(Circulation. 1995;92:2940-2943.)
© 1995 American Heart Association, Inc.
Articles |
From the Division of Cardiology, Department of Medicine, University of Washington School of Medicine, Seattle.
Correspondence to Gust H. Bardy, MD, Box 356422, University of Washington Medical Center, Seattle, WA 98195.
| Abstract |
|---|
|
|
|---|
Methods and Results Twenty-four consecutive patients
underwent prospective, randomized evaluation of the effect of ICD can
size on defibrillation efficacy during standard ICD surgery. Each
patient had the unipolar defibrillation threshold (DFT) measured with
80-cc, 60-cc, or 40-cc active can placed in the left subcutaneous
infraclavicular region. The system included a 10.5F tripolar right
ventricular electrode that served as the shock anode. The
shock waveform used in each instance was a single capacitor biphasic
65% tilt pulse delivered from a 120-µF capacitor. Stored energy at
the DFT for the 80-cc, 60-cc, and 40-cc cans were 8.1±4.7 J,
8.7±5.8
J, and 9.5±4.8 J, respectively. There was no statistical significant
difference between the DFTs for the three unipolar can electrodes
(P=.39). Leading edge voltage also did not differ
significantly among the three unipolar cans (356±92 V, 365±110
V, and
387±94 V, respectively, P=.29). There was, however, a
slight progressive increase in resistance with decreasing can size
(57±7
, 60±7
, and 65±9
, respectively,
P<.001).
Conclusions Decreasing can volume from 80 cc to 60 cc to 40 cc does not compromise unipolar defibrillation efficacy despite a slight rise in shock resistance. These findings indicate that technological advances that allow for smaller-volume ICDs will not compromise defibrillation efficacy for unipolar systems.
Key Words: death sudden fibrillation implantable cardioverter-defibrillator defibrillation
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
Defibrillation Testing
A tripolar 10.5F right ventricular
transvenous lead
(Medtronic model 6936) was the endocardial component of the unipolar
ICD system. This lead incorporated a 5-cm-long coil defibrillation
electrode with standard bipolar pace/sense electrodes at the tip. The
active can component of the unipolar ICD system was provided by 80-cc,
60-cc, and 40-cc titanium cans modeled after the Medtronic model 7219C
unipolar ICD. The surface areas of the active can electrodes were 112
cm2, 88 cm2, and 62
cm2, respectively. The active can was positioned
subcutaneously on the anterior fascia of the left pectoralis major
muscle, 2 to 3 cm inferior to the left clavicle and 2 to 3
cm medial to the humoral head (Fig 1
). The long axis of
the can was placed parallel to the long axis of the body, with the
opening of the portals in the connector block directed medially. The
skin edges of the incision site were approximated with multiple towel
clips to ensure that the entire can electrode was encapsulated by
tissue during testing. All air and serosanguinous fluid were expressed
or aspirated before defibrillation testing. Finally, all defibrillation
pulses were delivered at end expiration to avoid the possible
variable effect of respiration on pulse impedance.
|
Defibrillation Threshold Testing
Testing of the single-lead
unipolar defibrillation system
with each of the three active can electrodes was done in a prospective,
randomized fashion with the use of a 120-µF capacitor, asymmetrical
biphasic pulse delivered at 65% tilt for both phases. Order of can
size testing was based on a numerical randomization scheme derived from
the patient's hospital number. After determining the order of testing,
each can was inserted in the subcutanous pocket with closure of the
skin edges as described above. The right ventricular (RV)
electrode served as the anode for the initial phase of the biphasic
waveform in each instance. After VF was induced with alternating
current, defibrillation thresholds were obtained. The first
defibrillation pulse was given starting with a 400-V leading edge pulse
delivered 10 seconds after induction of VF. If the transvenous pulse
was unsuccessful, a 100- to 200-J transthoracic rescue
pulse was immediately delivered via a precharged external defibrillator
(Physio-control Lifepack 6s) between anterior-posterior
cutaneous pads (Darox Corporation).
Before reinduction of VF, a minimum rest period of 3 minutes was required. In addition, arterial blood pressure, heart rate, ECG ST segment morphology, and O2 saturation were monitored and were required to return to baseline values before reinduction of VF. If the initial 400-V shock was unsuccessful, pulse voltages were increased in 100-V steps up to 900 V. If the initial pulse shock was successful, shock strength was decreased in 50-V steps.
The defibrillation threshold was defined as the minimum energy that successfully terminated VF 10 seconds after its induction. After the defibrillation threshold was determined for one system, the remaining methods were tested in a likewise and consecutive fashion. The defibrillation threshold was measured only once for each method, given the concerns over repetitive fibrillation and defibrillation in human subjects.
Statistical Analysis
Repeated-measures ANOVA was used to
compare defibrillation
thresholds among the three active can electrodes as well as to evaluate
for linear correlation. Statistical significance was defined as a value
of P<.05. The population studied was designed to provide a
power of 80% to detect a difference of 2.5 J for defibrillation
threshold among the three can size groups.
| Results |
|---|
|
|
|---|
Defibrillation Threshold
Mean defibrillation threshold data
are shown in the
Table
. The mean defibrillation threshold stored energy
for the 80-cc can was 8.1±4.7 J, for the 60-cc can was 8.7±5.8
J, and
for the 40-cc can was 9.5±4.8 J. There was no statistically
significant difference among the three active can sizes in terms of
energy requirements for defibrillation (P=.39, Fig
2
). In addition, no significant linear correlation was
seen between can size and stored energy (P=.16). Similarly,
there was no statistically significant difference in leading edge
voltage (356±92 V, 365±110 V, and 387±94 V, respectively,
P=.29) or leading edge current (6.3±1.5 A,
6.2±1.8 A, and
6.0±8.8 A, respectively, P=.68) at the defibrillation
threshold among the three can sizes. There also was no linear
correlation between can size and leading edge voltage or current
(P=.10 and P=.35, respectively). There was,
however, a slight but significant difference in the pulsing resistance
at the defibrillation threshold between the different sized cans
(56.6±6.9
, 60.2±7.4
, and 64.5±8.8
,
respectively,
P<.001; Fig 3
). No significant difference in
defibrillation threshold was noted based on the order of testing: The
defibrillation threshold for the cans tested first was 7.9±3.8 J,
9.0±6.0 J for the cans tested second, and 9.4±5.3 J for those
tested
last (P=.63). The number of shocks delivered to measure the
defibrillation threshold for each can size was 3.2±0.6, 3.3±0.7,
and
3.3±0.8, respectively (P=.90).
|
|
|
With the use of a
defibrillation threshold
24 J as an implant
criterion, only one patient would not have met implantation criterion
and only in the case of the 60-cc size can (Fig 2
). This
finding is
probably consistent with statistical variation in
defibrillation efficacy.
| Discussion |
|---|
|
|
|---|
Previous defibrillation studies with epicardial lead systems, using either monophasic or biphasic pulses, have shown that the surface area of electrodes is a significant factor for defibrillation efficacy.9 10 11 12 Given that the current unipolar system uses the active can as an electrode, the surface area of the can could play a significant role in defibrillation efficacy. In this study, however, there was no significant difference in defibrillation efficacy among active cans varying in volume between 80 cc, 60 cc, and 40 cc with surface areas of 112 cm2, 88 cm2, and 62 cm2, respectively. The mean defibrillation threshold in each of the systems tested was below 10 J. In addition, there was no significant difference among the varying can sizes in leading edge voltage or current required for defibrillation. The one factor that did differ among the different sized cans was pathway resistance. Although the larger electrode afforded a lower pathway resistance, this did not result in improved defibrillation efficacy.
One explanation as to why there was no statistically significant change in the defibrillation threshold despite a 50% reduction in surface area may be related to the observation that the electrode tissue interface impedance is a small component of overall shock resistance, within practical limits.13 14 It should be noted, however, that electrode-tissue interface impedance is nonlinear.15 Consequently, as electrode surface area is reduced beyond that evaluated in this study, electrode-tissue interface impedance may become more important and lead to more significant differences in defibrillation efficacy than seen in this study. Hence, there probably will be limits as to how small an active can could be. The trend seen in our data suggests that eventually defibrillation energy could increase unacceptably as the can size falls below 40 cc.
The results of this study on ICD size also have implications for ICD energy output. Because defibrillator size is significantly affected by capacitor technology, it would not be possible to provide a 34-J, 40-cc can ICD at this time with the 120-µF K-film capacitor used in this study. Although smaller 60-µF and 90-µF K-film capacitors have a modest effect on improving defibrillation efficacy in unipolar ICD systems, it is not sufficient to decrease capacitor size and therefore ICD size to 40 cc while maintaining a 34-J output.16 17 The direct relationship between capacitor size and energy cancels the slight advantage provided by use of a smaller capacitance. Thus, without a change in the basic capacitor technology, a 40-cc unipolar active can system will not be practical in the near future unless the maximal output of the device is reduced to 25 J or less. This study therefore anticipates either a reduced output ICD or improved capacitor technology before ICDs can be smaller.
| Acknowledgments |
|---|
| Footnotes |
|---|
Received January 30, 1995; revision received April 24, 1995; accepted June 23, 1995.
| References |
|---|
|
|
|---|
2.
Yee R, Klein GH, Leithch JW, Guiraudon GM, Guiraudon
CM, Jones DL, Norris C. A permanent transvenous lead system for
an implantable pacemaker cardioverter-defibrillator: nonthoracotomy
approach to implantation. Circulation. 1992;85:196-204.
3.
Bardy GH, Allen MD, Mehra R, Johnson G, Feldman S,
Greene HL, Ivey TD. Transvenous defibrillation in humans via
coronary sinus. Circulation. 1990;81:1252-1259.
4.
Fromer M, Brachmann J, Block M. Efficacy
of automatic multimodal device therapy for ventricular
tachyarrhythmias as delivered by a new implantable
pacing cardioverter-defibrillator: results of a European
multicenter study of 102 implants.
Circulation. 1992;86:363-374.
5.
Bardy GH, Johnson G, Poole J, Dolack GL, Kudenchuk PJ,
Kelso D, Mitchell R, Mehra R, Hofer B. A simplified,
single-lead unipolar transvenous cardioversion-defibrillation
system. Circulation. 1993;88:543-547.
6.
Bardy GH, Dolack GL, Kudenchuk PJ, Poole JE, Johnson
G, Kelso D, Mitchell R, Mehra R, Hofer B. Prospective,
randomized comparison in humans of a unipolar defibrillation system
with that using an additional superior vena cava electrode.
Circulation. 1994;89:1090-1093.
7.
Kudenchuk PJ, Bardy GH, Dolack GL, Poole JE, Mehra R,
Johnson G. Efficacy of a single-lead unipolar defibrillator
compared with a system employing an additional coronary sinus
electrode. Circulation. 1994;89:2641-2644.
8. Bardy GH, Raitt MH, Jones GK. Unipolar defibrillation systems. In: Singer I, ed. Implantable Cardioverter-Defibrillator. Armonk, NY: Futura Publishing Co; 1994:365-376.
9.
Dixon EG, Tang AS, Wolf PD, Meador JT, Fine MJ, Calfee
RV, Ideker RE. Improved defibrillation thresholds with large
contoured epicardial electrodes and biphasic waveforms.
Circulation. 1987;76:1176-1184.
10. Troup PJ, Chapman PD, Olinger GN, Kleinman LH. The implanted defibrillator: relation of defibrillating lead configuration and clinical variables to defibrillation threshold. J Am Coll Cardiol. 1985;6:1315-1321. [Abstract]
11. Kallok MJ, Bourland JD, Tacker WA, Jones DL, Klein GJ, Wessale JL. Optimization of epicardial electrode size and implant site for reduced sequential pulse defibrillation thresholds. Med Instrm. 1986;20:36.
12. Mehra R, DeGroot P, Norenberg S. Three-dimensional finite element model of the heart for analysis of epicardial defibrillation: effect of surface area. PACE Pacing Clin Electrophysiol. 1989;12:652.
13. Jorgenson DB, Haynor GH, Bardy GH, Kim Y. Computational studies of transthoracic and transvenous defibrillation in detailed 3-D human thorax model. IEEE Trans Biomed Eng. 1994;42:172-184.
14. Schimpf PH, Johnson G, Jorgenson DB, Haynor DR, Bardy GH, Kim Y. Effects of electrode interface impedance on finite element modeling of transvenous defibrillation. Med Biol Eng Comput. In press.
15. Mehra R, Cybulski Z. Tachyarrhythmia termination: lead system and design. In: Singer I, ed. Implantable Cardioverter-Defibrillator. Armonk, NY: Futura Publishing Co; 1994:109-133.
16. Bardy GH, Poole JE, Kudenchuk PJH, Dolack GL, Mehra R, Raitt MH, Jones GK, Johnson G. A prospective randomized comparison in man of biphasic waveform 60-µF and 120-µF capacitance pulses using a unipolar defibrillation system. Circulation. 1995;90:90-95.
17. Bardy GH, Poole JE, Kudenchuk PJ, Dolack GL, Raitt MH, Jones GK, Mehra T, Troutman C, Anderson J, Johnson G. A prospective randomized comparison in humans of 90µF and 120µF 65% tilt biphasic pulse defibrillation using a unipolar pectoral transvenous defibrillation system. J Am Coll Cardiol. 1994;1A-484A:13A. Abstract.
This article has been cited by other articles:
![]() |
S. Windecker, R. E. Ideker, V. J. Plumb, G. N. Kay, G. P. Walcott, and A. E. Epstein The influence of ventricular fibrillation duration on defibrillation efficacy using biphasic waveforms in humans J. Am. Coll. Cardiol., January 1, 1999; 33(1): 33 - 38. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1995 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |