(Circulation. 2007;116:561-571.)
© 2007 American Heart Association, Inc.
Interventional Cardiac Electrophysiology |
From the Department of Medicine/Division of Cardiology, Northwestern University Feinberg School of Medicine, Chicago, Ill.
Correspondence to Alan Kadish, MD, Cardiology Division, Northwestern Memorial Hospital, 201 E Huron, Suite 10-240, Chicago, IL 60611. E-mail a-kadish{at}northwestern.edu
Key Words: death, sudden defibrillators, implantable pacemakers
| Introduction |
|---|
| Permanent Pacing for Sudden Cardiac Death Prevention |
|---|
| Implantable Defibrillators for Sudden Cardiac Death Prevention |
|---|
Patients in whom ICD implantation may be considered can generally be categorized into 1 of several broad groups: (1) patients who have been resuscitated from sudden cardiac arrest or have previously manifested potentially life-threatening ventricular arrhythmias not related to a transient or correctable cause; (2) patients with moderate to severe structural heart disease, including ischemic, nonischemic, and hypertrophic cardiomyopathy; and (3) those with miscellaneous cardiac disorders, genetic and otherwise, for which the risk of arrhythmic death is considered high. We will briefly describe the data supporting the use of ICDs in each of these groups, but caution must be exercised in extrapolating these results to individual patients, because these studies enrolled highly selected individuals who may not be entirely representative of the population at large. Furthermore, underrepresentation of certain cohorts, such as women, minorities, and patients with chronic kidney disease, underscores the challenges in generalizing these findings.
Secondary Prevention
Survivors of sudden cardiac death due to arrhythmia carry a high risk of recurrence.15 Attempts to significantly reduce this risk with antiarrhythmic drugs yielded disappointing results, except perhaps in the case of amiodarone.16 The recognition of the limitations of antiarrhythmic drugs for secondary prevention was paralleled by the development of smaller, transvenous ICDs with tiered therapies, bradycardia pacing, and success rates of >95% in terminating VT and VF. Thus, the stage was set for randomized trials comparing the 2 treatments. The Antiarrhythmics Versus Implantable Defibrillators (AVID) study provided the first controlled evidence of the superiority of ICDs compared with antiarrhythmic drug therapy. The study randomized patients with resuscitated VF, sustained VT with syncope, or sustained VT with ejection fraction
40% and evidence of hemodynamic compromise to receive either an ICD or antiarrhythmic therapy, predominantly amiodarone. Overall mortality in the ICD arm was decreased by 39% and 31% at 1 and 3 years, respectively.17 Subgroup analyses of the AVID data demonstrated that the survival benefit of ICDs was confined to the group with ejection fractions between 20% and 34%.18 Two additional trials, the Canadian Implantable Defibrillator Study (CIDS) and The Cardiac Arrest Study Hamburg (CASH), both had similar study designs as AVID but showed only nonsignificant trends toward reduced mortality in the ICD arms.19,20 A meta-analysis of all 3 trials showed a significant risk reduction of 28% in total mortality and a 51% reduction in the risk of arrhythmic death.21,22 Interestingly, whereas the AVID study excluded patients with a transient or correctable cause for their VT or VF from randomization, a registry of all screened patients was maintained that yielded important results. Adjustment for multiple variables revealed that patients with transient or correctable causes, such as new ischemia or infarction, proarrhythmic drug reaction, electrolyte imbalance, or other causes, actually had worse survival than those considered to have primary VT or VF, thus highlighting the inherent challenges in identifying patients at low risk of recurrent arrhythmias and the limited long-term success in treating presumed transient causes.23
Primary Prevention
Primary Prevention in Ischemic Cardiomyopathy
Although survivors of sudden death represent a high-risk group for recurrent events, at-risk patients who have not yet suffered an event are in far greater abundance. Furthermore, given the low survival rate of out-of-hospital cardiac arrest, the potential for impact in terms of absolute number of lives saved is greater overall for primary prevention. Perhaps the only challenge that has rivaled that of the development of the ICD has been in determining who benefits most from its use. Although it now seems obvious that patients with significant underlying structural heart disease stand to gain the most from prophylactic ICD implantation, the medical community was divided over the potential impact of ICDs in the early years after its inception. In light of the fact that pump failure deaths increase and the proportion of deaths attributable to arrhythmias decreases as heart failure worsens, some argued that the impact of ICDs on mortality would be limited in a heart failure population. However, the rationale for using post–myocardial infarction (MI) patients with low ejection fraction as the first group to study is supported by the pathophysiological and observational data. Prior MI may result in both reduced ejection fraction and abnormalities of conduction and refractoriness that serve as the substrate for ventricular tachyarrhythmias.24,25 It had been estimated that half of the deaths in the post-MI population are due to VT and VF.26,27 The relationship between left ventricular systolic dysfunction and deaths due to progressive heart failure and ventricular arrhythmias in patients who have had an MI is well established. Studies dating back to the advent of cardiac imaging were the first to observe the association between reduced ejection fraction and outcome, with the majority of studies concluding that an ejection fraction of
40% serves as the threshold for identifying high-risk individuals.28–33
Over the last decade, 6 major trials have evaluated the role of ICDs in the primary prevention of sudden cardiac death in post-MI patients with reduced ejection fraction.34–39 A summary of these studies is presented in Table 1. These studies have not only established the utility of ICDs for this purpose but have also demonstrated that the risk of sudden cardiac death in a high-risk cohort treated with current medical therapies is approximately 5% per year and is considerably less than previously estimated. More recently, the Multicenter Automatic Defibrillator Implantation Trial (MADIT) II trial required only prior MI >1 month before enrollment and an ejection fraction
30% for study inclusion, given the uncertainties surrounding the long-term predictive value of invasive electrophysiological testing and the near-ubiquitous finding of ambient arrhythmias in the ischemic cardiomyopathy cohort. This study of 1232 patients compared overall mortality in ICD-treated patients versus those treated with conventional medical therapy that included ß-adrenergic blockers and ACE inhibitors. Overall mortality was reduced by 31% in patients randomized to receive ICDs during a mean follow-up of 20 months.39 The mortality benefit resulted entirely from the observed reduction in the risk of arrhythmic death and appeared to be greatest for patients with more remote MIs.40,41 The sum of the clinical data suggests that the risk of arrhythmic death has a complex distribution, with an early peak apparently related to acute infarction or pump failure, followed by a gradual increase over time, possibly as the result of long-term electrical remodeling or the development of new ischemia.
|
The largest study to date demonstrating the superiority of ICDs to medical therapies, including antiarrhythmic therapies, is the Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT).34 This 3-armed study randomized 2521 patients with ischemic or nonischemic cardiomyopathy, an ejection fraction
35%, and class II to III congestive heart failure to conventional therapy, conventional therapy plus weight-adjusted amiodarone, or conventional therapy plus a single-zone, single-chamber ICD with backup pacing only. Over a mean follow-up of 45.5 months, survival was no different in the standard therapy arms with and without amiodarone but was 23% lower in the group treated with an ICD. These results were consistent regardless of the cause of heart failure.
Among the central themes of sudden cardiac death prevention has been the understanding that the highest risk of arrhythmic death occurs around the time of an acute MI. The development of coronary care units in the 1960s, the accessibility to rapid external defibrillation, and the associated decline in the rates of post-MI mortality further served as proof-of-concept. However, it has also been recognized that some sustained tachyarrhythmias occurring in the immediate peri-MI period may have little long-term prognostic value, especially when they occur in the first few hours of the event.42 With the exception of the Multicenter UnSustained Tachycardia Trial (MUSTT), most ICD studies excluded patients within 3 weeks of an MI. The Defibrillators IN Acute Myocardial Infarction Trial (DINAMIT) was designed to test the hypothesis that survivors of recent MI (6 to 40 days) with an ejection fraction of
35% and evidence of sympathetic activation as assessed by heart rate variability would derive a survival benefit from ICD implantation. DINAMIT randomized 674 patients to an ICD or standard medical therapies but surprisingly found no difference in overall mortality between the 2 groups during an average follow-up of 30 months. These results occurred despite a 58% reduction in the relative risk of arrhythmic death and were driven mainly by an increase in nonarrhythmic cardiovascular mortality in the ICD arm.38 Although, as the authors suggest, it is possible that VT/VF in this population may serve as a marker for patients at high risk of mechanical cardiovascular mortality from issues such as pump failure, these intriguing findings require further study before final conclusions can be drawn.
Primary Prevention in Nonischemic Cardiomyopathy
Although coronary artery disease accounts for
80% of the cases of dilated cardiomyopathy, patients with nonischemic dilated cardiomyopathy are also at increased risk of sudden cardiac death by less well-understood mechanisms and have served as the focus of several trials addressing this issue. In the aforementioned SCD-HeFT trial, the beneficial effects of ICD therapy applied equally to patients with ischemic and nonischemic cardiomyopathy.34 Other ICD studies limited to patients with nonischemic dilated cardiomyopathy include the CArdiomyopathy Trial (CAT), the AMIOdarone versus implantable cardioVERTer-defibrillator trial (AMIOVERT), and the DEFibrillator In NonIschemic cardiomyopathy Treatment Evaluation (DEFINITE).43–45 The largest of these trials, DEFINITE, randomized 458 patients with nonischemic dilated cardiomyopathy, ejection fraction
35%, ambient ventricular arrhythmias, and class I to III heart failure to receive either ICD or standard medical therapy. Throughout a mean follow-up of 29 months, there was a nonsignificant 35% relative risk reduction in all-cause mortality and a highly significant 80% reduction in risk of arrhythmic death.44 A meta-analysis of all primary prevention trials to date that included nonischemic cardiomyopathy patients found a statistically significant 26% overall reduction in total mortality with standard (non–cardiac resynchronization therapy [CRT]) ICD use.21 Recently, the Centers for Medicare and Medicaid Services approved ICD implantation only for patients with nonischemic dilated cardiomyopathy of at least 9 months duration, or of 3 months duration if the patient was enrolled in the American College of Cardiology/Heart Rhythm Society–sponsored ICD Registry.46–48 Although some nonischemic cardiomyopathies may resolve during this period, the potential benefit must be weighed against the low risk of sudden death in the first months after diagnosis. While an analysis of the DEFINITE data did not reveal any less benefit of ICD therapy in patients with more recent diagnosis of dilated cardiomyopathy, a watch-and-wait approach in the first few months after diagnosis would likely result in only a small number of events, provided care is taken to monitor patients closely during this time.
The myriad of clinical trials and subgroup analyses examining the role of ICD therapy for primary prophylaxis of sudden death in patients with cardiomyopathy may still leave some practicing physicians with 2 simple questions: Who should get a prophylactic ICD, and when should it be implanted? Table 2 shows a compilation of current indications. Clinical judgment must also be applied, because competing comorbidities that impact duration and/or quality of life, very advanced age, and psychological factors may limit the potential benefit of ICDs even in those patients who meet these broad implantation indications.
|
Primary Prevention of Sudden Death in Other Diseases
A number of diseases other than dilated cardiomyopathy have been associated with an increased incidence of sudden cardiac death. These include inherited diseases of ion channels, such as LQTS,49 Brugada syndrome,50 and catecholaminergic VT.51 In addition, other structural heart diseases, such as right ventricular dysplasia,52 hypertrophic cardiomyopathy,53 and certain types of congenital heart disease, may be associated with increased risk of sudden death. The frequency of these conditions is not as high as that of ischemic or nonischemic dilated cardiomyopathy, which makes prospective randomized, controlled trials difficult to perform. Nonrandomized observational studies suggest subgroups of high-risk patients may benefit from ICD therapy.
In the absence of large-scale trials for most of these conditions, risk-stratification algorithms based on retrospective studies have been created in an attempt to help select appropriate patients for ICD therapy. In addition, there are disease-specific markers, such as the length of the QT interval in the inherited LQTS,49 the presence of persistent rather than intermittent right precordial ST elevation in Brugada syndrome,50 and the degree of left ventricular hypertrophy in patients with hypertrophic cardiomyopathy,54–57 that appear to portend an increased risk of sudden death and may make aggressive use of the ICD for the primary prevention of sudden death appropriate.
| Cost-Effectiveness of ICD Therapy |
|---|
Overall, these studies support the assertion that the cost of ICD therapy in patients who meet the inclusion criteria of the aforementioned trials may fall within the acceptable range from a societal vantage point under certain assumptions. Given the up-front costs of the ICD combined with the potential for a greater accrual in survival over time, the cost-benefit ratio of ICD therapy may be even more attractive than reported from randomized trials. Furthermore, efforts to better risk-stratify patients for primary prophylaxis would, if successful, reduce the number needed to treat and favorably impact the cost-effectiveness of ICD therapy.
As previously stated, there are no randomized trials to date that evaluated the use of ICDs for the treatment of LQTS or hypertrophic cardiomyopathy. Thus, cost-effectiveness data can only be derived by analytical models. One such analysis found that ICD therapy is highly cost-effective for the treatment of patients with LQTS or hypertrophic cardiomyopathy when used for primary prophylaxis in high-risk patients or for secondary prevention, driven mainly by the cost savings of added years of gained productivity in these generally young, otherwise healthy individuals.61
| Quality of Life With ICD Therapy |
|---|
The experience of shock remains a distinguishing characteristic of ICD patients over and above device implantation itself. Nonetheless, the changes in the mental-emotional scales seen with a single shock are often small and unlikely to reach the threshold for a clinically observable change. The clinical experience of psychological distress secondary to ICD shock may be more likely after the additive effects of multiple shocks, which result in obvious decrements in quality of life. Data from randomized controlled trials in the secondary prevention population indicate that even 1 shock is associated with reduced quality of life.63 Other studies, however, have suggested that 5 or more shocks defined the threshold for a decreased quality-of-life outcome.62,66–68 These results suggest that enhanced algorithms aimed at reducing inappropriate shocks delivered for atrial tachyarrhythmias with rapid ventricular response may improve health-related quality of life and support the potential benefits of programming interventions designed to terminate more stable ventricular tachyarrhythmias first with antitachycardia pacing.69
| Risk Stratification for ICD Therapy |
|---|
| Device Selection, Programming, and Testing |
|---|
There are no guidelines concerning device programming for the treatment of ventricular tachyarrhythmias, because therapies must be tailored to each individuals clinical circumstances. Randomized trials of device therapy have generally included single-zone programming for shocks to be delivered for rates exceeding
180 bpm. Although concerns have been voiced about the possibility that rapid antitachycardia pacing designed to pace-terminate VT may actually accelerate the arrhythmia, reports have demonstrated that antitachycardia pacing terminates 78% to 94% of VTs <188 to 200 bpm with a 2% to 4% risk of acceleration.69 A randomized study compared a single empirical attempt at antitachycardia pacing versus shock for fast VT at 188 to 250 bpm and found an 81% efficacy rate of antitachycardia pacing and no difference in acceleration, episode duration, or sudden death between the 2 groups.69
As with device programming, there are no guidelines concerning the testing of defibrillation thresholds either at the time of implantation or routinely during follow-up. Although device evaluation at implantation appears reasonable, it also exposes the patient to some risk, possible discomfort, and longer procedure times. A 10-J safety margin has been the traditional "standard of care," but this number was derived before the introduction of current technologies designed to enhance the efficacy of defibrillation, and there is evidence that lower margins may be acceptable.82 Although the recently published SCD-HeFT trial did not perform routine defibrillation threshold testing during implantation, an observational study found that 9% of patients with standard-output devices and 3% to 4% of patients with high-output devices had a <10-J safety margin at implantation.83 Because nearly all of these patients received successful system modification at the time of implantation, the clinical outcomes of patients with <10-J safety margins are unknown. Additional study is needed before any evidence-based conclusions can be drawn on this issue. In the interim, device-based testing of defibrillation thresholds is reasonable, and an 8- to 10-J difference between the maximum delivered energy of the device and the energy required to terminate VF should be used to define an adequate safety margin.
Routine follow-up testing has also evolved over time. The ability to assess shocking-lead impedance and other parameters has enabled practitioners to evaluate lead integrity without delivering therapy. Because the diagnostic yield of long-term assessment of defibrillation thresholds in the modern ICD era is uncertain, it appears reasonable that the need for and timing of such testing should be at the discretion of the treating physician.84,85 The addition of medical therapies, the development of conditions known to affect defibrillation thresholds, changes in clinical status, or concerns about the efficacy of therapy should elicit additional consideration for defibrillation threshold testing.
| Who Should Implant ICDs, and What Are the Risks? |
|---|
The advent of transvenous leads and downsized generators has markedly changed the complexity of ICD implantation. Compared with the era of epicardial leads that required thoracotomy and abdominal generator implants, the current generation allows for very high implantation success rates, reduced operative times, and a substantial shift in the rate and types of complications. Despite these advances, ICD implantation still carries risks both in the perioperative period and long-term. In a case series of 3344 patients from the European Registry of Implantable Defibrillators, the most common complications in the first 3 months after implantation were pocket hematoma in 1.9% and lead dislodgement in 1.4%. Infections were noted in 0.3%, and 1.6% complained of chronic pain in the generator pocket.88 An analysis of more than 23 000 Medicare patients who underwent ICD implantation demonstrated an overall complication rate of 11%, which was dominated by mechanical complications of the ICD system such as lead dislodgement (4.8%), hematoma/hemorrhage (2.5%), infection (1.4%), and pneumothorax (1.0%). Pericardial effusion/tamponade occurred in 0.3% of patients, and overall mortality in that study was 0.9%, one third of which occurred in patients with 1 or more complications.89
Complication types and rates may be affected by numerous variables, including operator experience, access approach, the number of leads implanted, patient comorbidities, and medical therapies, including antiplatelet agents, anticoagulants, and steroids. A national ICD registry established by the Centers for Medicare and Medicaid Services will better define the outcomes of ICD recipients in clinical practice.
| Device Recalls |
|---|
20-fold higher than that of pacemaker malfunction, and although malfunction rates for the latter declined significantly over a 21-year period, ICD malfunctions actually increased in more recent years.92 These concerns have resulted in specific recommendations made by the Heart Rhythm Society to device manufacturers, the Food and Drug Administration (FDA), Congress, the Centers for Medicare and Medicaid Services, and physicians.91,93 The document sets forth standards for independent performance review, terminology, and physician notification and requests the inclusion of performance data and data on adverse device events in a national registry, among other important recommendations. The ultimate goal of this effort is to "enhance and strengthen knowledge, confidence, and trust in cardiac rhythm management devices through: (1) greater transparency in post-market surveillance, analysis, and reporting; (2) enhanced systems to increase the return of devices to manufacturers and to improve the analysis and reporting of device performance and malfunction information; and (3) cooperation among industry, the FDA, and physicians in an effort to prevent injuries and deaths due to device malfunction."94 Whether these efforts result in a substantial improvement in device reliability and the methods by which malfunctions are dealt with will require further study.
| Future Directions |
|---|
Although the absolute number of patients who experience sudden cardiac death in the population with known but less extensive degrees of structural heart disease is large, the relative risk is small. Thus, identifying those patients who currently have no indications for ICD therapy but who may benefit from its use is an important but challenging issue. Trials using existing risk-stratification techniques alone are unlikely to provide the necessary predictive value in patients at lower risk of events. A study is currently under way comparing ICD versus standard medical therapy in patients with ejection fractions above 30% to 35% but with large infarcts on cardiac magnetic resonance imaging. The hypotheses of this study are that patients with large areas of scar are at risk for sudden death even if left ventricular ejection fraction is preserved and that ICD therapy can reduce this risk. This study and other proposed studies like it have the potential for determining whether myocardial scar, recurrent ischemia, or other factors contribute to sudden death in patients with relatively preserved ventricular function. Although it is still unclear whether ICD therapy in this population can be cost-effective, the large public health nature of the problem suggests that further studies are appropriate.
The last decade has resulted in amazing improvements in the technological aspects of ICDs although problems with device reliability remain. ICD manufacturers should first focus on device reliability and on improved communication with physicians and patients. However, technological improvements that will result in significant patient benefit continue to be made. A leadless ICD system may one day simplify device implantation and eliminate lead-related complications. Remote monitoring of ICD therapy is now provided by a number of manufacturers, which allows patients who live far away from providers to still have their devices evaluated and those who have symptoms or shocks to have remote interrogation that can be useful in management. ICDs have already become cardiac management devices rather than simply defibrillators. They include sophisticated pacing algorithms, atrial antitachycardia pacing, heart failure monitoring, and other refinements. These trends are likely to continue, and automatic adjustments in ICD therapy and heart failure therapy based on physiological monitoring will be further refined. ICD size should continue to decrease, and battery life should continue to improve. Advances in ICD technology must be balanced with encouragements to limit the cost of this therapy. These developments may result in increasing adoption rates for established indications, as well as expansion of the use of ICDs to new populations, both of which will result in significant improvements in public health.
| Conclusions |
|---|
| Acknowledgments |
|---|
Dr Passman is funded by NIH K23 HL.688; Dr Kadish is funded by NIH R01HL075382-01A1.
Disclosures
Dr Passman has received grant support from Medtronic; Dr Kadish received has grant support from Medtronic and St. Jude.
| References |
|---|
2. Langworthy O, Kouwenhoven W. An experimental study of abnormalities produced in the organism by electricity. J Indust Hyg. 1930; 12: 31–65.
3. Hooker D, Kouwenhoven W, Langworthy O. The effect of alternating electrical currents on the heart. Electric Eng. 1936; 55: 444–454.
4. White PD. Heart Disease. New York, NY: Macmillan; 1931.
5. Beck C, Pritchard W, Heil H. Ventricular fibrillation of long duration abolished by electric shock. JAMA. 1947; 135: 985–986.
6. Mirowski M, Reid P, Mower M, Watkins L, Gott V, Schauble J, Langer A, Heilman M, Kolenik S, Fischell R, Weisfeldt M. Termination of malignant ventricular arrhythmias with an implanted automatic defibrillator in human beings. N Engl J Med. 1980; 303: 322–324.[Medline] [Order article via Infotrieve]
7. Eldar M, Griffin J, Abbott J, Benditt D, Bhandari A, Herre J, Benson D, Scheinman M. Permanent cardiac pacing in patients with the long QT syndrome. J Am Coll Cardiol. 1987; 10: 600–607.[Abstract]
8. Eldar M, Griffin J, Van Hare G, Witherell C, Bhandari A, Benditt D, Scheinman M. Combined use of beta-adrenergic blocking agents and long-term cardiac pacing for patients with the long QT syndrome. J Am Coll Cardiol. 1992; 20: 830–837.[Abstract]
9. Viskin S. Cardiac pacing in the long QT syndrome: review of available data and practical recommendations. J Cardiovasc Electrophysiol. 2000; 11: 593–600.[Medline] [Order article via Infotrieve]
10. Tan H, Bardai A, Shimizu W, Moss A, Schulze-Bahr E, Noda T, Wilde A. Genotype-specific onset of arrhythmias in congenital long-QT syndrome. Circulation. 2006; 114: 2096–2103.
11. Dorostkar P, Eldar M, Belhasse B, Scheinman M. Long-term follow-up of patients with long-QT syndrome treated with beta-blockers and continuous pacing. Circulation. 1999; 100: 2431–2436.
12. Gregoratos G, Abrams J, Epstein A, Freedman R, Hayes D, Hlatky M, Kerber R, Naccarelli G, Schoenfeld M, Silka M, Winters S, Gibbons R, Antman E, Alpert J, Gregoratos G, Hiratzka L, Faxon D, Jacobs A, Fuster V, Smith SJ. ACC/AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices: summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines. Circulation. 2002; 106: 2145–2161.
13. Zipes D, Camm A, Borggrefe M, Buxton A, Chaitman B, Fromer M, Gregoratos G, Klein G, Moss A, Myerburg R, Priori S, Quinones M, Roden D, Silka M, Tracy C, Smith SJ, Jacobs A, Adams C, Antman E, Anderson J, Hunt S, Halperin J, Nishimura R, Ornato J, Page R, Riegel B, Priori S, Blanc J, Budaj A, Camm A, Dean V, Deckers J, Despres C, Dickstein K, Lekakis J, McGregor K, Metra M, Morais J, Osterspey A, Tamargo J, Zamorano J; European Heart Rhythm Association, Heart Rhythm Society, American College of Cardiology, American Heart Association Task Force, European Society of Cardiology Committee for Practice Guidelines. ACC/AHA/ESC 2006 guidelines for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: a report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death). Circulation. 2006; 114: e385–e484.[CrossRef][Medline] [Order article via Infotrieve]
14. Myerburg RKK, Castellanos A. Sudden cardiac death: structure, function, and time-dependence of risk. Circulation. 1992; 85 (suppl I): I-2–I-10.[Medline] [Order article via Infotrieve]
15. Liberthson R, Nagel E, Hirschman J, Nussenfeld S. Prehospital ventricular defibrillation: prognosis and follow-up course. N Engl J Med. 1974; 291: 317–321.[Medline] [Order article via Infotrieve]
16. CASCADE I. Randomized antiarrhythmic drug therapy in survivors of cardiac arrest (the CASCADE Study). Am J Cardiol. 1993; 72: 280–287.[CrossRef][Medline] [Order article via Infotrieve]
17. AVID I: a comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. N Engl J Med. 1997; 337: 1576–1583.
18. Domanski M, Sakseena S, Epstein A, Hallstrom A, Brodsky M, Kim S, Lancaster S, Schron E. Relative effectiveness of the implantable cardioverter-defibrillator and antiarrhythmic drugs in patients with varying degrees of left ventricular dysfunction who have survived malignant ventricular arrhythmias. J Am Coll Cardiol. 1999; 34: 1090–1095.
19. Connolly S, Gent M, Roberts R, Dorian P, Roy D, Sheldon R, Mitchell L, Green M, Klein G, OBrien B. Canadian Implantable Defibrillator Study (CIDS): a randomized trial of the implantable cardioverter defibrillator against amiodarone. Circulation. 2000; 101: 1297–1307.
20. Kuck K, Cappato R, Siebels J, Ruppel R. Randomized comparison of antiarrhythmic drug therapy with implantable defibrillators in patients resuscitated from cardiac arrest: the Cardiac Arrest Study Hamburg (CASH). Circulation. 2000; 102: 748–754.
21. Desai A, Fang J, Maisel W, Baughman K. Implantable defibrillators for the prevention of mortality in patients with nonischemic cardiomyopathy: a meta-analysis of randomized controlled trials. JAMA. 2004; 292: 2874–2879.
22. Oseroff O, Retyk E, Bochoeyer A. Subanalyses of secondary prevention implantable cardioverter-defibrillator trials: Antiarrhythmics Versus Implantable Defibrillators (AVID), Canadian Implantable Defibrillator Study (CIDS), and Cardiac Arrest Study Hamburg (CASH). Curr Opin Cardiol. 2004; 19: 26–30.[CrossRef][Medline] [Order article via Infotrieve]
23. Wyse D, Friedman P, Brodsky M, Beckman K, Carlson M, Curtis A, Hallstrom A, Raitt M, Wilkoff B, Greene H, Investigators A. Life-threatening ventricular arrhythmias due to transient or correctable causes: high risk for death in follow-up. J Am Coll Cardiol. 2001; 38: 1718–1724.
24. de Bakker J, van Capelle F, Janse M, Wilde A, Coronel R, Becker A, Dingemans K, van Hemel N, Hauer R. Reentry as a cause of ventricular tachycardia in patients with chronic ischemic heart disease: electrophysiologic and anatomic correlation. Circulation. 1988; 77: 589–606.
25. Downar E, Saito J, Doig J, Chen T, Sevaptsidis E, Masse S, Kimber S, Mickleborough L, Harris L. Endocardial mapping of ventricular tachycardia in the intact human ventricle, III: evidence of multiuse reentry with spontaneous and induced block in portions of reentrant path complex. J Am Coll Cardiol. 1995; 25: 1591–1600.[Abstract]
26. Davis H, DeCamilla J, Bayer L, Moss A. Survivorship patterns in the posthospital phase of myocardial infarction. Circulation. 1979; 60: 1252–1258.
27. Greene H, Richardson D, Barker A, Roden D, Capone R, Echt D, Friedman L, Gillespie M, Hallstrom A, Verter J. Classification of deaths after myocardial infarction as arrhythmic or nonarrhythmic (the Cardiac Arrhythmia Pilot Study). Am J Cardiol. 1989; 63: 1–6.[CrossRef][Medline] [Order article via Infotrieve]
28. Ahnve S, Gilpin E, Henning H, Curtis G, Collins D, Ross JJ. Limitations and advantages of the ejection fraction for defining high risk after acute myocardial infarction. Am J Cardiol. 1986; 58: 872–878.[CrossRef][Medline] [Order article via Infotrieve]
29. Bigger JJ, Fleiss J, Kleiger R, Miller J, Rolnitzky L. The relationships among ventricular arrhythmias, left ventricular dysfunction, and mortality in the 2 years after myocardial infarction. Circulation. 1984; 69: 250–258.
30. Greenberg H, McMaster P, Dwyer EJ. Left ventricular dysfunction after acute myocardial infarction: results of a prospective multicenter study. J Am Coll Cardiol. 1984; 4: 867–874.[Abstract]
31. Mukharji J, Rude R, Poole W, Gustafson N, Thomas LJ, Strauss H, Jaffe A, Muller J, Roberts R, Raabe DJ. Risk factors for sudden death after acute myocardial infarction: two-year follow-up. Am J Cardiol. 1984; 54: 31–36.[CrossRef][Medline] [Order article via Infotrieve]
32. Sanz G, Castaner A, Betriu A, Magrina J, Roig E, Coll S, Pare J, Navarro-Lopez F. Determinants of prognosis in survivors of myocardial infarction: a prospective clinical angiographic study. N Engl J Med. 1982; 306: 1065–1070.[Abstract]
33. Schulze RJ, Strauss H, Pitt B. Sudden death in the year following myocardial infarction: relation to ventricular premature contractions in the late hospitals phase and left ventricular ejection fraction. Am J Med. 1977; 62: 192–199.[CrossRef][Medline] [Order article via Infotrieve]
34. Bardy G, Lee KL, Mark D, Poole J, Packer D, Boineau R, Domanski M, Troutman R, Anderson J, Johnson G, McNulty S, Clapp-Channing N, Davidson-Ray L, Fraulo E, Fishbein D, Luceri R, Ip J. Amiodarone or an implantable cardioverter- defibrillator for congestive heart failure. N Engl J Med. 2005; 352: 225–237.
35. Moss A, Hall W, Cannom D, Daubert J, Higgins S, Klein H, Levine J, Saksena S, Waldo A, Wilber D, Brown M, Heo M; Multicenter Automatic Defibrillator Implantation Trial Investigators. Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. N Engl J Med. 1996; 335: 1933–1940.
36. Bigger J; Coronary Artery Bypass Graft (CABG) Patch Trial Investigators. Prophylactic use of implanted cardiac defibrillators in patients at high risk for ventricular arrhythmias after coronary-artery bypass graft surgery. N Engl J Med. 1997; 337: 1569–1575.
37. Buxton A, Lee K, DiCarlo L, Gold M, Greer G, Prystowsky E, OToole M, Tang A, Fisher J, Coromilas J, Talajic M, Hafley G; Multicenter Unsustained Tachycardia Trial Investigators. Electrophysiologic testing to identify patients with coronary artery disease who are at risk for sudden death. N Engl J Med. 2000; 342: 1937–1945.
38. Hohnloser S, Kuck KH, Dorian P, Roberts R, Hampton J, Hatala R, Fain E, Gent M, Connolly S; DINAMIT Investigators. Prophylactic use of an implantable cardioverter-defibrillator after acute myocardial infarction. N Engl J Med. 2004; 351: 2481–2488.
39. Moss A, Zareba W, Hall W, Klein H, Wilber D, Cannom D, Daubert J, Higgins S, Brown M, Andrews M; for the MADIT II Investigators. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002; 346: 877–883.
40. Greenberg H, Case R, Moss A, Brown M, Carroll E, Andrews M; for the MADIT II Investigators. Analysis of mortality events in the Multicenter Automatic Defibrillator Implantation Trial (MADIT-II). J Am Coll Cardiol. 2004; 43: 1459–1465.
41. Wilber D, Zareba W, Hall W, Brown M, Lin A, Andrews M, Burke M, Moss A. Time dependence of mortality risk and defibrillator benefit after myocardial infarction. Circulation. 2004; 109: 1082–1084.
42. Behar S, Goldbourt U, Reicher-Reiss H, Kaplinsky E; Principal Investigators of the SPRINT Study. Prognosis of acute myocardial infarction complicated by primary ventricular fibrillation. Am J Cardiol. 1990; 66: 1208–1211.[CrossRef][Medline] [Order article via Infotrieve]
43. Bansch D, Antz M, Boczor S, Volkmer M, Tebbenjohanns J, Seidl K, Block M, Gietzen F, Berger J, Kuck KH. Primary prevention of sudden cardiac death in idiopathic dilated cardiomyopathy: the Cardiomyopathy Trial (CAT). Circulation. 2002; 105: 1453–1458.
44. Kadish A, Dyer A, Daubert JP, Quigg R, Estes M, Anderson KP, Calkins H, Hoch D, Goldberger J, Shalaby A, Sanders WE, Schaechter A, Levine JH; Defibrillators in Non-Ischemic Cardiomyopathy Treatment Evaluation (DEFINITE) Investigators. Prophylactic defibrillator implantation in patients with nonischemic dilated cardiomyopathy. N Engl J Med. 2004; 350: 2151–2158.
45. Strickberger SA, Hummel JD, Bartlett TG, Frumin HI, Schuger CD, Beau SL, Bitar C, Morady F. Amiodarone versus implantable cardioverter-defibrillator: randomized trial in patients with nonischemic dilated cardiomyopathy and asymptomatic nonsustained ventricular tachycardia: AMIOVIRT. J Am Coll Cardiol. 2003; 41: 1707–1712.
46. ICD Registry. Available at: http://www.accncdr.com/webncdr/ICD/Default.aspx. Accessed March 20, 2007.
47. Medicare expands coverage to implantable defibrillators to save lives and develop evidence to maximize benefits [press release]. Baltimore, Md: CMS; January 27, 2005. Available at: http://www.cms.hhs.gov/apps/media/press/release.asp?Counter=1331. Accessed January 5, 2007.
48. McClellan M, Tunis SR. Medicare coverage of ICDs. N Engl J Med. 2005; 352: 222–224.
49. Moss AJ. Long QT syndrome. JAMA. 2003; 289: 2041–2044.
50. Priori SG, Napolitano C, Gasparini M, Pappone C, Della Bella P, Giordano U, Bloise R, Giustetto C, De Nardis R, Grillo M, Ronchetti E, Faggiano G, Nastoli J. Natural history of Brugada syndrome: insights for risk stratification and management. Circulation. 2002; 105: 1342–1347.
51. Priori SG, Napolitano C, Memmi M, Columbi B, Drago F, Gasparini M, DeSimone L, Coltorti F, Bloise R, Keegan R, Cruz Filho F, Vignati G, Benatar A, DeLogu A. Clinical and molecular characterization of patients with catecholaminergic polymorphic ventricular tachycardia. Circulation. 2002; 106: 69–74.
52. Tavernier R, Gevaert S, De Sutter J, De Clercq A, Rottiers H, Jordaens L, Fonteyne W. Long term results of cardioverter-defibrillator implantation in patients with right ventricular dysplasia and malignant ventricular tachyarrhythmias. Heart. 2001; 85: 53–56.
53. Maron B, Shen W, Link M, Epstein A, Almquist A, Daubert J, Bardy G, Favale S, Rea R, Boriani G, Estes NR, Spirito P. Efficacy of implantable cardioverter-defibrillators for the prevention of sudden death in patients with hypertrophic cardiomyopathy. N Engl J Med. 2000; 342: 365–373.
54. Elliott P, Gimeno Blanes J, Mahon N, Poloniecki J, McKenna W. Relation between severity of left-ventricular hypertrophy and prognosis in patients with hypertrophic cardiomyopathy. Lancet. 2001; 357: 420–424.[CrossRef][Medline] [Order article via Infotrieve]
55. Elliott P, Poloniecki J, Dickie S, Sharma S, Monserrat L, Varnava A, Mahon N, McKenna W. Sudden death in hypertrophic cardiomyopathy: identification of high risk patients. J Am Coll Cardiol. 2000; 36: 2212–2218.
56. Maron B, McKenna W, Danielson G, Kappenberger L, Kuhn H, Seidman C, Shah P, Spencer WR, Spirito P, Ten Cate F, Wigle E. American College of Cardiology/European Society of Cardiology clinical expert consensus document on hypertrophic cardiomyopathy: a report of the American College of Cardiology Foundation Task Force on Clinical Expert Consensus Documents and the European Society of Cardiology Committee for Practice Guidelines. J Am Coll Cardiol. 2003; 42: 1687–1713.
57. Spirito P, Bellone P, Harris K, Bernabo P, Bruzzi P, Maron B. Magnitude of left ventricular hypertrophy and risk of sudden death in hypertrophic cardiomyopathy. N Engl J Med. 2000; 342: 1778–1785.
58. Mark D, Nelson C, Anstrom K, Al-Khatib S, Tsiatis A, Cowper P, Clapp-Channing N, Davidson-Ray L, Poole J, Johnson G, Anderson J, Lee K, Bardy G; Sudden Cardiac Death in Heart Failure Trial Investigators. Cost-effectiveness of defibrillator therapy or amiodarone in chronic stable heart failure: results from the Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT). Circulation. 2006; 114: 135–142.
59. Mushlin A, Hall W, Zwanziger J, Gajary E, Andrews M, Marron R, Zou K, Moss A. The cost-effectiveness of automatic implantable cardiac defibrillators: results from MADIT: Multicenter Automatic Defibrillator Implantation Trial. Circulation. 1998; 97: 2129–2135.
60. Zwanziger J, Hall W, Dick A, Zhao H, Mushlin A, Hahn R, Wang H, Andrews M, Mooney C, Wang H, Moss A. The cost effectiveness of implantable cardioverter-defibrillators: results from the Multicenter Automatic Defibrillator Implantation Trial (MADIT)-II. J Am Coll Cardiol. 2006; 47: 2310–2318.
61. Goldenberg I, Moss A, Maron B, Dick A, Zareba W. Cost-effectiveness of implanted defibrillators in young people with inherited cardiac arrhythmias. Ann Noninvasive Electrocardiol. 2005; 10: 67–83.[CrossRef][Medline] [Order article via Infotrieve]
62. Schron E, Exner D, Yao Q, Jenkins L, Steinberg J, Cook J, Kutalek S, Friedman P, Bubien R, Page R, Powell J. Quality of life in the Antiarrhythmics Versus Implantable Defibrillators trial: impact of therapy and influence of adverse symptoms and defibrillator shocks. Circulation. 2002; 105: 589–594.
63. Irvine J, Dorian P, Baker B, OBrien B, Roberts R, Gent M, Newman D, Connolly S. Quality of life in the Canadian Implantable Defibrillator Study (CIDS). Am Heart J. 2002; 144: 282–289.[Medline] [Order article via Infotrieve]
64. Burke J, Hallas C, Clark-Carter D, White D, Connelly D. The psychosocial impact of the implantable cardioverter defibrillator: a meta-analytic review. Br J Health Psychol. 2003; 8: 165–178.[CrossRef][Medline] [Order article via Infotrieve]
65. Duru F, Buchi S, Klaghofer R, Mattmann H, Sensky T, Buddeberg C, Candinas R. How different from pacemaker patients are recipients of implantable cardioverter-defibrillators with respect to psychosocial adaptation, affective disorders, and quality of life? Heart. 2001; 85: 375–379.
66. Herrmann C, von zur Muhen F, Schaumann A, Buss U, Kemper S, Wantzen C, Gonska B. Standardized assessment of psychological well-being and quality-of-life in patients with implanted defibrillators. Pacing Clin Electrophysiol. 1997; 20: 95–103.[CrossRef][Medline] [Order article via Infotrieve]
67. Moser D, Dracup K. Is anxiety early after myocardial infarction associated with subsequent ischemic and arrhythmic events? Psychosom Med. 1996; 58: 395–401.
68. Sears SJ, Conti J. Understanding implantable cardioverter defibrillator shocks and storms: medical and psychosocial considerations for research and clinical care. Clin Cardiol. 2003; 26: 107–111.[Medline] [Order article via Infotrieve]
69. Wathen M, DeGroot P, Sweeney M, Stark A, Otterness M, Adkisson W, Canby R, Khalighi K, Machado C, Rubenstein D, Volosin K; PainFREE Rx II Investigators. Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter-defibrillators: Pacing Fast Ventricular Tachycardia Reduces Shock Therapies (PainFREE Rx II) trial results. Circulation. 2004; 110: 2591–2596.
70. Daubert JP, Zareba W, Hall WJ, Schuger C, Corsello A, Leon AR, Andrews ML, McNitt SH, DT, Moss AJ; MADIT II Study Investigators. Predictive value of ventricular arrhythmia inducibility for subsequent ventricular tachycardia or ventricular fibrillation in Multicenter Automatic Defibrillator Implantation Trial (MADIT) II patients. J Am Coll Cardiol. 2006; 47: 98–107.
71. Bloomfield D, Steinman R, Namerow P, Parides M, Davidenko J, Kaufman E, Shinn T, Curtis A, Fontaine J, Holmes D, Russo A, Tang C, Bigger JJ. Microvolt T-wave alternans distinguishes between patients likely and patients not likely to benefit from implanted cardiac defibrillator therapy: a solution to the Multicenter Automatic Defibrillator Implantation Trial (MADIT) II conundrum. Circulation. 2004; 110: 1885–1889.
72. Non-invasive test to predict sudden cardiac death proven effective [late-breaking clinical trial news release 10]. Chicago, Ill: American Heart Association; November 15, 2006. Available at: http://scientificsessions.americanheart.org/portal/scientificsessions/ss/lbctnewsrelease200610. Accessed June 28, 2007.
73. Gold M, Ensley D, Chilson D, Ip J, Constantini O, Bloomfield D, J P, McNulty S, Lee K, Bardy G. T-wave alternans SCD HeFT study: primary endpoint analysis. Circulation. 2006; 114 (suppl II): II-428. Abstract.
74. Alter PWS, Plachta E, Moosdorf R, Grimm W. Complications of implantable cardioverter defibrillator therapy in 440 consecutive patients. Pacing Clin Electrophysiol. 2005; 28: 926–932.[CrossRef][Medline] [Order article via Infotrieve]
75. Deisenhofer I, Kolb C, Ndrepepa G, Schreieck J, Karch M, Schmieder S, Zrenner B, Schmitt C. Do current dual chamber cardioverter defibrillators have advantages over conventional single chamber cardioverter defibrillators in reducing inappropriate therapies? A randomized, prospective study. J Cardiovasc Electrophysiol. 2001; 12: 134–142.[CrossRef][Medline] [Order article via Infotrieve]
76. Kolb C, Deisenhofer I, Schmieder S, Barthel P, Zrenner B, Karch M, Schmitt C. Long-term follow-up of patients supplied with single-chamber or dual-chamber cardioverter defibrillators. Pacing Clin Electrophysiol. 2006; 29: 946–952.[CrossRef][Medline] [Order article via Infotrieve]
77. Theuns D, Klootwijk A, Goedhart D, Jordaens L. Prevention of inappropriate therapy in implantable cardioverter-defibrillators: results of a prospective, randomized study of tachyarrhythmia detection algorithms. J Am Coll Cardiol. 2004; 44: 2362–2367.
78. Wilkoff B, Cook J, Epstein A, Greene H, Hallstrom A, Hsia H, Kutalek S, Sharma A; Dual Chamber and VVI Implantable Defibrillator Investigators. Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial. JAMA. 2002; 288: 3115–3123.
79. Sweeney M, Ellenbogen K, Casavant D, Betzold R, Sheldon T, Tang F, Mueller M, Lingle J; The Marquis MVP Download Investigators. Multicenter, prospective, randomized safety and efficacy study of a new atrial-based managed ventricular pacing mode (MVP) in dual chamber ICDs. J Cardiovasc Electrophysiol. 2005; 16: 811–817.[CrossRef][Medline] [Order article via Infotrieve]
80. Friedman P, McClelland R, Bamlet W, Acosta H, Kessler D, Munger T, Kavesh N, Wood M, Daoud E, Massumi A, Schuger C, Shorofsky S, Wilkoff B, Glikson M. Dual-chamber versus single-chamber detection enhancements for implantable defibrillator rhythm diagnosis: the Detect Supraventricular Tachycardia Study. Circulation. 2006; 113: 2871–2879.
81. Olshansky B, Day J, Moore S, Gering L, Rosenbaum M, McGuire M, Brown S, Lerew D. Is dual-chamber programing inferior to single-chamber programming in an implantable cardioverter defibrillator? Results of the INTRINSIC RV Study. Circulation. 2007; 115: 9–16.
82. Gold M, Higgins S, Klein R, Gilliam F, Kopelman H, Hessen S, Payne J, Strickberger S, Breiter D, Hahn S. Efficacy and temporal stability of reduced safety margins for ventricular defibrillation: primary results from the Low Energy Safety Study (LESS). Circulation. 2002; 105: 2043–2048.
83. Russo A, Sauer W, Gerstenfeld E, Hsia H, Lin D, Cooper J, Dixit S, Verdino R, Nayak H, Callans D, Patel V, Marchlinski F. Defibrillation threshold testing: is it really necessary at the time of implantable cardioverter-defibrillator insertion? Heart Rhythm. 2005; 2: 456–461.[CrossRef][Medline] [Order article via Infotrieve]
84. Delvecchio A, Trivedi H, Fisher J, Kim S, Ferrick K, Gross J, Palma E. Value of pre-hospital discharge defibrillation testing in recipients of implanted cardioverter defibrillators. Pacing Clin Electrophysiol. 2005; 28: S260–S262.[CrossRef][Medline] [Order article via Infotrieve]
85. Goldberger J, Horvath G, Inbar S, Kadish A. Utility of predischarge and one-month transvenous implantable defibrillator tests. Am J Cardiol. 1997; 79: 822–826.[CrossRef][Medline] [Order article via Infotrieve]
86. Beller GA, Bonow RO, Fuster V. ACC revised recommendations for training in adult cardiovascular medicine core cardiology training II (COCATS 2) (revision of the 1995 COCATS training statement). 2002. American College of Cardiology Web site. Available at: http://www.acc.org/clinical/training/cocats2.pdf. Accessed November 1, 2006.
87. Curtis A, Ellenbogen K, Hammill S, Hayes D, Reynolds D, Wilber D, Cain M. Clinical competency statement: training pathways for implantation of cardioverter defibrillators and cardiac resynchronization devices. Heart Rhythm. 2004; 7: 371–375.
88. Gradaus R, Block M, Brachmann J, Breithardt G, Huber H, Jung W, Kranig W, Mletzko R, Schoels W, Seidl K, Senges J, Siebels J, Steinbeck G, Stellbrink C, Andresen D; German EURID Registry. Mortality, morbidity, and complications in 3344 patients with implantable cardioverter defibrillators: results from the German ICD Registry EURID. Pacing Clin Electrophysiol. 2003; 26 (pt 1): 1511–1518.[CrossRef][Medline] [Order article via Infotrieve]
89. Reynolds M, Cohen D, Kugelmass A, Brown P, Becker E, Culler S, Simon A. The frequency and incremental cost of major complications among Medicare beneficiaries receiving implantable cardioverter-defibrillators. J AM Coll Cardiol. 2006; 42: 2493–2497.
90. Hauser R, Maron B. Lessons from the failure and recall of an implantable cardioverter-defibrillator. Circulation. 2005; 112: 2040–2042.
91. Maisel W, Sweeney M, Stevenson W, Ellison K, Epstein L. Recalls and safety alerts involving pacemakers and implantable cardioverter-defibrillator generators. JAMA. 2001; 286: 793–799.
92. Maisel W. Pacemaker and ICD generator reliability: meta-analysis of device registries. JAMA. 2006; 295: 1929–1934.
93. Myerburg RJ, Feigal DW, Lindsay BD. Life-threatening malfunction of implantable cardiac devices. N Engl J Med. 2006; 354: 2309–2311.
94. Carlson M, Wilkoff B, Maisel W, Carlson M, Ellenbogen K, Saxon L, Prystowsky E, Alpert J, Cain M, Ching E, Curtis A, Davies D, Hammill S, Hauser R, Lampert R, Zipes D. Recommendations from the Heart Rhythm Society Task Force on Device Performance Policies and Guidelines endorsed by the American College of Cardiology Foundation (ACCF) and the American Heart Association (AHA) and the International Coalition of Pacing and Electrophysiology Organizations (COPE). Heart Rhythm. 2006; 3: 1250–1273.[CrossRef][Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
R. J. Myerburg, V. Reddy, and A. Castellanos Indications for implantable cardioverter-defibrillators based on evidence and judgment. J. Am. Coll. Cardiol., August 25, 2009; 54(9): 747 - 763. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. G. Strauss, R. H. Selvester, J. A.C. Lima, H. Arheden, J. M. Miller, G. Gerstenblith, E. Marban, R. G. Weiss, G. F. Tomaselli, G. S. Wagner, et al. ECG Quantification of Myocardial Scar in Cardiomyopathy Patients With or Without Conduction Defects: Correlation With Cardiac Magnetic Resonance and Arrhythmogenesis Circ Arrhythm Electrophysiol, December 1, 2008; 1(5): 327 - 336. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2007 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |