(Circulation. 2006;114:2148-2153.)
© 2006 American Heart Association, Inc.
Interventional Cardiology |
From Asan Medical Center (Y.-H.K., S.-W.P., S.-W.L., D.-W.P., S.-C.Y., C.W.L., M.-K.H., S.-J.P.), University of Ulsan College of Medicine, Seoul, Korea; Chonbuk National University Hospital (H.-S.K., J.-K.K.), Jeonju, Korea; Chungnam National University Hospital (J.-H.P., J.-H.L., S.W.C., I.-W.S.), Daejeon, Korea; Soonchunhyang University Bucheon Hospital (Y.H.C., N.-H.L.), Soonchunhyang, Korea; and Pusan National University Hospital (J.H.K., K.-J.C.), Busan, Korea.
Correspondence to Seung-Jung Park, MD, PhD, Department of Medicine, University of Ulsan College of Medicine, Asan Medical Center, 388-1 Poongnap-dong, Songpa-gu, Seoul, 138-736, Korea. E-mail sjpark{at}amc.seoul.kr
Received September 22, 2006; revision received October 6, 2006; accepted October 13, 2006.
| Abstract |
|---|
|
|
|---|
Methods and Results The present randomized, multicenter, prospective study compared the use of long (
32 mm) SES with PES in 500 patients with long (
25 mm) native coronary lesions. The primary end point of the trial was the rate of binary in-segment restenosis according to follow-up angiography at 6 months. The SES and PES groups had similar baseline characteristics. Lesion length was 33.9±11.6 mm in the SES group and 34.5±12.6 mm in the PES group (P=0.527). The in-segment binary restenosis rate was significantly lower in the SES group than in the PES group (3.3% versus 14.6%; relative risk 0.23; P<0.001). In-stent late loss of lumen diameter was 0.09±0.37 mm in the SES group and 0.45±0.55 mm in the PES group (P<0.001). In patients with restenoses, a pattern of focal restenosis was more common in the SES group than in the PES group (100% versus 53.3%, P=0.031). Consequently, SES patients had a lower rate of target-lesion revascularization at 9 months (2.4% versus 7.2%, P=0.012). The incidence of death (0.8% in SES versus 0% in PES, P=0.499) or myocardial infarction (8.8% in SES versus 10.8% in PES, P=0.452) at 9 months of follow-up was not statistically different between the 2 groups.
Conclusions For patients with long native coronary artery disease, SES implantation was associated with a reduced incidence of angiographic restenosis and a reduced need for target-lesion revascularization compared with PES implantation.
Key Words: coronary disease stents restenosis
| Introduction |
|---|
|
|
|---|
Clinical Perspective p 2153
Although the use of DES has decreased the effect of lesion length on restenosis, long coronary lesions remain at a higher risk of unfavorable outcomes after percutaneous coronary intervention.9,1418 In the Long-DES Registry Study, SES was associated with a lower angiographic restenosis rate than PES in patients with lesions >24 mm in length.9 That study, however, used a nonrandomized observational methodology.
The present study, the Long-DES-II Study, compared long-term angiographic and clinical outcomes in patients with long coronary lesions treated with SES or PES. The study used a randomized, multicenter, controlled design approach.
| Methods |
|---|
|
|
|---|
18 years of age who had angina pectoris or a positive stress test and a native coronary lesion. The study involved 5 cardiac centers in Korea between August 2004 and August 2005. Angiographic eligibility for inclusion consisted of a target lesion with a diameter stenosis
50%, visual reference diameter
2.5 mm, and lesion length
25 mm. Inclusion required that the planned total stent length was
32 mm. Patients were excluded if they had (1) contraindication to aspirin, clopidogrel, or cilostazol; (2) left main disease (diameter stenosis
50% by visual estimate); (3) graft vessel disease; (4) left ventricular ejection fraction <30%; (5) recent history of hematologic disease or leukocyte count <3000/mm3 and/or platelet <100 000/mm3; (6) hepatic dysfunction with AST or ALT
3 times the upper normal reference limit; (7) history of renal dysfunction or a serum creatinine level
2.0 mg/dL; (8) serious noncardiac comorbid disease with a life expectancy <1 year; (9) planned bifurcation stenting in the side branch; (10) primary angioplasty for acute myocardial infarction (MI) within 24 hours; or (11) inability to follow the protocol. The institutional review board at each participating center approved the protocol. All patients provided written informed consent.
Randomization and Procedures
Once the guidewire had crossed the target lesion, patients were randomly assigned to either SES or PES implantation on a 1:1 basis with sealed envelopes that contained a computer-generated randomization sequence. Stratified and block randomization was done according to participation sites. In patients with multiple lesions that fulfilled the inclusion and exclusion criteria, the operator determined the hierarchy of lesions and declared the target lesion for each patient before the procedure. The same type of allocated stent was used for all lesions in patients with multiple lesions.
Coronary stenting was performed according to the standard technique.9 The decision of predilation or direct stenting was made by the operator. Beginning at least 24 hours before the procedure and continuing thereafter, all patients received aspirin (200 mg daily) and clopidogrel (a loading dose of 300 mg and then 75 mg daily for at least 6 months). To evaluate the effect of cilostazol on the reduction of in-stent late loss of lumen diameter after DES placement, cilostazol (a loading dose of 200 mg immediately after randomization and 100 mg twice a day for 6 months) was administered to 250 patients who were randomly allocated to receive this drug after the procedure on the basis of a 2-by-2 factorial design. The use of intravenous glycoprotein IIb/IIIa inhibitors was at the operators discretion. A 12-lead ECG was obtained after the procedure and before discharge from the hospital. Serum levels of creatine kinase, its MB isoenzyme, and troponin I were assessed 8, 12, and 24 hours after the procedure and thereafter if considered necessary.
Study End Point and Definitions
The primary end point of the trial was the rate of binary in-segment restenosis, defined as a diameter stenosis >50% with quantitative coronary angiography (QCA) at 6 months after the index procedure. The secondary end points included 9-month clinical and 6-month angiographic outcomes, such as rates of death; MI; target-lesion revascularization (TLR); target-vessel revascularization (TVR); composites of death, MI, and TLR/TVR; stent thrombosis; device success, defined as an in-segment final diameter stenosis <50% by QCA with the assigned device only; in-stent binary restenosis rate; and late loss, both in-stent and in-segment. Q-wave MI was defined by the postprocedural presence of new Q waves of >0.04 second in 2 contiguous leads. NonQ-wave MI was defined as a creatine kinase-MB fraction or troponin I concentration >3 times the normal upper limit. TLR was considered clinically driven if prompted by symptoms or signs consistent with myocardial ischemia or if the lesion diameter stenosis was >70% at follow-up.7 Stent thrombosis was defined as any of the following after the procedure19,20: angiographic documentation of stent occlusion with or without the presence of thrombus associated with an acute ischemic event, unexplained sudden death, or MI not clearly attributable to another coronary lesion.
Follow-Up
Repeat coronary angiography was routinely recommended at 6 months after stenting or earlier if indicated by clinical symptoms or evidence of myocardial ischemia. Clinical follow-up visits were scheduled at 30, 90, 180, and 270 days. At every visit, physical examinations, ECGs, and laboratory examinations were conducted, and the occurrence of adverse cardiac events and the recurrence of angina were monitored. At each participating center, patient data were recorded prospectively on standard case report forms and gathered in the central data management center (Asan Medical Center, Seoul, Korea). All adverse clinical events were adjudicated by an independent events committee blinded to the treatment groups.
QCA Analysis
Coronary angiograms were obtained before the procedure (baseline), after the procedure, and at follow-up and were submitted to the angiographic core analysis center (Asan Medical Center, Seoul, Korea) for analysis by independent angiographers. Digital angiograms were analyzed with an automated edge-detection system (CASS II, Pie Medical, Maastricht, the Netherlands). Angiographic variables included absolute lesion length, stent length, reference-vessel diameter, minimum lumen diameter, percent diameter stenosis, binary restenosis rate, immediate gain, late loss, and the patterns of recurrent restenosis. QCA measurements of target lesions were obtained for both the stented segment only (in-stent) and the region that included the stented segment and the margins 5 mm proximal and distal to the stent (in-segment). Lesion morphology was defined according to the guidelines of American College of Cardiology and American Heart Association.21 Patterns of angiographic restenosis were quantitatively assessed with the Mehran classification.22
Statistical Analysis
On the basis of the Long-DES Registry Study,9 we assumed an in-segment angiographic restenosis rate of 9% in the SES group and 21% in the PES group. Using a 2-sided 5% significance level, we estimated that 201 patients per group were needed to detect this difference with a statistical power of 90%. Expecting that
20% of the patients would not return for angiographic follow-up, total sample size was estimated to be 500 patients (250 patients per group). Analyses of the 2 groups were performed according to the intention-to-treat principle. Because of a 2-by-2 factorial design, a possible interaction between types of DES and use of cilostazol was evaluated by logistic regression analysis. Continuous variables are presented as mean±SD or median (interquartile range), and compared with Student unpaired t or Mann-Whitney U tests and repeated-measures ANOVA with the Bonferroni correction for post hoc comparisons as appropriate. Categorical variables are presented as numbers or percentages and were compared with
2 or Fishers exact tests. The relative risk and its 95% CI were computed for outcome measures. The Breslow-Day test was performed to assess treatment by participating center interaction.23 The adjusted relative risk and CI after we controlled for the center interaction was computed by the Mantel-Haenszel method. A probability value <0.05 was considered to indicate a significant difference. Statistical analysis was performed with commercially available software (SPSS 11 for Windows, SPSS Inc, Chicago, Ill).
The authors had full access to the data and take full responsibility for the integrity of the data. All authors have read and agree to the manuscript as written.
| Results |
|---|
|
|
|---|
|
|
Angiographic Outcomes
Baseline and postprocedural QCA outcomes for the 2 groups are shown in Table 3. Both groups had similar baseline and postprocedural QCA characteristics in terms of lesion length, reference diameter, minimum lumen diameter, diameter stenosis, and immediate gain. Follow-up angiography was performed in 210 SES patients (84.0%) and 205 PES patients (82.0%; P=0.552). The median duration of angiographic follow-up was 188 (interquartile range 178 to 200) and 186 (interquartile range 177 to 205) days for the SES and PES groups, respectively (P=0.857). Results of QCA measurements at follow-up are shown in Table 3.
|
In-segment restenosis, the prespecified primary end point of the present study, was identified in 7 SES patients (3.3%) and 30 PES patients (14.6%; relative risk, 0.23; 95% confidence interval [CI], 0.10 to 0.51; P<0.001). The in-stent restenosis rate was also lower in the SES group than in the PES group (2.9% versus 11.7%; relative risk, 0.24; 95% CI, 0.10 to 0.58; P=0.001). After we controlled for the center interaction, almost identical relative risks for in-segment (Breslow-Day test, P=0.054; relative risk, 0.25; 95% CI, 0.12 to 0.56) and in-stent restenosis (Breslow-Day test, P=0.139; relative risk, 0.27; 95% CI, 0.12 to 0.64) were found between the 2 groups. The interaction between types of DES and use of cilostazol was not significant (P=0.219 for in-segment restenosis and P=0.191 for in-stent restenosis). Patterns of in-stent restenosis are shown in Table 4. In patients with restenoses, a pattern of focal restenosis (type I) was more common in the SES group than in the PES group (100% versus 53.3%, P=0.031). Similarly, length of in-stent restenosis was shorter in the SES group than in the PES group.
|
Late loss of in-stent and in-segment lumen diameter was lower for the SES group than for the PES group. Consequently, minimum lumen diameter at follow-up was significantly larger in the SES group than in the PES group. The Figure shows the cumulative percent of in-segment diameter stenosis before and after the procedure and at follow-up angiography. Serial angiographic changes of in-segment minimum lumen diameter (P<0.001) and percent diameter stenosis (P<0.001) were significant over time, before and after the procedure and at follow-up in each group. Serial changes were also significantly different between the 2 groups with regard to in-segment minimum lumen diameter (P=0.021) and diameter stenosis (P=0.001).
|
Clinical Outcomes
A minimum 9-month clinical follow-up was performed in 249 SES patients (99.6%) and 245 PES patients (98.0%; P=0.216). Clinical outcomes at 30 days and 9 months are shown in Table 5. Two SES patients (0.8%) and no PES patients died during the study period (P=0.499). MI occurred in 22 SES patients (8.8%) and 27 PES patients (10.8%; P=0.452). Forty-seven patients with MI (9.4%) were observed during hospitalization as having nonQ-wave infarctions related to the procedure. Stent thrombosis occurred in 2 SES patients and in no PES patients (P=0.499). Of the 2 cases of stent thrombosis, 1 was angiographically documented. Consequently, the rates of TLR (2.4% versus 7.2%; relative risk, 0.33; 95% CI, 0.13 to 0.83; P=0.012) and TVR (3.2% versus 7.6%; relative risk, 0.42; 95% CI, 0.19 to 0.94; P=0.030) were lower in the SES group than in the PES group. Clinically driven TLR (1.6% versus 5.6%, P=0.028) and TVR (2.4% versus 7.2%, P=0.012) rates were also lower in the SES group than in the PES group.
|
| Discussion |
|---|
|
|
|---|
Several previous reports of randomized studies, registry data, and a meta-analysis showed that SES were more effective than PES in reducing the restenosis rate.35,811 Those studies assumed that the lower late loss associated with SES might contribute to lower incidences of angiographic restenosis and repeat revascularization than PES.35,811,24,25 Not all studies, however, found SES to be superior.6,7,12,13 Some studies challenged the expected relationship between late loss and clinical outcome.12,13,26 The present study supported the formal predictive model that late loss is closely related to long-term angiographic and clinical outcomes.2,24,25 In the present study, SES consistently reduced late loss, the incidence of angiographic restenosis, and the need for TLR compared with PES in patients with long coronary lesions. In addition, the fact that a focal restenosis pattern was more common in SES patients may be an additional benefit, because the focal pattern is a predictor of benign clinical prognosis compared with the diffuse pattern.27
In terms of the prespecified primary end point, the present study found a 77.2% relative risk reduction (11.3% of absolute) in the in-segment restenosis rate in SES compared with PES patients. This relative reduction is greater than that reported in previous randomized studies that showed a 7% to 42% relative reduction,8 possibly because the present study involved very complex coronary lesions with a long diseased segment, which may make differences between the performances of the 2 stents more pronounced. The previous studies with bare-metal stents suggested that the potential risk of restenosis in stent trials may be strongly dependent on the inclusion of patients with complex lesions.28,29 Similarly, randomized trials involving more complex patients/lesions, such as the ISAR-DIABETES (Intracoronary Stenting and Angiographic Results: Do Diabetic Patients Derive Similar Benefit from Paclitaxel-Eluting and Sirolimus-Eluting Stents?) and the ISAR-DESIRE (ISAR: Drug-Eluting Stents for In-Stent Restenosis) study, showed a more significant benefit of SES over PES than trials that involved relatively simple lesions, such as the REALITY (Randomized multicenter head-to-head comparison of the sirolimus-eluting stent and the paclitaxel-eluting stent) and TAXi (A prospective randomized comparison between paclitaxel and sirolimus stents in the real world of interventional cardiology) trials.47
In the present study, the incidence of serious adverse clinical outcomes, such as death, MI, or stent thrombosis, was low for both groups. Although angiographic and clinical stent thrombosis occurred in 2 SES patients, the incidence of these events was not found to differ significantly between the 2 groups. This finding is consistent with previous randomized studies that showed that both types of DES were safe in terms of an acceptably low incidence of cardiac mortality and stent thrombosis.8 The relatively high incidence of periprocedural nonQ-wave MI was related to the inclusion of patients with complex lesions and a high prevalence of acute coronary syndrome.1618
The late loss and restenosis rates for both stents were relatively low in the present study compared with the previous Long-DES Registry Study.9 In the Long-DES Registry Study, in-stent late loss was 0.26 mm for SES and 0.78 mm for PES. In contrast, in the present study, in-stent late loss was 0.10 mm for SES and 0.43 mm for PES. Consequently, the in-stent angiographic restenosis rate was lower in the present study for both SES (2.9% versus 7.6%) and PES (11.7% versus 16.0%) implantations. The different outcomes in the 2 studies might be in part a result of differences in study design, enrolled patient/lesion characteristics, and stenting procedures. A significant interaction by 2-by-2 factorial design was not introduced into the primary outcome, however.
The present study has some limitations. First, the routine 6-month angiography performed in the study might have resulted in an underestimation of the rates of restenosis and TLR compared with a study with a longer angiographic follow-up period. The present median duration of angiographic follow-up of 187 days, however, was similar to that of previous studies.4,5 Second, the present study was underpowered to detect differences in serious adverse clinical outcomes between the 2 groups. A large number of patients with longer follow-up may be required.
In conclusion, the present study showed that use of an SES resulted in a reduced incidence of angiographic restenosis and a reduced need for TLR compared with use of a PES.
| Acknowledgments |
|---|
This study was supported by the Cardiovascular Research Foundation (Korea), a grant from the Korean Ministry of Health & Welfare as part of the Korea Health 21 Research & Development Project (0412-CR02-0704-0001), and a grant from Cordis, a Johnson & Johnson company (Miami, Fla).
Disclosures
Young-Hak Kim and Seung-Jung Park have received research grants from Cordis.
| References |
|---|
|
|
|---|
2. Kereiakes DJ, Kuntz RE, Mauri L, Krucoff MW. Surrogates, substudies, and real clinical end points in trials of drug-eluting stents. J Am Coll Cardiol. 2005; 45: 12061212.
3. Windecker S, Remondino A, Eberli FR, Juni P, Raber L, Wenaweser P, Togni M, Billinger M, Tuller D, Seiler C, Roffi M, Corti R, Sutsch G, Maier W, Luscher T, Hess OM, Egger M, Meier B. Sirolimus-eluting and paclitaxel-eluting stents for coronary revascularization. N Engl J Med. 2005; 353: 653662.
4. Dibra A, Kastrati A, Mehilli J, Pache J, Schuhlen H, von Beckerath N, Ulm K, Wessely R, Dirschinger J, Schomig A; ISAR-DIABETES Study Investigators. Paclitaxel-eluting or sirolimus-eluting stents to prevent restenosis in diabetic patients. N Engl J Med. 2005; 353: 663670.
5. Kastrati A, Mehilli J, von Beckerath N, Dibra A, Hausleiter J, Pache J, Schuhlen H, Schmitt C, Dirschinger J, Schomig A; ISAR-DESIRE Study Investigators. Sirolimus-eluting stent or paclitaxel-eluting stent vs balloon angioplasty for prevention of recurrences in patients with coronary in-stent restenosis: a randomized controlled trial. JAMA. 2005; 293: 165171.
6. Goy JJ, Stauffer JC, Siegenthaler M, Benoit A, Seydoux C. A prospective randomized comparison between paclitaxel and sirolimus stents in the real world of interventional cardiology: the TAXi trial. J Am Coll Cardiol. 2005; 45: 308311.
7. Morice MC, Colombo A, Meier B, Serruys P, Tamburino C, Guagliumi G, Sousa E, Stoll HP; REALITY Trial Investigators. Sirolimus- vs paclitaxel-eluting stents in de novo coronary artery lesions: the REALITY trial: a randomized controlled trial. JAMA. 2006; 295: 895904.
8. Kastrati A, Dibra A, Eberle S, Mehilli J, Suarez de Lezo J, Goy JJ, Ulm K, Schomig A. Sirolimus-eluting stents vs paclitaxel-eluting stents in patients with coronary artery disease: meta-analysis of randomized trials. JAMA. 2005; 294: 819825.
9. Kim YH, Park SW, Lee CW, Hong MK, Gwon HC, Jang YS, Lee MM, Koo BK, Oh DJ, Seung KB, Tahk SJ, Yoon JH, Park SJ. Comparison of sirolimus-eluting stent, paclitaxel-eluting stent and bare metal stent in the treatment of long coronary lesions. Catheter Cardiovasc Interv. 2006; 67: 181187.[CrossRef][Medline] [Order article via Infotrieve]
10. Park KH, Park SW, Hong MK, Kim YH, Lee BK, Park DW, Choi BR, Kim MJ, Park KM, Lee CW, Cheong SS, Kim JJ, Park SJ. Comparison of the effectiveness of sirolimus- and paclitaxel-eluting stents for small coronary artery lesions. Catheter Cardiovasc Interv. 2006; 67: 589594.[CrossRef][Medline] [Order article via Infotrieve]
11. Jang JS, Hong MK, Lee CW, Park DW, Lee BK, Kim YH, Han KH, Kim JJ, Park SW, Park SJ. Comparison between sirolimus- and paclitaxel-eluting stents for the treatment of chronic total occlusions. J Invasive Cardiol. 2006; 18: 205208.[Medline] [Order article via Infotrieve]
12. Ong AT, Serruys PW, Aoki J, Hoye A, van Mieghem CA, Rodriguez-Granillo GA, Valgimigli M, Sonnenschein K, Regar E, van der Ent M, de Jaegere PP, McFadden EP, Sianos G, van der Giessen WJ, de Feyter PJ, van Domburg RT. The unrestricted use of paclitaxel- versus sirolimus-eluting stents for coronary artery disease in an unselected population: one-year results of the Taxus-Stent Evaluated at Rotterdam Cardiology Hospital (T-SEARCH) registry. J Am Coll Cardiol. 2005; 45: 11351141.
13. Airoldi F, Briguori C, Iakovou I, Stankovic G, Biondi-Zoccai G, Carlino M, Chieffo A, Montorfano M, Cosgrave J, Michev I, Rogacka R, Sangiorgi GM, Colombo A. Comparison of sirolimus versus paclitaxel eluting stents for treatment of coronary in-stent restenosis. Am J Cardiol. 2006; 97: 11821187.[CrossRef][Medline] [Order article via Infotrieve]
14. Lee CW, Park DW, Lee BK, Kim YH, Hong MK, Kim JJ, Park SW, Jung Park SJ. Predictors of restenosis after placement of drug-eluting stents in one or more coronary arteries. Am J Cardiol. 2006; 97: 506511.[CrossRef][Medline] [Order article via Infotrieve]
15. Kastrati A, Dibra A, Mehilli J, Mayer S, Pinieck S, Pache J, Dirschinger J, Schomig A. Predictive factors of restenosis after coronary implantation of sirolimus- or paclitaxel-eluting stents. Circulation. 2006; 113: 22932300.
16. Degertekin M, Arampatzis CA, Lemos PA, Saia F, Hoye A, Daemen J, Tanabe K, Lee CH, Hofma SJ, Sianos G, McFadden E, van der Giessen W, Smits PC, de Feyter PJ, van Domburg RT, Serruys PW. Very long sirolimus-eluting stent implantation for de novo coronary lesions. Am J Cardiol. 2004; 93: 826829.[CrossRef][Medline] [Order article via Infotrieve]
17. Tsagalou E, Chieffo A, Iakovou I, Ge L, Sangiorgi GM, Corvaja N, Airoldi F, Montorfano M, Michev I, Colombo A. Multiple overlapping drug-eluting stents to treat diffuse disease of the left anterior descending coronary artery. J Am Coll Cardiol. 2005; 45: 15701573.
18. Dawkins KD, Grube E, Guagliumi G, Banning AP, Zmudka K, Colombo A, Thuesen L, Hauptman K, Marco J, Wijns W, Popma JJ, Koglin J, Russell ME; TAXUS VI Investigators. Clinical efficacy of polymer-based paclitaxel-eluting stents in the treatment of complex, long coronary artery lesions from a multicenter, randomized trial: support for the use of drug-eluting stents in contemporary clinical practice. Circulation. 2005; 112: 33063313.
19. Iakovou I, Schmidt T, Bonizzoni E, Ge L, Sangiorgi GM, Stankovic G, Airoldi F, Chieffo A, Montorfano M, Carlino M, Michev I, Corvaja N, Briguori C, Gerckens U, Grube E, Colombo A. Incidence, predictors, and outcome of thrombosis after successful implantation of drug-eluting stents. JAMA. 2005; 293: 21262130.
20. Park DW, Park SW, Park KH, Lee BK, Kim YH, Lee CW, Hong MK, Kim JJ, Park SJ. Frequency of and risk factors for stent thrombosis after drug-eluting stent implantation during long-term follow-up. Am J Cardiol. 2006; 98: 352356.[CrossRef][Medline] [Order article via Infotrieve]
21. Ellis SG, Vandormael MG, Cowley MJ, DiSciascio G, Deligonul U, Topol EJ, Bulle TM. Coronary morphologic and clinical determinants of procedural outcome with angioplasty for multivessel coronary disease: implications for patient selection. Circulation. 1990; 82: 11931202.
22. Mehran R, Dangas G, Abizaid AS, Mintz GS, Lansky AJ, Satler LF, Pichard AD, Kent KM, Stone GW, Leon MB. Angiographic patterns of in-stent restenosis: classification and implications for long-term outcome. Circulation. 1999; 100: 18721878.
23. Breslow NE, Day NE. Statistical methods in cancer research: volume II: the design and analysis of cohort studies. IARC Sci Publ. 1987; 82: 1406.
24. Mauri L, Orav EJ, Candia SC, Cutlip DE, Kuntz RE. Robustness of late lumen loss in discriminating drug-eluting stents across variable observational and randomized trials. Circulation. 2005; 112: 28332839.
25. Mauri L, Orav EJ, Kuntz RE. Late loss in lumen diameter and binary restenosis for drug-eluting stent comparison. Circulation. 2005; 111: 34353442.
26. Ellis SG, Popma JJ, Lasala JM, Koglin JJ, Cox DA, Hermiller J, OShaughnessy C, Mann JT, Turco M, Caputo R, Bergin P, Greenberg J, Stone GW. Relationship between angiographic late loss and target lesion revascularization after coronary stent implantation: analysis from the TAXUS-IV trial. J Am Coll Cardiol. 2005; 45: 11931200.
27. Cosgrave J, Melzi G, Biondi-Zoccai GG, Airoldi F, Chieffo A, Sangiorgi GM, Montorfano M, Michev I, Carlino M, Bonizzoni E, Colombo A. Drug-eluting stent restenosis the pattern predicts the outcome. J Am Coll Cardiol. 2006; 47: 23992404.
28. Edelman ER, Rogers C. Stent-versus-stent equivalency trials: are some stents more equal than others? Circulation. 1999; 100: 896898.
29. Hausleiter J, Kastrati A, Mehilli J, Schuhlen H, Pache J, Dotzer F, Sattelberger U, Dirschinger J, Schomig A. Impact of lesion complexity on the capacity of a trial to detect differences in stent performance: results from the ISAR-STEREO trial. Am Heart J. 2003; 146: 882886.[CrossRef][Medline] [Order article via Infotrieve]
![]() |
S.-J. Park and Y.-H. Kim Current Status of Percutaneous Coronary Intervention With Drug-Eluting Stents in Asia Circulation, December 16, 2008; 118(25): 2730 - 2737. [Full Text] [PDF] |
||||
![]() |
S.-W. Lee, S.-W. Park, Y.-H. Kim, S.-C. Yun, D.-W. Park, C. W. Lee, M.-K. Hong, K.-S. Rhee, J. K. Chae, J.-K. Ko, et al. A Randomized Comparison of Sirolimus- Versus Paclitaxel-Eluting Stent Implantation in Patients With Diabetes Mellitus J. Am. Coll. Cardiol., August 26, 2008; 52(9): 727 - 733. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Billinger, J. Beutler, K. R. Taghetchian, A. Remondino, P. Wenaweser, S. Cook, M. Togni, C. Seiler, C. Stettler, F. R. Eberli, et al. Two-year clinical outcome after implantation of sirolimus-eluting and paclitaxel-eluting stents in diabetic patients Eur. Heart J., March 2, 2008; 29(6): 718 - 725. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. Galloe, L. Thuesen, H. Kelbaek, P. Thayssen, K. Rasmussen, P. R. Hansen, N. Bligaard, K. Saunamaki, A. Junker, J. Aaroe, et al. Comparison of Paclitaxel- and Sirolimus-Eluting Stents in Everyday Clinical Practice: The SORT OUT II Randomized Trial JAMA, January 30, 2008; 299(4): 409 - 416. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Schomig, A. Dibra, S. Windecker, J. Mehilli, J. Suarez de Lezo, C. Kaiser, S.-J. Park, J.-J. Goy, J.-H. Lee, E. Di Lorenzo, et al. A Meta-Analysis of 16 Randomized Trials of Sirolimus-Eluting Stents Versus Paclitaxel-Eluting Stents in Patients With Coronary Artery Disease J. Am. Coll. Cardiol., October 2, 2007; 50(14): 1373 - 1380. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Kereiakes The Emperor's New Clothes: Another Cypher Versus Taxus Post-Hoc Meta-Analysis J. Am. Coll. Cardiol., October 2, 2007; 50(14): 1381 - 1385. [Full Text] [PDF] |
||||
![]() |
N. M M Pires, D. Eefting, M. R de Vries, P. H A Quax, and J W. Jukema Sirolimus and paclitaxel provoke different vascular pathological responses after local delivery in a murine model for restenosis on underlying atherosclerotic arteries Heart, August 1, 2007; 93(8): 922 - 927. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. R. Dixon, C. L. Grines, and W. W. O'Neill The Year in Interventional Cardiology J. Am. Coll. Cardiol., July 17, 2007; 50(3): 270 - 285. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2006 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |