(Circulation. 1995;91:91-95.)
© 1995 American Heart Association, Inc.
From the Department of Medicine, University of Washington (Seattle).
Correspondence to Gust H. Bardy, MD, Mail Stop RG-22, University Hospital, Seattle, WA 98195.
| Abstract |
|---|
|
|
|---|
Methods and Results We prospectively and randomly compared the
defibrillation efficacy of a 60-µF and a 120-µF capacitance
asymmetrical 65% tilt biphasic waveform using a unipolar
defibrillation system in 38 consecutive cardiac arrest survivors before
implantation of a presently available standard transvenous
defibrillation system. The right ventricular defibrillation electrode
had a 5-cm coil located on a 10.5F lead and was used as the anode. The
system cathode was the electrically active 108-cm2 surface
area shell (or "can") of a prototype titanium alloy pulse
generator placed in a left infraclavicular pocket. The defibrillation
pulse was derived from either a 60-µF or a 120-µF capacitance and
was delivered from RV
CAN. Defibrillation threshold (DFT) stored
energy, delivered energy, leading-edge voltage and current, pulse
resistance, and pulse width were measured for both capacitances
examined. The 60-µF capacitance biphasic pulse resulted in a
stored-energy DFT of 8.5±4.1 J and a delivered-energy DFT of
8.4±4.0
J. In 34 of 38 patients (89%), the stored-energy DFT was <15 J.
Leading-edge voltage at the DFT was 517±128 V. Mean pulse impedance
for the 60-µF waveform was 60.6±7.1
. The 120-µF
capacitance
biphasic pulse resulted in a stored-energy DFT of 10.1±7.4 J and a
delivered-energy DFT of 10.0±7.2 J (P=.13 and .13,
respectively). In 28 of 38 patients (74%), the stored-energy DFT was
<15 J (P=.052). Leading-edge voltage at the DFT with the
120-µF capacitance pulse was 386±142 (P<.00001). Mean
pulse impedance for the 120-µF waveform was 60.7±7.0
(P=.80).
Conclusions The results of the present study suggest that a relatively small capacitance, 60 µF, can be used for unipolar defibrillation systems without compromising defibrillation energy requirements compared with more typical ICD capacitance values, but this will require a higher circuit voltage. The use of lower capacitance also provides a modest increase in the percent of patients who have very low energy defibrillation requirements, an important issue should maximum ICD energy be decreased from the present level of 34 J. Such a move to smaller output devices could allow significant decreases in device size, a necessary feature of making cardioverter-defibrillator implantation comparable to that of standard pacemaker surgery.
Key Words: death, sudden fibrillation tachycardia defibrillation
| Introduction |
|---|
|
|
|---|
Improving defibrillation efficacy is one way to favorably influence
defibrillator size. If a defibrillation pulse and lead system were
capable of terminating ventricular fibrillation in 90% of individuals
using
15 J, maximum ICD shock strength could decrease to 25 J from
its present standard of 34 J and still provide an adequate safety
margin in most patients. An ICD with a maximum output of 25 J would
reduce overall ICD size by using a smaller capacitor and possibly a
smaller battery.
Previous efforts to improve defibrillation efficacy of unipolar defibrillation systems have not proved helpful and have included the addition of a superior vena caval or a coronary sinus electrode to the unipolar system as well as a change in the waveform tilt.2 3 4 Capacitance has previously been explored systematically as a defibrillation variable as early as 19705 ; however, the effect of capacitance with the unipolar ICD system remains unexplored. Thus, the specific purpose of this study was to test in a prospective randomized fashion in humans whether a 60-µF capacitance biphasic waveform defibrillates better than the more standard 120-µF capacitance biphasic waveform when used with a unipolar defibrillation system. If so, it might be possible to further decrease defibrillator size.
| Methods |
|---|
|
|
|---|
|
Unipolar Defibrillation Lead System
The unipolar transvenous
defibrillation system has been fully
described previously.1 Briefly, it consists of a single
10.5F anodal 5-cm-long endocardial right ventricular (RV)
defibrillation electrode (Medtronic model 6966) and a
108-cm2-surface-area cathodal pulse generator titanium
shell electrode (Medtronic model 7219C) that was positioned in a left
infraclavicular pocket. The RV endocardial electrode was inserted into
the left cephalic vein when possible; otherwise, this lead was inserted
into the left subclavian vein. The endocardial RV lead also had
standard bipolar pace/sense electrodes at the tip. Ultimately, the RV
lead was used as part of the permanently implanted standard transvenous
defibrillation system.6 7
Defibrillation Threshold Testing
Defibrillation threshold
testing was performed in a randomized
fashion for both waveforms under study. The first transvenous
defibrillation test began with a 400-V leading-edge voltage pulse
delivered 10 seconds after VF onset, including the time period during
which alternating current was applied. If the transvenous pulse was
unsuccessful, a 100- to 200-J transthoracic rescue pulse was
delivered immediately via a precharged external defibrillator
(Physio-Control LifePak 6s) and anteroposterior cutaneous pads (Darox
Corporation).
After a minimum rest period of 3 minutes between VF inductions, pulse output was increased or decreased, depending on transvenous shock failure or success. Pulse voltages were changed in 100-V steps between voltages of 400 to 900 V and in 50-V steps below 400 V. Between each induction and termination of VF, care was taken to ensure that ECG ST-T segments, QRS duration, and arterial pressure had returned to baseline values before VF was reinitiated.
The defibrillation threshold was defined as the lowest pulse amplitude that could successfully terminate VF 10 seconds after its initiation. After the defibrillation threshold was determined for the first capacitance tested, the defibrillation protocol was repeated for the other capacitance. All defibrillation pulse characteristics were measured from oscilloscopic recordings of voltage and current waveforms as previously described.8
Statistical Analysis
The size of the study population was
determined for a power of
80% likelihood to observe a 25% difference at a P<.05
level. The 25% difference, or 2.3 J, was deemed a clinically
significant improvement from the present average defibrillation
threshold value (9.3±6.0 J) for 120 µF 65% tilt unipolar biphasic
RV
CAN pulses.1 The estimated standard deviation of
the differences between the defibrillation threshold for the two
capacitances in this study was 5.0 J. The minimum population sample
size needed with these assumptions was 38 patients.
A paired two-tailed
t test was used to compare
defibrillation thresholds for the two biphasic waveform capacitances
under study. A
2 comparison for the paired data
was done for percent defibrillation efficacy at 15 J for each
capacitance. A linear regression analysis was conducted to
determine whether a relation existed between pulse resistance and
defibrillation efficacy.
| Results |
|---|
|
|
|---|
Defibrillation Thresholds
The defibrillation threshold data
are given in detail in the
Table
. The defibrillation threshold stored-energy for
the reference 120-µF capacitance biphasic 65% tilt unipolar
defibrillation system was 10.1±7.4 J, with a range of 1.7 to 38.4 J
(Fig 2
). The defibrillation threshold delivered energy
for the 120-µF capacitance biphasic pulse was 10.0±7.2 J, with a
range of 1.7 to 37.8 J. The measured leading-edge voltage
defibrillation threshold for the 120-µF capacitance biphasic unipolar
defibrillation system was 386±142 V, with a range of 169 to 800 V (Fig
3
). The measured leading-edge current defibrillation
threshold for the 120-mF capacitance biphasic pulse was 6.5±2.6 A,
with a range of 2.6 to 13.8 A. The measured-leading edge resistance at
the defibrillation threshold was 60.7±7.0
, with a range of 38.8
to
73.1
. Total pulse width was 17.6±1.8 milliseconds.
|
|
|
In the
case of the 60-µF capacitance 65% tilt biphasic waveform, the
mean defibrillation threshold stored-energy was 8.5±4.1 J, with a
range of 2.2 to 19.2 J (P=.13) (Fig 2
). The
defibrillation
threshold delivered-energy for the 60-µF capacitance biphasic pulse
was 8.4±4.0 J, with a range of 2.2 to 18.9 J. The measured
leading-edge voltage at the defibrillation threshold for the 60-µF
capacitance biphasic waveform was 517±128 V, with a range of 273 to
800 V (P<.00001) (Fig 3
). The measured leading edge
current
defibrillation threshold for the 60-F capacitance biphasic pulse was
8.7±2.5 A, with a range of 4.3 to 14.0 A (P<.00001). Pulse
resistance at the defibrillation threshold was 60.6±7.1
, with a
range of 39.5 to 79.5
(P=.80). Total pulse width was
8.8±1.1 milliseconds (P<.00001).
Percent Defibrillation Efficacy
For the 120-µF
capacitance biphasic pulse, 28 of 38 patients
(74%) were defibrillated by <15 J stored energy. For the 60-µF
capacitance biphasic pulse, 34 of 38 patients (89%) were defibrillated
by <15 J stored energy (P=.052).
Effect of Resistance on Defibrillation Efficacy
Because
resistance can alter the waveform decay constant, the
relation of resistance to defibrillation efficacy was examined. There
was no relative benefit of one capacitance over the other regarding
stored-energy defibrillation threshold and pulse resistance. The ratio
of stored-energy defibrillation thresholds for the two capacitance
outputs failed to show a relation to pulse impedance
(r=-.10, P=.56; Fig 4
).
|
| Discussion |
|---|
|
|
|---|
On a practical level, one of the goals of using a smaller capacitance in the present study was to allow for a reduction in ICD size. However, altering capacitance will not decrease ICD size unless the energy required for storage on the capacitor can be decreased because capacitor size is proportional to the energy stored on it. Therefore, reducing the capacitance to 60 µF is not meaningful if it does not reduce the stored-energy needed for defibrillation.
Although the
outcome of the study with respect to defibrillation
thresholds (DFTs) does not strongly argue for the use of lower
capacitance, it is interesting to note that a greater percentage of
patients could be defibrillated with 60-µF than with 120-µF output
capacitance at a 15-J output (89% versus 74%, P=.052). A
comparison of capacitance effect based on a 15-J cutoff is important
when considering ICD size. For example, the 60-µF output capacitance
would have allowed device implantation in a greater percentage of
patients than a 120-µF capacitance device if one were to use an ICD
with a maximum output that had been reduced to 25 J. For patients with
a DFT
15 J, a 25-J device would be clinically acceptable as it would
provide a defibrillation safety margin within the context of current
practice. A lower maximum ICD energy would result in a smaller
capacitor volume and therefore a smaller ICD.
Corresponding Effect of Tilt and Pulse Width
In any study of
capacitance on defibrillation efficacy, one must
consider the independent defibrillation variables of tilt and pulse
width as each affects the amount of charge and energy delivered to the
tissues. Consequently, the optimal tilt and pulse width for a 60-µF
capacitance pulse may be different than that for a 120-µF capacitance
output. Use of a different tilt than the 65% used in the present
study, or use of a fixed pulse width, may have changed our study
outcome. The selection of a 65% tilt for this study was based on an
earlier investigation examining the effect of tilt on RV
CAN
defibrillation: it showed no significant difference between 65% and
50% tilt at 120-µF capacitance.4 The pertinent question
for the present study is whether the defibrillation thresholds with
60-µF capacitance and 120-µF capacitance would be different if
measurements would have been made at their respective optimal tilts or
pulse widths. The optimal tilt or pulse width at each of these
capacitances for unipolar defibrillation systems is not known and could
differ. That said, it is important to recognize that most of the
clinical studies indicate that there is no significant difference in
defibrillation thresholds with 120-µF capacitance when 50% and 65%
tilts are compared using several different lead
systems.4 10 11 Another recent clinical
study differs from
these reports, however, arguing that 42% tilt biphasic pulses reduce
the defibrillation threshold compared with 65% tilt biphasic pulses
when a 120-µF capacitance was used.12 This finding may
have been a consequence of the significantly lower average resistance
seen in this study (28±6
, J.F. Swartz, personal communication,
unpublished data), which may have resulted from the simultaneous,
dual-pathway coronary sinus defibrillation that was used. The average
impedance in the other studies ranged between 54 and 65
.4 10 11 These findings suggest
that tilt may need to
be customized to lead system impedance whenever capacitance is
altered.13
There have been few clinical studies controlling for the effect of pulse width on defibrillation, a waveform characteristic that can be pertinent to the present study if one considers the potential for high capacitance pulses to prolong the pulse duration. In one earlier clinical study, we examined the effect on defibrillation of varying pulse width for sequential 60-µF monophasic pulses when using an epicardial three-patch system.14 We were able to demonstrate a progressive deterioration in defibrillation efficacy as a function of delivered-energy as pulse width increased from 2.5 to 5.0 to 7.5 milliseconds. Although the lead system and waveform were considerably different than those used in the present study, this study does supply evidence to suggest that a pulse that is too long is detrimental for defibrillation. These findings are consistent with studies in dogs indicating that excessively long pulses may be less effective.5 15 In this light, one must consider the possibility that high capacitance shocks can prolong pulse width for any particular tilt and may prove to be a disadvantageous waveform design despite yielding higher shock energies.
Can Size and Defibrillation Efficacy
Any effort to diminish
ICD size for an active can system,
regardless of means, has the potential to affect defibrillation
adversely by diminishing the electrically active can surface area. If
one were to use a smaller capacitance and ICD size were consequently
reduced, the resulting reduction in ICD surface area could be countered
to some degree by changing ICD shape, for example, by flattening the
ICD more than diminishing its frontal area or by increasing the surface
complexity, ie, with fractal folds. Although this problem will
ultimately require attention, it is unlikely to be clinically evident
for several years. We have looked at defibrillation efficacy for 40-mL
and 60-mL cans of similar size, shape, and surface material to the
80-mL can used in the present study and have not found a decrease
in defibrillation efficacy.16 We suspect the knee of the
defibrillation efficacy curve, as referenced to ICD volume, will
eventually be encountered, although, as can volumes get smaller than 40
mL.
Study Limitations
One limitation of the present study relates
to the type of
pulsing technique and lead system used to examine the relative efficacy
of the two capacitances under study. It is conceivable that a different
electrode configuration with different impedance characteristics or a
more dramatic variation in waveform capacitance or fixation of pulse
width rather than tilt would have resulted in a different conclusion to
that found in the present study. This said, the system used is a
pragmatic one and has the advantage of being in widespread clinical
practice.
An additional limitation regards the statistical power of our results. Because of the high variability of the difference in defibrillation thresholds for 60 µF and 120 µF, the statistical power of the present study is such that the minimum detectable difference is 2.9 J, or 29% of the mean defibrillation threshold for 120 µF. This is higher than the stated goal of detecting a 2.3-J difference and is significantly greater than the 1.6-J mean difference actually observed in the study. In fact, using the results for power calculations rather than previously published data, as done at the outset of this study, the power of establishing the stated goal is 61%. To significantly establish a difference of <2.9 J, a greater number of patients would have been required than our power calculations initially indicated.
Conclusions
The results suggest that a relatively small
capacitance can be
used for unipolar defibrillation systems without compromising
defibrillation energy requirements. It does not, however, significantly
decrease DFT stored-energy requirements. The use of lower capacitance,
on the other hand, is potentially important in that it appears to
increase the percentage of patients with very low defibrillation
requirements. This is important should maximum ICD energy capabilities
be decreased from their present 34-J level to a 25-J level. Such a
move to lower energy output ICDs would, in turn, decrease device size,
an essential feature of making ICD surgery comparable to that of
standard pacemaker surgery. The practical procedural advantages of a
smaller ICD are obvious: local anesthesia, a subcutaneous pocket,
outpatient insertion, decreased morbidity, and decreased costs. Smaller
capacitance devices may modestly favor such a development.
| Acknowledgments |
|---|
Received May 27, 1994; accepted July 31, 1994.
| References |
|---|
|
|
|---|
2.
Bardy GH, Dolack GL, Kudenchuk PJ, Poole JE, Mehra R, Johnson
G. Prospective randomized comparison in man of a unipolar
defibrillation system with that utilizing an additional superior vena
cava electrode. Circulation. 1994;89:1090-1093.
3.
Kudenchuk PJ, Bardy GH, Dolack GL, Poole JE, Mehra R, Johnson
G. Efficacy of a single lead unipolar transvenous defibrillator
compared with a system employing an additional coronary sinus
electrode: a prospective randomized study.
Circulation. 1994;89:2641-2644.
4. Bardy GH, Dolack GL, Poole JE, Kudenchuk PJ, Johnson G, Raitt MH, Mehra R, DeGroot P, Hofer BO. A prospective randomized comparison in humans of 50% vs 65% tilt biphasic pulse defibrillation using the unipolar pectoral transvenous defibrillation system. Circulation. 1993;88(suppl I):I-113. Abstract.
5.
Geddes LA, Tacker WA, McFarlane J, Bourland J.
Strength-duration curves for ventricular defibrillation in dogs.
Circ Res.. 1970;27:551-560.
6.
Yee R, Klein GJ, Leitch JW, Guiraudon GM, Guiraudon CM, Jones
DL, Norris C. A permanent transvenous lead system for an implantable
pacemaker cardioverter-defibrillator: nonthoracotomy approach to
implantation. Circulation. 1992;85:196-204.
7.
Bardy GH, Hofer B, Johnson G, Kudenchuk PJ, Poole JE, Dolack
GL, Gleva M, Mitchell R, Kelso D. Implantable transvenous
cardioverter-defibrillators. Circulation. 1993;87:1152-1168.
8. Bardy GH, Ivey TD, Allen MD, Johnson G, Greene HL. Prospective comparison of sequential pulse and single pulse defibrillation using two different clinically available systems in man. J Am Coll Cardiol.. 1989;14:165-171. [Abstract]
9. Bourland JD, Tacker WA, Geddes LA. Strength duration curves for trapezoidal waveforms of various tilts for transchest defibrillation in animals. Med Instr.. 1978;12:38-41. [Medline] [Order article via Infotrieve]
10. Li HG, Yee R, Mehra R, DeGroot P, Zandini M, Morillo CA, Thakur RK, Klein GJ. The effect of biphasic shock waveform tilt upon defibrillation efficacy. Circulation. 1993;88(suppl I):I-113. Abstract.
11. Natale A, Van Hout W, DeGroot P, Deshpande S, Axtell K, Krum D, Sheth M, Sra J. Effects of tilt on biphasic defibrillation efficacy in patients. Circulation. 1993;88(suppl I):I-114. Abstract.
12.
Swartz JF, Fletcher RD, Karasik PE. Optimization of biphasic
waveforms for human nonthoracotomy defibrillation.
Circulation. 1993;88:2646-2654.
13. Swerdlow CD, Kass RM, Hwang C, Chen PS. Effect of capacitor size and pathway resistance on defibrillation threshold in humans. PACE Pacing Clin Electrophysiol.. 1994;17:851. Abstract.
14. Bardy GH, Ivey TD, Stewart RB, Fishbein DP, Poole JE, Kudenchuk PJ, Adhar GC, Greene HL. Sequential pulse defibrillation thresholds in man: effect of pulse width variation. J Am Coll Cardiol. 1987;9:166A. Abstract.
15. Chapman PD, Wetherbee JN, Vetter JW, Troup P, Souza J. Strength-duration curves of fixed pulse width varable tile truncated exponential waveforms for nonthoracotomy internal defibrillation in dogs. PACE Pacing Clin Electrophysiol.. 1988;11:1045-1050. [Medline] [Order article via Infotrieve]
16. Jones GK, Poole JE, Kudenchuk PJ, Dolack GL, Johnson G, DeGroot P, Gleva M, Raitt M, Bardy GH. A prospective randomized evaluation of implantable cardioverter-defibrillator size in unipolar defibrillation system efficacy in man. Circulation 1994;90(suppl I):I-227. Abstract.
This article has been cited by other articles:
![]() |
J. A. Blatt, J. E. Poole, G. W. Johnson, D. J. Callans, M. H. Raitt, R. K. Reddy, F. E. Marchlinski, R. Yee, T. Guarnieri, M. Talajic, et al. No Benefit From Defibrillation Threshold Testing in the SCD-HeFT (Sudden Cardiac Death in Heart Failure Trial) J. Am. Coll. Cardiol., August 12, 2008; 52(7): 551 - 556. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. A. Mowrey, Y. Cheng, P. J. Tchou, and I. R. Efimov Kinetics of defibrillation shock-induced response: design implications for the optimal defibrillation waveform Europace, January 1, 2002; 4(1): 27 - 39. [Full Text] [PDF] |
||||
![]() |
J. Brugada, B. Herse, B. Sandsted, U. Michel, B. D. Schubert, and S. J. Hahn Clinical evaluation of defibrillation efficacy with a new single-capacitor biphasic waveform in patients undergoing implantation of an implantable cardioverter defibrillator Europace, January 1, 2001; 3(4): 278 - 284. [Abstract] [PDF] |
||||
![]() |
P. N. Schauerte, K. Ziegert, M. Waldmann, F. A. Schondube, F. Birkenhauer, K. Mischke, M. Grossmann, P. Hanrath, and C. Stellbrink Effect of Biphasic Shock Duration on Defibrillation Threshold With Different Electrode Configurations and Phase 2 Capacitances : Prediction by Upper-Limit-of-Vulnerability Determination Circulation, March 23, 1999; 99(11): 1516 - 1522. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Schauerte, F. A. Schondube, M. Grossmann, H. Dorge, F. Stein, B. Dohmen, A. Moumen, K. Erena, B. J. Messmer, P. Hanrath, et al. Influence of Phase Duration of Biphasic Waveforms on Defibrillation Energy Requirements With a 70-µF Capacitance Circulation, May 26, 1998; 97(20): 2073 - 2078. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. D. Swerdlow, J. E. Brewer, R. M. Kass, and M. W. Kroll Application of Models of Defibrillation to Human Defibrillation Data : Implications for Optimizing Implantable Defibrillator Capacitance Circulation, November 4, 1997; 96(9): 2813 - 2822. [Abstract] [Full Text] |
||||
![]() |
C. D. Swerdlow, W. Fan, and J. E. Brewer Charge-Burping Theory Correctly Predicts Optimal Ratios of Phase Duration for Biphasic Defibrillation Waveforms Circulation, November 1, 1996; 94(9): 2278 - 2284. [Abstract] [Full Text] |
||||
![]() |
Y. Yamanouchi, K. A. Mowrey, G. R. Nadzam, D. G. Hills, M. W. Kroll, J. E. Brewer, A. M. Donohoo, B. L. Wilkoff, and P. J. Tchou Large Change in Voltage at Phase Reversal Improves Biphasic Defibrillation Thresholds: Parallel-Series Mode Switching Circulation, October 1, 1996; 94(7): 1768 - 1773. [Abstract] [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1995 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |