(Circulation. 2007;116:2866-2877.)
© 2007 American Heart Association, Inc.
Advances in Interventional Cardiology |
From Evanston Hospital (T.F.), Evanston, Ill, and Columbia University Medical Center (M.B.L.), New York, NY.
Address correspondence to Ted Feldman, MD, Evanston Hospital, Cardiology Division, Walgreen Bldg 3rd Floor, 2650 Ridge Ave, Evanston, IL 60201. E-mail tfeldman{at}enh.org
Key Words: aortic stenosis mitral regurgitation mitral valve aortic valve
| Introduction |
|---|
|
|
|---|
| Unmet Clinical Needs |
|---|
|
|
|---|
| Preclinical Assessments |
|---|
|
|
|---|
| Clinical Trial Methodologies |
|---|
|
|
|---|
Most assuredly, to propose transcatheter valve alternatives will require the demonstration of incremental clinical benefits in well-characterized patient cohorts studied with the use of rigorous clinical trial methodologies. Thus, these new therapy approaches must have substantive clinical value and cannot simply be the fashionable extrapolation of previous catheter-based treatments for vascular disease.
| Percutaneous Pulmonary Valve Replacement |
|---|
|
|
|---|
200 patients worldwide with congenital heart disease. Seventeen of these have been stent-within-stent procedures. Infrequent procedural complications (
6% of cases) have included homograft rupture, device dislodgement, coronary compression, and jailing of the right pulmonary artery. One- and 5-year freedom from subsequent surgical procedures to treat device failures has been 90% and 80%, respectively (Philipp Bonhoeffer, MD, personal communication, 2007). In addition to sparing these patients the rigors of an additional or repeat high-risk surgical cardiac procedure, pulmonic valve replacement in this population has been shown to have a profound impact on the physiology of right ventricular outflow tract obstruction occurring late after repair for other congenital heart defects, with improvements in symptoms, aerobic and anaerobic exercise capacity, right ventricular volumes, and systolic and diastolic function.10,11 | Surgical Aortic Valve Replacement—The "Gold Standard," but Not for Everyone |
|---|
|
|
|---|
Despite clear benefits of AVR in patients with aortic stenosis, morbidity and mortality of valve replacement surgery remains significant in several subgroups. Average perioperative mortality reported from the recent Society of Thoracic Surgeons database is 5.5% to 6.8% for AVR combined with coronary bypass surgery.16,17 Surgical mortality increases by 33% in low-volume centers18,19 and increases to >10% in octogenarians.20,21 Other traditional operative risk factors include female gender, pulmonary hypertension, chronic pulmonary disease, prior cardiac surgical procedures, chronic renal insufficiency, reduced ejection fraction (especially an ejection fraction <30% with prior myocardial infarction), and Ney York Heart Association class III and IV congestive heart failure.22–24 In addition, other imponderables contribute to surgical mortality, such as heavy calcification of the thoracic aorta (so-called porcelain aorta), chest-wall radiation exposure, and chest-wall deformities. These higher risks are amplified in the elderly population, in which additional postoperative concerns are present, including increased neurocognitive dysfunction, prolonged hospital stays, frequent discharge to a rehabilitation facility, and high readmission rates in the first 30 days, primarily as a result of heart failure, arrhythmias, and wound infections.25–27 Thus, an expanding population exists of patients with severe aortic stenosis in whom aortic valve surgery is either not considered or results in disproportionate perioperative mortality and morbidity. This so-called high-risk group has been the focus for new transcatheter AVR therapies.
| Transcatheter Aortic Valve Replacement |
|---|
|
|
|---|
Successful application of catheter-based aortic valve technology depends greatly on careful attention to operator technique and procedural methodology. All current catheter-based aortic stent valve procedures begin with conventional balloon aortic valvuloplasty to provide an enlarged passageway for the insertion of the larger stent-valve device. Careful patient screening for vascular access and aortic annulus dimensions for correct valve sizing are critical. Because of increased periprocedural complications associated with the earlier transvenous antegrade approach, the transarterial retrograde approach is now the preferred technique. This requires femoral artery access (and usually a surgical arterial repair for closure); negotiation of the femoral, iliac, and aortic vasculature; and retrograde crossing of the calcified aortic valve. In general, the implanted catheter-based aortic valves are oversized relative to the annulus to ensure stability and immobility of the implanted stent and to minimize paravalvular regurgitation. During valve deployment, to ensure hemodynamic stability and to minimize movement effects of pulsatile aortic flow heart on stent-valve positioning, additional hemodynamic support devices or temporary rapid right ventricular pacing (to heart rates
200 beats per minute) may be necessary. After implantation of the stent valve, postdilatation with slightly larger balloon catheter may be needed to reduce paravalvular regurgitation.
The Cribier-Edwards Valve
The Cribier-Edwards Aortic Bioprosthesis (Edwards Lifesciences Inc., Orange, Calif) has been in development for 8 years.29 The leaflet material is currently thin, durable equine pericardium. The 3 valve leaflets are hand-sewn to a stainless steel, tubular, slotted balloon expandable stent (Figure 1, left). Currently stent valves are available that are 23 mm and 26 mm in diameter and that accommodate aortic annulus sizes between 18 mm and 24 mm. At the time of implantation, the sterile stent valve is carefully mounted and crimped (with a specialized crimping tool) onto conventional balloon dilatation catheters. At present, during retrograde procedures, the balloon catheter and stent valve assembly are placed within a tip-deflecting catheter, which is then inserted into a 22F to 24F flexible arterial sheath.30 Active flexion of the tip-deflecting catheter navigates aortic tortuosity and assists with native valve crossing. After successful balloon valvuloplasty and retrograde positioning of the stent valve, rapid ventricular pacing is initiated and balloon inflation deploys the stent valve in the subannular position.
|
An alternative to the percutaneous access approach for the Cribier-Edwards bioprosthesis involves the use of a transthoracic, transapical device access strategy.31 A left anterolateral intercostal incision is used to expose the left ventricular apex. Using direct needle puncture, a 33F sheath is inserted into the left ventricle. The stent valve, which is identical to the percutaneous transfemoral version, is mounted in the antegrade direction on a shorter catheter and is positioned in either a catheterization laboratory or an operating room with the use of fluoroscopy, aortography, and echocardiography. This technique encourages a close collaboration between surgeons and interventionalists and avoids the sometimes forbidding iliofemoral arterial access anatomy that may preclude the percutaneous retrograde approach.
The first successful percutaneous aortic stent valve implantation was performed by Cribier on April 16, 2002, in a patient with critical aortic stenosis and cardiogenic shock, multiple comorbidities, and no therapy options because he was refused surgical AVR.28 The antegrade approach was used to implant a 23-mm bovine pericardial stent valve. The valve performed flawlessly, and the patient experienced dramatic hemodynamic improvement, but he died 17 weeks later after a lower limb amputation for peripheral vascular disease. After 3 additional "compassionate use" cases performed by Cribiers team in Rouen, France, a European registry was initiated for patients without surgical options (Initial Registry of EndoVascular Implantation of Valves in Europe (I-REVIVE)). This was followed by another single-center registry in Rouen that used the antegrade approach in high-risk patients who had been refused surgical AVR (Registry of Endovascular Critical Aortic Stenosis Treatment (RECAST)). Cribier recently reported the procedural results, clinical outcomes, and impressions from the 36 patients treated in his center who enrolled in these registries.32 In 3 patients, because of procedural complications, valve implantation was never attempted. Overall, successful stent-valve placement was achieved in 27 patients (75%); 23 patients were treated using the antegrade approach and 4 were treated using the retrograde approach. Reasons for failure to successfully implant the stent valve included hemodynamic instability, stent-valve embolization into the aorta after deployment, and failure to cross the native valve from the retrograde direction. Hemodynamic results after successful stent-valve implantation were uniformly excellent; echocardiography mean aortic valve gradients decreased from 37 to 9 mm Hg (P<0.0001) and mean aortic valve areas increased from 0.6 to 1.7 cm2 (P<0.0001). Ejection fraction increased from 45% pretreatment to 53% at 1 week (P=0.02). Postprocedural aortic regurgitation (moderate or severe) was present in 17 patients (63%) and was always paravalvular in origin. Aortic insufficiency was more common when the only available device size was 23 mm in diameter. Aortic insufficiency has been diminished with the availability of the larger, 26-mm diameter device and the use of additional balloon dilatation after initial prosthesis deployment when needed. Procedure-related in-hospital complications were disturbing, including 6 deaths and 1 stroke (26%). An additional 10 patients died in the ensuing 6 months. In all cases, out-of-hospital mortality was noncardiac and caused by comorbid conditions. Importantly, extended follow-up in the surviving patients shows continued excellent valve function and maintained symptom relief. The longest survivor from this series received the stent valve >3 years ago.
An important advance in transcatheter Cribier-Edwards AVR was led by Webb and colleagues, who refined the retrograde implantation approach with the use of improvements in equipment and procedural techniques. The availability of a 26-mm valve prosthesis has reduced the likelihood of valve embolization and the severity of paravalvular regurgitation. The use of a supportive tip-deflecting loading catheter assists with catheter navigation and native valve crossing. Webb et al reported the initial 18 cases performed with the retrograde technique.30 Fourteen successful stent valve implants occurred, with 2 procedure-related deaths (1 death as a result of complications from an iliac artery rupture and the other related to inadvertent obstruction of the left main coronary artery by the prosthesis). In the expanded patient series from Webb et al in Vancouver of >40 retrograde cases (John Webb, MD, unpublished observations, 2007), procedure-related 30-day mortality has been <10%. In the United States, a 3-center phase I safety and feasibility study, Percutaneous Endovascular Implantation of Valves (REVIVAL), is ongoing in patients with severe aortic stenosis and high–surgical risk characteristics as determined by a logistic EuroSCORE of >20% and independent surgical assessments of risk profile. Thus far, in the first 55 patients,33 in whom the retrograde approach was used, there have been 47 successful valve implants (87%) and 4 (7.3%) procedure-related deaths at 30 days (1 death was caused by procedural valve embolization, 1 death was caused by refractory heart failure 2 weeks after failed valve implantation, and the other 2 deaths were caused by complications of acute thoracic aorta dissection during failed valve crossing). Other periprocedural complications have included stroke (5 patients) and iliac artery rupture (5 patients). The major limitation of the current Cribier-Edwards aortic bioprosthesis resides in the large device profiles rendering transfemoral access treacherous in patients with diseased arterial vasculature. As in prior series, short-term hemodynamic results have been excellent and the frequency of moderate or severe paravalvular regurgitation has been markedly reduced (to <10%).
The major limitation of the current Cribier-Edwards aortic bioprosthesis is the large device profile. Nevertheless, the clinical trial process using this device has evolved significantly over the past 4 years and current clinical outcomes are sufficiently encouraging to justify a randomized clinical trial versus an appropriate control group in critical aortic stenosis patients who are at high risk for standard surgical AVR.
The CoreValve System
The ReValving System (CoreValve, Paris, France) consists of 3 components: a self-expanding frame with a trileaflet porcine pericardial valve, a delivery catheter that has been reduced in size from 25F to most recently 18F, and a loading system.34 The self-expanding nitinol support frame (total length 45 mm) has a diamond-cell configuration and incorporates 3 different areas of radial force (Figure 1, right). The upper-frame portion anchors in the aorta above the coronary sinuses with low radial force, the middle portion contains the valve leaflets, and the lower portion has high radial force and implants in the subannular region. Despite crossing the coronary ostia, the frame geometry is narrowed and unapposed to the aortic wall in its middle portion, thus allowing normal flow and free access to the coronary ostia through the stent struts. The single-layer porcine pericardial tissue valve is sewn to the frame with polytetrafluoroethylene (PTFE) sutures. The present system can be used in patients with distal aortas <45 mm in diameter and with aortic annulus sizes <23 mm in diameter. After balloon aortic valvuloplasty, the CoreValve ReValving system is positioned in a transaortic valve location and the overriding sheath is withdrawn over several minutes to expose and anchor the frame-valve unit. During the critical several minutes of valve deployment, percutaneous external hemodynamic support is used to maintain patient stability. Because of the self-expanding nature of the nitinol frame, continuous further expansion of the system occurs over 30 to 60 minutes, which provides apposition at the proximal and distal implantation zones, minimizing paravalvular regurgitation.
Since the initial report of the first successful implantation of the CoreValve stent valve prosthesis in 2005,35 significant progress has been made in device design, operator technique, and clinical outcomes. The largest reported clinical series of CoreValve cases36 includes 25 cases; the initial 10 cases used the first-generation 25F system (bovine pericardial valve material), and the subsequent 15 patients were treated with the second-generation 21F system (porcine pericardial valve material). The average baseline peak and mean transvalvular aortic pressure gradients were 69.3 mm Hg and 44.2 mm Hg, respectively, and after successful CoreValve implantation, gradients were reduced to 21.3 mm Hg and 12.4 mm Hg, respectively (both P<0.0001). Postprocedure aortic regurgitation grade was unchanged (0.9 preprocedure and 0.7 postprocedure), and no cases of moderate or severe paravalvular regurgitation occurred. Major in-hospital cardiovascular and cerebral events occurred in 8 (32%) patients, including death in 5 patients (20%). Among the 8 patients with device success who survived to discharge, no adverse events occurred within 30 days after discharge.
An informal report of the complete CoreValve clinical experience includes the aforementioned Siegburg series and the international ongoing multicenter trial (unpublished CoreValve company records, 2007). A total of >100 patients have been treated; 14 were treated with the 25F first-generation device and 63 were treated with the 21F second-generation device, and the remainder with the most recent 18F device. Patients were at high risk for surgical AVR (either compassionate use or logistic EuroSCORE >20%), and significant improvement occurred between first- and second-generation patient cohorts with device success increasing from 86% to 93% and 30-day death rate decreasing from 43% to 15%. The third-generation 18F device is now in use and promises to expand the treatment potential of this prosthesis. A limitation of the CoreValve approach has been the requirement for extracorporeal cardiopulmonary support during deployment. An alternative has been the use of a percutaneous left ventricular assist device for ventricular unloading, which obviates the need for ventilatory support and extracorporeal oxygenation. The lower-profile 18F device may not require as much dependence on temporary cardiopulmonary support, thus simplifying the procedure.
Other New Technologies
Several other new percutaneous AVR technologies are being developed. These new devices offer creative design features and attempt to address some of the deficiencies of the first generation technologies. Common characteristics include retrograde valve crossing, reduced profiles, pericardial tissue valve leaflets, and the ability to retrieve and reposition the device before final deployment and release. Some of the most promising of these new aortic valve technologies are described below.
The Direct Flow percutaneous aortic valve (Santa Rosa, Calif) is a stentless, inflatable, fabric-cuff, equine pericardial tissue valve. After advancement across the diseased native aortic valve, the distal ring is positioned and inflated with contrast to secure fixation. Tethers are used to align the valve correctly and the proximal ring is then deployed. Once valve position, function, and sealing are confirmed, a permanent polymer is infused to replace the contrast, followed by detachment of the control tethers (Figure 2, left).
|
The Lotus percutaneous aortic valve by Sadra Medical (Saratoga, Calif) comprises a nitinol continuous braid frame with a bovine pericardial trileaflet tissue valve (Figure 2, middle). The 19F outer diameter catheter is delivered across the aortic valve and unsheathed. The self-expanding nitinol prosthesis passively shortens and self-centers with low radial force, which allows the valve to begin functioning. When it is optimally located, the nitinol frame is actively shortened and locked to its final height (19 mm), which increases the radial force and secures the bioprosthesis. The frame-valve assembly is attached to the catheter deployment system by 15 arms and the device can be re-elongated, retrieved, and repositioned at any time before final release from the catheter.
Finally, the AorTx percutaneous aortic valve (Palo Alto, Calif) consists of a low-profile, folded, metallic frame that incorporates a pericardial tissue valve (Figure 2, right). The frame is positioned, springs open to unfold the trileaflet valve, and with high radial force is securely implanted. As with the previously mentioned designs, the system is fully retrievable and can be repositioned before final detachment from the catheter delivery system.
Transcatheter Aortic Valve Replacement Clinical Trial Considerations
Given the success of surgical AVR for normal-risk patients, the focus for new transcatheter AVR technologies has been on high–surgical risk patients. Optimal characterization of patient risk requires a combination of clinical judgment and application of standard quantitative risk assessment algorithms. The 2 most commonly applied risk assessment tools are the EuroSCORE and the Society of Thoracic Surgeons risk scoring systems. In general, high surgical risk for AVR is defined as the upper 10% risk decile or, alternatively, as a 30-day procedure death rate in excess of 15%. Less tangible risks, such as patient frailty, chest-wall pathology, and thoracic aorta calcification are often omitted from these risk assessment algorithms and must be considered separately when overall patient risk is determined.
Undoubtedly, successful validation of transcatheter AVR as a meaningful clinical procedure will require appropriately designed randomized clinical trials. A template clinical trial pathway might involve randomization of high surgical risk patients to either transcatheter versus surgical AVR by use of a noninferiority methodology and a primary end point of all-cause death at 1 year. In critical aortic stenosis patients deemed inoperable by current standards, a different randomized clinical trial design would be appropriate, using best medical therapy and/or balloon aortic valvuloplasty as the control arm (versus transcatheter AVR). Nevertheless, secondary quality-of-life measures and other death end points would be important in such clinical trials, and long-term follow-up should be required to assess valve durability of these new devices. In the future, if transcatheter AVR is proven to be safe and effective in the high surgical risk patients, additional clinical trials comparing transcatheter AVR to analogous surgical procedures in lower risk aortic stenosis patients, patients with aortic stenosis and coronary disease, and patients with aortic regurgitation may be considered.
| Surgical Mitral Valve Repair |
|---|
|
|
|---|
| Transcatheter Mitral Valve Therapies |
|---|
|
|
|---|
Surgical edge-to-edge repair has had mixed clinical results.43,44 It has been used as a bail-out procedure in cases of both functional and degenerative MR when more conventional surgical approaches had suboptimal outcomes. Isolated edge-to-edge surgical repair in a patient cohort with optimal leaflet morphology had 90% freedom from reoperation and recurrent MR >2+ at 5 years, and almost 80% freedom from reoperation and recurrent MR >2+ after 12 years.45 This latter report clearly demonstrates that isolated surgical edge-to-edge repair can be durable in selected patients.
Edge-to-edge repair has been duplicated using percutaneous clip- and suture-based devices.46–50 After transseptal puncture, MitraClip (Evalve, San Francisco, Calif) is delivered to the left atrium via a 24F guide catheter and positioned in the mid–left atrial cavity above the mitral orifice (Figure 3). The clip must be aligned in the center of the valve orifice, with the clip arms perpendicular to the line of coaptation. The process of steering the MitralClip guide catheter into optimal position is accomplished using steering knobs on the guide catheter and clip-delivery catheter, using both fluoroscopic and transesophageal echocardiographic guidance. When the clip is centered above the origin of the regurgitant jet along the line of leaflet coaptation, the clip is opened. The opened clip arms are passed through the mitral orifice; the open arms minimize the chance for chordal entanglement. After the clip is passed into the left ventricle below the mitral leaflets, it is pulled back, the leaflets are grasped, and the clip arms are closed to create a double orifice. The device can be repositioned if control of the MR is not adequate and removed if it appears to be unsuccessful. A second clip can also be placed if a first clip appears inadequate in decreasing the magnitude of MR.
|
This device approach has been successfully used in a phase I clinical trial in the United States, with results at 6 and 12 months reported recently.51,52 Surgical candidates with moderately severe or severe MR and cardiac symptoms or no symptoms with signs of left ventricular dysfunction were included in the clinical Endovascular Valve Edge-to-Edge Repair Study (EVEREST-I). Patients fulfilled the American Heart Association/American College of Cardiology guidelines criteria for surgical treatment of MR,12 and echocardiograms were evaluated using the American Society for Echocardiography methods for assessment of MR severity.53 Mitral leaflet morphology and MR jet origin must be suited to this approach. The regurgitant jet must arise from the central two thirds of the line of coaptation. Leaflet coaptation length and depth must be
2 mm and
11 mm, respectively. When flail segments are present, the flail gap must be <10 mm and the flail width <15 mm. These rigorous clinical and morphological criteria effectively exclude patients with severe annular dilatation. Less than 20% of echocardiograms evaluated by the core laboratory are considered appropriate for treatment with the MitralClip device.
A total of 70 patients were enrolled in this phase I trial and in the run-in portion of the subsequent EVEREST-II trial (see below) with >6 months follow-up in 27 patients. Compared with results from the most recent Society of Thoracic Surgeons database, patients referred for this percutaneous procedure were significantly older; median age was 71 years for the clip procedure compared with 59 years for surgical repairs. In these 70 patients, clips were successfully implanted in 90% and no intraprocedural major complications occurred. Acute procedure success, defined as successful clip placement with reduction in MR severity to
2+, was 73%. Major adverse events within 30 days included partial clip detachment without embolization in 7% of patients, all of whom underwent successful elective valve surgery, and a postprocedure stroke in 1 patient, which resolved in 1 month. Average length of hospital stay was <2 days. Even when a clip was placed and the results were suboptimal and required surgery afterward, mitral leaflet repair using standard surgical techniques has been possible as late as 18 months after the index interventional procedure.54 Actuarial 2-year freedom from death, mitral valve surgery, or recurrent MR >2+ has been 80% among patients discharged with successful clip therapy.52 The mitral clip procedure may be difficult to generalize to most interventional operators, as a significant operator learning curve is required to reduce procedure time and achieve consistent results, in addition to a familiarity with transesophageal echocardiographic techniques for therapy guidance.
The encouraging success of the Evalve clip procedure, particularly the favorable procedural safety results, has led to a randomized trial comparing this device with mitral valve surgery in the United States. EVEREST II is currently randomizing patients to percutaneous repair versus a standard surgical approach (2:1 allocation), with clinical and echocardiographic safety and efficacy end points. Importantly, in the surgical literature there has never been a prospective, echocardiography core laboratory–evaluated, intention-to-treat trial of mitral valve repair therapy. The EVEREST II trial will be important not only in the assessment of a new percutaneous mitral valve therapy but also in defining the contemporary results of surgery for mitral valve disease.
The edge-to-edge mitral valve repair can also be accomplished using the Mobius percutaneous suture device (Edwards Lifesciences Inc., Orange, Calif). A 16F guide catheter is used to deliver a 10F therapy catheter. The therapy catheter uses suction to capture and immobilize a portion of the mitral leaflet, and then deliver a suture. The device is rotated and the procedure is repeated on the second leaflet. A nitinol fastener is delivered over the suture and the suture is clipped. Early first-in-man experiences that use this suture-based edge-to-edge approach at multiple investigator sites are accumulating. Percutaneous suture-based edge-to-edge mitral valve repair has been successful in reducing MR and creating a double-orifice mitral inlet, but additional device enhancements and refinement in patient selection and operator technique are required to achieve consistent procedure results.
Coronary Sinus Annuloplasty
The mainstay of surgical therapy for MR has been ring reduction annuloplasty, either as a stand-alone treatment for MR or in conjunction with mitral leaflet repair. A simplified interventional approach to simulate surgical annuloplasty has been to work from within the coronary sinus to geometrically deform the anteroposterior dimension of the mitral annulus. This procedure is dependent on the anatomic juxtaposition of the mitral annulus and the circumnavigating coronary sinus.55 Anchors or stents can be placed in both the distal coronary sinus or great cardiac vein and in the coronary sinus ostium, with a bridging connector, which either immediately or over time constrains the coronary sinus and reduces the cross-sectional area of the mitral annulus, thereby improving MR.
The Monarc device (Edwards Lifesciences Inc.) has been implanted in >80 patients outside the United States. The coronary sinus and anterior interventricular vein are cannulated via the right internal jugular vein with deployment of distal and proximal self-expanding stent anchors that are separated by a connecting bridge element. The connecting bridge is a coiled spring, held in an open position by biodegradable material in the spring interstices. Thus, tension on the coronary sinus (and mitral annulus) develops over 3 to 6 weeks as the spring shortens as a result of biodegradation of the embedded material. This device is in early phase I clinical trials, with published reports on only the first few patients treated. In the earliest experience, bridge fractures between the 2 anchors occurred in 3 of the 4 implanted patients (detected at days 22, 28, and 81 after device implantation) but were not associated with clinical sequelae other than worsening MR.56 After a redesign of the bridge element, >80 additional procedures have been performed without fractures or other device failures. Importantly, because tension on the mitral annulus is slow and gradual (coincident with bridge shortening), the efficacy in reducing MR can only be determined during follow-up assessments.
The Carillon mitral contour system (Cardiac Dimensions, Kirkland, Wash) consists of a nitinol wire shaping ribbon between proximal and distal anchors that are placed in the coronary sinus.57–59 Tension applied to the shaping ribbon between the 2 anchors deforms the coronary sinus geometry, thereby reducing the mitral annulus dimensions (Figure 4). The Carillon device, delivered via a transjugular puncture and requiring a 9F guiding catheter, is progressively shortened before final deployment of the proximal anchor on the basis of the reduction in MR measured using echocardiography. This device has been inserted successfully in a small number of patients both outside and inside the United States and has already undergone a device redesign to improve coronary sinus securement and fixation. The first treated patient had ischemic MR, with sustained reduction of MR after 2 years. A continuation of the first-in-humans clinical trial is underway.
|
The percutaneous transvenous mitral annuloplasty system (Viacor, Wilmington, Mass) is the third in this genre of coronary sinus shape deforming devices. Percutaneous transvenous mitral annuloplasty was invented by cardiac surgeons and consists of a 7F multilumen polytetrafluoroethylene catheter, within which variable stiffness rods are inserted. The rods deform the shape of the midportion of the coronary sinus, which diminishes the septal to lateral dimension of the mitral annulus and reduces the severity of MR in animal models.60 After the optimal number and stiffness of rods have been inserted in a temporary diagnostic catheter, a permanent version of the device is implanted. Importantly, the system shape and stiffness can be adjusted over time by addition or substitution of rods, depending on the patient response and changes in the severity of MR. A small series of temporary implants were inserted in the operating room to establish proof of concept. Presently, a 30-patient clinical trial is ongoing at 3 centers outside the United States.
Another similar coronary sinus approach is the so-called percutaneous septal-sinus shortening procedure. An anchor is placed in the coronary sinus and a cord traversing the left atrium is attached to a septal occluder in the fossa ovalis and tensioned to modify mitral annulus geometry.61 This system is in the process of initiating first-in-humans clinical trials.
Important limitations may be associated with coronary sinus annuloplasty as a methodology to reduce MR. The coronary sinus does not directly parallel the mitral annulus in many patients, but rather is positioned about 1 cm cranial and somewhat tangential to the annulus plane.62 Thus, tension created within the coronary sinus is transmitted imprecisely and indirectly to the annulus and might dissipate into the surrounding tissues over time, reducing the efficacy of MR reduction. Moreover, the coronary sinus crosses over branches of the circumflex coronary artery in approximately half of the patients. It remains unclear whether any of the aforementioned devices will induce important circumflex artery compression and ischemia. The coronary sinus has become a frequently utilized anatomic space over the past several years and it would be important to maintain access for lead placements during procedures such as resynchronization therapy. Finally, issues of erosion and thrombosis of the coronary sinus with these devices can only be ascertained after increased clinical experiences. Despite these potential disadvantages, if a safe and predictable device reduces MR by use of a simple transvenous coronary sinus implant, this would become a worthwhile therapy for many patients.
Direct Mitral Annuloplasty
To overcome some of the potential limitations of indirect annuloplasty via the coronary sinus, direct approaches to the mitral annulus are being developed. The Mitralign device (Mitralign, Tewksbury, Mass) uses anchor pledgets that are placed directly into the mitral annulus, and a drawstring to cinch the annulus in a manner analogous to surgical plication annuloplasty (Figure 5). A relatively small (20%) reduction of the posterior annulus can normalize the septal–lateral dimension and eliminate ischemic MR.63 The surgical version of this approach was initially described in 1977,64 and acceptable clinical surgical results have been reported recently.65,66 The Mitralign annuloplasty system places anchors directly into the mitral annulus from the left ventricular side and tethers them with a plication suture. Transventricular access to the periannular space on the left ventricular side of the posterior leaflet is achieved with retrograde left ventricular catheterization by use of standard guiding catheter shapes. Clinical studies with the Mitralign device will commence in the next several months.
|
Transpericardial Annulus and Left Ventricular Remodeling
The Coapsys surgical system (Myocor, Maple Grove, Minn) uses pads placed on either side of the ventricle with a cord passing through the left ventricular cavity to apply tension to both the mitral annulus and the basal left ventricular cavity (Figure 6).67 Uniquely, this off-pump surgical procedure is a direct approach to achieve both left ventricular remodeling and an associated mitral annuloplasty. Initial results of the Coapsys surgical system implanted in patients with ischemic MR during coronary revascularization, have shown sustained reductions in MR and improved ventricular chamber dimensions for as long as 1 year after the procedure. A percutaneous transpericardial method to simulate this surgical procedure (iCoapsys) is under development in preclinical models.68
|
Percutaneous Mitral Repair Trial Considerations
Patient selection and trial design for the development of these novel technologies remains challenging.69 There have been no prior randomized trials and little organized follow-up to evaluate surgical approaches to treat MR. Thus, no established methods exist to compare percutaneous and surgical valve therapy approaches. Percutaneous therapy is inherently different, and the standards for measuring clinical outcomes after percutaneous therapy may not be comparable to surgery. Clinical trials designed to demonstrate noninferiority for efficacy end points and superiority for safety end points may be reasonable goals for catheter-based versus surgical mitral valve procedures.
It has been demonstrated that asymptomatic patients with severe MR have a poor prognosis compared with those with less severe MR.70 A clinical trial is urgently needed to demonstrate that early mitral valve therapy might benefit these patients. The lower morbidity of percutaneous therapies may be well suited for such an asymptomatic, lower-risk population. Alternatively, evidence exists to indicate favorable outcomes with watchful waiting in this patient group; thus, a randomized clinical trial versus either medical or surgical therapy would be critical to evaluate the utility of any therapy in this patient population.71 Indirect or direct annuloplasty approaches are better suited to patients with functional MR as a result of heart failure or coronary ischemia. This population is often not treated surgically, and comparisons with medical therapy may be more appropriate.
A paucity of data on the results of mitral repair surgery complicates comparisons with catheter-based technologies. There have been no prospective intention-to-treat trials to define the rate of prosthetic replacement when valve repair is intended, and this conversion rate will impact the relative merits of surgical and percutaneous methods. The usual end point in long-term surgical reports is freedom from reoperation rather than the combination of freedom from death, reoperation, and recurrent MR. Patient selection and surgical outcomes have never been reported with the use of rigorous intention-to-treat principles, core laboratory evaluation of baseline echocardiograms, use of American Society for Echocardiography criteria for grading severity of MR, or prospective echocardiographic follow-up.
Combinations of mitral annuloplasty and leaflet repair approaches may ultimately be necessary for optimal percutaneous therapy in some patients with MR. The use of device combinations will be extremely difficult to evaluate from a clinical trial perspective. The potential for novel nondevice therapies is also just being recognized. Recently, autologous myoblast transplantation that results in left ventricular remodeling has been used to reduce MR in a sheep myocardial infarction model designed to induce chronic ischemic MR.72
| Conclusions |
|---|
|
|
|---|
| Acknowledgments |
|---|
Dr Feldman is on the scientific advisory board of Myocor and has received grant suport from Cardiac Dimensions, Edwards Lifesciences, and Myocor. He is principal investigator for the Evalve EVEREST trial. Dr Leon is on the scientific advisory boards of Edwards Lifesciences, Guided Delivery Systems, Mitralign, and Sadra Medical. He has received grant support from Edwards Lifesciences and owns equity in Guided Delivery Systems, Mitralign, and Sadra Medical. He is principal investigator for the Edwards Lifesciences PARTNER trial.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
2. Andersen HR, Knudsen LL, Hasenkam JM. Transluminal implantation of artificial heart valves. Description of a new expandable aortic valve and initial results with implantation by catheter technique in closed-chest pigs. Eur Heart J. 1992; 13: 704–708.
3. Alfieri O, Maisano F, Colombo A. Future of transcatheter repair of the mitral valve. Amer J Cardiol. 2005; 96: 71–75.[Medline] [Order article via Infotrieve]
4. Liddicoat JR, MacNeill BD, Gillinov AM, Cohn WE, Chin CH, Prado AD, Pandian NG, Oesterle SN. Percutaneous mitral valve repair: a feasibility study in an ovine model of acute ischemic mitral regurgitation. Catheter Cardiovasc Interv. 2003; 60: 410–416.[CrossRef][Medline] [Order article via Infotrieve]
5. Iung B, Baron G, Butchart EG, Delahaye F, Gohlke-Barwolf C, Levang OW, Tornos P, Vanoverschelde JL, Vermeer F, Boersma E, Ravaud P, Vahanian A. A prospective survey of patients with valvular heart disease in Europe: the Euro Heart Survery on Valvular Heart Disease. Eur Heart J. 2003; 24: 1231–1243.
6. Varadarajan P, Kapoor N, Bansal RC, Pai RG. Clinical profile and natural history of 453 nonsurgically managed patients with severe aortic stenosis. Ann Thorac Surg. 2006; 82: 2111–2115.
7. Bando K, Danielson GK, Schaff HV, Mair DD, Julsrud PR, Puga FJ. Outcome of pulmonary and aortic homografts for right ventricular outflow tract reconstruction. J Thorac Cardiovasc Surg. 1995; 109: 509–519.
8. Sugiyama H, Williams W, Benson LN. Implantation of endovascular stents for the obstructive right ventricular outlow tract. Heart. 2005; 91: 1058–1063.
9. Bonhoeffer P, Boudjemline Y, Saliba Z, Merckx J, Aggoun Y, Bonnet D, Acar P, Le Bidois J, Sidi D, Kachaner J. Percutaneous replacement of pulmonary valve in a right-ventricle to pulmonary-artery prosthetic conduit with valve dysfunction. Lancet. 2000; 356: 1403–1405.[CrossRef][Medline] [Order article via Infotrieve]
10. Coats L, Khambadkone S, Derrick G, Sridharan S, Schievano S, Mist B, Jones R, Deanfield JE, Pellerin D, Bonhoeffer P. Taylor AM. Physiological and clinical consequences of relief of right ventricular outflow tract obstruction late after repair of congenital heart defects. Circulation. 2006; 113: 2037–2044.
11. Khambadkone S, Coats L, Taylor A, Boudjemline Y, Derrick G, Tsang V, Cooper J, Muthurangu V, Hegde SR, Razavi RS, Pellerin D, Deanfield J, Bonhoeffer P. Percutaneous pulmonary valve implantation in humans: results in 59 consecutive patients. Circulation. 2005; 112: 1189–1197.
12. Bonow O, Carabello B, Chaterjee K, DeLeon AC Jr, Faxon DP, Freed MD, Gaasch WH, Lytle BW, Nishimura RA, OGara PT, ORourke RA, Otto CM, Shah PM, Shanewise J. ACC/AHA 2006 Guidelines for the Management of Patients with Valvular Heart Disease. Executive Summary. Circulation. 2006; 114: 450–527.
13. Schwarz F, Baumann P, Manthey J, Hoffmann M, Schuler G, Mehmel HC, Schmitz W, Kubler W. The effects of aortic valve replacement on survival. Circulation. 1982; 66: 1105–1110.
14. Smith N, McAnulty JH, Rahimtoola SH. Severe aortic stenosis with impaired left ventricular function and clinical hear failure: results of valve replacement. Circulation. 1978; 58: 255–264.
15. Kvidal P, Bergstrom R, Horte LG, Stahle E. Observed and relative survival after aortic valve replacement. J Am Coll Cardiol. 2000; 35: 747–756.
16. Edwards FH, Peterson ED, Coombs LP, DeLong ER, Jamieson WR, Shroyer ALW, Grover FL. Prediction of operative mortality after valve replacement surgery. J Am Coll Cardiol. 2001; 37: 885–892.
17. Society of Thoracic Surgeons National Cardiac Surgery Database. Available at: http://www.sts.org/documents/pdf/Spring2005STS-ExecutiveSummary.pdf. Accessed November 2005.
18. Birkmeyer JD, Siewers AE, Finlayson EV, Stukel TA, Lucas FL, Batista I, Welch HG, Wennberg DE. Hospital volume and surgical mortality in the United States. N Engl J Med. 2002; 346: 1128–1137.
19. Goodney PP, OConnor GT, Wennberg DE, Birkmeyer JD. Do hospitals with low mortality rates in coronary artery bypass also perform well in valve replacement? Ann Thorac Surg. 2003; 76: 1131–1136.
20. Levinson JR, Akins CW, Buckley MJ, Newell JB, Palacios IF, Block PC, Fifer MA. Octogenarians with aortic stenosis. Outcome after aortic valve replacement. Circulation. 1989; 80: I49–I56.[Medline] [Order article via Infotrieve]
21. Kolh P, Kerzmann A, Lahaye L, Gerard P, Limet R. Cardiac surgery in octogenarians; peri-operative outcome and long-term results. Eur Heart J. 2001; 22: 1235–1243.
22. Powell DE, Tunick PA, Rosenzweig BP, Freedberg RS, Katz ES, Applebaum RM, Perez JL, Kronzon I. Aortic valve replacement in patients with aortic stenosis and severe left ventricular dysfunction. Arch Intern Med. 2000; 160: 1337–1341.
23. Connolly HM, Oh JK, Schaff HV, Roger VL, Osborn SL, Hodge DO, Tajik AJ. Severe aortic stenosis with low transvalvular gradient and severe left ventricular dysfunction: result of aortic valve replacement in 52 patients. Circulation. 2000; 101: 1940–1946.
24. Pereira JJ, Lauer MS, Bashir M, Afridi I, Blackstone EH, Stewart WJ, McCarthy PM, Thomas JD, Asher CR. Survival after aortic valve replacement for severe aortic stenosis with low transvalvular gradients and severe left ventricular dysfunction. J Am Coll Cardiol. 2002; 39: 1356–1363.
25. Goodney PP, Stukel TA, Lucas FL, Finlayson EV, Birkmeyer JD. Hospital volume, length of stay, and readmission rates in high-risk surgery. Ann Surg. 2003; 238: 161–167.[CrossRef][Medline] [Order article via Infotrieve]
26. Sundt TM, Bailey MS, Moon MR, Mendeloff EN, Huddleston CB, Pasque MK, Barner HB, Gay WA Jr. Quality of life after aortic valve replacement at the age of >80 years. Circulation. 2000; 102: III70–III74.[Medline] [Order article via Infotrieve]
27. Zimpfer D, Czerny M, Kilo J, Kasimir MT, Madl C, Kramer L, Wieselthaler GM, Wolner E, Grimm M. Cognitive deficit after aortic valve replacement. Ann Thorac Surg. 2002; 74: 407–412.
28. Cribier A, Eltchaninoff H, Bash A, Borenstein N, Tron C, Bauer F, Derumeaux G, Anselme F, Laborde F, Leon MB. Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis: first human case description. Circulation. 2002; 106: 3006–3008.
29. Eltchaninoff H, Tron C, and Cribier A. Percutaneous implantation of aortic valve prosthesis in patients with calcific aortic stenosis: technical aspects. J Interv Cardiol. 2003; 16: 515–521.[CrossRef][Medline] [Order article via Infotrieve]
30. Webb JG, Chandavimol M, Thompson CR, Ricci DR, Carere RG, Munt BI, Buller CE, Pasupati S, Lichtenstein S. Percutaneous aortic valve implantation retrograde from the femoral artery. Circulation. 2006; 113: 842–850.
31. Lichtenstein SV, Cheung A, Ye J, Thompson CR, Carere RG, Pasupati S, Webb JG. Transapical transcatheter aortic valve implantation in humans: initial clinical experience. Circulation. 2006; 114: 591–596.
32. Cribier A, Eltchaninoff H, Tron C, Bauer F, Agatiello C, Nercolini D, Tapiero S, Litzler PY, Bessou JP, Babaliaros V. Treatment of calcific aortic stenosis with the percutaneous heart valve. Mid-term follow-up from the initial feasibility studies: the French experience. J Am Coll Cardiol. 2006; 47: 1214–1223.
33. Kodali, SK, ONeill W, Moses JW, Kapadia S, Williams M, Hanzel G, Stewart A, Tuzcu ME, Collins M, Leon MB. Preliminary Results from the PeRcutaneous EndoVascular Implantation of VALves Trial in High Risk Patients with Critical Aortic Stenosis. Am J Cardiol. 2006; 98 (suppl 1); 48M.
34. Laborde JC, Borenstein N, Behr L, Farah B, Fajadet J. Percutaneous implantation of the Corevalve aortic valve prosthesis for patients presenting high risk for surgical valve replacement. Eurointervention. 2006; 1: 472–474.
35. Grube E, Laborde JC, Zickmann B, Gerckens U, Felderhoff T, Sauren B, Bootsveld A, Buellesfeld L, Iversen S. First report on a human percutaneous transluminal implantation of a self-expanding valve prosthesis for interventional treatment of aortic valve stenosis. Catheter Cardiovasc Interv. 2005; 66: 465–469.[CrossRef][Medline] [Order article via Infotrieve]
36. Grube E, Laborde JC, Gerckens U, Felderhoff T, Sauren B, Buellesfeld L, Mueller R, Menichelli M, Schmidt T, Zickmann B, Iversen S, Stone GW. Percutaneous implantation of the CoreValve self-expanding valve prosthesis in high-risk patients with aortic valve disease: the Siegburg first-in-man study. Circulation. 2006; 114: 1616–1624.
37. Harris KM, Sundt TM III, Aeppli D, Sharma R, Barzilai B. Can late survival of patients with moderate ischemic mitral regurgitation be impacted by intervention on the valve? Ann Thorac Surg. 2002; 74: 1468–1475.
38. Diodato MD, Moon MR, Pasque MK, Barner HB, Moazami N, Lawton JS, Bailey MS, Guthrie TJ, Meyers BF, Damiano RJ Jr. Repair of ischemic mitral regurgitation does not increase mortality or improve long-term survival in patients undergoing coronary artery revascularization: a propensity analysis. Ann Thorac Surg. 2004; 78: 794–799.
39. Wu AH, Aaronson KD, Bolling SF, Pagani FD, Welch K, Koelling TM. Impact of mitral valve annuloplasty on mortality risk in patients with mitral regurgitation and left ventricular systolic dysfunction. J Am Coll Cardiol. 2005; 45: 381–387.
40. Enomoto Y, Gorman JH 3rd, Moainie SL, Guy TS, Jackson BM, Parish LM, Plappert T, Zeeshan A, St John-Sutton MG, Gorman RC. Surgical treatment of ischemic mitral regurgitation may not influence ventricular remodeling. J Thorac Cardiovasc Surg. 2005; 129: 504–511.
41. Alfieri O, Elefteriades JA, Chapolini RJ, Steckel R, Allen WJ, Reed SW, Shreck S. Novel suture device for beating-heart mitral leaflet approximation. Ann Thorac Surg. 2002; 74: 1488–1493.
42. Alfieri O, Maisano F, DeBonis M, Stefano PL, Torracca L, Oppizzi M, LaCanna G. The edge-to-edge technique in mitral valve repair: a simple solution for complex problems. J Thorac Cardiovasc Surg. 2001; 122: 674–681.
43. Bhudia SK, McCarthy PM, Smedira NG, Lam BK, Rajeswaran J, Blackstone EH. Edge-to-edge (Alfieri) mitral repair: results in diverse clinical settings. Ann Thorac Surg. 2004; 77: 1598–1606.
44. Kherani AR, Cheema FH, Casher J, Fal JM, Mutrie CJ, Chen JM, Morgan JA, Vigilance DW, Garrido MJ, Smith CR, Oz MC. Edge-to-edge mitral valve repair: the Columbia Presbyterian experience. Ann Thorac Surg. 2004; 78: 73–76.
45. Maisano F, Vigano G, Blasio A, Columbo A, Calabrese C, Alfieri O. Surgical isolated edge-to-edge mitral repair without annuloplasty-clinical proof of principle for an endovascular approach. Eurointervention. 2006; 2: 181–186.
46. St Goar FG, James FI, Komtebedde J, Foster E, Oz MC, Fogarty TJ, Feldman T, Block PC. Endovascular edge-to-edge mitral valve repair: short-term results in a porcine model. Circulation. 2003; 108: 1990–1993.
47. Fann JI, St Goar FG, Komtebedde J, Oz MC, Block PC, Foster E, Feldman T, Burdon TA. Off-pump edge-to-edge mitral valve technique using a mechanical clip in a chronic model. Circulation. 2003; 108 (supp IV): 493. Abstract.
48. St Goar FG, Fann JI, Feldman TE, Block PC, Herrmann HC. Percutaneous mitral valve repair with the edge-to-edge technique. In: Herrman HC, ed. Contemporary Cardiology: Interventional Cardiology-Percutaneous Noncoronary Intervention. Totowa, NJ. Humana Press; 2005: 87–95.
49. Herrmann HC, Feldman T. Percutaneous mitral valve edge-to-edge repair with the Evalve Mitraclip system: rationale and phase I results. EuroIntervention. 2006; 1 (suppl A): A36–A39.
50. Feldman T, Alfieri O, St Goar F: Percutaneous leaflet repair for mitral regurgitation using the Evalve edge-to-edge clip technique. In: Hijazi Z, Bonhoeffoer P, Feldman T, Ruiz C, Martin Dunitz, eds. Transcatheter Valve Repair. London, United Kingdom. Parthenon Publishing/Taylor & Francis Medical Books; 2006: 275–284.
51. Feldman T, Wasserman HS, Herrmann HC, Gray W, Block PC, Whiltlow PL, St Goar F, Rodriguez L, Silvestry F, Schwartz A, Sanborn TA, Condado JA, Foster E. Percutaneous mitral valve repair using the edge-to-edge technique: six-month results of the EVEREST phase I clinical trial. J Am Coll Cardiol. 2005; 46: 2134–2140.
52. Feldman T, Wasserman HS, Herrmann HC, Whitlow PL, Block PC, Gray WA, Foster E, St Goar F: Edge-to-edge mitral valve repair using the Evalve MitraClip: One year results of the EVEREST phase I clinical trial. Am J Cardiol. 2005; 96 (supp): 49H.[CrossRef]
53. Zoghbi WA, Enriquez-Sarano M, Foster E, Grayburn PA, Kraft CD, Levine RA, Nihoyannopoulos P, Otto CM, Quinones MA, Rakowski H, Stewart WJ, Waggoner A, Weissman NJ. Recommendations for evaluation of the severity of native valvular regurgitation with two-dimensional and Doppler echocardiography. J Am Soc Echocardiogr. 2003; 16: 777–802.[CrossRef][Medline] [Order article via Infotrieve]
54. Dang NC, Aboodi MS, Sakaguchi T, Wasserman HS, Argenziano M, Cosgrove DM, Rosengart TK, Feldman T, Block PC, Oz MC. Surgical revision after percutaneous mitral valve repair with a clip: initial multicenter experience. Ann Thorac Surg. 2005; 80: 2338–2342.
55. Yamanouchi Y, Igawa O, Hisatome I. Activation mapping from the coronary sinus may be limited by anatomic variations. Pacing Clin Electrophysiol. 1998; 21: 2522–2526.[CrossRef][Medline] [Order article via Infotrieve]
56. Webb JG, Harnek J, Munt BI, Kimblad PO, Chandavimol M, Thompson CR, Solem JO. Percutaneous transvenous mitral annuloplasty; initial human experience with device implantation in the coronary sinus. Circulation. 2006; 113: 851–855.
57. Kaye DM, Byrne M, Alferness C, Power J. Feasibility and short-term efficacy of percutaneous mitral annular reduction for the therapy of heart failure—induced mitral regurgitation. Circulation. 2003; 108: 1795–1797.
58. Maniu CV, Patel JB, Reuter DG, Meyer DM, Edwards WD, Rihal CS, Redfield MM. Acute and chronic reduction of functional mitral regurgitation in experimental heart failure by percutaneous mitral annuloplasty. J Am Coll Cardiol. 2004; 44: 1652–1661.
59. Byrne MJ, Kaye DM, Mathis M, Reuter DG, Alferness CA, Power JM. Percutaneous mitral annular reduction provides continued benefit in an ovine model of dilated cardiomyopathy. Circulation. 2004; 110: 3088–3092.
60. Liddicoat JR, MacNeill BD, Gillinov AM, Cohn WE, Chin CH, Prado AD, Pandian NG, Oesterle SN. Percutaneous mitral valve repair: a feasibility study in an ovine model of acute ischemic mitral regurgitation. Catheter Cardiovasc Interv. 2003; 60: 410–416.[CrossRef][Medline] [Order article via Infotrieve]
61. Rogers JH, Macoviak JA, Rahdert DA, Takeda PA, Palacios IF, Low RI. Percutaneous septal sinus shortening: a novel procedure for the treatment of functional mitral regurgitation. Circulation. 2006; 113: 2329–2334.
62. Maselli D, Guarracino F, Chiaramonti F, Mangia F, Borelli G, Minzioni G. Percutaneous mitral annuloplasty: an anatomic study of human coronary sinus and its relation with mitral valve annulus and coronary arteries. Circulation. 2006; 114: 377–380.
63. Tibayan FA, Rodriguez F, Liang D, Daughters GT Ingels NB Jr, Miller DC. Paneth suture annuloplasty abolishes acute ischemic mitral regurgitation but preserves annular and leaflet dynamics. Circulation. 2003; 108 (suppl 1): II-128–II-33.[Medline] [Order article via Infotrieve]
64. Burr LH, Krayenbuhl C, Sutton MS. The mitral plication suture: a new technique of mitral valve repair. J Thorac Cardiovasc Surg. 1977; 73: 589–595.[Abstract]
65. Nagy ZL, Bodi A, Vazily M, Szerafin T, Horvath A, Peterffy A. Five-year experience with a suture annuloplasty for mitral valve repair. Scand Cardiovasc J. 2000; 34: 528–532.[CrossRef][Medline] [Order article via Infotrieve]
66. Aybek T, Risteski P, Miskovic A, Simon A, Dogan S, Abdel-Rahman U, Moritz A. Seven years experience with suture anuloplasty for mitral valve repair. J Thorac Cardiovasc Surg. 2006; 101: 99–106.
67. Grossi EA, Saunders PC, Woo J, Gangahar DM, Laschinger JC, Kress DC, Caskey MP, Schwartz CF, Wudel J. Intraoperative effects of the Coapsys annuloplasty system in a randomized evaluation (RESTOR-MV) of functional ischemic mitral regurgitation. Ann Thorac Surg. 2005; 80: 1706–1711.
68. Pedersen WR, Block P, Feldman T. The iCoapsys repair system for the percutaneous treatment of functional mitral insufficiency. Eurointervention. 2006; 1 (suppl A): A44–A48.
69. Vassiliades TA Jr, Block PC, Cohn LH, Adams DH, Borer JS, Feldman T, Holmes DR, Laskey WK, Lytle BW, Mack MJ, Williams DO. The clinical development of percutaneous heart valve technology: an interdisciplinary position statement. The Society of Thoracic Surgeons (STS), The American Association for Thoracic Surgery (AATS), The American College of Cardiology (ACC), and The Society of Cardiovascular Angiography and Intervention (SCAI). J Am Coll Cardiol. 2005; 49: 1554–1560.
70. Enriquez-Sarano M, Avierinos JF, Messika-Zeitoun D, Detaint D, Capps M, Nkomo V, Scott C, Schaff HV, Tajik AJ. Quantitative determinants of the outcome of asymptomatic mitral regurgitation. N Engl J Med. 2005; 352: 9: 875–883.
71. Rosenhek R, Rader F, Klaar U, Gabriel H, Krejc M, Kalbeck D, Schemper M, Maurer G, Baumgartner H. Outcome of watchful waiting in asymptomatic severe mitral regurgitation. Circulation. 2006; 113: 2238–2244.
72. Messas E, Bel A, Cortes Morichetti M, Carrion C, Handschumacher MD, Peyrard S, Vilquin JT, Desnos M, Bruneval P, Carpentier A, Menasché P, Levine RA, Hagège AA. Autologous myoblast transplantation for chronic ischemic mitral regurgitation. J Am Coll Cardiol. 2006; 47: 2086–2093.
73. Feldman T. Percutaneous valve repair and replacement: challenges encountered, challenges met, challenges ahead. Circulation. 2006; 113: 771–772.
This article has been cited by other articles:
![]() |
J. L. Quill, A. J. Hill, T. G. Laske, O. Alfieri, and P. A. Iaizzo Mitral leaflet anatomy revisited. J. Thorac. Cardiovasc. Surg., May 1, 2009; 137(5): 1077 - 1081. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-B. Masson and J. G. Webb Percutaneous Treatment of Mitral Regurgitation Circ Cardiovasc Intervent, April 1, 2009; 2(2): 140 - 146. [Full Text] [PDF] |
||||
![]() |
H. C. Herrmann Transcatheter Aortic Valve Implantation: Past, Present, and Future Circ Cardiovasc Intervent, December 1, 2008; 1(3): 159 - 160. [Full Text] [PDF] |
||||
![]() |
M. Mack Fool me once, shame on you; fool me twice, shame on me! A perspective on the emerging world of percutaneous heart valve therapy. J. Thorac. Cardiovasc. Surg., October 1, 2008; 136(4): 816 - 819. [Full Text] [PDF] |
||||
![]() |
J. S. Gammie, L. S. Krowsoski, J. M. Brown, P. N. Odonkor, C. A. Young, M. J. Santos, J. S. Gottdiener, and B. P. Griffith Aortic Valve Bypass Surgery: Midterm Clinical Outcomes in a High-Risk Aortic Stenosis Population Circulation, September 30, 2008; 118(14): 1460 - 1466. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2007 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |