(Circulation. 2008;117:1873-1882.)
© 2008 American Heart Association, Inc.
New Drugs and Technologies |
From the Division of Vascular Surgery, Northwestern University Feinberg School of Medicine, Chicago, Ill.
Correspondence to Melina R. Kibbe, MD, Northwestern University, Division of Vascular Surgery, Galter 10–105, 201 E Huron Street, Chicago, IL 60611. E-mail mkibbe{at}nmh.org
| Abstract |
|---|
|
|
|---|
Key Words: atherosclerosis bypass peripheral vascular disease restenosis revascularization
| Introduction |
|---|
|
|
|---|
| Prosthetic Bypass Grafting |
|---|
|
|
|---|
|
|
Several studies have explored side-by-side comparisons of Dacron and ePTFE for above-knee applications and found no statistically significant differences in patency between the 2 conduits.4,5 However, when separated by location of distal anastomosis, the patency rate of ePTFE for infragenicular anastomoses tended to be better compared with Dacron (60% versus 46%, respectively); however, this difference was not statistically significant.5 Hobson et al6 also examined the difference between ePTFE and composite Dacron-autogenous vein grafts to the below-knee popliteal artery. Over a 30-month period, they found that in patients with poor arterial runoff, 45% of ePTFE grafts remained patent; however, all Dacron-autogenous vein grafts occluded within the first 10 months. Because of these poor outcomes, Dacron typically is not used for small-vessel anastomoses.
In smaller-diameter conducting vessels, radial compliance is more important than strength for vascular conduits.2 In an attempt to meet this requirement, polyurethane, which has the highest elasticity among existing polymers, has been developed into vascular grafts. Despite this promising characteristic, 2 properties have limited the clinical application of virtually all polyurethane vascular grafts: poor biostability and loss of compliance after implantation. To overcome this, Seifalian and coworkers7 developed a compliant poly(carbonate-urea)urethane vascular graft, MyoLink (CardioTech International Inc, Wilmington, Mass). Studies showed that it possessed a compliance profile similar to human arteries and was more resistant to biodegradation than earlier versions of polyurethane grafts. Currently, the only polyurethane graft available for use in the United States for vascular surgical procedures is the Vectra graft (Thoratec Corp, Pleasanton, Calif), which has been approved by the US Food and Drug Administration (FDA) for hemodialysis access.
| Cause of Prosthetic Graft Failure |
|---|
|
|
|---|
Poor patency of prosthetic bypass grafts has led several researchers to concentrate efforts on improving available prosthetic materials. With these objectives in mind, a great deal of attention has been focused on surgical modifications, materials and coatings, protein modifications, endothelial cell seeding, and nitric oxide (NO) modifications of existing prosthetic grafts to improve graft patency and clinical outcomes (Table 2). Most of these modifications are designed to improve thrombogenicity, to decrease the development of neointimal hyperplasia, and/or to improve host incorporation and healing. These approaches are not designed to effect graft compliance, suturability, or tensile strength, the former of which remains an important issue to date.
|
| Surgical Modifications of Prosthetic Vascular Grafts |
|---|
|
|
|---|
|
| Graft Materials and Coatings |
|---|
|
|
|---|
An approach for improving long-term graft outcomes is to use coatings to decrease thrombogenicity and neointimal hyperplasia. One such strategy is the use of carbon-impregnated ePTFE grafts. Carbon is biocompatible and has been used in other applications to reduce thrombogenicity. The Impra Carboflo graft (Bard Peripheral Vascular, Inc, Tempe, Ariz), a carbon-modified ePTFE graft, was evaluated in a prospective, randomized multicenter trial compared with standard ePTFE.16 All bypasses were to the anterior tibial artery, and a vein patch or cuff was used in each case. At 36 months, for the carbon and standard ePTFE groups, the primary patency (33% and 30%, respectively) and limb salvage (67% and 58%, respectively) rates were not statistically different.
Several additional graft coatings have undergone preliminary testing. Karrer et al17 coated ePTFE with polypropylene sulfide–polyethylene glycol (PEG) and tested these grafts in a porcine extracorporeal femorofemoral arteriovenous shunt. Grafts perfused for 3 and 9 minutes with and without heparin were assessed for cellular and microthrombi deposition with scanning electron microscopy. The polypropylene sulfide–PEG–coated ePTFE showed decreased thrombogenicity when used in combination with heparin. Yoneyama et al18 used the phospholipid polymer 2-methacryloyloxyethyl phosphorylcholine and polyurethane to form a 2-mm vascular graft. These grafts were anastomosed to rabbit carotid arteries and harvested at 8 weeks. Neither thrombus nor neointimal formation was observed. Jordan et al19 produced a membrane-mimetic film on the luminal surface of an ePTFE graft by in situ photopolymerization of an acrylate phospholipid. The grafts were assessed in vivo in a baboon femoral arteriovenous shunt model for 1 hour and demonstrated negligible platelet and fibrinogen deposition. Recently, this group of investigators coated ePTFE with an elastin protein polymer that showed decreased thrombogenicity in a primate ex vivo circuit model.20 San Roman et al21 coated Dacron grafts with hydrophilic acrylic polymers, which release salicylic acid, and tested these grafts in an ex vivo canine circuit. This coating appears to decrease the thrombogenicity of Dacron grafts. Yang et al22 coated the luminal surface of ePTFE grafts with a biodegradable elastomer poly(1,8-octanediol citrate) (POC) using a spin-shearing technique. The POC coating did not alter graft compliance and resulted in delayed plasma clotting in vitro compared with controls. Additionally, POC-ePTFE grafts supported attachment and proliferation of porcine endothelium-like cells (isolated from blood) in vitro; within 10 days, cells became confluent on POC-ePTFE grafts, whereas they were only in random clusters and patches on ePTFE controls. Although these and other polymeric applications to vascular grafts are preliminary, they certainly have potential to improve long-term vascular bypass graft outcomes.
| Protein and Drug Modifications |
|---|
|
|
|---|
Another anticoagulant protein is thrombomodulin, an endothelial cell membrane–bound protein that complexes with thrombin and activates the protein C anticoagulation pathway. Tseng et al24 immobilized both thrombomodulin and heparin on the surface of membrane mimetic films and demonstrated in a continuous-flow circuit that thrombin activity was significantly inhibited. The investigators found that combining thrombomodulin with heparin resulted in an additive effect compared with heparin alone or thrombomodulin alone. This material demonstrates great potential to improve the thrombogenic profile of blood-contacting devices.
Hirudin, a direct thrombin antagonist, is yet another anticoagulant under investigation. Wyers et al25 bound hirudin to Dacron using an albumin base coat. Treated Dacron patches were implanted into thoracic canine aortas and exposed to nonheparinized blood for 2 hours and showed less gross thrombus and less pseudointimal development compared with control patches. Hirudin also has been bound to a carboxylated polyurethane with albumin to decrease polyurethane thrombogenicity. This material was shown to retain antithrombin properties in vitro; however, no in vivo studies have been conducted to date.26 Hirudin, in combination with iloprost, an inhibitor of platelet aggregation and promoter of vasodilation, has been used to modify ePTFE grafts. Heise et al27 used these 2 drugs, along with PEG, as the delivery mechanism to coat 4-mm ePTFE grafts. As PEG is hydrolyzed, there is a slow, continuous drug release over several weeks to months. These investigators found that the PEG-hirudin/iloprost group maintained nearly equivalent flow rates at 6 weeks compared with baseline, whereas the 2 control groups, untreated ePTFE and grafts coated with polylactide polymer alone, had markedly reduced flow. Furthermore, although neointimal hyperplasia developed at the distal anastomosis in these grafts, the PEG-hirudin/iloprost grafts were noted to have less restenosis.
An antithrombotic agent, tissue plasminogen activator, which degrades fibrin complexes, has been used to treat graft thrombosis. Greco et al28 implanted 1-mm tissue plasminogen activator–iloprost–modified ePTFE grafts into rat aortas and found that 4 of 10 control grafts were patent at 1 week and 9 of 10 treated grafts were patent. Fibroblast growth factor–modified ePTFE grafts have been studied in animal models. Greisler et al29 pretreated ePTFE grafts with 125I-labeled fibroblast growth factor-1/fibrin glue mixtures with heparin through pressure perfusion. When used as aortoiliac or thoracoabdominal bypass grafts in a canine model, the authors found that fibroblast growth factor-1/fibrin glue promoted greater endothelialization compared with control groups.29,30
Another protein that has been applied to vascular grafts and studied in vivo is P15. This cell-binding peptide was discovered as the cell adhesion domain within type I collagen. The rationale for coating graft surfaces with P15 was that it would promote endothelial cell adhesion and result in a more biomimetic graft. These grafts were evaluated in a sheep model in which 2 P15-coated ePTFE grafts and 1 ePTFE graft were implanted as arteriovenous shunts into each of 2 sheep. At the end of the study, the P15-coated grafts showed a higher degree of endothelialization and significantly less neointimal hyperplasia at the venous anastomosis.31
Sirolimus- and paclitaxel-eluting stents have been used successfully in coronary interventions and have been shown to reduce the incidence of restenosis. Cagiannos et al32 coated 6-mm ePTFE grafts with sirolimus for 1 cm at each end of the graft and used a porcine iliac artery bypass model to examine efficacy. The authors found a statistically significant reduction in percent of cross-sectional narrowing in the sirolimus group compared with untreated and adhesive-treated ePTFE grafts (28.5% and 16.2%; P=0.007). Lee et al33 examined paclitaxel-coated ePTFE grafts using a porcine model for arteriovenous grafting. Although the hemodynamics are not the same as an arterial bypass graft, the authors found that the mean percent of luminal stenosis in the paclitaxel-coated ePTFE graft was 10.4% compared with 60.5% in the control ePTFE graft (P<0.05).
| Endothelial Cell Seeding |
|---|
|
|
|---|
Two-stage seeding techniques involve procuring vein or artery from the patient, harvesting the endothelial cells, and expanding them in culture before seeding them onto the graft. Several groups have shown success with 2-stage seeding in animal models and in clinical trials. Magometschnigg et al34 harvested venous endothelial cells, seeded them onto ePTFE, and used the grafts for crural reconstruction. They showed a 30-day graft patency of 92% compared with 53% in the control ePTFE group. Deutsch et al35 reported their results with this procedure for infrainguinal bypasses in a 2-phase study. Phase I involved 27 endothelial cell–seeded ePTFE grafts and 17 unseeded ePTFE grafts. The 9-year patency rate was significantly better for the experimental group (65% versus 16%; P=0.002). Phase II did not have a control arm, but of the 86 endothelial cell–lined ePTFE grafts, 68% were patent at 5 years. Laube et al36 reported this technique for coronary artery revascularization using 4-mm ePTFE seeded with venous endothelial cells. Of 21 coronary grafts, 91% remained patent after a mean follow-up time of 28 months. Although 2-stage procedures clearly have impressive results, there are also significant limitations. Because of the several-week cell culture delay, these grafts cannot be used in an emergency situation. The processes of culture and seeding are labor intensive and costly and carry the risk of infection. Additionally, endothelial cell growth requires several growth factors that may cause unwanted effects after the grafts are implanted.37
In contrast to 2-stage procedures, 1-stage procedures can be completed during 1 surgical procedure. One-stage procedures involve isolating endothelial cells, usually from venipuncture. Herring et al38 first successfully introduced the concept of single-stage seeding in 1978 using a canine model. Subsequently, their group, along with others, reported clinical results in which venous endothelial cells were seeded on preclotted grafts, but results were mostly disappointing.39–41 A randomized multicenter prospective trial compared seeded femoropopliteal grafts with saphenous vein grafts.40 The 30-month patency rates were 38% and 92%, respectively (P=0.006). Possible reasons for failure included low endothelial cell seeding density, lack of endothelial cell attachment, and lower propensity for endothelialization compared with the canine model.
More recently, investigators have used endothelial progenitor cells (EPCs) isolated from peripheral blood or bone marrow to seed prosthetic vascular grafts. In 1999, Fujita et al42 seeded 4-mm Dacron grafts with preclotted autologous bone marrow blood and implanted them into the canine carotid circulation. At 4 weeks, treated grafts exhibited 80% coverage with endothelium-like cells, whereas control grafts had no coverage. Others have seeded Dacron or ePTFE grafts with bone marrow–derived CD34+ EPCs and have demonstrated 40% to 92% graft endothelialization at 4 weeks in canine and rabbit models.43,44 The challenges of using EPCs include cell culture and the associated time, cost, and infection risks that are inherent to that process. Furthermore, whether these cells will function as true endogenous endothelial cells instead of taking on the characteristics of fibroblasts, smooth muscle cells, or myofibroblasts remains to be determined. An exciting alternative approach that deserves mention is the approach used by Rotmans et al.45 This group coated ePTFE grafts with anti-CD34 antibodies to capture circulating EPCs to the grafts in vivo. Implanted grafts demonstrated enhanced endothelialization at 72 hours; however, intimal hyperplasia still developed at the distal anastomosis. Refinements in this technique may help to recruit EPCs to prosthetic grafts to enhance endogenous endothelialization.
Other sources for seeding have included microvascular endothelial cells derived from fatty tissue such as subcutaneous fat or omentum. The advantages of these cells include ease of obtaining them in large quantities and the lack of required cell culture. However, results with seeding of these cells also have been disappointing to date.46,47 Overall, endothelial cell seeding of prosthetic grafts remains a very attractive approach. The problems of cell attachment and obtaining a sufficient number of cells to seed the grafts remain challenges that must be overcome.
| NO Modifications |
|---|
|
|
|---|
|
Diazeniumdiolates are NO donors formed by the reaction of secondary amine structures with NO under high pressures. The major advantages of diazeniumdiolates are that they are stable solids, that the half-life of NO release can be tailored to the need, and that NO release is easily triggered in an aqueous environment.49 Smith et al50 were the first to use the combination of a diazeniumdiolate and a polymer (polyethyleneamine) incorporated into vascular grafts. These grafts produced NO for several weeks in vitro. Using a baboon arteriovenous shunt model, these investigators demonstrated that after 1 hour, NO-releasing grafts showed significantly less platelet deposition compared with untreated grafts. Unfortunately, the coating process altered graft architecture, and the resultant compliance mismatch may have created long-term biocompatibility issues. To address the issues of compliance mismatch, Pulfer et al51 incorporated polymeric diazeniumdiolate polyethyleneamine/NO microspheres into the pores of an ePTFE graft in an attempt to release NO but not alter the compliance of the graft. They showed that these grafts retained their physical properties and released NO in vitro for >150 hours.
Although these studies provided efficient methods for delivering NO, it was subsequently shown that some of the diazeniumdiolate polymers leach out of the polymer matrixes and form measurable levels of nitrosamines, a well-known class of carcinogens.52 To address this issue, Batchelor et al53 reported the preparation of a more lipophilic, discrete diazeniumdiolate species that is resistant to leaching. Polyurethane vascular grafts (5 mm, Vectra) were coated with this new diazeniumdiolate dispersed in a polyvinyl chloride matrix. Using a sheep arteriovenous shunt model, these investigators reported that the NO-coated grafts were patent at 21 days and had a mean luminal thrombus-free surface area of 95%, whereas the control grafts had all occluded.
Recently, NO-producing polyurethanes were created by covalently binding diazeniumdiolates to a polyurethane backbone.54,55 This approach eliminates the need for reaction additives, thus avoiding leaching of undesired byproducts. Jun et al54 showed that NO production by these polyurethane films occurred for
2 months under physiological conditions in vitro and that mechanical properties of the material were suitable for vascular graft applications. Furthermore, platelet adhesion was greatly diminished, vascular smooth muscle cell growth was inhibited, and endothelial cell growth was stimulated across the polyurethane graft. To date, there are no published in vivo studies using these grafts.
The second class of NO donors that have been used in developing NO-releasing polymers is S-nitrosothiols. S-nitrosothiols serve as a reservoir and transporter of NO within biological systems.56 Examples of endogenous S-nitrosothiols include S-nitrosoalbumin, S-nitrosoglutathione, and S-nitrosocysteine (Figure 3), which are present in circulating blood and within cells.57 Several polymers have been developed by covalently linking S-nitrosothiols to a polymer to prevent leaching of reaction byproducts. Bohl and West58 created NO-releasing hydrogels that showed decreased smooth muscle cell proliferation and decreased platelet adhesion on collagen-coated slides. They speculated that these materials could be used to coat vascular grafts to provide local, sustained NO delivery.
|
An exciting property of S-nitrosothiols is their ability to release NO when exposed to visible light, which provides an opportunity to control the timing of NO release. To test this property, Frost et al59 anchored the NO donor S-nitroso-N-acetylpenicillamine to fumed silica particles, which were then sandwiched between layers of silicone rubber and polyurethane films. They demonstrated that S-nitroso-N-acetylpenicillamine released NO in the presence of light but not when exposed to a physiological solution (ie, blood). They concluded that this represented an external on/off trigger for NO release and that this material has the potential to be made into intravascular devices that are biocompatible.
The main limitation to using polymers loaded with either S-nitrosothiols or diazeniumdiolates is the finite reservoir of NO that exists within these materials.60 New materials are being developed that rely on S-nitrosothiols and/or nitrates that are already circulating in blood. Gappa-Fahlenkamp et al61 immobilized L-cysteine on Dacron and polyurethane surfaces. This yields a free thiol group that can undergo NO exchange reactions to release NO. The potential advantage of this type of material is the essentially unlimited source of NO that could be generated locally when in contact with circulating blood.
| Challenges for Clinical Development |
|---|
|
|
|---|
Other questions that must be considered and for which there are currently no set precedents include how much drug should be incorporated into the graft and how long it should be available in vivo. Furthermore, an additional question that must be addressed is the adverse effects that could result from the drug-releasing nature of these grafts. If the drug were covalently bonded to the grafts, little systemic effects would be anticipated because the drug would not be released into the circulation. If the drug were not covalently attached to the backbone of the graft, side effects of the drug would need to be considered. With respect to adverse events, drug-eluting stents may serve as a precedent. Despite early enthusiasm for the decreased restenosis rates of both the sirolimus and paclitaxel drug-eluting stents, the stents have recently been shown to have increased rates of late in-stent thrombosis.62,63 The drug-eluting stents inhibit endothelial cell proliferation; therefore, they remain thrombogenic.64 FDA approval was based on short-term (<1 year) trials; however, most late in-stent thromboses occurred >1 year after implantation. This finding underscores the importance of extensive preclinical evaluation.
| Conclusions |
|---|
|
|
|---|
| Acknowledgments |
|---|
None.
| References |
|---|
|
|
|---|
1. Friedman SG. History of Vascular Surgery. 2nd ed. Malden, Mass: Futura; 2005.
2. Kannan RY, Salacinski HJ, Butler PE, Hamilton G, Seifalian AM. Current status of prosthetic bypass grafts: a review. J Biomed Mater Res B Appl Biomater. 2005; 74B: 570–581.[Medline] [Order article via Infotrieve]
3. Campbell CD, Brooks DH, Webster MW, Bahnson HT. Use of expanded microporous polytetrafluoroethylene for limb salvage: a preliminary report. Surgery. 1976; 79: 485–491.[Medline] [Order article via Infotrieve]
4. Abbott WM, Green RM, Matsumoto T, Wheeler JR, Miller N, Veith FJ, Suggs WD, Hollier L, Money S, Garrett HE, Moore WS, Dean RH, Cronenwett JL, Carter S, Greenhalgh RM, ODonnell TF. Prosthetic above-knee femoropopliteal bypass grafting: results of a multicenter randomized prospective trial. J Vasc Surg. 1997; 25: 19–28.[CrossRef][Medline] [Order article via Infotrieve]
5. Post S, Kraus T, Muller-Reinartz U, Weiss C, Kortmann H, Quentmeier A, Winkler M, Husfeldt KJ. Dacron vs polytetrafluoroethylene grafts for femoropopliteal bypass: a prospective randomised multicentre trial. Eur J Vasc Endovasc Surg. 2001; 22: 226–231.[CrossRef][Medline] [Order article via Infotrieve]
6. Hobson RW, Odonnell JA, Jamil Z, Mehta K. Below-knee bypass for limb salvage: comparison of autogenous saphenous-vein, polytetrafluoroethylene, and composite Dacron-autogenous vein grafts. Arch Surg. 1980; 115: 833–837.
7. Seifalian AM, Salacinski HJ, Tiwari A, Edwards A, Bowald S, Hamilton G. In vivo biostability of a poly(carbonate-urea)urethane graft. Biomaterials. 2003; 24: 2549–2557.[CrossRef][Medline] [Order article via Infotrieve]
8. Abbott WM, Megerman J, Hasson JE, Litalien G, Warnock DF. Effect of compliance mismatch on vascular graft patency. J Vasc Surg. 1987; 5: 376–382.[CrossRef][Medline] [Order article via Infotrieve]
9. Stonebridge PA, Prescott RJ, Ruckley CV. Randomized trial comparing infrainguinal polytetrafluoroethylene bypass grafting with and without vein interposition cuff at the distal anastomosis. J Vasc Surg. 1997; 26: 543–550.[CrossRef][Medline] [Order article via Infotrieve]
10. Griffiths GD, Nagy J, Black D, Stonebridge PA. Randomized clinical trial of distal anastomotic interposition vein cuff in infrainguinal polytetrafluoroethylene bypass grafting. Br J Surg. 2004; 91: 560–562.[CrossRef][Medline] [Order article via Infotrieve]
11. Taylor RS, Loh A, Mcfarland RJ, Cox M, Chester JF. Improved technique for polytetrafluoroethylene bypass-grafting: long-term results using anastomotic vein patches. Br J Surg. 1992; 79: 348–354.[Medline] [Order article via Infotrieve]
12. Batson RC, Sottiurai VS, Craighead CC. Linton patch angioplasty: an adjunct to distal bypass with polytetrafluoroethylene grafts. Ann Surg. 1984; 199: 684–693.[Medline] [Order article via Infotrieve]
13. Ascer E, Gennaro M, Pollina RM, Ivanov M, Yorkovich WR, Ivanov M, Lorensen E. Complementary distal arteriovenous fistula and deep vein interposition: a five-year experience with a new technique to improve infrapopliteal prosthetic bypass patency. J Vasc Surg. 1996; 24: 134–143.[CrossRef][Medline] [Order article via Infotrieve]
14. Syrek JR, Calligaro KD, Dougherty MJ, Raviola CA, Rua I, DeLaurentis DA. Do distal arteriovenous fistulae improve patency rates of prosthetic infrapopliteal arterial bypasses? Ann Vasc Surg. 1998; 12: 148–152.[CrossRef][Medline] [Order article via Infotrieve]
15. Farrar DJ. Development of a prosthetic coronary artery bypass graft. Heart Surg Forum. 2000; 3: 36–40.[Medline] [Order article via Infotrieve]
16. Kapfer X, Meichelboeck W, Groegler FM. Comparison of carbon-impregnated and standard ePTFE prostheses in extra-anatomical anterior tibial artery bypass: a prospective randomized multicenter study. Eur J Vasc Endovasc Surg. 2006; 32: 155–168.[CrossRef][Medline] [Order article via Infotrieve]
17. Karrer L, Duwe J, Zisch AH, Khabiri E, Cikirikcioglu M, Napoli A, Goessl A, Schaffner T, Hess OM, Carrel T, Kalangos A, Hubbell JA, Walpoth BH. PPS-PEG surface coating to reduce thrombogenicity of small diameter ePTFE vascular grafts. Int J Artif Organs. 2005; 28: 993–1002.[Medline] [Order article via Infotrieve]
18. Yoneyama T, Sugihara K, Ishihara K, Iwasaki Y, Nakabayashi N. The vascular prosthesis without pseudointima prepared by anti thrombogenic phospholipid polymer. Biomaterials. 2002; 23: 1455–1459.[CrossRef][Medline] [Order article via Infotrieve]
19. Jordan SW, Faucher KM, Caves J, Apkarian R, Rele S, Sun X, Hanson S, Chaikof E. Fabrication of a phospholipid membrane-mimetic film on the luminal surface of an ePTFE vascular graft. Biomaterials. 2006; 27: 3473–3481.[CrossRef][Medline] [Order article via Infotrieve]
20. Jordan SW, Haller CA, Sallach RE, Apkarian RP, Hanson SR, Chaikof EL. The effect of a recombinant elastin-mimetic coating of an ePTFE prosthesis on acute thrombogenicity in a baboon arteriovenous shunt. Biomaterials. 2007; 28: 1191–1197.[CrossRef][Medline] [Order article via Infotrieve]
21. San Roman J, Bujan J, Bellon JM, Gallardo A, Escudero MC, Jorge E, de Haro J, Alvarez L, Castillo-Olivares JL. Experimental study of the antithrombogenic behavior of Dacron vascular grafts coated with hydrophilic acrylic copolymers bearing salicylic acid residues. J Biomed Mater Res. 1996; 32: 19–27.[CrossRef][Medline] [Order article via Infotrieve]
22. Yang J, Motlagh D, Allen J, Webb A, Kibbe M, Aalami O, Kapadia M, Carroll T, Ameer GA. Modulating ePTFE vascular graft host response via citric acid-based biodegradable elastomers. Adv Mater. 2006; 18: 1493–1498.[CrossRef]
23. Bosiers M, Deloose K, Verbist J, Schroe H, Lauwers G, Lansink W, Peeters P. Heparin-bonded expanded polytetrafluoroethylene vascular graft for femoropopliteal and femorocrural bypass grafting: 1-year results. J Vasc Surgery. 2006; 43: 313–318.[CrossRef][Medline] [Order article via Infotrieve]
24. Tseng PY, Rele SS, Sun XL, Chaikof EL. Membrane-mimetic films containing thrombomodulin and heparin inhibit tissue factor-induced thrombin generation in a flow model. Biomaterials. 2006; 27: 2637–2650.[CrossRef][Medline] [Order article via Infotrieve]
25. Wyers MC, Phaneuf MD, Rzucidlo EM, Contreras MA, LoGerfo FW, Quist WC. In vivo assessment of a novel Dacron surface with covalently bound recombinant hirudin. Cardiovasc Pathol. 1999; 8: 153–159.[CrossRef][Medline] [Order article via Infotrieve]
26. Phaneuf MD, Szycher M, Berceli SA, Dempsey DJ, Quist WC, LoGerfo FW. Covalent linkage of recombinant hirudin to a novel ionic poly(carbonate) urethane polymer with protein binding sites: Determination of surface antithrombin activity. Artif Organs. 1998; 22: 657–665.[CrossRef][Medline] [Order article via Infotrieve]
27. Heise M, Schmidmaier G, Husmann I, Heidenhain C, Schmidt J, Neuhaus P, Settmacher U. PEG-hirudin/iloprost coating of small diameter ePTFE grafts effectively prevents pseudointima and intimal hyperplasia development. Eur J Vasc Endovasc Surg. 2006; 32: 418–424.[CrossRef][Medline] [Order article via Infotrieve]
28. Greco RS, Kim HC, Donetz AP, Harvey RA. Patency of a small vessel prosthesis bonded to tissue-plasminogen activator and iloprost. Ann Vasc Surg. 1995; 9: 140–145.[CrossRef][Medline] [Order article via Infotrieve]
29. Greisler HP, Cziperle DJ, Kim DU, Garfield JD, Petsikas D, Murchan PM, Applegren EO, Drohan W, Burgess WH. Enhanced endothelialization of expanded polytetrafluoroethylene grafts by fibroblast growth-factor type-1 pretreatment. Surgery. 1992; 112: 244–255.[Medline] [Order article via Infotrieve]
30. Gray JL, Kang SS, Zenni GC, Kim DU, Kim PI, Burgess WH, Drohan W, Winkles JA, Haudenschild CC, Greisler HP. FGF-1 affixation stimulates ePTFE endothelialization without intimal hyperplasia. J Surg Res. 1994; 57: 596–612.[CrossRef][Medline] [Order article via Infotrieve]
31. Li C, Hill A, Imran M. In vitro and in vivo studies of ePTFE vascular grafts treated with P15 peptide. J Biomater Sci Polym Ed. 2005; 16: 875–891.[CrossRef][Medline] [Order article via Infotrieve]
32. Cagiannos C, Abul-Khoudoud OR, DeRijk W, Shell DH, Jennings LK, Tolley EA, Handorf CR, Fabian TC. Rapamycin-coated expanded polytetrafluoroethylene bypass grafts exhibit decreased anastomotic neointimal hyperplasia in a porcine model. J Vasc Surg. 2005; 42: 980–987.[CrossRef][Medline] [Order article via Infotrieve]
33. Lee BH, Nam HY, Kwon T, Kim SJ, Kwon GY, Jeon HJ, Lim HJLWK, Park JS, Ko JY, Kim DJ. Paclitaxel-coated expanded polytetrafluoroethylene haemodialysis grafts inhibit neointimal hyperplasia in porcine model of graft stenosis. Nephrol Dial Transplant. 2006; 21: 2432–2438.
34. Magometschnigg H, Kadletz M, Vodrazka M, Dock W, Grimm M, Grabenwoger M, Minar E, Staudacher M, Fenzl G, Wolner E. Prospective clinical-study with invitro endothelial-cell lining of expanded polytetrafluoroethylene grafts in crural repeat reconstruction. J Vasc Surg. 1992; 15: 527–535.[CrossRef][Medline] [Order article via Infotrieve]
35. Deutsch M, Meinhart J, Fischlein T, Preiss P, Zilla P. Clinical autologous in vitro endothelialization of infrainguinal ePTFE grafts in 100 patients: a 9-year experience. Surgery. 1999; 126: 847–855.[Medline] [Order article via Infotrieve]
36. Laube HR, Duwe J, Rutsch W, Konertz W. Clinical experience with autologous endothelial cell-seeded polytetrafluoroethylene coronary artery bypass grafts. J Thorac Cardiovasc Surg. 2000; 120: 134–141.
37. Heyligers JMM, Arts CHP, Verhagen HJM, de Groot PG, Moll FL. Improving small-diameter vascular grafts: From the application of an endothelial cell lining to the construction of a tissue-engineered blood vessel. Ann Vasc Surg. 2005; 19: 448–456.[CrossRef][Medline] [Order article via Infotrieve]
38. Herring M, Gardner A, Glover J. Single-staged technique for seeding vascular grafts with autogenous endothelium. Surgery. 1978; 84: 498–504.[Medline] [Order article via Infotrieve]
39. Zilla P, Fasol R, Deutsch M, Fischlein T, Minar E, Hammerle A, Krupicka O, Kadletz M. Endothelial-cell seeding of polytetrafluoroethylene vascular grafts in humans: a preliminary report. J Vasc Surg. 1987; 6: 535–541.[CrossRef][Medline] [Order article via Infotrieve]
40. Herring M, Smith J, Dalsing M, Glover J, Compton R, Etchberger K, Zollinger T. Endothelial seeding of polytetrafluoroethylene femoral popliteal bypasses: the failure of low-density seeding to improve patency. J Vasc Surg. 1994; 20: 650–655.[Medline] [Order article via Infotrieve]
41. Jensen N, Lindblad B, Bergqvist D. Endothelial-cell seeded Dacron aortobifurcated grafts: platelet deposition and long-term follow-up. J Cardiovasc Surg. 1994; 35: 425–429.[Medline] [Order article via Infotrieve]
42. Fujita Y, Hong-De Wu M, Ishida A, Shi Q, Walker M, Hammond WP, Sauvage LR. Accelerated healing of Dacron grafts seeded by preclotting with autologous bone marrow blood. Ann Vasc Surg. 1999; 13: 402–412.[CrossRef][Medline] [Order article via Infotrieve]
43. Bhattacharya V, McSweeney PA, Shi Q, Bruno B, Ishida A, Nash R, Storb RF, Sauvage LR, Hammond WP, Wu MHD. Enhanced endothelialization and microvessel formation in polyester grafts seeded with CD34(+) bone marrow cells. Blood. 2000; 95: 581–585.
44. Griese DP, Ehsan A, Melo LG, Kong DL, Zhang LN, Mann MJ, Pratt RE, Mulligan RC, Dzau VJ. Isolation and transplantation of autologous circulating endothelial cells into denuded vessels and prosthetic grafts: implications for cell-based vascular therapy. Circulation. 2003; 108: 2710–2715.
45. Rotmans JI, Heyligers JMM, Verhagen HJM, Velema E, Nagtegaal MM, de Kleijn DPV, de Groot FG, Stroes ESG, Pasterkamp G. In vivo cell seeding with anti-CD34 antibodies successfully accelerates endothelialization but stimulates intimal hyperplasia in porcine arteriovenous expanded polytetrafluoroethylene grafts. Circulation. 2005; 112: 12–18.
46. Meerbaum SO, Sharp WV, Schmidt SP. Lower extremity revascularization with polytertrafluoroethykebe grafts seeded with microvascular endothelial cells. In: Zilla P, Fasol R, Callow P, eds. Applied Cardiovascular Biology. Basel, Switzerland: Karger; 1992: 107–119.
47. Williams SK. Human clinical trials of microvascular endothelial cell sodding. In: Zilla P, Greisler HP, eds. Tissue Engineering of Vascular Prosthetic Grafts. Austin, Tex: R.G. Landes; 1999: 143–147.
48. Kibbe M, Billiar T, Tzeng E. Inducible nitric oxide synthase and vascular injury. Cardiovasc Res. 1999; 43: 650–657.
49. Keefer LK. Nitric oxide (NO)- and nitroxyl (HNO)-generating diazeniumdiolates (NONOates): emerging commercial opportunities. Curr Top Med Chem. 2005; 5: 625–634.[CrossRef][Medline] [Order article via Infotrieve]
50. Smith DJ, Chakravarthy D, Pulfer S, Simmons ML, Hrabie JA, Citro ML, Saavedra JE, Davies KM, Hutsell TC, Mooradian DL, Hanson SR, Keefer LK. Nitric oxide-releasing polymers containing the [N(O)NO](-) group. J Med Chem. 1996; 39: 1148–1156.[CrossRef][Medline] [Order article via Infotrieve]
51. Pulfer SK, Ott D, Smith DJ. Incorporation of nitric oxide-releasing crosslinked polyethyleneimine microspheres into vascular grafts. J Biomed Mater Res. 1997; 37: 182–189.[CrossRef][Medline] [Order article via Infotrieve]
52. Mowery KA, Schoenfisch MH, Saavedra JE, Keefer LK, Meyerhoff ME. Preparation and characterization of hydrophobic polymeric films that are thromboresistant via nitric oxide release. Biomaterials. 2000; 21: 9–21.[CrossRef][Medline] [Order article via Infotrieve]
53. Batchelor MM, Reoma SL, Fleser PS, Nuthakki VK, Callahan RE, Shanley CJ, Politis JK, Elmore J, Merz SI, Meyerhoff ME. More lipophilic dialkyldiamine-based diazeniumdiolates: synthesis, characterization, and application in preparing thromboresistant nitric oxide release polymeric coatings. J Med Chem. 2003; 46: 5153–5161.[CrossRef][Medline] [Order article via Infotrieve]
54. Jun HW, Taite LJ, West JL. Nitric oxide-producing polyurethanes. Biomacromolecules. 2005; 6: 838–844.[CrossRef][Medline] [Order article via Infotrieve]
55. Reynolds MM, Hrabie JA, Oh BK, Politis JK, Citro ML, Keefer LK, Meyerhoff ME. Nitric oxide releasing polyurethanes with covalently linked diazeniumdiolated secondary amines. Biomacromolecules. 2006; 7: 987–994.[CrossRef][Medline] [Order article via Infotrieve]
56. Jourdheuil D, Hallen K, Feelisch M, Grisham MB. Dynamic state of S-nitrosothiols in human plasma and whole blood. Free Radic Biol Med. 2000; 28: 409–417.[CrossRef][Medline] [Order article via Infotrieve]
57. Tyurin VA, Tyurina YY, Liu SX, Bayir H, Hubel CA, Kagan VE. Quantitation of S-nitrosothiols in cells and biological fluids. Methods Enzymol. 2002; 352: 347–360.[Medline] [Order article via Infotrieve]
58. Bohl KS, West JL. Nitric oxide-generating polymers reduce platelet adhesion and smooth muscle cell proliferation. Biomaterials. 2000; 21: 2273–2278.[CrossRef][Medline] [Order article via Infotrieve]
59. Frost MC, Meyerhoff ME. Controlled photoinitiated release of nitric oxide from polymer films containing S-nitroso-N-acetyl-DL-penicillamine derivatized fumed silica filler. J Am Chem Soc. 2004; 126: 1348–1349.[CrossRef][Medline] [Order article via Infotrieve]
60. Reynolds MM, Frost MC, Meyerhoff ME. Nitric oxide-releasing hydrophobic polymers: preparation, characterization, and potential biomedical applications. Free Radic Biol Med. 2004; 37: 926–936.[CrossRef][Medline] [Order article via Infotrieve]
61. Gappa-Fahlenkamp H, Duan X, Lewis RS. Analysis of immobilized L-cysteine on polymers. J Biomed Mater Res A. 2004; 71A: 519–527.[Medline] [Order article via Infotrieve]
62. Spaulding C, Daemen J, Boersma E, Cutlip DE, Serruys PW. A pooled analysis of data comparing sirolimus-eluting stents with bare-metal stents. N Engl J Med. 2007; 356: 989–997.
63. Lagerqvist B, James SK, Stenestrand U, Lindback J, Nilsson T, Wallentin L. Long-term outcomes with drug-eluting stents versus bare-metal stents in Sweden. N Engl J Med. 2007; 356: 1009–1019.
64. Maisel WH. Unanswered questions: drug-eluting stents and the risk of late thrombosis. N Engl J Med. 2007; 356: 981–984.
This article has been cited by other articles:
![]() |
M. Mutsuga, Y. Narita, A. Yamawaki, M. Satake, H. Kaneko, Y. Suematsu, A. Usui, and Y. Ueda A new strategy for prevention of anastomotic stricture using tacrolimus-eluting biodegradable nanofiber. J. Thorac. Cardiovasc. Surg., March 1, 2009; 137(3): 703 - 709. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2008 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |