(Circulation. 2007;115:638-653.)
© 2007 American Heart Association, Inc.
Contemporary Reviews in Cardiovascular Medicine |
From the Cardiac Electrophysiology and Pacemaker Laboratory, Hospital of Saint Raphael, Yale University School of Medicine, New Haven, Conn.
Correspondence to Mark H. Schoenfeld, MD, FACC, FAHA, FHRS, Cardiac Electrophysiology and Pacemaker Laboratory, Hospital of Saint Raphael, Yale University School of Medicine, 330 Orchard St, Suite #210, New Haven, CT 06511. E-mail mschoenfeld{at}srhs.org
Key Words: pacemakers pacing cardioverter-defibrillator, implantable
| Introduction |
|---|
|
|
|---|
| Changing Indications for Device Therapy |
|---|
|
|
|---|
Potential new applications of cardiac pacing have been recognized, such as pacing in patients with neurocardiogenic syncope.3 Ironically, this application has been the subject of more recent debate: No benefit was found in the Second Vasovagal Pacemaker Study.12 Cardiac resynchronization therapy (CRT) has been identified as a class IIA therapy for patients with medically refractory New York Heart Association class III or IV CHF patients with cardiomyopathy, QRS interval of greater than or equal to 130 ms, and EF less than or equal to 35%.13,14 This latter application has allowed for a "marriage" of technologies in 1 device (CRT-ICD) that allows for treatment of CHF and protection against sudden cardiac death in selected patients.
Indications may flourish and, almost as rapidly, languish. The time-honored use of dual-chamber pacing to achieve physiological pacing15 has come under new scrutiny. This concept, as originally proposed, referred to pacing techniques that were designed to allow for normal cardiovascular physiology, preserve atrioventricular (AV) synchrony, and provide heart rate increases (rate adaptiveness) as required by exercise or other physiological demands.16 Although this improved symptoms in many patients and reduced the incidence of atrial fibrillation (AF) compared with single-chamber pacing, no benefit was identified in certain patient subsets,1719 as in the PASE (Pacemaker Selection in the Elderly)17 and UKPACE (United Kingdom Pacing and Cardiovascular Events)18 trials. Furthermore, potential adverse consequences of right ventricular apical pacing have led to the evaluation of alternative sites for pacing the heart,20 and our definition of physiological pacing continues to evolve.21 All of these points attest to the importance of periodically revisiting and updating implantation indications and guidelines as new data from randomized clinical trials become available.7,8,22
| Follow-Up of the Patient After Device Implantation |
|---|
|
|
|---|
Nonetheless, there is no substitute for direct patient contact, with an essential role for the physician to uncover device-related issues.24 In addition to periodic routine visits, more urgent examination may be required, as in the case of an ICD patient with increasingly frequent ICD discharges (as opposed to a single ICD shock, which could be addressed through remote monitoring). History taking is crucial to ensure that symptoms prompting the original implantation remain ameliorated and that device therapy has not resulted in the creation of new symptoms, as in the so-called pacemaker syndrome.31 This syndrome, most commonly caused by the loss of optimal AV synchrony in patients with single-chamber ventricular pacing, consists of cardiovascular and neurological symptoms that include neck/abdominal pulsations, palpitation, fatigue, dyspnea, and/or presyncope; the pathophysiology reflects diminished cardiac output and is associated with signs of systemic hypotension, AV valvular regurgitation, unpleasant cannon A waves in the neck, and pulmonary/hepatic congestion.24
Examination of the wound is important early after implantation to address wound-healing as well as potential infection; the latter is particularly serious because it is rare for a patient to overcome device-related infections without extraction of the entire system.24 Pacing-induced complications such as diaphragmatic stimulation or myopectoral stimulation/inhibition also may be identified. Examination of chest radiographs is occasionally necessary to confirm lead integrity and position.32 The importance of 12-lead electrocardiography cannot be understated, not only for assessment of rhythm but also for confirmation of the appropriate QRS complexpaced morphology associated with various pacing lead locations (typically left bundle-branch block morphology and left-axis deviation with conventional right ventricular apical pacing). This examination is particularly important with the advent of alternative-site pacing, where less typical ECG patterns are observed.20,33
Magnet application to a pacemaker generator is a useful adjunct to follow-up and converts the pacemaker to the asynchronous pacing mode (not inactivation).23 The resulting asynchronous rate is manufacturer specific and changes over the life of the battery as it approaches a preset elective replacement indicator rate. Thus, magnet application is useful in following device longevity and in confirming capture at the programmed output. It may be used to temporarily induce asynchronous pacing in cases of electromagnetic interference (EMI), such as electrocautery. With EMI, external signals may be detected by the device and misconstrued as intracardiac, resulting in inhibition of pacing output, unless sensing is eliminated by either magnet application or device reprogramming. In contrast, magnet application to ICD generators does not affect pacing, but it may temporarily or permanently suspend antitachycardia therapies.
The importance of device interrogation and programmability has been appreciated since the inception of pacemaker therapy.34,35 Unfortunately, there is no universal pacemaker/ICD programmer to date; the manufacturer-specific nature of programmers constrains the nonelectrophysiologist who may not be familiar with or have access to the seemingly vast array of programmers required for follow-up. Ironically, allied professionals, often industry employed, may know more about a particular device and its associated programmer than a given physician; guidelines for such assisted follow-up have been established, and the physician must remain responsible for the device management and the patients overall care.36
Interrogation of a device provides a wealth of information, including but not limited to programmed parameters, such as the pacing rates and pacing mode (Table 1)37; real-time lead and battery impedance, as well as trends; event counters and histograms detailing the percentage of time sensed/paced, automatic mode-switch episodes, and atrial and/or ventricular arrhythmias; real-time and stored intracardiac electrograms; and marker channels that indicate what a device thinks it has seen or is seeing (Figure 1). Interrogation of programmed settings is essential in explaining phenomena that may be misinterpreted as device malfunction. For example, if a pacing rate is lower than the programmed lower tracking rate in a correctly functioning pacemaker, the phenomenon may be understood if interrogation also reveals that a hysteresis function has been programmed, allowing the patients native rhythm to be maintained and to dip down to lower rates before pacing is triggered at a faster rate; this reduces current drain and preserves native rhythm.
|
|
The inability to perform telemetry could indicate that the wrong manufacturers programmer was used, that the telemetry wand has not been positioned adequately over the generator (some models also require a magnet applied to the wand), or that the generator has reached anticipated (or premature) end of life.38 Elevated lead impedances may suggest a fracture in the conductor coil(s), a loose-set screw site at the lead insertion into the header block, or a problem at the tissueelectrode interface39 (Figure 2). Reduction of lead impedance may indicate a breach in lead insulation, often seen at the site of transvenous insertionthe so-called subclavian crush syndrome.40 Analysis of impedance trends may actually provide historical insight as to when lead disruption may have occurred (Figure 3). Reduction in battery voltage and a rise in battery impedance are characteristic of generators over time, providing a method by which to assess generator longevity.
|
|
Device counters may allow for the quantification of events such as high atrial rates, high ventricular rates, and automatic mode-switch episodes, suggesting the interim occurrence of arrhythmias such as AF, supraventricular tachycardia (SVT), ventricular tachycardia (VT), or ventricular fibrillation (VF). Far-field sensing of ventricular depolarization, T waves, or noise by the atrial channel may result, however, in a false-positive incidence of high atrial rate or mode-switch episodes (Figure 4). Thus, pacemaker memory functions do not obviate the need for Holter or event monitoring in all patients with suspected arrhythmias. However, longer durations of individual episodes, on the order of many minutes to days, especially if confirmed by intracardiac electrograms, are more likely to correspond to true atrial dysrhythmias and an increased association with clinical events such as thromboemboli.41,42 Although device counters give a rough indication as to arrhythmia history, stored intracardiac electrograms have been increasingly used for validation of the true arrhythmia in pacemakers43 and ICDs44 (Figure 5).
|
|
| Automatic Algorithms for Detection and Treatment: A Brief Synopsis |
|---|
|
|
|---|
Automatic algorithms that are intended to enhance sensitivity and optimize arrhythmia detection invariably encounter challenges regarding under- and oversensing.45 This is particularly important for ICDs where low-amplitude fibrillatory signals from the ventricle must be adequately sensed while avoiding oversensing in sinus rhythm. ICDs typically have algorithms to automatically adjust the gain or sensing threshold to sense reduced signals; it is, therefore, uncommon for current ICDs to miss VF. Conversely, oversensing may result in spurious discharge or inhibition of pacing output in ICD patients who are pacemaker dependent (Figure 6).
|
Algorithms that allow for rhythm discrimination have become essential to ICD management. Inappropriate or spurious device discharges may be observed in up to 40% of patients, typically from misclassification of sinus tachycardia or from rapidly conducting supraventricular arrhythmias.46 For single-chamber devices, ventricular rate has served as the primary determinant of arrhythmia detection. Additional features such as sudden-onset criteria have aided discrimination of sinus tachycardia (gradual in onset) from VT or VF (typically sudden in onset) by comparing the interval at the onset of the tachycardia against an average of the previous intervals. This will not, however, distinguish VT from SVT. Rate-stability enhancements allow for measurement of beat-to-beat intervals, noting that monomorphic VT is typically regular, whereas AF and polymorphic VT or VF have wider variability when beat-to-beat variability is assessed. This feature proves to be particularly useful in discriminating AF from monomorphic VT. These enhancements reduce the likelihood of delivering spurious shocks for SVT without undersensing or delaying treatment for VT. Beyond this, certain algorithms have been developed to assess morphological characteristics of the electrogram during a patients baseline rhythm. The resulting template may then be compared with the morphological characteristics obtained during a tachycardia, thereby aiding in the discrimination of supraventricular and ventricular rhythms.47 Although it is unpleasant for a patient to receive spurious shocks for misclassified SVT, it is far more hazardous for VT to be undetected; this has led manufacturers to incorporate a programmable time limit for tachycardia duration that, when exceeded, will result in the delivery of VT therapy.
In dual-chamber ICDs, the presence of an atrial lead allows for comparison of ventricular and atrial rates during baseline rhythm and tachycardia, as well as the relative timing of atrial and ventricular activity. Ventricular rates exceeding atrial rates are typical for ventricular tachyarrhythmias with AV dissociation. Intracardiac electrograms are particularly useful in this regard (Figure 5). The Detect SVT trial reported a reduction of inappropriate SVT detection from 39.5% in the single-chamber detection arm compared with 30.9% in the dual-chamber arm.46 Arrhythmia detection by dual-chamber ICDs may be enhanced by combining various discriminators.48
A recent multicenter trial, Comparison of Empiric to Physician-Tailored Programming of Implantable Cardioverter-Defibrillators (EMPIRIC), reported that standardized empirical ICD programming was as effective as physician-tailored programming as long as the empirical strategy included avoidance of detection of nonsustained tacyhcardia, avoidance of detection of SVTs as VT, empirical antitachycardia pacing for slow and fast VTs, and high-output first shocks.49 Nonetheless, misclassification of arrhythmic events is an ongoing concern, and much work remains in the refinement of detection algorithms, particularly in delineating rapid SVTs.50,51
For the sake of completeness, it should be acknowledged that automatic sensing algorithms have been applied to a host of sensors unrelated to either atrial or ventricular depolarization, allowing for rate-adaptive pacing that is particularly well suited to patients exhibiting chronotropic incompetence.24,52 Thus, activity sensors, accelerometers, minute ventilation, and QT-interval sensing have been used alone or in combination to provide for more physiological rate modulation.
Beyond sensing and detection, automatic algorithms have also been increasingly used in the treatment and prevention of arrhythmias. Pacing for bradycardia has been facilitated by automatic capture threshold assessment. The determination of ventricular capture threshold may be made at programmed intervals during the day or on a beat-by-beat basis with corresponding adjustments in output.53 Periodic and frequent assessments of thresholds are important from a safety standpoint because thresholds may change as a function of multiple factors, including changes in autonomic tone, electrolyte and metabolic abnormalities, and concomitant antiarrhythmic drug therapy. On the other hand, the ability to automatically down-regulate outputs when improved thresholds are detected extends device longevity.
Automatic modulation of the AV interval may be accomplished in various ways in current pacing systems. It has long been appreciated that optimal hemodynamics are achieved with shorter AV delay intervals after atrial-sensed compared with atrial-paced events,54 and hemodynamics may be further enhanced by rate-adaptive shortening of the paced AV interval at faster heart rates.55
Conventional right ventricular apical pacing may contribute to cardiac desynchronization and foster or exacerbate CHF in pacer patients.20,56,57 Thus, in the DAVID (Dual Chamber and VVI Implantable Defibrillator) trial,56 committed dual-chamber pacing was associated with increased death or hospitalization for CHF in ICD patients with depressed EF compared with patients programmed to an intentionally slow backup ventricular-paced mode. MOST (Mode Selection Trial) identified an increased rate of CHF hospitalization associated with ventricular-paced rhythm in pacemaker patients with sinus node dysfunction.57 What options, then, are available to minimize cardiac desynchronization induced by right ventricular apical pacing? Reprogramming to atrial pacing mode is one alternative, recognizing, however, that there is a finite, albeit small progression to AV block in patients with sick sinus syndrome, approximating 0.6% of patients per year.58 Programming a fixed long AV delay may minimize ventricular pacing, but it is not always feasible.59,60 In recognition of this, some manufacturers have incorporated conduction search algorithms that automatically search for intrinsic ventricular events that, if not sensed during an extension of the AV interval, will resume stimulation at the programmed AV delay61 (Figure 7).
|
In an effort to mimic naturally occurring circadian rhythms, some devices allow for programming of algorithms that will allow for slower pacing rates nocturnally.62 Another hysteresis function allows for pacing to kick in only when an abrupt rate drop is detected, applied with mixed results in the management of patients with neurocardiogenic syncope.12 Rate-smoothing programs are designed to prevent sudden changes in ventricular cycle length that may be responsible for either inducing pause-dependent arrhythmias or exacerbating symptoms of cardiac irregularity that are often encountered in AF.
Automatic mode switching is, undoubtedly, one of the most useful and widely applied algorithms currently available; it provides for the detection of atrial tachyarrhythmias and conversion from dual-chamber to single-chamber ventricular pacing to preempt pacer-mediated tachycardia, with reversion to dual-chamber pacing once the sinus mechanism has reemerged63,64 (Figure 8). Oversensing, which results in false-positive automatic mode-switching responses, may be observed, most commonly from far-field sensing of the end of the QRS complex by the atrial channel, especially in the setting of noise and low right atrial implants63 (Figure 4). An analysis of automatic mode-switching episodes may provide the clinician with an estimate of the incidence of atrial dysrhythmias, prompting clinical decisions such as institution of anticoagulation, antiarrhythmic drug therapy, or even ablation; the above limitations of algorithm analysis regarding sensitivity and specificity must, however, be considered.
|
Finally, some commentary is warranted regarding the potential utility of pacing algorithms for the prevention and/or treatment of AF. A variety of different methodologies have been developed to increase the atrial pacing rate when the native rhythm emerges and to periodically reduce the search rate for intrinsic atrial activity (dynamic atrial overdrive),65 to transiently use higher rate pacing after mode-switch episodes or spontaneous atrial ectopy, and to pace rapidly alone or in combination with shock therapy to terminate established arrhythmias.66 In general, evidence-based data in this area are sparse, and much work is needed to clarify whether there is any role for the use of permanent pacing in the prevention or treatment of AF, as recently noted by a joint working group of the American Heart Association and the Heart Rhythm Society.67
| Alternative Site Pacing and CRT |
|---|
|
|
|---|
The general cardiologist may be called on to interpret chest radiographs and electrocardiograms that may become increasingly complex with these alternative pacing modalities. Recently, electrocardiographic demonstration of a ratio of 1 or greater in the R/S wave amplitude in lead V1 has been shown to reliably detect left ventricular capture in patients with cardiac resynchronization devices33 (Figure 9). Changes in morphology may also be appreciated during assessment of pacing thresholds in biventricular systems (Figure 10).
|
|
It is beyond the scope of the present article to detail the burgeoning area of CRT. Selection of optimal candidates for this therapy has typically included New York Heart Association class II or III CHF patients with a widened QRS complex. Although a wider baseline complex (greater than 150 ms versus 120 to 150 ms) in the PATH-CHF (Pacing Therapies in Congestive Heart Failure) study correlated with CRT benefit,74 other studies have demonstrated that QRS complex duration is a poor predictor of the response to CRT75; in addition, the extent of shortening of the QRS complex with biventricular pacing has not predicted better individual outcomes with CRT.76 Indeed, attention has been directed increasingly at baseline markers of mechanical dyssynchrony as assessed by echocardiography and tissue Doppler imaging.77,78 Thus, patients with a wide QRS complex, that is, electrical dyssychrony (but without evidence of mechanical dyssynchrony), may not prove to be ideal candidates for CRT. Mechanical dyssynchrony may exist in various forms, including AV dyssynchrony (delay between atrial and ventricular contraction), interventricular dyssynchrony (delayed activation of the left ventricle relative to the right ventricle), and intraventricular dyssynchrony (delayed activation of 1 segment of the left ventricle [LV] relative to another).
It remains unclear how the individual forms of dyssynchrony contribute to CHF; equally unclear is the extent to which echocardiographic optimization of these individual forms will ameliorate CHF, and when and how such optimization should be performed. Confounding this problem is the difficult issue of assessing the clinical response to CRT. There is no consensus on what constitutes an adequate response to CRT or on what defines a nonresponder.79,80 If, after 6 months of therapy, there is either a lack of symptomatic improvement or worsening CHF, a variety of contributing factors may be considered: poor patient selection, development of AF, change in cardiac substrate with worsening cardiomyopathy or ischemia, loss of LV or RV capture, and lack of ideal cardiac resynchronization (nonoptimal AV and VV delays, poor LV lead position).79 Optimization of the AV delay may increase stroke volume through an increase in preload and may be achieved by guiding programming AV-interval changes through simultaneous echocardiographic assessment of the mitral Doppler inflow pattern.77,81 Left bundle-branch block typically delays activation, resulting in fusion of the E and A waves. The optimal AV delay is the shortest possible delay that allows for complete diastolic ventricular filling without interfering with the atrial contribution; the echocardiographic representation of this is separation of the E and A waves without truncation of the A wave (Figure 11). This technique, though widely applied, has yet to be validated in LV-based pacing.82 Optimization of the VV interval, correcting for interventricular dyssynchrony between RV and LV contraction, is a programmable feature available in some devices that may contribute to hemodynamic and symptomatic improvement.83,84
|
Certain ICDs have the ability to assess changes in intrathoracic impedance as measured between the pacing lead and generator; an inverse correlation has been demonstrated between intrathoracic impedance and pulmonary capillary wedge pressure, often heralding the onset of fluid overload.85 Application of this technology may allow the practitioner to institute preemptive changes in medical therapy when such changes are detected, thereby avoiding hospitalization (Figure 12). Thus, management of CHF in patients with CRT devices may well require extensive collaboration among the general cardiologist, electrophysiologist, echocardiographer, and heart failure specialist.
|
| Drugs, Devices, and EMI |
|---|
|
|
|---|
25%; subsequent therapy occurred sooner and unpredictably, suggesting benefit of initiating antiarrhythmic drug therapy soon after the first symptomatic ICD therapy. If frequent nonsustained VT is evident at the time of implantation, such drugs may be considered even sooner. What is less clear is the role of prophylactic antiarrhythmic drugs in primary prevention patients who have yet to demonstrate a sustained arrhythmia. Although antiarrhythmic drugs such as sotalol have proven useful in reducing the incidence of arrhythmia occurrences and ICD discharges,89 rate slowing of VT may be observed, resulting in underdetection of arrhythmias by ICDs. Although some antiarrhythmic drugs such as sotalol may reduce defibrillation thresholds, other agents (notably amiodarone and mexiletine) may elevate defibrillation thresholds90; the resulting compromise of safety margin between the energy that is required and the energy that is deliverable for defibrillation may have disastrous consequences. Flecainide is particularly noteworthy for elevating pacing thresholds. Thus, the addition of antiarrhythmic agents to the medical program of a patient with a device warrants periodic attention to programmed settings and sensing/pacing/defibrillation thresholds.
Potential sources of EMI are ubiquitous, especially in the hospital environment and workplace; therefore, it is critical for the general cardiologist to be aware of potential risks to the device patient and precautions that may be warranted23,91,92 (Table 2). In the surgical arena, electrocautery may be oversensed by a patients intracardiac device, resulting in transient inhibition of pacer output or in triggering of inappropriate ICD shocks (Figure 6). Particularly in the pacemaker-dependent patient, devices should be programmed to an asynchronous or triggered mode to preempt inhibition of pacing output, and the tachycardia-detection mode of an ICD should be temporarily inactivated. One of the most frequent settings in which this scenario arises is the endoscopy suite; notably, no studies have been published examining whether, or how often, cautery during endoscopy procedures interacts with ICDs. In general, electrocautery should be minimized; when electrocautery is required, it should use short bursts of bipolar cautery with the grounding pad placed distant from the generator. Rarely, electrocautery and defibrillation may produce irreversible damage to the generator or reset the device to a backup or noise-reversion mode (reprogramming of a device to an alternative mode); transient undersensing and rises in pacing thresholds may also be observed. When undertaking external cardioversion, anteroposterior paddles with the least possible energy should be used to minimize trauma to pacemaker systems. Devices should be interrogated postoperatively or after cardioversion to assess for threshold rises, reset modes, or changes in the tachycardia mode.
|
Other sources of EMI include magnetic resonance imaging (MRI), lithotripsy, radiofrequency ablation, and radiation therapy. Potential interactions with MRI include pacing inhibition, inappropriate ICD discharges, rapid pacing, mechanical pull and rotation of the device, and device reprogramming.93 It has been suggested that certain pacemaker and ICD systems may be MRI safe.94 Alternatively, certain strategies to minimize complications have been suggested, including the use of less powerful MRI machines; imaging limited to extremities (ie, remote from the device); careful reprogramming of the intracardiac device, including asynchronous modes and maximal pacing output; selection of appropriate spin sequences; limitation of MRI to patients who are not pacemaker dependent; and careful, continuous periprocedure monitoring.95 In general, however, the present use of MRI should be considered contraindicated in the patient with a pacemaker or ICD.
Extracorporeal shockwave lithotripsy, in contrast, may be undertaken as long as there is appropriate distancing of the device from the focal point of lithotripsy, to minimize pacemaker inhibition or irregular sensing.96 Radiation therapy may pose the risk of random failure of individual components of the device, sudden loss of output, alternations in programming, rate runaway (nonphysiological rapid delivery of pacing stimuli to the heart), and premature battery depletion97,98; nonetheless, it may be unavoidable in certain device patients. Strategies to minimize this risk entail close discussion with the radiation therapist on appropriate shielding of the generator; rarely, repositioning of the generator or placement of a new system in the contralateral chest may be required.23 The EMI associated with radiofrequency current during catheter-ablation procedures may transiently cause reversion to asynchronous pacing or result in pacer inhibition. However, it is frequently the case that no significant impact on thresholds, impedances, or device function is observed.24,99 Reprogramming a pacemaker to asynchronous mode, inactivating antitachycardia therapies, and reassessing the device after the procedure are recommended approaches for patients undergoing radiofrequency ablation.
Although less common now, patients may have both a pacemaker and a separate ICD device. Clinical interactions between the 2 may include transient failure by the pacer to sense or capture post-ICD discharge; oversensing of the pacemaker stimulus by the ICD, leading to double counting; ICD failure to sense VF, resulting from pacer stimulus oversensing; and pacemaker reprogramming, caused by ICD discharge.100 In ICD patients who require a separate pacemaker, a bipolar pacing system is indicated to preempt these sensing issues. Other devices may rarely trigger device discharge in an ICD patient, notably, transcutaneous electronic nerve-stimulation units.101
In the nonmedical environment, it has been recognized that cellular telephones may interfere with the function of implanted devices, with the potential for any type of interference being 20% in one study with associated symptoms in 7.2% with the phone placed directly over the generator; the most common pacemaker interactions were tracking interference on the atrial channel, followed by asynchronous pacing and ventricular inhibition.102 Distancing the phone from the generator or using the contralateral ear is the recommended approach. Rarely, antitheft surveillance systems in retail centers may result in ICD discharge or pacemaker triggering. Patients should be advised to walk through the gates without leaning on them or lingering.103 When in doubt regarding the potential impact of any type of EMI, an inquiry with the manufacturer should be made.
| Determinants of Frequent ICD Shocks and the Management of VT Storm |
|---|
|
|
|---|
| Device Advisories |
|---|
|
|
|---|
| Participation in Sports and Driving by the Patient With a Device |
|---|
|
|
|---|
Patients presenting with impaired consciousness while driving a motorized vehicle, in the absence of intoxication or falling asleep, are frequently found to have arrhythmias as the putative cause, as identified by tilt-table testing and/or electrophysiological study.121 It is not uncommon for patients to have arrhythmia-type symptoms while driving; nonetheless, motor vehicle accidents are infrequent, even when associated with ICD discharge. Thus, in the AVID (Antiarrhythmics Versus Implantable Defibrillators) trial, 55 accidents were reported during 1619 patient-years of follow-up after resumption of driving by patients with documented sustained VT/VF (3.4% per patient-year), and only 11% of these accidents were preceded by symptoms of possible arrhythmia (0.4% per patient-year).122 Extensive recommendations regarding driving and related activities were published 1 decade ago.123 ICD patients with sustained VT or VF with structural heart disease were precluded from commercial driving but could resume private driving 6 months after device implantation if they were persistently free from ICD discharge. Pacemaker patients who were pacemaker dependent were only restricted from driving for the first week (private driving) or first month (commercial driving) after implantation. These recommendations predate the era of primary prevention ICD implantation, for which greater leniency may be anticipated; as such, the recommendations warrant updating and are, in any case, subject to the vagaries of local statutes. Of interest, when surveyed more than a decade ago, only 8 states in the United States had specific laws for patients with arrhythmias, and 74% of physician respondents (either general cardiologists or electrophysiologists) did not know their own states laws regarding driving by patients with ventricular arrhythmias.124
| Conclusion |
|---|
|
|
|---|
| Acknowledgments |
|---|
A member of Dr Schoenfelds household is a shareholder for Boston Scientific.
| References |
|---|
|
|
|---|
2. Mirowski M, Reid PR, Mower MM, Watkins L, Gott VL, Schauble JF, Langer A, Heilman MS, Kolenik SA, Fischell RE, Weisfeldt ML. Termination of malignant ventricular arrhythmias with an implanted automatic defibrillator in human beings. N Engl J Med. 1980; 303: 322324.[Medline] [Order article via Infotrieve]
3. Gregoratos G, Abrams J, Epstein AE, Freedman RA, Hayes DL, Hlatky MA, Kerber RE, Naccarelli GV, Schoenfeld MH, Silka MJ, Winters SL. ACC/AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices: summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/NASPE Committee to Update the 1998 Pacemaker Guidelines). Circulation. 2002; 106: 21452161.
4. Voigt A, Ezzeddine R, Barrington W, Obiaha-Ngwu O, Ganz LI, London B, Saba S. Utilization of implantable cardioverter-defibrillators in survivors of cardiac arrest in the United States from 1996 to 2001. J Am Coll Cardiol. 2004; 44: 855858.
5. Schoenfeld MH. Manpower concerns in cardiac electrophysiology and pacing. Pacing Clin Electrophysiol. 1995; 18: 19771979.[CrossRef][Medline] [Order article via Infotrieve]
6. Stevenson WG, Chaitman BR, Ellenbogen KA, Epstein AE, Gross WL, Hayes DL, Strickberger SA, Sweeney MO. Clinical assessment and management of patients with implanted cardioverter-defibrillators presenting to nonelectrophysiologists. Circulation. 2004; 110: 38663869.
7. Goldberger Z, Lampert R. Implantable cardioverter-defibrillators: expanding indications and technologies. JAMA. 2006; 295: 809818.
8. Schoenfeld MH. Updating the guidelines for pacemaker/defibrillator device implantation: an historical perspective and a guide to their interpretation. J Cardiovasc Electrophys. 2002; 13: 12001201.[CrossRef][Medline] [Order article via Infotrieve]
9. Moss AJ, Zareba W, Hall WJ, Klein H, Wilber DJ, Cannom DS, Daubert JP, Higgins SL, Brown MW, Andrews ML. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002; 346: 877883.
10. Priori SG, Napolitano C, Gasparini M, Pappone C, Della Bella P, Giordano U, Bloise R, Giustetto C, De Nardis R, Grillo M, Ronchetti E, Faggiano G, Nastoli J. Natural history of Brugada syndrome: insights for risk stratification and management. Circulation. 2002; 105: 13421347.
11. Bardy GH, Lee KL, Mark DB, Poole JE, Packer DL, Boineau R, Domanski M, Troutman C, Anderson J, Johnson G, McNulty SE, Clapp-Channing N, Davidson-Ray LD, Fraulo ES, Fishbein DP, Luceri RM, Ip JH. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med. 2005; 352: 225237.
12. Connolly SJ, Sheldon R, Thorpe KE, Roberts RS, Ellenbogen KA, Wilkoff BL, Morillo C, Gent M. Pacemaker therapy for prevention of syncope in patients with recurrent severe vasovagal syncope: Second Vasovagal Pacemaker Study (VPS II): a randomized trial. JAMA. 2003; 289: 22242229.
13. Cazeau S, Leclercq C, Lavergne T, Walker S, Varma C, Linde C, Garrigue S, Kappenberger L, Haywood GA, Santini M, Bailleul C, Daubert JC; Multisite Stimulation in Cardiomyopathies (MUSTIC) Study Investigators. Effects of multisite biventricular pacing in patients with heart failure and intraventricular conduction delay. N Engl J Med. 2001; 344: 873880.
14. Abraham WT, Fisher WG, Smith AL, Delurgio DB, Leon AR, Loh E, Kocovic DZ, Packer M, Clavell AL, Hayes DL, Ellestad M, Messenger J. Randomized controlled trial of cardiac resynchronization in chronic heart failure. N Engl J Med. 2002; 346: 18451853.
15. Connolly SJ, Kerr CR, Gent M, Roberts RS, Yusuf S, Gillis AM, Sami MH, Tlajic M, Tang AS, Klein GJ, Lau C, Newman DM. Effects of physiologic pacing versus ventricular pacing on the risk of stroke and death due to cardiovascular causes. Canadian Trial of Physiologic Pacing Investigators. N Engl J Med. 2000; 342: 13851391.
16. Tse H-F, Lau C-P. Clinical trials for cardiac pacing in bradycardia: the end or the beginning? Circulation. 2006: 114: 35.[CrossRef]
17. Lamas GA, Orav EJ, Stambler BS, Kellenbogen KA, Sgarbossa EG, Huang SKS, Marinchak RA, Estes NAM 3rd, Mitchell GF, Lieberman EH, Mangione CM, Goldman L. Quality of life and clinical outcomes in elderly patients treated with ventricular pacing as compared with dual-chamber pacing. N Engl J Med. 1998; 338: 10971104.
18. Toff WD, Camm AJ, Skehan JD. Single-chamber versus dual-chamber pacing for high-grade atrioventricular block. N Engl J Med. 2005; 353: 145155.
19. Healey JS, Toff WD, Lamas GA, Andersen HR, Thorpe KE, Ellenbogen KA, Lee KL, Skene AM, Schron EB, Skehan JD, Goldman L, Roberts RS, Camm AJ, Yusuf S, Connolly SJ. Cardiovascular outcomes with atrial-based pacing compared with ventricular pacing: meta-analysis of randomized trials, using individual patient data. Circulation. 2006; 114: 1117.
20. Schoenfeld MH. Alternative site pacing to promote cardiac synchrony: has conventional pacing become unconventional? J Am Coll Cardiol. 2006; 47: 19461948.
21. Ellenbogen KA, Wood MA. Pacemaker selection: the changing definition of physiologic pacing. N Engl J Med. 2005; 353: 202204.
22. Trohman RG, Kim MH, Pinski SL. Cardiac pacing: the state of the art. Lancet. 2004; 364: 17011719.[CrossRef][Medline] [Order article via Infotrieve]
23. Schoenfeld MH, Blitzer ML. Follow-up assessments of the pacemaker patient. In: Ellenbogen KA, Wood MA, eds. Cardiac Pacing and ICDs. 4th ed. Malden, Mass: Blackwell Publishing; 2005: 500543.
24. Schoenfeld MH. Pacemaker insertion, revision, extraction, and follow-up. In: Saksena S, Camm AJ, eds. Electrophysiological Disorders of the Heart. Philadelphia, Pa: Elsevier; 2005: 767807.
25. Hayes DL, Naccarelli GV, Furman S, Parsonnet V, Reynolds D, Goldschlager N, Gillette P, Maloney JD, Saxon L, Leon A, Daoud E. NASPE training requirements for cardiac implantable electronic devices: selection, implantation, and follow-up. Pacing Clin Electrophysiol. 2003; 26: 15561562.[CrossRef][Medline] [Order article via Infotrieve]
26. Curtis AB, Langberg JJ, Tracy CM. Clinical competency statement: implantation and follow-up of cardioverter-defibrillators. J Cardiovasc Electrophysiol. 2001; 12: 280284.[CrossRef][Medline] [Order article via Infotrieve]
27. Senges-Becker JC, Klostermann M, Becker R, Bauer A, Siegler KE, Katus HA, Schoels W. What is the "optimal" follow-up schedule for ICD patients? Europace. 2005; 7: 319326.
28. Schoenfeld MH, Compton SJ, Mead RH, Weiss DN, Sherfesee L, Englund J, Mongeon LR. Remote monitoring of implantable cardioverter-defibrillators: a prospective analysis. Pacing Clin Electrophysiol. 2004; 27: 757763.[CrossRef][Medline] [Order article via Infotrieve]
29. Schoenfeld MH, Reynolds DW. Sophisticated remote implantable cardioverter-defibrillator follow-up: a status report. Pacing Clin Electrophysiol. 2005; 28: 235240.[CrossRef][Medline] [Order article via Infotrieve]
30. Schoenfeld MH, Markowitz HT. Device follow-up in the age of automaticity. Pacing Clin Electrophysiol. 2000; 23: 803806.[CrossRef][Medline] [Order article via Infotrieve]
31. Ausubel K, Furman S. The pacemaker syndrome. Ann Intern Med. 1985; 103: 420429.
32. Burney K, Burchard F, Papouchado M, Wilde P. Cardiac pacing systems and implantable cardiac defibrillators (ICDs): a radiological perspective of equipment, anatomy and complications. Clin Radiol. 2004; 59: 699708.[CrossRef][Medline] [Order article via Infotrieve]
33. Ammann P, Sticherling C, Kalusche D, Eckstein J, Bernheim A, Schaer B, Osswald S. An electrocardiogram-based algorithm to detect loss of left ventricular capture during cardiac resynchronization therapy. Ann Intern Med. 2005; 142: 968973.
34. Schoenfeld MH. A primer on pacemaker programmers. Pacing Clin Electrophysiol. 1993; 16: 20442052.[CrossRef][Medline] [Order article via Infotrieve]
35. Schoenfeld MH. Pacemaker programmers: an updated synopsis. Cardiac Electrophysiol Rev. 1999; 3: 2023.[CrossRef]
36. Gura MT, Bubien RS, Belco KM, Taibi B, Schurig L, Wilkoff BL. North American Society of Pacing and Electrophysiology. Standards of professional practice for the allied professionals in pacing and electrophysiology. Pacing Clin Electrophysiol. 2003; 26: 127131.[CrossRef][Medline] [Order article via Infotrieve]
37. Bernstein AD, Daubert JC, Fletcher RD, Hayes DL, Luderitz B, Reynolds DW, Schoenfeld MH, Sutton R. The revised NASPE/BPEG generic code for antibradycardia, adaptive-rate and multisite pacing. Pacing Clin Electrophysiol. 2002; 25: 260264.[CrossRef][Medline] [Order article via Infotrieve]
38. Blitzer ML, Marieb MA, Schoenfeld MH. Inability to communicate with ICDs: an underreported failure mode. Pacing Clin Electrophysiol. 2001; 24: 1315.[CrossRef][Medline] [Order article via Infotrieve]
39. Haddad L, Padula LE, Moreau M, Schoenfeld MH. Troubleshooting implantable cardioverter defibrillator system malfunctions: the role of impedance measurements. Pacing Clin Electrophysiol. 1994; 17: 14561461.[CrossRef][Medline] [Order article via Infotrieve]
40. Magney JE, Flynn DM, Parsons JA, Staplin DH, Chin-Purcell MV, Milstein S, Hunter DW. Anatomical mechanisms explaining damage to pacemaker leads, defibrillator leads, and failure of central venous catheters adjacent to the sternoclavicular joint. Pacing Clin Electrophysiol. 1993; 16: 445457.[CrossRef][Medline] [Order article via Infotrieve]
41. Plummer CJ, McComb JM; STOP AF Trial. Detection of atrial fibrillation by permanent pacemakers: observations from the STOP AF trial. Card Electrophys Rev. 2003; 7: 333340.[CrossRef]
42. Capucci A, Santini M, Padeletti L, Gulizia M, Botto G, Boriani G, Ricci R, Favale S, Zolezzi F, Di Belardino N, Molon G, Drago F, Villani GQ, Mazzini E, Vimercati M, Grammatico A. Monitored atrial fibrillation duration predicts arterial embolic events in patients suffering from bradycardia and atrial fibrillation implanted with antitachycardia pacemakers. J Am Coll Cardiol. 2005; 46: 19131920.
43. Pollak, WM, Simmons JD, Interian A Jr, Castellanos A, Myerburg RJ, Mitrani RD. Pacemaker diagnostics: a critical appraisal of current technology. Pacing Clin Electrophysiol. 2003; 26: 7698.[CrossRef][Medline] [Order article via Infotrieve]
44. Swerdlow CD, Friedman PA. Advanced ICD troubleshooting: part I. Pacing Clin Electrophysiol. 2005; 28: 13221346.[CrossRef][Medline] [Order article via Infotrieve]
45. Richardson K, Al-Ahmad A, Wang PJ. Implantable defibrillator sensing and discrimination algorithms. In: Wang PJ, ed. New Arrhythmia Technologies. Malden, Mass: Blackwell Futura; 2005: 163177.
46. Friedman PA, McClelland RL, Bamlet WR, Acosta H, Kessler D, Munger TM, Kavesh NG, Wood M, Daoud E, Massumi A, Schuger C, Shorofsky S, Wilkoff B, Glikson M. Dual-chamber versus single-chamber detection enhancements for implantable defibrillator rhythm diagnosis: the Detect Supraventricular Tachycardia study. Circulation. 2006; 113: 28712879.
47. Swerdlow CD, Brown ML, Lurie K, Zhang J, Wood NM, Olson WH, Gillberg JM. Discrimination of ventricular tachycardia from supraventricular tachycardia by a downloaded wavelet-transform morphology algorithm: a paradigm for development of implantable cardioverter-defibrillator detection algorithms. J Cardiovasc Electrophysiol. 2002; 13: 432441.[CrossRef][Medline] [Order article via Infotrieve]
48. Glikson M, Swerdlow CD, Gurevitz OT, Daoud E, Shivkumar K, Wilkoff B, Shipman T, Friedman PA. Optimal combination of discriminators for differentiating ventricular from supraventricular tachycardia by dual-chamber defibrillators. J Cardiovasc Electrophysiol. 2005; 16: 732739.[CrossRef][Medline] [Order article via Infotrieve]
49. Wilkoff BL, Ousdigian KT, Sterns LD, Wang ZJ, Wilson RD, Morgan JM. A comparison of empiric to physician-tailored programming of implantable cardioverter-defibrillators: results from the prospective randomized multicenter EMPIRIC trial. J Am Coll Cardiol. 2006; 48: 330339.
50. Aliot E, Nitzsche R, Ripart A. Arrhythmia detection by dual-chamber implantable cardioverter-defibrillators: a review of current algorithms. Europace. 2004; 6: 273286.
51. Wilkoff BL, Kuhlkamp V, Volosin K, Ellenbogen K, Waldecker B, Kacet S, Gillberg JM, DeSouza CM. Critical analysis of dual-chamber implantable cardioverter-defibrillator arrhythmia detection: results and technical considerations. Circulation. 2001; 103: 381386.
52. Leung SK, Lau CP. Developments in sensor-driven pacing. Cardiol Clin. 2000; 18: 113155.[CrossRef][Medline] [Order article via Infotrieve]
53. Clarke M, Liu B, Schuller H, Binner L, Kennergren C, Guerola M, Weinmann P, Ohm OJ. Automatic adjustment of pacer stimulation output correlated with continuously monitored capture thresholds: a multicenter study. European Microny Study Group. Pacing Clin Electrophysiol. 1998; 21: 15671575.[CrossRef][Medline] [Order article via Infotrieve]
54. Janosik DL, Pearson AC, Buckingham TA, Labovitz AJ, Redd RM. The hemodynamic benefit of differential atrioventricular delay intervals for sensed and paced atrial events during physiologic pacing. J Am Coll Cardiol. 1989; 14: 499507.[Abstract]
55. Ritter P, Daubert C, Mabo P, Descaves C, Gouffault J. Haemodynamic benefit of a rate-adapted A-V delay in dual chamber pacing. Eur Heart J. 1989; 10: 637646.
56. The DAVID Trial Investigators. Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) trial. JAMA. 2002; 288: 31153123.
57. Sweeney MO, Hellkamp AS, Ellenbogen KA, Greenspon AJ, Freedman RA, Lee KL, Lamas GA. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation. 2003; 23: 29322937.
58. Andersen HR, Nielsen JC, Thomsen PE, Thuesen L, Vesterlund T, Pedersen AK, Mortensen PT. Atrioventricular conduction during long-term follow-up of patients with sick sinus syndrome. Circulation. 1998; 98: 13151321.
59. Nielsen JC, Pedersen AK, Mortensen PT, Andersen HR. Programming a fixed long atrioventricular delay is not effective in preventing ventricular pacing in patients with sick sinus syndrome. Europace. 1999; 1: 113120.
60. Sweeney MO, Prinzen FW. A new paradigm for physiologic ventricular pacing. J Am Coll Cardiol. 2006; 47: 282288.
61. Sweeney MO, Shea JB, Fox V, Adler S, Nelson L, Mullen TJ, Belk P, Casavant D, Sheldon T. Randomized pilot study of a new atrial-based minimal ventricular pacing mode in dual-chamber implantable cardioverter-defibrillators. Heart Rhythm. 2004; 1: 160167.[CrossRef][Medline] [Order article via Infotrieve]
62. Duru F, Bloch KE, Weilenmann D, Candinas R. Clinical evaluation of a pacemaker algorithm that adjusts the pacing rate during sleep using activity variance. Pacing Clin Electrophysiol. 2000; 23: 15091515.[CrossRef][Medline] [Order article via Infotrieve]
63. Lau C-P, Leung S-K, Tse H-F, Barold SS. Automatic mode switching of implantable pacemakers: I: principles of instrumentation, clinical, and hemodynamic considerations. Pacing Clin Electrophysiol. 2002; 25: 967985.[CrossRef][Medline] [Order article via Infotrieve]
64. Lau C-P, Leung S-K, Tse H-F, Barold SS. Automatic mode switching of implantable pacemakers: II: clinical performance of current algorithms and their programming. Pacing Clin Electrophysiol. 2002; 25: 10941113.[CrossRef][Medline] [Order article via Infotrieve]
65. Carlson MD, Ip J, Messenger J, Beau S, Kalbfleisch S, Gervais P, Cameron DA, Duran A, Val-Mejias J, Mackall J, Gold M. A new pacemaker algorithm for the treatment of atrial fibrillation: results of the atrial dynamic overdrive pacing trial (ADOPT). J Am Coll Cardiol. 2003; 42: 627633.
66. Lee MA, Weacher R, Pollak S, Kremers MS, Naik AM, Silverman R, Tuzi J, Wang W, Johnson LJ, Euler DE. The effect of atrial pacing therapies on atrial tachyarrhythmia burden and frequency. J Am Coll Cardiol. 2003; 41: 19261932.
67. Knight BP, Gersh BJ, Carlson MD, Friedman PA, McNamara RL, Strickberger SA, Tse HF, Waldo AL. Role of permanent pacing to prevent atrial fibrillation: science advisory from the American Heart Association Council on Clinical Cardiology (Subcommittee on Electrocardiography and Arrhythmias) and the Quality of Care and Outcomes Research Interdisciplinary Working Group, in Collaboration with the Heart Rhythm Society. Circulation. 2005; 111: 240243.
68. Schoenfeld MH. Upgrading from right ventricular apical pacing to biventricular pacing. In: Raviele A, ed. Cardiac Arrhythmias 2003: Proceedings of the 8th International Workshop on Cardiac Arrhythmias. Milan, Italy: Springer-Verlag; 2004: 815822.
69. Leon AR, Greenberg JM, Kanuru N, Baker CM, Mera FV, Smith AL, Langberg JJ, DeLurgio DB. Cardiac resynchronization in patients with congestive heart failure and chronic atrial fibrillation: effect of upgrading to biventricular pacing after chronic right ventricular pacing. J Am Coll Cardiol. 2002; 39: 12581263.
70. Occhetta E, Bortnik M, Magnani A, Francalacci G, Piccinino C, Plebani L, Marino P. Prevention of ventricular desynchronization by permanent parahisian pacing after atrioventricular node ablation in chronic atrial fibrillation: a cross-over, blinded, randomized study versus apical right ventricular pacing. J Am Coll Cardiol. 2006; 47: 19381945.
71. Bristow MR, Saxon LA, Boehmer J, Krueger S, Kass DA, De Marco T, Carson P, DiCarlo L, DeMets D, White BG, DeVries DW, Feldman AM. Comparison of Medical Therapy, Pacing and Defibrillation in Heart Failure (COMPANION) Investigators. Cardiac resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N Engl J Med. 2004; 350: 21402150.
72. Cleland J, Daubert JC, Erdmann E, Freemantle N, Gras D, Kappenberger L, Tavazzi L; Cardiac Resynchronization-Heart Failure (CARE-HF) Study Investigators. The effect of cardiac resynchronization on morbidity and mortality in heart failure. N Engl J Med. 2005; 352: 15391549.
73. Kindermann M, Hennen B, Jung J, Geisel J, Bohm M, Frohlig G. Biventricular versus conventional right ventricular stimulation for patients with standard pacing indication and left ventricular dysfunction: the Homburg Biventricular Pacing Evaluation (HOBIPACE). J Am Coll Cardiol. 2006; 47: 19271937.
74. Auricchio A, Stellbrink C, Butter C, Sack S, Vogt J, Misier AR, Bocker D, Block M, Kirkels JH, Kramer A, Huvelle E. Pacing Therapies in Congestive Heart Failure II Study Group; Guidant Heart Failure Research Group. Clinical efficacy of cardiac resynchronization therapy using left ventricular pacing in heart failure patients stratified by severity of ventricular conduction delay. J Am Coll Cardiol. 2003; 42: 21092116.
75. Achilli A, Sassara M, Ficili S, Pontillo D, Achillia P, Alessi C, De Spirito SD, Guerra R, Patruno N, Serra F. Long-term effectiveness of cardiac resynchronization therapy in patients with refractory heart failure and "narrow" QRS. J Am Coll Cardiol. 2003; 42: 21172124.
76. Molhoek SG, Bax JJ, Boersma E, Van Erven L, Bootsma M, Steendijk P, Van Der Wall EE, Schalij MJ. QRS duration and shortening to predict clinical response to cardiac resynchronization therapy in patients with end-stage heart failure. Pacing Clin Electrophysiol. 2004; 27: 308313.[CrossRef][Medline] [Order article via Infotrieve]
77. Bax JJ, Ansalone G, Breithardt OA, Derumeaux G, Leclercq C, Schalij MJ, Sogaard P, St. John Sutton M, Nihoyannopoulos P. Echocardiographic evaluation of cardiac resynchronization therapy: ready for routine clinical use? A critical appraisal. J Am Coll Cardiol. 2004; 44: 19.
78. Hawkins NM, Petrie MC, MacDonald MR, Hogg KJ, McMurray JJV. Selecting patients for cardiac resynchronization therapy: electrical or mechanical dyssynchrony? European Heart J. 2006; 27: 12701281.
79. Aranda JM Jr, Woo GW, Schofield RS, Handberg EM, Hill JA, Curtis AB, Sears SF, Goff JS, Pauly DF, Conti JB. Management of heart failure after cardiac resynchronization therapy: integrating advanced heart failure treatment with optimal device function. J Am Coll Cardiol. 2005; 46: 21932198.
80. Yu CM, Wing-Hong FJ, Zhang Q, Sandersen JE. Understanding nonresponders of cardiac resynchronization therapy: current and future perspectives. J Cardiovascular Electrophysiol. 2005; 16: 11171124.[CrossRef][Medline] [Order article via Infotrieve]
81. Auricchio A, Stellbrink C, Block M, Sack S, Vogt J, Bakker P, Klein H, Kramer A, Ding J, Salo R, Tockman B, Pochet T, Spinelli J. Effect of pacing chamber and atrioventricular delay on acute systolic function of paced patients with congestive heart failure. Circulation. 1999; 99: 29933001.
82. Kerlan JE, Sawhney NS, Waggoner AD, Chawla MK, Garhwal S, Osborn JL, Faddis MN. Prospective comparison of echocardiographic atrioventricular delay optimization methods for cardiac resynchronization therapy. Heart Rhythm. 2006; 3: 148154.[CrossRef][Medline] [Order article via Infotrieve]
83. Bordachar P, Lafitte S, Reuter S, Sanders P, Jais P, Haissaguerre M, Roudaut R, Garrigue S, Clementy J. Echocardiographic parameters of ventricular dyssynchrony validation in patients with heart failure using sequential biventricular pacing. J Am Coll Cardiol. 2004; 44: 21572165.
84. Sogaard P, Egeblad H, Pedersen AK, Kim WY, Kristensen BO, Hansen PS, Mortensen PT. Sequential versus simultaneous biventricular resynchronization for severe heart failure: evaluation by tissue Doppler imaging. Circulation. 2002; 106: 20782084.
85. Yu CM, Wang L, Chau E, Chan RH, Kong SL, Tang MO, Christensen J, Stadler RW, Lau CP. Intrathoracic impedance monitoring in patients with heart failure: correlation with fluid status and feasibility of early warning preceding hospitalization. Circulation. 2005; 112: 841848.
86. Bollmann A, Husser D, Cannom DS. Antiarrhythmic drugs in patients with implantable cardioverter-defibrillators. Am J Cardiovasc Drugs. 2005; 5: 371378.[CrossRef][Medline] [Order article via Infotrieve]
87. Connolly SJ, Dorian P, Roberts RS, Gent M, Bailin S, Fain ES, Thorpe K, Champagne J, Talajic M, Coutu B, Gronefeld GC, Hohnloser SH. Comparison of b-blockers, amiodarone plus b-blockers, or sotalol for prevention of shocks from implantable cardioverter defibrillators: the OPTIC Study: a randomized trial. JAMA. 2006; 295: 165171.
88. Freedberg NA, Hill JN, Fogel RI, Prystowsky EN. Recurrence of symptomatic ventricular arrhythmias in patients with implantable cardioverter defibrillator after the first device therapy. J Am Coll Cardiol. 2001; 37: 19101915.
89. Pacifico A, Hohnloser SH, Williams JH, Tao B, Saksena S, Henry PD, Prystowsky EN. Prevention of implantable defibrillator shock by treatment with sotalol. N Engl J Med. 1999; 340: 18551862.
90. Pinter A, Dorian P. Interactions between implantable devices and antiarrhythmic drugs. Card Electrophysiol Rev. 2000; 4: 222226.[CrossRef]
91. Blitzer ML, Schoenfeld MH. Effects of electromagnetic interference on implanted cardiac devices. In: Raviele A, ed. Cardiac Arrhythmias. Milan, Italy: Springer-Verlag; 2004: 911918.
92. American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices. Practice advisory for the perioperative management of patients with cardiac rhythm management devices: pacemakers and implantable cardioverter-defibrillators. Anesthesiology. 2005; 103: 186198.[CrossRef][Medline] [Order article via Infotrieve]
93. Kalin R, Stanton MS. Current clinical issues for MRI scanning of pacemaker and defibrillator patients. Pacing Clin Electrophysiol. 2005; 28: 326328.[CrossRef][Medline] [Order article via Infotrieve]
94. Roguin A, Zviman MM, Meininger GR, Rodrigues ER, Dickfeld TM, Bluemke DA, Lardo A, Berger RD, Calkins H, Halperin HR. Modern pacemaker and implantable cardioverter/defibrillator systems can be magnetic resonance imaging safe. In vitro and in vivo assessment of safety and function at 1.5 T. Circulation. 2004; 110: 475482.
95. Gimbel JR, Bailey SM, Tchou PJ, Ruggieri PM, Wilkoff BL. Strategies for the safe magnetic resonance imaging of pacemaker-dependent patients. Pacing Clin Electrophysiol. 2005; 28: 10411046.[CrossRef][Medline] [Order article via Infotrieve]
96. Cooper D, Wilkoff B, Masterson M, Castle L, Belco K, Simmons T, Morant V, Streem S, Maloney J. Effects of extracorporeal shock wave lithotripsy on cardiac pacemakers and its safety in patients with implanted cardiac pacemakers. Pacing Clin Electrophysiol. 1988; 11: 16071616.[CrossRef][Medline] [Order article via Infotrieve]
97. Shehata WM, Daoud GL, Meyer RL. Radiotherapy for patients with cardiac pacemakers: possible risks. Pacing Clin Electrophysiol. 1986; 9: 919922.[CrossRef][Medline] [Order article via Infotrieve]
98. Pinski SL, Trohman RG. Interference in implanted cardiac devices, part II. Pacing Clin Electrophysiol. 2002; 25: 14961509.[CrossRef][Medline] [Order article via Infotrieve]
99. Ellenbogen KA, Wood MA, Stambler BS. Acute effects of radiofrequency ablation of atrial arrhythmias on implanted permanent pacing systems. Pacing Clin Electrophysiol. 1996; 19: 12871295.[CrossRef][Medline] [Order article via Infotrieve]
100. Calkins H, Brinker J, Veltri EP, Guarnieri T, Levine JH. Clinical interactions between pacemakers and automatic implantable cardioverter-defibrillators. J Am Coll Cardiol. 1990; 16: 666673.[Abstract]
101. Philbin D, Marieb MA, Aithal K, Schoenfeld MH. Inappropriate therapy delivered by an ICD as a result of sensed potentials from a transcutaneous electronic nerve stimulation unit. Pacing Clin Electrophysiol. 1998; 21: 20102011.[CrossRef][Medline] [Order article via Infotrieve]
102. Hayes DL, Wang PJ, Reynolds DW, Estes M 3rdI, Griffith JL, Steffens RA, Carlo GL, Findlay GK, Johnson CM. Interference with cardiac pacemakers by cellular telephones. N Engl J Med. 1997; 336: 14731479.
103. Groh WJ, Boschee SA, Engelstein ED, Miles WM, Burton ME, Foster PR, Crevey BJ, Zipes DP. Interactions between electronic article surveillance systems and implantable cardioverter-defibrillators. Circulation. 1999; 100: 387392.
104. Brigadeau F, Kouakam C, Klug D, Marquie C, Duhamel A, Mizon-Gerard F, Lacroix D, Kacet S. Clinical predictors and prognostic significance of electrical storm in patients with implantable cardioverter-defibrillators. Eur Heart J. 2006; 27: 700707.
105. Pratt CM, Greenway PS, Schoenfeld MH, Hibben ML, Reiffel JA. Exploration of the precision of classifying sudden cardiac death: implications for the interpretation of clinical trials. Circulation. 1996; 93: 519524.
106. Singh JP, Hall WJ, McNitt S, Wang H, Daubert JP, Zareba W, Ruskin JN, Moss AJ; MADIT-II Investigators. Factors influencing appropriate firing of the implantable defibrillator for ventricular tachycardia/fibrillation: findings from the Multicenter Automatic Defibrillator Implantation Trial II (MADIT-II). J Am Coll Cardiol. 2005; 46: 17121720.
107. Whang W, Albert CM, Sears SF Jr, Lampert R, Congti JB, Wang PJ, Singh JP, Ruskin JN, Muller JE, Mittleman MA. Depression as a predictor for appropriate shocks among patients with implantable cardioverter-defibrillators: results from the Triggers Of Ventricular Arrhythmias (TOVA) study. J Am Coll Cardiol. 2005; 45: 10901095.
108. Daubert JP, Zareba W, Hall WJ, Schuger C, Corsello A, Leon AR, Andrews ML, McNitt S, Huang DT, Moss AJ. Predictive value of ventricular arrhythmia inducibility for subsequent ventricular tachycardia or ventricular fibrillation in Multicenter Automatic Defibrillator Implantation Trial (MADIT-II) patients. J Am Coll Cardiol. 2006; 47: 98107.
109. Credner SC, Klingenheben T, Mauss O, Sticherling C, Hohnloser SH. Electrical storm in patients with transvenous implantable cardioverter-defibrillators: incidence, management and prognostic implications. J Am Coll Cardiol. 1998; 32: 19091915.
110. Sweeney MO, Wathen MS, Volosin K, Abdalla I, DeGroot PJ, Otterness MF, Stark AJ. Appropriate and inappropriate ventricular therapies, quality of life, and mortality among primary and secondary prevention implantable cardioverter defibrillator patients: results from the pacing fast VT reduces shock therapies (Pain FREE RxII) trial. Circulation. 2005; 111: 28982905.
111. Sadoul N, Mletzko R, Anselme F, Bowes R, Schols W, Kouakam C, Casteigneau G, Luise R, Iscolo N, Aliot E. Incidence and clinical relevance of slow ventricular tachycardia in implantable cardioverter-defibrillator recipients. Circulation. 2005; 112: 946953.
112. Hauser RG, Kallinen L. Deaths associated with implantable cardioverter defibrillator failure and deactivation reported in the United States Food and Drug Administration manufacturer and user facility device experience database. Heart Rhythm. 2004; 4: 399405.[CrossRef]
113. Ellenbogen KA, Wood MA, Shepard RK, Clemo HF, Vaughn T, Holloman K, Dow M, Leffler J, Abeyratne A, Verness D. Detection and management of an implantable cardioverter defibrillator lead failure: incidence and clinical implications. J Am Coll Cardiol. 2003; 41: 7380.
114. Maisel WH, Moynahan M, Zuckerman BD, Gross TP, Tovar OH, Tillman D-B, Schultz DB. Pacemaker and ICD generator malfunctions: analysis of Food and Drug Administration annual reports. JAMA. 2006; 295: 19011906.
115. Schoenfeld MH. Recommendations for implementation of a North American multicenter arrhythmia device/lead database. Pacing Clin Electrophysiol. 1992; 15: 16321636.[CrossRef][Medline] [Order article via Infotrieve]
116. Maisel WH. Pacemaker and ICD generator reliability: meta-analysis of device registries. JAMA. 2006; 295: 19291934.
117. Gould PA, Krahn AD. Complications associated with implantable cardioverter-defibrillator replacement in response to device advisories. JAMA. 2006; 295: 19071911.
118. Carlson MD, Wilkoff BL, Maisel WH, Carlson MD, Ellenbogen KA, Saxon LA, Prystowsky EN, Alpert JS, Cain ME, Ching EA, Curtis AB, Davies DW, Hammill SC, Hauser RG, Lampert RL, Zipes DP. Recommendations from the Heart Rhythm Society task force on device performance policies and guidelines. Heart Rhythm. 2006; 3: 12501273.[CrossRef][Medline] [Order article via Infotrieve]
119. Lampert R, Cannom D, Olshansky B. Safety of sports participation in patients with implantable cardioverter defibrillators: a survey of Heart Rhythm Society members. J Cardiovasc Electrophysiol. 2006; 17: 1115.[CrossRef][Medline] [Order article via Infotrieve]
120. Maron BJ, Zipes DP. Introduction: eligibility recommendations for competitive athletes with cardiovascular abnormalities: general considerations. J Am Coll Cardiol. 2005; 45: 13181321.
121. Blitzer ML, Saliba BC, Ghantous AE, Marieb MA, Schoenfeld MH. Causes of impaired consciousness while driving a motorized vehicle. Am J Cardiol. 2003; 91: 13731374.[CrossRef][Medline] [Order article via Infotrieve]
122. Akiyama T, Powell JL, Mitchell LB, Ehlert FA, Baessler C. Resumption of driving after life-threatening ventricular tachyarrhythmia. N Engl J Med. 2001; 345: 391397.
123. Epstein AE, Miles WM, Benditt DG, Camm AJ, Darling EJ, Friedman PL, Garson A Jr, Harvey JC, Kidwell GA, Klein GJ, Levine PA, Marchlinski FE, Prystowsky EN, Wilkoff BL. Personal and public safety issues related to arrhythmias that may affect consciousness: implications for regulation and physician recommendations: a medical/scientific statement from the American Heart Association and the North American Society of Pacing and Electrophysiology. Circulation. 1996; 94: 11471166.
124. Strickberger SA, Cantillon CO, Friedman PL. When should patients with lethal ventricular arrhythmia resume driving? An analysis of state regulations and physician practices. Ann Int Med. 1991; 115: 560563.
This article has been cited by other articles:
![]() |
M. H. Schoenfeld Transtelephonic Versus Remote Monitoring of Cardiovascular Implantable Electronic Devices Is One Approach to Be Preferred? J. Am. Coll. Cardiol., November 24, 2009; 54(22): 2020 - 2022. [Full Text] [PDF] |
||||
![]() |
A. Proclemer, M. Ghidina, D. Gregori, D. Facchin, L. Rebellato, P. Fioretti, and M. Brignole Impact of the main implantable cardioverter-defibrillator trials in clinical practice: data from the Italian ICD Registry for the years 2005-07 Europace, April 1, 2009; 11(4): 465 - 475. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Swedberg, J. Cleland, M. R. Cowie, M. Nieminen, S. G. Priori, L. Tavazzi, D. J. van Veldhuisen, L. Alonso-Pulpon, J. Camm, K. Dickstein, et al. Successful treatment of heart failure with devices requires collaboration Eur J Heart Fail, December 1, 2008; 10(12): 1229 - 1235. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. E. Epstein, J. P. DiMarco, K. A. Ellenbogen, N.A. M. Estes III, R. A. Freedman, L. S. Gettes, A. M. Gillinov, G. Gregoratos, S. C. Hammill, D. L. Hayes, et al. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices) Developed in Collaboration With the American Association for Thoracic Surgery and Society of Thoracic Surgeons J. Am. Coll. Cardiol., May 27, 2008; 51(21): e1 - e62. [Full Text] [PDF] |
||||
![]() |
Writing Committee Members, A. E. Epstein, J. P. DiMarco, K. A. Ellenbogen, N.A. M. Estes III, R. A. Freedman, L. S. Gettes, A. M. Gillinov, G. Gregoratos, S. C. Hammill, et al. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): Developed in Collaboration With the American Association for Thoracic Surgery and Society of Thoracic Surgeons Circulation, May 27, 2008; 117(21): e350 - e408. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2007 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |