(Circulation. 2007;115:1392-1397.)
© 2007 American Heart Association, Inc.
Interventional Cardiology |
From the University College London Institute of Child Health and Great Ormond Street Hospital for Children (J.N., S.K., L.C., S.S., P.L. G.P., A.M.T., P.B.), London, United Kingdom, and Department of Cardiology (J.E.L.), Childrens Hospital, Boston, Mass.
Correspondence to Dr Philipp Bonhoeffer, Cardiothoracic Unit, Great Ormond Street Hospital for Children, Great Ormond Street, London WC1N 3JH, United Kingdom. E-mail bonhop{at}gosh.nhs.uk
Received November 3, 2006; accepted January 5, 2007.
| Abstract |
|---|
|
|
|---|
Methods and Results After PPVI, 123 patients had chest x-ray in anteroposterior and lateral projection, echocardiography, and clinical evaluation during structured follow-up. Of these 123 patients, 26 (21.1%) developed stent fracture 0 to 843 days after PPVI (stent fracturefree survival at 1 year, 85.1%; at 2 years, 74.5%; and at 3 years, 69.2%). Stent fracture was classified as type I: no loss of stent integrity (n=17); type II: loss of integrity with restenosis on echocardiography (n=8); and type III: separation of fragments or embolization (n=1). In a multivariate Cox regression, we analyzed various factors, of which 3 were associated with a higher risk of stent fracture: implantation into "native" right ventricular outflow tract (P=0.04), no calcification along the right ventricular outflow tract (judged with fluoroscopy, P=0.02), recoil of PPV (qualitatively, PPV diameter in frontal or lateral plane with fully inflated balloon > diameter after balloon deflation, P=0.03). Substernal PPV location, high-pressure post-PPVI dilatation of PPV, pre-PPVI right ventricular outflow tract gradients, and other indicators of PPV compression or asymmetry did not pose increased risk. Patients with type I fracture remain under follow-up. Patients with type II fracture had 2nd PPVI or are awaiting such procedure, and 1 patient with type III fracture required surgical explantation.
Conclusion Stent fracture after PPVI can be managed effectively by risk stratification, systematic classification, and anticipatory management strategies. Serial x-ray and echocardiography are recommended for surveillance.
Key Words: catheterization echocardiography heart defects, congenital pulmonary valve
| Introduction |
|---|
|
|
|---|
Clinical Perspective p 1397
In September 2000, we performed the first PPVI,2,3 and during the study period 123 patients have been treated with this approach. The valved stents were implanted in patients with RVOT obstruction, with or without pulmonary regurgitation, that occurred as a late consequence of previous surgery.4,5 In the vast majority, the valves were implanted into dysfunctional conduits that connected the right ventricle to the pulmonary artery.4,5 Before this, stenotic conduits have been treated by bare stent implantation by our team and others,68 whereas regurgitation often required reoperation. Complications of stenting, such as stent fracture,610 are well known with a reported incidence of up to 43%.8
With the introduction of the novel treatment of PPVI, which has its basis in endovascular stenting, we set out to establish risk factors for stent fracture in our population and to devise management strategies.
| Methods |
|---|
|
|
|---|
Classification
Stent fracture was classified on the basis of chest x-ray appearance. Type I indicates fracture of
1 strut without loss of stent integrity. Type II indicates fracture with loss of stent integrity. Type III indicates fracture associated with separation of fragments or embolization (Figure 1).
|
Risk Stratification
Angiograms of the PPVI procedure or chest x-rays performed after the procedure were reviewed. Independent variables analyzed for association with risk of stent fracture were morphological characteristics of RVOT (type a: "native," any part of circumference formed by native tissue; type b: homograft or other valved or nonvalved conduit), presence of preexisting stent in RVOT, substernal location of the PPV, post-PPVI dilatation procedure with high-pressure balloon to relieve residual stenosis (16 to 22 mm, inflated to 8 to 12 atm; Mullins, NuMed), recoil of PPV (qualitatively, PPV diameter in frontal or lateral plane with fully inflated balloon > diameter after balloon deflation), RVOT calcification (circumferential, partial, or no calcifications along the RVOT, judged with fluoroscopy), and compression/asymmetry of PPV (assessed as ratio of anteroposterior and lateral PPV diameters at proximal, mid, and distal site). The PPV dimensions were measured (Medcon TCS Symphony image management) after calibration with the markers on Multitrack angiographic catheters (NuMed). RVOT gradients pre-PPVI (group I: RVOT gradient
30 mm Hg; group II: RVOT gradient <30 mm Hg) and changes in RVOT gradients post-PPVI were analyzed to assess the influence of hemodynamic values on the probability of stent fractures.
Statistics
Data are expressed as means±SEM unless otherwise specified. Last follow-up was defined as last clinical contact or explantation. Factors for risk stratification were analyzed with multivariate Cox regression. Survival curves were obtained by Kaplan-Meier plots, and log-rank tests were performed to compare the probabilities for stent fracturefree survival. Student t test was used to evaluate differences in the initial decrease of RVOT gradients between patients with and without stent fractures. The post-PPVI echocardiogram (before stent fracture) and the first echocardiogram after detection of stent fractures were compared with Student paired t test. P<0.05 was considered significant. Statistical analysis was performed on SPSS 13.0 (SPSS Inc, Chicago, Ill) and GraphPad InStat 3 (Graphpad Software, San Diego, Calif).
The authors had full access to and take full responsibility for the integrity of the data. All authors have read and agree to the manuscript as written.
| Results |
|---|
|
|
|---|
|
Incidence of Stent Fractures in Patients With PPV
Stent fracture was detected in 26 of 123 patients (21.1%) with the highest incidence in the first 400 days (21 of 26). The probability of stent fracture freesurvival was 85.1% at 1 year, 74.5% at 2 years, and 69.2% at 3 years (Figure 2A).
|
At initial presentation, 20 of 26 patients had type I, 5 of 26 had type II, and 1 of 26 had type III stent fracture. Of these 26 patients, 6 experienced clinical consequences. The patient with type III stent fracture had stent embolization into the right pulmonary artery, which required surgical explantation of the PPV. The diagnosis was made 2 days after the onset of symptoms and confirmed by x-ray. All patients who initially presented with type II stent fracture had second PPVI for relief of restenosis (movies I and II, online-only data supplement). Of the 20 patients with type I fracture, 85% (17 of 20) are under medical follow-up with serial chest x-ray and echocardiography for surveillance. In 3 of 20 patients, progression to type II fracture occurred; 1 patient had second PPVI and 2 patients are awaiting second PPVI. Overall probability of freedom from type II/III stent fracture was 96.3% at 1 year, 92.5% at 2 years, and 86.5% at 3 years (Figure 2B).
Hemodynamic Analysis
RVOT velocities and tricuspid regurgitation jet velocities measured by echocardiography showed no difference before and after stent fracture (RVOT, 3.0±0.2 versus 3.2±0.2 m/s, P=0.06; tricuspid regurgitation jet velocity, 3.3±0.2 versus 3.6±0.2 m/s, P=0.07) (Figure 3A and 3B), although individual patients showed increasing velocities (Figure 3A and 3B).
|
Risk Factors for Stent Fracture
In 97 of 123 patients, the original angiograms of PPVI procedure or chest x-rays performed after the procedure were available for review. In a multivariate Cox regression analysis (Table 2), 3 factors were associated with higher risk of stent fracture: PPVI into "native" RVOT (B=1.8, P=0.04), noncalcified RVOT (B=0.7, P=0.02; log-rank test, P<0.001) (Figure 4A), and recoil of PPV during balloon deflation (B=1.4, P=0.03). Presence of preexisting stent in RVOT (B=0.3, P=0.69), substernal PPV location (B=0.6, P=0.35), post-PPVI dilatation procedure with high-pressure balloon (B=0.1, P=0.86; log-rank test, P=0.97) (Figure 4B), other indicators for PPV compression or asymmetry [anteroposterior-to-lateral PPV diameter ratios at proximal site (B=3.7, P=0.07), mid-site (B=0.01, P=1.00), and distal site (B=2.2, P=0.41)], and significant RVOT gradient pre-PPVI (B=0.02, P=0.97) did not pose increased risk. Changes in RVOT gradients immediately after PPVI, measured by catheterization, were not significantly different in patients with and without stent fractures (16.9±2.1 versus 18.9±4.5 mm Hg, P=0.63).
|
|
Freedom From Reintervention
To distinguish the probability of freedom from reintervention, we excluded 7 patients with stent fractures from the analysis because it was impossible to establish a causal relationship between stent fracture and the need for reintervention, or stent fracture was even secondary to reintervention for other causes (Table 3). Patients with stent fractures had significantly shorter freedom from catheter reintervention (51.0% versus 94.2% at 3 years, P=0.002), but not from explantation (93.8% versus 81.8% at 3 years, P=0.32).
|
Detection Accuracy
Good agreement existed in detection of stent fracture in a small sample of 13 patients, whose follow-up chest x-rays with follow-up fluoroscopy (10 patients positive on x-ray and positive on fluoroscopy; 3 patients negative on x-ray and negative on fluoroscopy) were compared. However, the degree of displacement of stent components can be better judged with fluoroscopy compared with chest x-ray (Figure 5), unless a clear distortion exists.
|
| Discussion |
|---|
|
|
|---|
Peng et al reported stent fracture in 56 of 126 patients (43%) who underwent cardiac catheterization before conduit replacement surgery.8 They describe the potential for underestimation of stent fracture, as minor grades may be clinically silent, difficult to detect by plain radiography, or not detected at surgery if the conduit excision was en masse without dissection into it. Their cohort had 2 different types of balloon-expandable stainless steel stents: the Palmaz-Schatz and the Palmaz-Schatz Genesis.8 Nearly 90% of fractured stents had substernal location, which was identified as an important risk factor for fracture.8
The stent used for PPVI was specifically designed for applications in congenital heart disease, particularly for use over a large range of diameters. We chose the platinum-iridium stent because of its crimpability, limited shortening during deployment, and the belief that it was the most fracture-resistant stent available. Other stents routinely used were designed for peripheral vascular applications, and shortcomings have included sharp edges that lead to balloon rupture, excessive shortening, and fractures when implanted into the RVOT.8,12
For PPV, the detection accuracy of chest x-ray surveillance compared well to fluoroscopy in a small sample of patients because the PPV stent material (platinum-iridium) is very radio-opaque. However, displacement of stent components can be missed with plain radiography. Fluoroscopy has therefore an added benefit in facilitating the grading of stent fractures.
Of the 123 patients, 21.1% (26 of 123) developed stent fracture during follow-up, which represents the most common early complication after PPVI. We identified 3 variables that pose high risk: implantation of PPV into native RVOTs, noncalcified RVOTs, and presence of PPV recoil during balloon deflation. All these factors might implicate higher cyclic compressive stresses on the PPV. In our analysis, substernal stent location or asymmetry of the PPV did not pose higher risk on multivariate analysis. This is in contrast to findings in the bare metal stent population, where both mechanisms contributed to stent fractures.8 Our analysis might be underpowered, because the more robust multivariate analysis might be too rigorous to detect statistical significance in a relatively small population. Also, we believe that not all substernal calcified conduits may be subject to cyclic compressive stress. Some conduits that are circumferentially calcified and adherent to sternum lack mobility and deformation during the cardiac cycle. This situation is often detected during angiography, where, although the cardiac chambers move during the cardiac cycle, the conduit often remains rigidly immobile behind the sternum. This immobility may, however, contribute to the risk of residual stenosis caused by failure to expand the stent to its intended dimension.
Importantly, higher grades of stent fracture are associated with restenosis, which can also be seen in other modes of early PPVI failure.4 This restenosis leads to significantly shorter freedom from catheter reintervention, which underscores the need for continued surveillance. If stent fracture is diagnosed at an early stage and reintervention is warranted, repeat PPVI prolongs the lifespan of the conduit and provides a reasonable alternative to surgery, thus reducing surgical explantation. A similar strategy can be found in other studies in which patients with stent fracture received additional stents to restore stent patency and enhance integrity.8,9,11 In our experience, successful treatment of clinical consequences of stent fracture and freedom from reoperation justify continuous surveillance.
Limitations
Although all chest x-rays from our institution were reviewed by members of our team (J.N., S.K., L.C., P.L.) and underwent quality control, a relatively large proportion of the study population (approximately 40%) was followed-up by other specialized centers as per the recommended protocol. For these patients, image resolution and less diligent interpretation could have affected the quality of reporting. Three patients were lost to follow-up. Fluoroscopy was not performed in all patients because it was not part of the ethically approved study protocol, but fluoroscopy may influence the classification of stent fractures. The assessment of stent recoil was subjective and may reflect bias. Long-term follow-up will provide further data on late incidence of stent fractures.
Conclusion
Stent fractures occur after PPVI, but they can be managed effectively by risk stratification, systematic classification, and anticipatory management strategies. Serial x-ray and echocardiography are recommended for surveillance. Detection of stent fractures at an early stage could avert serious sequelae and prolong conduit lifespan. As more stent-based applications are introduced to clinical practice, stent fracture is likely to become a frequent clinical event that requires surveillance and early intervention.
| Acknowledgments |
|---|
This study was supported by the British Heart Foundation (RG/03/006).
Disclosures
Dr Khambadkone, Dr Coates, and Dr Bonhoeffer have received honoraria from Medtronic Inc. Dr Khambadkone, Dr Taylor, and Dr Bonhoeffer are consultants/advisory board members of Medtronic Inc. Dr Coates has received a British Heart Foundation Junior Fellowship Grant. Dr Bonhoeffer has an ownership interest in the Melody Valve (Medtronic Inc) and has received a British Heart Foundation Program Grant. The remaining authors report no conflicts.
| References |
|---|
|
|
|---|
2. Bonhoeffer P, Boudjemline Y, Saliba Z, Merckx J, Aggoun Y, Bonnet D, Acar P, Le Bidois J, Sidi D, Kachaner J. Percutaneous replacement of pulmonary valve in a right-ventricle to pulmonary-artery prosthetic conduit with valve dysfunction. Lancet. 2000; 356: 14031405.[CrossRef][Medline] [Order article via Infotrieve]
3. Bonhoeffer P, Boudjemline Y, Saliba Z, Hausse AO, Aggoun Y, Bonnet D, Sidi D, Kachaner J. Transcatheter implantation of a bovine valve in pulmonary position: a lamb study. Circulation. 2000; 102: 813816.
4. Khambadkone S, Coats L, Taylor A, Boudjemline Y, Derrick G, Tsang V, Cooper J, Muthurangu V, Hegde SR, Razavi RS, Pellerin D, Deanfield J, Bonhoeffer P. Percutaneous pulmonary valve implantation in humans: results in 59 consecutive patients. Circulation. 2005; 112: 11891197.
5. Coats L, Tsang V, Khambadkone S, van Doorn C, Cullen S, Deanfield J, de Leval MR, Bonhoeffer P. The potential impact of percutaneous pulmonary valve stent implantation on right ventricular outflow tract re-intervention. Eur J Cardiothorac Surg. 2005; 27: 536542.
6. Murugan SJ, Dickinson DF, Gibbs JL. Images in cardiology. Complete fracture of a right ventricle to pulmonary artery shunt. Heart. 2005; 91: 1441.
7. Ovaert C, Caldarone CA, McCrindle BW, Nykanen D, Freedom RM, Coles JG, Williams WG, Benson LN. Endovascular stent implantation for the management of postoperative right ventricular outflow tract obstruction: clinical efficacy. J Thorac Cardiovasc Surg. 1999; 118: 886893.
8. Peng LF, McElhinney DB, Nugent AW, Powell AJ, Marshall AC, Bacha EA, Lock JE. Endovascular stenting of obstructed right ventricle-to-pulmonary artery conduits: a 15-year experience. Circulation. 2006; 113: 25982605.
9. Knirsch W, Haas NA, Lewin MAG, Uhlemann F. Longitudinal stent fracture 11 months after implantation in the left pulmonary artery and successful management by a stent-in-stent maneuver. Catheter Cardiovasc Interv. 2003; 58: 116118.[CrossRef][Medline] [Order article via Infotrieve]
10. Olaughlin MP, Slack MC, Grifka RG, Perry SB, Lock JE, Mullins CE. Implantation and intermediate-term follow-up of stents in congenital heart-disease. Circulation. 1993; 88: 605614.
11. Scheinert D, Scheinert S, Sax J, Piokowski C, Braunlich S, Ulrich M, Biamino G, Schmidt A. Prevalence and clinical impact of stent fractures after femoropopliteal stenting. J Am Coll Cardiol. 2005; 45: 312315.
12. Cheatham JP. Stenting of coarctation of the aorta. Catheter Cardiovasc Interv. 2001; 54: 112125.[CrossRef][Medline] [Order article via Infotrieve]
| Footnotes |
|---|
This article has been cited by other articles:
![]() |
E. M. Zahn, W. E. Hellenbrand, J. E. Lock, and D. B. McElhinney Implantation of the melody transcatheter pulmonary valve in patients with a dysfunctional right ventricular outflow tract conduit early results from the u.s. Clinical trial. J. Am. Coll. Cardiol., October 27, 2009; 54(18): 1722 - 1729. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Nordmeyer, P. Lurz, V. T. Tsang, L. Coats, F. Walker, A. M. Taylor, S. Khambadkone, M. R. de Leval, and P. Bonhoeffer Effective transcatheter valve implantation after pulmonary homograft failure: A new perspective on the Ross operation J. Thorac. Cardiovasc. Surg., July 1, 2009; 138(1): 84 - 88. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Lurz, J. Nordmeyer, V. Muthurangu, S. Khambadkone, G. Derrick, R. Yates, M. Sury, P. Bonhoeffer, and A. M. Taylor Comparison of Bare Metal Stenting and Percutaneous Pulmonary Valve Implantation for Treatment of Right Ventricular Outflow Tract Obstruction: Use of an X-Ray/Magnetic Resonance Hybrid Laboratory for Acute Physiological Assessment Circulation, June 16, 2009; 119(23): 2995 - 3001. [Abstract] [Full Text] [PDF] |
||||
![]() |
P Lurz, J Nordmeyer, L Coats, A M Taylor, P Bonhoeffer, and I Schulze-Neick Immediate clinical and haemodynamic benefits of restoration of pulmonary valvar competence in patients with pulmonary hypertension Heart, April 1, 2009; 95(8): 646 - 650. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. T.L. Chiam and C. E. Ruiz Percutaneous Transcatheter Aortic Valve Implantation: Assessing Results, Judging Outcomes, and Planning Trials: The Interventionalist Perspective J. Am. Coll. Cardiol. Intv., August 1, 2008; 1(4): 341 - 350. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Lurz, L. Coats, S. Khambadkone, J. Nordmeyer, Y. Boudjemline, S. Schievano, V. Muthurangu, T. Y. Lee, G. Parenzan, G. Derrick, et al. Percutaneous Pulmonary Valve Implantation: Impact of Evolving Technology and Learning Curve on Clinical Outcome Circulation, April 15, 2008; 117(15): 1964 - 1972. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Nordmeyer, L. Coats, P. Lurz, T.-Y. Lee, G. Derrick, P. Rees, S. Cullen, A. M. Taylor, S. Khambadkone, and P. Bonhoeffer Percutaneous pulmonary valve-in-valve implantation: a successful treatment concept for early device failure Eur. Heart J., March 3, 2008; (2008) ehn073v1. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2007 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |