissections or tears in the plaque were abrupt, focal interruptions in the continuity of the plaque or intima that spanned normal tissue planes radially, axially, or circumferentially but did not necessarily extend to the media. Dissection planes were classified according to location: within calcium, at the junction of calcified and noncalcified plaque, and within noncalcified plaque.
Statistical Analysis
Statistical analysis was performed with
statview version
4.01. Quantitative data were presented as mean±SD. Qualitative
data were presented as frequencies. Comparisons between groups
were performed by use of paired and unpaired t tests for
continuous variables or
2 statistics and
Fisher's exact test for categorical variables. Sequential
measurements were compared by use of ANOVA for repeated measures with
post hoc analysis and by use of Fisher's protected
least-significant differences test. The level of significance was
defined as P<.05.
| Results |
|---|
|
|
|---|
|
IVUS Analysis After ELCA
By IVUS analysis, 36 (73%) of the
lesions contained
calcium (P<.0001 versus angiography), and 29 (59%) of the
lesions contained superficial calcium.
A comparison of IVUS images in
the 49 lesions studied both before and
after ELCA is shown in Table 2
and Fig 1
.
The lumen improvement after ELCA (from 1.4±0.5 to 2.7±0.8
mm2) was the result of both tissue ablation (a decrease in
P+M CSA from 16.8±7.1 to 15.9±6.7 mm2,
P<.0001) and vessel expansion (increase in EEM CSA from
18.2±7.1 to 18.6±6.8 mm2, P=.0245).
Despite these small changes, 96% of the lesions showed an increase in
lumen CSA, 67% showed an increase in EEM CSA, and 78% showed a
decrease in P+M CSA. There was no change in the arc of calcium or of
superficial calcium, nor was there subjective evidence of calcium
removal (eg, increased ultrasound penetration of deeper
arterial structures).
|
|
After ELCA, the IVUS measurement of minimum lumen diameter correlated only fairly with the quantitative angiography (r=.433, P=.0059).
The post-ELCA lumen CSA averaged 17% larger
than the CSA of the
largest laser catheter used; the decrease in P+M CSA averaged only 39%
of the CSA of the largest laser catheter used. However, these were
extremely variable and were not predicted by any angiographic or
IVUS lesion characteristic. In some lesions, lumen improvement was due
entirely to plaque ablation (Fig 2
); in
others, it was due entirely to vessel expansion (Fig 3
).
|
|
Dissections were present in 39% of the lesions after ELCA. Most of
the dissections (84%) were within superficial calcified deposits;
after ELCA, calcified deposits often developed a characteristic
"fragmented" or "shattered" appearance: newly created,
sharp-edged gaps within a previously solid calcium deposit (Fig
4
). The resulting calcium fragments often
were displaced circumferentially. Conversely, dissections occurred in
14% of the lesions that contained only deep calcium and in 23% that
contained no calcium (P=.0471).
|
IVUS Analysis After Adjunct Device Use
Table 3
summarizes the quantitative IVUS
analysis after adjunct device therapy. The mechanism of
progressive lumen enlargement after adjunctive PTCA was vessel
expansion (increase in EEM CSA); after adjunctive DCA, it was a
combination of vessel expansion and additional atheroablation (decrease
in P+M CSA). There was no significant change in lesion
calcification.
|
Overall, the number of lesions with dissections increased to 30 of 41 (73%), and there was almost a threefold increase in the number of dissection planes per lesion site. New dissections were almost entirely at the junction of calcified and noncalcified plaque.
| Discussion |
|---|
|
|
|---|
IVUS imaging provides high-quality tomographic images of the coronary artery lumen, lumen-intima interface, atherosclerotic plaque, and vessel wall in vivo. It has proved useful for evaluating mechanisms and results of various transcatheter therapies, including PTCA, DCA, rotational atherectomy, and endovascular stent placement.8 9 10 11 12 13 (For example, all of the lumen improvement after rotational atherectomy has been shown to be the result of tissue ablation11 ; 75% of the lumen improvement after primary DCA has been shown to be the result of tissue removal.9 10 ) IVUS images acquired before intervention are essential for the accurate assessment of changes in plaque morphology that occur in direct response to intervention. We have been routinely obtaining IVUS images in our patients with angiographically severe stenoses before transcatheter interventions without adverse sequelae.38 Similarly, sequential imaging has been shown to be essential for separating primary and adjunct device effects.11 The use of the motorized pullback device assisted in comparing sequential imaging studies. The steps used to make these measurements are described in detail above. Because the transducer always was pulled back to an easily reproducible landmark (eg, the aortoostial junction, a large proximal side branch, or a characteristic proximal calcific deposit), the same tomographic image slice could be identified after each imaging run. This facilitated comparative measurements.
Baseline Lesion Characteristics
Patients treated with ELCA in
this study had stenoses that
were angiographically severe (minimum lumen diameter, 0.86±0.51 mm
with 6% total occlusions) and moderately long (9.8±5.4 mm). The
severity, length, complexity, and location of these target lesions
before intervention were comparable to those reported in other studies
using angiography to evaluate the effect of
ELCA.1 2 3 14 15
Mechanisms of Lumen Improvement After ELCA
Comparison of the
sequential IVUS images before and after ELCA
showed that the contribution of tissue ablation (decrease in P+M CSA)
to lumen improvement averaged 76%, and the contribution of vessel
expansion (increase in vessel or EEM CSA) averaged 24%. Furthermore,
the decrease in P+M CSA averaged only 39% of the CSA of the laser
catheter. There were two potential mechanisms of ELCA-induced vessel
expansion. The presence of dead space between the laser fibers may have
contributed to inadequate tissue ablation and could have resulted in a
Dotter effect.37 However, a Dotter effect would not have
produced a lumen larger than the laser catheter used.
Alternatively, laser-induced shock waves and forceful expansion of vapor bubbles into tissue could have caused acute vessel expansion.37 38 39 The relative contributions of tissue ablation and vessel expansion to lumen enlargement after ELCA varied among lesions. It was not possible to relate the relative contributions of these two potential mechanisms to any qualitative or quantitative angiographic or ultrasound variable.
Sequential imaging shows that the magnitude of lumen improvement after ELCA is at most modest and perhaps less than suggested by angiography. The only fair correlation between quantitative angiographic and IVUS assessment of post-ELCA minimum lumen diameter (r=.433) may explain some of this discrepancy.
Sequential IVUS analysis consistently failed to demonstrate calcium ablation either quantitatively (by a measurable decrease in the arc of calcium or of superficial calcium) or qualitatively (eg, visual evidence of reduced shadowing and increased ultrasound penetration into deeper tissue planes). This is in distinct contrast to sequential IVUS analysis before and after rotational atherectomy, which was able to demonstrate quantitative and qualitative calcium ablation.11
Dissections after ELCA almost always occurred within superficial fibrocalcific plaque. The visual appearance of these dissections (a fragmenting of superficial fibrocalcific deposits) was, in our experience, unique to ELCA. PTCA caused dissections at the junction of calcified and noncalcified plaque.13 Rotational atherectomy caused fissuring of superficial calcium in approximately 25% of lesions.11 The fissures caused by rotational atherectomy differed from the ELCA-induced fragmentation of calcific deposits in that fissures were barely perceptible hairline spaces or cracks within otherwise unbroken and displaced calcium deposits. In experimental models, it has been shown that ELCA-induced vapor bubble expansion caused significant dissections.38 39 Thus, forced lumen and vessel expansion and plaque fragmentation after ELCA in human coronary arteries may be related.40
Mechanisms of Adjunctive Device Therapy
Although it was not
the main purpose of this study to
evaluate and compare the mechanisms of adjunctive PTCA and DCA after
ELCA, several observations are noteworthy. Angiographic minimum lumen
diameters and IVUS lumen CSA were larger and angiographic percent
diameter stenoses were smaller after adjunct DCA than after
adjunct PTCA. The mechanism of progressive lumen improvement after
adjunct PTCA was additional vessel expansion; after adjunct DCA, it was
a combination of vessel expansion and tissue removal. After both
adjunctive PTCA and DCA, there was an increase in frequency and number
of dissections, primarily at the junction of calcified and noncalcified
plaque. Thus, the mechanism of adjunctive PTCA was similar to that of
primary PTCA, and the mechanism of adjunctive DCA was similar to that
of primary
DCA.8 9 10 11 12 13 41
Study Limitations
This study presented a heterogeneous
patient
and lesion population, including both native vessel and vein graft
lesion location. However, the population was similar to that of most
reports using angiography to assess the effects of ELCA. The changes
observed were small, perhaps smaller than the resolution of sequential
IVUS analysis. Furthermore, these observations depend on the
ability of IVUS imaging to sequentially study the same anatomic image
slice. However, this methodology has been used to study other
transcatheter
devices8 10 11 28 29 30 31 32 ;
extreme
care was taken to analyze the same cross section on repeated
studies. Furthermore, the small changes noted only served to emphasize
the conclusion that lumen improvement and atheroablation after ELCA
are, at most, modest. Mostly small (
1.8 mm) laser fiber catheters
were used. A greater degree of tissue ablation might have been
demonstrated if 1.3-mm catheters had been consistently followed
by
2.0-mm catheters.42 However, the treatment strategy
was designed to minimize dissections and complications; the
angiographic percent diameter stenosis after ELCA alone
averaged 50%.
Conclusions
ELCA increased lumen CSA by a combination of
atheroablation and
vessel expansion. The amount of tissue ablation was small,
variable, and averaged
40% of the laser catheter CSA. A
concomitant increase in vessel CSA accounted for a lumen CSA larger
than that of the laser catheter. Dissections occurred within
superficial fibrocalcific plaque that developed a characteristic
fragmented appearance. These findings support both photoablation and
forced vessel expansion (mediated by either acoustic shock wave or
vapor bubble) as mechanisms of lumen enlargement and plaque dissection
after ELCA in humans in vivo.
| Selected Abbreviations and Acronyms |
|---|
|
| Acknowledgments |
|---|
| Footnotes |
|---|
Received August 3, 1994; revision received August 1, 1995; accepted August 6, 1995.
| References |
|---|
|
|
|---|
2. Bittl JA, Sanborn TA, Tcheng JE, Siegel RM, Ellis SG. Clinical success, complications and restenosis rates with excimer laser coronary angioplasty. Am J Cardiol. 1992;70:1533-1539. [Medline] [Order article via Infotrieve]
3.
Cook SL, Eigler NL, Shefer A, Goldenberg T, Forrester
JS, Litvack F. Percutaneous excimer laser
coronary angioplasty of lesions not ideal for balloon
angioplasty. Circulation. 1991;84:632-643.
4. Grundfest WS, Litvack F, Forrester JS, Goldenberg T, Swan HJC, Morgenstern L, Fishbein M, McDermid IS, Rider DM, Pacala TJ, Laudenslager JB. Laser ablation of human atherosclerotic plaque without adjacent tissue injury. J Am Coll Cardiol. 1985;5:929-933. [Abstract]
5. Litvack F, Grundfest WS, Segalowitz J, Papaioannou T, Goldenberg T, Laudenslager J, Hestrin L, Forrester J, Eigler NS, Cook SA. Interventional cardiovascular therapy by laser and thermal angioplasty. Circulation. 1990;81(suppl IV):IV-109-IV-116.
6. Litvack F, Grundfest WS, Papaioannou T, Mohr FW, Jakubowski AT, Forrester JS. Role of laser and thermal ablation devices in the treatment of vascular diseases. Am J Cardiol. 1988;61:81G-86G. [Medline] [Order article via Infotrieve]
7. Shelton ME, Hoxworth B, Shelton JA, Virmani R, Friesinger GC. A new model to study quantitative effects of laser angioplasty on human atherosclerotic plaque. J Am Coll Cardiol. 1986;7:909-915. [Abstract]
8. Potkin BN, Keren G, Mintz GS, Douek P, Pichard AD, Satler LF, Kent KM, Leon MB. Arterial responses to balloon coronary angioplasty: an intravascular ultrasound study. J Am Coll Cardiol. 1992;20:942-951. [Abstract]
9. Tenaglia AN, Buller CE, Kisslo KB, Stack RS, Davidson CJ. Mechanisms of balloon angioplasty and directional coronary atherectomy as assessed by intracoronary ultrasound. J Am Coll Cardiol. 1992;20:685-691. [Abstract]
10. Matar FA, Mintz GS, Farb A, Douek P, Pichard AD, Kent KM, Satler LF, Popma JJ, Keller MB, Pinnow E, Merritt A, Lindsay JJ, Leon MB. The contribution of tissue removal to lumen improvement after directional coronary atherectomy. Am J Cardiol. 1994;74:647-650. [Medline] [Order article via Infotrieve]
11. Kovach JA, Mintz GS, Pichard AD, Kent KM, Popma JJ, Satler LF, Leon MB. Sequential intravascular ultrasound characterization of the mechanisms of rotational atherectomy and adjunct balloon angioplasty. J Am Coll Cardiol. 1993;22:1024-1032. [Abstract]
12.
Tobis JM, Mallery JA, Gessert J, Griffith J, Mahon D,
Bessen M, Moriuchi M, McLeay L, McRae M, Henry WL. Intravascular
ultrasound cross-sectional arterial imaging before and
after balloon angioplasty in vitro.
Circulation. 1989;80:873-882.
13.
Fitzgerald PF, Ports TA, Yock PG. Contribution
of localized calcium deposits to dissection after angioplasty: an
observational study using intravascular ultrasound.
Circulation. 1992;86:64-70.
14. Litvack F, Eigler NL, Margolis JR, Grundfest WS, Rothbaum D, Linnemeier T, Hestrin LB, Tsoi D, Cook SL, Krauthamer D, Goldenberg T, Laudenslager JR, Segalowitz J, Forrester JS. Percutaneous excimer laser coronary angioplasty. Am J Cardiol. 1990;66:1027-1032. [Medline] [Order article via Infotrieve]
15. Sanborn TA, Bittl JA, Hershman RA, Siegel RS. Percutaneous coronary excimer laser-assisted angioplasty: initial multicenter experience in 141 patients. J Am Coll Cardiol. 1991;17:169B-173B.
16. Popma JA, Bashore TD. Qualitative and quantitative angiography. In: Topol E, ed. Textbook of Interventional Cardiology. Philadelphia, PA: WB Saunders; 1993:1052-1068.
17. Mancini GBL, Simon SB, McGillem MJ, LeFree MT, Friedman HZ, Vogel RA. Automated quantitative coronary angiography: morphologic and physiologic validation of a rapid digital angiographic method. Circulation. 1987;76:452-460.
18. Pandian NG, Kreis A, Brockway B, Isner JM, Sacharoff A, Boleza E, Caro R, Muller D. Ultrasound angioscopy: real-time, two-dimensional, intraluminal ultrasound imaging of blood vessels. Am J Cardiol. 1988;62:493-494. [Medline] [Order article via Infotrieve]
19. Hodgson JM, Graham SP, Sarakus AD, Dame SG, Stephens DN, Dhillon PS, Brands D, Sheehan H, Eberle MJ. Clinical percutaneous imaging of coronary anatomy using an over-the-wire ultrasound catheter system. Int J Cardiac Imaging. 1989;4:186-193.
20. Gussenhoven EJ, Essed CE, Lancee CT, Mastik F, Frietman P, van Egmond FC, Reiber J, Bosch H, van Urk H, Roelandt J, Bom N. Arterial wall characteristics determined by intravascular ultrasound imaging: an in vitro study. J Am Coll Cardiol. 1989;14:947-952. [Abstract]
21. Nishimura RA, Edwards WD, Warnes CA, Reeder GS, Holmes DJ Jr, Tajik AJ, Yock PG. Intravascular ultrasound imaging: in vitro validation and pathologic correlation. J Am Coll Cardiol. 1990;16:145-154. [Abstract]
22.
Potkin BN, Bartorelli AL, Gessert JM, Neville RF,
Almagor Y, Roberts WC, Leon MB. Coronary artery imaging
with intravascular high-frequency ultrasound.
Circulation. 1990;81:1575-1585.
23.
Nissen SE, Grines CL, Gurley JC, Sublett K, Haynie D,
Diaz C, Booth DC, DeMaria AN. Application of a new
phased-array ultrasound imaging catheter in the assessment of
vascular dimensions: in vivo comparison to cineangiography.
Circulation. 1990;81:660-666.
24.
Tobis JM, Mallery J, Mahon D, Lehmann K, Zalesky P,
Griffith J, Gessert J, Moriuchi M, McRae M, Dwyer ML, Greep N, Henry
WL. Intravascular ultrasound imaging of human coronary
arteries in vivo: analysis of tissue characterizations with
comparison to in vitro histological specimens.
Circulation. 1991;83:913-926.
25.
Nissen SE, Gurley JC, Grines CL, Booth DC, McClure R,
Berk M, Fischer C, DeMaria AN. Intravascular ultrasound
assessment of lumen size and wall morphology in normal subjects and
patients with coronary artery disease.
Circulation. 1991;84:1087-1099.
26. Ebel RL. Estimation of the reliability of rating. Psychometrika. 1951;16:407-424.
27. Mintz GS, Griffin J, Chuang YC, Pichard AD, Kent KM, Satler LF, Popma JJ, Leon MB. The reproducibility of the intravascular ultrasound assessment of stent implantation in saphenous vein grafts. Am J Cardiol. 1995;75:1267-1270. [Medline] [Order article via Infotrieve]
28.
Mintz GS, Potkin BN, Keren G, Satler LF, Pichard AD,
Kent KM, Popma JJ, Leon MB. Intravascular ultrasound evaluation
of the effect of rotational atherectomy in obstructive atherosclerotic
coronary artery disease.
Circulation. 1992;86:1383-1393.
29. Matar FA, Mintz GS, Pinnow E, Javier SP, Popma JJ, Kent KM, Satler LF, Pichard AD, Leon MB. Multivariate predictors of intravascular ultrasound endpoints after directional coronary atherectomy. J Am Coll Cardiol. 1995;25:318-324. [Abstract]
30. Painter JA, Mintz GS, Wong SC, Popma JJ, Pichard AD, Kent KM, Satler LF, Leon MB. Serial intravascular ultrasound studies fail to show evidence of chronic Palmaz-Schatz stent recoil. Am J Cardiol. 1995;75:398-400. [Medline] [Order article via Infotrieve]
31. Dussaillant GR, Mintz GS, Popma JJ, Kent KM, Pichard AD, Satler LF, Leon MB. Intravascular ultrasound, directional coronary atherectomy, and the Optimal Atherectomy Restenosis Trial (OARS). Coron Artery Dis. In press.
32. Dussaillant GR, Mintz GS, Pichard AD, Kent KM, Satler LF, Popma JJ, Wong SC, Leon MB. Small stent size and intimal hyperplasia contribute to restenosis: a volumetric intravascular ultrasound analysis. J Am Coll Cardiol. 1995;26:720-724. [Abstract]
33. Javier SP, Mintz GS, Popma JJ, Pichard AD, Kent KM, Satler LF, Leon MB. Intravascular ultrasound assessment of the magnitude and mechanism of coronary artery and lumen tapering. Am J Cardiol. 1995;75:177-180.[Medline] [Order article via Infotrieve]
34. Mintz GS, Douek P, Pichard AD, Kent KM, Satler LF, Popma JJ, Leon MB. Target lesion calcification in coronary artery disease: an intravascular ultrasound study. J Am Coll Cardiol. 1992;20:1149-1155. [Abstract]
35. Rosenfeldt FL, Chi L, Black AJR, Waugh JR, Pedersen JS, Levatter J. Excimer laser angioplasty in the atherosclerotic rabbit: comparison with balloon angioplasty. Am Heart J. 1992;124:349-355. [Medline] [Order article via Infotrieve]
36.
Isner JM, Rosenfield K, Losordo DW. Eximer laser
atherectomy (the greening of Sisyphus).
Circulation. 1990;81:2018-2020.
37. Isner JM, Pickering JG, Mosseri M. Laser-induced dissections: pathogenesis and implications for therapy. J Am Coll Cardiol. 1992;19:1619-1621. [Medline] [Order article via Infotrieve]
38. van Leeuwen TG, van Erven L, Meertens JH, Motamedi M, Post MJ, Borst C. Origin of arterial wall dissections induced by pulsed excimer and mid-infrared laser ablation in the pig. J Am Coll Cardiol. 1992;19:1610-1618. [Abstract]
39.
van Leeuwen TG, Meertens JH, Velema E, Post MJ, Borst
C. Intraluminal vapor bubble induced by excimer laser pulse
causes microsecond arterial dilation and invagination
leading to extensive wall damage in the rabbit.
Circulation. 1993;87:1258-1263.
40. Mintz GS, Pichard AD, Kovach JA, Kent KM, Satler LF, Javier SP, Popma JJ, Leon MB. Impact of pre-intervention intravascular ultrasound imaging on transcatheter treatment stategies in coronary artery disease. Am J Cardiol. 1994;73:423-430. [Medline] [Order article via Infotrieve]
41. Mintz GS, Pichard AD, Popma JJ, Kent KM, Satler LF, Leon MB. Preliminary experience with adjunct directional coronary atherectomy following high-speed rotational atherectomy in the treatment of calcific coronary artery disease. Am J Cardiol. 1993;71:799-804. [Medline] [Order article via Infotrieve]
42. Freed M, Lichtenberg A, May MA, Juran N, Grines CL, O'Neill WW. Are residual stenoses after eximer laser angioplasty and coronary atherectomy due to inefficient or small devices? Comparison with balloon angioplasty. J Am Coll Cardiol. 1993;22:1628-1634.[Abstract]
This article has been cited by other articles:
![]() |
J. M. Ahmed, G. S. Mintz, N. J. Weissman, A. J. Lansky, A. D. Pichard, L. F. Satler, and K. M. Kent Mechanism of Lumen Enlargement During Intracoronary Stent Implantation : An Intravascular Ultrasound Study Circulation, July 4, 2000; 102(1): 7 - 10. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Koster, J. Kahler, W. Terres, J. Reimers, S. Baldus, D. Hartig, J.u. Berger, T. Meinertz, and C. W. Hamm Six-month clinical and angiographic outcome after successful excimer laser angioplasty for in-stent restenosis J. Am. Coll. Cardiol., July 1, 2000; 36(1): 69 - 74. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Mehran, G. Dangas, G. S. Mintz, R. Waksman, A. Abizaid, L. F. Satler, A. D. Pichard, K. M. Kent, A. J. Lansky, G. W. Stone, et al. Treatment of In-Stent Restenosis With Excimer Laser Coronary Angioplasty Versus Rotational Atherectomy : Comparative Mechanisms and Results Circulation, May 30, 2000; 101(21): 2484 - 2489. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Mehran, G. S. Mintz, L. F. Satler, A. D. Pichard, K. M. Kent, T. A. Bucher, J. J. Popma, and M. B. Leon Treatment of In-Stent Restenosis With Excimer Laser Coronary Angioplasty : Mechanisms and Results Compared With PTCA Alone Circulation, October 7, 1997; 96(7): 2183 - 2189. [Abstract] [Full Text] |
||||
![]() |
N. Reifart, M. Vandormael, M. Krajcar, S. Gohring, W. Preusler, F. Schwarz, H. Storger, M. Hofmann, J. Klopper, S. Muller, et al. Randomized Comparison of Angioplasty of Complex Coronary Lesions at a Single Center : Excimer Laser, Rotational Atherectomy, and Balloon Angioplasty Comparison (ERBAC) Study Circulation, July 1, 1997; 96(1): 91 - 98. [Abstract] [Full Text] |
||||
![]() |
G. S. Mintz, K. M. Kent, A. D. Pichard, L. F. Satler, J. J. Popma, and M. B. Leon Contribution of Inadequate Arterial Remodeling to the Development of Focal Coronary Artery Stenoses : An Intravascular Ultrasound Study Circulation, April 1, 1997; 95(7): 1791 - 1798. [Abstract] [Full Text] |
||||
![]() |
J. A. Bittl Advances in Coronary Angioplasty N. Engl. J. Med., October 24, 1996; 335(17): 1290 - 1302. [Full Text] [PDF] |
||||
![]() |
G. S. Mintz, J. J. Popma, A. D. Pichard, K. M. Kent, L. F. Satler, S. Chiu Wong, M. K. Hong, J. A. Kovach, and M. B. Leon Arterial Remodeling After Coronary Angioplasty : A Serial Intravascular Ultrasound Study Circulation, July 1, 1996; 94(1): 35 - 43. [Abstract] [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1995 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |