Circulation. 2006;114:I-425-I-429
doi: 10.1161/CIRCULATIONAHA.105.000695
(Circulation. 2006;114:I-425 I-429.)
© 2006 American Heart Association, Inc.
Surgery for Coronary Artery Disease |
Long-Term Outcomes of the Symmetry Vein Graft Anastomosis Device
A Matched Case-Control Analysis
Radha G. Kachhy, MD;
David F. Kong, MD, AM;
Emily Honeycutt, MBI;
Linda K. Shaw, MS;
R. Duane Davis, MD
From Duke Clinical Research Institute (R.G.K., D.F.K., E.H., L.K.S.), the Department of Medicine, Division of Cardiovascular Medicine (R.G.K., D.F.K.), and Department of Surgery, Division of Cardiothoracic Surgery (R.D.D.), Duke University Medical Center, Durham, NC.
Correspondence to David F. Kong, Box 3850, Division of Cardiology, Duke University Medical Center, Durham, NC 27710. E-mail David.F.Kong{at}duke.edu
 |
Abstract
|
|---|
Background The Symmetry Bypass Connector (St. Jude Medical)
was developed to rapidly anastomose saphenous vein grafts to
the aorta during coronary bypass surgery (CABG) without cross-clamping.
Previous uncontrolled studies of this device observed vein graft
closures at six months, possibly attributable to neointimal
hyperplasia.
Methods and Results To assess the long-term clinical outcomes of the Symmetry device, we performed a retrospective matched casecontrol analysis of patients who underwent CABG at Duke Medical Center between January 1, 2002 and December 31, 2003. In 121 patients, at least one proximal anastomosis used a Symmetry device. Traditional suture methods were used in 178 control patients, matched by age group, gender, use of cardiopulmonary bypass, and Hannan perioperative risk score. One-year outcomes were compared using the log-rank test and Cox proportional hazards regression models. Major adverse events were more frequent among cases compared with controls. By unadjusted KaplanMeier analysis for the composite end point of death, nonfatal myocardial infarction, repeat cardiac catheterization, or repeat CABG, there was a trend towards increased events in the Symmetry device group (P=0.053). No significant differences were detected for stroke, all cause mortality, or the combined end point of death or nonfatal myocardial infarction.
Conclusions Use of the Symmetry Bypass Connector was associated with increased risk for major adverse events at 1 year, suggestive of early graft closure. The potential reductions in operative stroke risk (from the elimination of aortic cross-clamping) must be weighed against the potential risk for later adverse events. These findings encourage close follow-up of patients who received this device.
Key Words: anastomosis atherosclerosis bypass surgery veins
 |
Introduction
|
|---|
More than 500 000 coronary artery bypass grafting (CABG) procedures
are performed in the United States each year. Perioperative
stroke is a devastating complication of CABG surgery, affecting
between 1.6% and 6% of patients.
1,2 This risk increases with
age. Patients aged 75 years or older had postoperative neurologic
events more than twice as frequently as patients aged 65 to
74, and 9-times more frequently than patients younger than 65.
3 Cross-clamping and manipulation of the aorta contribute to the
risk of perioperative stroke. CABG without cardiopulmonary bypass
(off-pump CABG) is associated with lower stroke risk and mortality
compared with the conventional cardiopulmonary bypass approach
(on-pump CABG.)
4,5 The increasing number of off-pump CABG procedures
inspired rapid methods for constructing proximal and distal
graft anastomoses using sutureless vascular connectors.
6 The
Symmetry Bypass connector (St. Jude Medical) was a nitinol,
star-shaped device designed to rapidly anastomose saphenous
vein grafts to the aorta during CABG without use of side-biting
or cross- clamping of the aorta. The Symmetry device was approved
by the United States Food and Drug Administration in May 2001,
and was the first such device marketed in the United States.
More than 30 000 connector devices have since been implanted
worldwide.
After the introduction of the Symmetry device, several case series reported early vein graft closures.712 Several of these were uncontrolled reports,79 whereas others used historical controls.1012 It is often difficult to control for many factors that contribute to graft failure, including operator experience, vein quality, distal anastomotic technique, diabetes mellitus, and other patient characteristics. Consequently, we conducted a retrospective study using matched contemporaneous controls to assess the long-term outcome of patients receiving the St. Jude Medical symmetry aortic connector system for proximal vein graft anastomoses in coronary artery bypass grafting.
 |
Methods
|
|---|
A retrospective matched case control analysis was performed
using the Duke Cardiovascular Disease Databank. The null hypothesis
was that no difference in graft patency existed between patients
who received the Symmetry aortic connector system and patients
who had proximal vein graft anastomoses created by traditional
suturing methods. The study protocol was approved by the Duke
Institutional Review Board before initiation of the study. The
authors had full access to the data and take full responsibility
for their integrity. All authors have read and agree to the
manuscript as written.
Patient Population
The organization of the Duke computerized cardiovascular database and follow-up methods have been previously described in detail.13 Patients were included in the case group if they had undergone CABG surgery with placement of at least one St. Jude Medical symmetry aortic connector at Duke Medical Center between the dates of January 1, 2002 and December 31, 2003. Control patients who had undergone CABG by traditional suture methods (saphenous vein grafts with or without internal mammary arteries) at Duke Medical Center between the dates of January 1, 2002 and December 31, 2003 were randomly identified from the Cardiovascular Databank and matched according to age, gender, on-/off-pump and the Hannan model of perioperative risk (Table 1).14 Patients were excluded if they received only an internal mammary graft, or if they underwent valve replacement at the time of CABG.
Statistical Analysis
Baseline characteristics were compiled for the 2 patient groups and compared using the Wilcoxon rank-sum test for continuous variables and the Pearson
2 test for discrete variables. Time to the composite end point of major adverse cardiac events (death, nonfatal myocardial infarction, repeat cardiac catheterization, repeat CABG) was generated using Kaplan-Meier methodology and compared using the log-rank test. Secondary end points included the composite of death and nonfatal myocardial infarction, stroke, and all-cause mortality. Unadjusted Cox proportional hazards regression models estimated the unadjusted effect of each baseline characteristic on the primary end point. A multivariable Cox proportional hazards regression model estimated the hazard ratios for the Symmetry aortic connector adjusted for baseline patient characteristics. The regression model parameter estimates were validated internally using bootstrap estimation of the entire patient sample. Analyses were performed using SAS statistical software (SAS Institute, Cary, NC).
 |
Results
|
|---|
In the Duke Cardiovascular Database, 121 patients received at
least 1 proximal Symmetry aortic connector during the study
period. There were 178 matched contemporaneous control patients
who underwent CABG with vein grafts attached via traditional
hand suturing techniques. There was no difference between the
proportion of cases and controls with regards to diabetes, smoking,
number of diseased coronary arteries, cerebrovascular disease,
or renal failure (
Table 2). Overall, 66% of the population was
male, and 31% had diabetes. Eighty-six percent of the CABG procedures
were performed off-pump.
Figure 1 displays the percent number of bypass grafts used for the cases and controls. In the control group, there was a higher percentage of patients who received 2 or 3 bypass grafts when compared with the case group. More patients in the case group received 4 bypass grafts. These differences achieved statistical significance by Wilcoxon 2-sample test (P=0.0002). In the case group, >80% of patients received 1 or 2 Symmetry devices during coronary artery bypass surgery (Figure 2). Additionally, 85.12% of case patients received an internal mammary artery graft during bypass surgery, compared with 91.57% of control patients (P=0.08).
At 30 days, 5 case patients and 3 control patients underwent repeat cardiac catheterization (P=0.198). Two case patients had repeat CABG at 30 days, but no control patients had repeat CABG in this time period (P=0.085).
Within 1 year of initial CABG, 7 patients in each group died (5.8% cases versus 3.9% controls), and 11 patients in each group had death or myocardial infarction (9.1% cases versus 6.2% controls). At 1 year, 26 patients (21.5%) in the Symmetry connector group had sustained death, myocardial infarction, repeat catheterization, or repeat CABG, compared with 26 controls (13.5%).
There was a trend toward an increased event rate in the cases when compared with the controls for the composite end point of death, nonfatal myocardial infarction, repeat cardiac catheterization, or repeat CABG (P=0.053), as seen in the Kaplan Meier estimates presented in Figure 3. No significant differences were seen for all-cause mortality (P=0.44), stroke (P=0.19), or the combined end point of death or nonfatal myocardial infarction (P=0.33).

View larger version (10K):
[in this window]
[in a new window]
|
Figure 3. Unadjusted KaplanMeier estimates for the composite end point (death, nonfatal myocardial infarction, repeat cardiac catheterization, repeat CABG).
|
|
Unadjusted hazard ratios for the Symmetry aortic connector were calculated for multiple end points and are shown in Table 3. The Symmetry aortic connector was not a significant independent predictor of any of the end points. Additionally, a multivariable Cox proportional hazards model for the composite end point of death, nonfatal myocardial infarction, repeat cardiac catheterization, or repeat CABG was developed to assess the impact of the Symmetry aortic connector in an adjusted setting. Significant predictors of the composite end point included hyperlipidemia and peripheral vascular disease. When the Symmetry aortic connector variable was added to the model, it was not significant (P=0.19).
To assess the reliability of estimates from the Duke patient series, and to assess the possible influence of outliers on the model results, a bootstrap internal validation was performed. As expected, use of the St. Jude connector did not meet criteria (alpha=0.05) for entry into bootstrap-resampled models for death or the composite end point of death or myocardial infarction. For the composite end point of death, myocardial infarction, repeat CABG, or repeat catheterization, use of the St. Jude connector met entry criteria in 66 of 200 bootstrapped samples, compared with 159 for hyperlipidemia, 115 for peripheral vascular disease, 84 for female gender, and 74 for diabetes.
 |
Discussion
|
|---|
The analysis of the results suggest that use of the Symmetry
aortic connector system is associated with increased major adverse
cardiac events at 1 year, suggestive of early graft closure.
Although the determinants of postoperative vein graft patency
are variable (including conduit quality, distal vessel run-off,
progression of native vessel disease, and surgical technique),
these results suggest that graft closure events occur with the
Symmetry device more frequently than might be expected otherwise.
The precise contribution of the device to early graft failure
is unknown, but suspected to be intimal hyperplasia. These findings
encourage close clinical follow-up of patients who have received
this device. This may include more aggressive noninvasive testing
postoperatively.
Despite initial promising reports,1517 several other case series have also reported early vein graft closures with the Symmetry device.712 Cavendish et al reported a series of 121 CABG patients with Symmetry implants, of which 5 presented with an acute coronary syndrome 2 to 5 months after operation.7 In each patient all saphenous vein grafts were totally occluded or compromised with ostial stenoses. After percutaneous intervention of the occluded saphenous vein grafts, 2 patients presented again with near-occlusive ostial restenosis. Intravascular ultrasound showed severe neointimal hyperplasia only at the interface between the stent and the connector device.
Carrel et al reviewed 107 patients who had received at least 1 Symmetry connector device.8 At 6 months, 14 patients had routine angiography. Of the 24 vein grafts studied, 11 were anastomosed with a running suture and 13 with the proximal connector system. Five of the 13 patients had a 30% to 90% stenosis within the connector, and there was one complete occlusion. One hand-sewn anastomosis had occluded, but the remainder did not show any significant stenosis.
Reuthebuch et al reported a series of 61 patients who received a total of 77 Symmetry connectors.9 Eight patients with 12 implanted connectors developed ischemic symptoms within 8 months. Angiography revealed significant stenosis or occlusion within all connectors in these patients.
Schoettle reported a retrospective review of two patient groups undergoing CABG.12 In group A, 323 patients underwent surgery from October 2000 through September 2001, with hand-sewn proximal anastomoses. In group B, 305 patients underwent surgery from October 2001 through September 2002, with proximal anastomoses constructed with the Symmetry device. Both groups received aspirin postoperatively and received clopidogrel daily for 2 months. Forty patients in group A underwent repeat cardiac catheterization, and of 71 venous grafts studied, 61 were widely patent. Forty-five patients in group B underwent repeat catheterization, and of 88 venous grafts studied, 39 were totally occluded. Seventeen had
90% stenosis, and 15 had
50% stenosis. These stenoses uniformly occurred at the connector site. Among the patients studied, this represented an 80% rate of occlusion or stenosis with the proximal venous graft connectors.
Dewey et al reported a series of 166 patients who underwent off-pump bypass grafting using at least 1 St. Jude symmetry aortic connector between May 2001 and December 2001.11 A control group of 159 patients was identified from a cohort of patients having beating heart surgery with 1 or more sutured proximal vein graft anastomosis in the preceding year. Patients with connectors showed an accelerated number of major adverse events beginning &180 days from the time of surgery and stabilizing at &300 days. Logistic regression analysis identified the presence of diabetes as a significant preoperative risk factor predisposing patients to earlier onset of adverse events (P=0.03).
Nishizaki et al reviewed 17 patients who received the aortic Symmetry connector with 19 patients who received saphenous vein grafts by traditional suture techniques.10 They observed that 4 of a total of 30 aortic symmetry connectors were occluded because of proximal vein graft kinking. They concluded that when using the aortic Symmetry connector, the proximal vein take-off is an important consideration to prevent early vein graft occlusion.
Although the St. Jude Symmetry aortic connector device was voluntarily withdrawn from the market in December 2004, these results have important implication for future aorto-saphenous vein graft connectors and their development. Before marketing a new medical device, manufacturers must provide the Food and Drug Administration (FDA) with "reasonable assurance" that the device is both safe and effective. There are 2 mechanisms for providing this assurance: the premarket notification [510(k)] process, and the premarket approval (PMA) process.
The FDAs PMA process requires examination of detailed safety and effectiveness information for each device. The average PMA submission requires 1200 hours of rigorous review. The Medical Device Amendments to the Federal Food, Drug, and Cosmetic Act (21 US C.
360c et seq.) allows manufacturers to seek approval for a new device by showing that the new device is substantially equivalent to a device introduced to the market before May 28, 1976, when the amendments took effect. This 510(k) mechanism merely establishes whether a pre-1976 device and a post-1976 device are equivalent, regardless of how unsafe or ineffective the grandfathered device happens to be. The FDA completes the average 510(k) review within 20 hours. Not surprisingly, the PMA process represents a much more significant financial barrier to the market ($111 000 to $828 000 per device) than the 510(k) process ($50 to $2000 per device).18
The Symmetry device was approved through a 510(k) mechanism, intended to demonstrate similarity with a predicate device, the US Surgical AutoSuture VCS. In retrospect, the devices fundamentally differ in that the Symmetry connector is an intravascular device, whereas the AutoSuture VCS is an extravascular device. In its review of 510(k) premarket notifications for aortic connector devices, the FDA required limited clinical studies, supported by extensive preclinical data. The clinical studies were designed to substantiate equivalence to historical data for conduit patency, using suture anastomosis as the gold standard control. The FDA accepted a relatively short follow-up period (6 to 9 months) to focus on the immediate adequacy of the anastomosis, rather than other factors contributing to graft failure. A 2004 meeting of the Circulatory System Devices Panel evaluated design parameters for future trials of anastomotic devices. The legacy of the Symmetry device will likely be more rigorous clinical evaluation of future connectors, with a continuing emphasis on long-term graft patency. Although more rigorous trial designs will increase research costs and lengthen development timelines, they will ultimately yield better clinical outcomes for patients receiving approved anastomotic devices.
 |
Acknowledgments
|
|---|
We express our gratitude to the Duke Heart center for their
generous funding and support for this research.
Sources of Funding
R.D. Davis received research funding from St. Jude Medical.
Disclosures
None.
 |
Footnotes
|
|---|
Presented at the American Heart Association Scientific Sessions,
Dallas, Tex, November 1316, 2005.
 |
References
|
|---|
- Dacey LJ, Likosky DS, Leavitt BJ, Lahey SJ, Quinn RD, Hernandez F Jr, Quinton HB, Desimone JP, Ross CS, OConnor GT. Perioperative stroke and long-term survival after coronary bypass graft surgery. The Ann Thorac Surg. 2005; 79: 532536.
- Roach GW, Kanchuger M, Mangano CM, Newman M, Nussmeier N, Wolman R, Aggarwal A, Marschall K, Graham SH, Ley C. Adverse cerebral outcomes after coronary bypass surgery. N Engl J Med. 1996; 335: 18571863.[Abstract/Free Full Text]
- Tuman KJ, McCarthy RJ, Najafi H, Ivankovich AD. Differential effects of advanced age on neurologic and cardiac risks of coronary artery operations. J Thorac Cardiovasc Surg. 1992; 104: 15101517.[Abstract]
- Stamou SC, Jablonski KA, Hill PC, Bafi AS, Boyce SW, Corso PJ. Coronary revascularization without cardiopulmonary bypass versus the conventional approach in high-risk patients. The Ann Thorac Surg. 2005; 79: 552557.
- Zangrillo A, Crescenzi G, Landoni G, Leoni A, Marino G, Calabro MG, Corno C, Pappalardo F, Alfieri O. Off-pump coronary artery bypass grafting reduces postoperative neurologic complications. J Cardiothorac Vasc Anesth. 2005; 19: 193196.[Medline]
[Order article via Infotrieve]
- Swanson WJ, Wahlberg MD, Galdonik JA, Berg TA, Thome SP. St. Jude Medical Cardiovascular Group I, editor. Medical anastomosis apparatus. Minneapolis, MN patent 6,113,612. September 2000.
- Cavendish JJ, Penny WF, Madani MM, Keramati S, Ben-Yehuda O, Blanchard DG, Mahmud E, Perricone A, Tsimikas S. Severe ostial saphenous vein graft disease leading to acute coronary syndromes following proximal aorto-saphenous anastomoses with the symmetry bypass connector device: is it a suture device or a "stent"? J Am Coll Cardiol. 2004; 43: 133139.[Abstract/Free Full Text]
- Carrel TP, Eckstein FS, Englberger L, Windecker S, Meier B. Pitfalls and key lessons with the symmetry proximal anastomotic device in coronary artery bypass surgery. Ann Thorac Surg. 2003; 75: 14341436.[Abstract/Free Full Text]
- Reuthebuch O, Lachat M, Kadner, Turina M. Early bypass occlusion in patients with the St Jude Medical Symmetry connector. J Thoracic Cardiovasc Surg. 2003; 125: 443444.[Free Full Text]
- Nishazaki K, Seki T. Early results of aortic connector system for proximal anastmoses of saphenous vein graft. Ann Thorac Cardiovasc Surg. 2005; 11: 98103.[Medline]
[Order article via Infotrieve]
- Dewey TM, Crumrine K, Herbert MA, Leonard A, Prince SL, Worley C, Edgerton JR, Magee MJ, Mack MJ. First-year outcomes of beating heart coronary artery bypass grafting using proximal mechanical connectors. The Ann Thorac Surg. 2004; 77: 15421549.
- Schoettle GP. Use of a proximal anastomotic device in coronary artery bypass surgery: a word of caution. J Thorac Cardiovasc Surg. 2004; 127: 269270.[Free Full Text]
- Rosati RA, McNeer JF, Starmer CF, Mittler BS, Morris JJ, Wallace AG. A new information system for medical practice. Arch Intern Med. 1975; 135: 10171024.[Abstract]
- Hannan EL, Kilburn, H Jr, Racz M, Shields E, Chassin MR. Improving the outcomes of coronary artery bypass surgery in New York State. J Am Med Assoc. 1994; 271: 761766.[Abstract]
- Eckstein FS, Bonilla LF, Englberger L, Immer FF, Berg TA, Schmidli J, Carrel TP. The St Jude Medical Symmetry aortic connector system for proximal vein graft anastomoses in coronary artery bypass grafting. J Thoracic Cardiovasc Surg. 2002; 123: 777782.[Abstract/Free Full Text]
- Endo M, Benhameid O, Morin J, Shennib H. Avoiding aortic clamping during coronary artery bypass using an automated anastomotic device. Ann Thorac Surg. 2002; 73: 10001001.[Abstract/Free Full Text]
- Calafiore AM, Bar-El Y, Vitolla G, Di Giammarco G, Teodori G, Iaco AL, DAlessandro S, DiMauro M. Early clinical experience with a new sutureless anastomotic device for proximal anastomosis of the saphenous vein to the aorta. J Thoracic Cardiovasc Surg. 2001; 121: 854858.[Abstract/Free Full Text]
- Goodlin v Medtronic, Inc. [No. 97-5801]. February 18, 1999. 11th Circulation.