(Circulation. 2007;116:2802-2808.)
© 2007 American Heart Association, Inc.
Interventional Cardiology |
From the Department of Cardiology, Odense University Hospital, Odense (L.O.J., P.T., H.S.H., A.J., K.N.H., K.E.P.); Department of Cardiology, Aarhus University Hospital, Skejby Sygehus, Aarhus (L.T., J.F.L., H.E.B., L.R.K.); and Department of Cardiology, Rigshospitalet, Copenhagen (H.K.), Denmark.
Correspondence to Lisette Okkels Jensen, MD, PhD, Department of Cardiology, Catheterization Laboratory, Odense University Hospital, Sdr Blvd 29, 5000 Odense C, Denmark. E-mail okkels{at}dadlnet.dk
Received March 20, 2007; accepted September 14, 2007.
| Abstract |
|---|
|
|
|---|
Methods and Results— In 98 patients with angina pectoris, 1 de novo coronary lesion was treated with a bare-metal stent. After stent implantation, pressure wire measurements (Pd=mean hyperemic coronary pressure and Pa=mean aortic pressure) were performed in the target vessel: (1) Pd/Pa as distal to the artery as possible (fractional flow reserve per definition); (2) Pd/Pa just distal to the stent; (3) Pd/Pa just proximal to the stent; and (4) Pd/Pa at the ostium. Residual abnormal Pd/Pa was defined as a pressure drop between Pd/Pa measured at points 1 and 2. Fractional flow reserve distal to the artery after stenting was significantly lower (0.88±0.21 versus 0.97±0.05; P<0.001), and angiographic in-stent binary restenosis rate was significantly higher (44.0% versus 8.1%; P<0.001) in vessels with a residual abnormal Pd/Pa. Residual abnormal Pd/Pa (odds ratio, 4.39; 95% confidence interval, 1.10 to 18.16; P=0.034), reference vessel size (odds ratio, 0.17; 95% confidence interval, 0.04 to 0.69; P=0.013), and stent length (odds ratio, 1.11; 95% confidence interval, 1.03 to 1.21; P=0.009) were predictors of angiographic in-stent restenosis after 9 months.
Conclusions— A residual abnormal Pd/Pa distal to a bare-metal stent was an independent predictor of in-stent restenosis after implantation of a coronary bare-metal stent.
Key Words: atherosclerosis collateral circulation myocardial fractional flow reserve restenosis stents
| Introduction |
|---|
|
|
|---|
Clinical Perspective p 2808
Coronary pressure measurement with determination of fractional flow reserve (FFR)6,7 has been proposed as a supplementary technique for optimizing PCI results.8–10 In addition, FFR has been shown to have a predictive value after PCI.11,12 FFR measured after stenting with the pressure wire located distal to the stent indicates the effects on maximal flow of the stented segment and of the remaining part of the artery. A complete analysis of the coronary artery after PCI can be achieved by a pressure wire pullback during sustained hyperemia induced by intravenous adenosine. A pullback pressure recording in an epicardial coronary artery reflects the conductance of the entire artery as well as of the individual segments. In diffusely diseased coronary arteries, a marked pressure drop may occur between an implanted stent and the distal part of the artery.
The aim of the present study was to assess the in-stent restenosis rate in the setting of optimal stent implantation and a persisting pressure gradient in the vessel distal to the stent as assessed by a pressure wire pullback recording in the stented coronary artery.
| Methods |
|---|
|
|
|---|
Intracoronary Pressure Measurements
Before pressure measurements and intervention, the patients had a 200-µg intracoronary nitroglycerin and a 5000- to 10 000-U intravenous heparin administration. A 0.014-inch pressure wire (PressureWire, Radi Medical Systems, Uppsala, Sweden, or WaweWire, JoMed, Helsinborg, Sweden) was passed through the target lesion and placed as distal to the coronary artery as possible. Maximal hyperemia was induced by intravenous adenosine (140 µg/kg per min), and the ratio Pd/Pa was calculated. Here, Pd represents mean hyperemic coronary pressure of the index vessel measured by the pressure wire, and Pa represents mean aortic pressure measured by the guiding catheter. After stenting, a slow manual pullback of the pressure sensor from the most distal position to the proximal part of the artery was performed at maximal hyperemia and recorded on paper. Pressure measurements were performed in the whole length of the artery: Pd/Pa was measured as (1) distal in the artery as possible (per definition of FFR), (2) just distal to the stent, (3) just proximal to the stent, and (4) and at the ostium (Figure 1). A residual abnormal Pd/Pa was defined as a pressure drop between hyperemic Pd/Pa measured at points 1 and 2. On the basis of the pressure measurements, 2 groups of patients were defined. In group I, hyperemic Pd/Pa point 1-hyperemic Pd/Pa point 2 was <0, and in group II, hyperemic Pd/Pa point 1=hyperemic Pd/Pa point 2.
|
Quantitative Coronary Angiography
Angiographic studies performed at baseline, after the procedure, and at follow-up were assessed at the Angiographic Core Laboratory (Catheterization Laboratory, Odense University Hospital, Odense, Denmark). The computer-based ACOM.PC V3.1 (Siemens Medical Systems, Inc) was used for quantitative coronary angiography. Quantitative analysis was performed offline by experienced personnel unaware of the pressure measurements. The same 2 projections were used at all time points. The following angiographic measurements were measured: reference diameter of the vessel, minimal luminal diameter, percent diameter stenosis [1–(minimal luminal diameter/reference segment diameter)x100], and late lumen loss (difference between minimal luminal diameter at the end of the procedure and at follow-up).
Study End Point and Definitions
The primary end point of the study was binary angiographic restenosis
50% after 9 months. Optimal stenting was based on visual estimates by the operator and defined as a residual stenosis of 0. The core laboratory qualitative comparative analysis could overrule the visual operator assessment.
Statistical Analysis
Statistical analysis was performed with the use of SPSS 14.0. Categorical data were presented as counts and percentages and compared by the Pearson
2 test or the Fisher exact test. Continuous data were expressed as mean±SD and compared by t test. Two-way ANOVA with 4 repeated measurements was used to test whether a significant change occurred in measurements at different wire positions. If this test was significant, a paired t test was used to compare the values between the relevant wire positions. Separate logistic regression analyses were performed to identify univariate predictors of binary angiographic restenosis, and a subsequent stepwise (forward conditional) regression analysis was performed with entry and removal criteria of 0.05 and 0.10, respectively. Logistic regression analyses were presented as odds ratio (OR) with 95% confidence intervals. All statistical tests were 2-tailed.
With anticipation of a mean binary angiographic restenosis (
50%) rate of 25% after 9 months, the 166 patients enrolled provided 80% power and a 5%
level to detect a difference of 20% (event rate of 15% and 35%, respectively). A probability value <0.05 (2-sided) was considered statistically significant.
The authors had full access to and take full responsibility for the integrity of the data. All authors have read and agree to the manuscript as written.
| Results |
|---|
|
|
|---|
|
Lesion Characteristics
The occurrence of 1- and 2-vessel coronary artery disease was similar in the 2 groups. Diameter stenosis and lesion type did not differ between the 2 groups. Lesions tended to be longer in patients with residual abnormal Pd/Pa (Table 2). Patients with residual abnormal Pd/Pa had more left anterior descending artery lesions and fewer left circumflex artery lesions treated. The reference diameter was significantly lower in patients with residual abnormal Pd/Pa compared with patients without residual abnormal Pd/Pa (2.9±0.6 versus 3.3±0.5; P<0.001).
|
FFR and Pd/Pa at Different Pressure Wire Positions
Preinterventional FFR was significantly lower in vessels with residual abnormal Pd/Pa compared with vessels without residual abnormal Pd/Pa (Table 2). FFR before intervention correlated inversely with the angiographic diameter stenosis (r=–0.562, P<0.001) (Figure 2). FFR increased significantly after PCI (0.95±0.09 versus 0.65±0.20; P<0.001). No artery showed a sudden increase in distal coronary pressure by pressure wire pullback. This indicated that the pressure gradient, observed in the distal part of the artery, was due to a continuous loss of pressure in the artery distal to the stented segment and was not caused by an angiographically undetected focal narrowing. After the intervention, a significant within-subject change of Pd/Pa occurred in the entire artery in patients with a residual abnormal Pd/Pa (group I) (P<0.001, 2-way ANOVA analysis; Figure 3), whereas the within-subject change of Pd/Pa did not differ significantly along the entire length of the artery in patients without a residual abnormal Pd/Pa (group II). In group I, FFR (pressure wire placed in the distal vessel) was significantly lower than hyperemic Pd/Pa just distal to the stent (0.88±0.12 versus 0.94±0.11; P<0.001). In both groups, a small trans-stent gradient was present (Pd/Pa just distal to the stent versus Pd/Pa just proximal to the stent) (group I, 0.94±0.10 versus 0.97±0.08, P<0.001; group II, 0.97±0.05 versus 0.99±0.04, P=0.001). In group I, 29.3% (n=17) of the lesions had no trans-stent pressure gradient, and in group II, 67.5% (n=27) of the lesions had no trans-stent pressure gradient (P=0.001).
|
|
FFR After PCI in the Individual Vessels With and Without a Residual Abnormal Pd/Pa Ratio Distal to the Stent
No significant differences in Pd/Pa were present just distal to the stent for the 3 major coronary arteries in group I (left anterior descending artery, 0.90±0.13; left circumflex artery, 0.98±0.08; right coronary artery, 0.94±0.07; P=NS) or group II (left anterior descending artery, 0.94±0.04; left circumflex artery, 0.98±0.06; right coronary artery, 0.97±0.04; P=NS). FFR after PCI was significantly lower in vessels with a residual abnormal Pd/Pa ratio compared with vessels without a residual abnormal Pd/Pa ratio (0.88±0.21 versus 0.97±0.05; P<0.001).
Angiographic Follow-Up and Event Rate
At 9-month follow-up, in-stent binary angiographic restenosis was demonstrated in 28.6% of patients with angiographic follow-up (Table 3) (25.7% of all patients). Target lesion revascularization was performed in 24.2% of patients with angiographic follow-up (21.8% of all patients). The in-stent binary angiographic restenosis rate was 21.1% (n=8) for right coronary artery lesions, 30.4% (n=7) for left anterior descending artery lesions, and 37.5% (n=9) for left circumflex artery lesions. In vessels with a residual abnormal Pd/Pa ratio, binary angiographic restenosis was seen in 44.0% compared with 8.1% in vessels without a residual abnormal Pd/Pa ratio (P<0.001). During the 9-month follow-up, no stent thrombosis was seen, and none of the patients died or suffered an acute myocardial infarction.
|
Predictors of In-Stent Binary Angiographic Restenosis
Logistic regression analysis was used to assess independent predictors of binary angiographic restenosis at 9 months. The parameters examined with the use of univariate logistic regression analysis are shown in Table 4. A residual abnormal Pd/Pa, reference vessel diameter, minimal luminal diameter after stent implantation, lesion length, and stent length were significantly associated with an increased rate of binary angiographic restenosis. To adjust for differences in lesion factors, we performed a multiple logistic regression analysis including the abnormal residual Pd/Pa ratio (as the variable of primary interest) and reference vessel diameter (as a well-known factor influencing binary angiographic restenosis) by forced entry and parameters with P<0.20 (from the univariate analysis) in a forward stepwise procedure. Included in the forward procedure were minimal luminal diameter after stent implantation, lesion length, stent length, FFR before PCI, and FFR after PCI. After these adjustments, a residual abnormal Pd/Pa ratio, reference vessel diameter, and stent length were found to be predictors of binary angiographic restenosis at 9 months (Table 5). Performing a backward stepwise procedure showed the same predictors of binary angiographic restenosis. The c statistic (area under the receiver operating characteristic curve) in the final model was 0.83.
|
|
Number of Patients Included
According to the power calculation, 166 patients were expected to be included. However, during the enrollment period the drug-eluting stents were implemented for clinical use, and the study was stopped before enrollment of the 166 patients in order not to have a selected cohort with bare-metal stents.
In regard to calculating the power with the observed difference of 44.0%–8.1%=35.9% in the event rate with binary angiographic restenosis, for a comparison of 2 independent binomial proportions with the use of the likelihood ratio statistic with a
2 approximation with a 2-sided significance level of 0.05, group sample sizes of 50 and 38 have an approximate power of 0.974 when the proportions are 0.44 and 0.081.
| Discussion |
|---|
|
|
|---|
In normal coronary arteries, no gradient exists between the ostium and the distal part of the artery despite induced hyperemia. In patients with angiographically mild disease, Gould et al13 demonstrated a base-to-apex myocardial perfusion gradient after dipyridamole administration by positron emission tomography without significant regional perfusion defects. It has been shown by coronary pressure measurements that a base-to-apex perfusion gradient can be due to abnormal resistance in atherosclerotic epicardial coronary arteries without segmental stenosis5 because diffuse coronary atherosclerosis with no focal stenoses results in a graded, continuous pressure drop in the artery. This is in accordance with the findings of the present study, in which optimal coronary stenting resulted in a minimum hyperemic pressure drop within the stented coronary artery segment, whereas the transstenotic hyperemic gradient was not eliminated fully after stenting when the entire artery was evaluated. The minimum hyperemic pressure drop across the stented segment might indicate a small resistance to blood flow, but the present study was not designed to obtain a total elimination of the transstenotic hyperemic gradient after stenting.
After treatment with bare-metal stents, several studies have demonstrated that FFR predicts major cardiac events.11,12 However, when FFR is reduced after stenting, it is important to distinguish between a persistent hyperemic gradient due to incomplete deployment and a gradient caused by diffuse disease proximal or distal to the treated lesion. Consequently, evaluation of stent deployment by FFR can be improved by calculations of Pd/Pa distal and proximal to the stent during maximal hyperemia to assess the conductance of the stented segment. In the present study, FFR distal to the artery tended to be related to development of in-stent restenosis.
Reference Vessel Size
The second independent predictor of angiographic in-stent restenosis after 9 months was reference vessel size. The ability to distinguish between large and small coronary arteries on the basis of quantitative coronary angiography is essential in PCI, and stent implantation in small arteries is a well-known independent risk factor of restenosis and major adverse cardiac events after PCI.14–16 The mechanisms behind the unfavorable outcome for small vessels are not well understood. In addition to a small postprocedural lumen diameter, a high plaque burden and pronounced diffuse disease may be important factors.17 Pathological and intravascular ultrasound studies have shown that an angiographically documented stenosis is associated with diffusely atherosclerotic changes in other parts of the coronary artery, although this may not be identified by coronary arteriography.18–20 These findings are in accordance with the present study, in which coronary arteries without diffuse disease had a larger reference vessel diameter.
Pressure Pullback Recordings With Intravenous Adenosine
In the present study, maximum hyperemia was achieved with continuous intravenous adenosine infusion because intracoronary adenosine was too short-acting for a pullback recording. A pullback pressure recording in the epicardial coronary artery reflects the conductance of the entire artery as well as of every individual segment. A pullback recording during adenosine infusion has the potential to differentiate between diffusely diseased coronary arteries and a focal problem, such as a coronary lesion or an underexpanded stent.
Residual Abnormal Pd/Pa and Restenosis
Several mechanisms might influence our findings of increased binary restenosis rate in lesions with a residual abnormal Pd/Pa distal to the stent. Lesions with a residual abnormal Pd/Pa might not have been covered completely by the stent, and our intention to stent from disease-free to disease-free vessel might not have been achieved. In addition, a positive remodeling may have made a diffusely diseased artery look angiographically normal, and a diffusely diseased vessel might appear smaller because of lack of a normal reference segment. This might result in underexpansion or undersizing of the stent. The same degree of in-stent neointimal hyperplasia would contribute to a larger relative lumen reduction in a smaller stent size compared with a larger stent.
Study Limitations
Several limitations related to the study design should be taken into account. First, a low FFR and a low Pd/Pa without a focal step-up during a manual pullback through the artery is a pathophysiological finding that hypothetically can be applied to "diffuse disease," which is an anatomic description. However, we did not perform intravascular ultrasound imaging to confirm the presence of diffuse disease. Second, intravascular ultrasound imaging was not performed to confirm optimal stent expansion. Third, only disease distal to the stent was taken into account as a proximal and/or a distal "diffuse disease segment." Multiple segments within a patient should have been taken into account in the statistical analysis, and this could raise the question of whether either the distal or the proximal segment would be the strongest predictor, but the study was not powered for this determination. Fourth, the regression analyses for restenosis are based on 88 patients having 25 events only, as evidenced by the very wide confidence limits. In addition, we performed statistical tests with no accounting for multiple testing.
All patients were treated with bare-metal stents, and the overall angiographic restenosis rate was 25.7%, which is comparable to results with the placebo group in drug-eluting stent trials.21,22 The use of drug-eluting stents has reduced restenosis rates dramatically, and our results cannot be extended to patients treated with drug-eluting stents.
Conclusion
A pullback recording during maximal hyperemia of a coronary artery treated with a stent is a rapid and simple method to analyze residual hyperemic gradients after coronary stenting. By this technique, it is possible to differentiate between a persistent gradient caused by incomplete stent deployment or by diffuse disease proximal or distal to the stent. Furthermore, a distal residual abnormal Pd/Pa seems to be a predictor of in-stent restenosis.
| Acknowledgments |
|---|
None.
| References |
|---|
|
|
|---|
2. Fischman DL, Leon MB, Baim DS, Schatz RA, Savage MP, Penn I, Detre K, Veltri L, Ricci D, Nobuyoshi M, Cleman M, Heuser R, Almond D, Teirstein PS, Fish RD, Colombo A, Brinker J, Moses J, Shaknovich A, Hirshfeld J, Bailey S, Ellis S, Rake R, Goldberg S; Stent Restenosis Study Investigators. A randomized comparison of coronary-stent placement and balloon angioplasty in the treatment of coronary artery disease. N Engl J Med. 1994; 331: 496–501.
3. Serruys PW, van Hout B, Bonnier H, Legrand V, Garcia E, Macaya C, Sousa E, van der Giessen W, Colombo A, Seabra-Gomes R, Kiemeneij F, Ruygrok P, Ormiston J, Emanuelsson H, Fajadet J, Haude M, Klugmann S, Morel MA. Randomised comparison of implantation of heparin-coated stents with balloon angioplasty in selected patients with coronary artery disease (Benestent II). Lancet. 1998; 352: 673–681.[CrossRef][Medline] [Order article via Infotrieve]
4. Jensen LO, Thayssen P, Kassis E, Rasmussen K, Saunamaki K, Thuesen L; Danish Percutaneous Transluminal Coronary Angioplasty Registry. Target vessel revascularization following percutaneous coronary intervention: a 10-year report from the Danish Percutaneous Transluminal Coronary Angioplasty Registry. Scand Cardiovasc J. 2005; 39: 30–35.[CrossRef][Medline] [Order article via Infotrieve]
5. De Bruyne B, Hersbach F, Pijls NH, Bartunek J, Bech JW, Heyndrickx GR, Gould KL, Wijns W. Abnormal epicardial coronary resistance in patients with diffuse atherosclerosis but "normal" coronary angiography. Circulation. 2001; 104: 2401–2406.
6. Pijls NH, Van Gelder B, Van der Voort P, Peels K, Bracke FA, Bonnier HJ, el Gamal MI. Fractional flow reserve: a useful index to evaluate the influence of an epicardial coronary stenosis on myocardial blood flow. Circulation. 1995; 92: 3183–3193.
7. Pijls NH, De Bruyne B, Peels K, Van Der Voort PH, Bonnier HJ, Bartunek J, Koolen JJ. Measurement of fractional flow reserve to assess the functional severity of coronary artery stenoses. N Engl J Med. 1996; 334: 2401–2406.
8. Hanekamp CE, Koolen JJ, Pijls NH, Michels HR, Bonnier HJ. Comparison of quantitative coronary angiography, intravascular ultrasound, and coronary pressure measurement to assess optimum stent deployment. Circulation. 1999; 99: 1015–1021.
9. Fearon WF, Luna J, Samady H, Powers ER, Feldman T, Dib N, Tuzcu EM, Cleman MW, Chou TM, Cohen DJ, Ragosta M, Takagi A, Jeremias A, Fitzgerald PJ, Yeung AC, Kern MJ, Yock PG. Fractional flow reserve compared with intravascular ultrasound guidance for optimizing stent deployment. Circulation. 2001; 104: 1917–1922.
10. Bech GJ, De Bruyne B, Pijls NH, de Muinck ED, Hoorntje JC, Escaned J, Stella PR, Boersma E, Bartunek J, Koolen JJ, Wijns W. Fractional flow reserve to determine the appropriateness of angioplasty in moderate coronary stenosis: a randomized trial. Circulation. 2001; 103: 2928–2934.
11. Pijls NH, Klauss V, Siebert U, Powers E, Takazawa K, Fearon WF, Escaned J, Tsurumi Y, Akasaka T, Samady H, De Bruyne B; Fractional Flow Reserve (FFR) Post-Stent Registry Investigators. Coronary pressure measurement after stenting predicts adverse events at follow-up: a multicenter registry. Circulation. 2002; 105: 2950–2954.
12. Klauss V, Erdin P, Rieber J, Leibig M, Stempfle HU, Konig A, Baylacher M, Theisen K, Haufe MC, Sroczynski G, Schiele T, Siebert U. Fractional flow reserve for the prediction of cardiac events after coronary stent implantation: results of a multivariate analysis. Heart. 2005; 91: 203–206.
13. Gould KL, Nakagawa Y, Nakagawa K, Sdringola S, Hess MJ, Haynie M, Parker N, Mullani N, Kirkeeide R. Frequency and clinical implications of fluid dynamically significant diffuse coronary artery disease manifest as graded, longitudinal, base-to-apex myocardial perfusion abnormalities by noninvasive positron emission tomography. Circulation. 2000; 101: 1931–1939.
14. Elezi S, Kastrati A, Neumann FJ, Hadamitzky M, Dirschinger J, Schomig A. Vessel size and long-term outcome after coronary stent placement. Circulation. 1998; 98: 1875–1880.
15. Akiyama T, Moussa I, Reimers B, Ferraro M, Kobayashi Y, Blengino S, Di Francesco L, Finci L, Di Mario C, Colombo A. Angiographic and clinical outcome following coronary stenting of small vessels: a comparison with coronary stenting of large vessels. J Am Coll Cardiol. 1998; 32: 1610–1618.
16. Kasaoka S, Tobis JM, Akiyama T, Reimers B, Di Mario C, Wong ND, Colombo A. Angiographic and intravascular ultrasound predictors of in-stent restenosis. J Am Coll Cardiol. 1998; 32: 1630–1635.
17. Mintz GS, Popma JJ, Pichard AD, Kent KM, Salter LF, Chuang YC, Griffin J, Leon MB. Intravascular ultrasound predictors of restenosis after percutaneous transcatheter coronary revascularization. J Am Coll Cardiol. 1996; 27: 1678–1687.[Abstract]
18. Glagov S, Weisenberg E, Zarins CK, Stankunavicius R, Kolettis GJ. Compensatory enlargement of human atherosclerotic coronary arteries. N Engl J Med. 1987; 316: 1371–1375.[Abstract]
19. Nissen SE, Gurley JC, Grines CL, Booth DC, McClure R, Berk M, Fischer C, DeMaria AN. Intravascular ultrasound assessment of lumen size and wall morphology in normal subjects and patients with coronary artery disease. Circulation. 1991; 84: 1087–1099.
20. Mintz GS, Painter JA, Pichard AD, Kent KM, Satler LF, Popma JJ, Chuang YC, Bucher TA, Sokolowicz LE, Leon MB. Atherosclerosis in angiographically "normal" coronary artery reference segments: an intravascular ultrasound study with clinical correlations. J Am Coll Cardiol. 1995; 25: 1479–1485.[Abstract]
21. Morice MC, Serruys PW, Sousa JE, Fajadet J, Ban Hayashi E, Perin M, Colombo A, Schuler G, Barragan P, Guagliumi G, Molnar F, Falotico R; RAVEL Study Group. Randomized study with the sirolimus-coated BX velocity balloon-expandable stent in the treatment of patients with de novo native coronary artery lesions: a randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. N Engl J Med. 2002; 346: 1773–1780.
22. Stone GW, Ellis SG, Cox DA, Hermiller J, OShaughnessy C, Mann JT, Turco M, Caputo R, Bergin P, Greenberg J, Popma JJ, Russell ME; TAXUS-IV Investigators. A polymer-based, paclitaxel-eluting stent in patients with coronary artery disease. N Engl J Med. 2004; 350: 221–231.