Donate Help Contact The AHA Sign In Home
American Heart Association
Circulation
Search: search_blue_button Advanced Search
Circulation. 2008;117:714-716
doi: 10.1161/CIRCULATIONAHA.107.719617
This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Fiorentino, R. P.
Right arrow Articles by Uchida, T.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Fiorentino, R. P.
Right arrow Articles by Uchida, T.
Related Collections
Right arrow Catheter-based coronary interventions: stents
Right arrow Other Treatment

(Circulation. 2008;117:714-716.)
© 2008 American Heart Association, Inc.


Editorial

Regulatory Perspective on Embolic Protection Device Approval for Saphenous Vein Graft Stenting With a Single-Arm Trial Using Risk-Adjusted Prediction Model

Robert P. Fiorentino, MD, MPH; Bram Zuckerman, MD; Takahiro Uchida, MD, MSc

From the Center for Devices and Radiological Health, US Food and Drug Administration, Rockville, Md.

Correspondence to Takahiro Uchida, MD, MSc, Center for Devices and Radiological Health, 9200 Corporate Blvd, HFZ-450, Rockville, MD 20850. E-mail takahiro.uchida.md{at}gmail.com


Key Words: Editorials • embolism • grafting • equipment and supplies • stents • United States Food and Drug Administration • coronary disease


*    Introduction
up arrowTop
*Introduction
down arrowRegulatory Approaches to SVG...
down arrowImplementing the Prediction...
down arrowConclusions
down arrowReferences
 
The article by Coolong et al1 in this issue of Circulation identifies potential predictors of 30-day major adverse cardiac events (MACE) after saphenous vein graft (SVG) percutaneous coronary intervention (PCI) with embolic protection devices (EPDs). These predictors, angiographic estimates of plaque volume and SVG degeneration, were derived from patient-level data on 3958 patients enrolled in 6 clinical trials of SVG EPDs. As discussed in their article, the authors have incorporated these predictors into a model that seeks to accurately predict 30-day MACE rates for such devices. The authors make an intriguing proposal that their covariate-adjusted, historically derived model could be used to construct an objective performance goal for the evaluation of novel EPDs. The implementation of such a model seems an attractive goal not only because of savings in time and cost for future clinical trials but also because of the potential to allow effective devices to reach patients more expeditiously. However, putting the model into practice will likely need some additional forethought and further collaborative effort to ensure that the effectiveness and safety of these devices are adequately evaluated. Although the investigators did not submit the actual predictive model within their article, we appreciate this opportunity to provide our insight into how such a model could potentially be integrated into future trials supporting regulatory approval.

Article p 790


*    Regulatory Approaches to SVG EPDs for PCI
up arrowTop
up arrowIntroduction
*Regulatory Approaches to SVG...
down arrowImplementing the Prediction...
down arrowConclusions
down arrowReferences
 
The Food and Drug Administration (FDA) regulates devices in 3 classes, with the highest-risk devices in class III, moderate-risk devices in class II, and lowest-risk devices in class I. Medical devices in class III support or sustain human life, are of substantial importance in preventing impairment of human health, or present a potential, unreasonable risk of illness or injury. These devices reach the market through the premarket approval process. Because of the level of risk associated with class III devices, the FDA has determined that adequate clinical trials are required to ensure the safety and effectiveness of class III devices. Class II devices, which involve moderate risk, generally require 510(k) premarket submission. The 510(k) submission must demonstrate that the device to be marketed is at least as safe and effective as, ie, substantially equivalent to, a legally marketed device, or "predicate device," that is not subject to premarket approval. About 10% of devices submitted to the FDA through the 510(k) process require clinical data.

EPDs used during PCI in SVGs are regulated as class II devices, requiring clearance of a 510(k) before marketing. Although 510(k) devices may be shown through laboratory and bench testing to be substantially equivalent to a marketed device, clinical data also may be needed to establish safety and effectiveness and therefore substantial equivalence. In line with this regard, the FDA required adequate clinical trials to ensure reasonable safety and effectiveness for EPDs used in the setting of PCI for SVGs. On the initial introduction of these devices, there was some uncertainty regarding the most appropriate control group. It was not clear whether a concurrent control was needed or whether an adequate performance goal could be generated from historical data (or other means). This question was discussed by a Circulatory System Devices Advisory Panel meeting on February 5, 2001.2 The advisory panel agreed that the most appropriate control would be the "standard of practice," which at the time was no EPD use, and recommended that randomized controlled trials were necessary. After the first EPD was cleared and marketed, the standard of practice shifted to become use of an EPD; thus, ongoing and new clinical trials incorporated cleared devices as active controls in noninferiority comparisons.


*    Implementing the Prediction Model for New EPDs
up arrowTop
up arrowIntroduction
up arrowRegulatory Approaches to SVG...
*Implementing the Prediction...
down arrowConclusions
down arrowReferences
 
There are a number of important considerations regarding the validity of the proposed model, including confirmation of the poolability of the studies within the model, generalizability of the model itself to outside patient populations (external validity), and the modeling procedure (internal validity).

Trial poolability of the predictive model will be an important issue for a new EPD market approval using this method. The authors performed an analysis using a bayesian hierarchical model to estimate the intratrial correlation coefficient and concluded that there was little unexplained variation between trials used to generate the model. Although the bayesian analysis indicated that the between-trial variation is relatively small, those trials nevertheless may be somewhat heterogenic from a regulatory perspective. For instance, the CardioShield Application Protects During Transluminal Intervention of Vein Grafts by Reducing Emboli (CAPTIVE) trial,3 which evaluated the EmboShield device, failed to show both superiority to conventional guidewire and noninferiority to GuardWire. Second, the Embolic Protection Transluminally with the FilterWire EZ Device in Saphenous Vein Grafts (BLAZE) II study,4 unlike the others, was a single-arm trial in which the reported 30-day MACE rate was 3.8%, much lower than in the other 11 EPD treatment arms in the analysis (range, 8.2% to 11.4%).

Generalizability is another critical component for evaluating the prediction model for use in single-arm trials seeking market clearance of new EPDs for SVG PCI. The assumption that poolability of trials in this type of prediction model is appropriate rests on accepting that the study populations are homogeneous. In this respect, the prediction model may have powerful advantages for future trials with study populations homogeneous to those used to generate the prediction model. However, this advantage may not apply to specific indications for an EPD evaluated in populations dissimilar to those used to generate the model.

The 2 novel angiographic risk factors proposed, estimated plaque volume and SVG degeneration score, are conceptually intriguing and appear to be clinically sound.5 However, consistency and reproducibility in their measurement in practice may need to be further demonstrated before their acceptance. It also appears that these variables are not entirely objective, despite being measured by a quantitative coronary angiography system. This may introduce a level of bias, especially in single-arm studies, that could unpredictably affect the validity of the predictive model. The use of angiographic core laboratories with prospectively outlined and detailed performance procedures may minimize this potential measurement bias but with some uncertainty likely remaining.

Additionally, the univariate logistic regression analysis used to assess the relationship between the presence of diabetes and 30-day MACE showed that there was a protective relationship with this covariate. This is a curious finding in that diabetes is a significant risk factor for PCI procedures.6 However, the authors note that the univariate analysis was confounded by the finding that diabetics tended to be younger and had lower SVG degeneration scores and smaller estimated plaque volumes. Because the actual model was not provided in the article, we cannot confirm whether this bias was sufficiently adjusted for by the multiple logistic regression model selection; thus, it seems that the presence of diabetes should be included in multivariate logistic regression analysis.

Most critically, sensitivity analyses using external data sets such as other trials not used to generate the model might be a way to validate the prediction model. Such an analysis is important primarily to evaluate whether the model can actually predict 30-day MACE. Moreover, even after adoption of the prediction model, it would seem most appropriate for the model to be continuously updated as new data become available, understanding that clinical settings, adjunctive medical treatments, and novel or iterative devices may be introduced or change over time. Furthermore, it should be noted that although the authors’ study may appropriately conclude that EPDs used for PCI in SVGs are beneficial compared with conventional guidewire use, across multiple EPD types, this does not guarantee that future EPDs will have similar safety and effectiveness profiles.


*    Conclusions
up arrowTop
up arrowIntroduction
up arrowRegulatory Approaches to SVG...
up arrowImplementing the Prediction...
*Conclusions
down arrowReferences
 
A number of elements need to be considered when studies are designed to support medical device market approval. These include defining appropriate target populations, establishing the correct risk-to-benefit ratio based on safety and effectiveness, and choosing appropriate end points for the assessment of device performance. In regard to EPDs for PCI of SVG lesions, the FDA has historically considered randomized controlled trials most appropriate to address variation between studies, including differences in lesion or vessel characteristics, device types, treatment modalities (such as stent type or use of glycoprotein IIb/IIIa antagonists), and patient populations.

Coolong et al1 conclude that their covariate-adjusted prediction model of 30-day MACE can be used to generate patient- and trial-specific objective benchmarks for the evaluation of novel devices in single-arm studies. Accumulating data and increased knowledge about the performance of EPDs in SVGs have made this approach possible. It follows that our regulatory approach toward these devices should evolve as we gain more detailed understanding of the devices and diseases that they are designed to treat. Such a philosophy is incorporated into the "least burdensome" approach that the agency is mandated by law to follow.7 Discussions between the agency and its stakeholders is strongly recommended early in the planning stages of clinical trials to reach agreement on the most appropriate, least burdensome path.

However, in many circumstances, clinical data are held by proprietary interests, limiting the free exchange of data across the various stakeholders within the medical community. For predictive models to be adopted broadly, a cooperative approach between the relevant stakeholders is of paramount importance. Collaboration between industry, academia, and regulatory bodies will be necessary if we are to learn fully from previous clinical trials. Such cooperation will be key to developing novel regulatory pathways for promising medical devices.

We are pleased to see that such efforts are ongoing, as exemplified by the Coolong et al article, and we encourage all stakeholders to adopt and expand such collaboration. Moreover, we believe that such an approach can be applied to other areas of cardiovascular research and medical devices, and we welcome further discussion on the implementation of similar approaches.


*    Acknowledgments
 
We would like to thank our colleagues within the Center for Devices and Radiological Health for their contributions: Suzanne Kaiser; Ashley Boam, MSBE; Heng Li, PhD; and Laura Thompson, PhD.

Disclosures

None.


*    Footnotes
 
The opinions expressed in this article are not necessarily those of the editors, the American Heart Association, or the Food and Drug Administration.


*    References
up arrowTop
up arrowIntroduction
up arrowRegulatory Approaches to SVG...
up arrowImplementing the Prediction...
up arrowConclusions
*References
 
1. Coolong A, Baim DS, Kuntz RE, O’Malley AJ, Marulkar S, Cutlip DE, Popma JJ, Mauri L. Saphenous vein graft stenting and major adverse cardiac events: a predictive model derived from a pooled analysis of 3958 patients. Circulation. 2008; 117: 790–797.[Abstract/Free Full Text]

2. CDRH Advisory Committees. Available at: http://www.fda.gov/ohrms/dockets/ac/cdrh01.htm#Circulatory%20System. Accessed January 17, 2008.

3. Holmes DR. A prospective randomized controlled trial of distal protection with the third generation MedNova EmboShield compared to the GuardWire or FilterWire (CAPTIVE). Paper presented at: Transcatheter Cardiovascular Therapeutics; September 28, 2004; Washington, DC.

4. Cox DA. Stenting in saphenous vein grafts with distal protection using a second generation filter-based catheter: the combined BLAZE I and II registries. Paper presented at: Transcatheter Cardiovascular Therapeutics; October 17, 2005; Washington, DC.

5. Giugliano GR, Kuntz RE, Popma JJ, Cutlip DE, Baim DS. Determinants of 30-day adverse events following saphenous vein graft intervention with and without a distal occlusion embolic protection device. Am J Cardiol. 2005; 95: 173–1777.[CrossRef][Medline] [Order article via Infotrieve]

6. Smith SC Jr, Faxon D, Cascio W, Schaff H, Gardner T, Jacobs A, Nissen S, Stouffer R. Prevention Conference VI: Diabetes and Cardiovascular Disease: Writing Group VI: revascularization in diabetic patients. Circulation. 2002; 105: e165–e169.[CrossRef][Medline] [Order article via Infotrieve]

7. Least Burdensome Provision of the FDA Modernization Act of 1997. Available at: http://www.fda.gov/cdrh/ode/guidance/1332.pdf. Accessed January 17, 2008.




This article has been cited by other articles:


Home page
J Am Coll Cardiol IntvHome page
E. R. Bates
Aspirating and Filtering Atherothrombotic Debris During Percutaneous Coronary Intervention
J. Am. Coll. Cardiol. Intv., June 1, 2008; 1(3): 265 - 267.
[Full Text] [PDF]


This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Fiorentino, R. P.
Right arrow Articles by Uchida, T.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Fiorentino, R. P.
Right arrow Articles by Uchida, T.
Related Collections
Right arrow Catheter-based coronary interventions: stents
Right arrow Other Treatment