(Circulation. 2007;116:I-276 – I-281.)
© 2007 American Heart Association, Inc.
Surgery for Valvular Heart Disease |
From the Department of Cardiothoracic Surgery (T.A.T., D.T.L., G.T.D., N.B.I., D.C.M.) and the Division of Cardiovascular Medicine (D.L.), Stanford University School of Medicine, Stanford, Calif; the Department of Cardiothoracic and Vascular Surgery (S.L.N.), Aarhus University, Aarhus, Denmark; and the Laboratory of Cardiovascular Physiology and Biophysics (G.T.D., N.B.I.), Research Institute of the Palo Alto Medical Foundation, Palo Alto, Calif.
Correspondence to D. Craig Miller, MD, Department of Cardiothoracic Surgery, Falk Cardiovascular Research Center, Stanford University School of Medicine, Stanford, California 94305-5247. E-mail dcm{at}stanford.edu
| Abstract |
|---|
|
|
|---|
Methods and Results— Eight sheep were studied under open-chest conditions immediately after surgical placement of a miniature force transducer to approximate the leaflets and implantation of radiopaque markers on the LV and mitral annulus (MA). Chronotropic stimulation was induced with atrial pacing at 130 minutes–1 (n=5) whereas inotropic state was increased with IV CaCl2 bolus (n=8). Hemodynamic data, stitch tension, and 3-D marker coordinates were obtained throughout the cardiac cycle before and during each intervention. Peak stitch tension (TMAX) under all conditions was observed in diastole and temporally correlated with peak annular SL (SLMAX) size. Atrial pacing did not change peak transducer tension or annular size. Calcium infusion also did not alter peak transducer tension (0.29±0.11 versus 0.32±0.10 N; P=NS) and only slightly reduced SL dimension (29.9±3.3 versus 29.3±3.5 mm; P<0.05).
Conclusion— Isolated increase in heart rate or inotropic state did not alter peak stitch tension whereas enhanced contractile state decreased SL diameter minimally. These data, combined with those from our previous study, suggest that geometric (SL diameter) rather than hemodynamic parameters are the main determinants of "Alfieri stitch" tension. This implies that any interventional or surgical edge-to-edge repair performed without concomitant annular reduction to limit the SL dimension could expose the leaflet junction to forces which could limit repair durability.
Key Words: mitral valve valvuloplasty valves
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
|
After marker implantation, the heart was defibrillated, the animals weaned from cardiopulmonary bypass and transferred immediately to the experimental animal catheterization laboratory where they were studied intubated in open-chest, anesthetized condition. Simultaneous biplane videofluoroscopy, hemodynamic data recordings, and force transducer tension readings were measured before and after 2 experimental conditions: Rapid atrial pacing at 130 minutes–1 (Apace) and bolus infusion of 1 gm of calcium chloride (Calcium). Because of pacer noncapture, pacing data were available for only 5 animals, whereas all 8 animals had successful data acquisition during calcium infusion.
All animals received humane care in compliance with the Principles of Laboratory Animal Care formulated by the National Society for Medical Research and the Guide for Care and Use of Laboratory Animals prepared by the National Academy of Sciences and published by the National Institutes of Health (DHEW NIHG publication 85-23, revised 1985). This study was approved by the Stanford Medical Center Laboratory Research Animal Review committee and conducted according to Stanford University policy. The authors had full access to and take full responsibility for the integrity of the data. All authors have read and agree to the manuscript as written
Data Acquisition
Images were acquired with the animal in the right lateral decubitus position using a Philips Optimus 2000 biplane Lateral ARC 2/Poly DIAGNOST C2 system (Phillips Medical Systems, North America Company) with the image intensifier in the 9" fluoroscopic mode. Data from the 2 radiographic views were digitized and merged using custom-designed software18 to yield the 3-dimensional x,y,z coordinates for each of the radiopaque markers every 16.7 ms throughout the cardiac cycle. Force transducer tension, ascending aortic pressure, LV pressure, and ECG voltage signals were also digitized and recorded simultaneously.
Data Analysis
Two to 3 consecutive steady state beats before and after atrial pacing and calcium infusion were averaged and considered "Pre-Apace", "Apace", "Pre-calcium", and "Calcium" data for each animal. During each cardiac cycle, end-systole (ES) was defined as the frame containing minimum LV volume; end-diastole (ED) was defined as the videofluoroscopic preceding the frame containing the peak LV volume. Instantaneous LV volume was computed from the LV markers as described previously.19 Stroke volume (SV) was calculated as the difference between end-diastolic LV volume (EDV) and end-systolic LV volume (ESV), whereas cardiac output was calculated as heart rate multiplied by stroke volume (CO= HRxSV).
Mitral Annular Dynamics
Mitral annular area (MAA) was computed from the 3-D coordinates of the 8 markers sutured to the mitral annulus using an annular centroid. The septal-lateral (SL) annular diameter was calculated as the distance in 3-D space between markers placed on the midanterior and midposterior mitral annulus.
Statistical Analysis
All data are reported as mean±1SD, unless otherwise shown. Hemodynamic and marker-derived data from 3 consecutive steady-state beats from each heart were time-aligned at end-diastole. Marker data were calculated over 20 frames before and after end-diastole, thus allowing evaluation over a time period of 650 ms. The mean and SD for each variable at each sampling instant were then computed for all conditions. Data were compared using Student t test for paired comparisons.
| Results |
|---|
|
|
|---|
|
Force Transducer Tension
Force transducer tension, annular septal-lateral diameter, and annular area throughout the cardiac cycle before and after rapid atrial pacing are illustrated in Figure 2. Before pacing, maximum transducer tension was observed in diastole and temporally coincided with peak SL diameter and annular area. Atrial pacing did not alter transducer tension and annular SL diameter, and area also did not change. As shown in Figure 3, baseline dynamics of leaflet approximating point tension and annular diameter and area before calcium infusion were similar to those observed before atrial pacing. With calcium infusion and significant increase in inotropic state and flow (Table 1), transducer tension did not change and annular dimensions were altered only slightly. Table 2 summarizes peak transducer tension, annular septal-lateral diameter, and annular area before and after the 2 experimental interventions. Maximum annular septal-lateral diameter decreased only by 0.6 mm with calcium infusion, whereas no significant change in maximum transducer tension was observed.
|
|
|
| Discussion |
|---|
|
|
|---|
The current data are corroborated by our prior findings from the same sheep, where peak transducer tension occurred in diastole and coincided with peak annular SL diameter.13,14 Dobutamine infusion significantly augmented flow, heart rate, and inotropic state but caused transducer tension to fall as the annular SL size became smaller.14 As a corollary, in the current study, calcium infusion increased inotropic state and flow without augmenting heart rate, yet transducer tension did not change and no clinically significant alteration of annular area or SL size was induced. Thus, it appears, in this model at least, that hemodynamic interventions which do not affect annular SL dimension significantly will probably not increase Alfieri stitch tension. This is not to say that hemodynamic interventions may not be associated with changes in force transducer tension, but this is seen in the setting of concurrent reduction of annular size as witnessed previously.14 Conversely, a small increase in SL annular size accompanying acute posterolateral ischemia impressively increased the tension on the approximating stitch.13 Computational models of edge-to-edge repair support these findings as a 20% increase in annular size increased longitudinal stitch stress by more than 2-fold,23 whereas reduced annular area decreased anterior leaflet stress near the approximating suture.24 It is therefore reasonable to conjecture that addition of annuloplasty to an Alfieri stitch should reduce stitch tension and any subsequent rise in leaflet stress imposed by annular dilatation. Whether this would translate into improved clinical outcomes remains speculative.
Although current data favor annular geometry over hemodynamic conditions as the primary determinant of Alfieri stitch tension, the effect of flow and transmitral gradients cannot be ignored. A recent elegant in vitro study15 found peak approximating suture tension to be bimodally distributed during the cardiac cycle with the diastolic component predominating under normal annular geometry, supporting our results. With annular dilatation, on the other hand, systolic tension became predominant and diastolic tension varied inversely with annular size. This discordance with our current results may lie in the differences in transducer positioning, species differences (porcine versus ovine), in vitro versus in vivo conditions, annular diameter ratio (commissure-commissure to septal-lateral), or magnitude of 4-D perturbations in papillary muscle positions and shape of the mitral annulus. Mitral annular area and septal-lateral annular size in conjunction with mitral flow rate and transmitral pressure, however, were independent predictors of systolic force transducer tension. Annular septal-lateral dilatation of 20% in this in vitro model significantly increased transducer tension and mitral regurgitation supporting the notion that the edge-to-edge repair should be performed along with some type of reducing annuloplasty.
Whether the results of our study can be extrapolated to the clinical arena remains to be demonstrated, although a recent phase I trial of percutaneous Alfieri edge-to-edge repair offers some insight.11 In this multicenter study, among the 24 pts who had a percutaneous Alfieri clip placed, 3 had clip detachments and another 3 developed recurrent 3+ MR after initial reduction of mitral insufficiency.11 It is difficult to speculate whether the modes of repair failure and recurrent MR were directly related to excessive stitch tension, but at the same time it is reasonable to conjecture that in patients discharged with intact repairs and no mitral regurgitation without annular rings high stitch tension attributable to annular dilatation could be a contributing factor. Although the measured Alfieri stitch tension in the current study appears small (range 0.25 to 0.35 N), it is consistent with computational values reported in the literature23 and in vivo measurements of tension in the first-order chordae tendineae during acute LV ischemia and mitral regurgitation.25 Therefore, we believe these tension values are clinically relevant during the cardiac cycle, especially over the long-term. Potentially a second percutaneous annuloplasty device26 could be added to the leaflet clip, akin to surgical practice, to reduce annular SL diameter and size, suture tension, and possibly improve repair durability, but we need to learn more about how interventional edge-to-edge repairs fail. The results of the current study argue for using an annular reducing procedure in conjunction with the Alfieri repair as annular size appears to be the chief determinant of stitch tension which could increase with progressive dilatation of an unsupported annulus. A direct causal link between these biomechanical findings and clinical pathophysiology, however, remains to be established.
Limitations
The major limitations of this study are inherent to the animal model. Our tension measurements were obtained in sheep which were anesthetized and studied in open-chest conditions shortly after cardiopulmonary bypass. The surgical procedure necessitated implantation of small metallic markers around the mitral annulus which could theoretically affect valvular function. Echocardiographic studies, however, indicate that the markers do not interfere with mitral annular or leaflet motion as they are very small (aggregate mass =20±6 mg). However, the force transducer, although small, is considerably larger than the implanted markers and could potentially distort leaflet geometry. Our previous experiments with dobutamine27 revealed that leaflet approximation with the force transducer permits normal annular motion and hence there is no "girdling" effect of leaflet approximation. Furthermore, leaflet edge displacement from the annular plane near each commissure at end-systole was found to be similar in control animals and those who had undergone leaflet approximation with the miniature force transducer.28 Therefore, a significant effect of the transducer on leaflet geometry, though feasible, is probably unlikely. Although annular size affects Alfieri stitch tension, and annular dilatation, especially in the septal-lateral dimension, increases suture tension, it is yet not known what effect increased suture tension has on the durability of the edge-to-edge repair. The current study investigated Alfieri stitch tension with approximation of the central edges of both leaflets; whether the current findings are applicable when an "asymmetric" double orifice repair is performed (ie, suture placed toward either commissure) is not known. Lastly, although neither baseline nor peak tension changed with the hemodynamic interventions used, it is possible that tension-time integral across the cardiac cycle could uncover subtle differences. We, however, did not explore this possibility, as we believe that peak tension would most likely be related to the clinical durability of the procedure and therefore the most pertinent parameter to examine and compare.
Differences between human and ovine mitral anatomy must also be considered in extrapolating current data to the clinical arena. In particular, leaflet occlusion area relative to the mitral orifice is smaller in sheep than in humans29 and annular dilatation may exhaust this maximal coaptation area in an ovine heart, but this may not be true in human hearts. Sheep have a less well defined posterior annulus than human and more atrial tissue above and below the line of leaflet insertion,30 but mitral annular dynamics are similar in both sheep and human hearts.16,31 Despite the limitations inherent in this particular experimental preparation, reliable models of cardiac pathophysiology have been established in ovine models.32,33 In the current study we used healthy animals with normal mitral valves, which is a departure from the clinical scenario of severe mitral regurgitation that is often associated with LV chamber enlargement, annular dialatation, and leaflet pathology.
| Acknowledgments |
|---|
Sources of Funding
This work was supported by Grants HL-29589 and HL-67025 from the National Heart, Lung and Blood Institute; Drs Timek, and Lai are Carl and Leah McConnell Cardiovascular Surgical Research Fellows; Dr Timek is a recipient of the Thoracic Surgery Foundation Research Fellowship Award; Dr Nielsen was supported by grants from the Danish Heart Foundation; Dr Timek was also supported by NHLBI INRSA HL-10452; Dr Lai was supported by a fellowship from the American Heart Association, Western States Affiliate.
Disclosures
None.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
2. Enriquez-Sarano M, Schaff HV, Orszulak TA, Tajik AJ, Bailey KR, Frye RL. Valve repair improves the outcome of surgery for mitral regurgitation. A multivariate analysis. Circulation. 1995; 91: 1022–1028.
3. Maisano F, Schreuder JJ, Oppizzi M, Fiorani B, Fino C, Alfieri O. The double-orifice technique as a standardized approach to treat mitral regurgitation due to severe myxomatous disease: surgical technique. Eur J Cardiothorac Surg. 2000; 17: 201–205.
4. De Bonis M, Lapenna E, La Canna G, Ficarra E, Pagliaro M, Torracca L, Maisano F, Alfieri O. Mitral valve repair for functional mitral regurgitation in end-stage dilated cardiomyopathy: role of the "edge-to-edge" technique. Circulation. 2005; 112 (9 Suppl): I402–I408.[Medline] [Order article via Infotrieve]
5. Totaro P, Tulumello E, Fellini P, Rambaldini M, La Canna G, Coletti G, Zogno M, Lorusso R. Mitral valve repair for isolated prolapse of the anterior leaflet: an 11-year follow-up. Eur J Cardiothorac Surg. 1999; 15: 119–126.
6. Umana JP, Salehizadeh B, DeRose JJ, Jr., Nahar T, Lotvin A, Homma S, Oz MC. "Bow-tie" mitral valve repair: an adjuvant technique for ischemic mitral regurgitation. Ann Thorac Surg. 1998; 66: 1640–1646.
7. Alfieri O, De Bonis M, Lapenna E, Regesta T, Maisano F, Torracca L, La Canna G. "Edge-to-edge" repair for anterior mitral leaflet prolapse. Semin Thorac Cardiovasc Surg. 2004; 16: 182–187.[CrossRef][Medline] [Order article via Infotrieve]
8. Alfieri O, Maisano F, De Bonis M, Stefano PL, Torracca L, Oppizzi M, La Canna G. The double-orifice technique in mitral valve repair: a simple solution for complex problems. J Thorac Cardiovasc Surg. 2001; 122: 674–681.
9. Alfieri O, Elefteriades JA, Chapolini RJ, Steckel R, Allen WJ, Reed SW, Schreck S. Novel suture device for beating-heart mitral leaflet approximation. Ann Thorac Surg. 2002; 74: 1488–1493.
10. Fann JI, St Goar FG, Komtebedde J, Oz MC, Block PC, Foster E, Butany J, Feldman T, Burdon TA. Beating heart catheter-based edge-to-edge mitral valve procedure in a porcine model: efficacy and healing response. Circulation. 2004; 110: 988–993.
11. Feldman T, Wasserman HS, Herrmann HC, Gray W, Block PC, Whitlow P, St Goar F, Rodriguez L, Silvestry F, Schwartz A, Sanborn TA, Condado JA, Foster E. Percutaneous mitral valve repair using the edge-to-edge technique: six-month results of the EVEREST Phase I Clinical Trial. J Am Coll Cardiol. 2005; 46: 2134–2140.
12. Maisano F, Caldarola A, Blasio A, De Bonis M, La Canna G, Alfieri O. Midterm results of edge-to-edge mitral valve repair without annuloplasty. J Thorac Cardiovasc Surg. 2003; 126: 1987–1997.
13. Nielsen SL, Timek TA, Lai DT, Daughters GT, Liang D, Hasenkam JM, Ingels NB, Miller DC. Edge-to-edge mitral repair: tension on the approximating suture and leaflet deformation during acute ischemic mitral regurgitation in the ovine heart. Circulation. 2001; 104 (12 Suppl 1): I29–I35.[Medline] [Order article via Infotrieve]
14. Timek TA, Nielsen SL, Lai DT, Tibayan F, Liang D, Daughters GT, Beineke P, Hastie T, Ingels NB, Jr., Miller DC. Mitral annular size predicts Alfieri stitch tension in mitral edge-to-edge repair. J Heart Valve Dis. 2004; 13: 165–173.[Medline] [Order article via Infotrieve]
15. Jimenez JH, Forbess J, Croft LR, Small L, He Z, Yoganathan AP. Effects of annular size, transmitral pressure, and mitral flow rate on the edge-to-edge repair: an in vitro study. Ann Thorac Surg. 2006; 82: 1362–1368.
16. Glasson JR, Komeda M, Daughters GT, Foppiano LE, Bolger AF, Tye TL, Ingels NB, Jr., Miller DC. Most ovine mitral annular three-dimensional size reduction occurs before ventricular systole and is abolished with ventricular pacing. Circulation. 1997; 96 (9 Suppl): II-115–I122;discussion II-123.
17. Nielsen SL, Nygaard H, Fontaine AA, Hasenkam JM, He S, Andersen NT, Yoganathan AP. Chordal force distribution determines systolic mitral leaflet configuration and severity of functional mitral regurgitation. J Am Coll Cardiol. 1999; 33: 843–853.
18. Daughters GT MD, Schwartzkopf A, Mead CW, and Ingels NB. A comparison of two analytical systems for three-dimensional reconstruction from biplane videoradiograms. Proc Comp Cardiol (IEEE). 1988: 79–82.
19. Moon MR, DeAnda A, Jr., Daughters GT, 2nd, Ingels NB, Jr., Miller DC. Experimental evaluation of different chordal preservation methods during mitral valve replacement. Ann Thorac Surg. 1994; 58: 931–943;discussion 943–934.
20. Lapenna E, Torracca L, De Bonis M, La Canna G, Crescenzi G, Alfieri O. Minimally invasive mitral valve repair in the context of Barlows disease. Ann Thorac Surg. May 2005; 79: 1496–1499.
21. St Goar FG, Fann JI, Komtebedde J, Foster E, Oz MC, Fogarty TJ, Feldman T, Block PC. Endovascular edge-to-edge mitral valve repair: short-term results in a porcine model. Circulation. 2003; 108: 1990–1993.
22. McCarthy PM, Starling RC, Wong J, Scalia GM, Buda T, Vargo RL, Goormastic M, Thomas JD, Smedira NG, Young JB. Early results with partial left ventriculectomy. J Thorac Cardiovasc Surg. 1997; 114: 755–763;discussion 763–755.
23. Votta E, Maisano F, Soncini M, Redaelli A, Montevecchi FM, Alfieri O. 3-D computational analysis of the stress distribution on the leaflets after edge-to-edge repair of mitral regurgitation. J Heart Valve Dis. 2002; 11: 810–822.[Medline] [Order article via Infotrieve]
24. Dal Pan F, Donzella G, Fucci C, Schreiber M. Structural effects of an innovative surgical technique to repair heart valve defects. J Biomech. 2005; 38: 2460–2471.[CrossRef][Medline] [Order article via Infotrieve]
25. Nielsen SL, Hansen SB, Nielsen KO, Nygaard H, Paulsen PK, Hasenkam JM. Imbalanced chordal force distribution causes acute ischemic mitral regurgitation: mechanistic insights from chordae tendineae force measurements in pigs. J Thorac Cardiovasc Surg. 2005; 129: 525–531.
26. Daimon M, Shiota T, Gillinov AM, Hayase M, Ruel M, Cohn WE, Blacker SJ, Liddicoat JR. Percutaneous mitral valve repair for chronic ischemic mitral regurgitation: a real-time three-dimensional echocardiographic study in an ovine model. Circulation. 2005; 111: 2183–2189.
27. Timek TA, Liang D, Lai DT, Dagum P, Daughters GT, Ingels NB, Miller DC. Edge-to-edge mitral repair: gradients and three-dimensional annular dynamics in vivo during inotropic stimulation. Eur J Cardiothorac Surg. 2001; 19: 431–437.
28. Timek TA, Nielsen SL, Lai DT, Tibayan FA, Liang D, Rodriguez F, Daughters GT, ingels NB, Miller DC. Edge-to-edge mitral valve repair without ring annuloplasty for acute ischemic mitral regurgitation. Circulation. 2003; 108 (suppl II) II-122–II-127.[Medline] [Order article via Infotrieve]
29. Gorman JH, Jackson BM, Gorman RC, Kelley ST, Gikakis N, Edmunds LH. Papillary muscle discoordination rather than increased annular area facilitates mitral regurgitation after acute posterior myocardial infarction. Circulation. 1997; 96: II-124–II-1247.
30. Walmsley R. Anatomy of human mitral valve in adult cadaver and comparative anatomy of the valve. Br Heart J. 1978; 40: 351–366.
31. Ormiston JA, Shah PM, Tei C, Wong M. Size and motion of the mitral valve annulus in man. I. A two-dimensional echocardiographic method and findings in normal subjects. Circulation. 1982; 64: 113–120.
32. Markovitz LJ, Ratcliffe MB, Bavaria JE, Kreiner G, Iozzo RV, Hargrove WC, Bogen DK, Edmunds LH. Large animal model of left ventricular aneurysm. Ann Thorac Surg. 1989; 48: 838–845.[Abstract]
33. Llaneras MR, Nance ML, Streicher JT, Lima JA, Savino JS, Bogen DK, Deac RF, Ratcliffe MB, Edmunds LH. Large animal model of ischemic mitral regurgitation. Ann Thorac Surg. 1994; 57: 432–439.[Abstract]
This article has been cited by other articles:
![]() |
L. P. Dasi, P. Sucosky, and A. P. Yoganathan Letter by Dasi et al Regarding Article, "Effect of Chronotropy and Inotropy on Stitch Tension in the Edge-to-Edge Mitral Repair" Circulation, July 22, 2008; 118(4): e78 - e78. [Full Text] [PDF] |
||||
![]() |
T. A. Timek, D. T. Lai, D. Liang, D. C. Miller, S. L. Nielsen, G. T. Daughters, and N. B. Ingels Response to Letter Regarding Article, "Effect of Chronotropy and Inotropy on Stitch Tension in the Edge-to-Edge Mitral Repair" Circulation, July 22, 2008; 118(4): e79 - e79. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2007 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |