(Circulation. 1995;91:1629-1632.)
© 1995 American Heart Association, Inc.
Articles |
From the Cardiac Arrhythmia Center at the University of Minnesota, Minneapolis.
Correspondence and reprint requests to Keith G. Lurie, MD, Box 508 UMHC, 420 Delaware St SE, Minneapolis, MN 55455.
| Abstract |
|---|
|
|
|---|
Methods and Results ACD CPR was performed with a pneumatically driven automated compression-decompression device in a porcine model of ventricular fibrillation (VF). Nine pigs were randomized to receive ACD CPR alone, while 8 pigs received ACD CPR plus intermittent impedance to inspiratory gas exchange with a threshold valve set to 40 cm H2O. Results comparing 2 minutes of ACD CPR alone versus ACD CPR with the inspiratory impedance threshold valve (ITV) revealed significantly higher mean (±SEM) coronary perfusion pressures (diastolic aortic minus diastolic right atrial pressures) in the ITV (31.0±2.3 mm Hg) group versus with ACD CPR alone (21±3.6 mm Hg) (P<.05). Total left ventricular and cerebral blood flows, determined by radiolabeled microspheres, were 0.77±0.095 and 0.47±0.06 mL/min per gram, respectively, with ACD CPR plus the ITV versus 0.45±0.1 and 0.32±0.016 mL/min per gram, respectively, with ACD CPR alone (P<.05). Similar improvements in the ITV group were observed after 7 minutes of ACD CPR. After 16 minutes of VF and 13 minutes of ACD CPR, 6 of 8 pigs in the ITV group were successfully resuscitated with less than three successive 150-J shocks, whereas only 2 of 9 pigs with ACD CPR alone were resuscitated with equivalent energy levels (P<.02). With up to three additional and successive 200-J shocks, all pigs in the ITV group and 7 of 9 pigs with ACD CPR alone were resuscitated (P=.18).
Conclusions Intermittent impedance to inspiratory flow of respiratory gases during ACD CPR significantly improves coronary perfusion pressures and vital organ blood flow and lowers defibrillation energy requirements in a porcine model of VF.
Key Words: cardiopulmonary resuscitation fibrillation
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
The preparatory phase, which included cannulation of both femoral arteries and the right jugular vein, as well as calibration of all instruments, took approximately 2 hours. Once venous access was obtained, animals received normal saline solution at approximately 300 to 400 mL/h to maintain diastolic right atrial pressures of 3 to 5 mm Hg. Arterial blood gases were analyzed (Instrumentation Laboratory SpA) every 30 minutes to ensure adequate acid base status and oxygenation. Left ventricular and ascending aortic arch blood pressures were monitored using a single high-fidelity micromanometer catheter (Millar). This aortoleft ventricular catheter had a lumen for injecting radiolabeled microspheres and it was positioned, under fluoroscopic guidance, 15 minutes before initiation of VF. Right atrial pressures were also monitored using a micromanometer catheter (Millar) inserted through a right jugular vein sheath. The micromanometer catheters were calibrated to atmospheric pressure immediately before they were inserted into the pig. A 5F bipolar pacing catheter (Daig, Inc), used to induce VF with alternating current at 7 V and 60 Hz, was inserted through a second right jugular vein sheath and positioned by fluoroscopy in the right ventricular apex. For withdrawal of reference blood samples to measure organ blood flow, a 7F catheter was advanced by femoral arterial access to the aortic arch. Body temperature was monitored continuously via a rectal probe (Yellow Springs Instrument Co). Core temperatures were maintained between 36.5 and 38.5°C with a heating pad. Five minutes before induction of VF, 5000 U IV of sodium heparin was administered.
ACD CPR was performed with a 9.0-cm silicon suction cup positioned
fluoroscopically over the right and left ventricles. It was attached to
a pneumatically driven automatic piston device for ACD CPR (ACD
Controller, Ambu International) as previously described.3
The automatic piston device was modified such that it was attached to
the base of the surgical table with a rigid steel pole (diameter=7 cm)
as shown in Fig 1
. Compression and decompression forces
were measured continuously using a piezoelectric force transducer
(model 208A02, PCB Piezotronics) mounted directly in the force path
between the plunger cup and piston arm of the automatic
compression-decompression device. Compression and decompression
excursion was measured continuously by the voltage output of a linear
variable differential transformer (model 3000 HR-DC, Lucas Schaevitz
DC-LVDT). The transformer core moved orthogonally to the chest wall
such that voltage output was linearly proportional to distance traveled
above and below the neutral chest wall position. The compression depth
was also monitored visually with a ruler attached to the piston
housing. Both compression depth and force were monitored on a chart
paper recorder (ES 1000, Gould Instrument Systems) and digitized
on-line (SUPERSCOPE II v1.295, GW Instruments) with a
computerized recording system (Power Macintosh 7100/66 computer) to
allow for instantaneous feedback to the operators.
|
Digitized data were analyzed electronically to provide hemodynamic measurements. The compression-decompression phases were divided in half by using the point at which the suction cup passed the neutral position at the beginning of compression and the beginning of active decompression. Coronary perfusion pressure was calculated during systole and diastole and was defined as the aortoright atrial pressure difference (time-coincident difference between aortic and right atrial diastolic pressures) as previously defined.2 3 4
The protocol was designed to compare ACD CPR alone with ACD CPR plus an inspiratory ITV. Before induction of VF, baseline hemodynamic and arterial blood gas measurements were obtained. Radiolabeled microspheres were injected into the left ventricle to determine baseline vital organ blood flows, as previously described.4 VF was induced with a single 5-second application of alternating current. The endotracheal tube (ET Tube Hi-Lo Jet, Mallincrott, Inc) was immediately disconnected from the mechanical ventilator and the cuff pressure was assessed to ensure that it was adequate to seal the trachea. The animals were then randomized to receive either ACD CPR alone or ACD CPR with the ITV. After 3 minutes of VF, during which time no CPR or ventilation was performed, ACD CPR was performed with the automatic ACD device. The compression-decompression rate was 80 per minute with a 50% duty cycle, a depth of 25% of the anterior-posterior diameter of the chest wall, and a velocity of 7.5 in/s. Active decompression was performed to produce a sternal displacement of 10% greater than the resting anteroposterior diameter, for which a suction force of approximately 200 N was applied. Compression and decompression excursion were continuously monitored and adjusted with the control module, as necessary, during the experiment.
Ventilatory support and ACD CPR were performed simultaneously. During ACD CPR, ventilation was provided by manual bag ventilation (Ambu bag) with 10 L/min oxygen. Respirations were delivered as previously described4 at a rate of one breath every five compressions with a constant tidal volume of approximately 500 mL. In the eight pigs randomized to the ITV group, respiratory gas exchange was limited by the inspiratory valve. In these experiments two 20 cm H2O threshold valves (Ambu, Inc) were connected in series between the endotracheal tube and the Ambu bag such that during active decompression, but in the absence of a manual ventilation, the valves would open only after 40 cm H2O of inspiratory pressure. In this manner, more than -40 cm H2O of intrathoracic pressure was required for inspiration of respiratory gases during four of every five compression-decompression cycles.
Once either ACD CPR alone or ACD CPR plus the ITV was initiated, the same method was performed continuously for 13 minutes during VF. Arterial blood gases were measured immediately before VF and after 2 and 7 minutes of ACD CPR. Radiolabeled microspheres were injected into the left ventricle after 2 minutes and 7 minutes of CPR and after 15 minutes of ROSC. After 16 minutes of VF and 13 minutes of ACD CPR, defibrillation was attempted by transthoracic shock. Defibrillation (43100 A Defibrillator, Hewlett-Packard) was performed initially with a 150-J shock delivered immediately after the automatic ACD CPR machine was turned off. At this time the pig was reconnected to the mechanical ventilator. If the 150-J shock was not successful, two additional 150-J shocks were delivered within the next 15 seconds. Assessment of heart rate and arterial pressure was performed on a continuous basis to monitor for ROSC. If after three successive 150-J shocks spontaneous circulation was not restored, ACD CPR was performed for an additional 10 seconds before delivery of a 200-J shock. A total of three successive 200-J shocks were delivered, as necessary. If ROSC was not possible after the third 200-J shock, no further efforts were made to resuscitate the animals. When the animals were resuscitated, pigs were mechanically ventilated as before the cardiac arrest and continued to receive 300 mL of normal saline solution for the subsequent 15 minutes. At that time, the final radiolabeled microspheres were injected. Myocardial, brain, and kidney blood flows were determined with radiolabeled microspheres as previously described.4 Animals were then sacrificed 5 minutes after reinduction of VF.
All values are expressed as mean±SEM. Statistical analysis was performed using unpaired Student's t test. Statistical significance was considered to be at the P<.05 level.
| Results |
|---|
|
|
|---|
|
Baseline blood flows were similar in both groups. For ACD CPR alone, baseline total left ventricular, right ventricular, and cerebral blood flows were 1.99±0.26, 1.51±0.21, and 0.4±0.04 mL/min per gram, respectively. In the ITV group, baseline total left ventricular, right ventricular, and cerebral blood flows were 1.76±0.38, 1.54±0.36, and 0.38±0.05 mL/min per gram, respectively.
As shown in Fig 2
, total myocardial blood flow was
significantly higher in pigs after ACD CPR with the ITV when compared
with pigs receiving ACD CPR alone. After 2 minutes of ACD CPR, total
left ventricular blood flow was 0.77±0.095 mL/min per gram, or 56% of
baseline, with the ITV versus 0.45±0.10 mL/min per gram, or 24% of
baseline, with ACD CPR alone (P<.01). After 2 minutes of
ACD CPR, right ventricular blood flow was 1.04±0.13 mL/min per gram
with the ITV versus 0.6±0.16 mL/min per gram with ACD CPR alone
(P<.01). Cerebral blood flow was also significantly
increased in animals treated with the ITV, as shown in Fig 2
.
Similar
improvements with the ITV were observed after 7 minutes of ACD CPR.
Renal blood flow was not altered by the ITV. After 2 minutes of ACD CPR
plus ITV, it was 0.44±0.07 mL/min per gram versus 0.38±0.10
mL/min
per gram with ACD CPR alone. The endocardial-to-epicardial blood flow
ratios were also similar between groups. When the data were analyzed by
normalizing blood flow during CPR to baseline blood flows on a
per-animal basis, there was a statistically significant improvement in
myocardial and cerebral blood flow in animals treated with the ITV (Fig
2
).
|
ROSC was achieved in all 8 pigs with the ITV and 7 of 9 pigs treated with ACD CPR alone. However, 6 of 8 pigs in the ITV group and only 2 of 9 pigs with ACD CPR alone had an ROSC after one to three 150-J shocks (P<.03). In animals that had an ROSC, blood flow to the vital organs was similar after 15 minutes between groups. In the ITV group (n=8), left ventricular and cerebral blood flows were 2.1±0.42 and 0.45±0.06 mL/min per gram, respectively, versus 2.7±0.40 and 0.37±0.03 mL/min per gram without the ITV (n=6). One pig in the ACD CPR alone group died within 10 minutes after ROSC.
Arterial blood gas data are shown in the Table
. There were
no
statistically significant differences in arterial blood pH,
PO2, or PCO2
after 2 minutes of ACD CPR. By 7 minutes of ACD CPR, the arterial pH
and PO2 levels were significantly lower in the
ITV group.
| Discussion |
|---|
|
|
|---|
Relatively little is known about the effects of ACD CPR on lung function. We have previously observed that ACD CPR alone significantly enhances minute ventilation compared with standard CPR.1 2 No evidence of lung damage was found in either group at autopsy. A theoretical disadvantage of a very negative intrathoracic pressure during active decompression is negative pressure pulmonary edema. However, this was not observed grossly and is generally observed only with negative inspiratory pressures greater than -50 cm H2O.
Insertion of the ITV during ACD CPR eventually leads to lower pH and PO2 values. When minute ventilation was decreased with the ITV, progressive acidosis occurred to a greater extent when compared with ACD CPR alone. However, these values remained within normal physiological limits throughout the 16 minutes of VF and ACD CPR when supplemental O2 was used. Although the blood pH was >7.3 in both groups after 16 minutes of VF and pH differences did not alter the chances for successful defibrillation with lower energy requirements in the ITV group, it will be important to assess whether the observed decreases in PO2 and pH are significant in the clinical setting.
Venous blood flow is increased clinically by the Mueller maneuver, a technique in which inspiration is performed when the trachea is simultaneously occluded by the epiglottis.6 By using a mechanical valve in an analogous physiological fashion during ACD CPR, we demonstrated in this study that blood flow to the vital organs and the chances for ROSC with defibrillation were significantly increased. Additional mechanisms may also contribute to these improvements. On the basis of these favorable results, the potential benefits of intermittent impedance of inspiratory gas exchange during ACD CPR should be studied in patients in cardiac arrest.
| Acknowledgments |
|---|
Received November 28, 1994; revision received January 4, 1995; accepted January 17, 1995.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
V. Srinivasan, V. M. Nadkarni, D. Yannopoulos, B. S. Marino, G. Sigurdsson, S. H. McKnite, M. Zook, D. G. Benditt, and K. G. Lurie Rapid Induction of Cerebral Hypothermia Is Enhanced With Active Compression-Decompression Plus Inspiratory Impedance Threshold Device Cardiopulmonary Resusitation in a Porcine Model of Cardiac Arrest J. Am. Coll. Cardiol., February 21, 2006; 47(4): 835 - 841. [Abstract] [Full Text] [PDF] |
||||
![]() |
Part 4: Advanced Life Support Circulation, November 29, 2005; 112(22_suppl): III-25 - III-54. [Full Text] [PDF] |
||||
![]() |
B. B. Wolcke, D. K. Mauer, M. F. Schoefmann, H. Teichmann, T. A. Provo, K. H. Lindner, W. F. Dick, D. Aeppli, and K. G. Lurie Comparison of Standard Cardiopulmonary Resuscitation Versus the Combination of Active Compression-Decompression Cardiopulmonary Resuscitation and an Inspiratory Impedance Threshold Device for Out-of-Hospital Cardiac Arrest Circulation, November 4, 2003; 108(18): 2201 - 2205. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Raedler, W. G. Voelckel, V. Wenzel, L. Bahlmann, W. Baumeier, C. A. Schmittinger, H. Herff, A. C. Krismer, K. H. Lindner, and K. G. Lurie Vasopressor Response in a Porcine Model of Hypothermic Cardiac Arrest Is Improved with Active Compression-Decompression Cardiopulmonary Resuscitation Using the Inspiratory Impedance Threshold Valve Anesth. Analg., December 1, 2002; 95(6): 1496 - 1502. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. G. Lurie, T. Zielinski, S. McKnite, T. Aufderheide, and W. Voelckel Use of an Inspiratory Impedance Valve Improves Neurologically Intact Survival in a Porcine Model of Ventricular Fibrillation Circulation, January 1, 2002; 105(1): 124 - 129. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. G. Lurie, W. G. Voelckel, T. Zielinski, S. McKnite, P. Lindstrom, C. Peterson, V. Wenzel, K. H. Lindner, N. Samniah, and D. Benditt Improving Standard Cardiopulmonary Resuscitation with an Inspiratory Impedance Threshold Valve in a Porcine Model of Cardiac Arrest Anesth. Analg., September 1, 2001; 93(3): 649 - 655. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. G. Voelckel, K. G. Lurie, T. Zielinski, S. McKnite, P. Plaisance, V. Wenzel, and K. H. Lindner The Effects of Positive End-Expiratory Pressure During Active Compression Decompression Cardiopulmonary Resuscitation with the Inspiratory Threshold Valve Anesth. Analg., April 1, 2001; 92(4): 967 - 974. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. G. Voelckel, K. G. Lurie, K. H. Lindner, T. Zielinski, S. McKnite, A. C. Krismer, and V. Wenzel Vasopressin Improves Survival After Cardiac Arrest in Hypovolemic Shock Anesth. Analg., September 1, 2000; 91(3): 627 - 634. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. A. Ewy and J. P. Ornato Emergency cardiac care: introduction J. Am. Coll. Cardiol., March 15, 2000; 35(4): 825 - 880. [Full Text] [PDF] |
||||
![]() |
P. Plaisance, K. G. Lurie, and D. Payen Inspiratory Impedance During Active Compression-Decompression Cardiopulmonary Resuscitation : A Randomized Evaluation in Patients in Cardiac Arrest Circulation, March 7, 2000; 101(9): 989 - 994. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|