(Circulation. 1995;91:1996-2001.)
© 1995 American Heart Association, Inc.
Articles |
From the Division of Cardiology, Department of Medicine, University of Washington, Seattle.
Correspondence to Gust H. Bardy, MD, Division of Cardiology RG-22, 1959 NE Pacific, University of Washington, Seattle, WA 98195.
| Abstract |
|---|
|
|
|---|
Methods and Results The records of 111 consecutive patients who had undergone transvenous cardioverter/defibrillator (ICD) implantation for malignant ventricular arrhythmias were reviewed retrospectively. For each patient, all device tachyarrhythmia detections were examined and classified as VF, MVT, rapid polymorphic VT, or other. The number of events, time to first arrhythmia detection, and cycle length of MVTs were recorded. There were 55 patients with a history of only VF and 56 with a history that included an episode of MVT. Over 14 months of follow-up, with all patients initially off of antiarrhythmic medications, MVT was detected by only 18% of patients with a history of only VF compared with 54% of those with a history that included MVT (P=.002). Among patients who did detect MVT, those with a history of only VF had fewer episodes (7±7 versus 20±31, P=.001) and a shorter mean MVT cycle length (279 versus 314 ms, P=.03) than those with a clinical history of MVT. Abrupt onset of VF not preceded by MVT was detected in 11% of patients with VF only. In addition to a history of MVT, male sex, age <60 years, and MVT inducible on EPS were all significantly associated with an increased likelihood of MVT detection. On multivariate analysis, the inducibility of MVT was the primary independent predictor of MVT detection but was of minimal incremental predictive value in the subgroup of patients with a history of only VF. When EPS results were not considered, arrhythmia history was the primary independent predictor of MVT detection.
Conclusions Patients with a history of only VF infrequently have MVT detected by their defibrillators. When these patients do detect MVT, it is faster than that detected in patients with a clinical history of MVT before ICD surgery. A significant percentage of VF survivors detected the abrupt onset of VF not preceded by MVT, suggesting that the deterioration of rapid MVT to VF is not the only clinically important mechanism of VF induction. These findings may have important implications for the understanding of the mechanism of VF induction and for use of an implantable defibrillator.
Key Words: defibrillation fibrillation tachycardia electrophysiology
| Introduction |
|---|
|
|
|---|
Despite these differences, there is a common impression among physicians that VF survivors simply represent a more malignant end of the spectrum of the manifestations of MVT.8 9 10 11 12 13 14 It is postulated that in some patients, MVT-induced hypotension, ischemia, left ventricular dysfunction, or electrical instability eventually causes the rhythm to deteriorate to VF.8 9 10 In favor of this argument are reports of MVT deteriorating to VF during ambulatory ECG monitoring.11 12 13 14 The incidence of spontaneous MVT and VF in an unselected group of patients resuscitated from VF not taking antiarrhythmic medications is unknown. Determining the incidence of these rhythms in such a patient group might provide insights into the mechanisms of cardiac arrest in VF survivors and have important implications for implantable cardioverter/defibrillator (ICD) use. If VF survivors are unlikely to have MVT, an ICD capable of specific VT detection and therapy may be unnecessary. This patient group may need only defibrillation capabilities.
The purpose of this study was to use the memory and therapeutic capacity of the ICD to document the incidence of MVT and VF in a group of patients with a history of only VF and compare them with patients with a history that included at least one episode of sustained MVT.
| Methods |
|---|
|
|
|---|
Baseline data gathered for each patient included age, left ventricular ejection fraction, presence of coronary artery disease, history of myocardial infarction, past history of CABG, and results of EPS. EPSs including up to triple extrastimuli at two drive train cycle lengths at two right ventricular pacing sites were performed as a separate procedure before ICD implantation in 51 (91%) of the MVT patients and 28 (51%) of the VF patients. An EPS via the defibrillator consisting of up to triple extrastimuli at two drive train cycle lengths from the right ventricular apex was performed in 98 (88%) of the patients (48 VF and 50 MVT patients) before hospital discharge, usually 3 days after ICD implantation. Six patients did not have EPS performed either before implantation or before hospital discharge and were excluded from the analysis of the predictive value of EPS testing. MVT induced on either of these studies was considered inducibility in the analysis.
Implantable Cardioverter Defibrillator
A tiered-therapy ICD
(model 7216A or 7217B PCD, Medtronic) was
implanted in all patients via a transvenous lead system incorporating a
right ventricular lead, a subcutaneous patch, and a superior vena cava
or coronary sinus lead. A report of implantation methods and survival
data in many of these patients was published elsewhere.15
This device detects rhythms for treatment based on the cycle length of
ventricular activity as detected by the right ventricular lead. The
details of our use of the detection, treatment, and memory capacity of
this device have been described.16 The VF zone is the
programmed heart rate range within which a tachycardia is detected and
treated as VF. VF detection and therapy was programmed as on in all
patients. The VT zone is the programmed heart rate range within which a
tachycardia is detected and treated as VT. The VT zone was activated
only in patients with a clinical history of MVT, reproducible MVT
during EPS, or frequent episodes of rapid MVT detected in the VF zone.
Antiarrhythmic drugs were discontinued at the time of device
implantation and were used in an attempt to reduce the frequency of
device discharges in only five patients (all in the MVT group) a mean
of 6 months after ICD implantation.
Follow-up
Defibrillators were interrogated after all
clinically apparent
device treatment events and on routine follow-up every 3 months. The
date of an event was defined as the date of symptomatic device therapy,
or if asymptomatic, the date of device interrogation. The ICD memory
contains the cycle length of the 20 cycles before the most recent
arrhythmia detection and the 10 cycle lengths after the last therapy
delivered. The device does not store electrograms. Arrhythmias were
identified on the basis of these cycle length data in conjunction with
clinical information. A rapid arrhythmia detected in the VF or VT zone
was classified as MVT if the cycle length of the tachycardia did not
vary by >30 ms once initiated. More irregular tachycardias, with a
mean cycle length <200 ms or those in which 75% of recorded cycle
lengths were <260 ms, were defined as VF. Slower irregular
tachycardias were defined as polymorphic VT unless atrial fibrillation
was strongly suspected on clinical grounds. Rapid atrial fibrillation
could be recognized by variations in the cycle length of a tachycardia,
with a mean cycle length usually >300 ms in a patient known to have
chronic or paroxysmal atrial fibrillation. Sinus tachycardia was
suspected in the proper clinical setting when the cycle length of the
tachycardia detected equaled the lower limit of the VT zone, suggesting
acceleration of the sinus rate into the VT zone. The cycle length of
the 10 beats after the last therapy is also recorded by the device and
was used to differentiate tachycardia type. For instance, variable
cycle lengths persisting after therapy just below the VT zone detection
interval would help confirm atrial fibrillation if suspected.
Similarly, a regular posttherapy cycle length just below the VT zone
cycle length would help confirm sinus tachycardia if suggested by the
clinical history and the pretherapy cycle length data. Supraventricular
tachycardias, including sinus tachycardia and atrial fibrillation, were
eliminated from this analysis.
If on interrogation more than one MVT or VF episode had been detected, then all the episodes were judged to have been of the mechanism determined from the cycle length data available for the most recent tachycardia detected. If both MVT and VF had been detected, then cycle length data would be available only for the last detected tachycardia. Frequently, this occurred when the last episode was detected in the VF zone and resulted in a high-energy shock prompting the interrogation. In most of these instances, examination of the available cycle length data revealed that it had actually been rapid MVT that had precipitated the "VF" detection. In such an instance, the MVT episodes for which no cycle length data were available were presumed to be true MVT detections. If the last episode had been MVT and there were VF episodes for which no cycle length data were available, differentiation between acceleration of MVT by antitachycardia pacing, rapid MVT in the VF zone, and primary VF could not be made, and these episodes were classified as "unknown." This occurred in eight MVT patients and prevents the accurate estimation of the frequency of primary VF in the MVT patients; it may lead to a slight undercounting of the number of MVT episodes in some MVT patients. The infrequency of MVT in VF survivors was such that this occurred in only two patients. In both patients, primary VF episodes were documented on separate occasions. Similarly, in no VF survivors did VF episodes prevent examination of cycle length data for evaluation of possible MVT in a patient who had not had MVT documented on another occasion. The cycle length of an MVT episode was determined as the mean of the cycle lengths recorded before therapy once MVT was established. Episodes of VT for which cycle length data were not available were not used in the calculation of the mean MVT cycle length.
Statistics
The
2 test was used for analysis
of
categorical variables, and Student's t test was used for
analysis of continuous variables. Life table analysis was
performed according to the Kaplan-Meier method. Stepwise Cox regression
was used to identify independent predictors of the presence or absence
of MVT detection on follow-up. Analysis of the independent predictors
of the number of MVT episodes detected was performed by stepwise linear
regression.
| Results |
|---|
|
|
|---|
|
Tachycardia Detection
MVT was detected significantly more
often in MVT than in VF
patients (Table 2
). MVT was detected in 18% (10 of 55)
of VF patients and 54% (30 of 56) of MVT patients (P=.002).
Life table analysis of the time to first MVT detection revealed
that MVT was detected earlier in MVT than in VF patients
(Figure
). Among patients detecting MVT, patients with a
clinical history of MVT had more episodes than VF patients. The mean
number of MVTs detected by MVT patients was 20±31 compared with
7±7
in VF survivors (P<.001). Only 5% (3 of 55) of VF
survivors detected more than five episodes of MVT compared with 38%
(21 of 56) of MVT patients (P=.002). No clinically apparent
episodes of sustained MVT not treated by the device occurred in the VF
patient group. The mean cycle length of the MVT detected in VF
survivors was significantly shorter than that in MVT patients, 279±39
versus 314±43 ms (P=.03). There was no significant
difference in MVT detection or MVT cycle length among the patients with
a history of MVT as a function of whether they had experienced an
episode of VF before ICD implant. Abrupt-onset VF was detected in 11%
(6 of 55) of the VF survivors, including 2 patients who also detected
MVT. The mean cycle length of these episodes of VF was 194 ms.
Polymorphic VT with a mean cycle length of 265 ms was detected in 7%
(4 of 55) of the patients with a history of VF only. All but 1 of these
4 patients also detected MVT. The acceleration of MVT with
antitachycardia pacing in 14 MVT patients with multiple episodes of MVT
hampered accurate determination of the incidence of primary VF and
polymorphic VT in the MVT patients. Sinus tachycardia was detected in 3
patients (1 VF and 2 MVT). Atrial fibrillation was detected in 8
patients (1 VF and 7 MVT). True MVT was documented at other times in
all but 1 of these 11 patients. Thus, these false detections tended to
occur in patients who had detected true ventricular arrhythmias, and in
most cases, their VT detection zone had been expanded as a result of
frequent episodes of slow MVT.
|
|
Clinical Predictors of MVT Detection
The univariate relative
risk of MVT detection as a function of
baseline characteristics in all patients is presented in Table
3
. A clinical arrhythmia history of MVT, age <60 years,
male sex, and MVT induced on EPS were associated with an increased risk
of MVT detection. A history of coronary artery disease, ejection
fraction <30%, and a history of myocardial infarction were not
statistically associated with an increased risk of MVT. On stepwise Cox
regression, inducibility of MVT on EPS was the most
powerful independent predictor of MVT detection (Table 4
).
After the
results of EPS were taken into account, a lower ejection fraction and
no history of myocardial infarction were independently associated with
a higher risk of MVT detection. If inducibility on EPS is left out of
the model, then arrhythmia history becomes the most powerful predictor
of MVT detection. Virtually identical results were found when stepwise
linear regression was used to determine the independent predictors of
the number of MVT episodes on follow-up. If inducibility is included in
the model, then it becomes the only independent predictor; otherwise,
the arrhythmia history is the only independent predictor. Only 6 of 14
VF survivors (43%) with MVT induced on EPS had MVT detected on
follow-up. Furthermore, 4 of the 10 VF survivors who detected MVT had
not had MVT induced on EPS; 2 of these 4 patients had undergone EPS
both before and after ICD implant. Thus, EPS was of minimal incremental
predictive value once the arrhythmia history had been taken into
consideration.
|
|
| Discussion |
|---|
|
|
|---|
Our findings can also be interpreted to support the hypothesis that some episodes of VF are the end result of MVT deteriorating to VF. We found that MVT detected in patients with a history of only VF was faster than in patients with a history of only MVT, 279±39 versus 314±43 ms (P=.03). This faster MVT may be more likely to precipitate VF.8 9 10 Because the ICD treated MVTs early in their course, we were not able to document rapid MVT deteriorating to VF and can make no assessment as to the frequency of such an event. Thus, there may be several ECG mechanisms by which VF occurs clinically.
Only a few investigators have examined the differences between patients who present with MVT and those who present with VF. As in our patients, those studies showed that VF survivors tend to have a higher ejection fraction1 and are less likely to have MVT induced with EPS1 2 3 4 5 6 than are MVT patients. Adhar et al1 reported the results of EPS performed on 66 survivors of out-of-hospital VF and compared the results with those in 35 patients presenting with hemodynamically well-tolerated sustained VT. Sustained MVT was induced in only 25% of the VF survivors, compared with 69% of the patients with a history of VT, whereas polymorphic VT or VF was induced in 50% and 12%, respectively. Although, perhaps, fewer of the patients with a history of VT had MVT induced than one might expect,4 the difference in inducibility between them and the VF survivors persists. On multivariate analysis, the inducibility of rapid polymorphic rhythms was the only clinical or EPS variable that independently distinguished the two groups. These findings again suggest that a basic difference exists in the electrophysiology of patients who present with VF compared with those who have had sustained VT.
To investigate the value of EPS-guided therapy in cardiac arrest patients, Poole et al5 reviewed mortality and arrhythmia recurrence rates as a function of the results of EPS in 241 VF survivors followed for 30±15 months. VF was induced in 39 patients (16%) and sustained VT in 66 (27%). The suppression of inducible VT or VF with type 1 antiarrhythmic medications was not associated with either improved arrhythmia-free survival or improved total mortality compared with patients who were not suppressed. One interpretation of these results is that the rhythm induced and suppressed in most of these patients, MVT, was not the rhythm responsible for either the patients' presenting cardiac arrest or their recurrent episode and that antiarrhythmic drug suppression of MVT inducibility is not a marker of effective therapy in VF survivors. This hypothesis is supported by our present findings that on follow-up after defibrillator implantation, spontaneous MVT is much less common in VF survivors compared with patients who present with MVT and that VF survivors have a relatively high incidence of primary VF not precipitated by MVT.
Clinical Predictors of MVT Detection
Although the
inducibility of MVT on EPS was the most powerful
independent predictor of the detection of MVT on follow-up in the
patient population as a whole, when the analysis was restricted to
patients with a clinical history of only VF, the results of EPS were
less helpful. In only 6 of 14 VF survivors with MVT induced on EPS was
it detected during follow-up, whereas 4 of 10 patients who detected MVT
had not had MVT induced with EPS. Thus, the results of EPS were of
minimal incremental value in assessing the risk of MVT detection once
the clinical arrhythmia history was considered.
Past myocardial infarction has been identified as a marker for the inducibility of MVT in patients undergoing EPS. Similarly, in our study, on univariate analysis both coronary artery disease and a history of myocardial infarction were associated with a trend toward an increased likelihood of MVT detection on follow-up. But by stepwise Cox regression, after correction for ejection fraction and the results of EPS, a history of myocardial infarction was associated with a reduced likelihood of MVT detection. This finding should not be interpreted to suggest that a history of myocardial infarction somehow reduces the risk of MVT on follow-up. More likely, the ejection fraction and the results of EPS are in part surrogate markers for a history of myocardial infarction; once these factors are corrected for the absence of a history of myocardial infarction, they may identify patients with idiopathic nonreentrant MVT who may not be readily induced with EPS but are still at significant risk for spontaneous MVT on follow-up.
We were not able to assess the risk of MVT or VF detection as a function of the degree of inducible ischemia or the severity of coronary artery disease beyond the information contained in the ejection fraction and the history of past myocardial infarction. Our patient population included some patients with inducible ischemia, but we were careful to exclude patients who were judged to have had an ischemic trigger to their presenting arrhythmia. We excluded patients undergoing epicardial ICD implantation with or without CABG because of concerns that the potential proarrhythmic effects of such procedures might bias outcome. Our findings, therefore, cannot necessarily be extrapolated to patients who undergo epicardial ICD implantation or have revascularization as part of their arrhythmia therapy, but they are applicable to patients who have previously undergone CABG (40% percent of our population).
Implications for ICD Use
The observed differences between VF
survivors and MVT patients may
have significant implications for ICD selection and programming. If VF
survivors are unlikely to have MVT on follow-up, then a tiered-therapy
ICD capable of antitachycardia pacing and/or low-energy cardioversion
may be unnecessary. When VF survivors in our study did have MVT, the
cycle length was short (279 ms). Because available defibrillators
classify arrhythmias largely on the basis of their cycle lengths, MVTs
with cycle lengths <300 ms frequently fall into the programmed VF
detection zone. Thus, differentiating rapid MVT from VF becomes a
significant problem in patients with fast MVT and further limits the
value of implanting a tiered-therapy ICD in VF patients. Finally, as
the cycle length of MVT shortens, the likelihood of success of
antitachycardia pacing therapy decreases.17 Thus, the
infrequent occurrence and the rapid rate of the MVT detected in
patients with a history of only VF may significantly limit the utility
of tiered-therapy ICDs in this patient group. On the other hand, the
results of EPS were of minimal incremental value in predicting the
occurrence or frequency of MVT in VF survivors after consideration of
clinical arrhythmia history. This makes EPS of doubtful utility in
assessing the value of implanting a tiered-therapy ICD versus a
shock-only ICD in VF survivors. Any cost savings achieved by using a
shock-only device would be offset by the cost of EPS used to determine
the type of ICD to implant.
Limitations
Several aspects of this study may limit the power
of our
conclusions. First, the ICD used stores cycle length data but does not
store electrograms or time of therapy. The classification of
arrhythmias on the basis of these cycle length data is most likely to
cause errors in determining the timing and number of episodes per
patient and in differentiating rapid atrial fibrillation from
polymorphic VT. We did not use polymorphic VT as a major end point and
classified detected tachycardias as atrial fibrillation in only eight
patients. All but one of these eight patients had true MVT documented
at another time. Furthermore, the relative infrequency of events in VF
survivors combined with routine interrogation, usually within 24 hours
of clinical events in all patients, led us to be able to examine cycle
length data such that an ensuing episode of VF or other arrhythmia did
not prevent examination of a possible episode of MVT in any patient who
had not otherwise had MVT documented.
A second potential limitation is that our assessment of the predictive value of EPS may have been affected by the fact that EPS was not performed preoperatively in our later VF patients. Instead, single-site EPS was performed through the ICD in the first post-operative week. Although 20 (40%) of the VF survivors and 4 (7%) of the patients with a history of MVT included in the analysis of EPS had EPS performed at only a single site at a postoperative study, only 1 patient from each group detected MVT after a negative single-site EPS study. Thus, although the yield of pacing at a single site is less than that of pacing at two sites,4 we do not feel that this difference in yield had a significant impact on our analysis of the predictive value of EPS for MVT detection either in MVT or in VF patients. We have chosen not to routinely perform EPS before ICD implantation in patients presenting with VF because the results of EPS in such patients have not proved to provide information that is critical to therapeutic decision making.18 The yield of EPS for finding otherwise unsuspected, treatable causes of VF, such as inducible rapid supraventricular tachycardia, has been very low in patients without other historical or clinical factors to suggest it. Moreover, the results of the CASCADE study19 suggested that EPS-guided drug therapy in VF survivors is inferior to empirical amiodarone therapy, making the demonstration of inducible arrhythmias relatively unimportant in therapeutic decision making. In addition, we have not found the success or failure of attempts at MVT pace termination in the electrophysiology laboratory to be predictive of the efficacy of ICD antitachycardia pacing therapies after device implantation.
Another potential limitation of our study is related to our population. As a referral center in an area served by an emergency medical system with rapid response times, it is possible that our patient group may include more true spontaneous VF survivors than would be found in other areas in which prompt resuscitation is not available. In areas with less rapid emergency response, more survivors of out-of-hospital VF may have had MVT that, over time, deteriorated to VF. If this is the case, physicians in other regions may find a higher incidence of spontaneous MVT in VF patients. Last, the 14-month follow-up time is relatively short; longer follow-up might have resulted in a higher or lower relative incidence of MVT and VF.
Conclusions
The results of this study strongly suggest that
after ICD
implantation, MVT is much less likely to be detected in patients with a
history of only VF than in those with a clinical history of MVT. This
lower incidence of MVT detection, the shorter cycle length of the MVT
that is detected, and the detection of spontaneous VF in patients with
a history of only VF support the concept that these patients form a
unique population that should be distinguished from patients with a
history of MVT. Furthermore, the results of EPS were found to be of
minimal value in predicting the detection of MVT in VF-only patients.
These findings may have important implications for understanding the
natural history of patients resuscitated from VF, the value of EPS
before defibrillator implantation, and the type of defibrillator best
suited for certain patient groups.
| Acknowledgments |
|---|
Received July 7, 1994; revision received October 17, 1994; accepted November 6, 1994.
| References |
|---|
|
|
|---|
2. Schoenfeld MH, McGovern B, Garan H, Kelly E, Grant G, Ruskin JN. Determinants of the outcome of electrophysiologic study of patients with ventricular tachyarrhythmias. J Am Coll Cardiol. 1985;6:298-306. [Abstract]
3. Dolack GL, Callahan DB, Bardy GH, Greene HL. Signal-averaged electrocardiographic late potentials in resuscitated survivors of out of hospital ventricular fibrillation. Am J Cardiol. 1990;65:1102-1104. [Medline] [Order article via Infotrieve]
4. Josephson ME. Recurrent ventricular tachycardia. In: Josephson ME. Clinical Cardiac Electrophysiology: Techniques and Interpretations. Philadelphia, Pa: Lea & Febiger; 1993:181-274.
5. Poole JE, Mathisen TL, Kudenchuk PJ, McAnulty JH, Swerdlow CD, Bardy GH, Greene HL. Long-term outcome in patients who survive out of hospital ventricular fibrillation and undergo electrophysiologic studies: evaluation by electrophysiologic subgroups. J Am Coll Cardiol. 1990;16:657-665. [Abstract]
6.
Stevenson WG, Brugada P, Waldecher B, Zehender M, Wellens
HJJ. Clinical, angiographic and electrophysiologic findings in patients
with aborted sudden death as compared with patients with sustained
ventricular tachycardia after myocardial infarction.
Circulation. 1985;71:1146-1152.
7. Wilber DJ, Garan H, Finklestein D, Kelly E, Newell J, McGovern B, Ruskin J. Out-of-hospital cardiac arrest: use of electrophysiologic testing in the prediction of long-term outcome. N Engl J Med. 1988;318:19-24. [Abstract]
8. Hamer AWF, Rubin SA, Peter T, Mandel WJ. Factors that predict syncope during ventricular tachycardia in patients. Am Heart J. 1984;107:997-1005. [Medline] [Order article via Infotrieve]
9.
Lima JAC, Weiss JL, Guzman PA, Weisfeldt ML, Reid PR, Traill
TA. Incomplete filling and incoordinate contraction as mechanism of
hypotension during ventricular tachycardia in man.
Circulation. 1983;68:928-938.
10. Raichlen JS, Links JM, Reid PF. Effect of electrical activation site on left ventricular performance in ventricular tachycardia patients with coronary heart disease. Am J Cardiol. 1985;55:84-88. [Medline] [Order article via Infotrieve]
11.
Gradman AH, Bell PA, DeBusk RF. Sudden death using ambulatory
monitoring. Circulation. 1977;55:210-211.
12. Lahiri A, Balasubramian V, Rafferty EB. Sudden death during ambulatory monitoring. Br Med J. 1979;1:1676-1678.
13.
Nikolic G, Bishop RL, Singh JB. Sudden death recorded during
Holter monitoring. Circulation. 1982;66:218-225.
14. Pratt CM, Francis MJ, Luck JC, Wyndham CR, Miller RR, Quinones MA. Analysis of ambulatory electrocardiograms in 15 patients during spontaneous ventricular fibrillation with special reference to preceding arrhythmic events. J Am Coll Cardiol. 1983;2:789-797. [Abstract]
15.
Bardy GH, Hofer B, Johnson G, Kudenchuk PJ, Poole JE, Dolack
GL, Geleva M, Mitchel R, Kelso D. Implantable transvenous
cardioverter-defibrillators. Circulation. 1993;87:1152-1168.
16. Bardy GH, Troutman C, Poole JE, Kudenchuk PJ, Dolack GL, Johnson G, Hofer B. Clinical experience with a tiered-therapy multiprogrammable antiarrhythmia device. Circulation. 1992;85: 1689-1698.
17. Newman D, Dorian P, Hardy J. Randomized controlled comparison of antitachycardia pacing algorithms for termination of ventricular tachycardia. J Am Coll Cardiol. 1993;21:1413-1418. [Abstract]
18. Dolack GL, Poole JE, Kudenchuk PJ, Johnson G, Mitchell R, Bardy GHQ. Management of ventricular fibrillation without baseline electrophysiologic testing off antiarrhythmic drugs using transvenous implantable cardioverter-defibrillators. Circulation. 1993;88(suppl I):I-167. Abstract.
19. Greene HL, the CASCADE Investigators. Randomized antiarrhythmic drug therapy in survivors of cardiac arrest (the CASCADE Study). Am J Cardiol. 1993;72:280-287.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
J. M Morgan PATIENTS WITH VENTRICULAR ARRHYTHMIAS: WHO SHOULD BE REFERRED TO AN ELECTROPHYSIOLOGIST? Heart, December 1, 2002; 88(5): 544 - 550. [Full Text] [PDF] |
||||
![]() |
A. E. Buxton, K. L. Lee, G. E. Hafley, D. G. Wyse, J. D. Fisher, M. H. Lehmann, L. A. Pires, M. R. Gold, D. L. Packer, M. E. Josephson, et al. Relation of Ejection Fraction and Inducible Ventricular Tachycardia to Mode of Death in Patients With Coronary Artery Disease: An Analysis of Patients Enrolled in the Multicenter Unsustained Tachycardia Trial Circulation, November 5, 2002; 106(19): 2466 - 2472. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. S. Wathen, M. O. Sweeney, P. J. DeGroot, A. J. Stark, J. L. Koehler, M. B. Chisner, C. Machado, and W. O. Adkisson Shock Reduction Using Antitachycardia Pacing for Spontaneous Rapid Ventricular Tachycardia in Patients With Coronary Artery Disease Circulation, August 14, 2001; 104(7): 796 - 801. [Abstract] [Full Text] [PDF] |
||||
![]() |
R.N.W. Hauer, E. Aliot, M. Block, A. Capucci, B. Luderitz, M. Santini, and P.E. Vardas Indications for implantable cardioverter defibrillator (ICD) therapy. Study Group on Guidelines on ICDs of the Working Group on Arrhythmias and the Working Group on Cardiac Pacing of the European Society of Cardiology Eur. Heart J., July 1, 2001; 22(13): 1074 - 1081. [PDF] |
||||
![]() |
G. Gronefeld and S. H Hohnloser What do implantable cardioverter/defibrillators teach us about the mechanisms of sudden cardiac death? Cardiovasc Res, May 1, 2001; 50(2): 232 - 241. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. D. Swerdlow Implantation of Cardioverter Defibrillators Without Induction of Ventricular Fibrillation Circulation, May 1, 2001; 103(17): 2159 - 2164. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. N. W. Hauer, E. Aliot, M. Block, A. Capucci, B. Luderitz, M. Santini, and P. E. Vardas Indications for implantable cardioverter defibrillator (ICD) therapy: Study Group on Guidelines on ICD of the Working Group on Arrhythmias and the Working Group on Cardiac Pacing of the European Society of Cardiology Europace, January 1, 2001; 3(3): 169 - 176. [PDF] |
||||
![]() |
D. Bansch, M. Castrucci, D. Bocker, G.u. Breithardt, and M. Block Ventricular tachycardias above the initially programmed tachycardia detection interval in patients with implantable cardioverter-defibrillators: Incidence, prediction and significance J. Am. Coll. Cardiol., August 1, 2000; 36(2): 557 - 565. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. P. Andrews, R. I. Fogel, G. Pelargonio, J. J. Evans, and E. N. Prystowsky Implantable defibrillator event rates in patients with unexplained syncope and inducible sustained ventricular tachyarrhythmias: A comparison with patients known to have sustained ventricular tachycardia J. Am. Coll. Cardiol., December 1, 1999; 34(7): 2023 - 2030. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Mont, M. Valentino, A. Sambola, M. Matas, L. Aguinaga, and J. Brugada Arrhythmia recurrence in patients with a healed myocardial infarction who received an implantable defibrillator: analysis according to the clinical presentation J. Am. Coll. Cardiol., August 1, 1999; 34(2): 351 - 357. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Ruppel, C. A. Schluter, S. Boczor, T. Meinertz, M. Schluter, K.-H. Kuck, and R. Cappato Ventricular tachycardia during follow-up in patients resuscitated from ventricular fibrillation: experience from stored electrograms of implantable cardioverter-defibrillators J. Am. Coll. Cardiol., November 15, 1998; 32(6): 1724 - 1730. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. D. Swerdlow, C. T. Peter, R. M. Kass, E. S. Gang, W. J. Mandel, C. Hwang, D. J. Martin, and P.-S. Chen Programming of Implantable Cardioverter-Defibrillators on the Basis of the Upper Limit of Vulnerability Circulation, March 18, 1997; 95(6): 1497 - 1504. [Abstract] [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1995 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |