(Circulation. 2007;116:1795-1800.)
© 2007 American Heart Association, Inc.
Valvular Heart Disease |
From the Department of Thoracic and Cardiovascular Surgery and Department of Cardiology, Rikshospitalet-Radiumhospitalet Medical Center, University of Oslo, Oslo, Norway.
Correspondence to Jan L. Svennevig, MD, PhD, Department of Thoracic and Cardiovascular Surgery, Sognsvannsveien 20, Rikshospitalet, 0027 Oslo, Norway. E-mail j.l.svennevig{at}medisin.uio.no
Received November 30, 2006; accepted July 18, 2007.
| Abstract |
|---|
|
|
|---|
Methods and Results— This is a retrospective cohort analysis of 816 consecutive patients undergoing aortic valve replacement with the Medtronic-Hall valve at Rikshospitalet, Oslo, Norway, from 1977 to 1987. All patients were contacted by means of questionnaires or telephone. Data were checked against hospital databases and medical records. Date of death was verified by the Norwegian civil registry. Follow-up was 99.6% complete. Survival analysis included operative deaths as well as late deaths. Survival at 25 years was 24.9%. No mechanical failures were found. Valve thrombosis was seen in 4 patients, in 1 case combined with pannus formation. Small valves (20 mm to 21 mm) were associated with reduced survival; however, when controlled for the confounding effects of age and gender, valve size did not remain a significant risk factor. Patient-related factors were important: Older age, female gender, and the need for concomitant coronary artery bypass surgery significantly reduced survival, whereas surgery of the ascending aorta did not. Linearized rates of thromboembolic complications, warfarin-related bleeding, and endocarditis were 1.5%, 0.7%, and 0.16%/patient-year, respectively. At follow-up, 79% of the patients were in New York Heart Association classes I to II.
Conclusions— This study confirms the excellent long-term outcome for patients with Medtronic-Hall valves in the aortic position.
Key Words: heart diseases prosthesis heart valves
| Introduction |
|---|
|
|
|---|
Editorial p 1759
Clinical Perspective p 1800
The MH valve is made of a single piece of titanium with no welds. The disk is made of tungsten-impregnated graphite with a carbon pyrolytic coating. The tungsten renders the disk radiopaque. In the aortic position, the maximal opening is 75°. A central aperture in the disk allows free rotation. The valve is rotatable in the sewing ring. The sewing ring is made of Teflon. Aortic sizes are 20 to 31 mm (outer diameter).
The initial clinical results were published in 1979.1 The cohort of patients has been followed up and analyzed at 5, 10, and 15 years.2–5 In the present study we report our 25-year follow-up data for the 816 patients undergoing aortic valve replacement (AVR) with special reference to survival, functional status, valve-related complications, and the impact of concomitant surgery.
| Methods |
|---|
|
|
|---|
|
In addition, patients who underwent repair of 1 valve in addition to AVR were included in this analysis. Of the 816 operations, 35 had undergone previous heart surgery. Closing date was May 1, 2005.
The authors had full access to and take full responsibility for the integrity of the data. All authors have read and agreed to the manuscript as written.
Operative Technique
The patients were operated on under conditions of moderate hypothermia, local cooling, and St Thomas Hospital cardioplegia. The prosthesis was in most cases implanted with interrupted Dacron sutures with the large valvular opening oriented toward the noncoronary cusp. The largest valve size possible was preferably implanted. In 700 patients it was possible to use the larger valve sizes, with external diameter varying from 23 to 31 mm, and in 116 patients the smallest valve sizes were used, with external diameter varying from 20 to 21 mm. The former group of patients were younger than the latter, with mean age 54.8±13.0 years versus 60.2±13.2 years, and had a higher male-to-female ratio (79% versus 21%).
Anticoagulation Therapy
Anticoagulation therapy was initiated on the first postoperative day by parenteral administration of warfarin and was continued lifelong with warfarin tablets. The efficacy of anticoagulation treatment was controlled by the TT method (Nycomed, Oslo, Norway), with the ideal target being 12% to 7%, corresponding to an international normalized ratio of 2.5 to 3.5.
Follow-Up
The standardized questionnaire2–5 on quality of life previously used in connection with the 1-, 5-, 10-, and 15-year follow-up studies was sent to all patients. Hospital records were scrutinized for valve-related complications.
Thromboembolism was defined as any sudden, focal neurological deficit persisting for >1 hour or the occurrence of any sudden symptoms related to obstruction of a systemic artery. Thrombotic obstruction of the prosthesis was defined as any thrombotic material impeding free movement of the disk. Anticoagulation-related hemorrhage included any nontraumatic bleeding necessitating hospital admission or blood transfusions. Periprosthetic leak included all paravalvular fistulas that necessitated reoperation. Mechanical valve failure was defined as any case of mechanical rupture of the valve. Prosthetic valve endocarditis was defined as any infection of the prosthesis documented by blood cultures, reoperation, or autopsy.
Statistical Analysis
Survival analysis was performed by the Kaplan-Meier method. Survival analysis included operative deaths as well as late deaths. Comparison of our cohort of AVR patients with the mortality of the total Norwegian population matched by age, gender, and time period was performed with the standard mortality ratio method. Comparison of survival was done with the use of Breslow and Mantel-Cox test statistics.6 Independent risk factors of mortality were estimated by the Cox regression model.6 Complications were expressed as linear rates. Comparison of linear rates was completed with a likelihood ratio test.
Study Limitation
Because this study began many years before the guidelines for reporting on valve mortality and morbidity were defined, our original definitions may deviate to some degree from the present definitions.7 During the first 10 years of our study, all incidents of morbidity (eg, thromboembolism, valve, thrombosis, bleeding, leak, endocarditis) were counted; however, the date of the incidents was not always stored. For this reason, only linear rates of mortality are given. Because of older age and the mental condition or physical handicap of patients (such as hip arthrosis), the functional class could not be assessed in 15 of 227 patients still alive at follow-up.
Strength of the Study
All patients have been operated on at our department and have been followed up for 25 years by the same group. Death dates could be verified from the Norwegian civil registry. A standard questionnaire was sent to the patients 1, 3, 5, 10, 15, 20, and 25 years after surgery. In the case of lack of a response, telephone interviews were performed. Collection of data ended on May 1, 2005. Three patients were lost to follow-up, and therefore follow-up was 99.6% complete and included a total of 10 805 patient-years. Average follow-up time was 13.2±8.2 years.
| Results |
|---|
|
|
|---|
|
Age
Survival was related to age at the time of operation (Figure 3). The youngest age group was characterized by a 94% survival for the first 15 years of follow-up, whereas the middle-aged and elderly patients showed a gradual decrease in survival recognizable during the whole observation period. The difference between the 3 age groups was significant during the whole observation period.
|
Gender
A comparison between male versus female patient survival showed the same pattern of a gradual decrease in survival with time but with a significantly better survival in the male population: 25.4±2.3% versus 16.3±2.8% at 25-year follow-up (Figure 4).
|
Valve Size
During the whole observation period, significantly better survival was seen in patients with the largest valve sizes of 22 to 31 compared with the smallest valve sizes of 20 to 21, with a 26.8±1.9% survival in the former group at 25-year follow-up compared with a 13.9±3.5% survival in the latter (Figure 5). However, patients receiving small valves were more often female and were older. When age and gender were controlled for their confounding effect by the Cox regression model, valve size did not remain a significant risk factor for survival. Thereby, the risk ratio for the smaller valve sizes was reduced from risk ratio=1.48 (95% confidence interval, 1.2 to 1.9; P=0.0001) to risk ratio=1.02 (95% confidence interval, 0.8 to 1.3; P=0.83).
|
Concomitant Coronary Artery Bypass Graft
Survival was significantly better in those who did not receive coronary artery bypass graft (CABG) surgery versus the 140 patients who needed CABG surgery at the time of valve prosthesis insertion: 27.7±1.9% versus 11.0±3.1% at 25-year follow-up (Figure 6).
|
Ascending Aortic Surgery
Twenty-four patients received a Dacron graft to replace the ascending aorta, and 22 patients underwent enlargement of the aortic annulus. The 20-year survival for these 2 groups (37.5±9.0% and 48.2±9.6%, respectively) did not differ significantly from the rest of the population (not enough patients were at risk for a 25-year analysis).
Functional Class
On the basis of the questionnaires, we were able to estimate New York Heart Association (NYHA) functional class in 212 of the 227 patients alive at follow-up (Figure 7). A total of 121 patients (57%) were without dyspnea or angina pectoris at rest or during exercise (NYHA class I); another 56 (22%) had dyspnea or angina pectoris during more strenuous exercise (NYHA class II), and the remaining 17% had significant symptoms during moderate exercise or at rest (NYHA class III to IV).
|
Morbidity
Various serious complications are shown in Figure 8.
|
No incidents of mechanical valve failure occurred. A total of 159 thromboembolic events were reported (1.5 events per 100 patient-years). Valve thrombosis was seen in 4 patients (0.04 per 100 patient-years), and in 1 case the thrombus was associated with pannus formation. Endocarditis was seen in 17 patients (0.16 events per 100 patient-years). Seventy-five patients needed hospitalization or blood transfusion for warfarin-related bleeding (0.7 events per 100 patient-years). Paravalvular leak required reoperation in 13 cases (0.1 events per 100 patient-years).
New operations were performed in 60 patients: 13 for perivalvular leak, 12 for endocarditis, 4 for valve thrombosis, 1 for subaortic stenosis, and 1 for a sinus valsalvae aneurysm. Twelve patients were reoperated on for dilated ascending aorta, 12 patients were reoperated for coronary artery disease, 2 patients needed a mitral valve replacement, and 3 patients underwent heart transplantation.
| Discussion |
|---|
|
|
|---|
25% after replacement of the aortic valve with a modern mechanical valve prosthesis. Comparable survival rates have been published for other valve types.8–14 The range, however, as shown in our study, is large, indicating the importance of prognostic factors other than the valve implantation itself. Patient-related factors seem in many studies to be most important for survival,15,16 although the type of prosthesis has also been found to be significant,11,16–18 especially in the small root.19,20
Age
Age at the time of the operation is the strongest predictor of long-term survival.18 As might be expected, the older the patients are at the time of implantation, the less 25-year survival is to be expected. The relatively small group of patients aged <18 years at the time of operation had almost no mortality for the first 15 years after operation, with a drop in survival to 88% after 20 years. For patients aged >18 years at the time of operation, survival curves show a gradual decline, recognizable within 5 years after operation and continuing for the whole observation period. This decline was more pronounced in those aged >70 years at the time of operation than in those in the group aged 16 to 70 years. One reason for this is of course the length of normal life (ie, "the more years one has lived, the shorter time is left"). Obviously, factors other than those related to age are also of importance for the long-term outcome. Older patients are more likely to have clinically significant associated comorbid conditions at the time of operation that are known to adversely affect survival after AVR failure.21 Thus, previous studies indicate that patient characteristics at baseline are of more importance as determinants of late mortality after valve replacement than the choice of prosthesis. Some of these factors are also elucidated in the present study.
Gender
During the whole observation study, male patients had better survival rates than female patients; the difference was significant with cumulative survival of 25% versus 16%, respectively, up to 25 years after operation. Some previous studies did not reveal gender to be of significance,19 and in 1 study male gender appeared to be a risk factor.21
Valve Size
Although reduction in the natural valve area after valve prosthesis insertion is usually mild to moderate in severity and is regarded of no immediate clinical significance, patients with the largest valve sizes (23 to 31) have been reported to have a better prognosis than those with the smallest valve sizes of 20 to 21. This might indicate that the smallest valve sizes represent a hazard to the life of the patients because of an anticipated higher transvalvular gradient. This, however, is less likely because hemodynamic studies with the MH prosthesis have shown small and hemodynamically acceptable gradients even in the smallest valve sizes,22–27 which may represent an alternative to root enlargement.26,28 The interpretations of our findings are, however, rather complex because patients with the smallest valve sizes tended to be older and included more female patients than those with the largest valve sizes. Both factors are known to affect long-term survival. However, when we performed a multivariate analysis correcting for gender and age, the risk ratio for the smaller valve sizes was reduced from risk ratio=1.48 to risk ratio=1.02, indicating that the risk of survival was the same in those with the large valve prosthesis as in those with the smallest valve sizes. This also means that the smallest MH valve prosthesis has an effective prosthetic valve area considered comparable to that of a normal human valve when inserted into a patient of appropriate body size and that the poorer survival seen in patients with the smallest valve sizes seen in our study can be attributed to the fact that they were representing a population older than those who received the larger valve sizes. In a randomized study, Fiore et al29 could not detect any difference in performance when comparing the St Jude and the MH valves.
Concomitant CABG
The present study shows that associated CABG surgery adversely affected survival. Associated CABG has been shown to represent an independent predictor of mortality.9,19 In the study of Connolly et al,30 the 5-year survival was 69% for patients without coronary artery disease and 39% for those with significant coronary artery disease. Our data at 10 years were 63.7% and 51.3%, respectively (Figure 4).
Concomitant Ascending Aorta Surgery
The good long-term survival for this group of patients was surprising, and we do not find much relevant data in the literature.31 On the other hand, many patients with a slight dilatation of the ascending aorta that was not dealt with during the aortic valve operation later required reoperation. In many previous studies, patients needing concomitant procedures were excluded from the analysis.9,32
Thromboembolic Complications
Patients who are at the lowest risk of thromboembolism are those with sinus rhythm, normal left ventricular function, and no previous history of thromboembolism. Our report shows that the risk of thromboembolism is very small with the use of the MH valve prosthesis and confirms the previous experiences with this valve prosthesis.18,21
Bleeding Complications
Because of the risk of thromboembolism, the major disadvantage with use of mechanical valves is the need for anticoagulation therapy, bleeding, and its consequences. The linearized rate of bleeding was 0.7 per 100 patient-years. Bleeding is more frequent in mechanical valves than in biological valves.11 However, bleeding complications can be further reduced by introducing international normalized ratio self-management that permits lower anticoagulation levels.33
Valve-Related Complications
Structural deterioration was not seen in this study. This is in agreement with previous studies.18 Pannus requiring reoperation was seen in 1 case. Hemolysis was only seen secondary to paravalvular leak.
Summary
In summary, this 25-year follow-up study confirms that the MH valve is a durable and reliable valve with a low rate of valve-related complications. The study further demonstrates the significance of patient-related risk factors for survival associated with coronary artery disease.
| Acknowledgments |
|---|
This study was supported by the Professor K.V. Hall Fund, Rikshospitalet, Oslo, Norway.
Disclosures
Dr Svennevig is presently (for a 4-year period) an elected member of the Medtronic Medical Advisory Board. The remaining authors report no conflicts.
| References |
|---|
|
|
|---|
2. Nitter-Hauge S, Semb B, Abdelnoor M, Hall KV. A 5 year experience with the Medtronic-Hall disc valve prosthesis. Circulation. 1983; 68 (suppl II): 169–174.
3. Abdelnoor M, Hall KV, Nitter-Hauge S, Golf S, Tretli S, Geiran O, Johansen A. Risk factors of morbidity and mortality in surgically treated chronic aortic valvular heart disease. Rev Epidem et Sante Publ. 1988; 36: 89–98.
4. Nitter-Hauge S, Abdelnoor M. Ten-year experience with the Medtronic-Hall valvular prosthesis: a study of 1104 patients. Circulation. 1989; 80 (suppl I): 43–48.
5. Nitter-Hauge S, Abdelnoor M, Svennevig JL. Fifteen-year experience with the Medtronic-Hall valve prosthesis: a follow-up study of 1104 consecutive patients. Circulation. 1996; 94: (suppl II): 105–108.
6. Lee ET. Statistical Methods for Survival Data Analysis. Belmont, Calif: Lifetime Learning Publications; 1980.
7. Edmunds LH Jr, Clark RE, Cohn LH, Miller C, Weisel RD. Guidelines for reporting morbidity and mortality after cardiac valvular operations. J Thorac Cardiovasc Surg. 1988; 96: 351–353.[Medline] [Order article via Infotrieve]
8. Godje OL, Fischlein T, Adelhard K, Nollert G, Klinner W, Reichart B. Thirty-year results of Starr-Edwards prostheses in the aortic and mitral position. Ann Thorac Surg. 1997; 63: 613–619.
9. Lund O, Nielsen SL, Arildsen H, Ilkjaer LB, Pilegaard HK. Standard aortic St. Jude valve at 18 years: performance profile and determinants of outcome. Ann Thorac Surg. 2000; 69: 1459–1465.
10. Mykèn P, Bech-Hanssen O, Phipps B, Caidahl K. Fifteen years follow up with the St. Jude Medical Biocor porcine bioprosthesis. J Heart Valve Dis. 2000; 9: 415–422.[Medline] [Order article via Infotrieve]
11. Oxenham H, Bloomfield P, Wheatley DJ, Lee RJ, Cunningham J, Prescott RJ, Miller HC. Twenty year comparison of a Bjork-Shiley mechanical heart valve with porcine bioprostheses. Heart. 2003; 89: 715–721.
12. Gao G, Wu Y, Grunkemeier GL, Furnary AP, Starr A. Forty-year survival with the Starr-Edwards heart valve prosthesis. J Heart Valve Dis. 2004; 13: 91–96.[Medline] [Order article via Infotrieve]
13. Emery RW, Krogh CC, Arom KV, Emery AM, Benyo-Albrecht K, Joyce LD, Nicoloff DM. The St. Jude Medical cardiac valve prosthesis: a 25-year experience with single valve replacement. Ann Thorac Surg. 2005; 79: 776–782.
14. Yankah CA, Schubel J, Buz S, Siniawski H, Hetzer R. Seventeen-year clinical results of 1,037 Mitroflow pericardial heart valve prostheses in the aortic position. J Heart Valve Dis. 2005; 14: 1–9.[Medline] [Order article via Infotrieve]
15. Bessel JR, Gower G, Craddock DR, Stubberfield J, Maddern GJ. Thirty years experience with heart valve surgery: isolated aortic valve replacement. Aust N Z J Surg. 1996; 66: 799–805.[Medline] [Order article via Infotrieve]
16. Khan SS, Trento A, DeRobertis M, Kass RM, Sandhu M, Czer LSC, Blanche C, Raissi S, Fontana GP, Cheng W, Chaux A, Matloff JM. Twenty-year comparison of tissue and mechanical valve replacement. J Thorac Cardiovasc Surg. 2001; 122: 257–269.
17. Murday AJ, Hochstitzky A, Mansfield J, Miles J, Taylor B, Whitley E, Treasure T. A prospective controlled trial of St. Jude versus Starr Edwards aortic and mitral valve prostheses. Ann Thorac Surg. 2003; 76: 66–73.
18. Akins CW. Long-term results with the Medtronic-Hall valvular prosthesis. Ann Thorac Surg. 1996; 61: 806–813.
19. He GW, Grunkemeier GL, Gately HL, Furnary AP, Starr A. Up to thirty-year survival after aortic valve replacement in the small aortic root. Ann Thorac Surg. 1995; 59: 1056–1062.
20. Rahimtoola SH. Choice of prosthetic heart valve for adult patients. J Am Coll Cardiol. 2003; 41: 893–904.
21. Butchart EG, Li HH, Payne N, Buchan K, Grunkemeier GL. Twenty years experience with the Medtronic Hall valve. J Thorac Cardiovasc Surg. 2001; 121: 1090–1100.
22. Wiseth R, Levang OW, Tangen G, Rein KA, Skjaerpe T, Hatle L. Exercise hemodynamics in small (
21 mm) aortic valve prostheses assessed by Doppler echocardiography. Am Heart J. 1993; 125: 138–146.[CrossRef][Medline]
[Order article via Infotrieve]
23. Anderson WA, Ilkowski DA, Eldredge J, Cha R, Chen C, Waters D, Mahan VL, Anolik G, Laub GW, Fernandez J, McGrath LB. The small aortic root and the Medtronic Hall valve: ultrafast computed tomography assessment of left ventricular mass following aortic valve replacement. J Heart Valve Dis. 1996; 5 (suppl 3): 329–335.
24. Aris A, Ramirez I, Camara ML, Carreras F, Borras X, Pons-Llado G. The 20 mm Medtronic Hall prosthesis in the small aortic root. J Heart Valve Dis. 1996; 5: 459–462.[Medline] [Order article via Infotrieve]
25. Brazao AJ, Prieto D, de Oliveira JF, Eugenio L, Antunes MJ. Aortic valve replacement with small-sized disc prostheses (Medtronic Hall). J Heart Valve Dis. 1999; 8: 680–686.[Medline] [Order article via Infotrieve]
26. Kleine P, Abdel-Rahman U, Klesius AA, Scherer M, Simon A, Moritz A. Comparison of hemodynamic performance of Medtronic Hall 21 mm versus St. Jude Medical 23 mm prostheses in pigs. J Heart Valve Dis. 2002; 11: 857–863.[Medline] [Order article via Infotrieve]
27. Sung K, Park PW, Park KH, Jun TG, Lee YT, Yang JH. Comparison of transprosthetic mean pressure gradients between Medtronic Hall and ATS valves in the aortic position. Int J Cardiol. 2005; 99: 29–35.[CrossRef][Medline] [Order article via Infotrieve]
28. Obadia JF, Martelloni YA, Bastien OH, Durand de Gevigney GM, Chassignolle JF. Long-term follow-up of small (size 20 and 21) Medtronic-Hall aortic valve prostheses. Ann Thorac Surg. 1997; 64: 421–425.
29. Fiore AC, Swartz M, Grunkemeier G, Dressler F, Peigh PS, McBride LR, Kaiser GC, Labovitz AJ, Barner HB. Valve replacement in the small aortic annulus: prospective randomized trial of St. Jude with Medtronic Hall. Eur J Cardiothorac Surg. 1997; 11: 485–492.[Abstract]
30. Connolly HM, Oh JK, Orszulak TA, Osborn SL, Roger VL, Hodge DO, Bailey KR, Seward JB, Tajik AJ. Aortic valve replacement for aortic stenosis with severe left ventricular dysfunction: prognostic indicators. Circulation. 1997; 95: 2395–2400.
31. Pracki P, Petri D, Kellner HJ, Struck E. Composite graft (Medtronic-Hall) replacement of the ascending aorta and aortic valve in aortic aneurysms: what is adequate follow-up? Thorac Cardiovasc Surg. 1995; 43: 104–107.[Medline] [Order article via Infotrieve]
32. Arom KV, Emery RW, Petersen B, Radosevich DM. St. Jude Medical valve prosthesis: health status of the patients after 15 years. Ann Thorac Cardiovasc Surg. 1996; 2: 43–49.
33. Koertke H, Minami K, Boething D, Breymann T, Seifert D, Wagner O, Atmacha N, Krian A, Ennker J, Taborski U, Klovekorn WP, Moosdorf R, Saggau W, Koerfer R. INR self-management permits lower anticoagulation levels after mechanical heart valve replacement. Circulation. 2003; 108 (suppl 1): 1175–1178.
Related Article:
This article has been cited by other articles:
CLINICAL PERSPECTIVE
The Medtronic-Hall valve, first implanted in 1977, is a mechanical heart valve made of a single piece of titanium. The disk is made of graphite with a carbon pyrolytic coating. We followed up with 816 patients who received Medtronic-Hall valves in the aortic position over a 25-year period. No mechanical failures had occurred, and the linearized rates of complications were low. The cumulative survival at 25 years, 24.9%, was lower than expected for a matched normal population; however, 79% of the patients were in New York Heart Association functional class I to II. Older age, female gender, and the need for concomitant coronary surgery were associated with reduced survival. This study confirms that the Medtronic-Hall valve is a durable and reliable heart valve with a low rate of valve-related complications. The study further demonstrates the significance of patient-related risk factors for survival.
Circulation 2007 116: 1755.
![]() |
S. H. Rahimtoola The Year in Valvular Heart Disease J. Am. Coll. Cardiol., May 19, 2009; 53(20): 1894 - 1908. [Full Text] [PDF] |
||||
![]() |
A. Vahanian, B. Iung, L. Piérard, R. Dion, and J. Pepper CHAPTER 21 Valvular Heart Disease ESC Textbook of Cardiovascular Medicine, January 1, 2009; 2(1): med-9780199566990-chapter - med-9780199566990-chapter. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. J. Vlahakes Mechanical Heart Valves: The Test of Time... Circulation, October 16, 2007; 116(16): 1759 - 1760. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2007 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |