(Circulation. 2006;114:591-596.)
© 2006 American Heart Association, Inc.
Valvular Heart Disease |
From the Divisions of Cardiology and Cardiac Surgery, St Pauls Hospital, University of British Columbia, Vancouver, Canada.
Correspondence to John G. Webb, MD, Director, Interventional Cardiology, St. Pauls Hospital, Room 5202A, 1081 Burrard St, Vancouver, BC, Canada V6Z 2E8. E-mail webb{at}providencehealth.bc.ca
Received April 14, 2006; revision received May 22, 2006; accepted May 30, 2006.
| Abstract |
|---|
|
|
|---|
Methods and Results A left anterolateral intercostal incision is used to expose the left ventricular apex. Direct needle puncture of the apex allows introduction of a hemostatic sheath into the left ventricle. The valve prosthesis, constructed from a stainless steel stent with an attached trileaflet equine pericardial valve, is crimped onto a valvuloplasty balloon. The prosthetic valve and balloon catheter are passed over a wire into the left ventricle. Positioning within the aortic annulus is confirmed by fluoroscopy, aortography, and echocardiography. Rapid ventricular pacing is used to reduce cardiac output while the balloon is inflated, deploying the prosthesis within the annulus. Transapical aortic valve implantation was successfully performed in 7 patients in whom surgical risk was deemed excessive because of comorbidities. Echocardiographic median aortic valve area increased from 0.7±0.1 cm2 (interquartile range) to 1.8±0.8 cm2 (interquartile range). There were no intraprocedural deaths. At a follow up of 87±56 days, 6 of 7 patients remain alive and well.
Conclusions This initial experience suggests that transapical aortic valve implantation without cardiopulmonary bypass is feasible in selected patients with aortic stenosis.
Key Words: aorta catheters stenosis valves valvuloplasty
| Introduction |
|---|
|
|
|---|
Editorial p 533
Clinical Perspective p 596
Several groups have pursued the development of transcatheter valves.713 Percutaneous aortic valve implantation in humans was first performed as a femoral transvenous procedure with antegrade access to the aortic valve.1416 Subsequently, we reported a femoral transarterial procedure.17,18 Our groups initial animal developmental work used direct balloon catheter implantation from the left ventricular apex.12 Recently, we reported a case in which the transarterial delivery system was used to implant an aortic valve via the apex of the left ventricle.19,20 The present study describes our clinical experience and technique of prosthetic aortic valve implantation by direct left ventricular access using a beating-heart, catheter-based approach in humans.
| Methods |
|---|
|
|
|---|
Prosthetic Valve System
The Cribier-Edwards valve (Edwards Lifesciences Inc, Irvine, Calif) is constructed from a tubular slotted stainless steel stent with an attached equine pericardial trileaflet valve (Figure 1). A sewn fabric cuff covers the left ventricular portion of the prosthesis. In vitro durability of the prosthetic valve has repeatedly been demonstrated to be >200 million cardiac cycles, corresponding to >5 years of life. Valves are supplied sterile in glutaraldehyde. At the time of this experience, only a 26-mm-external-diameter prosthesis was available for transapical use. On the basis of our percutaneous implantation experience, the 26-mm prosthesis was considered appropriate for an echocardiographic annulus diameter of 20 to 25 mm. Valve sizing in this situation differs from that during traditional open surgery because variable expansion of the native calcified valve will accommodate a single size quite well (ie, 26 mm) despite a large range of annular sizes.
|
A mechanical crimping device is used to attach the prosthesis onto a specially constructed balloon deployment catheter. For the transapical approach, the occlusive fabric skirt must be mounted proximally on the balloon catheter. In these initial apical access cases, we used the identical valve and catheter delivery system designed for percutaneous femoral arterial access.17 In this system, the balloon delivery catheter shaft is contained within a steerable guiding catheter that can be actively flexed by rotation of an external handle.
Procedure
Patients were premedicated with aspirin, clopidogrel, and vancomycin. Procedures were performed by a combined team of cardiologists and cardiac surgeons in an operating room setting equipped with fluoroscopy. General anesthesia with double-lumen intubation was used. The apex of the left ventricle was identified by palpation and fluoroscopy. The pleural space overlying the apex (usually the sixth intercostal space) was entered via a 5- to 8-cm anterolateral thoracotomy. The pericardium over the apex of the left ventricle was identified and opened. Temporary epicardial ventricular pacing wires were placed on the left ventricle. Reliable capture was ensured at a rate of between 160 and 220 bpm with a reduction in systemic systolic arterial pressure to <60 mm Hg.17 During subsequent valvuloplasty and prosthesis deployment, rapid pacing was used to minimize transaortic flow and cardiac motion (Figure 2).
|
The thin portion of the left ventricular apex was identified by finger palpation and confirmed by simultaneous transesophageal echocardiographic imaging (TEE). Two paired orthogonal U-shaped sutures with pledgets were placed into the myocardium and passed through tensioning tourniquets. An arterial needle puncture allowed placement of a 7F sheath through the apex into the left ventricular cavity using a standard over-the-wire technique (Figure 3). Positive ventricular pressure and care to bleed all sheaths through appropriate ports prevented air entry into the ventricle. Heparin was administered to achieve an activated clotting time of >250 seconds. A wire (0.035-in J-curve Amplatz extra-stiff, Cook, Inc, Bloomington, Ind) was advanced through the aortic valve, around the arch, and into the descending aorta for stability. If necessary, a balloon catheter (Berman catheter, Arrow International Inc, Reading, Pa) was used to direct the wire in the direction of arterial flow past mitral chordae and around the aortic arch. The previously placed sheath was then exchanged for a larger 14F sheath. A valvuloplasty balloon (20 to 22 mm, Z-Med, Numed Inc, Hopkington, NY) was advanced over the wire to the level of the ascending aorta, evacuated of air, and test inflated with dilute contrast. The balloon was withdrawn to the level of the valve as determined by fluoroscopy and TEE. Rapid pacing was initiated, and when an adequate reduction in systemic arterial pressure was observed, the balloon was rapidly inflated and deflated.
|
The previously placed 14F sheath was exchanged over the wire for a larger 24F sheath. The prosthesis, balloon, and steerable catheter were passed over the previously placed wire and introduced as a unit through the sheath. The prosthetic valve was advanced to the level of fluoroscopically visible valvular calcium. Aortic root angiography via a pigtail catheter introduced from the femoral artery and TEE were used to aid positioning. Rapid pacing was again used to reduce transvalvular flow and cardiac motion during valve deployment. After deployment, valve function was assessed by echocardiography and angiography. If paravalvular insufficiency was excessive, dilatation of the prosthesis was repeated. After removal of the left ventricular sheath, hemostasis was secured with the previously placed pledgeted sutures. The pericardium was approximated to prevent myocardial herniation but allow drainage. A left chest tube was placed. Patients were maintained on aspirin indefinitely and clopidogrel for 1 month. Transthoracic echocardiography was performed before discharge and at 1 month.
The authors had full access to the data and take responsibility for its integrity. All authors have read and agree to the manuscript as written.
| Results |
|---|
|
|
|---|
|
Procedure
Prosthetic valve implantation was successful in all cases (Figure 4). Procedural characteristics are shown in Table 2. Fluoroscopic assessment of valvular calcium was used for initial positioning of the prosthesis. However, the amount and location of calcification were variable, requiring further assessment with repeated injections of radiographic contrast within the aortic root and online TEE (Figure 5). Integration of all 3 modalities by the implanting physician facilitated optimal positioning.
|
|
|
Rapid ventricular pacing was an effective and reproducible means of reducing transvalvular flow during balloon inflation (Figure 2). During inflation of the deployment balloon, movement either was not fluoroscopically apparent or was limited to a few millimeters. Movement was always in the direction of flow, toward the aorta. Thus, the prosthesis was generally positioned slightly toward the ventricle before deployment so that if slight movement did occur, final positioning would be adequate.
All patients received 26-mm-diameter prostheses. Final positioning was judged optimal in 5 patients. Positioning was slightly high within the annulus but adequate in 2 patients. Some degree of paravalvular insufficiency after initial deployment was evident in all patients. Paravalvular insufficiency was judged excessive by TEE in 2 patients in whom redilation resulted in further expansion of the prosthesis and a satisfactory reduction in paravalvular regurgitation. At the completion of the procedure, aortic insufficiency grade was 1.0±0.8 by TEE.
Follow-Up
Postprocedural transthoracic echocardiography reported a median transaortic gradient of 9±6 mm Hg (interquartile range) and an aortic valve area of 1.8±0.8 cm2 (interquartile range). Echocardiography documented no change in valve function at 1 month (Table 3). In no patient did paravalvular insufficiency appear clinically significant. One patient died at day 12 of pneumonia despite echocardiographically normal prosthetic valve function. Thirty-day mortality rate was 14%. At a mean follow up of 87±56 days, 6 of 7 patients remained alive and well.
|
| Discussion |
|---|
|
|
|---|
Unfortunately, many potential surgical candidates have significant comorbidities, and open heart surgery with cardiopulmonary bypass may pose risks that are unacceptable to them or their physicians. According to the Society of Thoracic Surgery Database (1998 to 2001), surgical aortic valve replacement carries a rate of serious complication or mortality of 16.8%. Operative risk is increased in the setting of comorbidities such as advanced age and others.2327 In our patients, judged poor candidates for routine surgery with an estimated logistic Euroscore mortality of 35±26%, a 30-day mortality rate of 14% is encouraging.
The feasibility of percutaneous valve implantation was first demonstrated in an animal model by Anderson et al.7 Subsequently, several groups, including our own, pursued the development of percutaneous heart valves.713 Percutaneous aortic valve implantation in humans was first performed as a transvenous transseptal procedure with antegrade access to the aortic valve by Cribier et al.14,15 The technical complexity and risks associated with this procedure appear likely to limit its widespread application, however.
We reported the development of a transarterial retrograde technique that has had favorable initial results.17,18 Evaluation of this procedure is ongoing and remains encouraging. Although the transfemoral arterial procedure has proved successful, some patients are poorly suited to this approach because of femoral, iliac, or aortic size; tortuosity; or atheroma. Our initial percutaneous valve development had used direct balloon catheter implantation from the left ventricular apex in a porcine model.12 It was anticipated that in such individuals aortic valve implantation using a left ventricular apical approach would have advantages.19,20
Obviously, optimal positioning of the prosthetic valve is critical. In our early percutaneous implantation experience, we sought to position the midpoint of the prosthesis adjacent to the major leaflet calcification. However, aortography and TEE often further clarify the relative positions of the prosthesis, annulus, and leaflets. Both native valve and native annulus appear to play a role in secure valve seating. The supra-annular aorta appears to play a less important role in valve securement. Ideal positioning appears to be dependent on prosthesis apposition with the annulus and outflow tract while extending just above the native leaflets. Device embolization was an unpredictable concern during early percutaneous implantation experience. However, with accurate positioning, sizing, and deployment techniques, prosthesis embolization appears likely to be rare, as reflected in this initial apical access experience.
Some degree of paravalvular insufficiency was common. In these patients with primarily aortic stenosis, however, the degree of native valve insufficiency before the procedure was comparable to the degree of paravalvular insufficiency after valve implantation. In no patient did the residual paravalvular leak appear clinically significant.
TEE demonstrates that paravalvular insufficiency is commonly due to inadequate apposition of the fabric cuff of the prosthesis and the aortic annulus. Our percutaneous implantation experience demonstrated that relative oversizing of the prosthesis is useful for reducing concerns about paravalvular insufficiency. Aggressive overdilation of a prosthesis would be expected to result in poor leaflet apposition, valvular insufficiency, and potentially impaired leaflet durability. Modest overdilation appears safe and was used commonly to ensure full prosthesis expansion in this series. An additional mechanism of paravalvular insufficiency is positioning the prosthesis either too high or too low in relation to the native annulus so that the fabric cuff is nonocclusive. The lack of clinically important paravalvular insufficiency in this series likely reflects routine prosthesis oversizing, modest postdilation when necessary, and increasingly accurate positioning.
Coronary obstruction or stroke could conceivably occur as a result of embolization of friable valvular debris, although we have not observed this outcome with percutaneous valve implantation.17 Although positioning a prosthesis above the annulus could result in placement of the fabric cuff at the level of the coronary ostia, any resulting obstruction seems unlikely because the diameter of the stent is less than that of the sinus of Valsalva. However, the prosthesis can displace a bulky native coronary leaflet over a coronary ostium.17 An unusually bulky leaflet may represent a relative contraindication to implantation of a valved stent. Because the native calcified leaflets were removed from the hemodynamic stresses of the bloodstream, it is likely that further growth of the native leaflets will cease. Indeed, the calcium may possibly be reabsorbed and affect long-term valve fixation. Assessment of coronary patency seems prudent after implantation; in some cases, however, overlying stent struts may interfere with selective coronary intubation.
Valvuloplasty data suggest that stroke is a surprisingly infrequent occurrence despite heavy aortic valve calcification.6,22 It is likely that the calcific debris familiar to surgeons replacing the aortic valve results from violation of the aortic valve endothelium when the valve is resected. If this covering is kept intact during valvuloplasty and valve deployment, embolization does not occur, despite displacement or even crumbling of the calcium within this covering.
Potential complications unique to this procedure include a significant incidence of perivalvular leak. Myocardial perforation or mitral or aortic trauma could occur from misdirected stiff catheters. Extensive surgical experience with venting of the left ventricle and commissurotomy of the mitral valve via the apex would suggest that apical access can be managed safely without excessive blood loss or subsequent arrhythmogenic potential.
As currently conceived, this procedure crosses the traditional boundaries of the catheterization laboratory and operating room. The skills required for apical exposure and repair are those of the surgeon, and the skills required for transcatheter implantation and angiography are those of the interventional cardiologist. Although patients in this report were not considered candidates for, and did not require, cardiopulmonary bypass, it is possible to envision a situation in which conversion to traditional surgery might be required to deal with unforeseen complications. Similarly, should valve embolization or coronary compromise occur at the time of deployment, interventional management may be required.17 Successful replication of this favorable experience will likely require a practiced team approach to implementation.
| Conclusions |
|---|
|
|
|---|
| Acknowledgments |
|---|
Drs Webb, Carere, and Thompson have acted as consultants to and received honoraria from Edwards Lifesciences, Irvine, Calif. Dr Webb received funding from Edwards Lifesciences for the REVIVE trial. The other authors report no conflicts.
| References |
|---|
|
|
|---|
Related Article:
Circulation 2006 114: 529.
This article has been cited by other articles:
![]() |
A. Vahanian, O. R. Alfieri, N. Al-Attar, M. J. Antunes, J. Bax, B. Cormier, A. Cribier, P. De Jaegere, G. Fournial, A. P. Kappetein, et al. Transcatheter valve implantation for patients with aortic stenosis: a position statement from the European Association of Cardio-Thoracic Surgery (EACTS) and the European Society of Cardiology (ESC), in collaboration with the European Association of Percutaneous Cardiovascular Interventions (EAPCI). Eur. J. Cardiothorac. Surg., July 1, 2008; 34(1): 1 - 8. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. G. Svensson, T. Dewey, S. Kapadia, E. E. Roselli, A. Stewart, M. Williams, W. N. Anderson, D. Brown, M. Leon, B. Lytle, et al. United States feasibility study of transcatheter insertion of a stented aortic valve by the left ventricular apex. Ann. Thorac. Surg., July 1, 2008; 86(1): 46 - 55. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. V. Vasilyev, P. M. Novotny, J. F. Martinez, H. Loyola, I. S. Salgo, R. D. Howe, and P. J. del Nido Stereoscopic vision display technology in real-time three-dimensional echocardiography-guided intracardiac beating-heart surgery. J. Thorac. Cardiovasc. Surg., June 1, 2008; 135(6): 1334 - 1341. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Descoutures, D. Himbert, L. Lepage, B. Iung, D. Detaint, D. Tchetche, E. Brochet, Y. Castier, J.-P. Depoix, P. Nataf, et al. Contemporary surgical or percutaneous management of severe aortic stenosis in the elderly Eur. Heart J., June 1, 2008; 29(11): 1410 - 1417. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Vahanian, O. Alfieri, N. Al-Attar, M. Antunes, J. Bax, B. Cormier, A. Cribier, P. De Jaegere, G. Fournial, A. P. Kappetein, et al. Transcatheter valve implantation for patients with aortic stenosis: a position statement from the European Association of Cardio-Thoracic Surgery (EACTS) and the European Society of Cardiology (ESC), in collaboration with the European Association of Percutaneous Cardiovascular Interventions (EAPCI) Eur. Heart J., June 1, 2008; 29(11): 1463 - 1470. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Sinhal, L. Altwegg, S. Pasupati, K. H. Humphries, M. Allard, P. Martin, A. Cheung, J. Ye, C. Kerr, S. V. Lichtenstein, et al. Atrioventricular Block After Transcatheter Balloon Expandable Aortic Valve Implantation J. Am. Coll. Cardiol. Intv., June 1, 2008; 1(3): 305 - 309. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. Webb Percutaneous Aortic Valve Replacement Will Become a Common Treatment for Aortic Valve Disease J. Am. Coll. Cardiol. Intv., April 1, 2008; 1(2): 122 - 126. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. K. Rosengart, T. Feldman, M. A. Borger, T. A. Vassiliades Jr, A. M. Gillinov, K. J. Hoercher, A. Vahanian, R. O. Bonow, and W. O'Neill Percutaneous and Minimally Invasive Valve Procedures: A Scientific Statement From the American Heart Association Council on Cardiovascular Surgery and Anesthesia, Council on Clinical Cardiology, Functional Genomics and Translational Biology Interdisciplinary Working Group, and Quality of Care and Outcomes Research Interdisciplinary Working Group Circulation, April 1, 2008; 117(13): 1750 - 1767. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. V. Vasilyev, I. Melnychenko, K. Kitahori, F. P. Freudenthal, A. Phillips, R. Kozlik-Feldmann, I. S. Salgo, P. J. del Nido, and E. A. Bacha Beating-heart patch closure of muscular ventricular septal defects under real-time three-dimensional echocardiographic guidance: A preclinical study J. Thorac. Cardiovasc. Surg., March 1, 2008; 135(3): 603 - 609. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. H. Rahimtoola The Year in Valvular Heart Disease. J. Am. Coll. Cardiol., February 19, 2008; 51(7): 760 - 770. [Full Text] [PDF] |
||||
![]() |
M. J. Davidson and D. S. Baim Percutaneous Aortic Valve Interventions Card. Surg. Adult, January 1, 2008; 3(2008): 963 - 971. [Full Text] |
||||
![]() |
T. Feldman and M. B. Leon Prospects for Percutaneous Valve Therapies Circulation, December 11, 2007; 116(24): 2866 - 2877. [Full Text] [PDF] |
||||
![]() |
E. Berreklouw, B. Vogel, H. Fischer, F. Weinberg, G. L. van Rijk-Zwikker, G. J. Van Nooten, and L. H. Cohn Fast sutureless implantation of mechanical aortic valve prostheses using Nitinol attachment rings: Feasibility in acute pig experiments. J. Thorac. Cardiovasc. Surg., December 1, 2007; 134(6): 1508 - 1512. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Marcheix, Y. Lamarche, C. Berry, A. Asgar, J.-C. Laborde, A. Basmadjian, A. Ducharme, A. Denault, R. Bonan, and R. Cartier Surgical aspects of endovascular retrograde implantation of the aortic CoreValve bioprosthesis in high-risk older patients with severe symptomatic aortic stenosis. J. Thorac. Cardiovasc. Surg., November 1, 2007; 134(5): 1150 - 1156. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Walther, P. Simon, T. Dewey, G. Wimmer-Greinecker, V. Falk, M. T. Kasimir, M. Doss, M. A. Borger, G. Schuler, D. Glogar, et al. Transapical Minimally Invasive Aortic Valve Implantation: Multicenter Experience Circulation, September 11, 2007; 116(11_suppl): I-240 - I-245. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. M. Flecher, J. W. Curry, T. M. Joudinaud, C. L. Kegel, P. A. Weber, and C. M.G. Duran Coronary flow obstruction in percutaneous aortic valve replacement. An in vitro study Eur. J. Cardiothorac. Surg., August 1, 2007; 32(2): 291 - 294. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. R. Dixon, C. L. Grines, and W. W. O'Neill The Year in Interventional Cardiology J. Am. Coll. Cardiol., July 17, 2007; 50(3): 270 - 285. [Full Text] [PDF] |
||||
![]() |
E. Grube, G. Schuler, L. Buellesfeld, U. Gerckens, A. Linke, P. Wenaweser, B. Sauren, F.-W. Mohr, T. Walther, B. Zickmann, et al. Percutaneous Aortic Valve Replacement for Severe Aortic Stenosis in High-Risk Patients Using the Second- and Current Third-Generation Self-Expanding CoreValve Prosthesis: Device Success and 30-Day Clinical Outcome J. Am. Coll. Cardiol., July 3, 2007; 50(1): 69 - 76. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Walther, V. Falk, T. Dewey, J. Kempfert, F. Emrich, B. Pfannmuller, P. Broske, M. A. Borger, G. Schuler, M. Mack, et al. Valve-in-a-Valve Concept for Transcatheter Minimally Invasive Repeat Xenograft Implantation J. Am. Coll. Cardiol., July 3, 2007; 50(1): 56 - 60. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Grube, L. Buellesfeld, U. Gerckens, R. Mueller, M. Menichelli, T. Schmidt, J. C. Laborde, T. Felderhoff, S. Iversen, B. Sauren, et al. Response to Letter Regarding Article, "Percutaneous Implantation of the CoreValve Self-Expanding Valve Prosthesis in High-Risk Patients With Aortic Valve Disease: The Siegburg First-in-Man Study" Circulation, June 5, 2007; 115(22): e613 - e613. [Full Text] [PDF] |
||||
![]() |
M. J. Antunes Off-pump aortic valve replacement with catheter-mounted valved stents.: Is the future already here? Eur. J. Cardiothorac. Surg., January 1, 2007; 31(1): 1 - 3. [Full Text] [PDF] |
||||
![]() |
T. Walther and F. W. Mohr Aortic valve surgery: time to be open-minded and to rethink Eur. J. Cardiothorac. Surg., January 1, 2007; 31(1): 4 - 6. [Full Text] [PDF] |
||||
![]() |
F. Beyersdorf Transapical transcatheter aortic valve implantation Eur. J. Cardiothorac. Surg., January 1, 2007; 31(1): 7 - 8. [Full Text] [PDF] |
||||