(Circulation. 2008;117:1460-1473.)
© 2008 American Heart Association, Inc.
Aortic Diseases |
From the Department of Radiology, Division of Angiography and Interventional Radiology, University of Virginia Health System, Charlottesville, Va.
Correspondence to Michael D. Dake, MD, University of Virginia Health System, Department of Radiology, PO Box 800170, Charlottesville, VA 22908. E-mail mdd2n{at}virginia.edu
Key Words: aortic diseases disease management percutaneous transluminal coronary angioplasty revascularization stents
| Introduction |
|---|
The latest additions to the armamentarium to treat dissection have been based on percutaneous interventional techniques. The minimally invasive nature of these techniques makes them an attractive alternative to open surgical intervention; however, the exact role and long-term durability of these procedures remain to be proven. The earliest endovascular therapies were directed toward the complications of aortic dissections and included angioplasty of an obstructed aorta, stenting of obstructed branch vessels, and fenestration of the dissection flap to relieve mesenteric ischemia.6–8 More recently, the advent of the stent graft has led to a novel endovascular approach aimed at treating the inciting lesion of aortic dissection by obliterating the primary intimal tear.9,10 Since the first reports in 1999, which involved treatment of acute and chronic aortic dissections, investigators have extended the application of stent grafts to treat a variety of related pathologies, such as intramural hematoma (IMH), penetrating atherosclerotic ulcer (PAU), and traumatic dissection.11,12
In this report, we begin by reviewing key features of aortic dissection pertinent to endovascular management. We also review the natural history of aortic dissection and conventional medical and surgical management. Then, we introduce the principles and techniques of endovascular management and review the current literature. Specifically, we will discuss the use of stent grafts, fenestration, and branch-vessel stenting in the treatment of aortic dissection and related pathologies.
| Pertinent Features |
|---|
It is important to differentiate the primary or entry intimal tear from the secondary or reentry tear(s). Approximately two thirds of primary tears occur in the ascending aorta, with more than half of these located within the first 2 cm of the ascending aorta. The next most common site of the primary tear is the isthmus of the aorta just beyond the ligamentum arteriosum.14 These regions are presumably subjected to the greatest hemodynamic stress, which makes them more susceptible to injury.15 In either location, these tears are 5 times more likely to be transverse in orientation rather than longitudinal. Other sites of primary tears include the descending thoracic aorta, aortic arch, and abdominal aorta, with multiple primary tears seen in
8% of cases.14
Secondary tears are identified less frequently than primary tears. The secondary tears tend to occur at the ostia of branch vessels, where often circumferential shearing of the intima from the vessel origin has taken place. When present, they allow blood to communicate between the true lumen and the dissecting column.14
Whether a reentry tear is present or not, during the acute phase of a dissection, the local environment of the false lumen is highly thrombogenic owing to the exposed adventitial and medial layers. This, in addition to the morphology of the dissection and underlying hemodynamics, may cause the false channel to become completely or partially thrombosed.16
Commonly, the true lumen, which is bound by intima, will course along the inferomedial aspect of the distal arch and descending aorta. Typically, the celiac trunk, superior mesenteric artery, and right renal arteries arise from the true lumen, and the left renal artery arises from the false lumen; however, variation in this pattern is frequent. Although branch-vessel obstruction may occur, more commonly, flow to branch vessels remains unobstructed with supply from the true lumen, false lumen, or both the true and false lumens.
Related Pathologies
Rapid advances in imaging technology have led to a greater understanding of aortic dissection variants. In particular, 2 entities (IMH and PAU) have been recognized and, when diagnosed in symptomatic patients, are categorized under the umbrella term "acute aortic syndromes" alongside classic type A and type B dissection.17,18
IMH is defined as blood within the intramural space without identification of an intimal disruption. In the purest description, IMH is considered a precursor to dissection, originating from a ruptured vasa vasorum within the medial layer. Subsequent aortic wall weakening or infarction may lead to formation of an intimal tear and result in a classic dissection. However, it is possible that numerous cases of IMH actually represent cases of classic dissection in which the intimal tear is occult on imaging studies.19
PAU is defined as focal ulceration of an atherosclerotic plaque that penetrates a variable depth through the intima and may be associated with intramural blood. The hematoma may propagate along the media and lead to aortic rupture or, less frequently, to an aortic dissection with development of a true and false lumen.20,21
Malperfusion Syndrome
A major complication of aortic dissection is obstruction of flow to the aortic branch vessels. This may involve any aortic branch with 1 or more vascular territories threatened. Critical ischemia of the vascular territory is termed "malperfusion syndrome." It is usually related mechanistically to branch flow obstruction by an intimal flap. Two patterns of branch-vessel involvement are described, dynamic obstruction and static obstruction.22
Dynamic obstruction is a term that characterizes the phenomenon of aortic true-lumen collapse or obliteration. Imaging in the setting of dynamic branch-vessel obstruction shows a paper-thin, crescent-shaped true aortic lumen dwarfed by a larger false lumen. The aortic flap has a convex contour that appears to flatten or efface the true lumen. Consequently, flow to all downstream abdominal branches supplied by the true lumen may be compromised. Further obstruction to a particular branch vessel may occur as the dissection flap prolapses over the branch-vessel ostium. Because of the constantly changing position of the intimal flap, particularly in the acute phase, these obstructions can be total or subtotal with persistent or intermittent features and thus are referred to as dynamic obstructions. A clear understanding of the pathophysiology responsible for dynamic obstruction is not apparent; however, large proximal entry tears are frequently observed.22
In distinction, static obstruction refers to the effects of direct extension of the dissection process into an individual branch. Typically, flap progression into a branch is tolerated. This occurs when the false lumen within the branch vessel develops a distal reentry tear that allows double-barrelled flow to the vascular bed with both true- and false-lumen contribution. In most cases, biluminal perfusion of the branch-vessel tissue bed is sufficient and is not associated with serious ischemia or the risk of necrosis. However, despite the "reentry" branch configuration described, the physical presence of the flap may cause various degrees of static obstruction depending on the artery involved, the adequacy of flow, and the presence of underlying lesions.22
The potential for profound malperfusion exists when the false lumen within the branch does not have a reentry point in the vessel. As a result, the false lumen is a blind cul-de-sac that enlarges and compresses the true lumen. The severity of true lumen obstruction is maximum at the distal margin of false-lumen extension, where a prominent bulbous contour is frequently evident. This "no reentry" branch configuration of static obstruction is often associated with severe ischemia related to the absence of false-lumen flow within the branch and a severely compromised true lumen. Fortunately, static obstruction with no reentry of the false lumen occurs relatively infrequently compared with the more common pattern of false-lumen reentry.22
In addition, it is possible for a combination of dynamic and static obstruction with or without branch-vessel reentry to coexist within an individual patient, which creates an interesting "signature" profile depending on the particular branch vessels affected and mechanism(s) of involvement by the process. As we will see, differentiating between a dynamic or static obstruction has significant implications for endovascular management.9,22
| Classification Systems |
|---|
The convenience and prognostic value of the Stanford system has resulted in its popular use; however, an important feature that is not distinguished in the Stanford system is the location of the primary tear. In a typical type A dissection, the primary tear is located in the ascending aorta, whereas in a retrograde type A dissection, the tear is located in the descending aorta. This detail has a profound influence on the current feasibility of stent-graft management.
Aortic dissections are temporally classified as acute when identified
2 weeks from the onset of symptoms and chronic when identified >2 weeks from the onset of symptoms, with the highest morbidity and mortality occurring during the acute phase. Some investigators further classify the acute phase as early (<24 hours) and late (
24 hours to 2 weeks) from the onset of symptoms.
In addition, type B dissections are often classified as complicated or uncomplicated. Complicated cases involve rupture, lesion progression, impending rupture, refractory hypertension, localized large false aneurysm, continued pain, or malperfusion syndrome and are associated with worse outcomes.23,24
| Natural History and Conventional Management |
|---|
|
|
|---|
Acute Type A Dissection
Although shown to be superior to medical management alone, surgical management is still associated with alarmingly high rates of morbidity and mortality. In a review of 547 type A dissections, IRAD investigators demonstrated a hospital mortality rate of 27% for those patients treated surgically.26 In another IRAD report, surgical repair was associated with in-hospital mortality rates of 10% by 24 hours, 16% by 7 days, and nearly 20% by 14 days.5 Long-term survival for patients treated with surgery who were discharged alive has been shown to be 96% at 1 year and 91% at 3 years.27
Patients with retrograde type A dissection (DeBakey type IIId) represent an important subgroup that comprises 4% to 20% of all type A cases.16,23,28,29 In these individuals, the inciting primary tear is typically positioned in the distal arch, with retrograde extension of the dissection process to the ascending aorta. This poses a dilemma for surgical repair that involves either excision of the entry tear with replacement of both the ascending aorta and aortic arch, which is associated with high morbidity and mortality, or graft replacement of the ascending aorta alone without excision of the primary tear, which leaves the patient at risk of postoperative aortic rupture.16,23,28–30 Regardless, conventional therapy mandates immediate surgical repair in this patient subgroup.
Acute Type B Dissection
Owing to the small risk of aortic rupture and sudden death and the high morbidity and mortality associated with surgical repair of the descending aorta, medical treatment alone is advocated for uncomplicated type B dissection. Surgery is typically reserved for complicated type B dissections. IRAD investigators reviewed 175 patients treated according to this complication-specific approach and identified in-house mortality rates of 11% and 31% for medical and surgical treatment, respectively.5 Patients who undergo surgery also have a high rate of morbidity; in particular, paraplegia has been reported in 1.5% to 19% despite advances in surgical technique.31 IRAD investigators also performed long-term evaluation of outcomes for patients discharged after hospital management of type B dissection. This analysis identified 3-year survival rates of 78% and 83% for patients treated medically and surgically, respectively.32
Because medical therapy alone does not stop flow within the false lumen, 20% to 50% of patients who survive the acute phase develop aneurysmal dilatation of the false lumen within 1 to 5 years after onset.15,33,34 In this regard, the majority of late deaths that occur in patients with type B dissection initially managed by medical therapy are due to rupture, extension of dissection, and perioperative mortality of subsequent aortic or vascular surgeries. In fact, the long-term survival of patients with type B dissection remains worse than that of patients with type A dissection.27,32
Aortic Dissection Variants (IMH and PAU)
Intramural Hematoma
The natural history of IMH is not well understood; however, it accounts for 5% to 20% of acute aortic syndrome cases.13,21,35,36 IRAD investigators demonstrated an association between increased hospital mortality and proximity of IMH to the aortic valve, regardless of medical or surgical treatment.18 A meta-analysis of 11 studies identified an overall mortality of 34% for type A IMH, 24% for those treated surgically and 47% for those treated medically.19 Overall mortality for type B IMH was 14%, with little difference between surgical (15%) and medical (13%) groups. Thus, the treatment paradigm for IMH parallels that of classic aortic dissection, with surgical repair favored for type A IMH and medical management preferred for type B IMH.
Penetrating Atherosclerotic Ulcer
Although PAU is poorly understood, its prognosis is thought to be much poorer than that of classic aortic dissection.37 Coady et al20 reported the risk of aortic rupture in patients with PAU and acute symptoms to be 40% compared with patients with type A or type B dissection, for whom the rupture risks were 7% and 3.6%, respectively. Although no consensus treatment strategy exists, early surgical graft replacement of the aorta has been advocated in symptomatic patients.38,39
Chronic Dissection
Those patients who survive the acute stage of aortic dissection, which is associated with the greatest mortality, by definition have chronic dissection. The 30-day survival rate for this population is high at 90%, independent of whether they were managed medically or surgically.40 Medical therapy is therefore recommended for patients with both type A and type B chronic dissection. Surgery is reserved for those who develop an aneurysm, rupture, peripheral branch-vessel compromise, or other complications.41,42
Malperfusion Syndrome
It is estimated that branch-vessel involvement complicates approximately one third to one half of all aortic dissections.24,43 With control for both age and gender, branch-vessel involvement presents a nearly 3-fold increased likelihood of in-hospital morality in acute type B dissections.44
Primary surgical aortic repair results in successful revascularization in the vast majority of patients with type B dissection; however, the presence of renal or mesenteric ischemia has been correlated with especially high surgical mortality rates of 70% and 87%, respectively.24,43,45,46 For this reason, alternative surgical procedures have been explored to address organ ischemia specifically. In particular, surgical aortic fenestration has been shown to relieve malperfusion syndrome in 93% to 100% of cases, with an in-hospital mortality rate of 0% to 43%.47–49
| Endovascular Management |
|---|
Stent-Graft Technology
The development of thoracic aortic stent grafts has largely followed in the footsteps of abdominal aortic stent-graft technology used primarily to treat abdominal aortic aneurysms. The earliest feasibility and safety studies of thoracic stent grafts were performed in 1992 to treat thoracic aortic aneurysms. Since then, the number of applications of this technology has grown rapidly to include the management of aortic dissection and dissection variants. Although treatment of descending thoracic aneurysm with a stent graft (TAG; WL Gore and Associates, Flagstaff, Ariz) is currently approved by the US Food and Drug Administration, the use of stent grafts to treat other indications such as aortic dissection and its related pathologies remains off-label.
The first-generation stent grafts were primarily homemade devices that married graft materials such as polyester or polytetrafluoroethylene to modified self-expanding stents such as the Gianturco Z stent. Most delivery systems were large (24F to 27F), relatively rigid, and difficult to deploy smoothly and accurately owing to extensive frictional resistance. Significant improvements were made in the second generation of devices, which are largely manufactured commercially. However, as we will demonstrate, numerous design issues remain. Current devices include the TAG (WL Gore and Associates), Talent (Medtronic Inc, Santa Rosa, Calif), TX-2 (Cook Inc, Bloomington, Ind), Relay (Bolton Medical, Inc, Sunrise, Fla), and Valiant (Medtronic Inc).
Principles and Techniques of Endovascular Management
Stent-Graft Management
The rationale of stent-graft management is 2-fold. First, in the acute phase, the use of stent grafts may prevent imminent aortic rupture and relieve dynamic branch-vessel obstruction. Second, stent-graft management may promote thrombosis of the thoracic false lumen and decrease the long-term morbidity associated with patency of the false lumen, including aneurysmal dilatation, late aortic rupture, and late mortality.9,16
Stent-graft treatment is predicated on the ability to cover the primary intimal tear and create a seal to stop the flow of blood entering the false lumen and prevent the transmission of systemic pressure across the major intimal defect (Figures 1 and 2
). If the seal is adequate, cardiac output is redirected into the true lumen and rapid and the false lumen simultaneously decompresses (which relieves dynamic obstruction of branches supplied by a diminutive true lumen). As a result, within seconds, the true lumen diameter typically enlarges, with markedly improved flow. The immediate hemodynamic and morphological alterations may prevent imminent rupture and relieve aortic true-lumen collapse and branch-vessel ischemia.9
|
|
Furthermore, stent-graft management of retrograde type A dissection and type B dissection has been shown to decrease flow in the false lumen and induce false-lumen thrombosis.50,51 Again, this is a critical point, because natural history studies of type B dissection have shown 20% to 50% of patients who receive medical therapy alone and survive the acute phase ultimately develop aneurysmal dilatation of the false lumen within 1 to 5 years.15,33,34 Even if complete thrombosis of the false lumen does not occur, it is likely that partial thrombosis and decreased flow will limit the progression to aneurysmal dilatation.9
To obliterate the primary tear, an adequate seal zone is required. One of the anatomic requirements is a proximal landing zone (relative to the primary tear) of at least 15 to 20 mm. The ideal landing zone should be uniform in shape and free of significant disease; however, this ideal is rarely met, because the position of the primary tear and its proximity to branch vessels usually requires device deployment within a dissected segment. A common dilemma is selection of the "correct" device dimension, because the true lumen is generally crescentric or elliptical in shape and a fraction of the overall transaortic diameter. Most operators base their selection on more than 1 measurement, the most compelling of which is the diameter of the nondissected aorta immediately proximal to the entry tear. In the setting of a classic entry location just distal to the left subclavian artery origin, the segment between the left carotid and left subclavian arteries is used. This is the best estimate of the original size of the involved aorta before dissection. This measurement is oversized by
10% and used to select the stent-graft diameter. The oversizing factor ensures secure anchoring and a tight circumferential seal.
Depending on the type and size of the stent graft, currently available devices will require delivery systems that are 20F to 24F in size. The iliofemoral arteries should be assessed routinely to ensure that an adequate intraluminal diameter to accommodate introduction of the device exists. Access usually involves surgical exposure of the common femoral artery. In the case of small or heavily calcified femoral arteries, surgical exposure of the iliac arteries or aorta with or without placement of a graft conduit may be required. Recently, in select patients, stent-graft procedures have been performed entirely percutaneously, with the puncture sites closed by commercially available suture-mediated access-closure devices.52 The obvious benefits of this approach over surgical exposure are the decreased time to recovery and possible reduced risk of infection, lymphocele, seroma, and postoperative scar.
Fenestration
Fenestration of the intimal flap serves as an alternative endovascular treatment to stent-graft management of aortic dissection (Figure 3). Rather than treating the entry tear, which increases the resistance to false-lumen inflow, fenestration is aimed at artificially creating a distal reentry channel, which decreases the resistance to false-lumen outflow.53 The reentry channel is not dissimilar to spontaneously formed reentry tears that help to balance pressures within the aortic lumens. The equalization of pressure between the true and false lumen alone may relieve dynamic obstruction of the aorta and branch vessels. In addition, strategic positioning of fenestrations may locally redirect a sufficient amount of blood flow to perfuse compromised end organs.8,54–56 Although beneficial in the acute phase, the continued patency of the false lumen potentially predisposes the patient to less salutary long-term outcomes, such as false-lumen dilatation, aneurysm development, and rupture, compared with those after stent-graft placement and thoracic false-lumen thrombosis.16
|
Because stent-graft treatment generally achieves similar results with a greater likelihood of false-lumen thrombosis, stent-graft management is preferred over fenestration as the primary mode of treatment in patients with malperfusion caused by dynamic obstruction. However, fenestration continues to be a valuable option in poor surgical candidates and in cases in which stent-graft treatment is not available or not feasible owing to anatomic constraints. Fenestration may also serve as a valuable adjunctive treatment in cases in which surgical repair or stent-graft repair inadequately addresses the indications for intervention.
Currently, percutaneous balloon fenestration of the flap is most frequently performed distally in the abdominal aorta at the level of the aortic bifurcation, usually to manage unilateral lower-extremity ischemic symptoms. Several approaches to dissection flap fenestration have been described.53–55,57,58 The 2 most commonly detailed techniques involve combined intravascular ultrasound and fluoroscopic guidance or fluoroscopic guidance alone.
When intravascular ultrasound is used, the probe is typically advanced into the larger lumen (usually the false lumen) while a long curved coaxial cannula/needle set, precurved transseptal needle, or reentry catheter with needle (Outback, Cordis, Miami, Fla; Pioneer, Medtronic, Inc) is advanced into the other lumen. The dissection flap is perforated under ultrasound guidance. Subsequently, a guidewire is advanced through the needle into the opposite lumen. Over the wire, the intimal defect is enlarged with an angioplasty balloon (usually 12-mm to 20-mm diameter). After angioplasty of the fenestration, it is essential to reevaluate the patency of all branch vessels for compromise, because the resultant alteration in blood flow is not always predicted accurately. In the case of inadequate flow through the aorta or branch vessels, an additional fenestration or stenting may be required.53
When fluoroscopic guidance is used alone, a balloon catheter is placed in the larger lumen and used as a target when the intimal flap is transgressed. Preprocedural cross-sectional imaging provides valuable guidance by establishing the orientation of the flap at the fenestration target. Otherwise, the procedure is very similar to the intravascular ultrasound–guided procedure described above.55 Regardless of the technique used, successful and safe fenestration requires a high level of interventional skill and intimate familiarity with a wide range of devices, including intravascular ultrasound, angioplasty balloons, stents, snares, occlusion balloons, and transseptal needles.
Stent Placement
The use of uncovered stents in the aorta and side branches may improve flow in aortic dissection (Figure 4). The most common indications include (1) inadequate relief of dynamic obstruction after surgery, stent-graft treatment, or fenestration and (2) static obstruction of abdominal aortic branch vessels, which is typically unaffected by proximal aortic stent-graft treatment and fenestration.9,53,55
|
In most cases of symptomatic static branch-vessel compromise, attempts are made to restore more normal flow dynamics by placing an uncovered stent via the aortic true lumen into the true lumen of the involved branch. The stent usually covers the length of the dissected branch segment. On stent deployment, the flap is displaced, and a cylindrical true lumen is established. In some instances, stents may be placed from the false lumen of the aorta to the true lumen of the branch vessel to restore flow.53,55
Uncovered aortic stents may also be placed to prop open a severely collapsed true lumen and allow flow to reach distal branches supplied exclusively by the aortic true lumen.53,55 Similarly, uncovered aortic stents may be a useful adjunct to dissection fenestration. After fenestration, the intimal flap may take on a position that continues to obstruct the aorta and branches fed by the true lumen. In such a case, a stent may be placed across the fenestration to prop open the fenestration and tack down a distal tag of intimal flap.56 A variety of both balloon-expandable and self-expanding stents have been used in the aorta and in renal, iliac, superior mesenteric, and brachiocephalic arteries.53,55,56
| Outcomes of Endovascular Management |
|---|
Several investigators have reported hybrid surgical and endovascular treatment of acute type A aortic dissections with the primary tear located in the ascending aorta. The combined technique uses traditional open surgical methods to identify and resect the primary tear, followed by stent-graft deployment into the descending aorta under direct, open visualization. The rationale is based on the potential long-term benefits of false-lumen thrombosis and aortic remodeling associated with endograft treatment of the descending aorta; the benefits are thought to include lower rates of aneurysm formation, expansion, and rupture.61–63 Finally, in poor surgical candidates, primary endoluminal stent-graft management of type A dissection with stent-graft placement across an entry tear in the ascending aorta has been described.64–67
Acute Type B Dissection
Meta-Analysis
A meta-analysis by Eggebrecht et al68 reviewed all published data between January 1999 and May 2004 involving stent-graft placement for aortic dissection (96% type B and 4% retrograde type A) with an entry tear located in the descending aorta (Table). Only studies with a minimum of 3 patients were included. Cases of antegrade surgical ("open") stent-graft placement via the aortic arch were excluded. A total of 39 studies comprising 609 patients were reviewed.
|
Procedural success was obtained in 98.2% of cases. Neurological complications were seen in 2.9%, with 1.9% representing periprocedural stroke and 0.8% representing paraplegia. Retrograde extension of dissection into the ascending aorta was seen in 1.9% of cases. The total in-house surgical conversion rate was 2.3%. The overall in-hospital mortality rate was 5.2%, and the 30-day mortality rate was 5.3%. Interestingly, this 30-day mortality was compared with IRAD data, which detailed 30-day mortality rates for type B dissection managed medically and surgically to be 10.7% and 31.4%, respectively.5 A composite Kaplan-Meier table including all 3 interventions showed worse survival outcome in the surgical group and similar outcomes in the medical and stent-graft treatment groups.
A mean follow-up time of 19.5 months was available for 561 patients. During follow-up, false-lumen thrombosis was observed in 75.5%, late surgical conversion in 2.5%, and a supplemental endovascular stent-graft procedure in 4.6%. The total reintervention rate was 11.9%. Late death occurred in 2.8%.
European Collaborators on Stent Graft Techniques for Thoracic Aortic Aneurysm and Dissection Repair and the United Kingdom Thoracic Endograft Registries
The compilation of cases from the European Collaborators on Stent Graft Techniques for Thoracic Aortic Aneurysm and Dissection Repair (EUROSTAR) and the United Kingdom (UK) Thoracic Endograft registries represents the next-largest reported series to date69 (Table). This series included 131 patients treated for acute dissection, with all cases having an entry tear identified in the descending aorta (81% type B, 5% retrograde type A, and 14% classification unavailable). At the time of the procedure, 43% were asymptomatic, whereas 57% of cases were complicated by symptoms of rupture, aortic expansion, or side-branch occlusion. In the series, 42% of patients were at high risk for conventional open surgery, with an American Society of Anesthesiologists score of 3 or 4.
Procedural success was achieved in 89% of cases, with the remaining 11% having either incomplete coverage of the entry tear, persistent flow with only partial thrombosis of the thoracic false lumen, no expansion of the true lumen, or endoleaks. Periprocedural endoleak was seen in 6.7%. Neurological complications were seen in 2.3%, with 1.5% representing stroke and 0.8% representing paraplegia. The overall 30-day mortality rate was 8.4%.
One-year data were available for 67 patients, with late intervention required in 1.5% and late endoleak identified in 1.5%. Late death occurred in 1.5% of patients, with a 1-year cumulative survival rate of 90%.
Contemporary Studies
A list of more contemporary studies is provided in the Table.68–76 These studies report similar short-term and mid-term results compared with the meta-analysis by Eggebrecht et al68 and the EUROSTAR/UK registries.69
This collection of observational data demonstrates that stent-graft treatment of type B dissection is feasible, with high technical success rates and acceptable short-term outcomes. Around the world at aortic-disease–referral centers, stent-graft treatment is currently performed for complicated type B dissection in which conventional surgical repair continues to be associated with strikingly high morbidity and mortality. Endograft management can be used to treat most complications and avoids the major trauma of open surgery. Notably, reports of endograft management detail markedly lower rates of paraplegia than surgical management of the descending aorta.31
In contrast, the role of stent-graft management for uncomplicated type B dissection is uncertain. Currently, no conclusive data exist that demonstrate superiority of either stent-graft or medical management. To address this issue, the INSTEAD (INvestigation of STEnt grafts in patients with type B Aortic Dissection) trial is under way.
INSTEAD Study
The INSTEAD study is a prospective, randomized, controlled clinical trial that compares stent-graft treatment of uncomplicated chronic type B dissection (occurring 2 to 52 weeks before randomization) with optimal medical therapy alone.77 The study design is based on a retrospective comparison of 80 patients with type B dissection treated with a stent graft compared with 80 patients treated with medical management alone. A Kaplan-Meier life-table analysis showed that at 12 months, the risk of death in the stent-graft–treated group was 5.1% versus a historic mortality rate of 27.5% with conventional therapy. The INSTEAD study enrolled 136 patients at 11 European clinical sites. The primary end point is all-cause mortality at 1 year after enrollment. It is anticipated that the forthcoming results will help shed light on the role of stent grafts in chronic uncomplicated type B dissections.
Dissection Variants
Penetrating Atherosclerotic Ulcer
The application of stent grafts to aortic dissection variants has focused more on PAU than other lesions.11,37,78–80 PAU is typically a focal process that involves the descending thoracic aorta, which makes it anatomically well suited for endograft treatment.
Eggebrecht et al80 performed a systematic review of 54 patients accumulated from 13 studies. This report detailed complete sealing of the ulcer in 94% of cases, neurological complication in 6%, and an in-hospital mortality rate of 5%.
A more recent study by Demers et al11 evaluated 26 patients with type B PAU (half of whom were deemed inoperable) and found a primary success rate of 92% and perioperative mortality of 12%, with no cases of paraplegia. Estimated survival rates were 81% at 1 year and 65% at 5 years.
Recently, Eggebrecht et al79 presented 22 cases of PAU treated by stent-graft placement with a median follow-up of 27 months. A technical success rate of 96% was achieved. One patient (5%) had a minor stroke, and no other in-hospital complications or deaths occurred. The overall survival rates were 100% at 30 days, 100% at 1 year, 82.5% at 2 years, and 62% at 5 years.
These data suggest high technical success rates with favorable acute and midterm mortality rates compared with the natural history of the disease. Although only limited amounts of data exist, stent-graft treatment of PAU appears to be a viable treatment option.
Intramural Hematoma
The treatment of IMH with stent grafts has been reported; however, the precise definition of its current role in the repair of IMH is still in evolution.9,12,19,78
Malperfusion Syndrome
Endovascular management of malperfusion may involve 1 or more techniques, including stent-graft placement, fenestration, and side-branch stenting. Together, they represent a powerful and promising set of techniques. Yet, only a small collection of data on the outcomes of these techniques exists.
Dake et al9 investigated stent-graft management of malperfusion syndrome in aortic dissection. In this series from the Stanford University School of Medicine, 11 of 19 patients had malperfusion syndrome, which represented 38 of 114 infradiaphragmatic vascular beds examined. Malperfusion syndrome was caused by dynamic obstruction in 22 arteries, dynamic and static obstruction in 15 arteries, and static obstruction only in 1 artery. After stent-graft treatment of the primary tear in the thoracic aorta, all cases of dynamic obstruction resolved, whereas only 6 of 15 arteries with combined dynamic and static obstruction resolved. The remaining vascular beds were treated successfully with placement of uncovered stents in side branches.9 In a smaller series, Duebener et al81 treated 5 cases of malperfusion syndrome in aortic dissection using stent grafts; of these, 2 patients had persistent visceral ischemia that required surgical fenestration and ultimately had fatal courses.
The 2 largest series using fenestration and stenting (without stent-graft management) were from the University of Michigan and Stanford. In the Michigan study, 24 cases of malperfusion syndrome in aortic dissection (13 type A, 8 type B, and 3 atypical) were treated with fenestration and/or stent placement. A total of 92% of the vascular beds were reperfused successfully, with a 30-day mortality rate of 25%.53 At least 3 of the 6 deaths were related to irreversible ischemia of abdominal organs. In addition, 2 of the 18 surviving patients died of complications of false-lumen enlargement during the follow-up period. In a follow-up study, the same group reported decreased mortality rates after the incorporation of percutaneous revascularization procedures into a "surgical delay" strategy for selected patients with type A dissection and symptomatic ischemic manifestations; however, this management strategy has been debated.82,83
In the Stanford series, 40 patients with malperfusion syndrome (10 type A, 30 type B) were treated with fenestration and/or stent placement. Of these, 93% were reperfused successfully, with a 30-day mortality rate of 25%, which was often related to irreversible ischemia of intra-abdominal organs present before the procedure. Of the early survivors, 5 died during follow-up, with at least 1 having false-lumen rupture.55
The results of endovascular fenestration and stenting in these series compare favorably with conventional management with surgical fenestration, which has been associated with mortality rates between 0% and 43%.47–49 However, neither surgical nor percutaneous fenestration addresses the issue of late complications related to false-lumen expansion. Stent-graft management has a distinct advantage in both treating the malperfused vascular bed and treating the inciting intimal tear and promoting thrombosis and remodeling of the false lumen. Additionally, stent-graft treatment can be performed faster and generally requires less interventional experience than does fenestration and side-branch stenting. For vascular territories that do not respond to endograft management, usually involving static obstructions, Dake et al9 demonstrated the indispensable utility of placing uncovered stents to restore flow.
Chronic Dissection
The evolution of an acute dissection to a chronic dissection involves progressive thickening of the intimal flap due to fibrosis. Additionally, chronic dissections often possess multiple natural fenestrations in the septum. These characteristics have brought into question the ability of stent-graft treatment to induce false-lumen thrombosis and remodeling.
Nienaber et al10 prospectively evaluated stent-graft management in 12 patients with chronic type B dissection and compared the results with 12 matched surgical controls. Proximal entry closure and complete thrombosis of the false lumen at 3 months were achieved in all patients. Stent-graft treatment resulted in no mortality or morbidity, whereas surgical treatment resulted in 4 deaths and 5 adverse events. At 3 months, complete thrombosis of the false lumen was achieved in all patients, with clear evidence of true-lumen expansion and false-lumen shrinkage. A similar study by Kato et al84 also demonstrated favorable aortic remodeling. However, in a second study by the same group that evaluated chronic and acute dissections, complete false-lumen thrombosis was documented in all patients, but complete obliteration of the false lumen was found in only 38.5% of the chronic dissection patients versus 70% of the acute dissection patients (mean follow-up of 27 months).85
| Complications |
|---|
Two potentially devastating complications of both open surgical repair and endovascular stent-graft repair are stroke and paraplegia. The cause of stroke associated with stent-graft treatment is thought to be primarily atheroembolic, with a stroke rate comparing favorably to that of open surgical repair of the descending aorta.86 The cause of paraplegia after surgical repair of the thoracic aorta is thought to be multifactorial, influenced by the duration and severity of the ischemic insult, the neuronal metabolic rate during the ischemic period, postoperative hypotension, and subsequent reperfusion injury.31,86 Despite improvements in conventional surgical techniques for the descending aorta, paralysis occurs in 1.5% to 19% of cases.31 The apparent lower risk of paraplegia after endograft treatment, reported to be between 0.8% and 3.6%, is likely attributable to the avoidance of aortic cross clamping and subsequent reperfusion required during open surgery.31 Although great concern existed over stent-graft coverage of intercostal arteries, it appears to be better tolerated than predicted initially. Nonetheless, extensive coverage of intercostal arteries is discouraged because of the heightened risk of spinal cord ischemia.31
Retrograde dissection into the ascending aorta is a serious complication associated with the formation of new intimal tears, aneurysm formation, aortic rupture, tamponade, and death.76,87 It typically requires immediate open surgical repair.88 This complication is identified in
2% of cases during the in-hospital period and 2% of cases during follow-up.68 The causative mechanism implicated by most reports is a stent-graft–induced intimal injury due to limited flexibility along the outer edges of current devices.76,87,88 The maintenance of meticulous interventional technique is another important consideration that may decrease the frequency of injury to the intima by guidewire and catheter manipulations. It is likely that the rate of retrograde dissection will improve as manufacturers create tapered and precurved stent grafts with less rigid, pliable ends.
Stent-graft collapse has been reported in up to 3% of cases in 1 series87 (Figure 5). In a separate series, 6 cases of stent-graft collapse were reported after treatment of 5 traumatic aortic injuries and 1 dissection with the TAG device.89 Both reports attributed all cases to oversizing of the stent graft to the aorta. These cases may reflect the inherent difficulty of sizing stent grafts in patients with extremely small true lumens. Furthermore, these cases may reflect the tendency of more experienced operators to push the anatomic limits of stent-graft therapy by treating dissections for which no ideal commercially available device exists. Nonetheless, it is safe to assume that improvements in stent-graft technology will likely offset this phenomenon and ultimately lead to a decrease in the rate of early and late collapse.
|
Overall, periprocedural endoleaks are encountered in
6.7% of cases, and late endoleak is identified in 1.5%.69 Most type I endoleaks are identified in the periprocedural period, whereas most type II endoleaks are identified during follow-up.69 Extrapolating from stent-graft treatment of abdominal aortic aneurysm, type I endoleaks require immediate repair, usually involving placement of an additional stent extension, and type II endoleaks may be monitored.90 The consequence of endoleaks in stent-graft management of aortic dissection is not well known, although such leaks are associated with persistent flow in the false lumen and may predispose to complications of false-lumen expansion or rupture.31
Other reported complications include stent-graft migration, stent-graft torsion, aortoesophageal fistula, and mobile thrombus within the stent-graft lumen.31,73,87 These complications, in addition to retrograde dissection, stent-graft collapse, endoleak, aneurysm formation, and late rupture, which are described above, underscore the importance of close and lifelong imaging follow-up of these patients.
| Treatment Considerations |
|---|
Several studies have demonstrated higher in-hospital mortality rates and major complication rates for patients treated with stent-graft placement for type B dissection in the acute phase compared with the chronic phase.68,73,88 Kato et al88 speculated that worse outcomes in the acute phase after endograft treatment may reflect sequelae caused by unfavorable morphological alterations of a fragile dissection membrane compared with manipulation of a more fibrotic and stable flap in the chronic phase. This led some investigators to favor treatment in the chronic phase. However, Eggebrecht et al73 suggested that a more important factor was the preprocedural clinical status of the patient. They documented that patients with poor clinical health status (American Society of Anesthesiologists class >3) had a reduced life expectancy after stent-graft therapy compared with patients with moderate comorbidities (American Society of Anesthesiologists class
3).73 This may explain the poorer outcome in the acute-phase treatment groups described in several case series, which are often composed of sicker patients. In particular, both age and renal failure have been identified as independent determinants of postintervention mortality.73–75 Interestingly, the importance of preprocedural clinical status was supported by a study by Chen et al75 in which an opposite trend was identified, with a lower 30-day mortality rate after stent-graft management in the acute-phase treatment group compared with the rate for the chronic-phase treatment group; the authors attributed the difference to the fact that patients in the acute-treatment group of that study were healthier.
The potential long-term benefit of stent-graft management of aortic dissection is based on the induction of false-lumen thrombosis and remodeling. The overall rate of complete thrombosis of the thoracic false lumen is
75%.50,51,73 Identification of the determinants of false-lumen thrombosis and strategies to improve false-lumen thrombosis rates have been topics of interest. Kusagawa et al50 and Resch et al51 have both shown that thoracic false-lumen thrombosis rates and shrinkage of the false lumen occur more often in patients treated in the acute phase than in the chronic phase.
Whereas false-lumen thrombosis is consistently seen at the level of the implanted stent graft, thrombosis distal to the implant, particularly in the abdominal aorta, is less common.50 This may relate to the to-and-fro movement of the intimal flap along uncovered segments of the aorta, with retrograde flow through distal flap fenestrations or secondary distal tears. This has implications for the selection of device length. Most investigators use stent grafts longer than needed to cover the primary tear alone, usually 15 to 20 cm. The additional distal coverage provides a more normal anatomic configuration by placing the distal device margin within the midthoracic segment, where the aorta usually becomes less curved. Further distal extension may expedite the rate of false-lumen thrombosis; however, extension of the stent graft into the distal third of the thoracic aorta may increase the risk of spinal cord ischemia and paraplegia. In light of this, some investigators have advocated the placement of bare stents into the distal thoracic aorta to provide structural stability without risking occlusion of intercostal arteries.91
Finally, it is becoming clear that adoption of this relatively new treatment modality involves an inherent learning curve that affects not only physicians who perform endovascular procedures but also equipment designers and manufacturers. Eggebrecht et al,68 in their meta-analysis of stent-graft management of type B aortic dissection, reported improved technical success rates in studies published between 2002 and 2004 compared with those published between 1999 and 2001. Additionally, fewer overall complications occurred in centers that performed a total number of cases above the median number of cases performed by all centers in the analysis. These findings likely reflect improvements in device technology and increasing operator experience associated with stent-graft management of aortic dissections over the last decade.68
| Conclusions |
|---|
With rapidly accumulating data, a greater understanding has been reached of the treatment considerations and complications of endovascular techniques for aortic dissection. It is certain that advances in graft design will address these issues and lead to greater applicability of stent grafts in the management of dissection; however, randomized, controlled clinical trials are needed to demonstrate any advantage of endovascular treatment over conventional management.
| Acknowledgments |
|---|
Disclosures
Dr Dake is a member of the W.L. Gore and Associates Endovascular Medical Advisory Board and participated as a clinical investigator in thoracic endograft trials sponsored by W.L. Gore, Cook Inc, and Medtronic Inc.
| References |
|---|
2. Abbott OA. Clinical experiences with application of polythene cellophane upon aneurysms of thoracic vessels. J Thorac Surg. 1949; 18: 435.[Medline] [Order article via Infotrieve]
3. DeBakey ME, Cooley DA, Creech O Jr. Surgical considerations of dissecting aneurysm of the aorta. Ann Surg. 1955; 142: 586–610.[Medline] [Order article via Infotrieve]
4. Wheat MW Jr, Palmer RF, Bartley TD, Seelman RC. Treatment of dissecting aneurysms of the aorta without surgery. J Thorac Cardiovasc Surg. 1965; 50: 364–373.[Medline] [Order article via Infotrieve]
5. Hagan PG, Nienaber CA, Isselbacher EM, Bruckman D, Karavite DJ, Russman PL, Evangelista A, Fattori R, Suzuki T, Oh JK, Moore AG, Malouf JF, Pape LA, Gaca C, Sechtem U, Lenferink S, Deutsch HJ, Diedrichs H, Marcos y Robles J, Llovet A, Gilon D, Das SK, Armstrong WF, Deeb GM, Eagle KA. The International Registry of Acute Aortic Dissection (IRAD): new insights into an old disease. JAMA. 2000; 283: 897–903.
6. Shimshak TM, Giorgi LV, Hartzler GO. Successful percutaneous transluminal angioplasty of an obstructed abdominal aorta secondary to a chronic dissection. Am J Cardiol. 1988; 61: 486–487.[CrossRef][Medline] [Order article via Infotrieve]
7. Walker PJ, Dake MD, Mitchell RS, Miller DC. The use of endovascular techniques for the treatment of complications of aortic dissection. J Vasc Surg. 1993; 18: 1042–1051.[CrossRef][Medline] [Order article via Infotrieve]
8. Williams DM, Brothers TE, Messina LM. Relief of mesenteric ischemia in type III aortic dissection with percutaneous fenestration of the aortic septum. Radiology. 1990; 174: 450–452.
9. Dake MD, Kato N, Mitchell RS, Semba CP, Razavi MK, Shimono T, Hirano T, Takeda K, Yada I, Miller DC. Endovascular stent-graft placement for the treatment of acute aortic dissection. N Engl J Med. 1999; 340: 1546–1552.
10. Nienaber CA, Fattori R, Lund G, Dieckmann C, Wolf W, von Kodolitsch Y, Nicolas V, Pierangeli A. Nonsurgical reconstruction of thoracic aortic dissection by stent-graft placement. N Engl J Med. 1999; 340: 1539–1545.
11. Demers P, Miller DC, Mitchell RS, Kee ST, Chagonjian L, Dake MD. Stent-graft repair of penetrating atherosclerotic ulcers in the descending thoracic aorta: mid-term results. Ann Thorac Surg. 2004; 77: 81–86.
12. Kaya A, Heijmen RH, Overtoom TT, Vos JA, Morshuis WJ, Schepens MA. Thoracic stent grafting for acute aortic pathology. Ann Thorac Surg. 2006; 82: 560–566.
13. Nienaber CA, Eagle KA. Aortic dissection: new frontiers in diagnosis and management: part I: from etiology to diagnostic strategies. Circulation. 2003; 108: 628–635.
14. Hirst AE Jr, Johns VJ Jr, Kime SW Jr. Dissecting aneurysms of the aorta: a review of 505 cases. Medicine (Baltimore). 1958; 37: 217–279.[Medline] [Order article via Infotrieve]
15. Wheat MW Jr. Acute dissection of the aorta. Cardiovasc Clin. 1987; 17: 241–262.[Medline] [Order article via Infotrieve]
16. Erbel R, Oelert H, Meyer J, Puth M, Mohr-Katoly S, Hausmann D, Daniel W, Maffei S, Caruso A, Covino FE; the European Cooperative Study Group on Echocardiography. Effect of medical and surgical therapy on aortic dissection evaluated by transesophageal echocardiography: implications for prognosis and therapy. Circulation. 1993; 87: 1604–1615.
17. Vilacosta I, Roman JA. Acute aortic syndrome. Heart. 2001; 85: 365–368.
18. Tsai TT, Nienaber CA, Eagle KA. Acute aortic syndromes. Circulation. 2005; 112: 3802–3813.
19. Maraj R, Rerkpattanapipat P, Jacobs LE, Makornwattana P, Kotler MN. Meta-analysis of 143 reported cases of aortic intramural hematoma. Am J Cardiol. 2000; 86: 664–668.[CrossRef][Medline] [Order article via Infotrieve]
20. Coady MA, Rizzo JA, Hammomd GL, Pierce JG, Kopf GS, Elefteriades JA. Penetrating ulcer of the thoracic aorta: what is it? How do we recognize it? How do we manage it? J Vasc Surg. 1998; 27: 1006–1016.[CrossRef][Medline] [Order article via Infotrieve]
21. Ganaha F, Miller DC, Sugimoto K, Do YS, Minamiguchi H, Saito H, Mitchell RS, Dake MD. Prognosis of aortic intramural hematoma with and without penetrating atherosclerotic ulcer: a clinical and radiological analysis. Circulation. 2002; 106: 342–348.
22. Williams DM, Lee DT, Hamilton BH, Marx MV, Narasimham DL, Kazanjian SN, Prince MR, Andrews JC, Cho KJ, Deeb GM. The dissected aorta: part III: anatomy and radiologic diagnosis of branch-vessel compromise. Radiology. 1997; 203: 37–44.
23. Miller DC, Mitchell RS, Oyer PE, Stinson EB, Jamieson SW, Shumway NE. Independent determinants of operative mortality for patients with aortic dissection. Circulation. 1984; 70 (pt 2): I-153–I-164.[Medline] [Order article via Infotrieve]
24. Cambria RP, Brewster DC, Gertler JG, Moncure AC, Gusberg R, Tilson MD, Darling RC, Hammond G, Mergerman J, Abbott WM. Vascular complications associated with spontaneous aortic dissection. J Vasc Surg. 1988; 7: 199–209.[CrossRef][Medline] [Order article via Infotrieve]
25. Erbel R, Alfonso F, Boileau OD, Dirsch O, Eber B, Haverich A, Rakowski H, Struyven J, Radegran K, Sechtem U, Taylor J, Zollikofer C, Klein WW, Mulder B, Providencia LA; Task Force on Aortic Dissection, European Society of Cardiology. Diagnosis and management of aortic dissection. Eur Heart J. 2001; 22: 1642–1681.
26. Mehta RH, Suzuki T, Hagan PG, Bossone E, Gilon D, Llovet A, Maroto LC, Cooper JV, Smith DE, Armstrong WF, Nienaber CA, Eagle KA; International Registry of Acute Aortic Dissection (IRAD) Investigators. Predicting death in patients with acute type A aortic dissection. Circulation. 2002; 105: 200–206.
27. Tsai TT, Evangelista A, Nienaber CA; Trimarchi S, Sechtem U, Fattori R, Myrmel T, Pape L, Cooper JV, Smith DE, Fang J, Isselbacher E, Eagle KA; International Registry of Acute Aortic Dissection (IRAD). Long-term survival in patients presenting with type A acute aortic dissection: insights from the International Registry of Acute Aortic Dissection (IRAD). Circulation. 2006; 114 (suppl I): I-350–I-356.[Medline] [Order article via Infotrieve]
28. Kazui T, Tamiya Y, Tanaka T. Extended aortic replacement for acute type A dissection with the tear in the descending aorta. J Thorac Cardiovasc Surg. 1996; 112: 973–978.
29. Landsman SL, Galla JD, Schor JS. Subtypes of acute aortic dissection. J Card Surg. 1994; 9: 729–733.[CrossRef][Medline] [Order article via Infotrieve]
30. Von Segesser LK, Killer I, Ziswiler M, Linka A, Ritter M, Jenni R, Baumann PC, Turina MI. Dissection of the descending thoracic aorta extending into the ascending aorta: a therapeutic challenge. J Thorac Cardiovasc Surg. 1994; 108: 755–761.
31. Piffaretti G, Tozzi M, Lomazzi C, Rivolta N, Caronno R, Castelli P. Complications after endovascular stent-grafting of thoracic aortic diseases. J Cardiothorac Surg. 2006; 1: 26.[CrossRef][Medline] [Order article via Infotrieve]
32. Tsai TT, Fattori R, Trimarchi S, Isselbacher E, Myrmel T, Evangelista A, Hutchison S, Sechtem U, Cooper JV, Smith DE, Pape L, Froehlich J, Raghupathy A, Januzzi JL, Eagle KA, Nienaber CA; International Registry of Acute Aortic Dissection. Long-term survival in patients presenting with type B acute aortic dissection: insights from the International Registry of Acute Aortic Dissection. Circulation. 2006; 114: 2226–2231.
33. Doroghazi RM, Slater EE, DeSanctis RW, Buckley MJ, Austen WG, Rosenthal S. Long-term survival of patients with treated aortic dissection. J Am Coll Cardiol. 1984; 3: 1026–1034.[Abstract]
34. Richter GM, Allenberg JR, Schumacher H, Hansmann J, Vahl C, Hagl S. Aortic dissection: when operative treatment, when endoluminal therapy? [in German] Radiologe. 2001; 41: 660–667.[CrossRef][Medline] [Order article via Infotrieve]
35. Evangelista A, Mukherjee D, Mehta RH, OGara PT, Fattori R, Cooper JV, Smith DE, Oh JK, Hutchison S, Sechtem U, Isselbacher EM, Nienaber CA, Pape LA, Eagle KA; International Registry of Aortic Dissection (IRAD) Investigators. Acute intramural hematoma of the aorta: a mystery in evolution. Circulation. 2005; 111: 1063–1070.
36. Shimizu H, Yoshino H, Udagawa H, Watanuki A, Yano K, Ide H, Sudo K, Ishikawa K. Prognosis of aortic intramural hemorrhage compared with classic aortic dissection. Am J Cardiol. 2000; 85: 792–795.[CrossRef][Medline] [Order article via Infotrieve]
37. Pauls S, Orend KH, Sunder-Plassmann L, Kick J, Schelzig H. Endovascular repair of symptomatic penetrating atherosclerotic ulcer of the thoracic aorta. Eur J Vasc Endovasc Surg. 2007; 34: 66–73.[CrossRef][Medline] [Order article via Infotrieve]
38. Stanson AW, Kazmier FJ, Hollier LH, Edwards WD, Pairolero PC, Sheedy PF, Joyce JW, Johnson MC. Penetrating atherosclerotic ulcers of the thoracic aorta: natural history and clinicopathologic correlations. Ann Vasc Surg. 1986; 1: 15–23.[Medline] [Order article via Infotrieve]
39. Tittle SL, Lynch RJ, Cole PE, Singh HS, Rizzo JA, Kopf GS, Elefteriades JA. Midterm follow-up of penetrating ulcer and intramural hematoma of the aorta. J Thorac Cardiovasc Surg. 2002; 123: 1051–1059.
40. Crawford ES, Svensson LG, Coselli JS, Safi HJ, Hess KR. Aortic dissection and dissecting aortic aneurysm. Ann Surg. 1988; 208: 254–273.[Medline] [Order article via Infotrieve]
41. Kato M, Bai H, Sato K, Kawamoto S, Kaneko M, Ueda T, Kishi D, Ohnishi K. Determining surgical indications for acute type B dissection based on enlargement of aortic diameter during the chronic phase. Circulation. 1995; 92 (suppl II): II-107–II-112.[Medline] [Order article via Infotrieve]
42. Juvonen T, Ergin MA, Galla JD, Lansman SL, McCullough JN, Nguyen K, Bodian CA, Ehrlich MP, Spielvogel D, Klein JJ, Griepp RB. Risk factors for rupture of chronic type B dissections. J Thorac Cardiovasc Surg. 1999; 117: 776–786.
43. Fann JI, Sarris GE, Mitchell RS, Shumway NE, Stinson EB, Oyer PE, Miller DC. Treatment of patients with aortic dissection presenting with peripheral vascular complications. Ann Surg. 1990; 212: 705–713.[Medline] [Order article via Infotrieve]
44. Suzuki T, Mehta RH, Ince H, Nagai R, Sakomura Y, Weber F, Sumiyoshi T, Bossone E, Trimarchi S, Cooper JV, Smith DE, Isselbacher EM, Eagle KA, Nienaber CA; International Registry of Aortic Dissection. Clinical profiles and outcomes of acute type B aortic dissection in the current era: lessons from the International Registry of Aortic Dissection (IRAD). Circulation. 2003; 108 (suppl II): II-312–II-317.[Medline] [Order article via Infotrieve]
45. Borst HG, Laas J, Heinemann M. Type A dissection: diagnosis and management of malperfusion phenomena. Semin Thorac Cardiovasc Surg. 1991; 3: 238–241.[Medline] [Order article via Infotrieve]
46. Heinemann MK, Buehner B, Schaefers HJ, Jurmann MJ, Laas J, Borst HG. Malperfusion of the thoracoabdominal vasculature in aortic dissection. J Card Surg. 1994; 9: 748–755.[Medline] [Order article via Infotrieve]
47. Elefteriades JA, Hartleroad J, Gusberg RJ, Salazar AM, Black HR, Kopf GS, Baldwin JC, Hammond GL. Long-term experience with descending aortic dissection: the complication-specific approach. Ann Thorac Surg. 1992; 53: 11–21.[Abstract]
48. Panneton JM, The SH, Cherry KJ Jr, Hofer JM, Gloviczki P, Andrews JC, Bower TC, Pairolero PC, Hallett JW Jr. Aortic fenestration for acute or chronic aortic dissection: an uncommon but effective procedure. J Vasc Surg. 2000; 32: 711–721.[CrossRef][Medline] [Order article via Infotrieve]
49. Harms J, Hess U, Cavallaro A, Naundorf M, Maurer PC. The abdominal aortic fenestration procedure in acute thoraco-abdominal aortic dissection with aortic branch artery ischemia. J Cardiovasc Surg. 1998; 39: 273–280.[Medline] [Order article via Infotrieve]
50. Kusagawa H, Shimono T, Ishida M, Suzuki T, Yasuda F, Yuasa U, Onoda K, Yada I, Hirano T, Takeda K, Kato N. Changes in false lumen after transluminal stent-graft placement in aortic dissections: six years experience. Circulation. 2005; 111: 2951–2957.
51. Resch TA, Delle M, Falkenberg M, Ivancev K, Konrad P, Larzon T, Lönn L, Malina M, Nyman R, Sonesson B, Thelin S. Remodeling of the thoracic aorta after stent grafting of type B dissection: a Swedish multicenter study. J Cardiovasc Surg. 2006; 47: 503–508.[Medline] [Order article via Infotrieve]
52. Quinn SF, Duke DJ, Baldwin SS, Bascom TH, Ruff SJ, Swangard RJ, DeHaas DR, Padgett RC, Bergin PJ, Lau S. Percutaneous placement of a low-profile stent-graft device for aortic dissections. J Vasc Interv Radiol. 2002; 13: 791–798.[CrossRef][Medline] [Order article via Infotrieve]
53. Williams DM, Lee DT, Hamilton BH, Marx MV, Narasimham DL, Kazanjian SN, Prince MR, Andrews JC, Cho KJ, Deeb GM. The dissected aorta: percutaneous treatment of ischemic complications: principles and results. J Vasc Interv Radiol. 1997; 8: 605–625.[Medline] [Order article via Infotrieve]
54. Slonim SM, Nyman U, Semba CP, Miller DC, Mitchell RS, Dake MD. Aortic dissection: percutaneous management of ischemic complications with endovascular stents and balloon fenestration. J Vasc Surg. 1996; 23: 241–251.[CrossRef][Medline] [Order article via Infotrieve]
55. Slonim SM, Miller DC, Mitchell RS, Semba CP, Razavi MK, Dake MD. Percutaneous balloon fenestration and stenting for life-threatening ischemic complications in patients with acute aortic dissection. J Thorac Cardiovasc Surg. 1999; 117: 1118–1127.
56. Vedantham S, Picus D, Sanchez LA, Braverman A, Moon MR, Sundt T III, Kim P, Rubin B, Sicard GA. Percutaneous management of ischemic complications in patients with type-B dissection. J Vasc Interv Radiol. 2003; 14: 181–193.[Medline] [Order article via Infotrieve]
57. Trerotola SO. Use of a stone basket as a target during fenestration of aortic dissection. J Vasc Intervent Radiol. 1996; 7: 687–690.[Medline] [Order article via Infotrieve]
58. Kronzon I, Tunick PA, Riles T, Rosen R. Transesophageal echocardiography in intimal flap fenestration. J Am Soc Echocardiogr. 2001; 14: 934–936.[CrossRef][Medline] [Order article via Infotrieve]
59. Kato N, Shimono T, Hirano T, Ishida M, Yada I, Takeda K. Transluminal placement of endovascular stent grafts for the treatment of type A aortic dissection with an entry tear in the descending thoracic aorta. J Vasc Surg. 2001; 34: 1023–1028.[CrossRef][Medline] [Order article via Infotrieve]
60. Shimono T, Kato N, Tokui T, Onoda K, Hirano T, Takeda K, Yuasa H, Yada I. Endovascular stent-graft repair for acute type A aortic dissection with an intimal tear in the descending aorta. J Thorac Cardiovasc Surg. 1998; 116: 171–173.
61. Shah A, Coulon P, DeChaumaray T, Rosario R, Khanoyan P, Boukhris M, Tshiombo G, Gay J, Bergeron P. Novel technique: staged hybrid surgical and endovascular treatment of acute type A aortic dissection with aortic arch involvement. J Cardiovasc Surg (Torino). 2006; 47: 497–502.[Medline] [Order article via Infotrieve]
62. Fleck T, Hutschala D, Czerny M, Ehrlich MP, Kasimir MT, Cejna M, Wolner E, Grabenwoger M. Combined surgical and endovascular treatment of acute aortic dissection type A: preliminary results. Ann Thorac Surg. 2002; 74: 761–766.
63. Dietrich EB, Ghazoul M, Wheatley GH, Alpern J, Rodriguez-Lopez J, Ramaiah V, Williams J. Surgical correction of ascending type A thoracic aortic dissection: simultaneous endoluminal exclusion of the arch and distal aorta. J Endovasc Ther. 2005; 12: 660–666.[CrossRef][Medline] [Order article via Infotrieve]
64. Ihnken K, Sze D, Dake MD, Fleischmann D, Van der Starre P, Robbins R. Successful treatment of a Stanford type A dissection by percutaneous placement of a covered stent graft in the ascending aorta. J Thorac Cardiovasc Surg. 2004; 127: 1808–1810.
65. Zhang H, Li M, Jin W, Wang Z. Endoluminal and surgical treatment for the management of Stanford type A aortic dissection. Eur J Cardiothorac Surg. 2004; 26: 857–859.
66. Zimpfer D, Czerny M, Kettenbach J, Schoder M, Wolner E, Lammer J, Grimm M. Treatment of acute type A dissection by percutaneous endovascular stent-graft placement. Ann Thorac Surg. 2006; 82: 747–749.
67. Xu SD, Huang FJ, Yang JF, Du JH, Wang XY, Zhang ZG. Treatment of type A dissection with stent-graft. Asian Cardiovasc Thorac Ann. 2007; 15: 80–81.
68. Eggebrecht H, Nienaber CA, Neuhauser M, Baumgart D, Kische S, Schmermund A, Herold U, Rehders TC, Jakob HG, Erbel R. Endovascular stent-graft placement in aortic dissection: a meta-analysis. Eur Heart J. 2006; 27: 489–498.
69. Leurs LJ, Bell R, Degrieck MD, Thomas S, Hobo R, Lundbom J; EUROSTAR; UK Thoracic Endograft Registry Collaborators. Endovascular treatment of thoracic aortic diseases: combined experience from the EUROSTAR and United Kingdom thoracic endograft registries. J Vasc Surg. 2004; 40: 670–680.[CrossRef][Medline] [Order article via Infotrieve]
70. Bortone AS, DeCillis E, DAgostino D. Endovascular treatment of thoracic aortic disease: four years of experience. Circulation. 2004; 110 (suppl II): II-262–II-267.[Medline] [Order article via Infotrieve]
71. Dialetto G, Covino FE, Scognamiglio G, Manduca S, Della Corte A, Giannolo B, Scardone M, Cotrufo M. Treatment of type B aortic dissection: endoluminal repair or conventional medical therapy? Eur J Cardiothorac Surg. 2005; 27: 826–830.
72. Nathanson DR, Rodriguez-Lopez JA, Ramaiah VG, Williams J, Olsen DM, Wheatley GH, Diethrich EB. Endoluminal stent-graft stabilization for thoracic aortic dissection. J Endovasc Ther. 2005; 12: 354–359.[CrossRef][Medline] [Order article via Infotrieve]
73. Eggebrecht H, Herold U, Kuhnt O, Schmermund A, Bartel T, Martini S, Lind A, Naber CK, Kienbaum P, Kühl H, Peters J, Jakob H, Erbel R, Baumgart D. Endovascular stent-graft treatment of aortic dissection: determinants of post-interventional outcome. Eur Heart J. 2005; 26: 489–497.
74. Xu SD, Huang FJ, Yang JF, Li ZZ, Wang XY, Zhang ZG, Du JH. Endovascular repair of acute type B aortic dissection: early and mid-term results. J Vasc Surg. 2006; 43: 1090–1095.[CrossRef][Medline] [Order article via Infotrieve]
75. Chen S, Yei F, Zhou L, Luo J, Zhang J, Shan S, Tian N, Kwan TW. Endovascular stent-grafts treatment in acute aortic dissection (type B): clinical outcomes during early, late, or chronic phases. Catheter Cardiovasc Interv. 2006; 68: 319–325.[CrossRef][Medline] [Order article via Infotrieve]
76. Schoder M, Czerny M, Cejna M, Rand T, Stadler A, Sodeck GH, Gottardi R, Loewe C, Lammer J. Endovascular repair of acute type B aortic dissection: long-term follow-up of true and false lumen diameter changes. Ann Thorac Surg. 2007; 83: 1059–1066.
77. Nienaber CA, Zannetti S, Barbieri B, Kische S, Schareck W, Rehders TC; INSTEAD Study Collaborators. Investigation of stent grafts in patients with type B aortic dissection: design of the INSTEAD trial: a prospective, multicenter, European randomized trial. Am Heart J. 2005; 149: 592–599.[CrossRef][Medline] [Order article via Infotrieve]
78. Sundt TM. Intramural hematoma and penetrating atherosclerotic ulcer of the aorta. Ann Thorac Surg. 2007; 83: S835–S841.
79. Eggebrecht H, Herold U, Schmermund A, Lind AY, Kuhnt O, Martini S, Kühl H, Kienbaum P, Peters J, Jakob H, Erbel R, Baumgart D. Endovascular stent-graft treatment of penetrating aortic ulcers: results over a median follow-up of 27 months. Am Heart J. 2006; 151: 530–536.[CrossRef][Medline] [Order article via Infotrieve]
80. Eggebrecht H, Baumgart D, Schmermund A, Herold U, Hunold P, Jakob H, Erbel R. Penetrating atherosclerotic ulcer of the aorta: treatment by endovascular stent-graft placement. Curr Opin Cardiol. 2003; 18: 431–435.[CrossRef][Medline] [Order article via Infotrieve]
81. Duebener LF, Lorenzen P, Richardt G, Misfeld M, Nötzold A, Hartmann F, Sievers HH, Geist V. Emergency endovascular stent-grafting for life-threatening acute type B aortic dissections. Ann Thorac Surg. 2004; 78: 1261–1266.
82. Deeb GM, Williams DM, Bolling SF, Quint LE, Monaghan H, Sievers J, Karavite D, Shea M. Surgical delay for acute type A dissection with malperfusion. Ann Thorac Surg. 1997; 64: 1669–1677.
83. Geirsson A, Szeto WY, Pochettino A, McGarvey ML, Keane MG, Woo YJ, Augoustides JG, Bavaria JE. Significance of malperfusion syndromes prior to contemporary surgical repair for acute type A dissection: outcomes and need for additional revascularizations. Eur J Cardiothorac Surg. 2007; 32: 255–262.
84. Kato N, Hirano T, Shimono T, Ishida M, Takano K, Nishide Y, Kawaguchi T, Yada I, Takeda K. Treatment of chronic aortic dissection by transluminal endovascular stent-graft placement: preliminary results. J Vasc Interv Radiol. 2001; 12: 835–840.[Medline] [Order article via Infotrieve]
85. Shimono T, Kato N, Yasuda F, Suzuki T, Yuasa U, Onoda K, Hirano T, Takeda K, Yada I. Transluminal stent-graft placement for the treatments of acute onset and chronic aortic dissections. Circulation. 2002; 106 (suppl I): I-241–I-247.[Medline] [Order article via Infotrieve]
86. Khoynezhad A, Donayre CE, Bui H, Kopchok GE, Walot I, White RA. Risk factors of neurological deficit after thoracic aortic endografting. Ann Thorac Surg. 2007; 83: S882–S889.
87. Won JY, Suh S, Ko H, Lee KH, Shim WH, Chang BC, Choi DH, Park SJ, Lee do Y. Problems encountered during and after stent-graft treatment of aortic dissection. J Vasc Interv Radiol. 2006; 17: 271–281.[CrossRef][Medline] [Order article via Infotrieve]
88. Kato N, Shimono T, Hirano T, Suzuki T, Ishida M, Sakuma H, Yada I, Takeda K. Midterm results of stent-graft repair of acute and chronic aortic dissection with descending tear: the complication-specific approach. J Thorac Cardiovasc Surg. 2002; 124: 306–312.
89. Muhs BE, Balm R, White GH, Verhagen HJ. Anatomic factors associated with acute endograft collapse after Gore TAG treatment of thoracic aortic dissection or traumatic rupture. J Vasc Surg. 2007; 45: 655–661.[CrossRef][Medline] [Order article via Infotrieve]
90. Chuter TA, Faruqi RM, Sawhney R, Reilly LM, Kerlan RB, Canto CJ, Lukaszewicz GC, Laberge JM, Wilson MW, Gordon RL, Wall SD, Rapp J, Messina LM. Endoleak after endovascular repair of abdominal aortic aneurysm. J Vasc Surg. 2001; 34: 98–105.[CrossRef][Medline] [Order article via Infotrieve]
91. Mossop PJ, McLachlan CS, Amukotuwa SA, Nixon IK. Staged endovascular treatment for complicated type B aortic dissection. Nat Clin Pract Cardiovasc Med. 2005; 2: 316–321.[CrossRef][Medline] [Order article via Infotrieve]