(Circulation. 1995;91:2204-2213.)
© 1995 American Heart Association, Inc.
Articles |
From the Cardiac Arrhythmia Service and the Cardiac Surgery Service, Harvard Medical School and the Massachusetts General Hospital, Boston.
Correspondence to Jeremy Ruskin, MD, Massachusetts General Hospital, Fruit St, Boston, MA 02114.
| Abstract |
|---|
|
|
|---|
Methods and Results The study group comprises the first 154 patients undergoing implantation of tiered-therapy ICDs at our institution. Pulse generators from three different manufacturers were used. In 39 patients, a complete nonthoracotomy lead system was used. The perioperative mortality was 1.3%. Of these 154 patients, 37% experienced late postoperative problems. Twenty-one patients required system revision within 36.5 months (mean, 8.57±11.3) of surgery. Reasons for revision were spurious shocks due to electrode fractures (3) or electrode adapter malfunction (2), inadequate signal from endocardial rate-sensing electrodes (3), superior vena cava or right ventricular coil migration (5), failure to correct tachyarrhythmias due to a postimplant rise in defibrillation threshold (5), or pulse generator failure (3). One of these patients required system removal for infection after revision of an endocardial lead. A further 32 patients received inappropriate shocks for atrial fibrillation with a rapid ventricular response or sinus tachycardia. Two of these patients also received shocks for ventricular tachycardia initiated by antitachycardia pacing triggered by atrial fibrillation. Ventricular pacing for bradycardia was associated with inappropriate shocks due to excessive autogain in 2 patients.
Conclusions Despite the major diagnostic and therapeutic advantages of tiered-therapy ICDs, a significant proportion of patients continue to experience hardware-related complications or receive inappropriate shocks.
Key Words: pacing defibrillators arrhythmia defibrillation
| Introduction |
|---|
|
|
|---|
The first available ICD devices were capable of high-energy defibrillation only irrespective of the nature of the ventricular arrhythmia. The current third-generation, multiprogrammable ICD pulse generators, in addition to high-energy defibrillation, incorporate facilities for antitachycardia and antibradycardia pacing, low-energy cardioversion, telemetry and intracardiac electrogram storage, and noninvasive programmed stimulation. A "second look" feature allows the device to abort therapy for arrhythmias that terminate spontaneously during charging.5 Concomitant with these improvements in ICD pulse generators, newer lead systems that can be implanted transvenously are likely to widen the indications for ICD therapy.6 7 8
The initial reports on these newer third-generation defibrillators and transvenous lead systems have focused on their advantages over earlier devices.9 10 11 However, as the technical aspects of device implantation become simplified, it is important that the limitations and complications associated with their use be evaluated. This study examines these problems and illustrates how they can be anticipated and in some cases avoided.
| Methods |
|---|
|
|
|---|
|
Devices
Third-generation ICD pulse generators from three
different
commercial manufacturers were used: Medtronic PCD model 7217B (61
patients), Ventritex Cadence V 100 (76 patients), and CPI Ventak PRx
(17 patients). All of these pulse generators are multiprogrammable, and
their features are summarized in Table 2
.
|
Several medical as well as nonmedical factors influenced the choice of generator used. These factors included the compatibility of the device with previously implanted defibrillation patch electrodes, the advantage of stored intracardiac electrograms in the management of patients with complex arrhythmias,12 the potential advantage of biphasic shocks13 in patients with previously failed implantation due to high defibrillation thresholds (DFTs) (Cadence only), the desirability of implanting a nonthoracotomy lead system (Medtronic PCD), and device availability. The transvenous defibrillation lead system used was the Medtronic Transvene System, which for the purposes of this study was used with the Medtronic PCD only. This lead system comprises two separate high-voltage defibrillation electrodes. The larger of the two leads incorporates both the right ventricular bipolar sensing electrodes and a defibrillation coil. Active screw-in fixation is used for stability. The second defibrillation lead lies free in the superior vena cava at its junction with the right atrium.
Implantation
Primary system implantation was performed in 141
patients and
pulse generator replacement in the remaining 13. One hundred patients
underwent a planned thoracotomy approach. In these patients, epicardial
rate-sensing electrodes were implanted during thoracotomy in 29
patients and transvenously placed endocardial electrodes were used in
the remainder. In 54 patients, implantation of a complete
nonthoracotomy lead system was attempted. The attempted nonthoracotomy
approach was performed using a graded protocol. Intraoperative
defibrillation threshold testing was carried out in all patients
receiving a primary implant. Satisfactory DFTs were defined as
successful conversion of three of four episodes of VF using 18 J or
less. In some patients in whom the morbidity of thoracotomy or system
revision was judged to be prohibitive, DFTs of up to 24 J were
accepted. The fact that all three devices have a maximum stored energy
of at least 34 J allowed a safety margin of at least 10 J in all
patients. If satisfactory DFTs could not be obtained using the
transvenous defibrillation leads in combination with a subcutaneous
patch, a single left ventricular extrapericardial patch electrode was
implanted via thoracotomy. If this was unsuccessful, a second right
ventricular patch was placed behind the sternum. In all except one
patient, the right ventricular endocardial lead was used for rate
sensing. Perioperative prophylactic antibiotic coverage consisting
of vancomycin and cefazolin was administered routinely.
Before discharge from the hospital, patients underwent a predischarge electrophysiology study. In patients whose preoperative VT was pace-terminable, the efficacy of antitachycardia pacing (ATP) was tested. Burst rather than ramp ventricular pacing was used in the vast majority of patients. If ATP was unsuccessful, this function was not activated at discharge and further assessment was made at postdischarge noninvasive programmed stimulation (NIPS).
Follow-up
After discharge from the hospital, all patients
were seen at 1-
to 2-month intervals or more frequently if ICD discharges occurred
inappropriately or with a high frequency. Within 1 to 2 months of
discharge from the hospital, ICD efficacy was tested by noninvasive
programmed stimulation unless an appropriate spontaneous discharge had
confirmed device efficacy. At each visit, the device was interrogated
and the number of tachycardia detections, their therapy, and efficacy
were recorded. Patients who underwent device explantation for any
reason, for example, cardiac transplantation, were censored from the
study at the time of explantation.
Classification of Events
Therapies were classified as appropriate or inappropriate, based
on all the available data including patient symptoms, telemetry data
from the ICD including stored intracardiac electrograms, 12-lead
surface ECG data as well as Holter or ECG event recordings, or rhythm
strips recorded by emergency medical personnel. Therapy was classified
as inappropriate only when the evidence for this was unequivocal. When
the number of clinical events on the therapy log since the last
interrogation exceeded the ability of the device to provide stored
intracardiac electrograms for corroboration, the available stored
events were taken as representative of all therapies delivered
since the last visit for statistical purposes.
Classification of Deaths
Sudden cardiac death was
defined as death within 1 hour of onset
of symptoms or resuscitated cardiac arrest from which the patient did
not regain consciousness before death. Other cardiac deaths were
defined as nonsudden. Death from other causes was defined as
noncardiac.
| Results |
|---|
|
|
|---|
Follow-up data are available on all patients. During a mean follow-up
of 15.3±9.7 months (range, 1.5 to 42), 99 patients (65.1%) received
therapy from their devices (Fig 1
). In 56 patients
(36.8%), device therapies were always appropriate, that is, they were
delivered for sustained ventricular arrhythmias only. Nearly one fifth
(18.4%) of patients received both appropriate therapy for VT or VF and
inappropriate therapies for other reasons. In 15 patients (9.8%),
device therapy was delivered for inappropriate reasons only.
|
A total of 56 patients (36.8%) had late postoperative ICD-related
problems. Twenty-one of these patients required a system revision a
mean of 8.6±11.3 months after device implantation (see Table
3
). The majority of system revisions occurred in the
early stages after implantation (13 of 21 occurring within the first 6
months). Seven of these 21 patients received inappropriate shocks; a
further 34 patients (22.3%) received inappropriate shocks for the
reasons listed in Table 3
. The primary cause of inappropriate
therapy
was the occurrence of supraventricular arrhythmias. Three patients had
more than one device-related problem. Fig 2
shows the
Kaplan-Meier survival curves for these patients. The 2-year survival
without the need for a system revision and without inappropriate shocks
was 86% and 57%, respectively.
|
|
Not all of the problems listed in Table 3
are of equal
seriousness. The loss of sensing function and substantial rises in DFT
are the most life-threatening, as both problems may result in failure
to restore sinus rhythm in the event of a cardiac arrest. Repetitive
spurious shocks in response to supraventricular arrhythmias are also
potentially life-threatening in patients with precarious ventricular
function. Device infection also represents a major problem, as
it usually involves the removal of all device components. The rest of
the problems outlined in Table 3
are less serious, although
they may
still be associated with significant morbidity.
Major Therapeutic Advantages
Antitachycardia Pacing
Of 85 patients with pace-terminable VT at preimplantation
electrophysiological evaluation, effective ATP was demonstrated
in 83 patients (98%) either at predischarge or subsequent
electrophysiological testing. Among these 83 patients, ATP therapy was
used appropriately in 58 (69%) during follow-up. A total of 3325
episodes of VT were treated with ATP as initial therapy. ATP therapy
was successful in terminating 3089 of these episodes, giving an overall
success rate of 92.9%. There was, however, a wide variation in
interpatient effectiveness, with some patients having several hundred
events successfully pace-terminated while in others only a small
proportion of events responded to ATP therapy (Fig 3
).
However, on a per-patient basis, the majority of individuals (62%) had
more than 90% of all episodes successfully terminated by ATP, whereas
only 12% had fewer than 1 in 10 episodes terminated.
|
Aborted Shock Function
The aborted shock function
was used on 288 occasions in 47
patients (30.9%) (Fig 4
). The mean number of aborted
shocks per patient was 6.1 (range, 1 to 45). The total number of
aborted shocks for inappropriate triggers such as lead noise or atrial
fibrillation was almost twice that for nonsustained VT or VF (184
versus 104). The mean number of aborted shocks for nonsustained VT was
3.4 per patient (range, 1 to 24), while that for inappropriate
triggers was 6.3 per patient (range, 1 to 45).
|
Problems
Patient-Related Problems
Slow VT. In all except 1 patient, detection of
ventricular arrhythmias was appropriate for the programmed device
parameters. However, one of the most troublesome of problems not yet
adequately solved by current ICD technology is the inability to
reliably differentiate between VT and supraventricular rhythms when
there is rate overlap between the two.
Because of the constraints involved in choosing optimal rate cutoffs in patients with overlapping sinus and VT rates, 12 patients (7.9%) had spontaneous slow VT below the programmed detection rate of the device. In 8 patients (5.2%) in whom such recurrent slow VTs could not be satisfactorily managed despite multiple antiarrhythmic drug trials and complex device reprogramming, endocardial ventricular mapping with a view to transcatheter radiofrequency ablation was undertaken. In 5 patients (3.2%), ablation of the target slow VT was possible. In these patients, despite the fact that faster nonclinical tachycardias were often still inducible by programmed ventricular stimulation after ablation, the technique provided useful adjunctive therapy to reduce the frequency of multiple ICD discharges for slow VT.
Supraventricular tachycardias. At the other end of
the
spectrum, 32 patients (21%) received inappropriate shocks for either
sinus tachycardia or atrial fibrillation with a rapid ventricular
response. ATP was used as initial therapy in 26 patients (17%). In
some cases, ATP resulted in apparent successful termination of the
supraventricular tachycardia (SVT) due either to spontaneous
slowing of the heart rate or possibly by concealed retrograde
conduction in the AV node during ventricular pacing. This
"pseudotermination" of supraventricular arrhythmias by ATP
could be seen clearly in patients with stored intracardiac electrograms
(Fig 5
). In those with event counters only,
inappropriate therapies with pseudotermination were classified as
successful therapy by the device log. In this patient population,
therefore, the ICD log will tend to overestimate device efficacy.
|
In a
majority of patients, antitachycardia pacing for supraventricular
rhythms was simply ineffective. However, in 2 patients, burst
ventricular pacing during SVT induced VT, as has previously been
recognized.15 These episodes of VT but not the background
atrial arrhythmia were terminated by shock therapy (example in Fig
6
). The cycle of ATP, VT, and shock was then repeated
until the device could be either inactivated or reprogrammed. In
patients in whom ATP was not activated or failed to slow SVT, shocks of
increasing energy were delivered. However, in only 1 patient did these
shocks terminate the supraventricular arrhythmia. In 7 patients, a
series of 15 or more inappropriate shocks was delivered during a single
clinical episode. In 1 patient, this complication was sufficiently
disturbing to initiate requests for device inactivation and require
psychiatric counseling. In an additional patient, fortuitously wearing
a Holter monitor recorder at the time, there was evidence of
progressive ST-segment elevation after repeated shocks for sinus
tachycardia with subsequent ECG and enzymatic evidence of nontransmural
myocardial infarction.
|
Drug therapy to prevent spurious shocks. At the time of initial discharge from the hospital, all patients with a history of atrial fibrillation or at high risk for spurious discharges for SVTs (for example, young patients with high levels of activity) were discharged on sinus slowing or AV nodalblocking drugs (digoxin, 50%; ß-adrenergic blocker, 36%; calcium channel blocker, 7%; class Ia antiarrhythmic agents, 20%; and amiodarone, 6%).
Those patients receiving inappropriate therapy for SVTs were taking the following medications at the time of first inappropriate therapy: digoxin, 64%; ß-adrenergic blocker, 37%; and calcium channel blocker, 18%. Initially, 67% of patients were taking one, 18% two, and 3.0% three of these agents. After adjustments in therapy, all were receiving rate-lowering medications, with 52% receiving one, 42% two, and 6% three of the above agents, with 10% of patients also receiving amiodarone therapy.
Rise in defibrillation thresholds. In 4 patients, ICD
system
revision was required because of high DFTs. In 2 of the patients, both
with nonthoracotomy lead systems and initial DFTs of
18 J, multiple
episodes of VT required initiation of therapy with amiodarone.
When the DFTs were retested on amiodarone, 34-J shocks failed
to restore sinus rhythm, and the systems were therefore revised with
implantation of supplementary extrapericardial defibrillation patch
electrodes. In 1 patient, a rise in DFT led to a failure to convert a
clinical tachyarrhythmia, and this patient was successfully
resuscitated by emergency medical technicians.
Lead-Related Problems
Migration of Transvenous
Defibrillation Leads
Of 54 patients with transvenous defibrillation
electrodes,
lead displacement or retraction occurred in 5 patients (superior vena
cava lead, 2 patients; right ventricular lead, 3 patients). In 1
patient, perforation of the right ventricle occurred without
pericardial tamponade, which required a thoracotomy for system
revision. This patient developed severe pericarditis, which persisted
for several months after hospital discharge.
Sensing Lead
Problems
Sensing lead problems resulted from either insulation breaks,
Y-adaptor malfunction, or a marked diminution of sensed
R-wave amplitude (Table 3
). The former two problems led either
to
inappropriate or aborted shocks that were documented by intracardiac
electrogram recordings. In 1 patient, mechanical noise induced by
intermittent contact of a redundant endocardial rate-sensing lead
against a newly implanted sensing electrode (ringing) resulted in
multiple inappropriate shocks. Only one of these sensing problems
occurred with the nonthoracotomy lead system.
Pulse Generator Problems
In 4 patients, the implanted pulse
generator was prematurely
explanted and replaced. In 2 patients this was done because of a faulty
circuit capacitor and in another patient because of premature battery
depletion at 6 months (no patient in this series has yet required an
elective pulse generator change). In all cases, pulse generator
malfunction was detected at routine follow-up interrogation or testing.
No adverse clinical outcome resulted. In the third patient, the device,
for unknown reasons, failed to detect an ECG-documented episode of VT,
which was well within the programmed detection parameters. The device
was therefore replaced, but no abnormality was found on factory
testing.
Autogain-Related Sensing Problems
Two of the
three devices in this study incorporate an automatic
gain control feature. The autogain feature automatically maximizes gain
settings during bradycardia pacing to ensure that the underlying rhythm
interpreted as asystole is not in fact VF with low-amplitude
intracardiac signals and ignores the immediate postpacing signal
(postpacing refractoriness). When sinus rhythm resumes, however, the
transiently augmented amplitude of the intracardiac electrogram may
result in double or triple counting, with subsequent inappropriate
shock delivery during sinus rhythm. This was observed in 2 patients who
received inappropriate shocks after ventricular pacing. In 1 patient,
it was resolved by programming an increase in the number of tachycardia
intervals for VT or VF detection, thereby allowing sufficient time for
the autogain function to gradually reduce gain settings before
detection and therapy. In the second patient, this approach proved
ineffective and required placement of a separate permanent pacemaker
for bradycardia pacing.
Other Management Issues
Bradycardia-Associated VT
In 5 patients (3.2%), all of whom had coronary artery disease and
easily inducible VT at preimplant evaluation, 25 episodes of VT after
ventricular paced beats terminating prolonged pauses were
documented by stored intracardiac electrograms (Fig 7
). These
paced beats were all appropriately timed and
did not result from failure to sense sinus beats, as has previously
been described.16 This mechanism of VT induction,
recognized by careful analysis of stored intracardiac electrograms,
can easily be overlooked because of the very small size of the
ventricular pacing artifact in most cases (Fig 7
). These
episodes of VT
are probably promoted by the long cycle length of the pauses, since
they may be effectively prevented by increasing the backup bradycardia
pacing rate. By contrast, the use of antiarrhythmic drug therapy, which
may further slow the sinus rate, may be counterproductive.
|
Deaths and System Removal
Twenty-three patients
(15.1%) died during follow-up, 19 (12.5%)
from cardiac causes. Five other patients had their devices removed, 3
at the time of cardiac transplantation and 1 at the time of corrective
arrhythmia surgery. The final patient had the device removed because of
infection related to a sensing lead revision procedure. This patient
subsequently underwent successful ventricular arrhythmia surgery.
Cardiac deaths occurred a mean of 11.3±9.7 months after ICD implantation. The mean left ventricular ejection fraction of these patients was 27.1±10.7%. A total of 13 deaths were due to progressive heart failure. Six patients died suddenly 6.1±5.1 months after ICD implantation, resulting in a 96.4% 1-year actuarial sudden deathfree survival. All 6 patients had a history of clinically significant heart failure, with mean left ventricular ejection fraction of 31.8±8.9%. Two of these 6 patients suffered acute hemodynamic collapse, were resuscitated by emergency medical technicians, and subsequently died without regaining consciousness 48 and 72 hours later, respectively. In 1 patient, the first documented rhythm was ventricular pacing, with no evidence of shocks from his device. An additional patient awaiting cardiac transplantation was admitted to the hospital in VT with shock after his device had fired several times; he died almost immediately afterward in electromechanical dissociation after an external shock. One patient with ischemic heart disease died within 20 minutes of the onset of acute chest pain and could not be resuscitated. Two patients died suddenly at home with unwitnessed collapse, with device interrogation in 1 patient revealing several successful shock therapies followed by bradycardia pacing. In none of these patients was there any evidence that the device had failed to detect arrhythmias or deliver appropriate therapy.
| Discussion |
|---|
|
|
|---|
The majority of problems observed in this study occurred in patients who also had appropriate detection and termination of VT or VF. This is not surprising, since particularly in patients with relatively slow VT, tachyarrhythmia detection rates must be programmed in a zone where the risks of inappropriate therapy for sinus tachycardia or AF with a rapid ventricular response pose the greatest risk. Furthermore, in a large majority of patients, inappropriate therapy as observed in this study reflects the limitations of current ICD technology rather than device malfunction or failure. The problems associated with effective ICD therapy in these patients may well be an acceptable price to pay for the protection against sudden death, which the device offers. However, a smaller group of patients experience ICD-related problems without appropriate therapy. This factor will require consideration in the evaluation of clinical trials on the prophylactic use of ICD therapy in high-risk patients.
As with previous generations of devices, there are two broad categories of problems associated with current ICD therapy: those related to hardware complications and those related primarily to the limitations of tachyarrhythmia detection algorithms.18 19 Some hardware-related problems with rate-sensing and defibrillation leads have already been solved (for example, problems with noise from faulty Y-adaptors) and are no longer relevant. Similarly, the high incidence of lead dislodgment with the early use of nonthoracotomy lead systems seen in this and other series has been largely overcome by the use of better anchoring sleeves and the heightened awareness of the risks of displacement.6 20 These problems are likely, therefore, to arise less frequently in the future. However, at present, there are no data on the long-term durability of current transvenous lead systems. The early complications observed in this and other studies may be compounded by long-term integrity problems, as seen with previous transvenous lead systems.21 This concern is of particular importance in younger patients with well-preserved left ventricular function who may require defibrillator therapy for decades.
Another problem shared with earlier generations of devices is elevation of DFTs by antiarrhythmic therapy, particularly amiodarone. With the routine use of biphasic waveforms, this problem may be partially resolved.
The issue of inappropriate therapy for supraventricular arrhythmias remains one of the major problems associated with ICD therapy. The 26.6% rate of inappropriate shocks is similar to that reported for earlier generations of devices.2 3 Two of the three devices in this series incorporate an optional rate stability criterion as part of the programmable arrhythmia detection algorithm. The rate stability algorithm detects beat-to-beat variations in cycle length and was initially developed to distinguish regular supraventricular arrhythmias from AF. However, great caution must be exercised in using this algorithm to distinguish VT from AF with a rapid ventricular response, since VT may be associated with variability in cycle length, making safe and reliable differentiation between the two arrhythmias by rate stability alone difficult in some patients. For this reason and to ensure maximum sensitivity for VT detection, we did not use the rate stability function, and this accounts in part for the continuing high incidence of inappropriate therapies observed in this study. Bardy et al6 reported that only 3 of 84 patients followed for 11±7 months received inappropriate therapy for AF and that no patient had an undetected ventricular tachyarrhythmia when the rate stability criterion was used. The availability of ICDs with dual-chamber sensing capability will also undoubtedly improve the specificity of algorithms for VT and SVT discrimination.
Mode of Death
Of the six sudden deaths in this series, at
least three were
related to primary pump failure or ischemia. These sudden deaths in
patients with poorly compensated left ventricular function suggest that
sudden deterioration in hemodynamic state, mediated by bradyarrhythmia
or tachyarrhythmia, although appropriately treated by the ICD, may
still be fatal. It has been suggested that as many as 50% of sudden
deaths in patients with advanced congestive heart failure may be due to
bradyarrhythmias.22 23 The mechanisms of such
bradyarrhythmias may be inherently different from the primary
bradyarrhythmias associated with sick sinus syndrome or intermittent
complete heart block, which can be effectively countered by cardiac
pacing. Our observations suggest that even the most versatile ICD
devices with bradycardia pacing do not uniformly prevent sudden cardiac
death in patients with severely compromised left ventricular
function. Since current ICDs store only events that trigger
tachyarrhythmia but not bradyarrhythmia therapy, a more precise
understanding of the modes of death in patients with ICDs will be
gained only when these devices incorporate additional memory for
storage of bradyarrhythmic events as well.
Reduction in Shocks With ATP
The success of ATP in reducing
the need for ICD shocks in selected
patients with frequently recurring VT in this study is striking.
Without this function, implantation of an ICD in many of these patients
would have been impractical. The high success rate (92.9%) of
conversion of VT by ATP in this highly selected group of patients is
virtually identical to the 88% to 92.4% reported from other
centers.6 10 24 25 It is
likely that these early reports
of successful ATP also represent highly selected patient
subgroups. Furthermore, this high percentage of success is due in part
to a relatively small number of patients in whom numerous episodes of
VT were reliably terminated by ATP. The observation that 62% of all
patients had more than 90% of their VT episodes terminated by ATP may
better reflect the overall effectiveness of antitachycardia pacing.
Study Limitations
The patient population in this study
represents a highly
selected subgroup of patients in whom the results may not be applicable
to the broad spectrum of ICD recipients. When third-generation devices
first became available in limited numbers, their use was largely
reserved for patients with severe left ventricular dysfunction and
frequently recurring drug-resistant VT, which could be terminated by
ATP or low-energy cardioversion. In addition, these devices were
implanted as pulse generator replacements in patients who had already
demonstrated problems with earlier-generation devices. Thus, the
percentage of patients receiving inappropriate therapy for these
conditions could be greater in this study than might be expected in an
unselected group of patients receiving ICD therapy. Furthermore, the
inappropriate shock rate might have been lower if the rate stability
function, available in two of the three ICDs used, had been used
although at the potential cost of nondetection of VT in this high-risk
population.
Despite the increasing availability of telemetry and stored electrograms, it is not always possible to determine whether therapy was appropriate or inappropriate, especially when large numbers of events have occurred. In this study, therapy was classified as inappropriate only when the evidence was unequivocal. Therefore, in this patient group, as with previous studies, we may have overestimated the number of ventricular arrhythmia events effectively terminated by the device.
Conclusions
Third-generation ICDs offer significant
advantages over previously
available devices. In particular, ATP can be extremely effective in
reducing the number of shocks delivered for frequently occurring VT.
However, third-generation ICDs, including those with nonthoracotomy
lead systems, are still associated with mechanical problems and
inappropriate therapies. Given the extremely rapid evolution of ICD
technology, it is likely that most of the technical difficulties
associated with current ICD devices will be solved, leaving random
component failure and operator error as the primary sources of future
problems. Nevertheless, the results of this study underscore the
complexity of these advanced devices and emphasize the need for
additional clinical experience as well as prospective evaluation of
different detection algorithms and treatment options to minimize the
occurrence of inappropriate therapies while maintaining maximum
sensitivity for the detection of VT and VF.
| Footnotes |
|---|
Received September 22, 1994; accepted November 20, 1994.
| References |
|---|
|
|
|---|
2. Kelly PA, Cannom DS, Garan H, Mirabal GS, Harthorne JW, Hurvitz RJ, Vlahakes GJ, Jacobs ML, Ilvento JP, Buckley MJ, Ruskin JN. The automatic implantable cardioverter-defibrillator: efficacy, complications and survival in patients with malignant ventricular arrhythmias. J Am Coll Cardiol. 1988;11:1278-1286. [Abstract]
3. Winkle RA, Mead RH, Ruder MA, Gaudiani VA, Smith NA, Buch WA, Schmidt P, Shipman T. Long-term outcome with the automatic implantable cardioverter defibrillator. J Am Coll Cardiol. 1989;13:1353-1361. [Abstract]
4. Kim SG. Management of survivors of cardiac arrest: is electrophysiologic testing obsolete in the era of implantable defibrillators? J Am Coll Cardiol. 1990;16:756-761. [Medline] [Order article via Infotrieve]
5. Hurwitz JL, Hook BG, Flores BT, Marchlinski FE. Importance of abortive shock capability with electrogram storage in cardioverter-defibrillator devices. J Am Coll Cardiol. 1993;21:895-900. [Abstract]
6.
Bardy GH, Hofer B, Johnson G, Kudenchuk PJ, Poole JE, Dolack
L, Gleva M, Mitchell R, Kelso D. Implantable transvenous
cardioverter-defibrillators. Circulation. 1993;87:1152-1168.
7.
Yee R, Klein GJ, Leitch JW, Guiraudon GM, Guiraudon CM, Jones
DL, Norris C. A permanent transvenous lead system for an implantable
pacemaker cardioverter-defibrillator: nonthoracotomy approach to
implantation. Circulation. 1992;85:196-204.
8.
Jordaens L, Trouerbach J-W, Vertongen P, Herregods L, Polaert
J, Van Nooten G. Experience of cardioverter-defibrillators inserted
without thoracotomy: evaluation of transvenously inserted intracardiac
leads alone or with a subcutaneous axillary patch. Br Heart
J. 1993;69:14-19.
9. Saksena S, Mehta D, Krol RB, Tullo NG, Saxena A, Kaushik R, Neglia J. Experience with a third-generation implantable cardioverter-defibrillator. Am J Cardiol. 1991;67:1375-1384. [Medline] [Order article via Infotrieve]
10.
Fromer M, Brachmann J, Block M, Siebels J, Hoffmann E,
Almendral J, Ohm OJ, den Dulk K, Coumel P, Camm AJ, Toubol P. Efficacy
of automatic multimodal device therapy for ventricular tachyarrhythmias
as delivered by a new implantable pacing cardioverter-defibrillator:
results of a European multicenter study of 102 implants.
Circulation. 1992;86:363-374.
11.
Bardy GH, Troutman C, Poole JE, Kudenchuk PJ, Dolack GL,
Johnson G, Hofer B. Clinical experience with a tiered therapy,
multiprogrammable antiarrhythmia device.
Circulation. 1992;85:1689-1698.
12. Marchlinski FE, Gottlieb CD, Sarter B, Finkle J, Hook B, Callans D, Schwartzman D. ICD data storage: value in arrhythmia management. PACE Pacing Clin Electrophysiol. 1993;16(pt II):527-534.
13. Saksena S, Huanlin A, Mehra R, DeGroot P, Krol RS, Burkhardt E, Mehta D, John T. Prospective comparison of biphasic and monophasic shocks for implantable cardioverter-defibrillators using endocardial leads. Am J Cardiol. 1992;70:304-310. [Medline] [Order article via Infotrieve]
14. Kim SG, Fisher JD, Furman S, Gross J, Zilo P, Roth JA, Ferrick KJ, Brodman R. Exacerbation of ventricular arrhythmias during the postoperative period after implantation of an automatic defibrillator. J Am Coll Cardiol. 1991;18:1200-1206. [Abstract]
15. Johnson NJ, Marchlinski FE. Arrhythmias induced by device antitachycardia therapy due to diagnostic nonspecificity. J Am Coll Cardiol. 1991;18:1418-1425. [Abstract]
16. Callans DJ, Hook BG, Marchlinski FE. Paced beats following single nonsensed complexes in a `codependent' cardioverter defibrillator and bradycardia pacing system: potential for ventricular tachycardia induction. PACE Pacing Clin Electrophysiol. 1991;14:1281-1287. [Medline] [Order article via Infotrieve]
17.
Bardy GH, Johnson G, Poole JE, Dolack L, Kudenchuk PJ, Kelso
D, Mitchell R, Mehra R, Hofer B. A simplified, single-lead unipolar
transvenous cardioversion-defibrillation system.
Circulation. 1993;88:543-547.
18. Grimm W, Flores BF, Marchlinski FE. Electrocardiographically documented unnecessary, spontaneous shocks in 241 patients with implantable cardioverter defibrillators. PACE Pacing Clin Electrophysiol. 1992;15:1667-1673. [Medline] [Order article via Infotrieve]
19. Grimm W, Flores BF, Marchlinski FE. Complications of implantable cardioverter defibrillator therapy: follow-up of 241 patients. PACE Pacing Clin Electrophysiol. 1993;16:218-222. [Medline] [Order article via Infotrieve]
20. Paul VE, Anderson M, Jones S, Skehan DJ, Camm AJ, Ward DE. Cardiologist implanted cardioverter defibrillators: early experience of three systems. Eur J Cardiac Pacing Electrophysiol. 1991;1:21-26.
21. Tullo NG, Saksena S, Krol RB, Mauro AM, Kunecz D. Management of complications associated with a first generation endocardial defibrillation lead system for implantable cardioverter defibrillators. Am J Cardiol. 1990;66:411-415. [Medline] [Order article via Infotrieve]
22.
Luu M, Stevenson WG, Stevenson LW, Baron K, Walden J. Diverse
mechanisms of unexpected cardiac arrest in advanced heart failure.
Circulation. 1989;80:1675-1680.
23. Packer M. Lack of relation between ventricular arrhythmias and sudden death in patients with chronic heart failure. Circulation. 1992;85(suppl I):I-50-I-56.
24. Porterfield JG, Porterfield LM, Smith BA, Bray L. Experience with three different third generation cardioverter-defibrillators in patients with coronary artery disease or cardiomyopathy. Am J Cardiol. 1993;72:301-304. [Medline] [Order article via Infotrieve]
25. Leitch JW, Gillis AM, Wyse G, Yee R, Klein GJ, Guiraudon G, Sheldon RS, Duff HJ, Kieser TM, Mitchell LB. Reduction in defibrillator shocks with an implantable device combining antitachycardia pacing and shock therapy. J Am Coll Cardiol. 1991;18:145-151.[Abstract]
This article has been cited by other articles:
![]() |
N. Sadoul, R. Mletzko, F. Anselme, R. Bowes, W. Schols, C. Kouakam, G. Casteigneau, R. Luise, N. Iscolo, E. Aliot, et al. Incidence and Clinical Relevance of Slow Ventricular Tachycardia in Implantable Cardioverter-Defibrillator Recipients: An International Multicenter Prospective Study Circulation, August 16, 2005; 112(7): 946 - 953. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. Senges-Becker, M. Klostermann, R. Becker, A. Bauer, K. E. Siegler, H. A. Katus, and W. Schoels What is the "Optimal" follow-up schedule for ICD patients? Europace, January 1, 2005; 7(4): 319 - 326. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. A.M.J. Theuns, A. P. J. Klootwijk, D. M. Goedhart, and L. J.L.M. Jordaens Prevention of inappropriate therapy in implantable cardioverter-defibrillators: Results of a prospective, randomized study of tachyarrhythmia detection algorithms J. Am. Coll. Cardiol., December 21, 2004; 44(12): 2362 - 2367. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Becker, J. Ruf-Richter, J. C. Senges-Becker, A. Bauer, S. Weretka, F. Voss, H. A. Katus, and W. Schoels Patient alert in implantable cardioverter defibrillators: toy or tool? J. Am. Coll. Cardiol., July 7, 2004; 44(1): 95 - 98. [Abstract] [Full Text] [PDF] |
||||
![]() |
C A Rinaldi, R D Simon, A Baszko, J Bostock, D Elliot, C A Bucknall, and J S S Gill A 17 year experience of inappropriate shock therapy in patients with implantable cardioverter-defibrillators: are we getting any better? Heart, March 1, 2004; 90(3): 330 - 331. [Full Text] [PDF] |
||||
![]() |
C. Kouakam, S. Kacet, J.-R. Hazard, A. Ferraci, H. Mansour, P. Defaye, J.-M. Davy, M. Lambiez, and on behalf of the Ventak AV Investigators Performance of a dual-chamber implantable defibrillator algorithm for discrimination of ventricular from supraventricular tachycardia Europace, January 1, 2004; 6(1): 32 - 42. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Quesada, J.ús Almendral, F. Arribas, R. Ricci, C. Wolpert, P. Adragao, E. Cobo, X. Navarro, and on behalf DATAS investigators The DATAS rationale and design: a controlled, randomized trial to assess the clinical benefit of dual chamber (DDED) defibrillator Europace, January 1, 2004; 6(2): 142 - 150. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. A. L. E. Bracke, A. Meijer, and L. M. van Gelder Malfunction of endocardial defibrillator leads and lead extraction: where do they meet? Europace, January 1, 2002; 4(1): 19 - 24. [Full Text] [PDF] |
||||
![]() |
G. D. Sanders, M. A. Hlatky, N. R. Every, K. M. McDonald, P. A. Heidenreich, L. S. Parsons, and D. K. Owens Potential Cost-Effectiveness of Prophylactic Use of the Implantable Cardioverter Defibrillator or Amiodarone after Myocardial Infarction Ann Intern Med, November 20, 2001; 135(10): 870 - 883. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Gronefeld and S. H Hohnloser What do implantable cardioverter/defibrillators teach us about the mechanisms of sudden cardiac death? Cardiovasc Res, May 1, 2001; 50(2): 232 - 241. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. H. Gollob and J. J. Seger Current Status of the Implantable Cardioverter-Defibrillator Chest, April 1, 2001; 119(4): 1210 - 1221. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Nisam A Prophylactic ICD? Who are the patients? What is the device? Europace, January 1, 2001; 3(4): 269 - 274. [PDF] |
||||
![]() |
T. Korte, W. Jung, G. Ostermann, C. Wolpert, S. Spehl, B. Esmailzadeh, and B. Luderitz Hospital readmission after transvenous cardioverter/defibrillator implantation. A single centre study Eur. Heart J., July 2, 2000; 21(14): 1186 - 1191. [Abstract] [PDF] |
||||
![]() |
A. Oto, A. Yildirir, G. Kabakci, and E. Atalar Inappropriate Shocks Diagnosed by Stored Electrograms of Implantable Cardioverter Defibrillators: Two Case Reports Angiology, May 1, 2000; 51(5): 425 - 430. [Abstract] [PDF] |
||||
![]() |
M. Greene, D. Newman, M. Geist, M. Paquette, D. Heng, and P. Dorian Is electrical storm in ICD patients the sign of a dying heart?: Outcome of patients with clusters of ventricular tachyarrhythmias Europace, January 1, 2000; 2(3): 263 - 269. [Abstract] [PDF] |
||||
![]() |
L. A. Pires, M. H. Lehmann, R. T. Steinman, J. J. Baga, C. D. Schuger, and Participating Investigators Sudden death in implantable cardioverter-defibrillator recipients: clinical context, arrhythmic events and device responses J. Am. Coll. Cardiol., January 1, 1999; 33(1): 24 - 32. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Gradaus, B. Lamp, D. Hammel, M. Loick, H.H. Scheld, G. Breithardt, and M. Block Multiple hardware complications in a patient with an ICD Europace, January 1, 1999; 1(4): 270 - 274. [Abstract] [PDF] |
||||
![]() |
D Babuty, L Fauchier, and P Cosnay Inappropriate shocks delivered by implantable cardiac defibrillators during oversensing of activity of diaphagmatic muscle Heart, January 1, 1999; 81(1): 94 - 96. [Abstract] [Full Text] |
||||
![]() |
R. Ruppel, C. A. Schluter, S. Boczor, T. Meinertz, M. Schluter, K.-H. Kuck, and R. Cappato Ventricular tachycardia during follow-up in patients resuscitated from ventricular fibrillation: experience from stored electrograms of implantable cardioverter-defibrillators J. Am. Coll. Cardiol., November 15, 1998; 32(6): 1724 - 1730. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Rosenqvist, T. Beyer, M. Block, K. den Dulk, J. Minten, and F. Lindemans Adverse Events With Transvenous Implantable Cardioverter-Defibrillators : A Prospective Multicenter Study Circulation, August 18, 1998; 98(7): 663 - 670. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. E. Marchlinski Predicting Arrhythmic Death : A Plea for Standardized Reporting Techniques and Data Based on Continuous Electrocardiographic Monitoring Circulation, September 16, 1997; 96(6): 1713 - 1716. [Full Text] |
||||
![]() |
P. Spirito, C. E. Seidman, W. J. McKenna, and B. J. Maron The Management of Hypertrophic Cardiomyopathy N. Engl. J. Med., March 13, 1997; 336(11): 775 - 785. [Full Text] [PDF] |
||||
![]() |
M. Nair, N. Saoudi, D. Kroiss, and B. Letac Automatic Arrhythmia Identification Using Analysis of the Atrioventricular Association: Application to a New Generation of Implantable Defibrillators Circulation, February 18, 1997; 95(4): 967 - 973. [Abstract] [Full Text] |
||||
![]() |
G. J. Fahy, E. B. Sgarbossa, P. J. Tchou, and S. L. Pinski Hospital Readmission in Patients Treated With Tiered-Therapy Implantable Defibrillators Circulation, September 15, 1996; 94(6): 1350 - 1356. [Abstract] [Full Text] |
||||
![]() |
T. Ferguson Jr, C. Ferguson, K Crites, and P Crimmins-Reda THE ADDITIONAL HOSPITAL COSTS GENERATED IN THE MANAGEMENT OF COMPLICATIONS OF PACEMAKER AND DEFIBRILLATOR IMPLANTATIONS J. Thorac. Cardiovasc. Surg., April 1, 1996; 111(4): 742 - 752. [Abstract] [Full Text] |
||||
![]() |
IMPLANTABLE DEFIBRILLATORS: PROMISE AND PROBLEMS Journal Watch (General), April 21, 1995; 1995(421): 2 - 2. [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1995 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |