(Circulation. 2007;115:2474-2480.)
© 2007 American Heart Association, Inc.
Arrhythmia/Electrophysiology |
From the Herzzentrum Ludwigshafen, Klinikum Ludwigshafen, Germany.
Correspondence to Thomas Kleemann, MD, Herzzentrum Ludwigshafen, Medizinische Klinik B, Bremserstraße 79, D-67063 Ludwigshafen, Germany. E-mail thomas.kleemann.lu{at}t-online.de
Received September 11, 2006; accepted February 22, 2007.
| Abstract |
|---|
|
|
|---|
Methods and Results A total of 990 consecutive patients who underwent first implantation of an implantable cardioverter-defibrillator between 1992 and May 2005 were analyzed. Median follow-up time was 934 days (interquartile range, 368 to 1870). Overall, 148 defibrillation leads (15%) failed during the follow-up. The estimated lead survival rates at 5 and 8 years after implantation were 85% and 60%, respectively. The annual failure rate increased progressively with time after implantation and reached 20% in 10-year-old leads (P<0.001). Lead defects affected newer as well as older models. Patients with lead defects were 3 years younger at implantation and more often female. Multiple lead implantation was associated with a trend to a higher rate of defibrillation lead defects (P=0.06). The major lead complications were insulation defects (56%), lead fractures (12%), loss of ventricular capture (11%), abnormal lead impedance (10%), and sensing failure (10%).
Conclusions An increasing annual lead failure rate is noted primarily during long-term follow-up and reached 20% in 10-year-old leads. Patients with lead defects are younger and more often female.
Key Words: defibrillation cardioverter-defibrillators, implantable tachyarrhythmia
| Introduction |
|---|
|
|
|---|
Editorial p 2461
Clinical Perspective p 2480
| Methods |
|---|
|
|
|---|
|
Definition of Lead Defects and Follow-Up
Lead defect was defined as a severe lead failure that required surgical correction. A lead failure was diagnosed if one or more of the following criteria were fulfilled: (1) oversensing unrelated to the cardiac cycle was documented; (2) the lead impedance was out of normal range, and a surgical revision was suggested by the experts of the manufacturer; (3) a fracture was observed on x-ray; (4) evidence existed of a lead failure during electrical testing. A sensing failure was assumed if T-wave oversensing was present that could not be overcome by reprogramming of the device or the ventricular sensing signal decreased to <2 mV. A lead insulation defect was assumed if x-ray film showed no evidence of lead discontinuity and if oversensing of irregular unphysiological signals were demonstrated to be incompatible with myopotentials or noise from electrical devices. Abnormal lead impedance was assumed if the measured lead impedance was out of normal range, no visual abnormality was found by x-ray, and the manufacturer suggested a surgical revision. After discharge, all patients visited the defibrillator outpatient clinic every 3 months as well as in case of any adverse event. Evaluation in the ICD clinic consisted of interrogation of the ICD with retrieval of all stored events and intracardiac electrograms, measurement of pacing and sensing thresholds, measurement of lead impedance, recordings of real-time electrograms, and evaluation of nonsustained ventricular tachycardia counters assessment of nonphysiological events. Routine annual ICD testing of DFTs was not performed. The time to follow-up was the period up to the most recent follow-up visit, the time to death, or time to first lead failure. The follow-up ended in September 2005. Death was a censoring mechanism with respect to lead failure.
Statistical Analysis
Absolute numbers, percentages, means and standard deviation, medians, and upper/lower quartile were computed as appropriate. Categorical variables were compared with the
2 or Fisher exact test, as appropriate. For trend analysis the Cochran-Armitage test was used. Cumulative lead survival rates were assessed by Kaplan-Meier analysis. Log-rank tests were used for statistical comparison among groups. Cox regression analysis was performed to find predictors for lead failure. Results were adjusted for age, gender, ejection fraction, number of leads per ICD, and implantation year. All P values were 2-tailed. P<0.05 was considered to be statistically significant. The tests were performed with the SAS statistical package, version 8.02 (Cary, North Carolina).
The authors had full access to and take full responsibility for the integrity of the data. All authors have read and agree to the manuscript as written.
| Results |
|---|
|
|
|---|
|
Lead Survival Outcome and Cause of Lead Failure
The estimated lead survival rates at 5 and 8 years after implantation were 85% and 60%, respectively (Figure 1). The annual failure rate increased with progressive time after implantation and reached 20% in 10-year-old leads (Figure 2). This was confirmed by a trend analysis (P<0.001). The major lead complications were insulation defects (56%), lead fractures (12%), loss of ventricular capture (11%), abnormal lead impedance (10%), and sensing failure (10%). The incidence of insulation failure increased with time after implantation. Insulation failure accounted for 70% of lead failures in leads older than 6 years (Figure 3). Lead defects occurred in newer as well as older models. Kaplan-Meier lead survival curves showed a trend to a better lead survival of the lead models 6936 and 6966 compared with the pooled group of models implanted from 1997 onwards (Figure 4). A better lead survival of 6936 and 6966 leads was particularly observed during the first 5 years after implantation. The estimated survival rate of the models 6936 and 6966 after 5 years was 93% compared with 80% in the pooled model group (P<0.05) (Figure 4). The major cause of lead failure in both groups was insulation defects (56% in 6936/6966 leads versus 46% in the pooled model group; P=0.2). Younger age and female gender were associated with a higher incidence of lead defects, whereas single-chamber ICD and implantation before 1998 tended to be associated with a better lead survival (Figure 5).
|
|
|
|
|
Diagnosis of Lead Defect
Diagnosis of lead defect was made by clinical presentation (inappropriate shocks) in 33% of patients and in 65% by electrical parameters during routine device evaluation. Intraoperative ICD-testing performed during elective generator replacement detected 2% of lead defects. One lead failure was detected after the patient was resuscitated by external defibrillation as a result of ventricular fibrillation. This lead had a fracture; therefore, defibrillation shock could not be delivered. The number of inappropriate shocks in patients ranged from 1 to 17 shocks (median 2, 95% confidence interval 1 to 5 shocks). Inappropriate shocks were caused by artifact sensing in 77%, T-wave oversensing in 18%, and fracture in 5% of leads.
Clinical Outcome
During the median follow-up of 934 days (interquartile range, 368 to 1870), 207 patients (21%) died: 115 patients (55%) died from congestive heart failure, 4 (2%) from sudden death, 18 (9%) from other cardiovascular death, and 27 (13%) from noncardiac causes. In 45 patients (21%), the cause of death remained unknown. Seven patients underwent heart transplantation.
| Discussion |
|---|
|
|
|---|
Incidence and Cause of Lead Failure
Lead-related complications are reported in 1% to 14% of ICD patients.610 This widespread range derives from different data acquisition modalities: retrospective data analysis, short-term follow-up periods, and nonstandardized inclusion criteria. The present study shows that lead defect depends mainly on the follow-up time after the lead was implanted. The estimated lead survival rates at 5 and 8 years are 85% and 60%, respectively. The annual defect rate increases with time and reaches 20% in 10-year-old leads. This is important because many of these leads remain in service. Lead-related problems can occur at any time during long-term follow-up. In addition, in accordance with other studies, a growing number of insulation defects have to be expected with increasing time after implantation.11,12,15,16 In the present study, more than half of the lead defects resulted from insulation defects. Insulation defect is a typical cause of lead failure in coaxial polyurethane-insulated leads like the Medtronic models 6936 and 6966. Polyurethane breakdown, in most cases caused by metal ion oxidation, was shown to be the mode of failure of most coaxial polyurethane pacing leads.1719 The observed failure rate of the Medtronic 6936 and 6966 ICD leads is similar to that reported by others, as well as by the manufacturer.12,13,16,20 Both Hauser et al and Luria et al reported failure rates of close to 20% at 4 years for the 6936 and other similar coaxial polyurethane ICD leads.16,20 Dorwarth et al reported a lead survival rate of 62% at 8-year follow-up.12 The incidence of insulation defects caused by polyurethane breakdown increases during the late follow-up. In the present study, the decline in lead survival of the 6936/6966 models appears to accelerate after 5 years, which might be caused by polyurethane degradation of the 6936/6966 leads caused by metal ion oxidation.
Comparison Between Newer and Older Models
When the polyurethane ICD leads 6936 and 6966 are compared with the newer models implanted from 1997 onwards, the survival curves of the leads 6936 and 6966 show a trend to better lead survival. The newer lead models have a multilumen design: each conductor is individually insulated by silicone rubber, which should prevent the lead from degradation and injury by mechanical stress. The rate of insulation defects observed is less than in the Transvene 6936/6966 leads but still represents the main cause of lead failure. Silicone leads therefore also seem to be prone to insulation failure. Mewis et al reported a high incidence of insulation defects (12%) in silicone lead systems (CPI Endotak C models 0064 and 0072) that required operative revisions during a mean follow-up of 23.8±10.9 months.21 Abrasion of ICD lead insulation material was a frequent finding in silicone leads during generator replacement in a study by de Lurgio et al.22 After an average lead implant duration of 53±18 months, 79% of silicone leads showed abrasions, whereas only 1 of 11 polyurethane leads was affected.
Mechanisms of Lead Failure
Mechanisms of insulation defect that apply to all models should also be considered. These are chronic excessive pressure on the body of the lead from the ligature used for lead fixation12 or mechanical injury such as subclavian crush syndrome.18,23 In the present study, about 95% of the leads were implanted with the subclavian technique. Subclavian puncture is known to have a higher lead complication rate24 and might have led to an increased incidence of insulation failure caused by subclavian crush syndromes. Additional causes for chronic mechanical stress on the lead could be higher activity as in younger patients, restricted anatomic circumstances as in female patients, or multiple leads in the vein.25 Insulation damage caused by mechanical stress is time dependent. Therefore, lead failure becomes more common a problem in younger ICD patients and patients with better preserved left ventricular function who have a longer life expectancy. Insulation damage might also depend on the stability of the lead itself. Newer models have a smaller diameter than the Transvene 6936/6966 lead models (smaller than 10 F versus 12 F). One could speculate that the better lead survival during the first years of the 6936 or 6966 ICD leads is partly a result of the larger diameter.
Diagnosis of Lead Failure
Diagnosis of lead failure was made by clinical presentation with inappropriate shocks in 33% patients. The number of inappropriate shocks in patients ranged from 1 to 17 shocks (median 2, 95% confidence interval 1 to 5 shocks). Because routine ICD follow-up did not routinely include high voltage coil impedance measurements or DFT testing, it is possible that isolated high-voltage coil failures had been missed. Routine DFT testing might have led to an earlier diagnosis of lead failure and might decrease the rate of inappropriate shocks. The main causes of inappropriate shocks were artifact detection or T-wave oversensing. Routine DFT testing might have provoked postshock oversensing, which is reported to be a specific sign of coaxial polyurethane ICD lead failure.11 Routine 3-month ICD evaluation led to the diagnosis of 65% of lead defects. This result emphasizes the need for increased surveillance of patients with older leads. In addition to routine device interrogation, the evaluation should include testing for oversensing during provocative maneuvers7 and measurement of high voltage coil and pace/sense impedances if possible. Only intraoperative ICD testing performed during elective generator replacement detected 2% of lead defects. Preoperatively measured electrical parameters during routine ICD interrogation did not indicate lead failure.
One lead failure was detected after the patient was resuscitated by external defibrillation as a result of ventricular fibrillation. This lead had a fracture, and consecutive defibrillation shocks could not be delivered. If the patient had not been successfully resuscitated, he would have died from sudden death. Sudden death in patients with ICD is not uncommon. In the present study, 4 patients died from sudden death. Pooled data of randomized clinical trials that involve >3500 patients indicate an ICD-unresponsive sudden-death rate of nearly 5%.26 Few data are available on the mode of death in these patients. About one third is attributed to sudden cardiac death, whereas the mode of up to 25% of sudden death remains unclear.27 Future ICD studies should focus on mode of death in ICD patients and lead failure as a mode of death should be considered.
Conclusions
The annual rate of ICD lead defects that require intervention increases with time and reaches 20% in 10-year-old leads. Lead defect is an overall problem encountered in all lead models. Lead defects occur in newer as well as older models. Patients with lead defects are younger and more often female. Multiple lead implantations are associated with a trend to a higher rate of defibrillation lead defects. Nearly two thirds of lead defects can be detected by electrical parameters during routine ICD control, whereas in one third of patients lead defect is recognized only after the occurrence of inappropriate shocks.
These findings raise important issues for patients with expected longer life span at time of ICD implantation, such as younger patients, patients with preserved left ventricular function, or patients who have a prophylactic ICD. The high incidence of lead complications in these patients should be considered at time of implantation in the selection of device or mode of surgical approach for lead implantation. Furthermore, patients with longer follow-up after lead implantation should have increased surveillance. When the increasing rate of lead failure with time is considered, leads should be carefully evaluated at the time of pulse generator replacement. DFT testing, measurement of high-voltage coil and pace/sense impedances, and electrogram recordings should be performed before a decision is made to retain or replace the lead. The data that form the present study do not support routine replacement of intact, normally functioning, transvenous defibrillation leads. However, additional studies are needed to identify predictors of lead failure and safe management strategies to minimize inappropriate shocks and number of additional lead revisions. Consideration of novel approaches to these patients such as "leadless" ICDs deserves further attention.28
Study Limitations
As all implants originated from a single implanting center, our observations and conclusions may not necessarily be generalizable. However, ICD implantation procedures were performed by surgeons who had >10 years of experience with pacemaker and defibrillator implants and performed >95% of all implantations and generator replacements.
About 95% of the leads were implanted with the subclavian technique. As subclavian puncture is known to have a higher lead complication rate, these results may not be extended to leads that are implanted via more desirable access routes such as the cephalic vein.
Because of the long implant duration, lead extraction was not performed routinely. Therefore, the precise cause of lead failure could not be clarified in detail. The reliability of the estimated lead survival rates is decreased because of inconsistent follow-up, loss of patients over time to death from heart failure, and other causes. The number of lead failures has presumably been underestimated.
| Acknowledgments |
|---|
None.
| References |
|---|
|
|
|---|
2. Moss AJ, Hall J, David S, Cannom DS, James P, Dauber JP, Higgins SL, Klein H, Levine JH, Saksena S, Waldo AL, Wilber M, Brown MW, Moonsenong H. Improved survival with an implanted defibrillator in patients with coronary artery disease at high risk for ventricular arrhythmia. N Engl J Med. 1996; 335: 19331940.
3. Moss AJ, Zareba W, Hall WJ, Klein H, Wilber DJ, Cannom DS, Daubert JP, Higgins SL, Brown MW, Andrews ML; Multicenter Automatic Defibrillator Implantation Trial II Investigators. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002; 346: 877883.
4. Buxton AE, Lee KL, DiCarlo L, Gold MR, Greer GS, Prystowsky EN, OToole MF, Tang A, Fisher JD, Coromilas J, Talajic M, Hafley G; for the Multicenter Unsustained Tachycardia Trial Investigators. Electrophysiological testing to identify patients with coronary artery disease who are at risk for sudden death. N Eng J Med. 2000; 342: 19371945.
5. Bardy GH, Lee KL, Mark DB, Poole JE, Packer DL, Boineau R, Domanski M, Troutman C, Anderson J, Johnson G, McNulty SE, Clapp-Channing N, Davidson-Ray LD, Fraulo ES, Fishbein DP, Luceri RM, Ip JH; Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT) Investigators. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med. 2005; 352: 225237.
6. Schwartzman D, Nallamothu N, Callans DJ, Preminger MW, Gottlieb CD, Marchlinski FE. Postoperative lead-related complications in patients with nonthoracotomy defibrillation systems. J Am Coll Cardiol. 1995; 25: 776786.
7. Lawton JS, Ellenbogen KA, Wood MA, Stambler BS, Herre JM, Nath S, Bernstein RC, Di Marco JP, Haines DE, Szentpetery S, Baker LD, Damiano RJ Jr. Sensing lead-related complications in patients with transvenous implantable cardioverter-defibrillators. Am J Cardiol. 1996; 78: 647651.[CrossRef][Medline] [Order article via Infotrieve]
8. Helguera ME, Maloney JD, Pinski SJ, Woscoboinik JR, Wilkoff BL, Castle LW. Long-term performance of endocardial pacing leads. Pacing Clin Electrophysiol. 1994; 17: 5664.[CrossRef][Medline] [Order article via Infotrieve]
9. Mehta D, Nayak HM, Singson M, Chao S, Pe E, Camunas JL, Gomes JA. Late complications in patients with pectoral defibrillator implants with transvenous defibrillator lead systems: high incidence of insulation breakdown. Pacing Clin Electrophysiol. 1998; 21: 18931900.[CrossRef][Medline] [Order article via Infotrieve]
10. Schwacke H, Drögemüller A, Siemon G, Werling C, Saggau W, Senges J, Seidl K. Lead-related complications in 340 patients with an implantable cardioverter/defibrillator. Z Kardiol. 1999; 88: 559565.[CrossRef][Medline] [Order article via Infotrieve]
11. Ellenbogen KA, Wood MA, Shepard RK, Clemo HF, Vaughn T, Holloman K, Dow M, Leffler J, Abeyratne A, Verness D. Detection and management of an implantable cardioverter defibrillator lead failure: incidence and clinical implications. J Am Coll Cardiol. 2003; 41: 7380.
12. Dorwarth U, Frey B, Dugas M, Matis T, Fiek M, Schmoeckel M, Remp T, Durchlaub I, Gerth A, Steinbeck G, Hoffmann E. Transvenous defibrillation leads: high incidence of failure during long-term follow-up. J Cardiovasc Electrophysiol. 2003; 14: 3843.[CrossRef][Medline] [Order article via Infotrieve]
13. Medtronic Inc. Defibrillation leads reference. CRM Product Performance Report 2006. Available at: http://www.medtronic.com/crm/performance/references/defib_lead_reference.html. Accessed August 30, 2006.
14. St. Jude product performance report, April 2006. Available at: http://www.sjm.com/devices/deviceindex.aspx. Accessed August 30, 2006.
15. Kitamura S, Kazuhiro S, Takahashi K, Wataru S, Kazuhiro S, Naohiko A, Kazuo N, Junjiro K, Shiro K. Long-term follow-up of transvenous defibrillation leads: high incidence of fracture in coaxial polyurethane lead. Circ J. 2006; 70: 273277.[CrossRef][Medline] [Order article via Infotrieve]
16. Hauser RG, Cannom D, Hayes DL, Parsonnet V, Hayes J, Ratliff N, Tyers GF, Epstein AE, Vlay SC, Furman S, Gross J. Long-term structural failure of coaxial polyurethane implantable cardioverter defibrillator leads. Pacing Clin Electrophysiol. 2002; 25: 879882.[CrossRef][Medline] [Order article via Infotrieve]
17. Hayes DL, Graham KJ, Irwin M, Vidaillet H, Disler G, Sweesy M, Kincaid D, Osborn MJ, Suman VJ, Neubauer SA. Multicenter experience with a bipolar tined polyurethane ventricular lead. Pacing Clin Electrophysiol. 1995; 18: 9991004.[CrossRef][Medline] [Order article via Infotrieve]
18. Antonelli D, Rosenfeld T, Freeberg NA, Palma E, Gross JN, Furman S. Insulation lead malfunction: is it a matter of insulation coating, venous approach, or both? Pacing Clin Electropyhsiol. 1998; 21: 418421.[CrossRef]
19. Mond HG. Engineering and clinical aspects of pacing leads. In: Ellenbogen KA, Kay GN, Wilkoff BL, eds. Clinical Cardiac Pacing and Defibrillation. Philadelphia, Pa: WB Saunders Co; 2000: 127150.
20. Luria D, Glikson M, Brady PA, Lexvold NY, Rasmussen MJ, Hodge DO, Chugh SS, Rea RF, Hayes DL, Hammill SC, Friedman PA. Predictors and mode of detection of transvenous lead malfunction in implantable defibrillators. Am J Cardiol. 2001; 87: 901904.[CrossRef][Medline] [Order article via Infotrieve]
21. Mewis C, Kühlkamp V, Dörnberger V, Mermi J, Seipel L. High incidence of isolator fractures in transvenous implanted cardioverter/defibrillators. Z Kardiol. 1997; 86: 8594.[Medline] [Order article via Infotrieve]
22. De Lurgio DB, Sathavorn C, Mera F, Leon A, Walter PF, Langberg JJ. Incidence and complications of abrasion of implantable cardioverter-defibrillator leads. Am J Cardiol. 1997; 79: 14091411.[CrossRef][Medline] [Order article via Infotrieve]
23. Roelke M, ONunain SS, Osswald S, Garan H, Harthorne JW, Ruskin JN. Subclavian crush syndrome complication transvenous cardioverter defibrillator systems. Pacing Clin Electrophysiol. 1995; 18: 973979.[CrossRef][Medline] [Order article via Infotrieve]
24. Kron J, Herre J, Renfroe EG, Rizo-Patron C, Raitt M, Halperin B, Gold M, Goldner B, Wathen M, Wilkoff B, Olarte A, Yao Q. Lead- and device-related complications in the Antiarrhythmics Versus Implantable Defibrillators Trial. Am Heart J. 2001; 141: 9298.[CrossRef][Medline] [Order article via Infotrieve]
25. Takahashi T, Bhandari AK, Watanuki M, Cannom DS, Sakurada H, Hiraoka M. High incidence of device-related and lead-related complications in the dual-chamber implantable cardioverter defibrillator compared with the single-chamber version. Circ J. 2002; 66: 746750.[CrossRef][Medline] [Order article via Infotrieve]
26. Anderson KP. Sudden cardiac death unresponsive to implantable defibrillator therapy: an urgent target for clinicians, industry and government. J Interv Card Electrophysiol. 2005; 14: 7178.[CrossRef][Medline] [Order article via Infotrieve]
27. Mitchell LB, Pineda EA, Titus JL, Bartosch PM, Benditt DG. Sudden death in patients with implantable cardioverter defibrillators: the importance of post-shock electromechanical dissociation. J Am Coll Cardiol. 2002; 39: 13231328.
28. Burke MC, Coman JA, Cates AW, Lindstrom CC, Sandler DA, Kim SS, Knight BP. Defibrillation energy requirements using a left anterior chest cutaneous to subcutaneous shocking vector: implications for a total subcutaneous implantable defibrillator. Heart Rhythm. 2005; 2: 13321338.[CrossRef][Medline] [Order article via Infotrieve]
![]() |
M. Brignole Are complications of implantable defibrillators under-estimated and benefits over-estimated? Europace, July 4, 2009; (2009) eup174v1. [Full Text] [PDF] |
||||
![]() |
J. G. Gaca, B. Lima, C. A. Milano, S. S. Lin, R. D. Davis, J. E. Lowe, and P. K. Smith Laser-assisted extraction of pacemaker and defibrillator leads: the role of the cardiac surgeon. Ann. Thorac. Surg., May 1, 2009; 87(5): 1446 - 1450. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Grace Implantable defibrillators in hypertrophic cardiomyopathy Heart, May 1, 2009; 95(9): 695 - 696. [Full Text] [PDF] |
||||
![]() |
S. Spencker, N. Coban, L. Koch, A. Schirdewan, and D. Muller Potential role of home monitoring to reduce inappropriate shocks in implantable cardioverter-defibrillator patients due to lead failure Europace, April 1, 2009; 11(4): 483 - 488. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. E. Marine Remote monitoring for prevention of inappropriate implantable cardioverter defibrillator shocks: is there no place like home? Europace, April 1, 2009; 11(4): 409 - 411. [Full Text] [PDF] |
||||
![]() |
G. Z. Duray, J. Schmitt, S. Cicek-Hartvig, S. H. Hohnloser, and C. W. Israel Complications leading to surgical revision in implantable cardioverter defibrillator patients: comparison of patients with single-chamber, dual-chamber, and biventricular devices Europace, March 1, 2009; 11(3): 297 - 302. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Corrado, G. Gasparini, and A. Raviele Lead malfunctions in implantable cardioverter defibrillators: where are we and where should we go? Europace, March 1, 2009; 11(3): 276 - 277. [Full Text] [PDF] |
||||
![]() |
T. Kleemann and K. Seidl Letter by Kleemann and Seidl Regarding Article, "Necessity for Surgical Revision of Defibrillator Leads Implanted Long-Term: Causes and Management" Circulation, January 6, 2009; 119(1): e8 - e8. [Full Text] [PDF] |
||||
![]() |
C. Sticherling, J. Eckstein, B. A. Schaer, S. Osswald, M. T. Koller, D. Kalusche, and M. Zabel Response to Letter Regarding Article, "Necessity for Surgical Revision of Chronically Implanted Defibrillator Leads: Causes and Management" Circulation, January 6, 2009; 119(1): e9 - e9. [Full Text] [PDF] |
||||
![]() |
G. Boriani, M. Biffi, A. Marziali, I. Diemberger, and C. Martignani A changing scenario in the clinical use of implantable defibrillators: the need for long-term data on lead performance Europace, January 1, 2009; 11(1): 1 - 3. [Full Text] [PDF] |
||||
![]() |
A. Goette, F. Cantu, L. van Erven, P. Geelen, F. Halimi, J. L. Merino, J. M. Morgan, and on behalf of the Scientific Initiative Committee o Performance and survival of transvenous defibrillation leads: need for a European data registry Europace, January 1, 2009; 11(1): 31 - 34. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. G. Hauser A Better Method for Preventing Adverse Clinical Events Caused by Implantable Cardioverter-Defibrillator Lead Fractures? Circulation, November 18, 2008; 118(21): 2117 - 2119. [Full Text] [PDF] |
||||
![]() |
C. D. Swerdlow, B. D. Gunderson, K. T. Ousdigian, A. Abeyratne, R. W. Stadler, J. M. Gillberg, A. S. Patel, and K. A. Ellenbogen Downloadable Algorithm to Reduce Inappropriate Shocks Caused by Fractures of Implantable Cardioverter-Defibrillator Leads Circulation, November 18, 2008; 118(21): 2122 - 2129. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. C. Saumarez, M. Pytkowski, M. Sterlinski, J. P. Bourke, J. R. Clague, S. M. Cobbe, D. T. Connelly, M. J. Griffith, P. P. McKeown, K. McLeod, et al. Paced ventricular electrogram fractionation predicts sudden cardiac death in hypertrophic cardiomyopathy Eur. Heart J., July 1, 2008; 29(13): 1653 - 1661. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Chinushi, Y. Hosaka, N. Ikarashi, K. Iijima, H. Furushima, and Y. Aizawa Automatic R-wave and impedance testing with the modern patient alert system to reduce inappropriate implantable cardioverter defibrillator shocks due to lead fracture Europace, June 1, 2008; 10(6): 738 - 740. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. M. Scheinman and E. Keung The Year in Review of Clinical Cardiac Electrophysiology J. Am. Coll. Cardiol., May 27, 2008; 51(21): 2075 - 2081. [Full Text] [PDF] |
||||
![]() |
W. H. Maisel and D. B. Kramer Implantable Cardioverter-Defibrillator Lead Performance Circulation, May 27, 2008; 117(21): 2721 - 2723. [Full Text] [PDF] |
||||
![]() |
J. Eckstein, M. T. Koller, M. Zabel, D. Kalusche, B. A. Schaer, S. Osswald, and C. Sticherling Necessity for Surgical Revision of Defibrillator Leads Implanted Long-Term: Causes and Management Circulation, May 27, 2008; 117(21): 2727 - 2733. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. B. Johansen, S. S. Pedersen, H. Spindler, K. Andersen, J. C. Nielsen, and P. T. Mortensen Symptomatic heart failure is the most important clinical correlate of impaired quality of life, anxiety, and depression in implantable cardioverter-defibrillator patients: a single-centre, cross-sectional study in 610 patients Europace, May 1, 2008; 10(5): 545 - 551. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Sarkozy, P. Brugada, L. Mont, and J. Brugada Optimizing the clinical use of implantable defibrillators in patients with Brugada syndrome Eur. Heart J. Suppl., December 1, 2007; 9(suppl_I): I74 - I80. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Suri and S. Keller Letter by Suri and Keller Regarding Article, "Annual Rate of Transvenous Defibrillation Lead Defects in Cardioverter-Defibrillators Over a Period of >10 Years" Circulation, October 30, 2007; 116(18): e512 - e512. [Full Text] [PDF] |
||||
![]() |
T. Kleemann, T. Becker, K. Doenges, M. Vater, J. Senges, S. Schneider, W. Saggau, U. Weisse, and K. Seidl Response to Letter Regarding Article, "Annual Rate of Transvenous Defibrillation Lead Defects in Cardioverter-Defibrillators Over a Period of >10 Years" Circulation, October 30, 2007; 116(18): e513 - e513. [Full Text] [PDF] |
||||
![]() |
J. A. Ezekowitz, B. H. Rowe, D. M. Dryden, N. Hooton, B. Vandermeer, C. Spooner, and F. A. McAlister Systematic Review: Implantable Cardioverter Defibrillators for Adults with Left Ventricular Systolic Dysfunction Ann Intern Med, August 21, 2007; 147(4): 251 - 262. [Abstract] [Full Text] [PDF] |
||||
![]() |
Failure Rate of ICD Leads Are Greater Than Expected Journal Watch Cardiology, May 30, 2007; 2007(530): 2 - 2. [Full Text] |
||||
![]() |
W. H. Maisel Transvenous Implantable Cardioverter-Defibrillator Leads: The Weakest Link Circulation, May 15, 2007; 115(19): 2461 - 2463. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2007 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |